ASTRODYNE MPT-120D

1~4�Output�Medical Type
1 2 0 W 1 ~ 4 O u t p u t for Medical Type
0 W��M
m1p2 s
/ d / t e/ qd -i 1c 2a 0l sseerri ieess
Features
Universal AC�input�/�Full�range
Low�leakage�current�<180 A
Protections:Short�circuit/Over�load/Over�voltage
UL2601-1�medical�safety approved
With�power�good�and�fail�signal�output�(Optional)
100%�full�load�burn-in�test
Fixed�switching�frequency�at�45KHz
3�years�warranty
Specification
AC�INPUT�VOLTAGE
90~264�VAC,�47~440Hz�/�127~370VDC.
TEMP.�COEFFICIENT
0.04%/ on�all�outputs�at�full�load�between�0~50
AC�INPUT�CURRENT (Typ.)
Maximum�input�current�2.9A at�115VAC,�60Hz�or
1.7A at�230VAC,�60Hz�with�100%�output�load.
VIBRATION
2G�of�acceleration,�vibrating�frequency�adjust�from�10Hz�~500Hz�within�a
10-minute�cycle.�6�testing�cycles�(60�minutes)�each�along�X, Y,�Z�axes.
INRUSH�CURRENT (Typ.)
Inrush�current�is�less�than�22A at�115VAC�or�less�than
45A at�230VAC�under�cold�start�conditions.�Limiting
provided�by�internal�thermistors.
SAFETY STANDARDS
Medical : UL2601-1, TUV�EN60601-1�approved
Commercial�: Also�design�refer�to�UL60950-1, TUV�EN60950-1
SETUP,�RISE�TIME
MPS-120:
800ms,�20ms�/�230VAC�at�full�load
2000ms,�50ms�/�115VAC�at�full�load
MPD,T,Q-120:500ms,�20ms�/�230VAC�at�full�load
1200ms,�50ms�/�115VAC�at�full�load
HOLD-UP TIME (Typ.)
80ms�/�230VAC�at�full�load
16ms�/�115VAC�at�full�load
DC�OUTPUT ADJ.�RANGE
DC�output�voltage�(or�Ch1�of�multiple�output�models)
can�be�adjusted�between�-5%~+10%�rated�output
voltage�by�potential�meter.
OVER�LOAD�PROTECTION
Fully�protected�against�short�circuit�and�output�overload.
The�hiccup�type�protection will�be�activated�at�110~150%
(For�MPD,T,Q-120),120~160%(For�MPS-120)�rated�load
and�recovers�automatically�after�fault�condition�is removed.
OVER�VOLTAGE�PROTECTION
Provided�on�output�channel�1�only�at�115%~135%
rated�output�voltage.�(120%~140%�for�MPS-120-15/24/48).
Output�will�be�shut�down�when�this�protection�is�activated.
POWER�GOOD�/�FAIL SIGNAL (OPTIONAL)
TTL logic�high�for�power�good�and TTL low�for
power�fail.�When�the�output�voltage�reaches�90%�of
rated�value,�a�+5V TTL signal�will�be�sent�out�with�a
10~500ms�delay; At�least�1ms�before�the�output
voltage�goes�below�90%�of�the�rated�value,�the TTL
signal�will�be�turned�off.
*�MPS-120-3.3�does�not�have�this�optional�function.
WORKING�TEMP.
Whole�series�can�operate�from�-20~70
to�the�derating�curves.
WITHSTAND�VOLTAGE
4000VAC�between�input�and�output
1500VAC�between�input�and�F.G.
500VAC�between�output�and�F.G.
ISOLATION�RESISTANCE
>100M�Ohms�for�I/P-O/P,�I/P-FG,�O/P-FG�by�using�500VDC�test�voltage.
EMI�COMPLIANCE
EMI�Specifications
Conducted�&�Radiation
LEAKAGE�CURRENT
Leakage�current�is�less�than�180 A at�264VAC
.�Please�refer
of�ambient�temperature.
Compliance�Level
EN55011,�Class�B
EN55022,�Class�B
EN61000-3-2
EN61000-3-3
Harmonic�distortion
Voltage�flicker
EMS�COMPLIANCE
EMS�Specification
ESD�air
ESD�contact
RF�field�susceptibility
EFT(Electrical�Fast Transient)/Burst
Lightning/Surge
Conducted�RF�susceptibility
Magnetic�field�immunity
Voltage�dip,�interruption
Digital�phone�carrier�immunity
Compliance�Level
EN61000-4-2, Level�3,�8KV
EN61000-4-2, Level�2,�4KV
EN61000-4-3, Level�2,�3V/m
Level�3,�10V/m
EN61000-4-4, Level�2,�1KV/5KHz
Level�3,�2KV/5KHz
EN61000-4-5, Level�4,�2KV/Line-Line
4KV/Line-Earth
EN61000-4-6, Level�2,�3Vrms/m
Level�3,�10Vrms/m
EN61000-4-8, Level�2,�3A/m
Level�3,�10A/m
EN61000-4-11,Compliance
ENV50204, Level�2,�3V/m,�900MHz
Level�3,�10A/m,�900MHz
MTBF
262,100�hours�min.�at�full�load�and�25
calculated�per�MIL-HDBK-217F.
of�ambient��temperature,
DIMENSION�(L*W*H)
177.8x107.95x35.5mm�or�7"x4.25"x1.4"
PACKING
0.55Kg;�24pcs/14.5Kg/1.04CUFT
WORKING�HUMIDITY
20~90%�RH�non-condensing.
