ASM-SENSOR 3664

OPTICAL POWER METER 3664
Optical/Telecom Measurement
From Blue-Ray, High Definition DVD to Near-Infrared Rays
An Advanced Optical Power Meter to Meet
Next Generation Needs
■ Great Cost Performance in
a Compact Package
■ Ideal for all applications - from
Production to R&D to Maintenance
■ USB Ver.1.1 Compatible
■ Scaling
■ MAX/MIN/AVE Measurement
■ Large, easy-to-read LCD display
■ Analog Output
1
A Truly Flexible Instrument Catering to Applications in R&D,
Optical Power Meter 3664 serves as a convenient tool for the R&D, manufacture and maintenance of equipment
that incorporate laser light sources, such as DVD recorders, CD drives, copiers, and laser printers. With the
addition of the Optical Sensors 9743 and 9743-10 specifically designed for blue-violet optical rays that have low
dependency on the incidence angle and flat wave sensitivity characteristics over its 405nm bandwidth, the 3664
also proves to be a powerful instrument for inspecting next generation optical pickup drive devices.
Four Sensors to Accommodate All Types of Testing Needs
Choose from 4 different optical sensors according to wavelength and sensor structure to aptly meet
your application requirements.
■ Two Different Bandwidths
Blue-violet Optical Sensor 9743/9743-10
Incorporated with characteristics that are ideal for
measuring blue-violet lasers found in high-definition DVDs,
Models 9743 and 9743-10 offer the following features:
• Planar wavelength sensitivity characteristics at
the 405 nm bandwidth
• Low Incident Angle Dependency and Low Reflection
General-purpose Sensors 9742/9742-10
for an Extensive Range of Uses
Test DVD and CD drives, Copiers, Printers
and much, much more.
• Wide 320 to 1100 nm wavelength
■ Two Different Styles
Handheld model
9742/9743
Detachable model
Ideal for R&D and maintenance
work when functionality is a
priority.
9742-10/9743-10
Enhance production efficiency
by integrating the sensor into
the inspection line.
Model 9743.
Model 9743-10.
2
Production and Maintenance of Laser- and LED-Dependent Equipment
■ Superior cost performance
Powerful Features
Guaranteed accuracy of ±5%, ideal for the production and testing of optical pickup devices (in combination with the
Optical Sensor 9742/9742-10/9743/9743-10) Scaling functionality and USB compatibility all for one reasonable price.
■ Scaling
■ 2 Power settings
■ Wavelength setting
resolution up to 1 nm
■ Relative measurement
Adjust for sensitivity at the wavelength
level, and easily integrate the 3664 into an
inspection standard device.
Compatible to both DC power and AC
power. Choose a power method to suit
your application.
Display measurements as relative
values, i.e., the difference from a
set reference value. Load reference
values from a measured value or define
according to your requirements.
Up to 10 wavelength memory presets can be
configured, including the defaults for each
optical sensor. The default wavelengths for each
optional sensor are already stored in memory.
Add up to 6 more custom wavelengths and
call them up simply by pressing the λ key.
■ Upload data through the
USB interface
Program your PC to download captured
data, configure and even control the
3664 through the USB interface.
(USB driver software included)
■ MAX/MIN/AVE measurement
Display maximum, minimum, and averages,
as well as make relative measurements.
Specifications
Accuracy: 23°±5°C (73°±9°F), less than 80% rh, One year accuracy guarantee
■ OPTICAL POWER METER 3664 Specifications
Optical power measurement
Range
Accuracy
Calibration
wavelength
Scaling
Optical loss
measurement
Display
Measurement display
Wavelength display
Display refresh rate
MAX/MIN display
AVE display
Analog Output
Output resistance
Output connector
Interface
Included features
Applicable
standards
Power
Max. rated power
Operating time
Units W/dBm
Auto (manual settings available)
± 0.7 % (± 5 % when used with optional light sensor)
Resolution of 1 nm, automatic calibration of sensor, up to 10
wavelength presets available (including defaults for each sensor)
Configurable for each wavelength
Displays a measured value compared with a reference value
(Displayed value = measurement − reference)
* Reference value can be based on a measurement, or input manually
* Settings range: 0.001 nW to 1.9999 W (−90.00 dBm to 33.00 dBm)
4 1/2 digits, up to 19999. Display resolution: 0.01 dBm/0.01 dB
Units: nW/µW/mW/dBm/dB
4 digits, unit: nm
Approx. 330 ms
Displays MAX/MIN during measurement
Moving average, average count configurable from 2 to 100
According to optional Optical Sensor in use
9742/9742-10: Approx. 1V at sensor correction input
9743/9743-10: Approx. 0.7V at sensor correction input
50 Ω
ø3.5 mini jack
USB Ver1.1
Output of measurement data, configuration and control supported
Auto power save, configuration backup, battery check
Safety standard: EN61010-1
EMC:EN61326, EM61000-3-2, EM61000-3-3
LR6 (AA) Alkaline battery ✕ 4, AC adapter (9445-02/9445-03)
1.6 VA
Approx. 60 hours (when using battery, continuous use with
9742 optical sensor as correction input)
Operating conditions 0°C to -40°C (32°F to 104°F), up to 80 % rh (no condensation)
Storage conditions -10°C to 50°C (14°F to 122°F), up to 80 % rh (no condensation)
Operating environment Indoor, up to elevation of 2000 meters (6562 ft)
Approx. 85W✕160H✕35Dmm (3.35W✕6.30H✕1.38 D inches)
Dimensions and mass
(excluding protrusions), Approx. 270g (9.5 oz.) (without batteries)
■ OPTICAL SENSOR Specifications
9742/9742-10
Measured wavelength
Measured power
Maximum rated value
Receiving element
Receptor size
Accuracy
Wavelength configuration defaults
Operating conditions
Storage conditions
Operating environment
Mass
9743/9743-10
Measured wavelength
Measured power
Maximum rated value
Receiving element
Receptor size
Accuracy
Wavelength configuration defaults
Operating conditions
Storage conditions
Operating environment
Mass
320 to 1100 nm
−59 dBm to +17 dBm (correction wavelength)
50 mW (+17 dBm) (under direct lighting)
Si Photo diode
Approx. 9.6 mm ✕ 9.6 mm ( 0.38" ✕ 0.38")
± 4.3 % (± 5 % when used with OPTICAL POWER METER 3664)
[Correction conditions] correction wavelength 633 nm,
correction power 100 µW, when ø approx. 2 mm (0.08") parallel
beam strikes perpendicular to center of optical sensor, CW light
633 nm, 635 nm, 650 nm, 780 nm
0 °C to -40 °C (32°F to 104°F), up to 80 % rh (no condensation)
−10 °C to 50 °C (14°F to 122°F), up to 80 % rh (no condensation)
Indoor, elevation up to 2000 meters (6562 ft)
Approx. 100 g (3.5 oz.)
