AD AD7716

a
FEATURES
22-Bit Sigma-Delta ADC
Dynamic Range of 105 dB (146 Hz Input)
60.003% Integral Nonlinearity
On-Chip Low-Pass Digital Filter
Cutoff Programmable from 584 Hz to 36.5 Hz
Linear Phase Response
Five Line Serial I/O
Twos Complement Coding
Easy Interface to DSPs and Microcomputers
Software Control of Filter Cutoff
65 V Supply
Low Power Operation: 50 mW
APPLICATIONS
Biomedical Data Acquisition
ECG Machines
EEG Machines
Process Control
High Accuracy Instrumentation
Seismic Systems
GENERAL DESCRIPTION
The AD7716 is a signal processing block for data acquisition
systems. It is capable of processing four channels with bandwidths of up to 584 Hz. Resolution is 22 bits and the usable
dynamic range varies from 111 dB with an input bandwidth of
36.5 Hz to 99 dB with an input bandwidth of 584 Hz.
The device consists of four separate A/D converter channels that
are implemented using sigma-delta technology. Sigma-delta
ADCs include on-chip digital filtering and, thus, the system
filtering requirements are eased.
Three address pins program the device address. This allows a
data acquisition system with up to 32 channels to be set up in a
simple fashion. The output word from the device contains 32
bits of data. One bit is determined by the state of the DIN1 input and may be used, for example, in an ECG system with an
external pacemaker detect circuit to indicate that the output
word is invalid because of the presence of a pacemaker pulse.
LC2MOS
22-Bit Data Acquisition System
AD7716
FUNCTIONAL BLOCK DIAGRAM
AVDD DVDD AVSS
RESET
A0 A1
AD7716
AIN 1
AIN 2
AIN 3
ANALOG
MODULATOR
ANALOG
MODULATOR
ANALOG
MODULATOR
A2
CLKIN
CLKOUT
CLOCK
GENERATION
LOW PASS
DIGITAL
FILTER
MODE
LOW PASS
DIGITAL
FILTER
LOW PASS
DIGITAL
FILTER
CONTROL
LOGIC
CASCIN
OUTPUT
SHIFT
REGISTER
SDATA
CASCOUT
RFS
SCLK
DRDY
AIN 4
ANALOG
MODULATOR
VREF
AGND
LOW PASS
DIGITAL
FILTER
DGND
CONTROL
REGISTER
TFS
DIN 1
DOUT 1
DOUT 2
There are 22 bits of data corresponding to the analog input.
Two bits contain the channel address and 3 bits are the device
address. Thus, each channel in a 32-channel system would have
a discrete 5-bit address. The device also has a CASCOUT pin
and a CASCIN pin that allow simple networking of multiple
devices.
The on-chip control register is programmed using the SCLK,
SDATA and TFS pins. Three bits of the Control Register set
the digital filter cutoff frequency for the device. Selectable frequencies are 584 Hz, 292 Hz, 146 Hz, 73 Hz and 36.5 Hz. A
further 2 bits appear as outputs DOUT1 and DOUT2 and can be
used for controlling calibration at the front end. The device is
available in a 44-pin PQFP (Plastic Quad Flatpack) and 44-pin
PLCC.
REV. A
Information furnished by Analog Devices is believed to be accurate and
reliable. However, no responsibility is assumed by Analog Devices for its
use, nor for any infringements of patents or other rights of third parties
which may result from its use. No license is granted by implication or
otherwise under any patent or patent rights of Analog Devices.
One Technology Way, P.O. Box 9106, Norwood, MA 02062-9106, U.S.A.
Tel: 617/329-4700
Fax: 617/326-8703
1, 2 (f = 8 MHz; MODE Pin Is High (Slave Mode Operation); AV = DV = +5 V
AD7716–SPECIFICATIONS
6 5%; AV = –5 V 6 5%; AGND = DGND = 0 V; V = 2.5 V; Filter Cutoff = 146 Hz; Noise Measurement Bandwidth = 146 Hz; A Source
CLKIN
SS
DD
REF
Resistance = 750 V2 with 1 nF to AGND at each AIN. TA = TMIN to TMAX , unless otherwise noted.)
Parameter
STATIC PERFORMANCE
Resolution
Integral Linearity Error
Gain Error
Gain Match Between Channels
Gain TC
Offset Error
Offset Match Between Channels
Offset TC
Noise
DYNAMIC PERFORMANCE
Sampling Rate
Output Update Rate
Filter Cutoff Frequency
Settling Time
Usable Dynamic Range4
Total Harmonic Distortion
Absolute Group Delay3
Differential Group Delay3
Channel-to-Channel Isolation
ANALOG INPUT
Input Range
Input Capacitance
Input Bias Current
B Version
Units
22
0.003
0.006
1
0.5
30
0.2
0.1
4
11
Bits
% FSR typ
% FSR max
% FSR max
% FSR max
µV/°C typ
% FSR max
% FSR max
µV/°C typ
µV rms max
fCLKIN/14
fCLKIN/(14 3 256 3 2N )
fCLKIN/(3.81 3 14 3 256 3 2N)
(3 3 14 3 256 3 2N /fCLKIN )
See Table I
–90
–100
(3 3 14 3 256 3 2N )/2fCLKIN
10
–85
Test Conditions/Comments
Guaranteed No Missed Codes to 21 Bits3
See Table I for Typical Noise Performance vs. Programmed
Cutoff Frequency
570 kHz for fCLKIN = 8 MHz
N Is Decimal Equivalent of FC2, FC1, FC0 in Control Register
dB typ
dB typ
ns typ
dB typ
Input Frequency = 35 Hz
AIN = ± 10 mV p-p
Feedthrough from Any One Channel to the Other Three, with
35 Hz Full-Scale Sine Wave Applied to that Channel
± 2.5
10
1
Volts
pF typ
nA typ
2.4
0.8
V min
V max
+10/-130
+10/-650
± 10
10
µA max
µA max
µA max
pF max
Internal 50 kΩ Pull-Up Resistors
Internal 10 kΩ Pull-Up Resistor
LOGIC OUTPUTS
VOH, Output High Voltage
VOL, Output Low Voltage
2.4
0.4
V min
V max
|IOUT| ≤ 40 µA
|IOUT| ≤ 1.6 mA
POWER SUPPLIES
Reference Input
AV DD
DVDD
AV SS
IDD
ISS
Power Consumption
Power Supply Rejection5
2.4/2.6
4.75/5.25
4.75/5.25
–4.75/–5.25
7.5
2.5
50
–70
V min/V max
V min/V max
V min/V max
V min/V max
mA max
mA max
mW max
dB typ
LOGIC INPUTS
VINH, Input High Voltage
VINL, Input Low Voltage
IIN, Input Current
SDATA, RFS
TFS
All Other Inputs
CIN, Input Capacitance3
DD
IN
4.8 mA typ
1.8 mA typ
35 mW typ
NOTES
1
Operating temperature ranges as follows : B Version; –40°C to +85°C.
2
The AIN pins present a very high impedance dynamic load which varies with clock frequency.
