FUJI EDS3-110D

NDIR TYPE INFRARED GAS ANALYZER
DATA SHEET
ZRG
This NDIR gas analyzer features high accuracy, multiple
functions and easy operation through use of a microprocessor. It also utilizes a mass flow detector noted for its high
sensitivity and reliability.
This analyzer is used not only for measuring environmental pollution but is also used for various processes and/or
experiments.
This product, combined with sampling system (ZSP), has
passed the type-approval test based on Weight and Measure Act.
FEATURES
(1) Use of a microprocessor provides high accuracy, multiple functions and easy operation.
• A maximum of 3 gas components can be calibrated
with the built-in automatic calibrator (option).
• Using a signal from a zirconiz O2 sensor (ZFK3) or from
any other O2 sensor, the gas analyzer is able to output a value converted into O2 (option).
• Alarm function to emit high/low limit contact output
is also available (option).
• Measuring range can be selected by using external
single (option).
• Zero and span can be calibrated with high accuracy,
simply by pressing calibration keys.
• A self-diagnosis function is included.
(2) This analyzer utilizes mass flow detector featuring high
sensitivity and reliability. It has 2 standard measuring
ranges with a range ratio of up to 1:20.
(3) Three different gas components can be measured simultaneously by connecting a zirconia O2 sensor (ZFK3),
in addition to one-and two-component type sensors.
(4) Besides the standard measurement type, a sample switching type and a differential flow type are also available.
SPECIFICATIONS
General items
Power supply:
100V ± 10% AC, 50/60Hz
115, 220V AC also available on request
Note: Refer to "Code symbols"
Power consumption:
125VA max.
Ambient temperature:
–5 to +45°C
Ambient humidity:
90% RH or less
Enclosure:
Steel casing, indoor-use
Dimensions (H x W x D):
835 x 220 x 232 mm
Mass{weight}:
Approx. 24kg
Finish color:
Munsell 2.5Y 8.4/1.2
Indication:
4-digit LED for concentration
4-digit LED for sub-indication
Output hold:
Output hold prior to manual/auto calibration is possible. Hold ON/OFF is selectable.
Standard requirements for sample gas:
Temperature 0 to 50°C
(dehumidification 2°C saturation or less)
Dust
0.3µm or less
Pressure
9.8kPa{0.1kgf/cm2} or less
(flow rate 0.5R/min)
Standard control for sample gas:
Calibration gas Dry gas
Interfere control gas 2°C saturation
Warm-up time: About 8 hours (after power ON)
About 4 hours for sample switching type
Material of gas-contacting parts:
Sample cell; 304 stainless steel, neoprene
rubber
Infrared-ray transmitting window; CaF2 or
sapphire
Internal tubing; Teflon tube, silicone tube,
toaron tube
Gas inlet/outlet, purge gas inlet size:
Rc 1/4 (PT 1/4 internal thread) or NPT1/4
internal thread
EDS3-110d
Date
Jun. 29, 2005
ZRG
Purge gas flow rate:
1 ± 0.5R/min
Purging is required when measuring gas
is contained in the atmosphere or the
range of CO2 is 0 to less than 50ppm.
In other cases, purging should be made
as necessary.
Scope of delivery: Analyzer, test report, power fuse, cloth for
cleaning infrared-ray transmitting window
Mounting method:
Flush mounting (vertical mounting)
Installation condition:
Install the analyzer at a place not exposed
to direct sunlight or the radiation from a
high temperature object. Avoid vibration,
and select a clean place free from corrosive and/or combustible gases. If installing outdoors, provide a suitable casing or
cover to protect the analyzer from wind,
rain, etc.
Standard type
Measuring system:
Non-dispersion infrared-ray absorption
method, single light source, double beam
Measurable components and measuring range:
Standard single-component analyzer
Measurable component
CO
CO2
NO
SO2
CH4
Carbon monoxide
Carbon dioxide
Nitrogen monoxide
Sulfur dioxide
Methane
Min. measuring
range [ppm]
Measuring range
0 to 50
0 to 5
0 to 50
0 to 50
0 to 200
See Page 12
Measuring system:
Non-dispersion infrared-ray absorption
method, single light source, double beam,
sample switching system
Measurable component:
CO (carbon monoxide)
Measuring range:
1st range
2nd range, 3rd range, 4th range (*4) [ppm]
[ppm]
0 to 2
0 to 5
0 to 10
0 to 25
0 to 50
0 to 100
Output signal:
0 to 1V DC or 4 to 20mA DC, linear, step
output which changes every 50 seconds
Repeatability:
Within ± 1% of full scale
(*3) (within ± 2% of full scale)
Linearity:
Within ± 2% of full scale
Zero drift:
Within ± 0.5% of full scale/week
(*3) (within ± 1% of full scale/week)
Span drift:
Within ± 1.5% of full scale/week
(*3) (within ± 2.5% of full scale/week)
Response time: Within 120 seconds (for 90% response)
(*3) (within 150 seconds)
Measured gas flow rate:
1 ± 0.1R/min (reference gas)
1 ± 0.1R/min (sample gas)
Sample switching relay drive output:
3V DC, ON/OFF signal at 50 second interval (for solid state relay drive)
Note: *(3) (
*(4) (
Standard two-component analyzer
Measurable component
1st and 2nd component
NO + SO2
CO + CO2
NO + CO
Nitrogen monoxide +
sulfur dioxide
Carbon monoxide +
carbon dioxide
Nitrogen monoxide +
carbon monoxide
Min. measuring
range [ppm]
Measuring
range
0 to 100/0 to 100
0 to 100/0 to 100
Refer to
"Code
symbols"
0 to 250/0 to 250
Measuring range:Refer to table.
