CLAMP ON POWER LOGGER PW3360-20

CLAMP ON POWER LOGGER
PW3360-20
Power Measuring Instruments
Handy and Easy to Use
– Power Management Support
Now with
QUICK SET
Convenience
Reliable measurements start with
proper wiring.
The QUICK SET
function
guides you in
making the right
connections.
n Supports single to three-phase, 4-wire circuits
- Simultaneously measure up to three single-phase,
2-wire circuits (in the same power system).
n Measure up to 780V with a 1000V display range
n Broadly applicable for many jobs,
including leakage current measurement
- An optional clamp-on leakage sensor supports
measurements as low as 50 mA.
n Store months of data on SD cards
2
Begin with QUICK SET Convenience
Select your Wiring Type, Clamp and Destination, and Connect
Select wiring type (example: 3P4W) and connect
1
Connect the leads
to the PW3360-20.
Make proper
connections
simply by
observing
the colors of
the displayed
leads.
2 Connect the voltage clips.
▼
PASS
Double checks your
voltage input and phase
Proceed to the next
step when PASS appears.
Check wire
connection
status and
judgment
indicators.
FAIL
If FAIL appears, move the cursor
to the indicator and press the [ENTER] key.
Corrective action
tips appear
3 Connect the clamp sensors.
Select the
current range
Create a Graph to Clearly Grasp Power Consumption
Record power consumption on an
SD Card* at specific intervals. Load
the data into the PC.
Use Excel graph processing for
before and after comparisons.
Before and After Air Conditioning
and Power Consumption
Measurements
Before Power consumption (W)
After Power consumption (W)
Before Power conservation (Wh)
After Power conservation (Wh)
Time
* Store up to one year’s data acquired at one minute intervals. Performance cannot be guaranteed on storage media other than Hioki-specified SD card options.
3
Suits a Variety of Worksites
n Where no AC power is available
n In severe temperature environments
Battery* power provides about eight
hours of continuous operation. In
addition, a Voltage Line Power Adapter*
is available to power the PW3360-20
from the measurement lines.
The operating temperature range extends
from –10°C (14°F) to 50°C (122°F).
Even under battery operation,
measurements can be performed from
0 °C (32°F) to 40°C (104°F)
(0°C (32°F) to 50°C (122 °F) when using
LAN communication).
* Battery Set PW9002 and Voltage Line Power
Adapter PW9003 options are sold separately.
n Magnetic voltage adapters for hard-to-clip
terminals
Magnetic voltage adapters convertible
with the Voltage Cords L9438-53 let
you accurately detect voltage when
the circuit terminals are too shallow for
alligator clips to latch on.
* Magnetic Adapter 9804 option sold separately.
Magnet in the tip
n Fits in tight spaces
Battery Set PW9002
Generally compatible with M6 pan screws
Loaded with More Useful Functions
Simultaneous
Measurements
Simultaneously measures
three single-phase 2-wire
circuits in the same system.
The pulse input function can be used to record
power data and production volume counts
simultaneously. The power data and pulse volume
(production volume) information are useful for unit
cost production management.
Load
Power Source
Load Load
Power Source
Simultaneous
Measurements
Pulse
Input
▼
Load
Leakage Current
Measurement
As a 3-channel Leakage Current Logger
9657-10
Load
φ30mm
(1.18 in)
φ40mm
(1.57 in)
Ideal for quick investigation of intermittent leakage
by continuous calculation processing every 200 ms.
(Select to save the average, maximum and/or minimum at every interval.)
Useful
Tip
n Demand/Time Series Graph Displays
Load
9675
Leak Clamp on Sensor
Load
Options
Power Source
Simultaneous
With the optional leakage current clamp on sensors, turn the instrument into
Measurements
a 3-channel leakage current logger to help identify trouble spots.
(Demand Graph Display Example)
This function will be supported from version 2.00.
Demand graphs at specified times and power time series graphs can be
displayed on the color LCD. Observing on-site power fluctuations is useful
for confirming energy saving and related effects.
4
Efficient Power Management by Remote Measuring and PC Processing
Remote
Monitor
HTTP Server Function
Use a LAN cable to connect the PW3360-20 to a personal computer for
real-time remote monitoring and measurement display in a web browser.