3
Guaranteed by design and characterization. Digital filter has linear phase.
4
Usable dynamic range is guaranteed by measuring noise and relating this to the full-scale input range.
5
100 mV p-p, 120 Hz sine wave applied to each supply.
Specifications subject to change without notice.
–2–
REV. A
AD7716
Table I. Typical Usable Dynamic Range, RMS Noise and Filter Settling Time vs. Filter Cutoff Frequency
N
0
1
2
3
4
Programmed Cutoff
Frequency (Hz)
584
292
146
73
36.5
Output Update
Rate (Hz)
2232
1116
558
279
140
Usable Dynamic
Range (dB)
99
102
105
108
111
RMS Noise
(mV)
21
14
10
7
5
Filter Settling Time to
60.0007% FS (ms)
1.35
2.7
5.4
10.8
21.6
Absolute Group
Delay (ms)
0.675
1.35
2.7
5.4
10.8
NOTE
Usable Dynamic Range is defined as the ratio of the rms full-scale reading (sine wave input) to the rms noise of the converter.
CONTROL REGISTER TIMING CHARACTERISTICS1, 2 (AV
DD = DV DD = +5 V 6 5%; AVSS = –5 V 6 5%; AGND =
DGND = 0 V; fCLKIN = 8 MHz; Input Levels: Logic 0 = 0 V, Logic 1 = DVDD; unless otherwise noted)
Parameter
Limit at TMIN, T MAX
(B Version)
Units
Conditions/Comments
t1
t2
t3
t4
t5
t6
1/fCLKIN
77
30
20
10
20
ns min
ns min
ns min
ns min
ns min
ns min
SCLK Period
SCLK Width
TFS Setup Time
SDATA Setup Time
SDATA Hold Time
TFS Hold Time
NOTES
1
Sample tested at +25°C to ensure compliance. All input signals are specified with tr = tf = 5 ns (10% to 90% of 5 V) and timed from a voltage level of 1.6 V.
2
See Figure 2.
3
CLKIN Duty Cycle range is 40% to 60%.
1.6mA
TO
OUTPUT
PIN
IOL
+2.1V
CL
50pF
200µA
IOH
Figure 1. Load Circuit for Access Time and Bus Relinquish Time
t2
t1
SCLK (I)
t2
TFS (I)
t6
t3
t4
t5
SDATA (I)
DB0
(DB8)
DB1
(DB9)
DB2
(DB10)
DB3
(DB11)
DB4
(DB12)
DB5
(DB13)
DB6
(DB14)
Figure 2. Control Register Timing Diagram
REV. A
–3–
DB7
(DB15)
AD7716
MASTER MODE TIMING CHARACTERISTICS1, 2 (AV
DD = DVDD = +5 V 6 5%; AV SS = –5 V 6 5%; AGND = DGND = 0 V;
fCLKIN = 8 MHz; Input Levels: Logic 0 = 0 V, Logic 1 = DVDD; unless otherwise noted)
Parameter
fCLKIN3, 4
tr 5
tf 5
t7
t8
t9
t10
t11
t12
t13
t14
t15
t166
t177
t18
t19
t20
t21
Limit at TMIN, T MAX
(B Version)
Units
Conditions/Comments
400
8
40
40
1/fCLKIN
1/fCLKIN
1/2fCLKIN + 30
50
40
50
1/fCLKIN
40
1/fCLKIN
45
1/2fCLKIN + 50
1/2fCLKIN + 10
1/2fCLKIN + 60
50
20
1/2fCLKIN + 50
2/fCLKIN
kHz min
MHz max
ns max
ns max
ns min
ns min
ns max
ns max
ns max
ns min
ns
ns max
ns
ns max
ns max
ns min
ns max
ns max
ns min
ns max
ns
CLKIN Frequency
Digital Output Rise Time. Typically 20 ns
Digital Output Fall Time. Typically 20 ns
CASCIN Pulse Width
CASCIN to DRDY Setup Time
DRDY Low to SCLK Low Delay
CLKIN High to DRDY Low, SCLK Active, RFS Active
CLKIN High to SCLK High Delay
SCLK Width
SCLK Period
SCLK High to RFS High Delay
RFS Pulse Width
SCLK High to SDATA Valid Delay
SCLK Low to SDATA High Impedance Delay
CLKIN High to DRDY High Delay
CLKIN High to RFS High Impedance, SCLK High Impedance
SCLK Low to CASCOUT High Delay
CASCOUT Pulse Width
NOTES
1
Sample tested at +25°C to ensure compliance. All input signals are specified with tr = tf = 5 ns (10% to 90% of 5 V) and timed from a voltage level of 1.6 V.
2
See Figures 1 and 3.
3
CLKIN duty cycle range is 40% to 60%.
4
The AD7716 is production tested with f CLKIN at 8 MHz in the slave mode. It is guaranteed by characterization to operate at 400 kHz and 8 MHz in master mode.
5
Specified using 10% and 90% points on waveform of interest.
6
t16 is measured with the load circuit of Figure 1 and defined as the time required for an output to cross 0.8 V or 2.4 V.
7
t17 is derived from the measured time taken by the data outputs to change 0.5 V when loaded with the circuit of Figure 1. The measured number is then extrapolated
back to remove the effects of charging or discharging the 100 pF capacitor. This means that the time quoted in the timing characteristics is the true bus relinquish
time of the part and as such is independent of external bus loading capacitances.
t7
CASCIN (I)
t8
CLKIN (I)
t9
t18
t 11
DRDY (O)
t 10
t 12
t19
t 13
SCLK (O)
t 12
t 14
t19
t 15
RFS (O)
t17
t 16
SDATA (O)
DB31
CH1
DB30
CH1
DB29
CH1
DB25
CH1
DB24
CH1
DB23
CH1
DB2
CH4
DB1
CH4
DB0
CH4
t21
t20
CASCOUT (O)
Figure 3. Master Mode Timing Diagram
–4–
REV. A
AD7716
SLAVE MODE TIMING CHARACTERISTICS1, 2 (AV
DD = DV DD = +5 V 6 5%; AVSS = –5 V 6 5%; AGND = DGND = 0 V;
fCLKIN = 8 MHz; Input Levels: Logic 0 = 0 V, Logic 1 = DVDD; unless otherwise noted)
Parameter
(B Version)
Units
Conditions/Comments
fCLKIN3, 4
400
8
40
40
1/fCLKIN
50
125
1/fCLKIN +30
30
50
50
50
0
60
2/fCLKIN
kHz min
MHz max
ns max
ns max
ns min
ns min
ns min
ns min
ns min
ns max
ns min
ns max
ns min
ns max
ns max
CLKIN Frequency
tr 5
tf 5
t23
t24
t25
t26
t27
t286
t29
t307
t31
t32
Digital Output Rise Time. Typically 20 ns
Digital Output Fall Time. Typically 20 ns
CASCIN Pulse Width
SCLK Width
SCLK Period
CASCIN High to RFS Setup Time
RFS Low to SCLK High Setup Time
SCLK High to SDATA Valid Delay
RFS Hold Time After SCLK High
SCLK High to SDATA High Impedance Delay
SCLK High to CASCOUT High Delay.
