FUJI EDS3-123M

NDIR TYPE INFRARED GAS ANALYZER
(5-COMPONENT ANALYZER)
ZKJ
DATA SHEET
This gas analyzer (ZKJ) is capable of measuring the concentrations of NO, SO2, CO2, CO, CH4, N2O and O2 components in sample gas.
NO, SO 2, CO 2, CO, CH 4 and N 2O are measured by nondispersion infrared method (NDIR), while O 2 is measured
by built-in type paramagnetic method sensor or externalmount type zirconia method sensor. A maximum of 5
components including O2 (max. 4 components except for
O2 measurement) are simultaneously measurable.
The mass flow type twin detector of high sensitivity and
reliability adopted in the infrared ray method detection
block makes the measurement hardly affected by interfering components.
In addition, a microprocessor is built in and a large-size
liquid crystal display is equipped for easier operation,
higher accuracy and more functions.
Optimum as an analyzer unit of measurement system
for combustion exhaust gas from refuse incinerator and
boiler, or gas from different industrial furnaces.
FEATURES
1.Measure five components including O 2 simultaneously
and continuously
Simultaneously and continuously measures up to four
components out of NO, SO 2, CO, CO2, CH4 and N2O,
plus O2, or up to totally five components.
2. Hardly affected by interference by other gases
The mass flow type twin detector of high sensitivity
and reliability adopted makes the measurement hardly
affected by interfering components, ensuring a stable
operation.
3. Equipped with abundant functions
O2 conversion, average value computation, automatic
calibration, one touch calibration, upper/lower limit
alarm, remote measurement range changeover, range
identification signal output, etc. incorporated can
configure applications to match particular uses.
4. Easy-to-see large LCD unit
The large LCD unit adopted allows observing easily
the indication of all measured components and
computation values.
The interactive operation facilitates setting.
5. 19 inch rack mount structure
The mainframe unitized to 19 inch rack t ype and
electrical signal input/output terminal unit also unitized
easily configure a gas analyzer system.
6. Maximum range ratio is 1 to 25
Measuring ranges are changeable.
7. Drift +/-1% FS/week (more than 0 to 200ppm range)
SPECIFICATIONS
Standard Specifications
Principle of measurement:
NO, SO2, CO2, CO, CH4, N2O;
Non-dispersion infrared-ray absorption
method
Single light source and double beams
(double-beam system)
O2 ; Paramagnetic O2 sensor (built in) or zirconia O2 sensor (externally installed)
Measurable gas components and measuring range:
NO
SO2
CO2
CO
CH4
N2O
O2
(built in)
O2
(External Zirconia)
Minimum range
0 – 50ppm
0 – 50ppm
0 – 20ppm
0 – 50ppm
0 – 200ppm
0 – 200ppm
Maximam range
0 – 5000ppm
0 – 10vol%
0 – 100vol%
0 – 100vol%
0 – 100vol%
0 – 2000ppm
0 – 5vol%
0 – 25vol%
0 – 5vol%
0 – 25vol%
• Max. 5 components measurement including O2.
• Measuring range ratio ≤ 1:5 (O2)
≤ 1:25
(except for O2)
• Measuring ranges are changeable between
the specified minimum and maximum range
Settable one range or two ranges
• When only N2O analyzer is used, make sure
not to contain any components other than
N2O.
Multicomponent analyzers including N 2 O
analyzer + CO2 analyzer are used for sludge
incineration. In this case, the range of N 2O is
0 to 200ppm/500ppm, and the range of CO 2
is 0 to 10%/20%.
*For measurable components and possible combinations
of measuring ranges, refer to Tables 1-(1) to (3).
EDS3-123m
Date
May. 23, 2011
ZKJ
Measured value indication:
Digital indication in 4 digits
(LCD with back light)
• Instantaneous value of each component
• Instantaneous value after O2 conversion
(only in NO, SO2, CO measurement
with O2)
• Average value after O2 conversion
(only in NO, SO2, CO measurement
with O2)
• O2 average value
Analog output signals:
*Inputs/outputs of analog signals are
possible by combining with the input/
output terminal module.
4 to 20mA DC or 0 to 1V DC,
non-isolated output ; 12 points max.
max.load550Ω. for 4 to 20 mA DC
min.load 100kΩ. for 0 to 1V DC
*Refer to Table 2, for the channel No.
of displayed values and analog output
signals.
Analog input signal:
For signal input from externally installed
O2 sensor.
Signal requirement;
(1)Signal from Fuji’s Zirconia O2 sensor (TYPE: ZFK7)
(2)0 to 1V DC from an O2 sensor
Input section is not isolated. This feature is effective when an O2 sensor is
not built in.
(Depend on O 2 input signal, measured
concentration indication and O 2 conversion.)
Relay contact output:
1a contact (250V AC/2A, resistive load)
Instrument error, calibration error,
range identification, auto calibration
status, pump ON/OFF, peak alarm.
1c contact (250V AC/2A, resistive load
selectable 6 outputs)
High/Low limit alarm contact output.
Power disconnection alarm.
* All relay contacts are isolated mutually and from the internal circuit.
Contact input: No-voltage contact (ON/0V, OFF/5V
DC, 5mA flowing at ON)
Remote range switch, auto calibration remote start, remote holding,
average value resetting, pump ON/
OFF
Isolated from the internal circuit with
photocoupler. Contact inputs are not
isolated from one another.
Transmission output:
Solenoid valve drive signal for automatic calibration.
Transistor output (100mA or less)
Power supply: Voltage rating ; 100V to 240V AC
Allowable range; 85V to 264V AC
Frequency
; 50Hz/60Hz
Power consumption; 250VA max.
Inlet
; Conform to EN60320 Protection Class 1
Operating conditions:
Ambient temperature;-5°C to 45°C
Ambient humidity ;90% RH max.,
non-condensing
2
Storage conditions:
Ambient temperature; -20°C to 60°C
Ambient humidity ;100% RH max.,
non-condensing
Dimensions (H x W x D):
Analyzer main unit;
177 x 483 x 600mm
Input/output terminal module;
164 x 318 x 55mm
Mass:
Approx. 22 kg (only Analyzer)
Finish color: Front panel; Off-white (Munsell 10Y7.5/0.5
or equivalent)
Casing; Plating, Steel-blue (gray)
Enclosure: Steel casing, for indoor use
Material of gas-contacting parts:
Gas inlet/outlet; SUS304
Sample cell; SUS304,chloroprene rubber
Infrared-ray transmitting window; CaF 2
O2 sensor sample cell : SUS316
Internal piping; Toaron, Teflon
Gas inlet/outlet: Rc1/4 or NPT1/4 internal thread
Purge gas flow rate:1L/min ( when required)
Standard Functions
Output signal holding:
Output signals are held during manual
and auto calibrations by activation of
holding (turning “ON” its setting).