Load
Power Source
Load
Power Source
LAN
HUB
Files recorded in the PW3360-20 internal memory or SD card are accessible by LAN or USB
connection, and are downloadable using the free PW3360 Setup and Download Software.
PC
Processing
Power Logger Viewer SF1001 (option, sold separately)
Data saved to an SD card or internal memory can be loaded into a PC for expanded display, aggregation and analysis.
O n t h e s a m e t i m e a x i s , v i e w m e a s u re d p o w e r
consumption and equipment operating status at specific
intervals, along with equipment characteristics and
management details.
Simultaneously measure and record
separate loads using three PW3360-20s
Air Conditioner A
Stacked Graph Display Example
Use the [Stacked Display] to confirm at a glance comparative
power consumption at multiple locations simultaneously.
Power Source
Air Conditioner B
Power Source
Air Conditioner C
Power Source
Freeware for Model PW3360-20 (free download from Hioki’s website)
PW3360 Excel Graph Auto-Creation Software
Install the PW3360 Excel Graph Auto-Creation Software to create
graphs in Excel automatically using recorded measurement data.
PW3360 Setup and Download Software
Use with a LAN or USB connection to download data recorded in the
PW3360-20's internal memory or SD Card to a PC, and to change
instrument settings from the PC.
Recorded Data
USB or LAN
5
n PW3360-20 Specifications (product guaranteed for one year)
Input specifications
Measurement
line type
Measurement
line Frequency
Number of input
channels
Voltage range
Single-phase 2-wire, single-phase 3-wire, three-phase 3-wire,
three-phase 4-wire
General specifications
Operating
environment
Operating
temperature and
humidity
(no condensation)
Storage
temperature and
humidity
(no condensation)
3.5 inch TFT color LCD (320 × 240 pixel)
Japanese, English (supported from version 1.50),
Chinese (Simplified, supported from version 2.00)
Backlight auto-off function (after 2 minutes)
Indoors, Pollution degree 2, altitude up to 2000 m (6562-ft.)
-10°C to 50°C (14°F to 122°F), 80% RH or less
During LAN communication: 0°C to 50°C (32°F to 122°F), 80% RH or less
During battery operation: 0°C to 40°C (32°F to 104°F), 80% RH or less
During battery charging: 10°C to 40°C (50°F to 104°F), 80% RH or less
-20°C to 60°C (-4°F to 140°F), 80% RH or less
However, the batteryُs storage temperature range is -20°C to
30°C (-4°F to 86°F), 80% RH or less
4.29 kVrms AC (1 mA sense current) between voltage input
terminals and external terminals, 50/ 60 Hz for 60 sec.
Applicable standards Safety: EN61010, EMC: EN61326, EN61000-3-2, EN61000-3-3
•Z1006 AC Adapter (12 V, 1.25 A), Rated supply voltage 100 VAC
to 240 VAC, Rated power supply frequency 50/60 Hz
Power supply
•Model 9459 Battery Pack (Ni-MH DC7.2 V 2700 mAh)
Charges the battery regardless of whether the instrument is on or off.