CASCOUT Pulse Width
NOTES
1
Sample tested at +25°C to ensure compliance. All input signals are specified with tr = tf = 5 ns (10% to 90% of 5 V) and timed from a voltage level of 1.6 V.
2
See Figures 1 and 4.
3
CLKIN duty cycle range is 40% to 60%.
4
The AD7716 is production tested with f CLKIN at 8 MHz in the slave mode. It is guaranteed by characterization to operate at 400 kHz.
5
Specified using 10% and 90% points on waveform of interest.
6
t28 is measured with the load circuit of Figure 1 and defined as the time required for an output to cross 0.8 V or 2.4 V.
7
t30 is derived from the measured time taken by the data outputs to change 0.5 V when loaded with the circuit of Figure 1. The measured number is then extrapolated
back to remove the effects of charging or discharging the 100 pF capacitor. This means that the time quoted in the timing characteristics is the true bus relinquish
time of the part and as such is independent of external bus loading capacitances.
t23
CASCIN (I)
t 24
t 25
SCLK (I)
t 24
t 26
t29
t 27
RFS (I)
t 28
SDATA (O)
t30
DB31
CH1
DB30
CH1
DB29
CH1
DB28
CH1
DB27
CH1
DB2
CH4
DB1
CH4
t31
CASCOUT (O)
Figure 4. Slave Mode Timing Diagram
REV. A
–5–
DB0
CH4
t32
AD7716
ABSOLUTE MAXIMUM RATINGS 1
PQFP Package, Power Dissipation . . . . . . . . . . . . . . . 450 mW
θJA Thermal Impedance . . . . . . . . . . . . . . . . . . . . . . 95°C/W
Lead Temperature, Soldering
Vapor Phase (60 sec) . . . . . . . . . . . . . . . . . . . . . . +215°C
Infrared (15 sec) . . . . . . . . . . . . . . . . . . . . . . . . . . +220°C
PLCC Package, Power Dissipation . . . . . . . . . . . . . . . 500 mW
θJA Thermal Impedance . . . . . . . . . . . . . . . . . . . . . . 55°C/W
Lead Temperature, Soldering
Vapor Phase (60 sec) . . . . . . . . . . . . . . . . . . . . . . +215°C
Infrared (15 sec) . . . . . . . . . . . . . . . . . . . . . . . . . . +220°C
(TA = +25°C unless otherwise noted)
AVDD to AGND . . . . . . . . . . . . . . . . . . . . . . . . –0.3 V to +7 V
AVSS to AGND . . . . . . . . . . . . . . . . . . . . . . . +0.3 V to –7 V
AGND to DGND . . . . . . . . . . . . . . . . . . . . . –0.3 V to +0.3 V
AVDD to DVDD . . . . . . . . . . . . . . . . . . . . . . . –0.3 V to +0.3 V
Analog Inputs to AGND . . . . . . AVSS – 0.3 V to AV DD + 0.3 V
VREF to AGND . . . . . . . . . . . . AVSS – 0.3 V to AVDD + 0.3 V
Digital Inputs to DGND2 . . . . . . . . . . –0.3 V to DV DD + 0.3 V
Digital Outputs to DGND . . . . . . . . . . –0.3 V to DV DD + 0.3 V
Operating Temperature Range
Commercial Plastic (B Versions) . . . . . . . . . . . –40°C to +85°C
Storage Temperature Range . . . . . . . . . . . . . –65°C to +150°C
NOTES
1
Stresses above those listed under “Absolute Maximum Ratings” may cause
permanent damage to the device. This is a stress rating only and functional
operation of the device at these or any other conditions above those listed in the
operational sections of this specification is not implied. Exposure to absolute
maximum rating conditions for extended periods may affect device reliability.
2
Transient currents of up to 100 mA will not cause SCR latch-up.
CAUTION
ESD (electrostatic discharge) sensitive device. Electrostatic charges as high as 4000 V readily
accumulate on the human body and test equipment and can discharge without detection.
Although these devices feature proprietary ESD protection circuitry, permanent damage may
occur on devices subjected to high energy electrostatic discharges. Therefore, proper ESD
precautions are recommended to avoid performance degradation or loss of functionality.
33 MODE
39 MODE
38 NC
9
37 NC
31 NC
DGND 10
DGND 4
30 DVDD
36 DV DD
NC 11
29 D IN 1
TOP VIEW
(Not to Scale)
SCLK 13
27 CASCIN
26 CASCOUT
35 DIN 1
AD7716
RFS 12
28 NC
NC
44 43 42 41 40
SDATA
1
NC
TFS
DGND
2
DRDY
CLKOUT
DOUT 2
CLKIN
NC
3
8
32 NC
RESET 8
4
7
NC 2
TOP VIEW
(Not to Scale)
5
NC
D OUT 1 3
RFS 6
6
NC
DOUT 1
SCLK 7
NC
35 SDATA
34 NC
37 NC
36 DRDY
39 TFS
38 DGND
41 CLKOUT
40 D OUT 2
42 NC
44 NC
43 CLKIN
NC 1
AD7716
ESD SENSITIVE DEVICE
PLCC PINOUT
PQFP PINOUT
NC 5
WARNING!
34 NC
33
RESET 14
CASCIN
32 CASCOUT
AGND 9
25 VREF
AGND 15
31 VREF
AVDD 10
24 AVSS
AV DD 16
30 AV SS
AIN 1 11
23 AGND
AIN 1 17
29
AGND
A0
AIN 4
A1
AGND
AIN 3
AGND
AIN 2
AGND
AGND
A2
AGND
A0 22
AIN4 21
A1 19
AGND 20
AIN 3 17
AGND 18
AGND 16
AIN 2 15
AGND 14
A2 13
AGND 12
18 19 20 21 22 23 24 25 26 27 28
NC = NO CONNECT
NC = NO CONNECT
ORDERING GUIDE
Model
Temperature
Range
Output Noise
(Filter: 146 Hz)
Package
Option*
AD7716BP
AD7716BS
–40°C to +85°C
–40°C to +85°C
11 µV rms
11 µV rms
P-44A
S-44
*P = PLCC (Plastic Leaded Chip Carrier); S = PQFP (Plastic Quad Flatpack).
–6–
REV. A
AD7716
PIN DESCRIPTION
Pin
Description
AVDD
Analog Positive Supply, +5 V Nominal. This supplies +ve power to the analog modulators. AVDD & DV DD
must be tied together externally.
Digital Positive Supply, +5 V Nominal. This supplies +ve power to the digital filter and input/output registers.
Analog Negative Supply, –5 V nominal. This supplies –ve power to the analog modulators.
A high pulse on this input pin synchronizes the sampling point on the four input channels. It can be used in a
multichannel system to ensure simultaneous sampling. This also resets the digital interface to a known state.
The three address input pins, A0, A1 and A2 give the device a unique address. This information is contained in
the output data stream from the device.
Clock Input for External Clock.
Clock Output which is used to generate an internal master clock by connecting a crystal between CLKOUT and
CLKIN. If an external clock is used then CLKOUT is not connected.
