FUJI EDS3-61H

THERMAL CONDUCTIVITY GAS ANALYZER
THERMOMAT
DATA SHEET
ZAF
In modern industries, it is indispensable to analyze
components and measure concentrations of mixture gases
and combustion gases for strict quality control and to file data
for establishing production schedules which will allow proper
management of industrial processes.
Of the various analytical methods and measuring principles available for analyzing process gases, this gas analyzer
utilizes the fact that electric resistance of a heated platinum
wire changes depending on kind and concentration of gas.
Kind of Gas
Relative thermal conductivity
(at 0°C)expressed taking that
of air(5.572 × 10-5cal/cm.sec.
deg.)as 1
Sulfurdioxide SO2
Carbondioxide CO2
Argon Ar
Carbon monoxide CO
Water vapor(100°C)H2O
Air
Nitrogen N2
Oxygen O2
Methane CH4
Hydrogen H2
Measuring mode:
1 range or 2 ranges (for measuring the same
sample gas components with the same
FEATURES
measuring system)
Measuring range:
CO2
1. Continuous measurement possible.
2. Quick response.
3. Integral unit incorporating detector, current regulator and
indicator.
4. Simple and robust mechanism assuring high maintainability.
5. Minimum influence due to external and operating conditions.
6. Stable accuracy.
H2
0 to 10 ....100
100 to 90 .... 0
0 to 3 ....100
100 to 95 ....
Ar
0 to 10 ....100
100 to 90 .... 0
CH4
He
0 to 20 ....100
100 to 80 .... 0
0 to 5 ....100
100 to 90 ....
Output signal: DC 4 to 20 mA
(load resistance less
SPECIFICATIONS
Measuring principle:
Measurement of thermal conductivity
Measuring component:
Two-component gases (from viewpoint of
thermal conductivity)
such as CO2, H2, Ar, CH4, He, SF6
than 400Ω)
[
0
0
}
DC 0 to 10mV
(output resistance 100Ω)
Simultaneously
available
0 to 100µA (load resistance 100Ω) is
further available for special specification.
]
Repeatability:
Drift:
± 1% of F.S.
Within zero point ± 2.5% of full scale/week
Within span ± 2% of full scale/week
EDS3-61h
Date
Jun. 27, 2002
ZAF
Response speed:
Within 30 sec for 90% response (for 20%
CO2)
Within 60 sec for 90% response (for 0 to
10% H2)
(varies with kind of gas and range)
Sample gas flow rate:
0.4 ± 0.05 /min, constant
Standard requirements for sample gas:
Temperature 0 to 50°C
Dust
0.3µm or less
Pressure
10kPa or less
(flow rate 0.4L/min)
Materials of gas-contacting parts:
Lead, stainless steel JIS SUS316, platinum,
platinum iridium, epoxy resin, silver,
fluororubber
Power supply: AC 100V ± 10%, 50/60 Hz
110V, 200V, or 220V usable for special
specification
Power consumption:
Approx. 10VA
Mounting:
Indoor panel plug-in
CODE SYMBOLS
Z A F
P
Description
2 Reference
Sample
Measuring
gas
gas
range
3A
N2
CO2
0 to 20%
4A
N2
CO2
Others than
4B
H2
0 to20%
4C
Ar
4D
CH4
As specified
4E
He
by customer
4F
SF6
(See the table
4Z
Others
below)
6A
Others than
CO2
6B
N2 gas
H2
As specified
6C
Ar
by customer
6D
CH4
(See the table
6E
He
below)
6F
SF6
6Z
Others
Power supply
α
∠α = 90°
Warmup time: At least 30 min.
