Model HH - Honeywell Test and Measurement Sensors

Model HH
Portable LCD Display/Signal Conditioner
DESCRIPTION
The model HH handheld instrument features a full 3-1/2 digit
LCD display and is designed for use in remote field operations
as a portable calibrator or readout device. Shunt calibration offers the operator quick field setup with a minimum of warm-up. A
peak/hold feature is standard.
FEATURES
•
Strain-gage/bridge based transducer input
•
3 1/2 digit LCD display
•
Battery powered
•
Peak hold and shunt cal features
Model HH
General SPECIFICATIONS
Digital display characteristics
SPECIFICATIONS
Characteristic
Measure
Number of channels
1
Characteristic
Measure
Material
Hand-held plastic case
Number of display
characters
3½
Conversion/second
3
Scaling
0-1999
Scaling method
Zero and span adjustment
Max. display count
1999
Polarity indications
Yes
Programmable decimal
points
Yes
Display size
10,2 mm [0.4 in]
Overrange indication
Yes
Resolution
1/2000
ENVIRONMENTAL SPECIFICATIONS
Characteristic
Measure
Temperature, operating
0 °C to 54 °C [32 °F to 130 °F]
Temperature, storage
-29 °C to 66 °C [-20 °F to 150 °F]
Power requirement
Battery - one 9V cell
Transducer interface SPECIFICATIONS
Characteristic
Measure
Type
LCD
Transducer excitation
2.5 Vdc
Maximum sensitivity
0.1 microvolt/count
Types of inputs accepted
0.5 mV/V to 8 mV/V
Peak/hold
Yes
Transducer bridge
120 ohm min.
Pushbutton shunt cal
Yes
Calibration method
Manual
Electrical connection type
4 banana jacks
Zero balance
±15 % of full scale (min.)
Noise and ripple
30 microvolts
Amplifier characteristics SPECIFICATIONS
Characteristic
Measure
Non-linearity
±0.1 % of full scale (max.)
Drift, zero
±50 mV (max.)
Drift, span
±50 mV (max.)
Stability, zero
1.0 % of full scale/year
Stability, span
1.0 % of full scale/year
Frequency response
5 Hz
Fine span adjustment
±15 % of range
Fine zero adjustment
±15 % of range
Short circuit protection
Yes
Physical characteristics SPECIFICATIONS
Characteristic
Measure
Weight
0,5 kg [1 lb]
Power supply SPECIFICATIONS
Characteristic
Measure
Power requirements
One 9V cell
Front panel SPECIFICATIONS
Characteristic
Measure
Digital display
3-1/2 digits LCD
Not RoHS compliant
OPTION CODES
Many range/option combinations are available in
our quick-ship and fast-track manufacture programs. Please see http://sensing.honeywell.com/
TMsensor-ship for updated listings.
Input
2 Honeywell • Sensing and Control
52c. 10 mV/V to 40 mV/V
Model HH
Portable Display/Signal Conditioner
Mounting dimensions and characteristics
For reference only
TYPICAL SYSTEM DIAGRAM
Warranty. Honeywell warrants goods of its manufacture as
being free of defective materials and faulty workmanship.
Honeywell’s standard product warranty applies unless agreed
to otherwise by Honeywell in writing; please refer to your
order acknowledgement or consult your local sales office for
specific warranty details. If warranted goods are returned to
Honeywell during the period of coverage, Honeywell will repair
or replace, at its option, without charge those items it finds
defective. The foregoing is buyer’s sole remedy and is in lieu
of all warranties, expressed or implied, including those of
merchantability and fitness for a particular purpose. In no
event shall Honeywell be liable for consequential, special, or
indirect damages.
While we provide application assistance personally, through our
literature and the Honeywell web site, it is up to the customer to
determine the suitability of the product in the application.
Specifications may change without notice. The information we
supply is believed to be accurate and reliable as of this printing.
However, we assume no responsibility for its use.
For more information about Sensing and Control products, visit
www.honeywell.com/sensing or call +1-815-235-6847
Email inquiries to [email protected]
WARNING
PERSONAL INJURY
•DO NOT USE these products as safety or emergency
stop devices or in any other application where failure of
the product could result in personal injury.
Failure to comply with these instructions could result in
death or serious injury.
WARNING
MISUSE OF DOCUMENTATION
•The information presented in this catalogue is for
reference only. DO NOT USE this document as product
installation information.
•Complete installation, operation and maintenance
information is provided in the instructions supplied with
each product.
Failure to comply with these instructions could result in
death or serious injury.
Sensing and Control
Automation and Control Solutions
Honeywell
1985 Douglas Drive North
Golden Valley, MN 55422 USA
+1-815-235-6847
www.honeywell.com/sensing
008823-1-EN IL50 GLO June 2008
Copyright © 2008 Honeywell International Inc. All rights reserved.