EXTECH MN16

User's Guide
Autoranging Mini Multimeter
Model MN16
Introduction
Congratulations on your purchase of the Extech MN16 Autoranging Multimeter. This meter
measures AC/DC Voltage, AC/DC Current, Resistance, Capacitance, Frequency, Diode Test,
and Continuity plus Thermocouple Temperature. It features a rugged design for heavy duty
use This meter is shipped fully tested and calibrated and, with proper use, will provide years
of reliable service.
Safety
This symbol adjacent to another symbol, terminal or operating device
indicates that the operator must refer to an explanation in the Operating
Instructions to avoid personal injury or damage to the meter.
WARNING
This WARNING symbol indicates a potentially hazardous situation, which
if not avoided, could result in death or serious injury.
CAUTION
This CAUTION symbol indicates a potentially hazardous situation, which
if not avoided, may result damage to the product.
MAX
600V
This symbol advises the user that the terminal(s) so marked must not be
connected to a circuit point at which the voltage with respect to earth
ground exceeds (in this case) 600 VAC or VDC.
This symbol adjacent to one or more terminals identifies them as being
associated with ranges that may, in normal use, be subjected to
particularly hazardous voltages. For maximum safety, the meter and its
test leads should not be handled when these terminals are energized.
This symbol indicates that a device is protected throughout by double
insulation or reinforced insulation.
PER IEC1010 OVERVOLTAGE INSTALLATION CATEGORY
OVERVOLTAGE CATEGORY I
Equipment of OVERVOLTAGE CATEGORY I is equipment for connection to circuits in
which measures are taken to limit the transient overvoltages to an appropriate low level.
Note – Examples include protected electronic circuits.
OVERVOLTAGE CATEGORY II
Equipment of OVERVOLTAGE CATEGORY II is energy-consuming equipment to be
supplied from the fixed installation.
Note – Examples include household, office, and laboratory appliances.
OVERVOLTAGE CATEGORY III
Equipment of OVERVOLTAGE CATEGORY III is equipment in fixed installations.
Note – Examples include switches in the fixed installation and some equipment for industrial
use with permanent connection to the fixed installation.
OVERVOLTAGE CATEGORY IV
Equipment of OVERVOLTAGE CATEGORY IV is for use at the origin of the installation.
Note – Examples include electricity meters and primary over-current protection equipment
2
MN16 V1.2
7/08
CAUTIONS
•
Improper use of this meter can cause damage, shock, injury or death. Read and
understand this user manual before operating the meter.
•
Always remove the test leads before replacing the battery or fuses.
•
Inspect the condition of the test leads and the meter itself for any damage before
operating the meter. Repair or replace any damage before use.
•
Use great care when making measurements if the voltages are greater than 25VAC
rms or 35VDC. These voltages are considered a shock hazard.
•
Warning! This is a class A equipment. This equipment can cause interferences in the
living quarters; in this case the operator can be required to carry out adequate
measures.
•
Always discharge capacitors and remove power from the device under test before
performing Diode, Resistance or Continuity tests.
•
Voltage checks on electrical outlets can be difficult and misleading because of the
uncertainty of connection to the recessed electrical contacts. Other means should be
used to ensure that the terminals are not "live".
•
If the equipment is used in a manner not specified by the manufacturer, the protection
provided by the equipment may be impaired.
•
This device is not a toy and must not reach children’s hands. It contains hazardous
objects as well as small parts that the children could swallow. In case a child
swallows any of them, please contact a physician immediately
•
Do not leave batteries and packing material lying around unattended; they can be
dangerous for children if they use them as toys
•
In case the device is going to be unused for an extended period of time, remove the
batteries to prevent them from training
•
Expired or damaged batteries can cause cauterization on contact with the skin.
Always, therefore, use suitable hand gloves in such cases
•
See that the batteries are not short-circuited. Do not throw batteries into the fire.