Charge function
Charge time: Max. 6 hr. 10 min. (reference value at 23°C)
•When the Z1006 AC Adapter is used: 40 VA (including AC adapter),
Maximum rated
13 VA (PW3360-20 instrument only)
power
•When the 9459 Battery Pack is used: 3 VA
Continuous
Approx. 8 hr. (Continuous, backlight off)
battery
(when using the battery pack)
operation time
Backup battery life Clock and settings (Lithium battery), Approx. 10 years @23°C (@73.4°F)
Approx. 180W(7.09") × 100H(3.94") × 48D (1.89") mm (without PW9002)
Dimensions
Approx. 180W(7.09") × 100H(3.94") × 68D (2.68") mm (with PW9002)
Approx. 550g (19.4 oz) (without PW9002), Approx. 830g (29.3 oz) (with PW9002)
Mass
Voltage Cord L9438-53(1 set), AC Adapter Z1006 (1),
USB cable(1), instruction manual (1), measurement guide (1),
Accessories
color spiral tubes (1 set): red, yellow, blue/two each, for color-coding clamp
sensors, spiral tubes for grouping clamp sensor cords (5)
Dielectric strength
Connection
Single-phase 2-wire (1P2W, 1P2W 5 2 circuits, 1P2W 5 3 circuits)
Single-phase 3-wire (1P3W, 1P3W+I, 1P3W1U, 1P3W1U+I)
Three-phase 3-wire (3P3W2M, 3P3W2M+I, 3P3W3M)
Three-phase 4-wire (3P4W), Current only: 1 to 3 channels
Simultaneous
power/current
measurement modes
Measurement
items
1P3W+I: 1 power circuit and 1 current channel
3P3W2M+I: 1 power circuit and 1 current channel
Calculation
selection
Power factor, reactive and apparent power: rms calculation/
fundamental wave calculation
50/ 60 Hz
Voltage: 3 channels U1 to U3
Current: 3 channels I1 to I3
600 V AC (single range)
Total display area: 5V to 1000 V (less than 5 V displays as 0 V)
Effective measurement range: 90 V to 780 V, peak: ±1400V
[OVER] indicates over-range warning
Current ranges
Load current
CLAMP ON SENSOR 9694
: 500m/1/5/10/50 A
CLAMP ON SENSOR 9695-02
: 500m/1/5/10/50 A
CLAMP ON SENSOR 9660
: 5/10/50/100 A
CLAMP ON SENSOR 9695-03
: 5/10/50/100 A
CLAMP ON SENSOR 9661
: 5/10/50/100/500 A
CLAMP ON SENSOR 9669
: 100/200/1k A
FLEXIBLE CLAMP ON SENSOR CT9667 : 500/5k A
Leakage current
LEAK CLAMP ON SENSOR 9657-10 : 50m/100m/500m/1/5 A
LEAK CLAMP ON SENSOR 9675
: 50m/100m/500m/1/5 A
Total display range: Within 0.4 to 130% of the range
(zero is suppressed for less than 0.4%)
Effective measurement range: Within 5 to 110% of the range
peak: ±400% of range, however, maximum range is 200%.
[OVER] indicates over-range warning
300.00 W to 9.0000 MW
Power ranges
Depends on voltage/current combination and measured line type
(see Measurement Range Configuration Tables)
Total display range: Within 0 to 130% of the range
(“0W” display indicates zero rms voltage and/or current)
Effective measurement area: Within 5 to 110% of the range
Any (0.01 to 9999.99)
VT ratio settings
Selections (1/60/100/200/300/600/700/1000/2000/2500/5000)
Any (0.01 to 9999.99)
CT ratio settings
Selections (1/40/60/80/120/160/200/240/300/400/600/800/1200)
Voltage: Insolated inputs (except between U1, U2, U3 and N)
Input methods
Current: Isolated input using a clamp-on sensor
Voltage input part: 3 MΩ ±20% (50/ 60 Hz)
Input resistance
Maximum rated voltage Voltage input section: 1000 VAC, 1400 Vpeak
Current input section: 1.7 VAC, 2.4 Vpeak
between terminals
Voltage input section: 600V Measurement Category III
Maximum rated
300V Measurement Category IV
voltage to earth
Current input section: Depends on clamp sensor in use.
Display device
Accuracy guarantee period: One year 23°C ±3°C, 80%RH or less, (no condensation)
Measurement specifications
Voltage RMS, current RMS, voltage fundamental wave value,
current fundamental wave value, voltage fundamental wave phase
angle, current fundamental wave phase angle, frequency (U1),
voltage waveform peak (absolute value), current waveform peak
(absolute value), active power, reactive power (with lag/lead
display), apparent power, power factor (with lag/lead display) or
displacement power factor (with lag/lead display), active energy
(consumption, regeneration), reactive energy (consumption,
regeneration), electricity rate display (by means of planned future
function update), active power demand quantity (consumption,
regeneration), reactive power demand quantity (lag, lead), active
power demand value (consumption, regeneration), reactive power
demand value (lag, lead), power factor demand, pulse input
Voltage: ±0.3% rdg. ±0.1% f.s.
Current: ±0.3% rdg. ±0.1% f.s. + clamp sensor accuracy
(50/ 60Hz,
Active power: ±0.3% rdg. ±0.1% f.s. +clamp sensor accuracy
power factor = 1) Clamp-On Sensor 9661 accuracy: ±0.3% rdg. ±0.01% f.s.
