Platinum Resistance Temperature Detector L 622 DBC L series PRTDs, type L 622 DBC are designed for large volume applications where longterm stability, interchange ability and accuracy over a large temperature range are vital. The high resistance value of 1000 Ohm generates a high signal yield. Typical applications are Automotive, White goods, HVAC, Energy management, Medical and Industrial equipment. Nominal Resistance R0 Tolerance DIN EN 60751 1996-07 Tolerance DIN EN 60751 2009-05 Order Number Plastic Box Lead Length KL 1000 Ohm at 0°C Class B F 0.3 32 200 018 160mm 1000 Ohm at 0°C Class B F 0.3 32 200 023 250mm The measuring point for the nominal resistance is situated on the connections of the sensor body Specification DIN EN 60751 Temperature range -40°C to +160°C Tolerance Class B: -40°C up to +160°C Temperature coefficient TC = 3850 ppm/K Leads Cord, silicon isolated AWG 30 Wire lengths (KL) 70 to 500mm, customer connection engraved Long- term stability max. R0-Drift 0.04 % after 1500 h at 160°C Vibration resistance at least 40g acceleration at 10 to 2000 Hz, depends on installation Shock resistance at least 100g acceleration with 8ms half sine wave, depends on installation Environmental conditions max. 80% rel. humidity; durability against non-conductive media Self heating 0.4 K/mW at 0°C Response time water current (v= 0.4m/s): air stream(v= 2m/s): t0,5 = 0.20s t0,9 = 0.30s t0,5 = 3.4s t0,9 = 11.0s Measuring current 1000: 0.1 to 0.3mA (self heating has to be considered) Note Other tolerances, values of resistance and wire lengths are available on request. We reserve the right to make alterations and technical data printed. All technical data serves as a guideline and does not guarantee particular properties to any products. Heraeus Sensor Technology USA 1901 Route 130 North Brunswick, NJ 08902 Phone 732-940-4400 Fax 732-940-4445 Email [email protected] www.hst-us.com Name of document: 30910010 Index A Status: 06/2010