The values to be held are the ones just
before start calibration mode or setting
value.
It is selectable.
Indication of instantaneous values will
not be held.
Remote output holding:
Output signal is held at the latest value
or setting value by short-circuiting the
remote output holding input terminals.
Holding is maintained while the terminals are short-circuited. Indication of
instantaneous values will not be held.
Switch ranges :
The switch ranges is available in manual, auto, and remote modes. Only preset switch method is effective.
Manual:
Allows range to switch by key operation.
Auto:
Allows range to switch from low to high
range when 90%FS or more is available
in the low range.
Allows range to switch from high to
low range when 80%FS or less is available in the low range.
Remote:
No-voltage contact input (for measurable components)
Allows range to switch via an external
signal when remote range switch input
is received.
When the contact input terminals for
each component are short-circuited,
the first range is selected, and it is
switched to the second range when the
terminals are open.
Range identification signal:
The present measuring range is identified by a contact signal.
The contact output terminals for each
component are short-circuited when
the first range is selected, and when
the second range is selected, the terminals are open.
Auto calibration:
Auto calibration is carried out periodically at the preset cycle.
W h e n a s t a n d a r d g a s c y l i n d e r fo r
calibration and a solenoid valve for
opening/closing the gas flow line are
prepared externally by the customer,
calibration will be carried out with the
solenoid valve drive contacts for zero
calibration and each span calibration
turned on/off sequentially at the set
auto calibration timing.
Auto calibration cycle setting:
Auto calibration cycle is set.
Setting is variable within 1 to 99 hours (in
increments of 1 hour) or 1 to 40 days (in
increments of 1 day).
Gas flow time setting:
The time for flowing each calibration
gas in auto calibration is set.
Settable within 60 to 900 seconds (in
increments of 1 second)
Auto calibration remote start:
Auto calibration is carried out only once
according to an external input signal.
Calibration sequence is settable in the
same way as the general auto calibration.
Auto calibration is started by opening
the auto calibration remote start input
terminals after short-circuiting for 1.5
seconds or longer.
Auto zero calibration:
Auto zero calibration is carried out periodically at the preset cycle.
This cycle is independent on “Auto calibradion” cycle.
When zero calibration gas and solenoid
valve for opening/closing the calibration
gas flow line are prepared externally
by the customer, zero calibration will
be carried out with the solenoid valve
drive contact for zero calibration turned
on/off at the set auto zero calibration
timing.
Auto zero calibration cycle setting:
Auto zero calibration cycle is set.
Setting is variable within 1 to 99
hours (in increments of 1 hour) or
Setting is variable within 1 to 40 days
(in increments of 1 day)
Gas flow time setting:
The timing for flowing zero gas in
auto zero calibration is set.
Settable 60 to 900 seconds (in increments of 1 second)
High/Low limit alarm:
Alarm contact output turns on when
measurement value reach to the preset
high or low limit alarm value.
Contacts close when the channel value
of each channel becomes larger than
the high alarm limit value or smaller
than the low alarm limit value.
Instrument error contact output:
Contacts close at occurrence of analyzer error No. 1, 3 or 10.
Calibration error contact output:
Contacts close at occurrence of manual
or auto calibration error (any of errors
No. 4 to 9).
Auto calibration status contact outputs: Contacts close during auto calibration.
Pump ON/OFF contact output:
During measurement, this contact
close. While calibration gas is flowing,
this contact open. This contact is connected in power supply of pump, and
stop the sample gas while calibration
gas flowing.
3
ZKJ
Optional Functions
Performance
O2 conversion: Conversion of measured NO, SO 2 and
CO gas concentrations into values at
standard O2 concentration
Conversion formula: C =
21–On
× Cs
21–Os
C : Sample gas concentration after O 2
conversion
Cs: Measured concentration of sample
gas
Os: Measured O2 concentration
(Limit settable, 1 to 20%O 2)
On: Standard O 2 concentration (value
changeable by setting; 0 to 19%O2)
Average value after O2 conversion and O2 average valu
ecalculation: The result of O2 conversion or instantaneous O2 value can be outputted as an
average value in the preset period of
time.
Used for averaging is the moving average method in which sampling is carried out at intervals of 30 seconds.
(Output is updated every 30 seconds.
It is the average value in the determined period of time just before the
latest updating.)
Averaging time is settable within 1 to
59 minutes (in increments of 1 minute) or 1 to 4 hours (in increments of 1
hour).
Average value resetting:
The above-mentioned output of average
value is started from the initial state by
opening the average value resetting input terminals after short-circuiting for 1.5
seconds or longer.
Output is reset by short-circuiting and
restarted by opening.
CO concentration peak count alarm:
(added only for CO/O2 measurement)
Alarm output turns on according to the
preset concentration and count.
Whenever the instantaneous value of
CO exceeds the preset concentration
value, count increments. If the count
exceeds the preset value in one hour,
the alarm contacts close.
Communication function:
RS-232C (9pins D-sub)
Half-duplex bit serial
Start-stop synchronization
ModbusTM protcol
Contents: Read/Wright parameters
Read measurement concentration and instrument status
Remark: When connecting via RS-485 interface, a RS-232C
RS-485
converter should be used.
4
: ±0.5% of full scale
: ±1% of full scale
: ±1% of full scale/week
(±2% of full scale/week; range be
tween 0 to 50ppm and 0 to 200ppm)
(±2% of full scale/day; smaller than 0 to 50ppm range)
Span drift
: ±2% of full scale/week
(±2% of full scale/day; smaller than 0 to 50ppm range)
Response time :
(for 90% FS response)
15 sec electrical response
Within 60 seconds including replacement time of sampling gas (when gas
flow rate is 0.5L/min)
Gas replacement time depends on the
number of measuring components,and
measuring range
Repeatability
Linearity
Zero drift
Standard Requirements for Sample Gas
: 0.5L / min ±0.2L / min
: 0 to 50°C
: 10 kPa or less (Gas outlet side should
be open to the atmospheric air.)