This digital input determines the device interface mode. If it is hardwired low, then the Master Mode interface is
enabled whereas if it is high, the Slave Mode interface is enabled.
This is an active-high, level-triggered digital input which is used to enable the output data stream. This input
may be used to cascade several devices in a multichannel system.
Digital output which goes high at the end of a complete 4-channel data transfer. This can be connected to the
CASCIN of the next device in a multichannel system to ensure proper control of the data transfer.
Receive Frame Synchronization signal for the serial output data stream. This can be an input or output depending
on the interface mode.
Serial Data Input/Output Pin.
Serial Clock Input/Output. The SCLK pin is configured as an input or output, depending on the state of the
Mode pin.
Data Ready Output. A falling edge indicates that a new word is available for transmission. It will return high
when 4, 32-bit words have been transmitted. It also goes high for one clock cycle, when a new word is being
loaded into the output register. Data should not be read during this period.
Transmit Frame Sync input for programming the on-chip Control Register.
Digital Data Input. This is contained in the digital data stream sent from the device.
Digital Outputs. These two digital outputs can be programmed from the on-chip Control Register. They can
be used to control calibration signals at the front end.
Reference Input, Nominally 2.5 V.
Analog Ground. Ground reference for analog circuitry.
Digital Ground. Ground return for digital circuitry.
Analog Input Pins. The analog input range is ± 2.5 V.
DVDD
AVSS
RESET
A0–A2
CLKIN
CLKOUT
MODE
CASCIN
CASCOUT
RFS
SDATA
SCLK
DRDY
TFS
DIN1
DOUT1, DOUT2
VREF
AGND
DGND
AIN1–AIN4
REV. A
–7–
AD7716
TERMINOLOGY
OUTPUT UPDATE RATE
LINEARITY ERROR
This is the rate at which the digital filter updates the output shift
register. It is a function of the master clock frequency and the
programmed filter cutoff frequency.
This is the maximum deviation of any code from a straight line
passing through the endpoints of the transfer function. The
endpoints of the transfer function are zero scale (not to be confused with Bipolar Zero), a point 0.5 LSB below the first code
transition (000 . . . 000 to 000 . . . 001) and full scale, a point
0.5 LSB above the last code transition (111 . . . 110 to
111 . . . 111). The error is expressed as a percentage of full
scale.
FILTER CUTOFF FREQUENCY
The digital filter of the AD7716 can be programmed, in binary
steps, to 5 discrete cutoff frequencies, ranging from 584 Hz to
36.5 Hz (for a CLKIN frequency of 8 MHz).
SETTLING TIME
This is the settling time of the on-chip digital filter, to 0.0007%
of FSR, in response to a full-scale step at the input of the ADC.
It is proportional to the master clock frequency and the filter
cutoff frequency.
DIFFERENTIAL LINEARITY ERROR/NO MISSED CODES
This is the difference between any code’s actual width and the
ideal (1 LSB) width. Differential Linearity Error is expressed in
LSBs. A differential linearity specification of ±1 LSB or less
guarantees no missed codes to the full resolution of the device.
The AD7716 has no missed codes guaranteed to 21 bits with a
cutoff frequency of 146 Hz.
USABLE DYNAMIC RANGE
GAIN TC
The usable dynamic range is the ratio of the rms full-scale
reading (sine wave input) to the rms noise of the converter,
expressed in dBs. It determines the level to which it is possible
to resolve the input signal. For example, at a bandwidth of
146 Hz, the rms noise of the converter is 11 µV. The full-scale
rms is 1.77 volts. So, the usable dynamic range is 104 dB. Any
signal below this level will be indistinguishable from noise unless
extra post-filtering techniques are employed.
This is the variation of gain error with temperature and is expressed in µV/°C.
TOTAL HARMONIC DISTORTION
OFFSET ERROR
Total harmonic distortion (THD) is the ratio of the rms sum
of the harmonics to the fundamental. For the AD7716, it is
defined as:
GAIN ERROR
Gain Error is the deviation of the last code transition
(111 . . . 110 to 111 . . . 1) from the ideal (VREF –3/2 LSBs). It
is expressed as a percentage of full scale.
Offset Error is the deviation of the first code transition from the
ideal (–VREF + 0.5 LSB). It is expressed as a percentage of full
scale.
2
THD (dB) = 20 log
OFFSET TC
This is the variation of offset error with temperature and is expressed in µV/°C.
2
2
2
V 2 +V 3 +V 4 +V 5 +V 6
V1
2
where V1 is the rms amplitude of the fundamental and V2, V 3,
V4, V5 and V6 are the rms amplitudes of the second through
sixth harmonics.
NOISE
This is the converter rms noise expressed in µV. Because of the
digital filtering in the sigma delta converter, the noise performance is a function of the programmed filter cutoff.
ABSOLUTE GROUP DELAY
SAMPLING RATE
Absolute group delay is the rate of change of phase versus frequency, dφ/df and is expressed in seconds. For the AD7716,
it is dependent on master clock frequency and filter cutoff
frequency.
This is the modulator sampling rate. For the AD7716, it is
fCLKIN/14.
DIFFERENTIAL GROUP DELAY
Differential group delay is the total variation in absolute group
delay in the specified bandwidth. Since the digital filter in the
AD7716 has perfectly linear phase, the differential group delay
is almost zero. This is important in many signal processing applications where excessive differential group delay can cause
phase distortion.
–8–
REV. A
AD7716
The output of the comparator provides the digital input for the
1-bit DAC, so the system functions as a negative feedback loop
which minimizes the difference signal. The digital data that represents the analog input voltage is in the duty cycle of the pulse
train appearing at the output of the comparator. It can be retrieved as a parallel binary data word using a digital filter.
GENERAL DESCRIPTION
The AD7716 is a 4-channel 22-bit A/D converter with on-chip
digital filtering, intended for the measurement of wide dynamic
range, low frequency signals such as those representing ECG,
EEG, chemical, physical or biological processes. It contains
four sigma delta ADCs, a clock oscillator and a serial communications port.
C
Each of the analog input signals to the AD7716 is continuously
sampled at a rate determined by the frequency of the master
clock, CLKIN. Four sigma-delta modulators convert the
sampled signals into digital pulse trains whose duty cycles contain the digital information. These are followed by low-pass filters to process the output of the modulators and update the
output register at a maximum rate of 2.2 kHz. The output data
can be read from the serial port at any rate up to this.
CLOCK
R
AIN
EN
INTEGRATOR
STROBED
COMPARATOR
TO
DIGITAL
FILTER
R
+V REF
–VREF
1-BIT DAC
THEORY OF OPERATION
Figure 5. First Order Modulator
The general block diagram of a delta-sigma ADC is shown in
Figure 5. It contains the following elements:
Sigma-delta ADCs are generally described by the order of the
analog low-pass filter. A simple example of a first order sigmadelta ADC is shown in Figure 5. This contains only a firstorder low-pass filter or integrator.