Output signal:
0 to 1V DC or 4 to 20mA DC (allowable
load resistance, 550Ω or less), linear
Repeatability:
within ± 0.5% of full scale
(*1) (within ± 1% of full scale)
Linearity:
Within ± 2% of full scale
Zero drift:
Within ± 2% of full scale/week
(*1) (within ± 2% of full scale/day)
(*2) (within ± 3% of full scale/week)
Span drift:
Within ± 2% of full scale/week
(*1) (within ± 2% of full scale/day)
(*2) (within ± 3% of full scale/week)
Response time: Max. 50 seconds (for 90% response) including gas substitution time; time differs
with the length of sample cell.
Measured gas flow rate:
0.5 ± 0.25R/min
1
Note: *( ) (
*(2) (
): For the range less than 50ppm
): For 50ppm range
Sample switching type
(Sample switching type gas analyzer suited for measurement of low concentrations or for eliminating the effects of
interfering components.)
2
Select from 5, 10, 20, 25, 50, 100
Max. range rate within 1:10
1st range<2nd range<3rd range<4th range
): For 0 to 2ppm range
): Sample switching type with 4-ranges is available (Max.
range ratio within 1:10). In 4-range analyzer, optional functions of remote range, range identification, alarm and external hold are not available.
Differential flow type
(Gas analyzer suited for measurement in two modes, absolute concentration and concentration difference)
Measuring system:
Non-dispersion infrared-ray absorption
method, single light source, double beam,
differential flow system
Measurable components and measuring range:
1st range
[ppm]
2nd range
[ppm]
CO2
Carbon
dioxide
0 to 500
– 50 to +50
–100 to +100 0 to 1000
CO
Carbon
monoxide
0 to 50
0 to 100
0 to 200
0 to 250
None, 0 to 100, 0 to 200, 0 to 250, 0 to 500
None, 0 to 200, 0 to 250, 0 to 500
None, 0 to 500
None, 0 to 500
Note: There are restrictions on the reference gas conditions.
Output signal:
1st range
2nd range
CO2
– 1 to +1V DC
0 to 1V DC
CO
0 to 1V or 4 to 20mA DC
0 to 1V or 4 to 20mA DC
Note: Linear output
Within ± 0.5% of full scale
within ± 2% of full scale
Within ± 2% of full scale/week
Within ± 2% of full scale/week
Max. 50 seconds (for 90% response) including gas substitution time
Measured gas flow rate:
0.5 ± 0.25R/min (reference gas)
0.5 ± 0.25R/min (sample gas)
Repeatability:
Linearity:
Zero drift:
Span drift:
Response time:
Optional specifications
(There are added on request. Refer to the "Code symbols".)
O2 correction output:
An private O2 sensor is used for correcting the measured gas concentration into
the value at standard O2 concentration.
For obtaining the NOx exhaust standard
value, NOx concentration and residual
oxygen concentration in exhaust gas are
measured simultaneously, then it is corrected according to the following equation. (Application of this equation is mandatory for the NOx exhaust standard.)
C= 21 – On
21 – Os
• Cs
where C: Concentration after O2 correction
Cs: NOx measured concentration
Os: O2 measured concentration
On: O2 standard concentration
Block diagram (Example of NOx measurement)
NOx signal
21 – On
•Cs
21 – Os
O2 corrected
NOx [ppm]
O2 signal
The O2 measured signal is according to the private O2 sensor (ZFK) or external O2 meter (0 to 1V DC/0 to 25% O2).
O2 output signal: 0 to 1V or 4 to 20mA DC, linear
O2 correction output signal:
0 to 1V or 4 to 20mA DC, linear
Output can be provided for each of 1st
and 2nd components
Alarm output:
Upper limit alarm
Contact output, SPDT (1c) contact
Contact capacity, 250V AC, 2A
(resistive load)
Lower limit alarm
Contact output, SPDT (1c) contact
Contact capacity, 250V AC, 2A
(resistive load)
Remote range changeover:
Range is changeable via external signal.
