FUJI EDS3-111E

NDIR TYPE INFRARED GAS ANALYZER
DATA SHEET
ZRH
This NDIR gas analyzer is used for measuring CO2, CO,
SO2 and CH4 and features a high accuracy, multiple functions
and easy operation through use of a microprocessor. It is
housed in an easy-to-use horizontal case (can be flush
mounted on panel, mounted on 19 inch rack, placed on table
top), or vertical case (can be flush mounted on panel,
mounted on wall).
The analyzer provides a performance superior to the conventional double-beam system, is easy to maintain, and offers
an excellent long-term stability. It is thus optimum for continuous measurement in the combustion control of various
industrial furnaces, in research on garden-plants, and so on.
FEATURES
1. The use of a microprocessor provides high accuracy,
multiple functions and easy operation.
• Zero and span calibration is accurate and easy just by
pressing the calibrating keys.
• A self-diagnosis function is included.
• An automatic calibrating function can be provided as
an option.
• Range can be changed over by a external signal as an
option.
2. An improved optical system provides long term stability, and there is a minimum of drift caused by contamination of the measuring cell, so the long-term stability
is excellent.
3. Adopting a serial dual-layer type of transmission detector minimizes remarkable the interface from other gas
components.
4. Easy maintenance.
The single-beam photometric system uses a sample cell
only and eliminates the necessity of delicate adjustment
for optical balance. The instrument is designed as a
unit of simple construction featuring easy maintenance
and checks.
SPECIFICATIONS
Measurable gas components:
Single-component
CO 2 CO, SO2, CH4
Dual-component
CO2/CO:
Landfill garbage application
CO2: 0 to 50%
CH 4: 0 to 80%
Measuring range: Refer to measurable range combination
table (page 5).
Measuring system:
Non-dispersion infrared-ray absorption
method, deviation method, single light
source – single beam
Output signal:
Output 1; 0 to 1V DC
Output 2; 4 to 20mA DC (optional allowable load resistance 550Ω or less).
Repeatability:
1st range (low range)
Within ± 0.5% of full scale
2nd range (high range)
Within ± 1% of full scale
Linearity:
± 2% of full scale
Zero drift:
Within ± 2% of full scale/week
Span drift:
Within ± 2% of full scale/week
Response time: Within 15 seconds max. (for 90% response) depending on cell length
Power supply:
100V, 115V, 220V or 240V ± 10% AC, 50/
60Hz
Power consumption:
37VA max.
EDS3-111e
Date
Nov. 24, 2004
ZRH
Ambient temperature:
–5 to +45°C
Ambient humidity:
90% RH or less
Enclosure:
Steel casing, for indoor use
Storage condition:
Temperature; –20 to +60°C
Humidity; 100% RH max. (free from condensation)
Outer dimensions (H x W x D):
(1) Horizontal case
Rack mounting type; 133 x 483 x 435 mm
Panel flush mounting type;
133 x 443 x 435 mm
Table-top type; 145 x 443 x 435 mm
(2) Vertical case
Panel flush mounting type;
440 x 294 x 178 mm
Wall mounting type; 484 x 294 x 178 mm
Mass weight:
Approx. 12Kg
Finish color:
Munsell 2.5Y8.4/1.2
Display:
4 digit LED for concentration display
4 digit LED for sub-display
Output hold:
Output value before manual or automatic
calibration is hold. Whether or not to effect hold function can be selected.
Sample gas condition:
Temperature; 0 to 50°C
Dust; less than 0.3 µm
Pressure; less than 10 kPa
Standard adjustment gas:
Dry N2 Balance
Warm up time: Approx. 2 hours
Material of gas-contacting parts:
Sample cell, SUS304, neoprene rubber
Infrared-ray transmitting window; CaF2
or sapphire
Internal tubing; Toaron tube
Gas inlet/outlet, purge gas inlet size:
Rc1/4 (PT1/4 internal thread) or NPT1/4
internal thread
Measured gas flow rate:
1 ± 0.5 liter/minutes
Purge gas flow rate:
Approx. 1 liter/minute
Scope of delivery:Analyzer, power fuse, manual, mounting
bracket in the case of panel mounting type
Mounting method:
(1) Horizontal case
Mounted on 19 inch rack, or on panel, or
on table-top
approx. 250mm
90°
Support
Remark: 70% or more of the analyzer weight should
be supported at the bottom of the case.
(In case of mounting panel or 19 inch rack,
provide a support at the rear of casing.)