1. Continuously Sampling Integrator
2. A Differential Amplifier or Subtracter
3. A 1-Bit A/D Converter (Comparator)
The AD7716 uses a second-order sigma-delta modulator and a
digital filter that provides a rolling average of the sampled output. After power-up or if there is a step change in the input
voltage, there is a settling time before valid data is obtained.
4. A 1-Bit DAC
5. A Digital Low-Pass Filter
In operation, the sampled analog signal is fed to the subtracter,
along with the output of the 1-bit DAC. The filtered difference
signal is fed to the comparator, whose output samples the difference signal at a frequency many times that of the analog signal
frequency (oversampling).
DIGITAL FILTERING
The AD7716’s digital filter behaves like an analog filter, with a
few minor differences.
First, since digital filtering occurs after the A-to-D conversion
process, it can remove noise injected during the conversion process. Analog filtering cannot do this.
Oversampling is fundamental to the operation of delta-sigma
ADCs. Using the quantization noise formula for an ADC:
SNR = (6.02 3 number of bits + 1.76) dB,
On the other hand, analog filtering can remove noise superimposed on the analog signal before it reaches the ADC. Digital
filtering cannot do this and noise peaks riding on signals near
full scale have the potential to saturate the analog modulator
and digital filter, even though the average value of the signal is
within limits. If noise signals cause the input signal to exceed
the specified range, consideration should be given to analog input filtering, or to reducing the gain in the input channel to
bring the combination of signal and noise spike within the specified input range.
a 1-bit ADC or comparator yields an SNR of 7.78 dB.
When operating with a master clock of 8 MHz, the AD7716
samples the input signal at 570 kHz, which spreads the quantization noise from 0 kHz to 285 kHz. Since the specified analog
input bandwidth of the AD7716 is only 584 Hz maximum (it
can be programmed to be lower), the noise energy in this bandwidth would be only 1/488 of the total quantization noise, assuming that the noise energy was spread evenly throughout the
spectrum. This very high sampling with respect to the input
bandwidth is known as oversampling, and the ratio of 488:1 is
called the oversampling ratio. The noise is reduced still further
by analog filtering in the modulator loop, which shapes the
quantization noise spectrum to move most of the noise energy to
frequencies above 584 Hz. The SNR performance in the 0 Hz
to 584 Hz range is conditioned to the 99 dB level in this fashion
(see Table I). As the programmed bandwidth is reduced, the
oversampling ratio increases and the usable dynamic range also
increases. Thus, for example, with a programmed bandwidth
of 73 Hz, the oversampling ratio is 3904:1, and the usable dynamic range is 108 dB which corresponds to greater than 17-bit
resolution.
REV. A
Filter Characteristics
The cutoff frequency of the digital filter is determined by bits
FC2, FC1 and FC0 in the control register (See Table IV). The
cutoff frequency of the filter is fCLKIN /(3.81 3 14 3 256 3 2N),
where N is the decimal equivalent of FC2, FC1, FC0. At the
maximum clock frequency of 8 MHz, with all 0s loaded to FC2,
FC1, FC0, the cutoff frequency of the filter is 584 Hz and the
data update rate is 2232 Hz.
Since the AD7716 contains low-pass filtering, there is a settling
time associated with step function inputs, and data will be invalid after a step change until the settling time has elapsed. The
–9–
AD7716
relationship between input bandwidth and settling is given in
Table I. Because of this settling time, most sigma delta ADCs
are unsuitable for high speed multiplexing, where channels are
switched and converted sequentially at high rates, as switching
between channels can cause a step change in the input. However, the AD7716 is a sigma-delta solution to multichannel applications, since it can process four channels simultaneously. In
addition, it is easy to cascade several devices in order to increase
the number of channels being processed.
0
–20
In the AD7716, the on-chip modulator provides the digital filter
with samples at a rate of 570 kHz. The filter decimates these
samples to provide data at an output rate which corresponds to
the programmed first notch frequency of the filter.
If the user wants to reduce the output noise from the device for
bandwidths less than 36.5 Hz, then it is possible to employ extra
filtering after the AD7716. This extra digital filtering is called
post filtering. If a straight averaging filter is used, for example, a
reduction in bandwidth by a factor of 2 results in √2 reduction
in the rms noise. This additional filtering will also result in a
longer settling time.
–40
Antialias Considerations
–60
The digital filter does not provide any rejection at integer multiples of the modulator sampling frequency (n 3 570 kHz,
where n = 1, 2, 3, . . .). This means that there are frequency
bands, ± f3dB wide (f3dB is the cutoff frequency selected by FC0
to FC2) where noise passes unattenuated to the output. However, due to the AD7716’s high oversampling ratio, these bands
occupy only a small fraction of the spectrum and most broadband noise is filtered.
–80
GAIN – dB
Post Filtering
–100
–120
–140
–160
–180
–200
In spectral analysis applications, it is important to note that attenuation at half the output update rate is 16 dB. Extra frontend filtering or post filtering may be required to keep aliases in
this frequency band at an acceptable level.
–220
–240
0
73
278
556
834
1112
FREQUENCY – Hz
1390
1668
Figure 6. Frequency Response of AD7716 Filter
Figure 6 shows the filter frequency response for a cutoff frequency of 73 Hz. This is a (sinx/x)3 response (also called sinc3)
that provides greater than 100 dB rejection at the notch frequencies. The relationship between the programmed cutoff
frequency and the first notch is constant (fNOTCH = 3.81 3
fCUTOFF). The first notch frequency is also the output data rate.
The settling time to a full-scale step input is four times the output data period. Programming a different cutoff frequency via
FC0–FC2 does not alter the profile of the filter response, it simply changes the frequency of the notches.
In Figure 6, the first notch is at 278 Hz. This is also the output
data rate. Settling time to a full-scale step input is 10.8 ms.
The digital filter can be defined by the following equations.
 1 1– Z – N 
H (z) =  ×
–1 
 N 1– Z 
H(f)=
sinπf / f S )
πf / f S )
USING THE AD7716
SYSTEM DESIGN CONSIDERATIONS
The AD7716 operates differently from successive approximation ADCs or other integrating ADCs. Since it samples the signal continuously, like a tracking ADC, there is no need for a
start convert command. The output register is updated at a rate
dependent on the programmed cutoff frequency, and the output
can be read at any time.
Input Signal Conditioning
The input range for the AD7716 is ± VREF, where VREF = 2.5 V
± 10%. Other input ranges can be accommodated by input signal conditioning. This may take the form of gain to increase a
smaller signal range, or passive attenuation to reduce a larger input voltage range.
3
3
where N = Ratio of Modulator Sampling Frequency to Output
Rate
and
fS = Output Rate.
–10–
REV. A
AD7716
Source Resistance
–0.125
Active signal conditioning circuits such as op amps generally do
not suffer from problems of high source impedance. Their
open-loop output resistance is normally only tens of ohms and,
in any case, most modern general purpose op amps have
sufficiently fast closed-loop settling time for this not to be a
problem.
–0.375
–0.500
–0.625
20
30
40
50
60
70
TEMPERATURE – °C
80
90
Voltage Reference
The voltage applied to the VREF pin defines the analog input
range. The specified reference voltage is 2.5 V ± 10%.