Dust
: 100µg/Nm 3 or less in particle size of
1µm or less
Mist
: Unallowable
Moisture
: Below a level where saturation occurs
at 2°C (condensation unallowable).
Corrosive component:
1 ppm or less
Standard gas for calibration:
Zero gas; Dry N2
Span gas ; Each sample gas having
concentration 90 to 100%
of its measuring range (recommended). Gas beyond concentration
100%FS is unusable.
In case a zirconia O2 analyzer is installed
externally and calibration is carried out
on the same calibration gas line:
Zero gas ;Dry air or atmospheric air
(provided without CO 2 sensor)
Span gas ; For other than O2 measurem e n t , e a ch s a m p l e g a s
h av i n g c o n c e n t r a t i o n 9 0
to 100% of its measuring
range.
For O 2 measurement, O 2
gas of 1 to 2 vol%.
Flow rate
Temperature
Pressure
Installation Requirements
• Indoor use. (Select a place where the equipment does
not receive direct sunshine, draft/rain or radiation from
hot substances. If such a place cannot be found, a roof
or cover should be prepared for protection.)
• Avoide a place where receives heavy vibration
• Select a place where atmospheric air is clean
EC Directive Compliance
The product conforms to the requirements of the Low
Voltage Directive 73/23/EEC and EMC directive 89/336/
EEC (as amended by Directive 92/31/EEC), both as
amended by Directive 93/68/EEC.
It conforms to following standards for product safety and
electromagnetic compatibility ;
EN61010-1 : 2001 Safet y requirements for electrical
equipment for measurement, control
and laboratory use.
“Installation Category II”
“Pollution Degree 2”
EN61326-1 : 1997, AI: 1998, A2: 2001
Electrical equipment for measurement,
control and laboratory use — EMC requirements.
*The product mounted in a steel enclosure conforms
to the requirements of EMC directive.
5
ZKJ
Principle diagram of NDIR type measurement (For NO, SO2, CO2, CO, CH4, N2O)
Distribution cell
(interference filter)
Motor
Trimmer
Interference compensating
detector
Reference cell
Rotary sector
Sample cell
Sample gas
outlet
Sample gas
inlet
Gain changeover
Infrared (single)
light source
AC amplifier
Sector
sync. signal
O2 sensor
(option)
A/D
CPU
ROM
Output
0 to 1V or
4 to 20mA DC
Indication
Indicator
D/A
Operating keys
External input / Contact output
RAM
Principle diagram of paramagnetic type measurement (For O2)
Permanent
magnet
Mirror
Measuring
cell
Gas inlet
Magnetic field
Dumbbell
Gas outlet
Preamplifier
Permanent
magnet
LED
Photodiode
Indication
Output
6
Signal processing
and calculation block
Example configuration of gas sampling system
The following illustrates a typical system configuration for five component gas measurement for monitoring combustion
exhaust gas from boiler, refuse incinerator, etc.
Contact FUJI ELECTRIC for system configuration matching the particular use or further information.
(1) Gas extractor
ZKJ Unit
Gas inlet tube
(2) Mist filter
(8) Flowmeter
(7) Membrane filter
CO, CO2
(O2)
(4) Gas aspirator
15° or larger
(11) NO2/NO
converter
for NOx
measurement
(5)
Electronic
gas cooler
(7) Membrane
filter
Infrared gas
analyzer main
unit
NO, SO2
(3) Gas conditioner
(6) Solenoid valve
ZERO
NO
SO2
CO2
CO
O2
Pressure
reducing valve
Gas outlet
N2
or
Air
NO
/N2
CO2
/N2
CO
/N2
O2
/N2
(10) Zirconia
O2 sensor
(9) Standard gas
Drain
Functions of Individual Components
(1)Gas extractor:Gas extractor with a heating type
stainless steel filter of standard mesh
40µm
(2)Mist filter:
For separation of drain and removal of
dust and mist
(3)Safety drain trap:
Prevention of drain from being sucked
and composite operation of constantpressure bubbler
(4)Gas aspirator:For aspiration of sample gas (sample
gas flow rate approx. 2L/min)
(5)Electronic gas cooler:
Dries the moisture in sample gas to a
dew point of approx. 3°C.
(6)Solenoid valve:Used for introducing calibration gas.
(7)Membrane filter:
PTFE filter used to eliminate fine dust
particles and permit monitoring of dust
adhering condition on the front panel of
the gas analyzer.
SO2
/N2
O2
(8)Flowmeter: Adjusts and monitors the flow rate of sample gas.
(9)Standard gas: Reference gas used for calibrating zero
and span of the analyzer. Total 6 cylinders required for air, zero gas, span gas
NO, SO2, CO, CO2 and O2.
(10)Zirconia O2 sensor:
(This is not necessary in case when the
zirconia type O2 sensor is built-in.)
External zirconia oxygen sensor used
for measuring the oxygen concentration
(0 to 25%) in sample gas.
(11)Converter: Added to NOx analyzer.
A special catalyst material for efficient
conversion of NO2 gas to NO is used.
*(Note)
For each gas sampling device, refer to the separate Data Sheet for each gas sampling device.