3
MN16 V1.2
7/08
SAFETY INSTRUCTIONS
This meter has been designed for safe use, but must be operated with caution. The rules
listed below must be carefully followed for safe operation.
1.
NEVER apply voltage or current to the meter that exceeds the specified maximum:
Input Protection Limits
Function
Maximum Input
V DC or V AC
600VDC/AC
mA AC/DC
400mA AC/DC
A AC/DC
10A AC/DC (20A for 30 sec max every 15 min)
Frequency, Resistance, Capacitance,
Diode Test, Continuity, Temperature
250VDC/AC
2.
USE EXTREME CAUTION when working with high voltages.
3.
DO NOT measure voltage if the voltage on the "COM" input jack exceeds 600V above
earth ground.
4.
NEVER connect the meter leads across a voltage source while the function switch is in
the current, resistance, or diode mode. Doing so can damage the meter.
5.
ALWAYS discharge filter capacitors in power supplies and disconnect the power when
making resistance or diode tests.
6.
ALWAYS turn off the power and disconnect the test leads before opening the covers to
replace the fuse or batteries.
7.
NEVER operate the meter unless the back cover and the battery and fuse covers are in
place and fastened securely.
8.
If the equipment is used in a manner not specified by the manufacturer, the protection
provided by the equipment may be impaired.
4
MN16 V1.2
7/08
Controls and Jacks
1. 4,000 count LCD display
2. MODE button
3. HOLD button
4. Function switch
5. COM input jack
6. 10A input jack
7. V, Ω, CAP, Temp, Hz, μA and mA input jack
8. REL button
9. RANGE button
10. Battery compartment (rear)
Symbols and Annunciators
•)))
n
µ
m
A
k
F
M
Ω
Hz
%
AC
DC
ºF
Continuity
Diode test
Low Battery
-9
nano (10 ) (capacitance)
-6
micro (10 ) (amps)
-3
milli (10 ) (volts, amps)
Amps
3
kilo (10 ) (ohms)
Farads (capacitance)
6
mega (10 ) (ohms)
Ohms
Hertz (frequency)
Percent (duty ratio)
Alternating current
Direct current
Degrees Fahrenheit
V
REL
AUTO
HOLD
ºC
Volts
Relative
Autoranging
Display hold
Degrees Centigrade
5
MN16 V1.2
7/08
Operating Instructions
WARNING: Risk of electrocution. High-voltage circuits, both AC and DC, are very
dangerous and should be measured with great care.
1.
ALWAYS turn the function switch to the OFF position when the meter is not in use.
2.
If “OL” appears in the display during a measurement, the value exceeds the range
you have selected. Change to a higher range.
DC VOLTAGE MEASUREMENTS
CAUTION: Do not measure DC voltages if a motor on the circuit is being switched ON or
OFF. Large voltage surges may occur that can damage the meter.
1.
2.
3.
4.
Set the function switch to the VDC position.
Insert the black test lead banana plug into the negative COM jack.
Insert the red test lead banana plug into the positive V jack.
Touch the black test probe tip to the negative side of the circuit.
Touch the red test probe tip to the positive side of the circuit.
Read the voltage in the display.
AC VOLTAGE (FREQUENCY, DUTY CYCLE) MEASUREMENTS
WARNING: Risk of Electrocution. The probe tips may not be long enough to contact the
live parts inside some 240V outlets for appliances because the contacts are recessed
deep in the outlets. As a result, the reading may show 0 volts when the outlet actually has
voltage on it. Make sure the probe tips are touching the metal contacts inside the outlet
before assuming that no voltage is present.
CAUTION: Do not measure AC voltages if a motor on the circuit is being switched ON or
OFF. Large voltage surges may occur that can damage the meter.
1.
2.
3.
4.
5.
6.
7.
8.
Set the function switch to the VAC position.
Insert the black test lead banana plug into the negative COM jack.
Insert red test lead banana plug into the positive V jack.