The reference input presents exactly the same dynamic load as
the analog input, but in the case of the reference input, source
resistance and long settling time introduce gain errors rather
than offset errors. Most precision references however have sufficiently low output impedance and wide enough bandwidth to
settle to the required accuracy in the time allowed by the
AD7716.
Sigma-delta ADCs, like VFCs and other integrating ADCs, do
not contain any source of nonmonotonicity and inherently offer
no missing codes performance.
The AD7716 achieves excellent linearity by the use of high
quality, on-chip silicon dioxide capacitors, which have a very
low capacitance/voltage coefficient.
Drift Considerations
The AD7716 uses autozeroing techniques to minimize input
offset drift. Charge injection in the analog switches and leakage
currents at the sampling node are the primary sources of offset
voltage drift in the converter. Figure 7 indicates the typical offset due to temperature changes. Drift is relatively flat up to
85°C. Above this temperature, leakage current becomes the
main source of offset drift. Since leakage current doubles approximately every 10°C, the offset drifts accordingly. The value
of the voltage on the sample capacitor is updated at a rate determined by the master clock, therefore the amount of offset drift
which occurs will be proportional to the elapsed time between
samples.
REV. A
–0.25
Figure 7. Typical Offset Drift
Accuracy
Gain drift within the converter depends mainly upon the temperature tracking of the internal capacitors. It is not affected by
leakage currents.
OFFSET VOLTAGE – mV
If passive attenuators are used in front of the AD7716, care
must be taken to ensure that the source impedance is sufficiently low. The dc input resistance for the AD7716 is greater
than 1 GΩ. In parallel with this there is a small sampling capacitor. The dynamic load presented by this varies with the
clock frequency. The modulator sampling rate determines the
amount of time available for the sampling capacitor to be
charged. Any extra external impedances result in a longer overall charge time resulting in extra gain errors on the analog input.
The AD7716 has a quite large gain error (1% FSR) due to the
fact that there is no on-chip calibration. Thus, even an extra
10 kΩ source resistance and 50 pF source capacitance will have
no significant effect on this.
The reference should be chosen to have minimal noise in the
programmed passband. Recommended references are the
AD780 or the REF43 from Analog Devices. These low noise
references have typical noise spectral densities of 100 nV/√Hz at
600 Hz. This corresponds to an rms noise of 2.5 µV in this
band and is more than adequate for the AD7716.
Clock Generation
The device operates from a master clock which must be provided either from a crystal source or an external clock source. If
a crystal is used, it must be connected across the CLKIN and
CLKOUT pins. Typical loading capacitors of 15 pF are required on CLKIN, CLKOUT. The crystal manufacturers data
should be consulted for more information. An external clock
can also be used to drive the CLKIN input directly with a
CMOS compatible clock. In this case, CLKOUT is left unconnected. The nominal clock frequency for the device is 8 MHz.
–11–
AD7716
CONTROL REGISTER DESCRIPTION
The 16-bit control register is programmed in two 8-bit bytes;
the low byte is programmed first and the high byte second. The
loading format is LSB first (DB0 for the Least Significant Byte;
DB8 for the Most Significant Byte). Three control lines are
used: TFS, SCLK and SDATA. On initial application of
power to the AD7716, the control register will come up in an
undetermined state. Programming the control register requires
an SCLK input, a TFS input and an SDATA input. The
MODE pin on the device determines whether it is in the master
interface mode or the slave interface mode. In either mode, a
falling edge on TFS causes the part to relinquish control of the
SDATA and SCLK lines. When TFS goes low, data on the
SDATA line is clocked into the control register on each succeeding falling edge of SCLK. When 8 bits have been clocked
in, the transfer automatically stops. Only when another negative
going edge is detected on TFS will new information be written
into the control register. The control register programming
model is shown in Table II. Bits DB8 and DB0 allow the control register to identify whether the MS Byte or the LS Byte has
been programmed. Only when DB8 is a 1 and DB0 is a 0 will
the register recognize that a complete valid word has been
programmed.
Control register bit, DB15 (A3), acts as an extra address bit
which must always be set to 1 to enable programming of the
AD7716. If it is set to 0, then the programmed word is ignored.
This allows the user to bypass the AD7716 control register and
use the serial stream from the DSP or microcomputer to program other serial peripheral devices.
When a valid word has been received, the device interrogates
the M0 bit. If this is 0, then the digital filter cutoff frequencies
are programmed to the appropriate value if the device address
pins correspond to the A2, A1, A0 bits in the control register. If
the device address pins do not correspond to the A2, A1, A0
bits then the FC2, FC1, FC0 bits are ignored. If M0 is 1, then
the digital filter cutoff frequencies are programmed to the FC2,
FC1, FC0 value irrespective of the address bits. In a multichannel system this allows the user to either program all
AD7716s to have the same cutoff frequency or else to give each
device a separate cutoff frequency.
Control register bits FC2, FC1, FC0 program the digital filter
cutoff frequency, see Table VI.
Control register bits D2, D1 control the digital output pins D2
and D1. These are programmed in the same way as FC2, FC1,
FC0.
Table II. Control Register Programming Model
Most Significant Byte
DB15
A3
DB14
A2
DB13
A1
DB12 DB11 DB10
A0
M0
FC2
Least Significant Byte
DB9
FC1
DB8
1
DB7
FC0
DB6
DB5
DOUT2 DOUT1
DB4
X
DB3
X
DB2
X
DB1
X
DB0
0
Table IV. Cutoff Frequency Truth Table
Table III. M0 Truth Table
M0
Programming Mode
FC2
FC1
FC0
Cutoff Frequency (Hz)
0
A2, A1, A0 determine which device is addressed and
programmed with cutoff frequency and digital output.
A2, A1, A0 ignored. All devices are addressed and
programmed with common cutoff frequency and digital
output.
0
0
0
0
1
0
0
1
1
0
0
1
0
1
0
584
292
146
73
36.6
1
–12–
REV. A
AD7716
RESET
Master Mode Interface
The AD7716 has a hardware reset which can be used to synchronize many devices. When the RESET pin goes low after being
high for at least four CLKIN cycles, the modulator sampling
points and digital filter starting points are all synchronized. This
synchronizes all devices which receive the RESET pulse and
gives simultaneous sampling of all channels. It does not affect
the control register but restarts the interface. Also, it is necessary
to wait the requisite settling time after applying Reset to get valid
data from the device.
The device may be placed in the Master Mode by tying the
MODE pin low. In this mode, data is clocked out of the
AD7716 by an internally generated serial clock and frame synchronization pulse. Two signals initiate the transfer. These are
the input CASCIN and the internally generated DRDY signal.
