MEANWELL MPD-200

1~4 Output Medical Type
200W Medical
series
Features
Universal AC input / Full range
Low leakage current <180 A
Protections: Short circuit / Overload / Over voltage / Over temperature
Free air convection for 140W and forced air convection for 200W
UL60601-1 medical safety approved
With power good and fail signal output
Built-in remote ON-OFF control
Built-in remote sense function
Fixed switching frequency at 100KHz
3 years warranty
EN60601-1
IEC60601-1
Specification
AC INPUT VOLTAGE
90~264 VAC, 47~440Hz / 127~370VDC.
POWER FACTOR (Typ.)
PF>0.95/230VAC
PF>0.98/115VAC at full load
AC INPUT CURRENT (Typ.)
Maximum input current 3.5A at 115VAC, 60Hz or
1.6A at 230VAC, 60Hz with 100% output load.
INRUSH CURRENT (Typ.)
Inrush current is less than 25A at 115VAC or
less than 40A at 230VAC under cold start
conditions. Limiting provided by internal
thermistors.
SETUP, RISE TIME
1000ms, 20ms / 230VAC at full load
3000ms, 20ms / 115VAC at full load
HOLD-UP TIME (Typ.)
16ms / 230VAC at full load
16ms / 115VAC at full load
LEAKAGE CURRENT
Leakage current is less than 180 A at 264VAC for
earth leakage current
Leakage current is less than 100 A at 264VAC for
patient leakage current
DC OUTPUT ADJ. RANGE
DC output voltage (or CH1 of multiple output models)
can be adjusted between -5%~+10% rated output
voltage by potential meter.
OVERLOAD PROTECTION
Fully protected against short circuit and output overload.
The hiccup type protection will be activated at 120~160%
rated load and recovers automatically after fault condition
is removed.
OVER VOLTAGE PROTECTION
Provided on output channel 1 only at 115%~135%
rated output voltage.
Output will be shut down when this protection is activated.
OVER TEMPERATURE PROTECTION
When the temperature of TSW1 which detect on heat sink of
power transistor reaches 95
This protection is activated.
Then output will be shut down and recovers automatically
after temperature goes down.
POWER GOOD / FAIL SIGNAL
TTL logic high for power good and TTL low for
power fail. When the output voltage reaches 90% of
rated value, a +5V TTL signal will be sent out with a
10~500ms delay; At least 1ms before the output
voltage goes below 90% of the rated value, the TTL
signal will be turned off.
* MPS-200-3.3 does not have this function.
REMOTE CONTROL
RC+/RC-:0 ~ 0.8V=power on; 4 ~ 10V=power off sink current<4~10mA
WORKING TEMP.
Whole series can operate from -20~70 . Please refer
to the derating curves.
WORKING HUMIDITY
20~90% RH non-condensing.
STORAGE TEMP., HUMIDITY
-40~+85 , 10~90% RH
TEMP. COEFFICIENT
0.04%/ on all outputs at full load between 0~50
of ambient temperature.
VIBRATION
2G of acceleration, vibrating frequency adjust from 10Hz ~500Hz within a
10-minute cycle. 6 testing cycles (60 minutes) each along X, Y, Z axes.
SAFETY STANDARDS
Medical : UL60601-1, TUV EN60601-1, IEC60601-1 approved
Commercial : Also design refer to UL60950-1, TUV EN60950-1
WITHSTAND VOLTAGE
4000VAC between input and output
1500VAC between input and F.G.
1500VAC between output and F.G.
ISOLATION RESISTANCE
>100M Ohms for I/P-O/P, I/P-FG, O/P-FG by using 500VDC test voltage.
EMI COMPLIANCE
EMI Specifications
Conducted & Radiation
Harmonic distortion
Voltage flicker
EMS COMPLIANCE
EMS Specification
ESD air
ESD contact
RF field susceptibility
EFT(Electrical Fast Transient)/Burst
Lightning/Surge
Conducted RF susceptibility
Magnetic field immunity
Voltage dip, interruption
Digital phone carrier immunity
MTBF
262,100 hours min. at full load and 25
calculated per MIL-HDBK-217F.
Compliance Level
EN55011, Class B
EN55022, Class B
EN61000-3-2
EN61000-3-3
Compliance Level
EN61000-4-2, Level 3, 8KV
EN61000-4-2, Level 2, 4KV
EN61000-4-3, Level 2, 3V/m
Level 3, 10V/m
EN61000-4-4, Level 2, 1KV/5KHz
Level 3, 2KV/5KHz
EN61000-4-5, Level 4, 2KV/Line-Line
4KV/Line-Earth
EN61000-4-6, Level 2, 3Vrms/m
Level 3, 10Vrms/m
EN61000-4-8, Level 2, 3A/m
Level 3, 10A/m
EN61000-4-11,Compliance
ENV50204, Level 2, 3V/m, 900MHz
Level 3, 10A/m, 900MHz
of ambient temperature,
DIMENSION (L*W*H)
177.8x107.2x35.5mm or 7"x4.22"x1.4"
PACKING
0.66Kg; 24pcs/16.8Kg/1.04CUFT
File Name:200W MEDICAL-SPEC 2008-03-25
200W Medical
1~4 Output Medical Type
series
Output Chart
MODEL
OUTPUT
VOLTAGE
RATED
CURRENT
MPS-200-3.3
3.3V
40A
OUTPUT CURRENT
RIPPLE & NOISE VOLTAGE
PEAK LOAD (Max.)
