AD AD597AR

a
FEATURES
Low Cost
Operates with Type J (AD596) or Type K (AD597)
Thermocouples
Built-In Ice Point Compensation
Temperature Proportional Operation – 10 mV/8C
Temperature Setpoint Operation – ON/OFF
Programmable Switching Hysteresis
High Impedance Differential Input
Thermocouple Conditioner and
Setpoint Controller
AD596*/AD597*
FUNCTIONAL BLOCK DIAGRAM
TO-100
–ALM
+ALM
–IN
–
G
+
+IN
ICE
POINT
COMP
V+
–
G
+
+
GENERAL DESCRIPTION
The AD596/AD597 is a monolithic temperature setpoint controller that has been optimized for use at elevated temperatures
such as those found in oven control applications. The device
cold junction compensates and amplifies a type J or K thermocouple input to derive an internal signal proportional to temperature. The internal signal is then compared with an externally
applied setpoint voltage to yield a low impedance switched output
voltage. Dead-Band or switching hysteresis can be programmed
using a single external resistor. Alternately, the AD596/AD597 can
be configured to provide a voltage output (10 mV/°C) directly from
a type J or K thermocouple signal. It can also be used as a standalone voltage output temperature sensor.
The AD596/AD597 can be powered with a single supply from
+5 V to +30 V, or dual supplies up to a total span of 36 V.
Typical quiescent supply current is 160 µA, which minimizes
self-heating errors.
The AD596/AD597 H package option includes a thermocouple
failure alarm that indicates an open thermocouple lead when
operated in the temperature proportional measurement mode.
The alarm output has a flexible format which can be used to
drive relays, LEDs or TTL logic.
The device is packaged in a reliability qualified, cost effective
10-pin metal can or SOIC and is trimmed to operate over an
ambient temperature range from +25°C to +100°C. Operation
over an extended ambient temperature range is possible with
slightly reduced accuracy. The AD596 will amplify thermocouple signals covering the entire –200°C to +760°C temperature range recommended for type J thermocouples while the
AD597 can accommodate –200°C to +1250°C type K inputs.
The AD596/AD597 has a calibration accuracy of ±4°C at an
ambient temperature of 60°C and an ambient temperature
stability specification of 0.05°C/°C from +25°C to +100°C. If
higher accuracy, or a lower ambient operating temperature is
required, either the AD594 (J thermocouple) or AD595 (K
thermocouple) should be considered.
+A
AD596/
AD597
HYS
VOUT
FB
GND
V–
SOIC
AD597
8 –IN
+IN 1
–
+
G
7 V+
HYS 2
+A
+
GND 3
V– 4
+
G
–
ICE POINT
COMP
6 VOUT
5 FB
TOP VIEW
(Not to Scale)
PRODUCT HIGHLIGHTS
1. The AD596/AD597 provides cold junction compensation
and a high gain amplifier which can be used as a setpoint
comparator.
2. The input stage of the AD596/AD597 is a high quality instrumentation amplifier that allows the thermocouple to float
over most of the supply voltage range.
3. Linearization not required for thermocouple temperatures
close to 175°C (+100°C to +540°C for AD596).
4. Cold junction compensation is optimized for ambient temperatures ranging from +25°C to +100°C.
5. In the stand-alone mode, the AD596/AD597 produces an
output voltage that indicates its own temperature.
*Protected by U.S. Patent No. 4,029,974.
REV. B
Information furnished by Analog Devices is believed to be accurate and
reliable. However, no responsibility is assumed by Analog Devices for its
use, nor for any infringements of patents or other rights of third parties
which may result from its use. No license is granted by implication or
otherwise under any patent or patent rights of Analog Devices.
One Technology Way, P.O. Box 9106, Norwood, MA 02062-9106, U.S.A.