Measurement
accuracy
(Accuracy depends on clamp sensor. See page 6 for the accuracy of
each model, and page 7 for combined accuracy of Model PW3360-20
and each clamp sensor.)
Approx. 0.5 sec (except when accessing SD card or internal memory,
Display update rate
or during LAN/USB communication)
Digital sampling and zero cross synchronization calculation method
Measurement
Sampling: 10.24 kHz (2048 points)
method
Calculation processing
50 Hz: Continuous, gapless measurement at 10 cycles
60 Hz: Continuous, gapless measurement at 12 cycles
A/D converter resolution 16bit
Recording specifications
Save destination SD Card, internal memory (capacity: approx. 320 KB)
Save interval time 1/2/5/10/15/30 seconds, 1/2/5/10/15/20/30/60 minutes
* Available storage time is displayed on PW3360-20's setting screen
Measurement save: Average only / all (average, maximum, minimum)
Save items
Screen copy: ON/OFF (Saves the displayed screen as a BMP at a fixed
interval.) The minimum interval time for saving screen copies is 5 min.
If the setting is less than 5 min., screen copies will be saved every 5 min.
Waveform save: Stores binary waveform data (with shortest
interval 1 minute) Supported from version 2.00
Recording start methods Interval time, manual, or at specified time
Recording stop methods Manual, or at specified time (up to one year)
Pulse input
Input specifications No-voltage contact input (counts when shorted terminals open)
Voltage input (Hi: 2 V to 45 V, Lo: 0 V to 0.5 V, counts at Lo to Hi)
Maximum rated input between terminals: 45 V DC
Maximum rated input to ground: not isolated (GND is equipment common)
Measurement range 0 to 9999 (maximum pulse count per save interval)
Filter On (for mechanical contacts) 25 Hz or less, and at least 20
Filter
ms Hi and Lo pulse width
Filter Off (for solid-state contacts) 5 kHz or less, and at least 100
µs Hi and Lo pulse width
Displays product of pulse count and scaling factor setting
Scaling
Setting ranges: 0.001 to 1.000, and 1.000 to 100.00
Pulse output
Function
Pulse rate
Pulse width
Output signal
Output pulse rate is proportional to active power consumption
(WP+) when measuring integral power consumption
OFF/1Wh/10Wh/100Wh/1kWh/10kWh/100kWh/1000kWh
(Default: 1 kWh)
approx. 100 ms
Open-collector 30 V, 5 mA max (photocoupler isolated)
Active Low
External interfaces Specifications
SD card Interface Settings data, measurement data, screen data
Waveform data (support planned from version 2.00)
10BASE-T/100BASE-TX IEEE802.3 Compliance
LAN interface
- HTTP server function
- Download settings and data by communication application program
USB Ver 2.0, Windows 7 (32/64bit) / Vista (32bit) /XP
USB interface
- When connected to a computer, the SD Card and internal
memory are recognized as removable storage devices.
- Download settings and data by communication application program
6
n POWER LOGGER VIEWER SF1001 Specifications
Functions Specifications
General Specifications
Read-compatible model
Supported
computer operating
systems
PW3360-20
Windows 7 SP1 or later (32/64bit)
Windows Vista SP2 or later (32bit)
Windows XP SP3 or later (32bit)
Display items: Voltage, current, active power, reactive power, apparent
power, power factor, frequency, integrated active power, integrated
reactive power, demand volume, demand value, voltage disequilibrium
factor, pulse
*Measurement values can be displayed by the cursor
Stacked bar graph display: Up to 16 types of data series can be
displayed in an overlay graph
Time series
graph display
function
Displayed items are the same as for Time Series Graph Display
- Display and totalize monthly/weekly/daily reports for specified period
- Calculate load factor and demand factor for daily/weekly/monthly
reports, and displays results
- Hourly totalization (up to four segments)
Captures any display image to the clipboard
Summary display
function
Copy function
Preview and print content shown on the time series graph,
report, and settings displays.