When a high level is detected on CASCIN, the device checks
the state of DRDY. Note, that on initial power-up or after a reset has been applied, the CASCIN input is not necessary on device 000 for the first data transfer but is required thereafter. If
DRDY is low, then the 3-state output, RFS goes high on the
next rising edge of CLKIN and stays high for one CLKIN cycle
before going low again. The 3-state SCLK output is also activated on the same rising edge. As RFS goes low, DB31 is
clocked out on the rising edge of SCLK and is valid on the falling edge of SCLK. Data is transmitted in 8-bit bytes. For each
AIN, there are 4, 8-bit bytes and 4 RFS pulses. When DB0 of
AIN 4 has been clocked out, SCLK goes back into 3-state and
the CASCOUT output goes high for one master clock cycle.
DRDY also goes high at this point. Successive devices can be
networked together by tying the CASCOUT of one device to
the CASCIN on the next one.
CASCADING DEVICES
The AD7716 provides a facility for connecting multiple devices
in series. The CASCIN and CASCOUT pins allow this. Connecting CASCOUT to CASCIN of the succeeding device means
that the SDATA output of the second device will be disabled until the output register of the first device is empty.
In the case of the first device in the system, it is possible to drive
CASCIN from CASCOUT of the last device or, alternatively,
invert DRDY to drive it. If CASCIN is driven by CASCOUT,
then a reset must be applied after every write to the control
register. This also applies in single device systems that use
CASCOUT to drive CASCIN
DATA OUTPUT INTERFACE MODES
When the control register has been programmed, the device begins conversion. There is an initial delay to allow the digital filters to settle. As already stated, these filters are Sinc3, and so the
filter output update rate is directly related to the programmed
cutoff frequency. The ratio between these is 3.81. So, for a filter
cutoff frequency of 584 Hz, the output update is 2.22 kHz. The
falling edge of the DRDY output indicates that the output shift
register has been updated. There are two interface modes. One
is the master mode, where the AD7716 is the master in the system and the processor to which it is communicating is the slave.
The other mode is the slave mode, where the AD7716 is the
slave and the processor is the system master. In both of these
modes the data output stream contains 4 3 32 bits, corresponding to the four input channels. The output data format is given
in Table V. The conversion result DB21–DB0 occupies location
DB31–DB10 of the output register. DB21 is the MSB and is
transmitted first as shown in the timing diagrams. The channel
address is given by CA0 and CA1 which occupy DB9 and DB8
of the output register. The channel address format is given in
Table VI.
Note that on device 0 (A2, A1, A0 tied low), the CASCIN input
should be driven by the inverse of the DRDY output. This is
shown in the interface diagram of Figure 8.
The Master Mode interface is very suitable for loading data into
a serial-to-parallel shift register or for DSPs which can accept a
continuous stream of 8-bit bytes.
Slave Mode Interface
The device may be placed in the slave interface mode by tying
the MODE pin high. In this mode, the master processor controls the transfer of data from the signal processing block. It
starts the transfer by sending a frame synchronizations pulse and
serial clock to the AD7716. This could be in response to an interrupt generated by the DRDY output on the AD7716. If the
device has detected a high level on CASCIN or is device 000 on
its first transfer, it starts to send out data on the next rising edge
of SCLK. This data is then valid on the falling edge of SCLK.
When all the data bits have been clocked out, the CASCOUT
pin goes high for one CLKIN cycle and DRDY also goes high.
The slave mode interface is suited to both microcomputers like
the 8051 and 68HC11 and also DSPs like the TMS320C25,
ADSP-2101 family and the DSP56000 family.
Table V. Output Data Word Format
DB31 . . . DB10
DB9
DB8
DB21 . . . DB0
Conversion Result
CA0
CA1
Channel Address
DB7
DB6 DB5
A0
A1 A2
Device Address
Table VI. Channel Address Format
Channel
CA1 (DB8)
CA0 (DB9)
AIN 1
AIN 2
AIN 3
AIN 4
0
0
1
1
0
1
0
1
REV. A
–13–
DB4
DB3
DB2
DB1
DB0
DIN1
Pace Detect
OVFL
Overflow
X
X
X
Indeterminate
AD7716
MICROPROCESSOR INTERFACING
Interfacing the AD7716 to the ADSP-2100 Family
The SPORT0 Control Register setting is “7FCF.” This sets the
receive section for internal SCLK, continuous receive with alternate inverted framing.
The ADSP-2100 family of microcomputers from Analog Devices are high speed, high performance digital signal processors.
Many members of the family have serial ports (known as
SPORTs) which are compatible with the AD7716. These include the ADSP-2101, ADSP-2105, ADSP-2111 and ADSP2115. Full details of these are available in the ADSP-2100
Family User’s Manual available from Analog Devices.
The SPORT0 SCLKDIV Register (0 3 3FF5) determines the
SCLK frequency from the ADSP-2101. With “0000” loaded,
the SCLK output is at its maximum (1/2 the master clock of
12.5 MHz).
Figure 8 shows the hardware interface between two AD7716s
and SPORT 0 of the ADSP-2101 DSP. This yields a very efficient 8-channel data acquisition system. The AD7716 is set up
for slave interface mode by tying the MODE pin high. This
means that the ADSP-2101 is the master in the system and supplies the necessary frame synchronization and SCLK Signals to
the AD7716s when writing to and reading from the device.
On power up, the user should write to the AD7716 control register in order to set the filter cutoff frequencies. The appropriate SPORT 0 Control Register (0 3 3FF6) setting is “7EC7.”
This sets the transmit section for alternate inverted framing with
a word length of 8 bits. Two 8-bit words should then be written
to each AD7716 to program the filter cutoff frequencies. The
control register programming model is given in Table II. Note
that the LSB (DB0) must be loaded first as in the timing diagram of Figure 2.
When the write operation is complete, a reset pulse should be
applied to both devices. This ensures that the sampling and interface timing of the device are synchronized. The reset can be
under DSP control, in which case a flag output could be used.
After reset, the processor should jump to the read routine. For
this read routine, there are several registers that need to be set.
In normal operation, a SPORT generates an interrupt when it
has received a data word. Autobuffering provides a mechanism
for receiving or transmitting an entire block of serial data before
an interrupt is generated. Service routines can operate on the
entire block of data, rather than on a single word, reducing overhead significantly. This is ideal for use with a device like the
AD7716 where there is a requirement to read many bits of data
(256 in this case) for each sampling instant. The SPORT0
Autobuffer Control Register (0 3 3FF3) is loaded with “0001”
to enable the Receive Autobuffering.
The SPORT0 RFSDIV Register (0 3 3FF4) should be set to
the minimum value of “000F.” Finally the IRQ2 interrupt
should be enabled.
The DSP will now wait for an interrupt from the AD7716. This
interrupt is generated by the AD7716 DRDY line going low. If
the interrupt service routine is set for autobuffered mode with a
length of 16 (16-bit) words, then the DSP will read in the 256
bits from the two AD7716s in one continuous stream and then
stop. The data from the two devices will be contained in the
designated data memory area and the DSP can now go and operate on this as is necessary. Note that, because of the ADSP2101 framing, a one-bit shift left will be necessary on the data in
memory. For 16 data words, this will require 22 instruction
cycles.