TOLERANCE
WITH 25CFM FAN (Note 2)
(Note 3)
LINE
LOAD
EFFICIENCY
REGULATION REGULATION (typ.)
MINIMUM
LOAD
CONVECTION
(max.)
WITH FAN
(25CFM)
(Note 4)
0A
28A
40A
48A
80mVp-p
2.0%
0.5%
1.0%
77%
MPS-200-5
5V
40A
0A
28A
40A
48A
80mVp-p
2.0%
0.5%
1.0%
81%
MPS-200-12
12V
16.7A
0A
11.7A
16.7A
20A
100mVp-p
2.0%
0.5%
1.0%
84%
MPS-200-15
15V
13.4A
0A
9.4A
13.4A
16A
100mVp-p
2.0%
0.5%
1.0%
85%
MPS-200-24
24V
8.4A
0A
5.9A
8.4A
10A
150mVp-p
1.0%
0.5%
1.0%
86%
MPS-200-48
48V
4.2A
0A
3A
4.2A
5A
200mVp-p
1.0%
0.5%
1.0%
87%
MPD-200A
5V
20A
4A
15A
20A
24A
80mVp-p
2.0%
0.5%
1.0%
12V
8A
0.8A
5.4A
8A
9.6A
120mVp-p
+8,-5%
1.0%
4.0%
5V
20A
4A
15A
20A
24A
80mVp-p
2.0%
0.5%
1.0%
24V
4A
0.4A
2.7A
4A
4.8A
180mVp-p
6.5%
1.0%
+4,-6%
5V
20A
4A
15A
20A
24A
80mVp-p
2.0%
0.5%
1.0%
12V
7.5A
0.8A
5A
7.5A
9A
120mVp-p
8.0%
1.0%
5.0%
-5V
2A
0A
1A
2A
2.4A
80mVp-p
5.0%
0.5%
1.0%
5V
20A
4A
15A
20A
24A
80mVp-p
2.0%
0.5%
1.0%
12V
6A
0.6A
4.4A
6A
7.2A
120mVp-p
8.0%
1.0%
5.0%
-12V
2A
0A
1A
2A
2.4A
80mVp-p
5.0%
0.5%
1.0%
5V
20A
4A
15A
20A
24A
80mVp-p
2.0%
0.5%
1.0%
15V
4.7A
0.5A
3.3A
4.7A
5.6A
150mVp-p
8.0%
1.0%
5.0%
-15V
2A
0A
1A
2A
2.4A
80mVp-p
5.0%
0.5%
1.0%
5V
20A
4A
15A
20A
24A
80mVp-p
2.0%
0.5%
1.0%
24V
3A
0.3A
2.2A
3A
3.6A
180mVp-p
8.0%
1.0%
5.0%
12V
2A
0A
1A
2A
2.4A
80mVp-p
5.0%
0.5%
1.0%
5V
15A
3A
12A
15A
18A
80mVp-p
2.0%
0.5%
1.0%
12V
7A
0.7A
5.3A
7A
8.4A
120mVp-p
8.0%
1.0%
5.0%
-5V
2A
0A
1A
2A
2.4A
80mVp-p
5.0%
0.5%
1.0%
-12V
2A
0A
1A
2A
2.4A
80mVp-p
5.0%
0.5%
1.0%
5V
15A
3A
12A
15A
18A
80mVp-p
2.0%
0.5%
1.0%
15V
5A
0.5A
4A
5A
6A
150mVp-p
6.0%
1.0%
5.0%
-5V
2A
0A
1A
2A
2.4A
80mVp-p
5.0%
0.5%
1.0%
-15V
2A
0A
1A
2A
2.4A
80mVp-p
5.0%
0.5%
1.0%
5V
15A
3A
12A
15A
18A
80mVp-p
2.0%
0.5%
1.0%
24V
3A
0.3A
2.3A
3A
3.6A
180mVp-p
8.0%
1.0%
5.0%
12V
2A
0A
1A
2A
2.4A
80mVp-p
5.0%
0.5%
1.0%
-12V
2A
0A
1A
2A
2.4A
80mVp-p
5.0%
0.5%
1.0%
5V
15A
3A
12A
15A
18A
80mVp-p
2.0%
0.5%
1.0%
24V
2.7A
0.3A
2.1A
2.7A
3.3A
180mVp-p
8.0%
1.0%
5.0%
15V
2A
0A
1A
2A
2.4A
80mVp-p
5.0%
0.5%
1.0%
-15V
2A
0A
1A
2A
2.4A
80mVp-p
5.0%
0.5%
1.0%
MPD-200B
MPT-200A
MPT-200B
MPT-200C
MPT-200D
MPQ-200B
MPQ-200C
MPQ-200D
MPQ-200F
Notes :
82%
83%
80%
80%
80%
81%
78%
78%
79%
81%
1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature.