7
ZKJ
CODE SYMBOLS
1 2 3 4 5 6 7 8
Digit
Description
4
<Custom specifications>
Standard
5
<Measurable component (SO2, CO2, CO, CH4, N2O) >
1st
2nd
3rd
4th
componet componet componet componet
NO
SO2
CO2
CO
CH4
NO
SO2
NO
CO
CO2
CO
CO
NO
SO2
NO
SO2
CO2
CO
N2O
CO2
N2O
NO
N 2O
CO2
SO2
N2O
CO2
N2O
CO2
CO
N2O
CO2
CH4
NO
SO2
N2O
CO2
6
<Measurable component (O2)>
None
External zirconia type O2 sensor (Model : ZFK7)
External O2 analyzer
Built-in paramagnetic type O2 sensor
7
<Gas inlet/outlet>
Rc1/4
Rc1/4, with purging
NPT1/4
NPT1/4, with purging
Resin(ø6)
8
<Revision code>
9
<Accessories>
None
With relay board for auto caliblation, with cable
With slide rail
With slide rail, relay board, and cable
10 <Indication and power supply cable>
In Japanese, Power cable rated 125V
In English, Power cable rated 125V (UL)
In English, Power supply rated 250V (CEE)
11 <Measuring range> 1st component
12 Minimum range
Maximum range
0 to 20ppm
0 to 500ppm
0 to 50ppm
0 to 1000ppm
0 to 100ppm
0 to 2000ppm
0 to 200ppm
0 to 500ppm
0 to 200ppm
0 to 2000ppm
0 to 200ppm
0 to 5000ppm
0 to 500ppm
0 to 1%
0 to 1000ppm
0 to 2%
0 to 2000ppm
0 to 5%
0 to 5000ppm
0 to 10%
0 to 1%
0 to 20%
0 to 2%
0 to 10%
0 to 2%
0 to 50%
0 to 10%
0 to 20%
0 to 10%
0 to 100%
13 <Measuring range> 2nd component
14 Minimum range
Maximum range
None
None
0 to 50ppm
0 to 1000ppm
0 to 100ppm
0 to 2000ppm
0 to 200ppm
0 to 500ppm
0 to 200ppm
0 to 2000ppm
0 to 200ppm
0 to 5000ppm
0 to 500ppm
0 to 1%
0 to 1000ppm
0 to 2%
0 to 5000ppm
0 to 10%
0 to 1%
0 to 20%
0 to 2%
0 to 50%
0 to 10%
0 to 20%
0 to 10%
0 to 100%
15 <Measuring range> 3rd component
16 Minimum range
Maximum range
None
None
0 to 50ppm
0 to 1000ppm
0 to 100ppm
0 to 1000ppm
0 to 100ppm
0 to 2000ppm
0 to 200ppm
0 to 500ppm
0 to 200ppm
0 to 2000ppm
0 to 200ppm
0 to 5000ppm
0 to 500ppm
0 to 1%
0 to 1000ppm
0 to 2%
0 to 2000ppm
0 to 5%
0 to 5000ppm
0 to 10%
0 to 1%
0 to 20%
0 to 2%
0 to 10%
0 to 10%
0 to 20%
0 to 10%
0 to 100%
8
note
ZKJ F
9 10 11 12 13
14 15 16 17 18 19 20
4
F
note 10
note 10
note 10
note 10
note 10
note 10
note 10
note 1b)
note 1a)b)
note 2a)
note 2a)
note 2b)
P
A
D
B
E
F
H
G
L
M
Q
R
S
T
U
V
W
Y
A
B
C
0
1
2
3
4
4
Y
A
C
D
note 3
J
E
U
note 4
note 5
1 E
AF
BG
CE
CG
CH
E J
FK
GL
HM
JN
KM
KP
MN
MR
note 4
Y
A
B
C
C
C
E
F
H
J
K
M
M
Y
F
G
E
G
H
J
K
M
N
P
N
R
note 4
YY
AF
BF
BG
CE
CG
CH
EJ
FK
GL
HM
JN
KP
MN
MR
21 22 23
Digit No.
of code
1 2 3 4 5 6 7 8
Digit
Description
17 <Measuring range> 4th component
18 Minimum range
Maximum range
None
None
0 to 50ppm
0 to 1000ppm
0 to 100ppm
0 to 2000ppm
0 to 200ppm
0 to 500ppm
0 to 200ppm
0 to 2000ppm
0 to 200ppm
0 to 5000ppm
0 to 500ppm
0 to 1%
0 to 1000ppm
0 to 2%
0 to 5000ppm
0 to 10%
0 to 1%
0 to 20%
0 to 2%
0 to 50%
0 to 10%
0 to 20%
0 to 10%
0 to 100%
19 <O2 analyzer, 1st range>
20 Minimum range
Maximum range
None
None
0 to 5%
0 to 25%
0 to 10%
0 to 25%
Other
21 <Output>
4 to 20mA DC
0 to 1V DC
4 to 20mA DC + Communication function
0 to 1V DC + Communication function
22 <O2 conversion and O2 average value output>
None
With O2 conversion output
With peak alarm
With O2 conversion output and peak alarm
23 <Adjustment, Range designation>
For combustion exhaust gas (specified range)
For combustion exhaust gas
For combustion exhaust gas (specified range)
Unit mg/m3
For combustion exhaust gas, Unit mg/m3
Others
note
note 4
ZKJ F
9 10 11 12 13
14 15 16 17 18 19 20
21 22 23
4
Digit No.
of code
YY
AF
BG
CE
CG
CH
EJ
FK
HM
JN
KP
MN
MR
note 4
YY
LV
MV
ZZ
A
B
C
D
note 6
note 7
Y
A
B
C
note 8a)
note 8a)b)
B
E
F
note 8b)
note 9
G
Z
Note 1 a) When “B” is specified at the 6th digit, O2 sensor signal has to be set as 0-1V DC linear corresponding to full scale.
b) External Zirconia O2 sensor and external O2 analyzer are not included in the scope of supply, and has to be separately
ordered.
Note 2 a) When “H”, “L” and “M" with purging are specified at 5th digit, only one set of gas inlet/outlet is provided.
In this case, NO2/NO converter cannot be mounted in between of two measuring cells.
b) Resin coupling with purging cannot be manufactued.
Note 3 Rated voltage and plug type of the attached power cable is different depending on the code “J”, “E” and “U” in the 10th
digit.
Select appropriate code according to operationg power supply voltage in the final destination.
Note 4 Measuring range can be selected within either the minimum or maximum range.
(Within min. or max. range, you can change the actual range settings locally.)
Initial setting from Fuji factory is Min. range for the 1st range, and Max. range for the 2nd range.
When range preset at Fuji factory is required, please select “specified range” at 23rd digit and inform Fuji of specified
range table.
Refer to Table 1, for possible combination of measuring components and ranges in the data sheet.
Note 5 “1E” can be specified at 11th & 12th digits, ONLY for CO2 measurement.
In this case, be sure to select “with purging” at 7th digit.
Note 6 O2 conversion is calculated only for NO, SO2 and CO.
Both average value output after O2 calculation and O2 average value output are provided at the same time.
a) Peak count alarm can be added only for CO measurement.
Note 7 When “Y” is specified at the 6th digit, 22nd digit always has to be specified as “Y”.
Note 8 a) If you would like Fuji to deliver ZKJ analyzer with specific range setting, select “specified range” and separately inform
Fuji of the actual range of each component together with your purchase order.
b) In case that the measurement unit is specified as “mg/m3”, it is necessary to select “unit : mg/m3” (Code “F” or “G”)
at the 23rd digit.
