SLPOWER MW173KB1503F01

MW173KB
Universal 20 – 32 Watt Series
Medical Switch-Mode Power Supply
3 Year Warranty
•100-240VAC Universal Input
•Desktop Style
•5V to 24V Single Output Models, up to 32W
•Modified and Custom Designs
•Regulated Output with Low Ripple
•Impact-Resistant Polycarbonate Enclosure
•No load Power Consumption < 0.50W
•Designed to meet EISA Requirements (see page
3 for details).
Specifications
All Specifications are typical at nominal input, full load at 25°C unless otherwise stated.
Output Specifications
Line and Load
Voltage Regulation
Excluding Cord
General Specifications
Line: +/-1%
Load: +/-5%
Topology
Switching - Fixed Frequency
Flyback
Efficiency
Designed to meet EISA
Requirements - see page 3
Ripple
1% Vp-p max.
Transient Response
0.5mS for 50% Load Change,
typical.
Protection
Overcurrent Protection
(Hiccup).
Short Circuit Protection
Dialectric Withstand
4,000VAC or 5,656VDC
Primary-Secondary;
1,500VAC or 2,150VDC
Primary - F.G;
500VDC Secondary - F.G.
Input Voltage Range Universal Input
100-240VAC, -10%, +10%
Storage Temp.
-30° C to 85° C
Line Frequency
47-63Hz
Approvals and
Safety Standards
UL60601-1
IEC/EN60601-1
EMC: EN60601-1-2/
EN55024
MTBF
100,000 Calculated Hours
Case and
Dimensions
Desktop Style:
3.74"L x 2.13"W x 1.26"H
95mm L x 54mm W x 33mm
H
Case Material
Black 94V0 Polycarbonate
Cord and
Connectors
18AWG, 1,800mm 2
conductor. (5V, 6V model:
1,500mm). Ault #3 connector.
Other connectors are
available.
Hold-up Time
Input Specifications
Input Current
90 VAC
1.0A, max.
Internal Primary Current Fuse,
Inrush Limiting
Protection
Environmental Specifications
Thermal
Performance
Relative Humidity
Altitude
Operating Temperature 0° C to 40° C
full load, no derating,
convection cooling, Nonvented case
Non-condensing
5% to 95%
0 to 10,000 feet
@ 115VAC
18mS, min.
______________________________________________________________________________________________________________________________ 1 of 3
SL Power Electronics Corp • 6050 King Drive Ventura, CA 93003 • Phone 805.486.4565 • Fax 858.712.2040 • Email: [email protected] • www.slpower.com
MW173KB
Universal 20-32 Watt Series
Medical Switch Mode Power Supply
3 Year Warranty
OUTPUT PARAMETERS
Model Number
Volts (V)
Output Current (max)
Max Watts
Ripple (Vp-p max)
MW173KB0503F01
5V
4.00 A
20.0 W
50mV
MW173KB0603F01
7.5 V
3.00 A
22.5 W
75mV
MW173KB0903F01
9V
3.00 A
27.0 W
90mV
MW173KB1203F01
12 V
2.50 A
30.0 W
120mV
MW173KB1503F01
15 V
2.00 A
30.0 W
150mV
MW173KB1803F01
18 V
1.67 A
30.0 W
180mV
MW173KB2403F01
24 V
1.33 A
32.0 W
240mV
Note: Part numbers above include #3 output connector and IEC320 C14 grounded input receptacle. See below for other options.
Model Number Key
MW173 K B 05 03 F 01
Modification Designator. "01" is standared part, "02" and higher denotes a modified
model.
Input Receptacle. See Below for options. "F" receptacle (IEC320, C14 type,
grounded) is standard.
Output Connector. "03" connector is standard. See below. Other output connectors
available. Contact the factory for more details
Output Voltage: 05=5V, 09=9V, 12=12V, 15=15V, 18=18V, 24=24V, 48=48V.
Design Revision Changes
Manufacturing Location
Product Family
AC Input Receptacle Options
Desktop
IEC320 - C14
Class I
Grounded
IEC320 - C18
Class II
(F)
Wall-Plug
N. Am erica
Japan
Ungro unde d
IEC320 - C8
Class II
"Shaver"
(Q)
(N)
(B)
F ixe d
______________________________________________________________________________________________________________________________ 2 of 3
SL Power Electronics Corp • 6050 King Drive Ventura, CA 93003 • Phone 805.486.4565 • Fax 858.712.2040 • Email: [email protected] • www.slpower.com
MW173KB
Universal 20-32 Watt Power Supply
Medical Switch Mode Power Supply
3 Year Warranty
2007 Energy Independence and Security Act - EISA
The Energy Independence and Security Act of 2007 was passed in December of 2007 and addresses minimum efficiency standards
and standby levels for Class A external power supplies that are 250 Watts and under. This law stipulates that external power
supplies manufactured on July 1, 2008 and beyond meet certain minimum efficiency and standby criteria as defined below.
Minimum Efficiency Criteria:
Active mode is defined as when a power supply’s input is connected to a line voltage AC and it’s output is connected to a DC or AC
load, drawing a portion of the product’s power output. Depending upon the power rating for the power supply, it must meet the
minimum efficiency criteria outlined below.
Energy-Efficiency Criteria for Active Mode:
Output Power on Adapter Label
Minimum Average Efficiency Percentage
0 to < 1 Watt
≥ 0.50 * output power on the label
> 1 watt to ≤ 51 watts
≥ [0.09 * Ln (output power on adapter label)] + 0.50
> 51 watts
≥ 0.85
Energy Consumption Criteria for No Load Mode:
The power supply must also meet a requirement for when its input is connected to line voltage AC but its output is not connected to a
load. Depending upon the power output of the supply, it must keep its energy consumption below the following values:
Output Power on Adapter Label
Maximum Power Consumption in No-Load Mode
0 to < 250 Watts
≤ 0.50 watts
Output Connector Information
Notes:
1. Center Contact = Positive
2. Connector is Switchcraft 760 plug or equivalent.
3. Suggested Mating Connector is Switchcraft 712A jack or equivalent.
4. Other output connector options are available. Contact your local representative for
details.
Data Sheet © 2011 SL Power Electronics Corp. The information and specifications contained herein are believed to be correct at the time of publication.
Rev. 2-13-2011
However, SL Power accepts no responsibility for consequences arising from reproduction errors or inaccuracies. Specifications are subject to change without notice.
______________________________________________________________________________________________________________________________ 3 of 3
SL Power Electronics Corp • 6050 King Drive Ventura, CA 93003 • Phone 805.486.4565 • Fax 858.712.2040 • Email: [email protected] • www.slpower.com