Touch the black test probe tip to the neutral side of the circuit.
Touch the red test probe tip to the “hot” side of the circuit.
Read the voltage in the display.
Press the MODE button to indicate “Hz”.
Read the frequency in the display.
Press the MODE button again to indicate “%”.
Read the % of duty cycle in the display.
6
MN16 V1.2
7/08
DC CURRENT MEASUREMENTS
CAUTION: Do not make 20A current measurements for longer than 30 seconds.
Exceeding 30 seconds may cause damage to the meter and/or the test leads.
1.
2.
3.
4.
5.
6.
7.
8.
9.
Insert the black test lead banana plug into the negative COM jack.
For current measurements up to 4000µA DC;
Set the function switch to the µA position and insert the red test lead
banana plug into the µA jack.
For current measurements up to 400mA DC;
Set the function switch to the mA position and insert the red test lead
banana plug into the mA jack.
For current measurements up to 10A DC;
Set the function switch to the 10A position and insert the red test lead
banana plug into the 10A jack.
Press the MODE button to indicate “DC” on the display.
Remove power from the circuit under test, then open the circuit at the
point where you wish to measure current.
Touch the black test probe tip to the negative side of the circuit.
Touch the red test probe tip to the positive side of the circuit.
Apply power to the circuit.
Read the current in the display.
AC CURRENT MEASUREMENTS
CAUTION: Do not make 20A current measurements for longer than 30 seconds.
Exceeding 30 seconds may cause damage to the meter and/or the test leads.
1. Insert the black test lead banana plug into the negative COM jack.
2. For current measurements up to 4000µA AC;
Set the function switch to the µA position and insert the red test lead
banana plug into the µA jack.
3. For current measurements up to 400mA AC;
Set the function switch to the mA position and insert the red test lead
banana plug into the mA jack.
4. For current measurements up to 10A AC;
Set the function switch to the 10A position and insert the red test lead
banana plug into the 10A jack.
5. Press the MODE button to indicate “AC” on the display.
6. Remove power from the circuit under test, then open the circuit at the
point where you wish to measure current.
7. Touch the black test probe tip to the negative side of the circuit.
Touch the red test probe tip to the positive side of the circuit.
8. Apply power to the circuit.
9. Read the current in the display.
7
MN16 V1.2
7/08
RESISTANCE MEASUREMENTS
WARNING: To avoid electric shock, disconnect power to the unit under test and
discharge all capacitors before taking any resistance measurements. Remove the
batteries and unplug the line cords.
1.
2.
3.
4.
5.
Set the function switch to the Ω position.
Insert the black test lead banana plug into the negative COM jack.
Insert the red test lead banana plug into the positive Ω jack.
Press the MODE button to indicate “Ω" on the display.
Touch the test probe tips across the circuit or part under test.
Read the resistance in the display.
CONTINUITY CHECK
WARNING: To avoid electric shock, never measure continuity on circuits or wires that
have voltage on them.
1.
2.
3.
4.
5.
Set the function switch to the
position.
Insert the black lead banana plug into the negative COM jack.
Insert the red test lead banana plug into the positive Ω jack.
Press the MODE button to indicate”
"and “Ω” on the display
Touch the test probe tips to the circuit or wire you wish to check.
If the resistance is less than approximately 150Ω, the audible signal
will sound. If the circuit is open, the display will indicate “OL”.
DIODE TEST
1.
2.
Set the function switch to the
position.
Insert the black test lead banana plug into the negative COM jack and
the red test lead banana plug into the positive V jack.
3.
4.
Press the MODE button to indicate
and V on the display.
Touch the test probes to the diode under test. Forward voltage will
typically indicate 0.400 to 0.700V. Reverse voltage will indicate “OL”.
Shorted devices will indicate near 0V and an open device will indicate
“OL” in both polarities.
8
MN16 V1.2
7/08
TEMPERATURE MEASUREMENTS
1.