(2) Vertical case
Flush mounted on panel or mounted on
wall
or
90°
2
90°
Installation conditions:
Install the analyzer at a place not exposed
to direct sunlight or the radiation from a
high temperature object. A void vibration, and select a clean place free of corrosive and/or combustible gases. If installing outdoors, provide a suitable casing or cover to protect the analyzer from
wind, rain, etc.
Optional specifications
Remote output hold:
Analog output (DC0-1V, 4-20mA) is held
via external signal.
Input signal: 5V DC
Remote range changeover:
Range is changeable via external signal.
Range changeover input signal: 5V DC
Range identification signal output:
Contact output; 1 a contact
Contact capacity; 250V AC, 2A (resistive
load)
Automatic calibration:
Zero and span are automatically calibrated
at the preset cycle. Calibration gas is
supplied sequentially by driving an electromagnetic value installed outside.
Calibration channel:
Up to 2 components can be calibrated simultaneously.
Zero calibration point:
Fixed at 0%
Span calibration point:
50 to 100% full scale
Calibration start:
Via built-in timer or remote start signal
Output hold at calibration:
Possible
Calibration gas flow mode:
(1) Zero gas
(2) Zero gas – span gas 1
(3) Zero gas – span gas 2
(4) Zero gas – span gas 1 – span gas 2
Calibration gas flow time:
Settable from 100 to 599 sec.
Calibration cycle:
1 to 199 hours (in 1-hour step)
Calibration failure alarm:
Provided when fault occurs during auto
calibration.
Contact output:
During calibration; 1 a (N.O) contact, contact capacity 250V AC, 2A (resistive
load)
Calibration failure; 1 a (N.O) contact, contact capacity 250V AC, 2A (resistive
load)
Electromagnetic valve drive; 1 a (N.O) contact, contact capacity 250V AC, 2A (resistive load)
Remote start:
Remote start signal; voltage input 5V DC
CODE SYMBOLS
(1) Single-component analyzer
1 2 3 4 5 6 7 8
Z RH
9 10 11 12 13
Y 2 -
A
14 15 16 17
-
Y
18
Description
Y -
No. of measurable components
Single component Horizontal case
Single component Vertical case
1
3
Measurable component
SO2
CO
CO2
CH4
A
B
D
E
Measuring range (1st range) (Note 1)
0 to 500 ppm
0 to 1000 ppm
0 to 2000 ppm
0 to 2500 ppm
0 to 5000 ppm
0 to 1%
0 to 2%
0 to 3%
0 to 5%
0 to 10%
0 to 20%
0 to 40%
0 to 50%
0 to 70%
0 to 100%
Others
E
F
G
U
H
J
K
Q
L
M
N
W
P
X
R
Z
Power supply, Piping connection
100V AC 50Hz Rc 1/4
100V AC 60Hz Rc 1/4
115V AC 60Hz Rc 1/4
220V AC 50Hz Rc 1/4
100V AC 50Hz NPT 1/4
100V AC 60Hz NPT 1/4
115V AC 60Hz NPT 1/4
220V AC 50Hz NPT 1/4
240V AC 50Hz NPT 1/4
240V AC 50Hz Rc 1/4
0
1
2
3
4
5
6
7
8
9
Structure
Table top type (For horizontal case only)
19 inch rack mounting type (For horizontal case only)
Panel mounting type
Wall mounting type
A
B
C
D
Output signal (Output 1/2) (Notes 2, 3)
0 to 1V DC /4 to 20mA DC
A
Optional function (1)
None
With auto-calibration
Y
A
Optional function (2)
None
With remote range, range identification and
remote output hold
Y
A
Measuring range (2nd range) (Note 1)
None
0 to 1000 ppm
0 to 2000 ppm
0 to 2500 ppm
0 to 5000 ppm
0 to 1%
0 to 2%
0 to 5%
0 to 10%
0 to 20%
0 to 50%
0 to 100%
Others
Y
F
G
U
H
J
K
L
M
N
P
R
Z
Adjustment gas
Standard
for carburizing furnace
Others (Note 4)
A
B
Z
Paint
Fuji standard
Y
Z
Non standard specifications
Non standard
Notes :
(1) Refer to measurable range combined table
(2) Output signal are provided simultaneously
(3) 0 to 100mV, 0 to 10mV DC is available on request
(4) To be adviced components of sample gas
3
ZRH
(2) Dual-component analyzer
1 2 3 4 5 6 7 8
Z RH
G
9 10 11 12 13
2 -
A
14 15 16 17
-
1 2 3 4 5 6 7 8
18
-
Description
G
9 10 11 12 13
2 -
No. of measurable
components
Dual component
Horizontal case
Dual component
Vertical case
2
4
Measuring range
(1st comp. 1st range)
Note (5, 6)
0 to 5000 ppm
0 to 1%
0 to 2%
0 to 5%
0 to 10%
0 to 20%
0 to 50%
0 to 100%
Others
H
J
K
L
M
N
P
R
Z
Power supply, Piping
connection
100V AC 50Hz Rc 1/4
100V AC 60Hz Rc 1/4
115V AC 60Hz Rc 1/4
220V AC 50Hz Rc 1/4
100V AC 50Hz NPT 1/4
100V AC 60Hz NPT 1/4
115V AC 60Hz NPT 1/4
220V AC 50Hz NPT 1/4
240V AC 50Hz NPT 1/4
240V AC 50Hz Rc 1/4
0
1
2
3
4
5
6
7
8
9
Structure
Table top type
19 inch rack mounting
type
Panel mounting type
Wall mounting type
A
B
C
D
Output signal (Output
1/2) (Notes 2, 3)
0 to 1V/DC 4 to 20mA
DC
A
Optional function (1)
None
With auto-calibration
Y
A
Y
A
Optional function (2)
None
With remote range,
range identification and
remote output hold
Note (5) Refer to measurable range combination table
(6) 1st comp. is CO2, 2nd comp. is CO.