Please refer to the table shown below for the corresponding range code based on “mg/m3”.
Corresponding range in mg/m3 or g/m3
In ppm
Range code Min. range Max. range
AF
0-50ppm 0-1000ppm
BG
0-100ppm 0-2000ppm
CH
0-200ppm 0-5000ppm
NO
Min. range Max. range
0-70mg/m3 0-1300mg/m3
0-140mg/m3 0-2600mg/m3
0-280mg/m3 0-6600mg/m3
SO2
Min. range Max. range
0-150mg/m3 0-2800mg/m3
0-300mg/m3 0-5500mg/m3
0-600mg/m3 0-14g/m3
CO
Min. range Max. range
0-65mg/m3 0-1250mg/m3
0-130mg/m3 0-2500mg/m3
0-250mg/m3 0-6250mg/m3
Note 9 When “Z” is specified at the 23rd digit, a gas composition table of actual measued gas has to be sent to Fuji
together with your purchase order.
Note 10 When only N2O analyzer is used, make sure not to contain any components other than N2O .
Multicomponent analyzers including N2O analyzer + CO2 analyzer are used for sludge incineration.
In this case, the range of N2O is 0 to 200ppm/500ppm, and the range of CO2 is 0 to 10%/20%.
9
ZKJ
Table 1. Measurable component and range – availability check table –
(1) Components of single-component analyzer and double-component analyzer (NO/CO), and
CO of three-component analyzer (NO/SO2/CO)
As shown in the range code, when “P”, “A”, “D”, “B”, and “E” are specified at 5th digit,
each component is given at 11th and 12th digits. When “H” is specified, NO is given at
11th and 12th digits and CO at 13th and 14th digits. When “L” is specified, CO is given at
15th and16th digits.
Range code
Code symbol
5th digit
Components
P,H
NO
A
SO2
D
B,H,L
CO2
CO
E
CH4
Range code
Code symbol
5th digit
Components
P,H
NO
A
SO2
D
CO2
B,H,L
CO
E
CH4
Q
N2O
1E
AF
BG
CH
EJ
FK
Range
0-20ppm
0-50ppm
0-100ppm 0-200ppm 0-500ppm 0-1000ppm
0-500ppm 0-1000ppm 0-2000ppm 0-5000ppm
0-1%
0-2%
Range
0-2000ppm 0-5000ppm 0-1%
0-2%
0-2%
0-10%
0-200ppm
0-5%
0-10%
0-20%
0-10%
0-50%
0-100% 0-2000ppm
GL
HM
JN
KM
KP
MR
CG
: Measurable
(2) NO/SO2 of double-component analyzer (NO/SO2), three-component analyzer (NO/SO2/CO)
and four-component analyzer (NO/SO2/CO2/CO)
Selection of NO/SO2 when “F”, “L”, and “M” are specified at 5th digit of the code symbol.
Measurable components
Code symbol, 13th, and
14th digits.
Measurable
components
Code symbol, 11th,
and 12th digits.
2nd
1st
1st component, AF
0-50ppm
0-1000ppm
NO
BG
0-100ppm
0-2000ppm
CH
0-200ppm
0-5000ppm
2nd component SO2
AF
CH
0-50ppm
0-100ppm 0-200ppm
0-1000ppm 0-2000ppm 0-5000ppm
: Combination is available.
10
BG
(3) CO2/CO of double-component analyzer (CO2/CO) and 4-component analyzer (NO/SO2/CO2/CO)
When “G” is specified at 5th digit, CO2 is given at 11th and 12th digits, and CO at 13th and 14th digits.
When “M” is specified, CO2 is given at 15th and 16th digits, and CO at 17th and 17th digits.
Measurable
components
Measurable
components
Range code
Range
2nd
code 1st
1st component,
AF
0-50ppm
0-1000ppm
CO2
BG
0-100ppm
0-2000ppm
CH
0-200ppm
0-5000ppm
EJ
0-500ppm
0-1%
FK
0-1000ppm
0-2%
GL
0-2000ppm
0-5%
HM
0-5000ppm
0-10%
JN
0-1%
0-20%
KP
0-2%
0-50%
MP
0-10%
0-100%
MN
0-10%
0-20%
2nd component, CO
AF
BG
CH
EJ
FK
HM
JN
KP
MR
0-2%
0-10%
0-50ppm
0-100ppm 0-200ppm 0-500ppm 0-1000ppm 0-5000ppm 0-1%
0-10%
0-20%
0-50%
0-100%
0-1000ppm 0-2000ppm 0-5000ppm
0-1%
0-2%
: Combination is available.
(4) N2O/CO2 of 2-component analyzer N2O/CO2, 3-component analyzer NO/N2O/CO2, SO2/N2O/CO2,
N2O/CO2/CO, CH4/N2O/CO2 and 4-component analyzer (NO/SO2/N2O/CO2)
Range code: When code symbol is "R" or "U", N2O is 11th and 12th digit, CO2 is 13th and 14th digit.
When code symbol is "S", "T" or "V", N2O is 13th and 14th digit, CO2 is 15th and 16th digit
When code symbol is "W", N2O is 15th and 16th digit, CO2 is 17th and 18th digit
Measurable
components
2nd
component, CO2
Range code
Measurable
components
1st component,
N 2O
Range
code
CE
MN
0-10%
0-20%
0-200ppm
0-500ppm
: Combination is available.
(5) CO2 range selection of 3-component analyzer (N2O/CO2/CO)
Range code: N2O is 11th and 12th digit, CO2 is 13th and 14th, CO is 15th and 16th digit.
The range code of CO2 is "MN".
Measurable
components
Range code
Measurable
components
1st component,
N 2O
Range
code
CE
2nd
component, CO
BF
0-100ppm
0-1000ppm
0-200ppm
0-500ppm
: Combination is available.
11
ZKJ
(6) SO2 range selection of 3-component analyzer (SO2/N2O/CO2)
Range code: SO2 is 11th and 12th digit, N2O is 13th and 14th, CO2 is 15th and 16th digit.
The range code of CO2 is "MN".
Measurable
components
Range code
Measurable
components
1st component,
SO2
Range
code
AF
2nd
component, N2O
CE
0-200ppm
0-500ppm
0-50ppm
0-1000ppm
: Combination is available.
