MEDICAL & MILITARY PM41 Series 3 0 - 4 0 W AT T S S I N G L E & M U LT I P L E O U T P U T A C / D C M E D I C A L Features • Low safety ground leakage current. • Wide input range 85 to 264Vac. • Input surge current protection. • Overvoltage protection. • Overcurrent protection. • Compact, open PCB construction. • 100% burn-in at full rated load. Specifications OPERATING INPUT Input Voltage 85–264Vac wide input Frequency 47Hz–400Hz Input Current 0.7A (rms) for 115Vac. 0.4A (rms) for 230Vac. 600,000 hrs @ 25°C Efficiency 65%–70% ENVIRONMENTAL Operating Temperature 0°C to 70°C Leakage current 90µA max. @ 115Vac. 60Hz. 150µA max. @ 230Vac. 50Hz. Derating Derate to 50% at 70°C Isolation Input – Output: 4000VAC Input – Ground: 1500VAC Output – Ground: 500VAC Storage Temperature -40°C to 85°C Relative Humidity 5%–95% non-condensing Inrush Current 12A @ 115V, 24A @ 230V OUTPUT 332 MTBF STANDARDS AND APPROVALS Safety standards UL2601-1, CSA C22.2, TUV EN60601-1 C-Tick AS/NZS CISPR11 Group 1, Class A EMC standards EN60601-1-2: 2001,EN61000-3-2, -3, EN61000-4-2, -3, -4, -5, -6, -8, -11 EN55011 Class B, FCC Class B, VCCI Class B Output Voltage See table Output Voltage Adjustment V1 ±10% Output Current See table EMI standards Ripple & Noise 1% peak to peak MECHANICAL Line Regulation ±0.5% maximum at full load. Dimensions 101.6x50.8x38.1mm Load Regulation See table Weight Single output model: 190grams. Multi-output model: 175 grams. Transient Response Max. deviation 4% recovering to 1% within 500µs after 25% load charge Overvoltage Protection Provided on V1 only. Set at 110%–132% of nominal Overcurrent Protection All outputs protected for short circuit conditions Temperature Coefficient All outputs ±0.04% per °C Holdup Time Typically 10ms at 110VAC MEDICAL & MILITARY PM41 Series 3 0 - 4 0 W AT T S S I N G L E & M U LT I P L E O U T P U T A C / D C M E D I C A L Selection Table MAX. MODEL OUTPUT #1 OUTPUT NUMBER VNOM IMIN IMAX TOL POWER PM41-10A 5.1V 0A 6.0A 2% 30W PM41-12A 12V 0A 3.3A 2% 40W PM41-13A 15V 0A 2.7A 2% 40W PM41-13-1A 18V 0A 2.3A 2% 40W PM41-14A 24V 0A 1.7A 2% 40W PM41-15A 28V 0A 1.4A 2% 40W NOTES: Ripple and noise: Peak to Peak with 20 MHz bandwidth and 10F in parallel with a 0.1F capacitor at rated line voltage and load ranges. Pin Chart SINGLE OUTPUT MODELS MODEL/PIN PM41-10A PM41-12A PM41-13A PM41-13-1A PM41-14A PM41-15A 1 2 3 4 Output Output Return Return Technical Illustration Your dependable power partner – www.powerbox.com.au 333