NI NIPXI-5661

2.7 GHz RF Vector Signal Analyzer with Digital Downconversion
NI PXI-5661
• 9 kHz to 2.7 GHz
• 20 MHz real-time bandwidth with
quadrature digital downconversion
• Burst power triggering
• High-stability timebase (10 MHz OCXO)
• ±20 ppb frequency stability
• ±50 ppb frequency accuracy
• Up to 80 dB spurious-free dynamic range
• +30 dBm full signal input range
• 14-bit resolution, 100 MS/s digitizer
• 64 or 256 MB memory
Operating Systems
• Windows 2000/XP
Recommended Software
• LabVIEW
• LabWindows/CVI
• Modulation Toolkit
Application Software (included)
• Spectral Measurements Toolkit
Driver Software (included)
• NI-RFSA
Overview
Frequency Characteristics
The National Instruments PXI-5661 is a modular 2.7 GHz RF vector signal
analyzer with wide real-time bandwidth, optimized for automated test in a
compact, 3U PXI package. It is ideal for high-throughput RF acquisition and
measurements. The 20 MHz wide onboard quadrature digital downconverter
(DDC) significantly reduces the required load size and the time Needed to
transfer waveform data to the host controller or PC. This functionality
is ideal for engineers who require data streaming with rapid response time
for software-defined radio (SDR), real-time signal processing, or faster
transfer times for spectral monitoring applications. The module is also
compatible with Conduant StreamStor devices, enabling terabyte recording.
The NI PXI-5661 RF analyzer also includes a burst power trigger ideal
for communications applications and signal intelligence. The PXI-5661
features a highly stable timebase and flexible software tools, including
Express VIs for National Instruments LabVIEW.
The NI Spectral Measurements Toolkit for LabVIEW and
LabWindows/CVI accompanies the PXI-5661, providing common
measurements such as power spectrum, peak power and frequency, in-band
power, adjacent-channel power, and occupied bandwidth. Additionally,
it offers vector capabilities, such as 3D spectrograms, I/Q data for
modulation analysis, and analog modulation analysis functions for LabVIEW.
You can also use the PXI-5661 with the NI Modulation Toolkit to analyze
custom and standard modulation formats.
The PXI-5661 provides outstanding frequency characteristics over its
operating range of 9 kHz to 2.7 GHz. It offers a typical noise density of
<-140 dBm/Hz, extendable to -165 dBm/Hz with a PXI-5690 preamplifier,
and up to 80 dB of intermodulation spurious-free dynamic range.
Typical phase noise is <-94 dBc/Hz at a 10 kHz offset.
Superior Measurement Throughput
The most significant advantage of the PXI-5661 over traditional
RF instrumentation is measurement throughput. Figure 1 illustrates the
difference in throughput between the PXI-5661 and a traditional analyzer.
Two comparisons are shown. One graph is a spectral sweep that illustrates
the advantages of the wide real-time bandwidth of the downconverter.
The other graph is an in-band power measurement comparison that
shows the throughput advantages of the PXI-5661 as a whole.
Hardware
The PXI-5661 consists of a 2.7 GHz downconverter and a
high-spectral-purity digitizer.
Analog Input
The PXI-5661 can acquire a wide range of signal levels, from +30 dBm
to less than -130 dBm, and provides up to 50 dB of input attenuation,
selectable in 10 dB steps.
Figure 1. PXI-5661 Performance Benchmarks
2.7 GHz RF Vector Signal Analyzer with Digital Downconversion
Ultrahigh-Stability Timebase
The PXI-5661 offers an extremely stable timebase with frequency
stability of ±20 ppb and frequency accuracy of ±50 ppb, making it useful
for a range of automation applications.
Accuracy
The noise and distortion characteristics of the PXI-5661 are stable and
repeatable over time and a wide range of temperature. For example, with
a 1 MHz measurement bandwidth, a -10 dBm signal has a repeatability
of less than 0.1 dB, where resolution bandwidth (RBW) = 1 kHz and number
of averages = 10. The PXI-5661 employs amplitude equalization that
reduces amplitude error to less than 0.2 dB over a 20 °C change.