Range changeover input signal: 5V DC
(minimum range selection at 5V input)
Range identification signal output:
Contact output, SPST (1a) contact
Contact capacity, 250V AC, 2A
(resistive load)
External hold:
Output hold is possible with external input signal.
External hold signal input:
5V DC (hold at 5V input)
Average value output:
Moving average value or average value
is outputted at intervals of 1-hour or 4hour (only the 1-output type for 4-hour
average value is available).
Output signal: 0 to 1V or 4 to 20mA DC,
linear.
Automatic calibration:
Zero and span are automatically calibrated
at the present cycle.
Calibrating gas is supplied by operating
the external electromagnetic valve.
Calibration channels:
Up to 3 components can be calibrated
simultaneously.
Zero calibration point:
Fixed at 0% (air point and span point can
be set by zirconia O2 meter)
Span calibration point:
0 to 100% of full scale
Calibration start:
Built-in timer or remote start signal
Output hold during calibration:
Possible
Calibration gas flow mode:
(1) Zero gas
(2) Zero gas – span gas 1
(3) Zero gas – span gas 1 – span gas 2
(4) Zero gas – span gas 1 – span gas 3
(O2)
(5) Zero gas – span gas 1 – span gas 2
– span gas 3 (O2)
Calibration gas flow time:
Settable from 100 to 599 seconds
Calibration cycle:
1 to 99 hours (1-hour units) or 1 to 7 days
(1-day units)
Calibration failure alarm:
Provided when fault occurs during auto
calibration.
Contact output:
Under calibration; SPST (1a) contact, contact capacity 250V AC, 2A (resistive
load)
Calibration failure; SPST (1a) contact, contact capacity 250V AC, 2A (resistive
load)
Electromagnetic valve drive; SPST (1a)
contact, contact capacity 250V AC, 2A
(resistive load)
Remote start: Remote start signal
Voltage input 5V DC
(start at 100msec pulse input)
3
ZRG
FUNDAMENTAL DIAGRAM
(Standard type)
Distribution cell
(interference filter)
Interference compensating
detector
Reference cell
Trimmer
Infrared light source
(single light source)
Motor
Rotary sector
(one-point chopper)
Sample gas
Sample
gas inlet
Sector synchronous
signal
AC voltage
amplifier
Sample
gas outlet
Gain
changeover
O2 meter
(Option)
A/D
CPU
Contact input,
output
D/A
Indicator
Alarm output
Display
Range identification
signal
Contact output
O2
Various
during calibration
correction average
Remote range input
value
value
External hold input
(Option)
RAM
ROM
Operating key
Output,
0 to 1V or
4 to 20mA
DC
(Option)
(Differential flow type)
Distribution cell
(interference filter)
Reference
Trimmer
Reference gas outlet
Reference cell
gas inlet
Interference compensating
detector
Infrared light source
(single light source)
Motor
Rotary sector
(one-point chopper)
Sample gas
Sample
gas inlet
AC voltage
amplifier
Sample
gas outlet
Gain
changeover
Sector synchronous
signal
A/D
CPU
Contact input,output
D/A
Indicator
Alarm output
Indication
Range identification signal
Operating key
Calibration output
SV drive signal
Remote range input
External hold input
(Option)
4
Output,
0 to 1V or
4 to 20mA
DC
RAM
ROM
FUNDAMENTAL DIAGRAM
(Sample switching type)
Distribution cell
(interference filter)
Interference compensating
detector
Trimmer
Reference cell
Reference
gas inlet
Reference
gas outlet
Infrared light source
(single light source)
Motor
Sample cell
Rotary sector
(one-point chopper)
AC voltage
amplifier
Sample
gas inlet
Sample
gas outlet
Gain
changeover
Sector synchronous
signal
A/D
RAM
ROM
CPU
Contact input,output
D/A
Indicator
Relay drive signal
Alarm output
Indication
Range identification signal
Operating key
Calibration output
SV drive signal
Output,
0 to 1V or
4 to 20mA
DC
Remote range input
External hold input
(Option)
Description of sample switching system
The measured gas is divided into two, and in one of the
flow paths, a high performance converter is provided to convert carbon monoxide into carbonic acid gas. Here, the
carbon monoxide in the measured gas is eliminated and the
gas is led into one cell of the high-sensitivity infrared analyzer. In the other flow path, the gas is led directly into the
other cell of the analyzer.