Comment entry (Text comments can be entered in any printout)
Printing support: Any color or monochrome printing supported by the
operating system
Print (static) contents over a specific time period
Print function
Report printing
n CLAMP SENSOR Specifications
CLAMP ON SENSOR
9694
9660
9661
9669
9695-02
9695-03
Appearance
CONNECTION CORD 9219
Cord length: 3 m
(9.84ft)
Measurable conductor
diameter
Primary current rating
Accuracy
Amplitude (45 to 66 Hz)
Phase (45 Hz to 5 kHz)
Cord length: 3 m
(9.84ft)
Cord length: 3 m
(9.84ft)
Cord length: 3 m
(9.84ft)
φ15mm (0.59")
φ15mm (0.59")
φ46mm (0.81")
5A AC
±0.3% rdg.
±0.02% f.s.
Within ±2˚
100A AC
±0.3% rdg.
±0.02% f.s.
Within ±1˚
500A AC
±0.3% rdg.
±0.01% f.s.
Within ±0.5˚
Frequency characteristic
φ55mm (2.17"),
80 (3.15")×20 (0.79")mm
1000A AC
±1.0% rdg.
±0.01% f.s.
Within ±1˚
Connect with the
9695-02/-03,
Output BNC terminal
Cord length: 3 m (9.84ft)
φ15mm (0.59")
φ15mm (0.59")
50A AC
±0.3% rdg.
±0.02% f.s.
Within ±2˚
100A AC
±0.3% rdg.
±0.02% f.s.
Within ±1˚
40Hz to 5kHz
(deviation from accuracy)
Within ±1.0%
Within ±2.0%
Within ±1.0%
Effect of external magnetic field
Equivalent to 0.1 A or less
Equivalent to 1 A or less
Equivalent to 0.1 A or less
Within ±1.5%
Within ±0.5%
CAT III 600Vrms
CAT III 300Vrms
(with a magnetic field of 400 A/ m AC)
Within ±0.5%
Effect of conductor position
Maximum rated voltage to
CAT III 300Vrms
CAT III 300Vrms
CAT III 600Vrms
earth
130 A continuous
550 A continuous
Maximum input (45 to 66Hz) 50 A continuous
46W (1.81") 5 135H (5.31") 46W (1.81") 5 135H (5.31") 77W (3.03") 5 151H (5.94")
Dimensions
Mass
60 A continuous 130 A continuous
50.5W(2.28") 5 58H(2.28")
5 18.7D(0.74") mm
50g (1.8 oz)
5 21D (0.83") mm
5 21D (0.83") mm
5 42D(1.65") mm
230g (8.1 oz)
230g (8.1 oz)
380g (13.4 oz)
590g (20.8 oz)
FLEXIBLE CLAMP ON SENSOR
CT9667
Appearance
1000 A continuous
99.5W (3.92") 5 188H (7.40")
5 42D (1.65") mm
Cord length:
Sensor - circuit: 2 m (6.56ft)
Circuit - connector: 1 m (3.28ft)
CLAMP ON LEAK SENSOR (Leakage Current Measurement Only)
9657-10
9675
Insulated conductor
Appearance
Cord length: 3 m
(9.84ft)
Measurable conductor diameter
Primary current rating
Amplitude
Accuracy
(45 to 66Hz)
Phase
Frequency characteristic
φ254mm
500A AC/5,000A AC
±2.0% rdg. ±0.3% f.s.
Within ±1˚
10Hz to 20kHz
(deviation from accuracy)
Within ±3 dB
Effect of external magnetic field
1.5% / f.s. or less.
Effect of conductor position
Maximum rated voltage to earth
Maximum input
Within ±3.0%
CAT III 1000Vrms, CAT IV 600Vrms
Dimensions
Mass
Circuit box: 35W (1.38") × 120H (4.74") × 34D (1.34") mm
470g (16.6 oz.) (Sensor + Circuit Box, w/battery)
LR06 alkaline battery × 2 (continuous operation max. 7 days)
or AC ADAPTER 9445-02/9445-03 (optional)
(with a magnetic field of 400 A/ m AC)
(45 to 66Hz)
Power supply
10000 A continuous
Insulated conductor
φ40mm (1.57")
Measurable conductor diameter
10A AC*
Primary current rating
Accuracy Amplitude (45 to 66 Hz) ±1.0% rdg. ±0.05% f.s.
Phase angle (@50 or 60 Hz)
Within ±3˚
Frequency characteristic
Within ±5%
40Hz to 5kHz
(deviation from accuracy)
Cord length: 3 m
(9.84ft)
φ30mm (1.18")
10A AC*
±1.0% rdg. ±0.005% f.s.
