MEDICAL PM66 Series 5 1 - 6 5 W AT T S - A C / D C S I N G L E O U T P U T FEATURES • Low safety ground leakage current • Wide input range 90 to 264Vac • Input surge current protection • Overvoltage protection • Overcurrent protection • Compact, open PCB construction • 100% burn-in at full rated load SPECIFICATIONS INPUT OPERATING Input voltage 90–264Vac wide input MTBF 400,000 hrs @ 25°C Frequency 47Hz–67Hz Efficiency 75% minimum Input current 0.70A (rms) for 115Vac. 0.40A (rms) for 230Vac. 90µA max. @ 115Vac. 60Hz. 150µA max. @ 230Vac. 50Hz. Input – Output: 4000VAC Input – Ground: 1500VAC Output – Ground: 500VAC Leakage current Isolation Inrush Current 17A @ 115V, 40A @ 230V 0°C to 70°C Derating See curve Storage temperature -40°C to 85°C Relative humidity 5%–95% non-condensing STANDARDS AND APPROVALS Safety UL 60601-1, CSA C22.2 No. 601.1, TÜV EN 60601-1, UL 60950-1, CSA C22.2 No. 60950-1, TÜV EN 60950-1 Output voltage adjustment V1 ±10% C-Tick AS/NZS CISPR11 Group1, Class A Output current See table EMC standards Ripple & noise 1% peak to peak Line regulation ±0.5% maximum at full load. Load regulation Transient Response Temperature coefficient See table Max. deviation 4% recovering to 1% within 500µs after 25% load charge Provided on V1 only. Set at 110%–132% of nominal All outputs protected for short circuit conditions All outputs ±0.04% per °C IEC60601-1-2, EN55011 / EN55022 and FCC, VCCI: Class B conducted . class B radiated EN61000-3-2, EN61000-3-3, EN61000-4-2, EN61000-4-3, EN61000-4-4, EN61000-4-5, EN61000-4-6, EN61000-4-8, EN61000-4-11 Holdup time Typically 10mS at 110VAC OUTPUT Output voltage Overvoltage protection Overcurrent protection 396 ENVIRONMENTAL Operating temperature See table Immunity standards MECHANICAL Dimensions 127x76.2x31.6mm Weight 330grams. Doc. No. PBC9 Rev.C 28-4-11 PM66 Series SELECTION TABLE OUTPUT EFFICIENCY (TYPICAL) @ MAX. OUTPUT POWER 115 / 230 VAC MODEL V1 MIN. CURRENT MAX. CURRENT TOL. RIPPLE & NOISE(1) MAX. OUTPUT POWER PM66-10A PM66-12A 5.1 V 12 V 0A 0A 10.0 A 5.5 A ±3% ±2% 51 mV 120 mV 51 W 65 W 71 / 71 % 79 / 80 % PM66-13A PM66-13-1A 15 V 18 V 0A 0A 4.4 A 3.7 A ±2% ±2% 150 mV 180 mV 65 W 65 W 79 / 81 % 80 / 81 % PM66-14A PM66-15A 24 V 28 V 0A 0A 2.8 A 2.4 A ±2% ±2% 240 mV 280 mV 65 W 65 W 80 / 81 % 81 / 80 % MEDICAL 5 1 - 6 5 W AT T S - A C / D C S I N G L E O U T P U T NOTES: 1. Ripple and noise is maximum peak-to-peak voltage value measured at output within 20 MHz bandwidth, at rated line voltage and output load ranges, and with a 10 µF tantalum capacitor in parallel with a 0.1 µF ceramic capacitor across the output. OUTPUT POWER DERATING CURVE MECHANICAL SPECIFICATIONS NOTES: 1. 2. 3. 4. 5. 6. 7. Dimensions shown in inches [mm] Tolerance 0.02 [0.5] maximum Connector P1: Molex header 09-65-2038 or equivalent, mating with Molex housing 09-50-1031 or equivalent. Connector P2 is 0.25 [6.35] × 0.032 [0.8] for earth grounding. Connector P3: Molex 09-65-2068 or equivalent, mating with Molex 09-50-1061 or equivalent. Connector P4: Molex header 22-04-1021 or equivalent for ±sense connections, mating with Molex housing 22-01-1023 or equivalent. Weight: 330 grams (0.726 lbs.) approx. PIN CHART P1 MODEL PM66-10A PM66-12A PM66-13A PIN PM66-13-1A PM66-14A PM66-15A Doc. No. PBC9 Rev.C 28-4-11 1 2 AC Live Void P2 3 1 AC AC Neutral Ground P3 1 2 3 +V1 +V1 +V1 Powering Progress – www.powerbox.com.au P4 4 5 V1 Return 6 1 2 +Sense -Sense 397