These flow paths are changed over via changeover valves
SV1 and SV2 every 50 seconds by means of the changeover
valve drive signal transmitted from the analyzer. By carrying out this changeover cyclically, the output of the analyzer varies with the concentration of the carbon monoxide
in the measured gas. Use of the obtained variation width
as a measured value enables improving the S/N ratio and
eliminating the effect of interfering components plus zero
drift.
Temperature
controller
ZRG3
2ry filter
SV1
CO/CO2
converter
CO
measured 1ry filter
gas
Reference
cell
Flowrator
SV4
Sample
switching
type analyzer
Pump
Sample
cell
Reference gas
SV3
SV2
Relay
Output signal
SV1 SV2
5
ZRG
CODE SYMBOLS
(Standard single-component analyzer)
1 2 3 4 5 6 7 8
Z RG
9 10 11 12 13
Y 2 -
0
14 15 16 17 18 19 20
-
21
Description
Y
Measuring method
Single-component analyzer
Single-component analyzer
(approved by Weight and Measure Act)
1
5
Measurable components
SO2
CO
CO2
CH4
NO
Other non-standard items
A
B
D
E
P
Z
1st component, 1st range
0 to 5ppm
0 to 10ppm
0 to 15ppm
0 to 20ppm
0 to 25ppm
0 to 50ppm
0 to 100ppm
0 to 200ppm
0 to 250ppm
0 to 500ppm
0 to 1000ppm
0 to 2000ppm
0 to 5000ppm
0 to 1%
0 to 2%
0 to 5%
0 to 10%
0 to 20%
0 to 50%
0 to 100%
Other non-standard items
W
V
0
1
T
A
B
C
D
E
F
G
H
J
K
L
M
N
P
R
Z
0
2
3
4
A
B
Power supply
100V AC
115V AC
220V AC
220V AC
50/60Hz
50/60Hz
50/60Hz
50/60Hz
Output signal
0 to 1V DC
4 to 20mA DC
linear
linear
Optional functions [1]
Automatic
Average value
calibration
output
_
Y
A
B
D
F
H
K
M
™
_
_
™
™
_
™
Y
A
B
C
D
_
_
™
_
™
_
™
™
Note: (1) When selecting a range, be sure to refer to the
table of "Measuring range".
(2) The specifications for Dioxin CO analyzer of
200ppm max. are different in some points.
Piping connection
Rc 1/4
NPT 1/4
Rc 1/4
NPT 1/4
Range identification Note: (3) Details of average value
alarm, remote range,
output depend on 16th and
external hold
17th digits.
_
(4) Average value output
_
includes moving average
_
value output.
™
_
™
™
™
Optional functions [2]
O2 indication O2 indication
Range identification Note: (5) In the case of O2 indication,
(external O2 (ZFK connection) alarm, remote range
corrected O2 is outputted as
analyzer
standard specification.
connection)
(6) Standard type of O2 analyzer
_
_
_
has 2 ranges (0 to 10/25%)
_
_
™
_
_
™
_
™
™
_
™
™
Note: (7) Corrected O2 is designated by 18th digit.
(8) External O2 analyzer is required 0 to 1V linear output.
6
(Standard single-component analyzer) (cont'd)
1 2 3 4 5 6 7 8
Z RG
Y 2 -
9 10 11 12 13
0
14 15 16 17 18 19 20
-
21
Description
Y
Y
V
0
1
T
A
B
C
D
E
F
G
H
J
K
L
M
N
P
R
Z
0
1
4
5
6
Y
A
D
1st component, 2nd range
2nd range without
0 to 10ppm
0 to 15ppm
0 to 20ppm
0 to 25ppm
0 to 50ppm
0 to 100ppm
0 to 200ppm
0 to 250ppm
0 to 500ppm
0 to 1000ppm
0 to 2000ppm
0 to 5000ppm
0 to 1%
0 to 2%
0 to 5%
0 to 10%
0 to 20%
0 to 50%
0 to 1 00%
Other non-standard items
Note: (9) 2nd range>1st range
Range rate: within 1:20
Average value output time
Without
1-hour moving average value output
4-hour moving average value output
1-hour average value output
4-hour average value output
Note: (10) Average value output (option)
designated only by 12th digit.
Average value output object component
1st component
Instantaneous value
O2 corrected
_
_
_
™
_
Note: (11) Average value output (option)
designated only by 12th digit.
(12) For selection of 4-hour moving
average value output and 4-hour
average value output, only the
1-output type is available.
™
Note: (13) For designation of "Z", the data
O2 corrected standard value
must be put on the data code.
Without
4%
5%
6%
7%
10%
11%
12%
15%
Other non-standard items (specify within 0 to 19%)
Y
4
5
6
7
A
B
C
F
Z
EY
FY
GY
ZZ
Z
A
Measuring gas kind
Atmospheric gas
Combustible exhaust gas
(Dioxin included)
Converter exhaust gas
Other non-standard items
Note: (14) For designation of "Z", gas
components specifications should
be attached.