Ambient temperature:
-5 to 45°C
Ambient humidity:
Less than 90% RH
Weight:
Approx. 5kg
External dimensions (H x W x D):
240x197x 212 mm
Finish color:
Munsell 7.5BG 3.2/0.8
N2, O2,
Air, He
Ar
CH4
He
AC 220V 50/60Hz
1
1 range
2
2 ranges (available only for measuring
the same sample gas)
MEASURING PRINCIPLE
Range ratio
0-10, 20, 50, 100%
100-90%, 100-80% 1 : max. 5
Output
characteristic
0-10, 20% : Linear
Others : non-linear
Impossible
N2, CO2,
Ar
0-3, 5, 10, 20, 50,
100-90% : Linear
80, 100% 1 : max. 10 Others : non-linear
100-90%, 100-80%
Air, O2
Impossible
N2, O2,
Air, He
0-10, 20, 50, 80,
100% 1 : max. 5
100-90%, 100-80%
CO2
Impossible
N2, CO2,
Ar, He
0-20, 40, 50, 60,
80, 100% 1 : max. 5
100-80%
Air, O2
Impossible
N2, CO2,
Ar, O2,
Air
0-5, 10, 20, 30, 40,
50, 80, 100% 1 : max. 10 100-90% : Linear
Others : non-linear
100-90%, 100-80%
0-10, 20%,
100-90% : Linear
Others : non-linear
Non-linear
(Note 1) Linear output is ±2.5% FS or less.
(Note 2) On 2-range type, the zero point is used in common.
(Note 3) When sample gas consists of more than 3 components, contact
our office with a table of gas components submitted for confirmation of whether or not applicable analyzers are available.
2
AC 200V 50/60Hz
4
(Note 1) Reference gas components are as listed in the table below.
(Note 2) Sample gas, herby, means a gas components to be measured,
and reference gas stands for each component other than the
sample gas (remaining gas component).
Ar
H2
AC 110V 50/60Hz
3
Sample gas
Enclosed
gas inlet
DC Constant current
CO2
Measuring range
AC 100V 50/60Hz
2
Measuring
STANDARD MEASURING
CompariMeasur- son gas
ing gas component
1
Measuring
chamber
Reference
chamber
V-I
converter
V/I
DC 4 to 20mA
Fixed
resistor
DC
amplifier
Fixed
resistor
DC 0 to 10mV
or 0 to 100 A
GAS SAMPLING SYSTEM DIAGRAM (EXAMPLE)
Flow
checker
Pump
Membrane
filter
ZAF
Atmosphere
Note 1
0.4L/min
Note 2
, Standard gas for span
Peltier gas
cooler
, Standard gas for zero
Note 2
Drain pot
(Note 1) Dust must be purged adequately (for protection of the pump, flowmeter, etc.).
(Note 2) Unnecessary when sample gas is dry.
CONNECTION DIAGRAM
SCOPE OF DELIVERY
Analyzer main unit
Panel mounting brackets
Fuse 2 pieces
1
2
3
POWER
AC 100V 50/60Hz
OR
AC 200, 110, 220V
4
5
6
7
8
9
10
G
4 to 20mA
0 to 10mA
OR
0 to 100 A
Items to be prepared separately
Sampling equipment
Receiving instrument
3
ZAF
OUTLINE DIAGRAM (Unit:mm)
2
Panel
t 10
Indicator
140
t
GAS
INLET
THERMOMAT
8
ZERO
SPAN
240
RANGE
26
268
230
POWER
GAS
OUTLET
169
24.2
188
192
197
Terminal
board
Rear View
46
33.5
19
Terminal
cover
Panel Cutout Dimensions
184 +20
232 +20
130
INLET
OUTLET
Specification
card
Tapping Rc1/4
182
Asterisked * items: Non-standard
Caution on Safety
*Before using this product, be sure to read its instruction manual in advance.
Head Office
6-17, Sanbancho, Chiyoda-ku, Tokyo 102-0075, Japan
http://www.fesys.co.jp/eng
Sales Div.
International Sales Dept.
No.1, Fuji-machi, Hino-city, Tokyo, 191-8502 Japan
Phone: 81-42-585-6201, 6202 Fax: 81-42-585-6187
http: //www.fic-net.jp/eng
Information in this catalog is subject to change without notice.
Printed in Japan