+5V
A0 A1 A2
CASCIN
AD7716 #1
4.7kΩ
DRDY
IRQ2 (–VE EDGE TRIGGERED)
RFS
RFS
SCLK
RESET
+5V
CASCIN
SCLK
TFS
CASCOUT
ADSP-2101
4.7kΩ
SDATA
TFS
DT
MODE
DR
+5V
A0 A1 A2
SCLK
AD7716 #2
RESET
CONTROL
TFS
SDATA
RESET
DRDY
RFS
CASCOUT
MODE
+5V
Figure 8. 8-Channel Data Acquisition System Using the ADSP-2101 Digital Signal Processor
–14–
REV. A
AD7716
When the AD7716 is programmed for the maximum cutoff frequency of 584 Hz, the output data rate is 2.25 kHz. This
means that there is 440 µs available to read the data and do the
necessary number crunching before the next sample must be
read. Assuming that the ADSP-2101 is running from a master
clock of 12.5 MHz means that the maximum SCLK available is
6.25 MHz (1/2 the master clock frequency). It will then take 40
µs to transfer the 256 bits of data from the ADC to the DSP.
This leaves 400 µs for number crunching in the DSP. If the
programmed cutoff frequency is lower then this allows even
more time to the DSP.
The optimum setup for reading all four channels of the AD7716
into the DSP56001 is six 24-bit reads. This will provide 144 clock
edges to shift out the 128 bits of data in the AD7716 output shift
register. The first clock applied to the AD7716 will clock out
DB21 of AIN1. DRDY from the AD7716 can be used as an
interrupt input to the DSP56001 to control the data transfer.
Either IRQA or IRQB of the DSP56001 can be used to detect the
interrupt.
RESET
CONTROL
+5V
MC68HC11 Interface
MODE RESET CASCIN
The MC68HC11 microcomputer can be interfaced to the
AD7716 using the slave mode interface. Figure 9 shows a typical setup. The MODE pin on the AD7716 is tied high for slave
mode operation. The SPI port of the MC68HC11 is used. The
microcomputer is in its single chip mode. DRDY from the
AD7716 is connected to the IRQ input of the MC68HC11.
MISO and MOSI on the MC68HC11 should be configured for
wired-or operation. Depending on the interface configuration,
it may be necessary to provide bidirectional buffers between
these lines.
SC2
TFS
SC1
RFS
SCLK
DSP56001 SCK
AD7716
SC0
SDATA
STD
SRD
IRQ
DRDY
CASCOUT
Figure 10. DSP56001 Interface
+5V
+5V
TMS320C25 Interface
SS
PC0
PC1
PC2
IRQ
MODE
RFS
TFS
CASCIN
DRDY
AD7716
PC2
PC3
SCK
RESET
SCLK
MISO
MOSI
SDATA
MC68HC11
Figure 11 shows the AD7716 interfaced to the TMS320C25
DSP using the master mode interface. For initial programming
of the AD7716 control register, the external gated clock is required. FSX going low enables this. When the two 8-bit bytes
have been sent to the AD7716, the FSX should go permanently
high. The external gated clock will now be disabled and the
AD7716 will take control of the SDATA line. It will begin
transmitting data as soon as it becomes available. It also provides the clock and frame synchronization signals required by
the DSP.
CASCOUT
Figure 9. MC68HC11 to AD7716 Interface
The MC68HC11 is configured in the master mode with its
CPOL bit set to a logic zero and its CPHA bit set to a logic one.
With an 8 MHz CLKIN input on the AD7716, the device will
operate with all four serial clock rates of the MC68HC11.
Reset for the AD7716 is provided by one of the DSP flag
outputs.
Sixteen, 8-bit read operations are necessary to read the 128 bits
from the AD7716 output register. An extra read is necessary to
reset the output register. This means a total of 17 read operations are needed from the MC68HC11.
FLAG
OUTPUT
MODE RESET
TMS320C25
DR
DT
Figure 10 shows an interface to the DSP56001 digital signal
processor. The AD7716 is set up for the slave interface mode.
The DSP56001 is set up for asynchronous operation with gated
clock and normal framing. Data must be written to the
AD7716 control register in two 8-bit bytes. The first byte is
written to the DSP56001 SSI transmit data register (TX) and
this is automatically transferred to the transmit shift register
when the frame sync occurs. Data is shifted out to the STD pin
by the internal bit clock (SCK) when the associated frame sync
output is asserted.
–15–
CASCIN
RFS
FSR
DSP56001 Interface
REV. A
RESET CONTROL
SDATA
AD7716
TFS
FSX
SCLK
CLKR
CLKX
CASCOUT
GATED
EN
CLOCK
Figure 11. TMS320C25 to AD7716 Interface
AD7716
Multibandwidth System
OUTLINE DIMENSIONS
Some applications may require multiple AD7716’s with different bandwidths programmed to each device. The best way to
accomplish this is shown in Figure 12. The master mode interface is used for this example but the slave mode may also be
used. The example shows an 8-channel system with Device #0
in the system programmed for a 292 Hz cutoff frequency and
Device #1 programmed for a 146 Hz cutoff frequency.
Dimensions shown in inches and (mm).
0.048 (1.21)
0.042 (1.07)
0.056 (1.42)
0.042 (1.07)
6
RESET
0.048 (1.21)
0.042 (1.07)
TFS
#0
(292 Hz)
0.025 (0.63)
0.015 (0.38)
40
PIN 1
IDENTIFIER
7
CASCIN
0.180 (4.57)
0.165 (4.19)
39
0.021 (0.53)
0.013 (0.33)
C1920a–2–11/95
44-Lead PLCC
(P-44A)
0.63 (16.00)
0.59 (14.99)
RFS
DRDY
SCLK
RESET
SDATA
0.032 (0.81)
0.026 (0.66)
TOP VIEW
CASCOUT MODE
0.050
(1.27)
BSC
29
17
18
28
0.020
(0.50)
R
CASCIN
0.040 (1.01)
0.025 (0.64)
0.656 (16.66)
SQ
0.650 (16.51)
0.110 (2.79)
0.085 (2.16)
0.695 (17.65)
SQ
0.685 (17.40)
#1
(146 Hz)
44-Lead Plastic Quad Flatpack
(S-44)
DRDY
RESET
MODE
0.557 (14.148)
0.537 (13.640)
0.096 (2.44)
MAX
0.037 (0.94)
0.025 (0.64)
Figure 12. Multibandwidth System
The resultant output signals are also shown. Since Device #0
has a higher bandwidth it will also have a higher update rate.
The receiving processor will be getting samples from this device
at twice the rate of Device #1.
0.398 (10.11)
0.390 (9.91)
8°
0.8 °
33
23
34
22
0.398 (10.11)
0.390 (9.91)
TOP VIEW
RESET
PIN 1
44
12
1
DRDY #0
0.040 (1.02)
0.032 (0.81)
0.083 (2.11)
0.077 (1.96)
SDATA
#0
#0 #1
#0
#0 #1
#0
0.016 (0.41)
0.012 (0.30)
0.033 (0.84)
0.029 (0.74)
PRINTED IN U.S.A.
DRDY #1
11
0.040 (1.02)
0.032 (0.81)
#0 #1
Figure 13. Output Signals for Figure 12
–16–