Non-standard spec.
Other non-standard items
Quick response
Note: (15) Data should be put on data code.
(16) Quick response type is available
only when gas density is 20%
or more.
7
ZRG
(Standard two-component analyzer)
1 2 3 4 5 6 7 8
Z RG
9 10 11 12 13
2 -
0
14 15 16 17 18 19 20
21
Description
-
Measuring method
Two-component analyzer
Two-component analyzer
(approved by Weight and Measure Act)
2
6
Measurable components (1st component) + (2nd component)
NO + SO2
CO + CO2
NO + CO
Other non-standard items
F
G
H
Z
1st component, 1st range
Code
B
C
D
E
F
Note:
B
Z
Code Range value
Code Range value
Range value
0 to 10 %
M
0 to 2000ppm
G
0 to 100 ppm
0 to 20 %
N
0 to 5000ppm
H
0 to 200 ppm
0 to 50 %
P
0 to 1%
J
0 to 250 ppm
0 to 100 %
R
0 to 2%
K
0 to 500 ppm
Other non-standard items
Z
0 to 5%
L
0 to 1000 ppm
(1) For range selection, be sure to refer to the table of "Measuring range".
(2) NO + CO min. range: 0 to 250ppm
2nd component, 1st range
Code
B
C
D
E
F
Note:
B
Z
Range value
Code Range value
Code Range value
0 to 100 ppm
G
0 to 2000ppm
M
0 to 10 %
0 to 200 ppm
H
0 to 5000ppm
N
0 to 20 %
0 to 250 ppm
J
0 to 1%
P
0 to 50 %
0 to 500 ppm
K
0 to 2%
R
0 to 100 %
0 to 1000 ppm
L
0 to 5%
Z
Other non-standard items
(3) For range selection, be sure to refer to the table of "Measuring range".
(4) NO + CO min. range: 0 to 250ppm
Power supply
100V AC
115V AC
220V AC
220V AC
0
2
3
4
50/60Hz
50/60Hz
50/60Hz
50/60Hz
Piping connection
Rc 1/4
NPT 1/4
Rc 1/4
NPT 1/4
Output signal
0 to 1V DC linear
4 to 20mA DC linear
Other non-standard items
A
B
Z
Optional functions [1]
Automatic
Average value
calibration
output
_
Y
A
C
E
G
J
L
N
™
_
_
™
™
_
™
_
_
™
_
™
_
™
™
Range identification Note: (5) Details of average value
alarm, remote range,
output depend on 16th and
external hold
17th digits.
_
(6) Average value output
_
includes moving average
_
value output.
™
_
™
™
™
Note: (7) This function applies to 1st and 2nd components.
Optional functions [2]
O2 indication O2 indication
Range identification Note: (8) In the case of O2 indication,
(external O2 (ZFK connection) alarm, remote range
corrected O2 is outputted as
analyzer
standard specification.
connection)
(9) Standard type of O2 analyzer
_
_
_
has 2 ranges (0 to 10/25%)
_
_
™
_
_
™
_
™
™
_
Y
A
B
C
D
™
™
Note: (10) Corrected O2 is designated by 18th digit.
(11) External O2 analyzer is required 0 to 1V linear output.
1st component, 2nd range
Y
Z
8
Code
Y
C
D
E
F
Note:
Code Range value
Code Range value
Range value
0 to 10 %
M
0 to 2000ppm
G
Without
0 to 20 %
N
0 to 5000ppm
H
0 to 200 ppm
0 to 50 %
P
0 to 1%
J
0 to 250 ppm
0 to 100 %
R
0 to 2%
K
0 to 500 ppm
Other non-standard items
Z
0 to 5%
L
0 to 1000 ppm
(12) 2nd range>1st range range rate: within 1:20
(Standard two-component analyzer) (cont'd)
1 2 3 4 5 6 7 8
Z RG
2 -
9 10 11 12 13
0
14 15 16 17 18 19 20
21
Description
2nd component, 2nd range
Code
Code Range value
G
Without
Y
H
0 to 200 ppm
C
J
0 to 250 ppm
D
K
0 to 500 ppm
E
L
0 to 1000 ppm
F
Note: (13) 2nd range>1st range
Y
Z
Code
Range value
M
0 to 2000ppm
N
0 to 5000ppm
P
0 to 1%
R
0 to 2%
Z
0 to 5%
range rate: within 1:20
Average value output time
Without
1-hour moving average value output
4-hour moving average value output
1-hour average value output
4-hour average value output
0
1
4
5
6
Range value
0 to 10 %
0 to 20 %
0 to 50 %
0 to 100 %
Other non-standard items
Note: (14) Average value output (option)
designated only by 12th digit.