(7) CH4 range selection of 3-component analyzer (CH4/N2O/CO2)
Range code: CH4 is 11th and 12th digit, N2O is 13th and 14th, CO2 is 15th and 16th digit.
The range code of CO2 is "MN".
Measurable
components
Range code
Measurable
components
1st component,
CH4
Range
code
CH
2nd
component, N2O
CE
0-200ppm
0-500ppm
0-200ppm
0-5000ppm
: Combination is available.
(8) NO/SO2/N2O and 4-component analyzer (NO/SO2/N2O/CO2)
Range code: NO is 11th and 12th digit, SO2 is 13th and 14th, N2O is 15th and 16th , CO2 is 17th and 18th digit.
The range code of CO2 is "MN".
Measurable
components
Range code
Measurable
components
1st component,
NO
Range
code
AF
0-50ppm
0-1000ppm
BG
0-100ppm
0-2000ppm
2nd
component, SO2
AF
3nd
component, N2O
CE
0-50ppm
0-1000ppm
0-200ppm
0-500ppm
: Combination is available.
12
OUTLINE DIAGRAM (Unit: mm)
<Analyzer main unit>
<Side>
<Upper>
12.7
(22)
123.8
288.9
466.7
578 or less
315.8
M4(for mounting slide rail)
26.1
38
<Front>
Power switch
Sample gas outlet Rc1/4 or NPT1/4
INFRARED GAS ANALYZER
Sample gas inlet Rc1/4 or NPT1/4
(Purge gas inlet Rc1/4 or NPT1/4)
POWER
MODE
ZERO
ESC
ENT
SPAN
465
483
<Rear>
<Rack mounting hole>
429
Connector for input/output
terminal module
177
169
100.8
7.5
M5
JIS : 100mm
EIA : 101.6mm
Power inlet (100V to 240V AC, 50Hz/60Hz)
13
ZKJ
<Input / output terminal module>
Mounting hole
6-ø4.5hole
(Accessory)
316
8
5
150
154
164
(140)
12
150
Terminal Terminal Terminal
block
block
block
<TN1> <TN2> <TN3>
Terminal
block
<TN4>
Terminal
block
<TN5>
29
(54.5)
Screw terminals M3.5
8
(280)
Input / output terminal
module connector
<Cable for connecting input / output terminal>
(Accessory)
100030
<Dimensions for mounting input / output terminal module>
150
150
142
154
M4 screw
302
Cut M4 screw holes at 6 positions.
Drill a rectangular hole of 302 142mm or more in the center.
14
Connector
<CN3>
OUTLINE DIAGRAM OF ACCESSORY SLIDE RAIL (Unit: mm)
Model : 305A-24/Accuride International Inc. compatible.
*The slide rails are attached to this equipment when designated.
Closed
6323
Cabinet member
1.52t
The same or less
(22.7)
Intermediate
member
Open
(9.5)
Drawer member
1.27t
609.60.8
577.80.3
565.10.3
4540.3
15.90.5
111.10.3
12.70.3
(24.4)
35.30.3
25.40.3
12.70.3
4.5 × 5.3
4.5 × 5.3
123.80.3
288.90.3
466.70.3
15.90.5
19.10.3
606.60.8
Reinforcement plate
19-inch rack mounting method:
The mass of the instrument should be supported at the bottom of the unit (or the side of the unit when mounted with
the slide rails).
Also, for facilitate maintenance, a structure which allows extraction of the main unit by using the slide rail is recommended.
Rack size
Rack size
101.6
Guide rail mounted type
101.6
Slide rail mounted type
450 or more
450 or more
465
465
Mounting diagram
Slide rails
Mounting diagram
Guide rails
Guide rails
For the guide rail mounted type, a maintenance
space (200mm or more) should be provided upper
the main unit.
15
ZKJ
EXTERNAL CONNECTION DIAGRAM
Terminal block 1
<TN1>
Ch5 output –
(Ch5_OUT) +
1
11
2
12
Ch4 output –
(Ch4_OUT) +
3
13
4
14
Ch3 output –
(Ch3_OUT) +
5
15
6
16
Ch2 output –
(Ch2_OUT) +
7
17
8
18
Ch1 output –
(Ch1_OUT) +
9
19
10
20
Terminal block 2
<TN2>
– Ch10 output
+ (Ch10_OUT)
O2 sensor input* –
(O2_IN) +
1
11
2
12
– Ch9 output
+ (Ch9_OUT)
Ch12 output
(Ch12_OUT)
3
13
4
14
– Ch8 output
+ (Ch8_OUT)
Ch11 output
(Ch11_OUT)
5
15
6
16
– Ch7 output
+ (Ch7_OUT)
Unassigned
7
17
8
18
– Ch6 output
+ (Ch6_OUT)
Unassigned
9
19
10
20
(M3.5 screw)
1
11
2
12
3
13
4
14
Remote hold input
(R_HOLD)
5
15
6
16
Average value reset
input (RESET)
7
17
8
18
9
19
10
20
Unassigned
Auto calibration
remote start
input (R_CAL)
Unassigned
Alarm3 output
(ALM_3)
Alarm3 output
(ALM_2)
Alarm1 output
(ALM_1)
Ch3 remote range
Changeover input
(R_RNG_Ch3)
Ch2 remote range
Changeover input
(R_RNG_Ch2)
Ch1 remote range
Changeover input
(R_RNG_Ch1)
Terminal block 4
<TN4>
Ch5 range identification
signal output (RNG_IDCh5)
Peak count alarm
output (PEAK_ALM)
1
11
2
12
Ch4 range identification
signal output (RNG_IDCh4)
Auto calibration status
contact output
(ACAL/MNT)
3
13
4
14
Ch3 range identification
signal output (RNG_IDCh3)
Pump ON/OFF
(PUMP)
5
15
6
16
Ch2 range identification
signal output (RNG_IDCh2)
Calibration error
contact output
(CAL_ALM)
7
17
8
18
Ch1 range identification
signal output (RNG_IDCh1)
Instrument error
contact output
(FAULT)
9
19
10
20
Unassigned
Unassigned
Unassigned
Unassigned
Unassigned
(M3.5 screw)
(M3.5 screw)
Terminal block 5
<TN5>
Connector
<CN3>
Solenoid valve drive signal output for calibration
1
11
2
12
3
13
4
14
5
15
6
16
7
17
8
18
9
19
10
20
Unassigned
Alarm6 output
(ALM_6)
Alarm5 output
(ALM_5)
Contact output for
sample gas selecting 1
Contact output for Ch3
6 span calibration (DO)
Contact output for Ch4
7 span calibration (DO)
Contact output for Ch5
8 span calibration (DO)
2
Contact output for
zero calibration (DO) 3
Contact output for Ch1 4
span calibration (DO)
Contact output for Ch2 5
span calibration (DO)
Alarm4 output
(ALM_4)
(M3.5 screw)
Note 1) Unassigned terminals are used for internal connection.