2.
3.
4.
Set the function switch to the ºF or ºC position.
Insert the Temperature Probe into the input jacks, making sure to
observe the correct polarity.
Touch the Temperature Probe head to the part whose temperature you
wish to measure. Keep the probe touching the part under test until the
reading stabilizes (about 30 seconds).
Read the temperature in the display.
Note: The temperature probe is fitted with a type K mini connector. A mini
connector to banana connector adaptor is supplied for connection to the
input banana jacks.
CAPACITANCE MEASUREMENTS
WARNING: To avoid electric shock, disconnect power to the unit under test and
discharge all capacitors before taking any capacitance measurements. Remove the
batteries and unplug the line cords.
1.
2.
3.
4.
5.
Set the rotary function switch to the CAP position.
Insert the black test lead banana plug into the negative COM jack.
Insert the red test lead banana plug into the positive CAP jack.
Press the MODE button to indicate “nF” on the display.
Touch the test leads to the capacitor to be tested.
Read the capacitance value in the display
FREQUENCY/DUTY CYCLE MEASUREMENTS
1. Set the rotary function switch to the “Hz %” position.
2. Press the MODE button to indicate “Hz” in the display.
3. Insert the black lead banana plug into the negative COM jack
Insert the red test lead banana plug into the positive Hz jack.
4. Touch the test probe tips to the circuit under test.
5. Read the frequency on the display.
6. Press the MODE button again to indicate “%” on the display.
7. Read the % of duty cycle on the display.
9
MN16 V1.2
7/08
AUTORANGING/MANUAL RANGE SELECTION
When the meter is first turned on, it automatically goes into Autoranging. This automatically
selects the best range for the measurements being made and is generally the best mode
for most measurements. For measurement situations requiring that a range be manually
selected, perform the following:
1.
2.
Press the RANGE button. The “Auto” display indicator will turn off.
Press the RANGE key to step through the available ranges. Observe the decimal
point and units displayed until the preferred range is located.
3. To exit the Manual Ranging mode and return to Autoranging, press and hold the
RANGE key for 2 seconds.
Note: Manual ranging does not apply for the Capacitance and Frequency functions.
RELATIVE MODE
The relative measurement feature allows you to make measurements relative to a stored
zero reference value. A reference voltage, current, etc. can be stored and measurements
made in comparison to that value. The displayed value is the difference between the
reference value and the measured value.
1.
2.
Perform the measurement as described in the operating instructions.
Press the REL button to store (zero) the reading in the display. The "REL" indicator
will appear on the display.
3. The display will now indicate the difference between the stored value and the
measured value.
4. Press the REL button to exit the relative mode.
Note: The Relative function does not operate in the Frequency function.
HOLD
The hold function freezes the reading in the display. Press the HOLD key momentarily to
activate or to exit the HOLD function.
AUTO POWER OFF
The auto off feature will turn the meter off after 30 minutes.
LOW BATTERY INDICATION
icon will appear in the display when the battery voltage becomes low. Replace the
The
battery when this appears.
10
MN16 V1.2
7/08
Maintenance
WARNING: To avoid electric shock, disconnect the test leads from any source of voltage
before removing the back cover or the battery or fuse covers.
WARNING: To avoid electric shock, do not operate your meter until the battery and fuse
covers are in place and fastened securely.
This MultiMeter is designed to provide years of dependable service, if the following care
instructions are performed:
1. KEEP THE METER DRY. If it gets wet, wipe it off.
2. USE AND STORE THE METER IN NORMAL TEMPERATURES. Temperature
extremes can shorten the life of the electronic parts and distort or melt plastic parts.
3. HANDLE THE METER GENTLY AND CAREFULLY. Dropping it can damage the
electronic parts or the case.
4. KEEP THE METER CLEAN. Wipe the case occasionally with a damp cloth. DO NOT
use chemicals, cleaning solvents, or detergents.