14 15 16 17
-
18
Description
-
Measuring range
(1st comp. 2nd range)
None
0 to 1%
0 to 2%
0 to 5%
0 to 10%
0 to 20%
0 to 50%
0 to 100%
Others
Measuring range
(2nd comp. 2nd range)
None
0 to 1000 ppm
0 to 2000 ppm
0 to 5000 ppm
0 to 1%
0 to 2%
0 to 5%
0 to 10%
0 to 20%
0 to 25%
0 to 50%
0 to 100%
Others
Y
F
G
H
J
K
L
M
N
V
P
R
Z
Measuring range
(2nd comp. 1st range)
Note (5, 6)
0 to 500 ppm
0 to 1000 ppm
0 to 2000 ppm
0 to 5000 ppm
0 to 1%
0 to 2%
0 to 5%
0 to 10%
0 to 20%
0 to 25%
0 to 50%
0 to 100%
Others
E
F
G
H
J
K
L
M
N
V
P
R
Z
A
Y
J
K
L
M
N
P
R
Z
Measurable component
CO2/CO
G
4
Z RH
Adjustment gas
Standard
for carburizing furnace
Others
A
B
Z
Paint
Fuji standard
Y
Z
Non standard specifications
Non standard
The product conforms to the requirements of the
Electromagnetic compatibility Directive 89/336/EEC as
detailed within the technical construction file number
TZ734576. The applicable standards used to demonstrate
compliance are :
EN 50081-1 : 1991 CLASS A Conducted and Radiated emissions
EN 50082-1 : 1992
Radiated immunity, ESD and FBT
Note: Available for ZRH1, 2
Measurable range combination table
(1) Single-component (CO2• CO• CH4)
1st
E
F
G
U
H
J
K
Q
L
M
N
W
P
X
R
G
U
2nd range
F
range
0 to 1000 ppm 0 to 2000 ppm 0 to 2500 ppm
0 to 500 ppm
—
0 to 1000 ppm
—
—
0 to 2000 ppm
—
—
—
0 to 2500 ppm
—
—
—
0 to 5000 ppm
—
—
—
0 to 1%
—
—
—
0 to 2%
—
—
—
0 to 3%
—
—
—
0 to 5%
—
—
—
0 to 10%
—
—
—
0 to 20%
—
—
—
0 to 40%
—
—
—
0 to 50%
—
—
—
0 to 70%
—
—
—
0 to 100%
H
0 to 5000 ppm
J
0 to 1%
K
0 to 2%
L
0 to 5%
M
0 to 10%
N
0 to 20%
P
0 to 50%
R
0 to 100%
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
:CO2
:CO
: CH4
:SO2 —: Impossible
* Also single range is possible
(2) Dual-components (CO2/CO)
2nd component
1st range
E
F
G
H
1st component
0 to 500ppm 0 to 1000ppm 0 to 2000ppm 0 to 5000ppm
1st range
CO2
H 0 to 5000 ppm
J 0 to 1%
K 0 to 2%
L 0 to 5%
M 0 to 10%
N 0 to 20%
P 0 to 50%
R 0 to 100%
–
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
CO
J
0 to 1%
K
0 to 2%
L
0 to 5%
M
0 to 10%
N
0 to 20%
V
0 to 25%
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
™
P
R
0 to 50% 0 to 100%
™
™
*™
™
™
™
™
™
™
™
™
™
™
™
™
™
1st component is CO2, 2nd component is CO.