Average value output object component
1st component
2nd component
Instantaneous value O2 corrected Instantaneous value O2 corrected
_
_
_
_
_
_
_
™
_
_
™
™
_
_
™
™
_
_
_
™
_
_
™
™
_
_
™
™
_
_
_
™
_
_
_
Y
A
B
C
D
E
F
G
H
Note:
(15)
Average value output
(option) designated only
by 12th digit.
(16)
For selection of 4-hour
moving average value
output and 4-hour average
value output, only the
1-output type is available.
™
O2 correct standard value
Note: (17) For designation of "Z", the data
Without
must be put on the data code.
4%
5%
6%
7%
10%
11%
12%
15%
Other non-standard items (specify within 0 to 19%)
Y
4
5
6
7
A
B
C
F
Z
EY
FY
GY
ZZ
Z
A
Measuring gas kind
Atmospheric gas
Combustible exhaust gas
(Dioxin included)
Converter exhaust gas
Other non-standard items
Note: (18) For designation of "Z", gas
components specifications should
be attached.
Non-standard spec.
Other non-standard items
Quick response
Note: (19) Data should be put on data code.
(20) Quick response type is available
only when gas density is 20%
or more.
9
ZRG
(Sample switching type)
1 2 3 4 5 6 7 8
Z RG
9 10 11 12 13
Y 2 -
0
14 15 16 17 18 19 20
-
Y 0 Y Y
21
Description
Measuring method
Sample switching system
3
Measurable components
CO
B
1st range
0 to
2ppm
0 to
5ppm
0 to
10ppm
0 to
25ppm
0 to
50ppm
0 to
100ppm
Other non-standard items
X
W
V
T
A
B
Z
0
2
3
4
A
B
50/60Hz
50/60Hz
50/60Hz
50/60Hz
Output signal
0 to 1V DC
4 to 20mA DC
linear
linear
Piping connection
Rc 1/4
NPT 1/4
Rc 1/4
NPT 1/4
Optional functions [1]
Automatic
Range identification
calibration
alarm, remote range,
external hold
_
_
_
™
_
YY
DY
AY
HY
™
™
™
2nd range/3rd range/4th range
4th range
3rd range
2nd range
—
—
—
—
—
5
—
—
10
—
—
20
—
10
5
—
20
5
—
20
10
20
10
5
—
—
10
—
—
25
—
—
50
—
25
10
—
50
10
—
50
25
50
25
10
—
—
25
—
—
50
—
—
100
—
50
25
—
100
25
—
100
50
100
50
25
—
—
50
—
—
100
—
100
50
—
—
100
Y
W
V
1
2
3
4
5
V
T
A
6
7
8
9
T
A
B
8
M
N
P
A
B
N
B
EY
ZZ
Z
10
Power supply
100V AC
115V AC
220V AC
220V AC
[Unit: ppm]
For 1st range 2ppm
For 1st range 5 ppm
For 1st range 10ppm
For 1st range 25ppm
For 1st range 50ppm
Measuring gas kind
Atmospheric gas
Other non-standard items
Note: (1) For designation of "Z", gas components
specifications should be attached.
Non-standard spec.
Other non-standard items
Note: (2) Data should be put on data code.
(Differential flow system)
1 2 3 4 5 6 7 8
Z RG
9 10 11 12 13
Y 2 -
0
14 15 16 17 18 19 20
-
Y 0 Y Y
21
Description
Measuring method
Differential flow system
4
Measurable component
CO2
(Carbon dioxide)
CO
(Carbon monoxide)
DQ
DS
BA
BB
BC
BD
Power supply
100V AC
115V AC
220V AC
220V AC
0
2
3
4
1st range
350 ± 50ppm
350 ± 100ppm
0 to 50ppm
0 to 100ppm
0 to 200ppm
0 to 250ppm
Piping connection
Rc 1/4
NPT 1/4
Rc 1/4
NPT 1/4
50/60Hz
50/60Hz
50/60Hz
50/60Hz
Output signal
0 to 1V DC linear
4 to 20mA DC linear
A
B
Note: (1) Voltage output signal only for 5th
and 6th digits, DQ or DS
(1st range: –1V to +1V)
(2nd range: 0 to 1V)
Optional functions (1)
Automatic
Range identification,
calibration
alarm, remote range,
external hold
_
_
_
™
_
YY
DY
AY
HY
™
™
™
2nd range [Unit : ppm]
Without
500 . For 1st range CO2 350±50
1000 . For 1st range CO2 350±100
100
200
For 1st range CO 50ppm
250
500
200
250
For 1st range CO 100ppm
500
500 . For 1st range CO 200ppm
500 . For 1st range CO 250ppm
Y
E
F
B
C
D
E
C
D
E
E
E
EY
ZY
Z
Measuring gas kind
Atmospheric gas
Other non-standard items
Note: (2) For designation of "Z", gas
components specifications should
be attached.