5. USE ONLY FRESH BATTERIES OF THE RECOMMENDED SIZE AND TYPE. Remove
old or weak batteries so they do not leak and damage the unit.
6. IF THE METER IS TO BE STORED FOR A LONG PERIOD OF TIME, the batteries
should be removed to prevent damage to the unit.
BATTERY REPLACEMENT
1.
Disconnect the test leads from the meter.
2.
Remove the 2 Phillips head screws located on the back of the instrument and remove
the battery cover.
3.
Replace the 2 AAA batteries.
4.
Secure the fuse/battery compartment cover.
5.
Dispose of the old battery as required by local regulations.
FUSE REPLACEMENT
1.
Disconnect the test leads from the meter.
2.
Remove the 2 Phillips head screws located on the back of the instrument and remove
the battery cover.
3.
Gently remove the fuse(s) and install the new fuse(s) into the holder(s).
4.
Always use fuses of the proper size and value (500mA/660V ceramic fast blow for the
mA / µA ranges, 10A/250Vceramic fast blow for the A range).
5.
Secure the fuse/battery compartment cover.
WARNING: To avoid electric shock, do not operate your meter until the fuse cover is in
place and fastened securely.
11
MN16 V1.2
7/08
Specifications
Function
Range
Resolution
DC Voltage
4V
0.001V
(V DC)
40V
0.01V
400V
0.1V
600V
1V
±(1.5% reading + 2 digits)
400mV*
0.1mV
±(1.5% reading + 15 digits)
4V
0.001mV
±(1.2% reading + 3 digits)
40V
0.01V
400V
0.1V
600V
1V
AC Voltage
(V AC)
(50 / 60Hz)
Accuracy
±(1.2% reading + 2 digits)
±(1.5% reading + 3 digits)
±(2.0% reading + 4 digits
*400mV range is not autoranging
DC Current
(A DC)
AC Current
(A AC)
Resistance
Capacitance
Duty Cycle
400μA
±(1.0% reading + 3 digits)
0.1μA
4000μA
1μA
40mA
0.01mA
400mA
0.1mA
10A
0.01A
±(2.5% reading + 5 digits)
400μA
0.1μA
±(1.5% reading + 5 digits)
4000μA
1μA
40mA
0.01mA
400mA
0.1mA
10A
0.01A
±(3.0% reading + 7 digits)
400Ω
0.1Ω
±(1.2% reading + 4 digits)
4kΩ
1Ω
±(1.0% reading + 2 digits)
40kΩ
0.01kΩ
400kΩ
0.1kΩ
±(1.5% reading + 3 digits)
±(1.8% reading + 5 digits)
±(1.2% reading + 2 digits)
4MΩ
0.001MΩ
40MΩ
0.01MΩ
±(2.0% reading + 3 digits)
40nF
0.01nF
±(5.0% reading + 30 digits)
400nF
0.1nF
4μF
0.001μF
40μF
0.01μF
100μF
0.1μF
0.1%
0.1 to 99.9%
±(3.0% reading + 5 digits)
±(5.0% reading + 5 digits)
±(1.2% reading + 2 digits)
pulse width range: 100µs - 100ms (Frequency: 5Hz to 150kHz)
Temp (type-K)
-4 to 1400°F
1°F
-20 to 750°C
1°C
±(3.0% reading + 9°F/5°C)
(probe accuracy not included)
12
MN16 V1.2
7/08
Function
Range
Resolution
Frequency
5.000Hz
50.00Hz
500.0Hz
5.000kHz
50.00kHz
500.0kHz
5.000MHz
0.001Hz
0.01Hz
0.1Hz
0.001kHz
0.01kHz
0.1kHz
0.001MHz
Accuracy
±(1.5% reading + 5 digits)
±(1.2% reading + 3 digits)
±(1.5% reading + 4 digits)
Sensitivity: >0.5V rms min. [ 1MHz; >3V rms > 1MHz
NOTE:
o
o
o
o
Accuracy is stated at 65 F to 83 F (18 C to 28 C) and less than 75% RH.