Within ±5˚
Within ±5%
Effect of external magnetic field
7.5 mA max.
7.5 mA max.
Effect of conductor position
Maximum rated voltage to earth
Maximum input
Within ±0.1%
CAT III 300Vrms
Within ±0.1%
CAT III 300Vrms
30 A continuous
10 A continuous
Dimensions
74W(2.91") × 145H(5.71")
× 42D(1.65")
60W(2.36") × 112.5H(4.43")
× 23.6D(0.95")
(with a magnetic field of 400 A/ m AC)
(45 to 66Hz)
Mass
Notes
380g (13.4 oz)
160g (5.6 oz)
Not used for power measurements
* Maximum AC measurement range with PW3360-20 is 5A.
7
n Measurement Range Configurations
Current
Voltage
Connection
1P2W
1P3W
1P3W1U
600.00 V
3P3W2M
3P3W3M
3P4W
Total display range
CLAMP ON SENSOR 9694 (CAT III 300V) *1
CLAMP ON SENSOR 9695-02 (CAT III 300V)
500.00 mA
1.0000 A
5.0000 A
10.000 A
50.000 A
300.00 W
600.00 W
3.0000 kW
6.0000 kW
30.000 kW
600.00 kW
1.2000 kW
6.0000 kW
12.000 kW
60.000 kW
900.00 W
1.8000 kW
9.0000 kW
18.000 kW
90.000 kW
Voltage is displayed from 5 V to 1000 V, with less than
5 V displayed as 0 V.
Current is displayed from 0.4% to 130% of the selected
range, with less than 0.4% displayed as 0 A
Power is displayed from 0 to 130% of full scale, with 0
W displayed when voltage or current is zero.
The range configurations for apparent power (S) and
reactive power (Q) are the same, with units of [VA] and
[var], respectively.
*1. For the 9694 sensor, the range of guaranteed accuracy is from 500 mA to 5 A, and for the 9695-02, from 500 mA to 50 A.
Current
Voltage
When VT and CT ratios are set, the range configuration
is the product (VT ratio × CT ratio).
CLAMP ON SENSOR 9660, 9695-03 (CAT III 300V) *2
CLAMP ON SENSOR 9661
5.0000 A
10.000 A
50.000 A
100.00 A
3.0000 kW
6.0000 kW 30.000 kW
60.000 kW
Connection
500.00 A
300.00 kW
1P2W
1P3W
1P3W1U
6.0000 kW
12.000 kW 60.000 kW
120.00 kW
600.00 kW
600.00 V
3P3W2M
3P3W3M
9.0000 kW
18.000 kW 90.000 kW
180.00 kW
900.00 kW
3P4W
*2. For the 9660 and 9695-03 sensors, the range of guaranteed accuracy is from 5 A to 100 A, and for the 9661, from 5 A to 500 A.
Current
Voltage
Connection
1P2W
1P3W
1P3W1U
600.00 V
3P3W2M
3P3W3M
3P4W
CLAMP ON SENSOR 9669
100.00 A
60.000 kW
200.00 A
120.00 kW
1.0000 kA
600.00 kW
120.00 kW
240.00 kW
1.2000 MW
180.00 kW
360.00 kW
1.8000 MW
Current
Voltage
Connection
1P2W
1P3W
1P3W1U
600.00 V
3P3W2M
3P3W3M
3P4W
FLEXIBLE CLAMP ON SENSOR
CT9667
500.00 A
5.0000 kA
300.00 kW
3.0000 MW
For voltage, 90 to 780 V, with max. 1400 V peak.
For current, 5% to 110% of the selected range with
peak ±400% of range, but maximum range is ±200%.
For power, 5% to 110% of the selected range.
For frequency, 45 to 66 Hz.
n Current Display and Effective Measurement Ranges (typical)
Current
Range
Display range
650 A
Effective measurement range
2A
25 A
550 A
130 A
Effective measurement range
5A
110 A
65 A
Effective measurement range
500 A
100 A
0.4 A
50 A
0.2 A
2.5 A
0A
55 A
10 A
1A
100 A
1000 A
13 A
600.00 kW
6.0000 MW
10 A
Effective measurement range
0.04 A
0.5 A
5A
900.00 kW
9.0000 MW
Leak current: CLAMP ON LEAK SENSOR 9657-10, 9675
Range
Effective measurement range
6.5 A
11 A
Effective measurement range
0.02 A
0A
5.5 A
0.25 A
100 mA
1A
10 A
100 A
Input Current
50.000 mA/100.00 mA/500.00 mA/1.0000 A/5.0000 A
n Measurement accuracy
±0.3% rdg. ±0.1% f.s.