Non-standard spec.
Other non-standard items
Note: (3) Data should be put on data code.
11
ZRG
NDIR TYPE INFRARED GAS ANALYZER
<Standard single-component analyzer measuring range>
Measuable
component
CO
CO2
2nd
range
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
x 10
—
—
—
—
—
—
—
x 20
—
—
—
—
—
—
—
—
—
—
—
—
x2
—
—
—
—
—
x 2.5
—
—
—
—
—
x4
—
—
—
—
—
x5
—
—
—
—
—
x8
—
—
—
—
—
x 10
—
—
—
—
—
x 20
—
—
—
—
—
x0
x 20
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
x0
—
x2
—
x 2.5
—
x4
—
x5
—
x8
—
x 10
—
x 20
—
x0
—
x2
—
x 2.5
—
x4
—
x5
—
x8
—
x 10
—
x 20
—
x0
—
—
—
—
x2
—
—
—
x 2.5
—
—
—
x4
—
—
x5
—
x8
x 2.5
x4
x5
x8
x 10
SO2
CH4
10ppm 15ppm 20ppm 25ppm 50ppm 100ppm 200ppm 250ppm 500ppm 0.1%
x0
x2
NO
1st measuring range
5ppm
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
0.2%
0.5%
1%
2%
5%
10%
20%
50%
100%
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
—
—
Remarks: (1) ™ : Standard measuring range
(2)
: Consult us regarding manufacture availability, price and delivery period.
(3) — : Not manufactured
<Standard type two-component analyzer measuring range>
NO
SO2
100
200
250
500
ppm
ppm
ppm
ppm
™ ™ ™ ™
200ppm
™ ™ ™ ™
250ppm
™ ™ ™ ™
500ppm
™ ™ ™ ™
Remarks : (1) ™ : Scope of manufacture for 1st range
100ppm
(2) 2nd range type is available provided the
range ratio of 1st or 2nd component is
within 1:20.
(Max. 2000ppm for NO analyzer)
NO
CO
250ppm
500ppm
1000ppm
250
500
1000
ppm
ppm
ppm
™
™
™
™
™
™
™
™
™
Remarks : (1) 2nd range type is available provided the
range ratio of 1st or 2nd component is
within 1:20.
(Max. 2000ppm for NO analyzer)
12
100
ppm
200
ppm
250
ppm
500
ppm
0.1
%
0.2
%
0.5
%
1
%
2
%
5
%
10
%
20
%
50
%
100
%
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
⊗
⊗
⊗
—
—
—
—
—
—
—
⊗
⊗
—
—
—
—
—
—
—
⊗
⊗
⊗
—
—
—
—
—
—
—
—
—
—
—
⊗
⊗
⊗
—
—
—
—
⊗
⊗
™
™
™
⊗
⊗
™
™
™
⊗
⊗
™
™
™
™
0.1%
⊗
⊗
⊗
0.2%
⊗
⊗
⊗
⊗
⊗
—
—
—
—
—
⊗
⊗
⊗
⊗
⊗
⊗
—
—
—
—
—
⊗
⊗
⊗
⊗
⊗
⊗
—
—
—
—
—
⊗
⊗
⊗
™
™
™
™
™
™
™
⊗
2%
™
™
™
⊗
—
™
™
™
0.5%
—
1%
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
CO
CO2
100ppm
200ppm
250ppm
500ppm
5%
—
—
—
—
—
—
⊗
⊗
⊗
10%
—
—
—
—
—
—
—
⊗
⊗
20%
—
—
—
—
—
—
—
—
⊗
50%
—
—
—
—
—
—
—
—
—
100%
—
—
—
—
—
—
—
—
—
Remarks : (1)™ : Scope of manufacture for 1st range
⊗ : Scope of manufacture for 2nd range for CO and CO2 (maximum range)
(2) 2nd range type is available provided the range ratio of 1st or 2nd component is within 1:20.
OUTLINE DIAGRAM
(Unit:mm)
NOTE* Lid is mounted when
reference gas is not used.
Reference gas inlet*
Reference gas outlet*
Sample gas
inlet
Purge gas inlet
Rc or NPT1/4
Mounting size
Sample gas
outlet
813
< Top panel >
External terminal
A/O
Approx.230
OPTION 1
OPTION 2
AUTO CAL
M3.5 screw
140
3- 12hole
Mounting screw
M10 should be used.