So they should not be used as repeating terminals either.
Note 2) The allocation of eaCh Channel (Ch1 to Ch12) depends on
measured gas components. Refer to the table on the next page.
9 Power for relay drive
5V DC
<D-sub 9P>
Connector
<CN2>
For serial communicetion
1
2
3
4
5
16
Ch4 remote range
Changeover input
(R_RNG_Ch4)
(M3.5 screw)
* : For external O2 sensor input.
Terminal block 3
<TN3>
Unassigned
Ch5 remote range
Changeover input
(R_RNG_Ch5)
6
7
8
9
Dedicated relay board
This relay board receives signals from connector CN3 of
the ZKJ I/O terminal module and activates the calibration
solenoid valve directly.
•Relay contact : 1 normally closed contact
Contact capacity; 250V/2A AC
(resistive load)
OUTLINE DIAGRAM (Unit: mm)
15 or less
Mounting hole
4.5
75
65
CN13
CN9
CN1 CN2 CN3 CN4 CN5 CN6 CN7 CN8
140
150
CONNECTIONS
CN9
CN1
CN2
CN3
CN4
CN5
CN6
CN7
CN8
For zero gas
CH1
CH2
CH3
CH4
CH5
Reserved
Solenoid valve
driving power
Solenoid valve
For switching
sample
For span gas
CN13
I/O terminal module
Relay board
CN3
Dedicated cable (D-sub 9-pin straight cable 1.5 m)
Recommended connector
Contact action
•CN1 to CN9 : Housing; VHR-2N
(Nihon Solderless Terminal)
Contact ; SVH-21T-1.1
(Nihon Solderless Terminal)
•During measurement :CN1 ;ON
Others;OFF
•During calibration
:CN1 ;OFF
Others;Contact corresponding
to calibration timing is
ON
17
ZKJ
Table 2. Correspondence between measurement channels and measured value
The following table gives measurement channels and their contents according to the code symbols.
Code symbol
5th digit 6th digit 22nd digit
Y
Y
P
Y
Y
A
Y
Y
D
Y
Y
B
Y
Y
E
Y
Y
F
Y
Y
H
Y
Y
G
Y
Y
L
Y
Y
M
A, B, C A
P
A, B, C A
A
A, B, C A
B
A, B, C A
F
18
H
A, B, C
A
G
A, B, C
A
L
A, B, C
A
M
A, B, C
A
B
H
G
L
M
B
H
A, B, C
A, B, C
A, B, C
A, B, C
A, B, C
A, B, C
A, B, C
B
B
B
B
B
C
C
G
A, B, C
C
L
A, B, C
C
M
A, B, C
C
Q
R
S
T
U
V
W
S
T
U
V
W
S
Y
Y
Y
Y
Y
Y
Y
A, B, C
A, B, C
A, B, C
A, B, C
A, B, C
A, B, C
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y, B
Y
Y
A
T
A, B, C
A
U
A, B, C
A, C
V
W
A, B, C
A, B, C
A
A
Contents
Ch1: NO
Ch1: SO2
Ch1: CO2
Ch1: CO
Ch1: CH4
Ch1: NO, Ch2: SO2
Ch1: NO, Ch2: CO
Ch1: CO2, Ch2: CO
Ch1: NO, Ch2: SO2, Ch3: CO
Ch1: NO, Ch2: SO2, Ch3: CO2, Ch4: CO
Ch1: NOX, Ch2: O2, Ch3: Converted NOX, Ch4: Converted NOX average, Ch5: O2 average
Ch1: SO2, Ch2: O2, Ch3: Converted SO2, Ch4: Converted SO2 average, Ch5: O2 average
Ch1: CO, Ch2: O2, Ch3: Converted CO, Ch4: Converted CO average, Ch5: O2 average
Ch1: NOX, Ch2: SO2, Ch3: O2, Ch4: Converted NOX, Ch5: Converted SO2,
Ch6: Converted NOX average, Ch7: Converted SO2 average, Ch8: O2 average
Ch1: NOX, Ch2: CO, Ch3: O2, Ch4: Converted NOX, Ch5: Converted CO,
Ch6: Converted NOX average, Ch7: Converted CO average, Ch8: O2 average
Ch1: CO2, Ch2: CO, Ch3: O2, Ch4: Converted CO, Ch5: Converted CO average,
Ch6: O2 average
Ch1: NOX, Ch2: SO2, Ch3: CO, Ch4: O2, Ch5: Converted NOX, Ch6: Converted SO2,
Ch7: Converted CO, Ch8: Converted NOX average, Ch9: Converted SO2 average,
Ch10: Converted CO average, Ch11: O2 average
Ch1: NOX, Ch2: SO2, Ch3: CO2, Ch4: CO, Ch5: O2, Ch6: Converted NOX,
Ch7: Converted SO2, Ch8: Converted CO, Ch9: Converted NOX average,
Ch10: Converted SO2 average, Ch11: Converted CO average, Ch12: O2 average
Ch1: CO, Ch2: O2
Ch1: NO, Ch2: CO, Ch3: O2
Ch1: CO2, Ch2: CO, Ch3: O2
Ch1: NO, Ch2: SO2, Ch3: CO, Ch4: O2
Ch1: NO, Ch2: SO2, Ch3: CO2, Ch4: CO, Ch5: O2
Ch1: CO, Ch2: O2, Ch3: Converted CO, Ch4: Converted CO average, Ch5: O2 average
Ch1: NOX, Ch2: CO, Ch3: O2, Ch4: Converted NOX, Ch5: Converted CO,
Ch6: Converted NOX average, Ch7: Converted CO average, Ch8: O2 average
Ch1: CO2, Ch2: CO, Ch3: O2, Ch4: Converted CO, Ch5: Converted CO average,
Ch6: O2 average
Ch1: NOX, Ch2: SO2, Ch3: CO, Ch4: O2, Ch5: Converted NOX, Ch6: Converted SO2,
Ch7: Converted CO, Ch8: Converted NOX average, Ch9: Converted SO2 average,
Ch10: Converted CO average, Ch11: O2 average
Ch1: NOX, Ch2: SO2, Ch3: CO2, Ch4: CO, Ch5: O2, Ch6: Converted NOX,
Ch7: Converted SO2, Ch8: Converted CO, Ch9: Converted NOX average,