™ : Dual-components are possible
Both components are able to have 2nd range.
2nd range is x 2 or x 2.5 of 1st range, choose the Code symbols
*™ : Dual-components are possible
But only one component, CO2 or CO, is able to have 2nd range. 2nd range is x 2 or x 2.5 of 1st range, choose the Code Symbols.
— : Impossible.
FUNDAMENTAL PRINCIPLE DIAGRAM
Infrared–ray light source
Gas outlet
Gas inlet
Detector
M
Mass flow
sensor
Measuring cell
Synchronuous motor
Rotary sector
DC amplifier
CPU
Input
Remote
range
Remote
output
hold
Contact
output
D/A
Output
0 to 1V,
4 to 20mA
DC
Rectifier
AC amplifier
A/D
Operating
panel
Display
Operation key
Range identification signal
Output during calibration
5
ZRH
OUTLINE DIAGRAM
(Unit:mm)
(1) Horizontal case
483
465
30
380
25
7
443
124.5
57.2
7
133
ZRH
Power switch
Operating panel
(Rack mounting plate)
<Front View>
<Side View>
Sample gas inlet
(Rc 1/4 or NPT 1/4)
Panel cutout dimensions
(in case of panel mounting)
External connection
terminal
Purge gas inlet
(Rc 1/4 or NPT 1/4)
(1st component)
126+2
–0
(2nd component)
Power terminals
(AUTO CAL)
436 +2
–0
435
<Rear View>
6
Sample gas outlet
(Rc 1/4 or NPT 1/4)
(2) Vertical case
424
440
Panel flush mounting
Panel mounting fixture
294
Door open -200°
283 radius
35 35 35 30
80
280
430+20
Rc1/4 or NPT1/4
(Purge gas inlet)
430+20
Rc1/4 or NPT1/4
(Sample gas inlet)
281+20
150
Rc1/4 or NPT1/4
(Sample gas outlet)
Panel cutout and spacing
178
118
148
170
Wiring hole (ø22)
with rubber bushing
350
424
484
454
440
Wall mounting
10
70
294
5
Door open -200°
283 radius
Rc1/4 or NPT1/4
(Sample gas outlet)
Rc1/4 or NPT1/4
(Sample gas inlet)
Rc1/4 or NPT1/4
(Purge gas inlet)
35 35 35 30
178
118
148
170
Wiring hole (ø22)
with rubber bushing
80
280
7
ZRH
CONNECTION DIAGRAM (Same for horizontal and vertical cases)
M4 screws used
1st component
Output 1:
0 to 1V DC
Output 2:
4 to 20mA DC
Fault
1
7
2
8
3
9
2nd component
Range
identification
contact
output
(option)
4 10
5 11
6 12
Output 1:
0 to 1V DC
Output 2:
4 to 20mA DC
1
7
2
8
3
9
AUTO CAL
Range
identification
contact
output
(option)
4 10
Remote hold
input(option)
5 11
6 12
Remote range
input(option)
Remote range
input(option)
Under
calibration
contact
output
Auto
calibration
failure
contact
output
Remote
start input
(option)
1
7
2
8
3
9
4 10
5 11
6 12
Zero gas
contact
output
Span gas 1
contact
output
Span gas 2
contact
output
SCOPE OF DELIVERY
ORDERING INFORMATION
1
1
1
2
4
1. Analyzer type.
2. Maximum, normal and minimum concentrations of sample
gas as well as type and content (percent by volume) of concomitant gas.
3. Temperatures (maximum, normal and minimum), pressure
and humidity of sample gas.
4. Dust conditions (mg/Nm3 or particle size, characteristics, etc.)
and environmental conditions.
5. Other items
x
x
x
x
x
gas analyzer main unit
test report
instruction manual
Power fuse
panel mounting bracket
RELATED DEVICES
•
•
•
•
Gas sampling device
Accommodating locker
Standard gas (for calibration)
Receiving instrument
Caution on Safety
*Before using this product, be sure to read its instruction manual in advance.
Head Office
6-17, Sanbancho, Chiyoda-ku, Tokyo 102-0075, Japan
http://www.fesys.co.jp/eng
Sales Div.
International Sales Dept.
No.1, Fuji-machi, Hino-city, Tokyo, 191-8502 Japan
Phone: 81-42-585-6201, 6202 Fax: 81-42-585-6187
http: //www.fic-net.jp/eng
Information in this catalog is subject to change without notice.
Printed in Japan