Acquisition Memory
The PXI-5661 includes up to 256 MB of onboard memory. With this
memory, you can acquire up to 128 million real 16-bit samples,
or 64 million complex 16-bit samples. The PXI-5661 uses the bus master
capability of the NI MITE ASIC to move data to computer memory at
much higher speeds – up to 10 times faster – than traditional instrument
interfaces. Because this ASIC performs memory management functions
usually handled by the host CPU, all the computer power of the host CPU
can be devoted to data analysis.
Clock Generation and Triggering
The 10 MHz reference clock on the PXI-5661 can synchronize to any one
of three sources – the onboard high-precision OCXO reference clock, an
external reference clock, or the PXI backplane. Using the PXI backplane,
you can synchronize two or more PXI-5661 units with each other and other
PXI modules without using cables. You can synchronize the PXI-5661 to
an external source using front panel connectors. The PXI-5661 can import
and export TTL triggers from the PXI trigger bus, the PXI star trigger line, or
the front panel SMB connector. The PXI trigger bus greatly simplifies
synchronizing RF measurements with other PXI modules such as digital
multimeters, audio analyzers, and machine vision modules. The PXI-5661
also includes a burst power trigger with which you can trigger an acquisition
based on the power within the configured modulation bandwidth.
Digital Downconversion and Decimation
With the digital downconversion (DDC) functionality of the PXI-5661,
you can acquire signals at much less than the full digitization rate.
By downconverting channels of up to the full 20 MHz bandwidth to
baseband, the PXI-5661 dramatically reduces the sampling rate necessary
to acquire these signals. This feature results in dramatic throughput
improvements for many applications. For example, if you want to acquire
a signal with a 200 kHz bandwidth centered at 900 MHz, the rate at
which samples are stored can be as low as 250 kS/s. Digital downconversion
and decimation in combination with the high-throughput PXI backplane
provide RF and communication data streaming to a controller or PC for
real-time signal processing and analysis or record and playback applications.
Calibration
National Instruments calibrates the amplitude accuracy of the analog input
of the downconverter and digitizer modules. Temperature variations are
calibrated and corrected during normal operation, resulting in very high
stability and repeatability. The modules are shipped with NIST-traceable
and ISO-9002-certified calibration certificates. Visit ni.com/calibration
for more information about calibration services.
Software
The National Instruments Spectral Measurements Toolkit and NI-RFSA
driver software are included with the PXI-5661. The Spectral Measurements
Toolkit plugs directly into LabVIEW and LabWindows/CVI to offer high-level
measurement functionality. For a list of measurement functions, refer to
Table 1. The driver software provides a driver-level interface and integrates
with LabVIEW and LabWindows/CVI.
Zoom FFT
Zoom power spectrum
Averaged power spectrum
Averaged cross spectrum
Averaged frequency response
Amplitude calibration
Power spectral density
Peak frequency
Peak amplitude/power
Spectrum peak search
In-band power
Adjacent channel power
Occupied bandwidth
Demodulate AM
Demodulate FM
Demodulate PM
Downconvert passband
Table 1. Spectral Measurements Toolkit Functions
General-Purpose Test
Spectral analysis
Semiconductor ATE
Ultrasound/radar/lidar
RF component characterization
Military/aerospace
Signal Intelligence (Sig Int)
Commercial Electronics Test
Cable modem
Mobile phones and pager
Wireless LAN and Bluetooth
Commercial radio
Digital television
Table 2. Applications for the PXI-5661
Ordering Information
NI PXI-5661 with:
64 MB ..................................................................................779782-01
256 MB ..................................................................................779782-02
256 MB with Modulation Toolkit ..........................................779782-03
Includes modules, cables, NI-RFSA driver, and Spectral Measurements Toolkit for
LabVIEW and LabWindows/CVI.
BUY NOW!
For complete product specifications, pricing, and accessory
information, call (800) 813 3693 (U.S. only) or go to ni.com/rf.
BUY ONLINE at ni.com or CALL (800) 813 3693 (U.S.)