Diode Test
Continuity Check
Temperature Sensor
Input Impedance
AC Response
ACV Bandwidth
Display
Overrange indication
Auto Power Off
Polarity
Measurement Rate
Test current of 0.3mA typical, open circuit voltage 1.5V DC typical
Audible signal will sound if the resistance is less than 150Ω
(approx.), test current <0.3mA
Requires type K thermocouple
7.8MΩ
Average responding
50Hz to 400Hz
4,000 count liquid crystal
“OL” is displayed
30 minutes (approximately)
Automatic (no indication for positive); Minus (-) sign for negative
2 times per second, nominal
Low Battery Indication
“
” is displayed if battery voltage drops below operating
voltage
Battery
Two AAA batteries
Fuses
mA, µA ranges; 0.5A/660V ceramic fast blow
A range; 10A/250V ceramic fast blow
Operating Temperature 41°F to 104°F (5°C to 40°C)
Storage Temperature
-4°F to 140°F (-20°C to 60°C)
Operating Humidity
Max 80% up to 87°F (31°C) decreasing linearly to 50% at 104°F
(40°C)
Storage Humidity
<80%
Operating Altitude
7000ft. (2000meters) maximum.
Weight
0.5lb (225g) (includes holster).
Size
5.31” x 2.67” x 1.7” (135 x 68 x 43mm)
Safety
This meter is intended for indoor use and protected, against the
users, by double insulation per EN61010-1 and IEC61010-1 2nd
Edition (2001) to CAT II 1000V & CAT III 600V; Pollution Degree 2.
The meter also meets UL 61010-1, Second Edition (2004),
CAN/CSA C22.2 No. 61010-1, Second Edition (2004), and UL
61010B-2-031, First Edition (2003)
Approvals
UL CE
UL Listed
The UL mark does not indicate that this product has been
evaluated for the accuracy of its readings.
13
MN16 V1.2
7/08
Warranty
EXTECH INSTRUMENTS CORPORATION warrants the basic instrument to be free of
defects in parts and workmanship for three years from date of shipment (a six month limited
warranty applies on sensors and cables). If it should become necessary to return the
instrument for service during or beyond the warranty period, contact the Customer Service
Department at (781) 890-7440 EXTENSION 210 for authorization or visit www.extech.com for
more information. A Return Authorization (RA) number must be issued before any
product is returned to Extech. The sender is responsible for shipping charges, freight,
insurance and proper packaging to prevent damage in transit. This warranty does not apply to
defects resulting from action of the user such as misuse, improper wiring, operation outside of
specification, improper maintenance or repair, or unauthorized modification. Extech
specifically disclaims any implied warranties or merchantability or fitness for a specific
purpose and will not be liable for any direct, indirect, incidental or consequential damages.
Extech's total liability is limited to repair or replacement of the product. The warranty set forth
above is inclusive and no other warranty, whether written or oral, is expressed or implied.
Calibration and Repair Services
Extech offers repair and calibration services for the products we sell. Extech also provides
NIST certification for most products. Call the Customer Care Department for information on
calibration services available for this product. Extech recommends that annual calibrations be
performed to verify meter performance and accuracy.
Support line (781) 890-7440
Technical Support: Extension 200; E-mail: [email protected]
Repair & Returns: Extension 210; E-mail: [email protected]
Product specifications subject to change without notice
For the latest version of this User’s Guide, Software updates, and other up-to-the-minute
product information, visit our website: www.extech.com
Extech Instruments Corporation, 285 Bear Hill Rd., Waltham, MA 02451
Copyright © 2008 Extech Instruments Corporation (a FLIR company)
All rights reserved including the right of reproduction in whole or in part in any form.
14
MN16 V1.2
7/08