Voltage
±0.3% rdg. ±0.1% f.s. + clamp sensor accuracy
Current
Active power ±0.3% rdg. ±0.1% f.s. + clamp sensor accuracy (power factor = 1)
Combined accuracy of PW3360-20 + clamp sensors
Range
50.000 A
10.000 A
5.0000 A
1.0000 A
500.00 mA
Range
500.00 A
100.00 A
50.000 A
10.000 A
5.0000 A
Range
1.0000 kA
200.00 A
100.00 A
Range
5.0000 kA
500.00 A
9694
―
―
±0.6% rdg. ±0.12% f.s.
±0.6% rdg. ±0.2% f.s.
±0.6% rdg. ±0.3% f.s.
9695-02
±0.6% rdg. ±0.12% f.s.
±0.6% rdg. ±0.2% f.s.
±0.6% rdg. ±0.3% f.s.
±0.6% rdg. ±1.1% f.s.
±0.6% rdg. ±2.1% f.s.
9660, 9695-03
―
±0.6% rdg. ±0.12% f.s.
±0.6% rdg. ±0.14% f.s.
±0.6% rdg. ±0.3% f.s.
±0.6% rdg. ±0.5% f.s.
9661
±0.6% rdg. ±0.11% f.s.
±0.6% rdg. ±0.15% f.s.
±0.6% rdg. ±0.2% f.s.
±0.6% rdg. ±0.6% f.s.
±0.6% rdg. ±1.1% f.s.
9669
±1.3% rdg. ±0.11% f.s.
±1.3% rdg. ±0.15% f.s.
±1.3% rdg. ±0.2% f.s.
CT9667 5.000 kA range
±2.3% rdg. ±0.4% f.s.
―
CT9667 500 A range
―
±2.3% rdg. ±0.4% f.s.
Conditions of guaranteed accuracy After 30 minute warm-up, with 50/60 Hz sine wave input
Temperature and humidity
for guaranteed accuracy
Display area
of guaranteed accuracy
Period of guaranteed acuracy
23˚C ±5˚C (73 ± 9°F), 80%RH or less
(applies to all specifications unless otherwise noted)
Effective measurement range
1 year
Within ±0.3 sec/day (with power on, within specified operating
temperature and humidity ranges)
Temperature characteristic Within ±0.1% f.s./ ˚C (except 23 ±5°C)
Within ±0.2% f.s.
Effect of common mode
(600 V AC, 50/60 Hz, between voltage input terminal and case)
voltage
Effect of external magnetic field Within ±1.5% f.s. (in a magnetic field of 400 A/m rms AC, 50/60 Hz)
Effect of phase
Phase accuracy ±0.3° equivalent (with 50/60 Hz f.s. input)
Real-time clock accuracy
±1 dgt. for the calculation obtained from each measurement value
Fundamental waveform calculations
±0.3% rdg. ±0.1% f.s. + clamp-on sensor accuracy (w/power factor = 1)
Rms calculations
From each measurement applied to calculation ±1 dgt.
Active and reactive power measurement accuracies ±1 dgt.
Energy
From each measurement applied to calculation ±1 dgt.
Power factor
±0.5% rdg. (with 90 to 780 V sine wave input)
Frequency
Active and reactive power measurement accuracies ±1 dgt.
Demand value
Active and reactive power measurement accuracies ±1 dgt.
Demand quantity
±1 dgt. for the calculation obtained from each measurement value
Pulse input
Frequency characteristic At 50/60 Hz fundamental waveform frequency,
up to 1 kHz, ±3% rdg. ±0.2% f.s.
up to 3kHz, ±10% rdg. ±0.2% f.s.
For current and active power, add clamp-on sensor accuracy.
Note: only for 3P3W3M wiring, add ±0.5% rdg.