380 to 450 nm
−50 dBm to + 20 dBm (correction wavelength)
100 mW (+20 dBm) (under direct lighting)
Si Photo diode
Approx. 10 mm ✕ 10 mm ( 0.39" ✕ 0.39")
±4.3% (±5% when used with OPTICAL POWER METER 3664)
[Correction conditions] correction wavelength 405 nm
(Using a 405 ± 5 nm wavelength as the reference wavelength) , correction
power 100 µW, when ø approx. 1.5 mm (0.06") parallel
beam strikes perpendicular to center of optical sensor
400 nm, 403 nm, 405 nm, 408 nm
0 °C to -40 °C (32°F to 104°F), up to 80 % rh (no condensation)
−10 °C to 50 °C (14°F to 122°F), up to 80 % rh (no condensation)
Indoor, elevation up to 2000 meters (6562 ft)
[9743]: Approx. 110 g (3.9 oz.)
[9743-10]: Approx. 100 g (3.5 oz.)
Please refer to the external diagrams depicted on the next page for the exact dimensions of the optical sensors.
3
OPTICAL SENSOR exterior view
■ OPTICAL SENSOR 9742
■ OPTICAL SENSOR 9742-10
■ OPTICAL SENSOR 9743
(can be secured with screws)
■ OPTICAL SENSOR 9743-10
(can be secured with screws)
(can be secured with screws)
■ 3664 OPTICAL POWER METER
Accessories
Accessories :
AC ADAPTER 9445-02 (UL) (1) or AC ADAPTER 9445-03 (CEE) (1),
OUTPUT CORD 9094 (1), USB CABLE (1), USB Driver CD-R (1),
Strap (1), Batteries (AAx4)
The POWER METER 3664 must be used in conjunction
with either the 9742 or 9742-10 or 9743 or 9743-10
OPTICAL SENSOR, sold separately.
9445-02 (UL) or
9445-03 (CEE)
9445-02 /-03 AC ADAPTER
9094 OUTPUT CORD
USB CABLE
DC 9 V, 1 A
(AC 100 V to 240 V, 50/60 Hz)
Analog output
Cord length 1.5m (4.92 ft)
Cord length 1m
(3.28 ft)
Options
OPTICAL SENSOR 9742
Handheld sensor
OPTICAL SENSOR 9742-10
(detachable model)
OPTICAL SENSOR 9743
OPTICAL SENSOR 9743-10
Sensor specifically for
Handheld sensor
specifically for blue-violet blue-violet optical rays
optical rays
(detachable model)
CARRYING CASE 9246
Handy for travel
and storage
Optical Power Meters installed with firmware version 1.01 or earlier must be updated to support compatibility with the new Optical Sensor 9743/9743-10
When using Model 9743/9743-10 with an earlier version of Model 3664, upon connecting the sensor to the optical power meter and
turning on the power, "Err1" will be displayed.
To update your Model 3664, please visit our website at www.hioki.com to download the latest firmware, "HIOKI 3664 Setup Utility"
and refer to the instructions provided to learn how to set up the proprietary data for the new optical sensor.
HIOKI (Shanghai) Sales & Trading Co., Ltd. :
1904 Shanghai Times Square Office, 93 Huai Hai Zhong Road
Shanghai, P.R.China POSTCODE: 200021
TEL +86-21-6391-0090/0092 FAX +86-21-6391-0360
E-mail: [email protected]
HEAD OFFICE :
81 Koizumi, Ueda, Nagano, 386-1192, Japan
TEL +81-268-28-0562 / FAX +81-268-28-0568
E-mail: [email protected]
HIOKI USA CORPORATION :
6 Corporate Drive, Cranbury, NJ 08512 USA
TEL +1-609-409-9109 / FAX +1-609-409-9108
E-mail: [email protected]
DISTRIBUTED BY
Beijing Office :
A-2602 Freetown, 58 Dong San Huan Nan Road
Beijing, P.R.China POSTCODE: 100022
TEL +86-10-5867-4080/4081 FAX +86-10-5867-4090
E-mail: [email protected]
Guangzhou Office :
Room 303, Profit Plaza, No.76, West Huangpu Road
Guangzhou, P.R.China POSTCODE: 510623
TEL +86-20-38392673/2676 FAX +86-20-38392679
E-mail: [email protected]
All information correct as of Jun. 22, 2007. All specifications are subject to change without notice.
3664E2-76E-00
Printed in Japan