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1 f & 47 f parallel capacitor.
3. Tolerance : includes set up tolerance, line regulation and load regulation.
4. 33% Duty cycle maximum within every 30 seconds. Average output power should not exceed the rated power.
5. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets
EMC directives.
6. Derating may be needed under low input voltages. Please check the derating curve for more details.
File Name:200W MEDICAL-SPEC 2008-03-25
200W Medical
1~4 Output Medical Type
Mechanical Specification(MPS-200)
series
Unit:mm
6.1[0.24"]
177.8[7"]
165.1[6.5"]
6.35[0.25"]
SVR1
LED 1
4
5
6
10
9
8
7
6
5
4
3
2
1
95.25[3.75"]
107.2[4.22"]
2
3
CN1
3
2
1
CN4
20
19
18
17
16
15
14
13
12
11
CN2
FG
15cm
4- 3.96[0.156"]
25CFM min.
35.5[1.4"]
FAN
3[0.118"] max.
DC Output Connector (CN4) : TKP DH2I-2*3 or equivalent
AC Input Connector (CN1) : JST B3P-VH or equivalent
Pin No.
1
2
3
Assignment
AC/N
No Pin
AC/L
Mating Housing
Terminal
JST VHR
or equivalent
JST SVH-21T-P1.1
or equivalent
Pin No.
1
2
3
4
5
6
DC Output Connector (CN2) : MOLEX 5566-20 or equivalent
Pin No.
1~5,11~15
6~10,16~20
Assignment
DC OUTPUT -V
DC OUTPUT +V
Mating Housing
Terminal
MOLEX 5557
or equivalent
MOLEX 5556
or equivalent
Derating Curve (MPS-200)
25CFM forced air cooling
TKP DH2-2*3
or equivalent
TKP DHT
or equivalent
90
80
Convection cooling
70
LOAD(%)
LOAD (%)
Terminal
100
80
40
20
-20
Mating Housing
Static Characteristics (MPS-200)
100
60
Assignment
PG
RSGND
RC+
RS+
RC-
0
10
20
30
40
50
AMBIENT TEMPERATURE ( )
60
70 (HORIZONTAL)
60
50
40
90
100
115
140
160
180
200
220
240
264
INPUT VOLTAGE (V) 60Hz
File Name:200W MEDICAL-SPEC 2008-03-25
200W Medical
1~4 Output Medical Type
series
Mechanical Specification(MPD/T/Q-200)
6.1[0.24"]
177.8[7"]
165.1[6.5"]
6.35[0.25"]
SVR1
LED 1
2
3
4
5
6
CN4
95.25[3.75"]
107.2[4.22"]
1
CN2
CN1
3
2
1
CN3
16
FG
15cm
4- 3.96[0.156"]
25CFM min.
35.5[1.4"]
FAN
3[0.118"] max.
DC Output Connector (CN4) : TKP DH2I-2*3 or equivalent
AC Input Connector (CN1) : JST B3P-VH or equivalent
Pin No.
1
2
3
Assignment
AC/N
No Pin
AC/L
Mating Housing
Terminal
JST VHR
or equivalent
JST SVH-21T-P1.1
or equivalent
Pin No.
1
2
3
4
5
6
DC Output Connector (CN2,3) : JST B8P-VH*2 or equivalent
Pin No.
1,2,3,4
5~11
12,13
14
15
16
Assignment
V1
COM
V2
V3
No pin
V4
Mating Housing
Terminal
JST VHR
or equivalent
JST SVH-21T-P1.1
or equivalent
Derating Curve (MPD/T/Q-200)
25CFM forced air cooling
MPD-200
TKP DH2-2*3
or equivalent
TKP DHT
or equivalent
90
MPT/Q-200
80
Convection cooling
70
MPT/Q-200
LOAD(%)
LOAD (%)
Terminal
100
80
MPD-200
40
20
-20
Mating Housing
Static Characteristics (MPD,T,Q-200)
100
60
Assignment
PG
RSGND
RC+
RS+
RC-
0
10
20
30
40
50
AMBIENT TEMPERATURE ( )
60
70 (HORIZONTAL)
60
50
40
90
100
115
140
160
180
200
220
240
264
INPUT VOLTAGE (V) 60Hz
File Name:200W MEDICAL-SPEC 2008-03-25