STORAGE�TEMP.,�HUMIDITY
-40~+85 ,�10~90%�RH
WWW.ASTRODYNE.COM
300 Myles Standish Blvd., Taunton, MA 02780
File�Name:120W�MEDICAL-SPEC���2005-09-13
Ph: 1-800-823-8082
Fax: 508-823-8181
m1p2s0 /W��M
d / t /eqd-i1c2a0l
Type
1 2 0 W 1 ~ 4 O u 1~4�Output�Medical
t p u t for Medical Type
sseerriieess
Output�Chart
MODEL
OUTPUT
VOLTAGE
RATED
CURRENT
MPS-120-3.3
3.3V
MPS-120-5
OUTPUT�CURRENT
RIPPLE�&�NOISE VOLTAGE
PEAK�LOAD (Max.)
TOLERANCE
WITH�25CFM�FAN (Note�2)
(Note�3)
LINE
LOAD
REGULATION REGULATION EFFICIENCY
MINIMUM
LOAD
CONVECTION
(max.)
WITH�FAN
(25CFM)
(Note�4)
24A
0A
16A
24A
26A
80mVp-p
3.0%
1.0%
3.0%
68%
5V
22A
0A
14.7A
22A
26A
80mVp-p
3.0%
1.0%
3.0%
73%
MPS-120-12
12V
10A
0A
6.7A
10A
11A
100mVp-p
2.0%
1.0%
2.0%
77%
MPS-120-15
15V
8A
0A
5.3A
8A
8.8A
100mVp-p
2.0%
1.0%
2.0%
79%
MPS-120-24
24V
5A
0A
3.3A
5A
5.5A
120mVp-p
2.0%
1.0%
2.0%
81%
MPS-120-48
48V
2.5A
0A
1.7A
2.5A
2.8A
120mVp-p
2.0%
1.0%
2.0%
82%
MPD-120A
5V
10A
2A
7.3A
10A
12A
80mVp-p
2.0%
0.5%
0.5%
12V
5A
0.5A
3.6A
5A
6A
120mVp-p
7.0%
2.0%
3.5%
5V
10A
2A
7A
10A
12A
80mVp-p
2.0%
0.5%
0.5%
24V
2.9A
0.3A
1.9A
2.9A
3.2A
250mVp-p
8.0%
2.0%
4.0%
5V
10A
2A
7.3A
10A
12A
80mVp-p
2.0%
0.5%
1.0%
12V
4.8A
0.4A
3.5A
4.8A
5.8A
120mVp-p
+8,-6%
1.5%
3.5%
-5V
0.6A
0A
0.3A
0.6A
1A
80mVp-p
5.0%
0.5%
1.0%
5V
10A
2A
7.3A
10A
12A
80mVp-p
2.0%
0.5%
1.0%
12V
4.4A
0.4A
3.2A
4.4A
5.3A
120mVp-p
6.0%
1.5%
3.5%
-12V
0.6A
0A
0.4A
0.6A
1A
80mVp-p
5.0%
0.5%
1.0%
5V
10A
2A
7A
10A
11A
80mVp-p
2.0%
0.5%
1.0%
15V
4A
0.4A
2.6A
4A
4.4A
150mVp-p
+6,-7%
2.0%
3.5%
-15V
0.6A
0A
0.4A
0.6A
1A
80mVp-p
5.0%
0.5%
1.0%
5V
10A
2A
7.3A
10A
12A
80mVp-p
2.0%
0.5%
1.0%
24V
2.2A
0.4A
1.6A
2.2A
2.64A
300mVp-p
+8,-6%
3.0%
+4,-3%
12V
0.6A
0A
0.4A
0.6A
1A
120mVp-p
5.0%
0.5%
1.0%
5V
10A
2A
7.3A
10A
11A
80mVp-p
2.0%
0.5%
0.5%
12V
4.2A
0.5A
3.1A
4.2A
5A
120mVp-p
6.0%
1.5%
3.5%
-5V
0.6A
0A
0.4A
0.6A
1A
80mVp-p
5.0%
0.5%
1.0%
-12V
0.6A
0A
0.4A
0.6A
1A
80mVp-p
5.0%
0.5%
1.0%
5V
10A
2A
7.3A
10A
11A
80mVp-p
2.0%
0.5%
0.5%
15V
3.2A
0.5A
2.4A
3.2A
3.8A
150mVp-p
+6,-7%
2.0%
3.5%
-5V
0.6A
0A
0.4A
0.6A
1A
80mVp-p
5.0%
0.5%
1.0%
-15V
0.6A
0A
0.4A
0.6A
1A
80mVp-p
5.0%
0.5%
1.0%
5V
10A
2A
7A
10A
11A
80mVp-p
2.0%
0.5%
0.5%
12V
1A
0.2A
0.7A
1A
1.1A
150mVp-p
+8,-6%
2.0%
3.5%
24V
2.1A
0.3A
1.4A
2.1A
2.3A
300mVp-p
8.0%
2.0%
3.5%
-12V
0.6A
0A
0.3A
0.6A
1A
80mVp-p
5.0%
0.5%
1.0%
5V
10A
2A
7.3A
10A
11A
80mVp-p
2.0%
0.5%
0.5%
12V
3A
0.5A
2.3A
3A
3.3A
120mVp-p
6.0%
2.0%
3.0%
