SLPOWER GSM28-12

GSM28 Medical
28 Watt Global Performance Switchers
GLOBAL PERFORMANCE SWITCHERS
FEATURES:
DC Power Supplies
• Industry’s smallest 28 W medically approved switcher
• Compact size (4.00” x 2.59” x 0.92”)
• Wide-range ac input: 85-264 Vac
• Less than 25 μA leakage current @ 120 Vac
• Approved to UL2601-1, EN60601-1
• EMI to FCC, CISPR 11 Class B/IEC601-1-2
• Overvoltage protection standard
• RoHS Compliant Models Available (G suffix)
•
marked to LVD
SPECIFICATIONS
Ac Input
85-264 Vac, 47-63 Hz single phase.
Temperature Coefficient
0.03%/°C typical on all outputs.
Input Current
Maximum input current at 120 Vac, 60 Hz with full rated output load is 0.85 A.
Environmental
Designed for 0 to 50°C operation at full rated output power; derate output current
and total output power by 2.5% per °C above 50°C.
Hold-up Time
15 ms minimum from loss of ac input at full load, nominal line (120 Vac).
Medical EMI/EMC Compliance
All models include built-in EMI filtering to meet the following emissions requirements:
Output Power
Normal continuous output power is 28 W, 32 W peak for 60 s maximum duration, 10% duty cycle. Factory set to begin power limiting at approximately 35
W.
Overload Protection
Fully protected against short circuit and output overload. Short circuit protection is cycling type power limit.
Output Noise
0.5% rms, 1% pk-pk, 20 MHz bandwidth, differential mode. Measured with
scope probe directly across output terminals of the power supply with load
terminated with 0.1 μF capacitor.
Overvoltage Protection
Built in with firing point set per ratings table. OVP firing reduces voltage to
less than 50% of nominal voltage in 50 ms.
Voltage Adjustment
Factory set with fixed resistors to maximize reliability.
Efficiency
70% minimum for the 5.1 V model at full rated load, nominal input voltage.
Efficiency increases as output voltage increases.
Input Protection
Internal ac fuse provided on all units. Designed to open only if a catastrophic
failure occurs in the unit.
EMI SPECIFICATIONS
COMPLIANCE LEVEL
Conducted Emissions
Static Discharge
RF Field Susceptibility
Fast Transients/Bursts
Surge Susceptibility
Line Frequency Harmonics
EN55011 Class B; FCC Class B
EN61000-4-2, 6 kV contact, 8 kV air
EN61000-4-3, 3 V/meter
EN61000-4-4, 2 kV, 5 kHz
EN61000-4-5, 1 kV diff., 2 kV com.
EN61000-3-2 Class A
Earth Leakage Current
Leakage current measured in the Gnd wire connection when measured per
EN60601-1 or UL2601-1 is as follows:
Medical
Model
Normal Leakage
SingleFault Leakage
Test Voltage
GSM28
25μA
45 μA
132 Vca/60 Hz
UL2601-1
GSM28
50 μA
90 μA
264 Vca/50 Hz
IEC60601-1
Medical Medical Safety
SL Power Electronics Corp. declares under our sole responsibility that all GSM
models are in conformity with the applicable requirements of UL2601-1 Patient
Care Equipment, CSA-C22.2 No. 234 (with additional tests to C22.2 No. 601.1 per
T.I.L. CA-08), EN60601-1.
Inrush Current
Inrush limited by internal thermistor. The inrush at 230 Vac, averaged over the
first ac half-cycle under cold start conditions will not exceed 32 A.
Minimum Load Requirement
Test
Method
5% of full load rating
Transient Response 3.5% max. dev. 50% load step at 0.2 A/μs. Recovery to
0.5% within 500 μs.
All specifications are typical at nominal input, full load at 25OC unless otherwise stated
Medical Model
Output
Current
Load Regulation
Initial Setpoint Tolerance
OVP Setpoint
Ripple and Noise
GSM28-5
5.1 V
5.5 A
0.75%
2.5%
6.2 ± 0.6 V
1.4%
GSM28-12
12 V
2.3 A
0.75%
2.5%
14 ± 1.0 V
1%
GSM28-15
15 V
1.9 A
0.75%
2.5%
18.5 ± 1.5 V
1%
GSM28-24
24 V
1.2 A
0.75%
2.5%
28 ± 2.5 V
1%
GSM28-28
28 V
1.0 A
0.75%
2.5%
34 ± 2.8 V
1%
* Add “G” suffix to model number for RoHS compliant model.
GSM28 MECHANICAL SPECIFICATIONS
ENVIRONMENTAL SPECIFICATIONS
OPERATING
NON-OPERATING
Temperature (A)
See individual specs
-40 to +85°C
Humidity (A)
0 to 95% RH
0 to 95% RH
Shock (B)
20 gpk
40 gpk
Altitude
-500 to 10,000 ft
-500 to 40,000 ft
Vibration (C)
1.5 g rms’ 0.003 g2/Hz
5 g rms’ 0.026 g2/Hz
A. Units should be allowed to warm up/operate under non-condensing
conditions before application of power.
B. Shock testing—half-sinusoidal, 10 ± 3 ms duration, ± direction, 3 orthogonal axes, total 6 shocks.
C. , Random vibration—10 to 2000Hz, 6dB/octave roll-off from 350 to
2000Hz 3 orthogonal axes. Tested for 10 min./axis operating and 1 hr./axis
non-operating.
SL Power Electronics Corp., 6050 King Drive, Bldg. A, Ventura, CA 93003, USA. Phone:(805) 486 4565 Fax:(805) 487 8911 www.slpower.com Rev. 1/07.
Data Sheet © 2007 SL Power Electronics Corp. The information and specifications contained in this data sheet are believed to be correct at time of publication.