EXTECH 45170

User's Guide
4-in-1 Humidity, Temperature
Airflow and Light Meter
Model 45170
Introduction
Congratulations on your purchase of Extech’s Model 45170 4-in-1 Humidity,
Temperature, Airflow and Light Meter. With careful use, this meter will provide years of
reliable service.
Meter Description
1.
Vane Airflow Sensor
2.
Type K Thermocouple Input Jack
3.
Display
4.
Power Button
5.
Unit /Zero Button
6.
C°/F° /Lux/Ft-cd Button
7.
Light Sensor
1
8
2
8.
9.
3
9
4
RH and Air Temperature Sensors
5
Hold Button
10
11
6
10. Max/Min Button
7
11. Function Button
Warranty
EXTECH INSTRUMENTS CORPORATION (A FLIR COMPANY) warrants this instrument to
be free of defects in parts and workmanship for one year from date of shipment (a six
month limited warranty applies to sensors and cables). If it should become necessary to
return the instrument for service during or beyond the warranty period, contact the
Customer Service Department at (781) 890-7440 ext. 210 for authorization or visit our
website www.extech.com for contact information. A Return Authorization (RA) number
must be issued before any product is returned to Extech. The sender is responsible for
shipping charges, freight, insurance and proper packaging to prevent damage in transit.
This warranty does not apply to defects resulting from action of the user such as misuse,
improper wiring, operation outside of specification, improper maintenance or repair, or
unauthorized modification. Extech specifically disclaims any implied warranties or
merchantability or fitness for a specific purpose and will not be liable for any direct, indirect,
incidental or consequential damages. Extech's total liability is limited to repair or
replacement of the product. The warranty set forth above is inclusive and no other
warranty, whether written or oral, is expressed or implied.
2
45170 V1.1 5/09
Operation
Turn Power ON
Press the power button to turn power on. The meter will perform a short self-test.
A.
Air Velocity Measurements
1. Press the Function button to select the Anemometer mode. In Air Flow mode,
the display will show Air Velocity and Temperature.
2. Press Unit/Zero button to select desired unit of measure (FPM, MPH, KNOT,
KM/H, or M/S) and face the Air Flow Sensor towards the source of wind.
3. Allow time for the reading to stabilize and note value. From a practical point of
view the velocity may fluctuate slightly.
B.
Humidity and Ambient Air Temperature Measurements
1. Press the Function button to select RH mode.. In RH mode, the display will
show %RH and Temperature.
2. Press the °C/°F button to select the desired temperature unit of measure.
C.
Temperature Measurements Using Type K Thermocouple
1. Press the Function button to select the Temperature mode. In Temperature
mode, the display will show only temperature.
2. Plug the thermocouple into the meter's thermocouple input jack. If no
thermocouple is connected, or if the thermocouple is defective, the meter will
----display “ “. (Primary display temperature will only work if a thermocouple
is connected).
D.
Light Measurements
1. Press the Function button to select the Light Meter mode . The light values
are displayed and orientated 180° from the other functions for easy use.
2. Press the Lux/Ft-cd to select Lux or Footcandle units of measure.
3. Before making measurements “Zero” the meter out by blocking the light
sensor to simulate a no light condition and pressing the Unit/Zero button.The
reading should go to Zero.
4. Hold the sensor toward the light source and note reading.
MIN, MAX, Data Recording Function
1. Press the Max/Min button once. The RECORD symbol will appear on the
display.This starts the Min/Max Data Recording Mode.
2. Press the Max/Min again and the MAX symbol along with the maximum value will
appear on the display.
3. Press the Max/Min once more and the MIN symbol along with the minimum value
will appear on the display.
4. To Clear the recorded Max/Min values press the HOLD button once. Press and
Hold the Max/Min button for 3 seconds to exit Min/Max Data Record Mode and
return to normal measurement mode. Record and Max/Min will disappear from the
display once this is done.
3
45170 V1.1 5/09
Data Hold
Press the HOLD button to freeze the reading in the display. The 'HOLD' icon will appear on
the upper right-hand side of the display. Press the HOLD button again to return to normal
operation (the 'HOLD' hold icon will disappear).
Auto Power OFF
The meter will automatically turn off after 10 minutes if no keys are pressed during this
period. Auto power off is disabled in the max/min recording mode.
Turning Power OFF
Press the power button to turn power off.
Calibration and Repair Services
Extech offers repair and calibration services for the products we sell. Extech also
provides NIST certification for most products. Call the Customer Care Department for
information on calibration services available for this product. Extech recommends that
annual calibrations be performed to verify meter performance and accuracy.
Useful Equations
Area equations
R
H
W
A=WxH
A = pi x R
2
Cubic equations
3
2
CFM (ft /min) = Air Velocity (ft/min) x Area (ft )
3
2
CMM (m /min) = Air Velocity (m/sec) x Area (m ) x 60
4
45170 V1.1 5/09
Specifications
Range Specifications
Measurement
Range
Resolution Accuracy
MPH (Miles per hour)
0.9 to 67.0MPH
0.1MPH
km/hr (kilometers per hour)
1.4 to 108.0 km/h
0.1km/h
≤ 3937 ft/min: ±3% F.S.
Knots (nautical miles per hour) 0.8 to 58.3 knots
0.1knots
> 3937 ft/min: ±4% F.S
m/sec (meters per second)
0.4 to 30.0 m/s
0.1 m/s
ft/min (feet per minute)
80 to 5910 ft/min
Temperature/Thermistor
32 to 122 F (-0 to 50 C)
Temperature /Thermocouple
Relative Humidity
1ft/min
0.1oF/C
± 2.5oF (± 1.2oC)
-148 to 2372°F
0.1°F
± (1% + 2°F)
-100 to 1300°C
0.1°C
± (1% + 1°C)
10.0 to 95.0%
0.1%
±4% RH (from 10% to 70%RH)
o
o
±4%rdg +1.2% RH (> 70% RH)
Light (Auto Ranging )
0 to 2,200 Lux
1 Lux
1,800 to 20,000 Lux
10 Lux
0 to 204.0 Fc
0.1 Fc
170 to 1,860 Fc
±5% rdg +8 digits
1 Fc
General Specifications
Display
4 digit (9999 count) Dual Display LCD
Sensors
Thin film capacitance humidity sensor
Min/Max
Min/Max recalls the highest/lowest reading
Data Hold
Data Hold freezes the display
Operating conditions
32 to 122oF (0 to 50oC) / < 80% RH
Power supply
9 Volt Battery
Dimensions / Weight
Instrument: 6.14 x 2.36 x 1.29” (156 x 60 x 33mm)
Vane: 1” ( 24mm ) diameter / 5.7 oz. ( 160 g )
Support line (781) 890-7440
Technical Support: Extension 200; E-mail: [email protected]
Repair & Returns: Extension 210; E-mail: [email protected]
Product specifications subject to change without notice
For the latest version of this User Guide, Software updates, and other
up-to-the-minute product information, visit our website: www.extech.com
Extech Instruments Corporation, 285 Bear Hill Road, Waltham, MA 02451
Copyright © 2009 Extech Instruments Corporation (a FLIR company)
All rights reserved including the right of reproduction in whole or in part in any form.
5
45170 V1.1 5/09
6
45170 V1.1 5/09