15V
0.6A
0A
0.4A
0.6A
1A
80mVp-p
8.0%
2.0%
3.0%
24V
0.6A
0A
0.4A
0.6A
1A
80mVp-p
5.0%
0.5%
1.0%
MPD-120B
MPT-120A
MPT-120B
MPT-120C
MPT-120D
MPQ-120B
MPQ-120C
MPQ-120D
MPQ-120E
Notes�:
75%
76%
72%
73%
72%
74%
71%
71%
74%
73%
����������������������������������������������������������������������������������������� �����������������������
������������������������������������������������������������������������������������������������������������������������������������������
������������������������������������������������������������������������������
����������������������������������������������������������������������������������������������������������
����������������������������������������������������������������������������������������������������������������������������������������������������������
���������������
WWW.ASTRODYNE.COM
300 Myles Standish Blvd., Taunton, MA 02780
Ph: 1-800-823-8082
Fax: 508-823-8181
File�Name:120W�MEDICAL-SPEC���2005-09-13
m1p2 s
/ d / t e/ qd -i c1 a2 l0 sseer ri ei ess
0 W��M
1 2 0 W 1 ~ 4 O u t1~4�Output�Medical
p u t for Medical Type
Type
Mechanical�Specification
Unit:mm
177.8[7"]
165.1�[6.5"]
CN1
95.25[3.75"]
13
12
11
10
9
8
7
6
5
4
3
2
1
CN2
107.95[4.25"]
5
4
3
2
1
10cm
+V ADJ
4-
3.96[0.156"]
35.5�[1.4"]
FAN
Pin Assignment
AC�Intput�Connector�(CN1)�:�Molex�5273-05�or�equivalent
Pin�No.
1
2,4
3
5
MPS-120
FG
No�Pin
AC/N
AC/L
MPD-120
FG
No�Pin
AC/N
AC/L
MPT-120
FG
No�Pin
AC/N
AC/L
MPQ-120
FG
No�Pin
AC/N
AC/L
Mating�Housing
Terminal
Molex�5195
or�equivalent
Molex�5194
or�equivalent
MPQ-120
V1
V1
V1
COM
COM
COM
COM
V2
V2
P.F.D.
V3
No�Pin
V4
Mating�Housing
Terminal
Molex�5195
or�equivalent
Molex�5194
or�equivalent
DC�Output�Connector�(CN2)�:�Molex�5273-13�or�equivalent
Pin�No.
1
2
3
4
5
6
7
8
9
10
11
12
13
MPS-120
+V
+V
+V
-V
-V
-V
-V
+V
+V
P.F.D.
NC
No�Pin
NC
MPD-120
V1
V1
V1
COM
COM
COM
COM
V2
V2
P.F.D.
NC
No�Pin
NC
MPT-120
V1
V1
V1
COM
COM
COM
COM
V2
V2
P.F.D.
V3
No�Pin
NC
Derating�Curve
Static�Characteristics
100
100
25CFM�forced�air�cooling
90
80
40
70
LOAD(%)
LOAD�(%)
60
80
MPS/D-120
Convection�cooling
53W�MPS-120-3.3�only
73.7W�MPS-120-5�only
80W�others
MPT/Q-120
20
-20
0
10
20
30
40
AMBIENT�TEMPERATURE�(
WWW.ASTRODYNE.COM
50 55 60
)
70
60
50
40
90
100
115
135
155
230 264
INPUT�VOLTAGE�(V)�60Hz
300 Myles Standish Blvd., Taunton, MA 02780
Ph: 1-800-823-8082
Fax: 508-823-8181
File�Name:120W�MEDICAL-SPEC���2005-09-13