2
2.7 GHz RF Vector Signal Analyzer with Digital Downconversion
Specifications
Valid over specified operating environment (0 to 50 °C) unless otherwise
stated. See detailed NI PXI-5661 specifications for more information.
General
Ports .................................................... 1 RF
1 IF
Frequency
Frequency range.................................. 9 kHz to 2.7 GHz
Real-time bandwidth .......................... 20 MHz
Resolution bandwidth (RBW).............. Fully adjustable
(<1 Hz to 10 MHz)
Internal reference frequency .............. 10 MHz
Temperature stability .......................... ±20 ppb max1
Initial achievable accuracy ................. ±50 ppb max
Aging ................................................... ±100 ppb/year
Locking range...................................... >±0.5 ppm
Lock time to ext frequency
reference......................................... <10 s
Warm-up time (typical) ....................... 15 minutes
Selectivity (60 dB : 3 dB)..................... <2.5 (Flat Top) <4.1
(7-term Blackman-Harris)
Tuning resolution
Frequency-domain digitizer
(PXI-5142) ................................... 355 nHz
RF downconverter (PXI-5600) ......... 1 MHz, minimum
Tuning speed downconverter
(1% of step size) ............................. 10 ms max
(0.01% of step size) ........................ 20 ms max
(1 ppm of step size) ........................ 30 ms max
Spectral purity (noise sidebands) at 100 MHz, minimum
1 kHz offset..................................... <-80 dBc/Hz2
10 kHz offset................................... <-90 dBc/Hz
30 kHz offset................................... <-95 dBc/Hz
100 kHz offset................................. <-110 dBc/Hz
1 MHz offset ................................... <-120 dBc/Hz
Sideband spurs
>10 kHz offset................................. <-70 dBc
<10 kHz offset................................. <-55 dBc
Residual FM ........................................ <10 Hzpp in 10 ms
10
to 50 °C, referenced to 25 °C
spans 20 kHz; for spans >20 kHz the value is <-78 dBc/Hz at 1 Hz offset
2For
Amplitude
Input signal range ............................... <-130 to 30 dBm
Maximum safe input power (continuous) +30 dBm (atten ≥10 dB)
+20 dBm (no atten)
RF input attenuator ............................. 0 to 50 dB (10 dB steps)
Maximum DC input voltage ................ ±25 VDC3
Relative accuracy (to 100 MHz, 15 to 35 °C)
<2 GHz, calibrated .......................... ±0.75 dB, ±0.5 dB typical
>2 GHz, calibrated .......................... ±1.25 dB, ±0.9 dB typical
Absolute accuracy (15 to 35 °C)
<2 GHz, calibrated .......................... ±1 dB, ±0.6 dB typical
>2 GHz, calibrated .......................... ±1.5 dB, ±1 dB typical
Group delay variation.......................... 30 nspp
All signal levels are mixer stage levels at RF input port
1 dB gain compression
10 MHz to 1 GHz............................. >0 dBm
1 to 2.7 GHz .................................... >2 dBm
3DC
levels up to ±25 VDC at input will not damage the instrument; however, high transient
currents from low impedance DC step voltages at input can cause damage.
Spurious Response
2nd-order harmonic distortion (single -30 dBm tone)
20 MHz to 2.7 GHz.......................... <-80 dBc typical
3rd-order intermodulation distortion
(two -30 dBm tones, >200 kHz separation)
20 MHz to 2.7 GHz.......................... <-85 dBc typical
Input-related spurs
Signal level = -30 dBm, 0 dB attenuation
>5 MHz............................................ <-70 dBc
<5 MHz............................................ <-60 dBc
Residual response related spurs
Input terminated, 0 dB input attenuation
≥20 MHz ......................................... <-100 dBm
<20 MHz.......................................... <-70 dBm
Noise density
9 kHz to 1 GHz ................................ <-135 dBm/Hz
(-144 dBm/Hz typical)
1 to 2 GHz ....................................... <-134 dBm/Hz
(-142 dBm/Hz typical)
2 to 2.5 GHz .................................... <-130 dBm/Hz
(-139 dBm/Hz typical)
2.5 to 2.7 GHz ................................. <-129 dBm/Hz
(-133 dBm/Hz typical)
NI PXI-5600 RF Downconverter Module
Input
Connector........................................