Ch10: Converted SO2 average, Ch11: Converted CO average, Ch12: O2 average
Ch1:N2O
Ch1:N2O, Ch2:CO2
Ch1:NO, Ch2:N2O, Ch3:CO2
Ch1:SO2, Ch2:N2O, Ch3:CO2
Ch1:N2O, Ch2:CO2, Ch3:CO
Ch1:CH4, Ch2:N2O, Ch3:CO2
Ch1:NO, Ch2:SO2, Ch3:N2O, Ch4:CO2
Ch1:NO, Ch2:N2O, Ch3:CO2,Ch4:O2
Ch1:SO2, Ch2:N2O, Ch3:CO2,Ch4:O2
Ch1:N2O, Ch2:CO2, Ch3:CO, Ch4:O2
Ch1:CH4, Ch2:N2O, Ch3:CO2,Ch4:O2
Ch1:NO, Ch2:SO2, Ch3:N2O, Ch4:CO2, Ch5:O2
Ch1:NOx, Ch2:N2O, Ch3:CO2, Ch4:O2, Ch5:Converted NOx, Ch6:Converted NOx average,
Ch7:O2 average
Ch1:SO2, Ch2:N2O, Ch3:CO2,Ch4:O2, Ch5:Converted SO2, Ch6:Converted SO2 average,
Ch7:O2 average
Ch1:N2O, Ch2:CO2, Ch3:CO, Ch4:O2, Ch5:Converted CO, Ch6:Converted CO average,
Ch7:O2 average
Ch1:CH4, Ch2:N2O, Ch3:CO2,Ch4:O2,Ch5:O2 average
Ch1:NOx, Ch2:SO2, Ch3:N2O, Ch4:CO2, Ch5:O2, Ch6:Converted NOx, Ch7:Converted SO2,
Ch8:Converted NOx average, Ch9:Converted SO2 average, Ch10:O2 average
note
Exclusive Zirconia O2 Sensor
(to be purchased separately)
For O 2 correction, the gas analyzer ZKJ can accept
linealized 0 to 1V DC signal coming from analyzer calibrated
0 to 25% O2 full scale. If the analyzer is not available, Fuji
can supply exclusive Zirconia O2 sensor Model ZFK.
Measuring method:
Zirconia system
Measurable component and measuring range:
Measurable component
O 2
Oxygen
Range
0 to 25vol%
Repeatability:
Within ± 0.5% of full scale
Linearity:
Within ± 1% of full scale
Zero drift:
Within ± 1% of full scale/week
Span drift:
Within ± 2% of full scale/week
Response time: Approx. 20 seconds (for 90% response)
Measured gas flow rate:
0.5 ± 0.25L / min
Remark: T h e Z i r c o n i a s y s t e m , d u e t o i t s
principle, may produce a measuring
error due to relative concentration
versus the com-bustible O2 gas
concentration. Also, a corrosive gas
(SO 2 of 250 ppm or more, etc.) may
affect the life of the sensor.
Gas inlet/outlet size:
Rc1/4 or NPT1/4
Power supply:
Rated voltage
;100 to 115V AC or
200 to 240V AC
Rated frequency ;50Hz/60Hz
Max. rated power;215VA (during power
ON)
65VA (during steadystate operation)
Enclosure:
Steel casing, for indoor application
Indication:
Temperature indication (LED)
Temperature alarm output:
Contact output 1a contact,
Contact capacity 220V, 1A AC (resistive
load)
Outer dimensions (H x W x D):
141 x 170 x 190mm
Mass {weight}: Approx. 3kg
Finish color:
Munsell 5Y 7/1
CODE SYMBOLS
1 2 3 4 5 6 7 8
Z F K 7 Y Y
9 10 11 12 13
4 -
Description
Y 0 Y Y
Measuring method
Zirconia method
7YY
Power supply
100 to 115V AC 50/60Hz(Standard)
200 to 240V AC 50/60Hz(Standard)
200 to 240V AC 50/60Hz(CE mark)
9
B
C
1
8
Gas inlet/outlet size
Rc 1/4
NPT 1/4
SCOPE OF DELIVERY
ORDERING INFORMATION
•Gas analyzer ... 1 unit
•Input /output terminal module for external mounting
... 1 set
•Connection cable (1m) between main unit and input /
output terminal module ... 1 pc
•Power cable (standard inlet type 2m) ... 1 pc
•Spare fuses (250V, 3.15A AC, delay type) ... 2 pcs
•Instruction manual ... 1 copy
•Slide rails ... 2 pcs (when with slide rails are selected)
•Relay board ... 1 pc (when with relay board are selected)
•Ceble foe relay board ... 1 pc (when with relay board are
selected)
1. Code symbols
2. Application and composition of sample gas
Items to be prepared separately
•Various sampling devices (refer to Data Sheets for the
sampling devices)
•Dedicated zirconia O2 sensor (see Page 16)
19
ZKJ
OUTLINE DIAGRAM
(Unit:mm)
17
170
172.5
4
152
OXYGEN ANALYZER
US
2
1
AC250V
T3.15A
3
4
5
6
7
E
F
110
131
141
TEMP.
L
AC
INPUT
N
OUT
ALM
9TH DIGIT
1:RC1/4
8:NPT1/4
48
37
23
ø5
INLET
OUTLET
EXTERNAL CONNECTION DIAGRAM
1
2
3
4
5
+
AC power supply
E
6
7
Output
to ZKJ
Temperature
alarm output
Caution on Safety
*Before using this product, be sure to read its instruction manual in advance.
International Sales Div
Sales Group
Gate City Ohsaki, East Tower, 11-2, Osaki 1-chome,
Shinagawa-ku, Tokyo 141-0032, Japan
http://www.fujielectric.com
Phone: 81-3-5435-7280, 7281 Fax: 81-3-5435-7425
http://www.fjielectric.com/products/instruments/
Information in this catalog is subject to change without notice.
Printed in Japan