ZRG
Cable inlet 22
(with rubber grommet)
See from P
Operation panel
61.5
813
835
785
94
Rock bracket
< Front panel >
Power supply
terminal
(inside)
< Side panel >
140
220
20
232
Cable inlet 22
(with rubber grommet)
P
13
ZRG
External connection diagram
1st
component
2nd
component
0 to 1V or
4 to 20mA
DC (instantaneous value)
0 to 1V or
4 to 20mA
DC (moving
average value)
(option)
A/C
OPTION 1
External hold +
+
1
11
+
-
2
12
-
+
3
13
+
-
4
14
-
0 to 1V or
4 to 20mA
DC (instantaneous value)
+
5
15
+
-
6
16
-
0 to 1V or
4 to 20mA
DC (moving
average value)
(option)
+
7
17
+
-
8
18
-
O2 sensor input
O2
0 to 1V or
4 to 20mA
DC (instantaneous value)
1st component
O2 corrected value
0 to 1V DC or
4 to 20mA DC
O2 2nd component
corrected value
0 to 1V DC or
4 to 20mA DC
1st component
Remote range
input (option)
+
9
10
O2
Upper limit alarm
contact output
(option)
2
12
3
13
4
14
5
15
6
16
7
17
8
18
9
19
10
20
+
2nd component
R1 Remote range
input (option)
R1
1st component
Range identification
contact output
(option)
19
20
O2
Remote range
input (option)
OPTION 2
2nd component
Upper limit alarm
contact output
(option)
11
2nd component
Range identification
contact output
(option)
+
FAULT
1st component
Upper limit alarm
contact output
(option)
1
1
11
2
12
3
13
4
14
5
15
6
16
7
17
8
18
9
19
10
20
O2
Remote range
input (option)
AUTO CAL
1st component
Lower limit alarm
contact output
(option)
Calibration
contact output
Zero gas
contact output
2nd component
Lower limit alarm
contact output
(option)
O2
Lower limit alarm
contact output
(option)
1
11
2
12
3
13
4
14
5
15
6
16
7
17
+
8
18
-
9
19
+
10
20
-
Auto calibration
failure contact output
Pump ON/OFF contact
+
Span gas-1
contact output
Span gas-2
contact output
Span gas-3
contact output
Remote start input
-
SV2
Relay drive output signal
(sample switching type only)
SV
Relay drive
output signal
(Differential or
sample switching type only)
Note: For relay drive output signal terminal, only the sample switching type of differential flow
system is used.
Terminal block should be used at the position of AUTO-CAL terminal block.
Power supply (installed in the main unit)
Ground
Power supply
14
Private zirconia O2 sensor <option>
This sensor is used together with ZRG.
Measuring method:
Zirconia system
Measurable component and measuring range:
Measurable component
O2
Oxygen
1st range
2nd range
0 to 10 vol %
0 to 25 vol %
Repeatability:
Within ± 0.5% of full scale
Linearity:
Within ± 2% of full scale
Zero drift:
Within ± 1% of full scale/week
Span drift:
Within ± 2% of full scale/week
Response time: Approx. 20 sec (for 90% response)
Measured gas flow rate:
0.5 ± 0.25R/min
Indication:
Temperature indication (LED)
Temperature alarm output:
Contact output, SPST (1a) contact
Contact capacity, 220V AC 1A
(resistive load)
Dimensions (H x W x D):
140 x 170 x 190mm
Mass{weight}:
Approx. 3kg
Finish color:
Munsell 5 Y7/1
CODE SYMBOLS
1 2 3 4 5 6 7 8
Z F K
9 10 11 12 13
Note: The Zirconia system, due to its principle,
may produce a measuring error due to rela-
Measuring method
Zirconia system
Zirconia system, (approval by Weight
and Measure Act)
3YY
4YY
tive concentration versus combustible O2
gas concentration. Also, a corrosive gas
(SO2 of 250ppm or more, etc.) may affect
Power supply
90 to 126V AC 50/60Hz
200 to 240V AC 50/60Hz
1
3
the life of the sensor.
Gas inlet/outlet size:
Rc1/4
Power supply:
90 to 126V or 200 to 240V AC, 50/60Hz
Enclosure:
Steel casing, for indoor application
OUTLINE DIAGRAM
Description
4 - 1 Y 0 YY
(Unit:mm)
170
152
OXYGEN ANALYZER
800
110
131
140
TEMP
POWER FUSE3A
1
OFF
2
AC
3
E
4
+
OUT
6
7
ALM
17
7-M4
5
5
Gas inlet
Rc 1/4
172.3
Gas outlet
Rc 1/4
External connection diagram
(M4 screw terminal)
1
2
3
4
5
+
AC power
supply,
50/60Hz
E
6
7
-
Output
to ZRG
Alarm
output
15
ZRG
Caution on Safety
*Before using this product, be sure to read its instruction manual in advance.
Head Office