Impedance ......................................
Coupling..........................................
Output
Connector........................................
Impedance ......................................
Frequency........................................
Amplitude .......................................
Frequency reference input
Connector........................................
Impedance ......................................
Input amplitude...............................
Maximum safe input level..............
Maximum DC input voltage............
Input frequency range ....................
10 MHz output (2 ports)
Connector........................................
Impedance ......................................
Signal..............................................
Amplitude .......................................
SMA female
50 Ω
AC
SMA female
50 Ω
5 to 25 MHz
0 dBm full scale
SMA female
50 Ω
-5 to +15 dBm
+16 dBm
±5 V
10 MHz ±0.5 ppm
SMA female
50 Ω
Square wave
±0.5 V (+7 dBm) into 50 Ω
(±1 V into open circuit)
Accuracy ......................................... Refer to
Internal Frequency Reference
BUY ONLINE at ni.com or CALL (800) 813 3693 (U.S.)
3
2.7 GHz RF Vector Signal Analyzer with Digital Downconversion
PXI 10 MHz I/O
Connector........................................
Impedance ......................................
Frequency........................................
Output amplitude............................
Safety and Compliance
SMA female
50 Ω
-5 to +15 dBm
0.5 V (=7 dBm) into 50 Ω
NI PXI-5142 IF Digitizer Module
CH0, CH1
Connector........................................
Impedance ......................................
Maximum input overload....................
Clk in
Connector........................................
Impedance ......................................
Input amplitude...............................
BNC female
50 Ω
7 Vrms; ≤10 Vp (30 dBm)
Note: For UL and other safety certifications, refer to the product label
or visit ni.com/certification, search by model number or product line,
and click the appropriate link in the Certification column.
Electromagnetic Compatibility
This product is designed to meet the requirements of the following
standards of EMC for electrical equipment for measurement, control,
and laboratory use:
• EN 61326 EMC requirements; Minimum Immunity
• EN 55011 Emissions; Group 1, Class A
• CE, C-Tick, ICES, and FCC Part 15 Emissions; Class A
Note: For EMC compliance, operate this device according to
product documentation.
CE Compliance
This product meets the essential requirements of applicable European
Directives, as amended for CE marking, as follows:
• 73/23/EEC; Low-Voltage Directive (safety)
• 89/336/EEC; Electromagnetic Compatibility Directive (EMC)
Power Requirements
+3.3 VDC
920 mA
1A
This product is designed to meet the requirements of the following
standards of safety for electrical equipment for measurement, control,
and laboratory use:
• IEC 61010-1, EN 61010-1
• UL 61010-1, CSA 61010-1
SMB female
50 Ω
-10 dBm to +13 dBm
(square wave)
Maximum input overload.................... 7 Vrms; ≤10 Vp (30 dBm)
Clk out
Connector........................................ SMB
Output impedance .......................... 50 Ω
Logic type........................................ 3.3 V CMOS
Maximum driver current ................. ±48 mA
PFI 0, PFI 1
Connector........................................ 9-pin minicircular DIN
Logic type........................................ 3.3 V CMOS
Maximum input voltage ................. 5.5 V
Module
PXI-5600 RF downconverter
PXI-5142 IF digitizer
Safety
+5 VDC
2.3 A
1.7 A
+12 VDC
700 mA
800 mA
-12 VDC
115 mA
270 mA
Voltages ±5%
Calibration
Interval ................................................ 1 year for PXI-5600
2 years for PXI-5142
IF/Baseband
Resolution ...........................................
System IF frequency rate ....................
Sample rate.........................................
Digital downconverter (OSP) bandwidth
14 bits
5 to 25 MHz
100 MS/s
20 MHz to 4.88 kHz using
100 MS sample clock timebase
Onboard memory................................. 64 MB
256 MB
Note: Refer to the PXI-5142 specifications document for additional
IF/baseband and onboard signal processing (OSP) specifications.
