INTERSEMA MS5201-XD

MS5201-XD (RoHS*)
•
•
•
•
•
•
SMD PRESSURE SENSOR
1 bar gage pressure range
Uncompensated
Piezoresistive silicon micromachined sensor
Miniature surface mount 7.6 x 7.6 mm
Low noise, high sensitivity, high linearity
RoHS-compatible & Pb-free*
DESCRIPTION
The MS5201-XD is a gage type SMD pressure sensor, suitable for applications which require relative (gage)
pressure measurements. The device consists of a silicon micromachined pressure sensor die mounted on a 7.6 x
7.6 mm ceramic carrier protected by a plastic cap with a port to connect plastic tube. The MS5201-XD can be
delivered in a highly sensitive version giving a maximal output voltage or in a highly linear version giving a linear
output voltage directly proportional to the applied pressure.
High Sensitivity Version
Full Scale
Pressure
1 bar
High Linearity Version
Product Code
Full Scale Span / Linearity
Product Code
Full Scale Span / Linearity
MS5201-AD
240 mV / ±0.2% FS
MS5201-BD
150 mV / ±0.05% FS
FEATURES
•
•
Low cost SMD ceramic package
High reliability, low drift
•
•
-40°C to +125 °C operation range
Gel protection
•
Pneumatic controls
APPLICATION
•
Medical instrumentation
PIN CONFIGURATION
3: VS+
4: OUT+
1: OUT-
2: GND
Bottom
Top
(Pressure to be applied from the top)
*
The European RoHS directive 2002/95/EC (Restriction of the use of certain Hazardous Substances in electrical and electronic equipment)
bans the use of lead, mercury, cadmium, hexavalent chromium and polybrominated biphenyls (PBB) or polybrominated diphenyl ethers
(PBDE).
DA5201-D_004
00005201637 ECN 780
June 29th, 2005
1
PIN DESCRIPTION
Pin Name
Pin No.
OUTGND
VS+
OUT+
1
2
3
4
Function
Negative output voltage of Wheatstone bridge
Ground
Supply voltage of Wheatstone bridge
Positive output voltage of Wheatstone bridge
ABSOLUTE MAXIMUM RATINGS
Parameter
Supply voltage
Storage temperature
Overpressure
MS5201-AD
MS5201-BD
DA5201-D_004
00005201637 ECN 780
Symbol
Conditions
Min
Max
-40
-
20
+125
5
10
o
VS+
TS
Ta = 25 C
P
Ta = 25 oC
June 29th, 2005
Unit
V
C
o
bar
2
ELECTRICAL CHARACTERISTICS
(VS+ = 5 V; Ta = 25°C)
Parameter
Min
Typ
Max
Unit
0
-
1
bar g
Full-scale span (FS)
190
240
290
mV
Sensitivity
190
240
290
mV/bar
Linearity
-
±0.15
±0.4
% FS
Operating pressure range
0
-
1
bar g
Full-scale span (FS)
120
150
180
mV
Sensitivity
120
150
180
mV/bar
-
±0.05
±0.2
% FS
Operating temperature range
-40
-
125
°C
Zero pressure offset
-40
0
40
mV
Pressure hysteresis
-
±0.05
±0.15
% FS
2, 6
Temperature hysteresis
-
-
0.3
% FS
3, 6
Repeatability
-
±0.05
±0.15
% FS
4, 6
3.0
3.4
3.8
+2’400
-1’500
-80
2’900
-1’900
+3’300
-2’300
+80
kΩ
ppm/°C
ppm/°C
µV/°C
5, 6
5, 6
5, 6
Operating pressure range
MS5201-AD
High Sensitivity
MS5201-BD
High
Linearity
All Types
Linearity
Bridge resistance
Temperature coefficient of resistance
Temperature coefficient of span
Temperature coefficient of offset
Notes
1, 6
1
NOTES
1) Deviation at one half full-scale pressure from the least squares best line fit over pressure range.
2) Maximum difference of output voltage after 1 pressure cycle at any pressure within the operating pressure
range.
3) Maximum difference in offset after one thermal cycle from -40°C to +125°C.
4) Same as 2) after 10 pressure cycles.
5) Slope of the endpoint straight line from 25°C to 60°C.
6) Not 100% tested.
DA5201-D_004
00005201637 ECN 780
June 29th, 2005
3
APPLICATION INFORMATION
GENERAL
The MS5201-XD is a miniaturised absolute pressure sensor series which has been designed for surface mounting
applications. Its main advantages are the high reliability of the semiconductor sensor and a design which makes it
suitable for applications requiring small-scale and cost-efficient solutions.
The sensor element of the MS5201-XD consists of a silicon micromachined membrane with a Pyrex glass and a
backside vent hole. Implanted resistors make use of the piezo-resistive effect. The sensor is mounted in a special
process allowing best-offset stability making the part suitable for direct PCB assembly.
Typical applications for this miniaturised pressure sensor MS5201-XD are blood pressure monitors or vacuum
measurement equipment.
CONNECTION TO PCB
The package outline of the module allows the use of a flexible PCB to connect it. This can be important for certain
applications, and will also reduce mechanical stress on the device.
For applications subjected to mechanical shock, it is recommended to enhance the mechanical reliability of the
solder junctions by covering the rim or the corners of MS5201-XD ceramic substrate with glue or Globtop-like
material.
SOLDERING
Please refer to the application note AN808 for all soldering issues.
CLEANING
The MS5201-XD has been manufactured under cleanroom conditions. Each device has been inspected for the
homogeneity and the cleanness of the silicone gel. It is therefore recommended to assemble the sensor under
class 10’000 or better conditions. Should this not be possible, it is recommended to protect the sensor opening
during assembly from entering particles and dust. To avoid cleaning of the PCB, solder paste of type “no-clean”
shall be used. Cleaning might damage the sensor.
DA5201-D_004
00005201637 ECN 780
June 29th, 2005
4
PACKAGE OUTLINES
Device package outlines of MS5201-XD (plastic cap, gel protection of bonding wires)
DA5201-D_004
00005201637 ECN 780
June 29th, 2005
5
RECOMMENDED PAD LAYOUT
Recommended pad layout for soldering of the MS5201-XD on a printed circuit board
ORDERING INFORMATION
MS5201-AD = Pressure Sensor 1 bar, high sensitivity
MS5201-BD = Pressure Sensor 1 bar, high linearity
Art.-Nr. 325201000
Art.-Nr. 325201001
A = high sensitivity
B = high linearity
FACTORY CONTACTS
Intersema Sensoric SA
Ch. Chapons-des-Prés 11
CH-2022 BEVAIX
SWITZERLAND
Tel. 032 847 9550
Tel. Int. +41 32 847 9550
Telefax +41 32 847 9569
e-mail:
http://www.intersema.ch
NOTICE
Intersema reserves the right to make changes to the products contained in this data sheet in order to improve the design or performance and
to supply the best possible products. Intersema assumes no responsibility for the use of any circuits shown in this data sheet, conveys no
license under any patent or other rights unless otherwise specified in this data sheet, and makes no claim that the circuits are free from
patent infringement. Applications for any devices shown in this data sheet are for illustration only and Intersema makes no claim or warranty
that such applications will be suitable for the use specified without further testing or modification.
DA5201-D_004
00005201637 ECN 780
June 29th, 2005
6