Apparent power
Reactive power
CLAMP ON POWER LOGGER PW3360-20
Accessories
VOLTAGE CORD L9438-53 (1 set) , AC ADAPTER Z1006 (1) , USB cable (1),
instruction manual (1), measurement guide (1), color spiral tubes (1 set): red, yellow, blue/two
each, for color-coding clamp sensors, spiral tubes for grouping clamp sensor cords (5)
Clamp-On Power Logger PW3360-20 by itself does not support current and
power measurements. Current and power measurements require clamp-on
sensors, sold separately. Also, use only HIOKI-issued SD cards guaranteed
to work for saving measurement data, (options, sold separately).
AC ADAPTER Z1006
VOLTAGE CORD L9438-53
cord length: 3m (9.84 ft)
1 cord each of black, red yellow, and blue,
and five spiral tubes for bundling cords
n Options
CLAMP ON ADAPTER
CLAMP ON SENSOR (for load current measurement)
9290-10
CLAMP ON SENSOR 9694 (AC5A)
CLAMP ON SENSOR 9660 (AC100A)
CLAMP ON SENSOR 9661 (AC500A)
CLAMP ON SENSOR 9669 (AC1000A)
FLEXIBLE CLAMP ON SENSOR CT9667 (AC5000A)
CLAMP ON SENSOR 9695-02 (AC50A)
CLAMP ON SENSOR 9695-03 (AC100A)
CONNECTION CORD 9219 (for connection to 9695-02, 9695-03)
MAX. 1500A AC (continuous: 1000A)
Primary side
1000A
CAT III 600V
Cord length: 3m (9.84 ft)
When purchasing the 9695-02 and 9695-03, we recommend also purchasing the
separately sold 9219 Connection Cord.
Secondary side
100A
Measurable conductor diameter
φ55 mm (2.17in)
Bus bar: ■ 80 mm (3.46in) 5 20 mm (0.79 in)
CT ratio: 10:1
CLAMP ON LEAK SENSOR (for leakage current measurement)
CLAMP ON LEAK SENSOR 9657-10
CLAMP ON LEAK SENSOR 9675
VOLTAGE LINE POWER ADAPTER
SD MEMORY CARD 2GB
Z4001
CARRYING CASE
PW9003
Rated voltage: 240 V AC
(supplies power from measurement lines) Operating temperature and humidity range: -10 to 50°C, 80% RH or less
CAT III 300V
Power Source
Stores up to one year’s data when
acquired at one minute intervals.
Performance cannot be guaranteed
on storage media other than Hiokispecified SD card options.
BATTERY SET
BATTERY PACK 9459
For purchase as replacement
battery pack
POWER LOGGER VIEWER
MAGNET ADAPTER
C1005
Battery Case and Battery Pack Set
PW9002
9804-01 Red
SF1001
LAN CABLE
9642
9804-02 Black
φ11mm (0.43 in)
(generally compatible with M6 pan screws)
Magnetic tip for use with the standard
VOLTAGE CORD L9438-53
Red and black adapters sold separately.
Purchase the quantity and color appropriate for your
application.
(Example: 3P3W-3 adapters, 3P4W-4 adapters)
Note: Company names and Product names appearing in this catalog are trademarks or registered trademarks of various companies.
HIOKI (Shanghai) SALES & TRADING CO., LTD.:
TEL +86-21-63910090 FAX +86-21-63910360
http://www.hioki.cn / E-mail: [email protected]
DISTRIBUTED BY
HIOKI INDIA PRIVATE LIMITED:
TEL +91-731-6548081 FAX +91-731-4020083
HEADQUARTERS:
E-mail: [email protected]
81 Koizumi, Ueda, Nagano, 386-1192, Japan
TEL +81-268-28-0562 FAX +81-268-28-0568 HIOKI SINGAPORE PTE. LTD.:
http://www.hioki.com / E-mail: [email protected] TEL +65-6634-7677 FAX +65-6634-7477
E-mail: [email protected]
HIOKI USA CORPORATION:
HIOKI KOREA CO., LTD.:
TEL +1-609-409-9109 FAX +1-609-409-9108 TEL +82-42-936-1281 FAX +82-42-936-1284
http://www.hiokiusa.com / E-mail: [email protected] E-mail: [email protected]
All information correct as of Jan. 21, 2013. All specifications are subject to change without notice.
PW3360E1-31M
Printed in Japan