Note: Refer to the Declaration of Conformity (DoC) for this product for
any additional regulatory compliance information. To obtain the DoC for
this product, visit ni.com/certification, search by model number or
product line, and click the appropriate link in the Certification column.
Waste Electrical and Electronic Equipment (WEEE)
EU Customers: At the end of their life cycle, all products must be
sent to a WEEE recycling center. For more information about WEEE
recycling centers and National Instruments WEEE initiatives, visit
ni.com/environment/weee.htm.
Physical Dimensions
PXI-5600 (3 slots) ................................ 3U, PXI/cPCI module
6.0 by 13.0 by 21.6 cm
(2.4 by 5.1 by 8.5 in.)
PXI-5142 (1 slot).................................. 3U, PXI/cPCI module
2.0 by 13.0 by 21.6 cm
(0.8 by 5.1 by 8.5 in.)
Weight (combined unit) ...................... 1,165 g (41.1 oz)
BUY ONLINE at ni.com or CALL (800) 813 3693 (U.S.)
4
NI Services and Support
SERVICE
NEEDS
NI has the services and support to meet
your needs around the globe and through
the application life cycle – from planning
and development through deployment
and ongoing maintenance. We offer
services and service levels to meet
customer requirements in research,
design, validation, and manufacturing.
Visit ni.com/services.
Training and Certification
NI training is the fastest, most certain route to productivity with our
products. NI training can shorten your learning curve, save development
time, and reduce maintenance costs over the application life cycle. We
schedule instructor-led courses in cities worldwide, or we can hold a
course at your facility. We also offer a professional certification program
that identifies individuals who have high levels of skill and knowledge on
using NI products. Visit ni.com/training.
Professional Services
Our Professional Services Team is composed of NI applications engineers,
NI Consulting Services, and a worldwide National Instruments Alliance
Partner program of more than 600 independent consultants and
integrators. Services range from
start-up assistance to turnkey
system integration.
Visit ni.com/alliance.
OEM Support
We offer design-in consulting and product integration assistance if you
want to use our products for OEM applications. For information about
special pricing and services for OEM customers, visit ni.com/oem.
ni.com • (800) 813 3693
National Instruments • [email protected]
Local Sales and Technical Support
In offices worldwide, our staff is local to the country, giving you access
to engineers who speak your language. NI delivers industry-leading
technical support through online knowledge bases, our applications
engineers, and access to 14,000 measurement and automation
professionals within NI Developer Exchange forums. Find immediate
answers to your questions at ni.com/support.
We also offer service programs that provide automatic upgrades to
your application development environment and higher levels of technical
support. Visit ni.com/ssp.
Hardware Services
NI Factory Installation Services
NI Factory Installation Services (FIS) is the fastest and easiest way to
use your PXI or PXI/SCXI combination systems right out of the box.
Trained NI technicians install the software and hardware and configure
the system to your specifications. NI extends the standard warranty by
one year on hardware components (controllers, chassis, modules)
purchased with FIS. To use FIS, simply configure your system online
with ni.com/pxiadvisor.
Calibration Services
NI recognizes the need to maintain properly calibrated devices for
high-accuracy measurements. We provide manual calibration
procedures, services to recalibrate your products, and automated
calibration software specifically designed for use by metrology
laboratories. Visit ni.com/calibration.
Repair and Extended Warranty
NI provides complete repair services for our products. Express repair
and advance replacement services are also available. We offer
extended warranties to help you meet project life-cycle requirements.
Visit ni.com/services.
*351392A-01*
351392A-01
© 2006 National Instruments Corporation. All rights reserved. CVI, LabVIEW, MITE, National Instruments, National Instruments Alliance Partner, NI, ni.com,
and SCXI are trademarks of National Instruments. Other product and company names listed are trademarks or trade names of their respective companies.
A National Instruments Alliance Partner is a business entity independent from NI and has no agency, partnership, or joint-venture relationship with NI.
2006-7812-101-D