5 1 - 6 5 W AT T S - A C / D C S I N G L E O U T P U T
FEATURES
•
Low safety ground leakage current
•
Wide input range 90 to 264Vac
•
Input surge current protection
•
Overvoltage protection
•
Overcurrent protection
•
Compact, open PCB construction
•
100% burn-in at full rated load
SPECIFICATIONS
INPUT
OPERATING
Input voltage
90–264Vac wide input
MTBF
400,000 hrs @ 25°C
Frequency
47Hz–67Hz
Efficiency
75% minimum
Input current
0.70A (rms) for 115Vac.
0.40A (rms) for 230Vac.
90µA max. @ 115Vac. 60Hz.
150µA max. @ 230Vac. 50Hz.
Input – Output: 4000VAC
Input – Ground: 1500VAC
Output – Ground: 500VAC
Leakage current
Isolation
Inrush Current
17A @ 115V, 40A @ 230V
0°C to 70°C
Derating
See curve
Storage temperature
-40°C to 85°C
Relative humidity
5%–95% non-condensing
STANDARDS AND APPROVALS
Safety
UL 60601-1, CSA C22.2 No. 601.1, TÜV EN
60601-1,
UL 60950-1, CSA C22.2 No.
60950-1, TÜV EN 60950-1
Output voltage adjustment V1 ±10%
C-Tick
AS/NZS CISPR11 Group1, Class A
Output current
See table
EMC standards
Ripple & noise
1% peak to peak
Line regulation
±0.5% maximum at full load.
Load regulation
Transient Response
Temperature coefficient
See table
Max. deviation 4% recovering to 1% within
500µs after 25% load charge
Provided on V1 only. Set at 110%–132% of
nominal
All outputs protected for short circuit
conditions
All outputs ±0.04% per °C
IEC60601-1-2, EN55011 / EN55022 and FCC,
VCCI: Class B conducted . class B radiated
EN61000-3-2, EN61000-3-3, EN61000-4-2,
EN61000-4-3, EN61000-4-4, EN61000-4-5,
EN61000-4-6, EN61000-4-8, EN61000-4-11
Holdup time
Typically 10mS at 110VAC
OUTPUT
Output voltage
Overvoltage protection
Overcurrent protection
396
ENVIRONMENTAL
Operating temperature
See table
Immunity standards
MECHANICAL
Dimensions
127x76.2x31.6mm
Weight
330grams.
Doc. No. PBC9 Rev.C 28-4-11
PM66 Series
SELECTION TABLE
OUTPUT
EFFICIENCY (TYPICAL)
@ MAX. OUTPUT
POWER
115 / 230 VAC
MODEL
V1
MIN. CURRENT MAX. CURRENT
TOL.
RIPPLE &
NOISE(1)
MAX. OUTPUT
POWER
PM66-10A
PM66-12A
5.1 V
12 V
0A
0A
10.0 A
5.5 A
±3%
±2%
51 mV
120 mV
51 W
65 W
71 / 71 %
79 / 80 %
PM66-13A
PM66-13-1A
15 V
18 V
0A
0A
4.4 A
3.7 A
±2%
±2%
150 mV
180 mV
65 W
65 W
79 / 81 %
80 / 81 %
PM66-14A
PM66-15A
24 V
28 V
0A
0A
2.8 A
2.4 A
±2%
±2%
240 mV
280 mV
65 W
65 W
80 / 81 %
81 / 80 %
MEDICAL
5 1 - 6 5 W AT T S - A C / D C S I N G L E O U T P U T
NOTES:
1. Ripple and noise is maximum peak-to-peak voltage value measured at output within 20 MHz bandwidth, at rated line voltage and output load
ranges, and with a 10 µF tantalum capacitor in parallel with a 0.1 µF ceramic capacitor across the output.
OUTPUT POWER DERATING CURVE
MECHANICAL SPECIFICATIONS
NOTES:
1.
2.
3.
4.
5.
6.
7.
Dimensions shown in inches [mm]
Tolerance 0.02 [0.5] maximum
Connector P1: Molex header 09-65-2038 or equivalent, mating with Molex housing 09-50-1031 or equivalent.
Connector P2 is 0.25 [6.35] × 0.032 [0.8] for earth grounding.
Connector P3: Molex 09-65-2068 or equivalent, mating with Molex 09-50-1061 or equivalent.
Connector P4: Molex header 22-04-1021 or equivalent for ±sense connections, mating with Molex housing 22-01-1023 or equivalent.
Weight: 330 grams (0.726 lbs.) approx.
PIN CHART
P1
MODEL
PM66-10A
PM66-12A
PM66-13A
PIN
PM66-13-1A
PM66-14A
PM66-15A
Doc. No. PBC9 Rev.C 28-4-11
1
2
AC
Live
Void
P2
3
1
AC
AC
Neutral Ground
P3
1
2
3
+V1
+V1
+V1
Powering Progress – www.powerbox.com.au
P4
4
5
V1 Return
6
1
2
+Sense -Sense