Tel: 781/329-4700
World Wide Web Site: http://www.analog.com
Fax: 781/326-8703
© Analog Devices, Inc., 1998
(@ +608C and VS = 10 V, Type J (AD596), Type K (AD597) Thermocouple,
AD596/AD597–SPECIFICATIONS unless otherwise noted)
Model
AD596AH
Typ
Max
Min
ABSOLUTE MAXIMUM RATINGS
+VS to –VS
Common-Mode Input Voltage
Differential Input Voltage
Alarm Voltages
+ALM
–ALM
Operating Temperature Range
Output Short Circuit to Common
TEMPERATURE MEASUREMENT
(Specified Temperature Range
+25°C to +100°C)
Calibration Error1
Stability vs. Temperature2
Gain Error
Nominal Transfer Function
AMPLIFIER CHARACTERISTICS
Closed Loop Gain3
Input Offset Voltage
Input Bias Current
Differential Input Range
Common-Mode Range
Common-Mode Sensitivity–RTO
Power Supply Sensitivity–RTO
Output Voltage Range
Dual Supplies
Single Supply
Usable Output Current4
3 dB Bandwidth
(–VS – 0.15)
–VS
–VS
–VS
–55
Indefinite
–4
± 0.02
–1.5
36
+VS
+VS
AD597AH
Typ
Min
(–VS – 0.15)
–VS
(–VS +36)
+VS
+125
–VS
–VS
–55
Indefinite
+4
± 0.05
+1.5
–4
± 0.02
–1.5
10
Min
36
+VS
+VS
(–VS – 0.15)
–VS
(–VS +36)
+VS
+125
–VS
–VS
–40
Indefinite
+4
± 0.05
+1.5
–4
POWER REQUIREMENTS
Operating
Quiescent Current
+VS
–VS
–1.5
(+VS – 2)
(+VS – 2)
245.5
°C × 41.27 – 37
0.1
–10
(+VS – 0.15)
1
(–VS + 2.5)
0
±5
Max
Units
36
+VS
+VS
Volts
Volts
Volts
(–VS +36)
+VS
+125
Volts
Volts
°C
+4
± 0.05
+1.5
°C
°C/°C
%
mV/°C
10
245.5
°C × 41.27 – 37
0.1
+50
(+VS – 4)
10
10
–10
(–VS – 0.15)
(+VS – 2)
(+VS – 2)
(–VS + 2.5)
0
±5
1
15
+50
(+V S – 4)
10
10
(+VS – 2)
(+VS – 2)
15
5
ALARM CHARACTERISTICS
VCE(SAT) at 2 mA
Leakage Current
Operating Voltage at – ALM
Short Circuit Current
± 0.02
10
180.6
°C × 53.21 + 235
0.1
–10
+50
(–VS – 0.15)
(+V S – 4)
10
1
10
(–VS + 2.5)
0
±5
15
AD597AR
Typ
Max
V/V
µV
µA
mV
Volts
mV/V
mV/V
Volts
Volts
mA
kHz
Alarm Function Not Pinned Out
0.3
0.3
61
(+VS – 4)
20
20
(+VS to –VS) ≤ 30
160
100
Volts
µA
Volts
mA
61
(+VS – 4)
(+VS to –VS) ≤ 30
(+VS to –VS) ≤ 30
160
100
160
100
300
200
300
200
300
200
Volts
µA
µA
NOTES
1
This is a measure of the deviation from ideal with a measuring thermocouple junction of 175°C and a chip temperature of 60°C. The ideal transfer function is given by:
AD596: VOUT = 180.57 × (Vm – Va + (ambient in °C) × 53.21 µV/°C + 235 µV)
AD597: VOUT = 245.46 × (Vm – Va + (ambient in °C) × 41.27 µV/°C – 37 µV)
where Vm , and Va represent the measuring and ambient temperatures and are taken from the appropriate J or K thermocouple table. The ideal transfer function minimizes the
error over the ambient temperature range of 25°C to 100°C with a thermocouple temperature of approximately 175°C.
2
Defined as the slope of the line connecting the AD596/AD597 CJC errors measured at 25°C and 100°C ambient temperature.
3
Pin 6 shorted to Pin 7.
4
Current Sink Capability in single supply configuration is limited to current drawn to ground through a 50 kΩ resistor at output voltages below 2.5 V.
5
Alarm function available on H package option only.
Specifications subject to change without notice.
Specifications shown in boldface are tested on all production units at final electrical test. Results from those tests are used to calculate outgoing quality levels. All min and max
specifications are guaranteed, although only those shown in boldface are tested on all production units.
ORDERING GUIDE
Model
Package Description
Package Options
AD596AH
AD597AH
AD597AR*
TO-100
TO-100
Plastic SOIC
H-10A
H-10A
SO-8
*Consult factory for availability.
–2–
REV. B
AD596/AD597
Table I. Output Voltage vs. Thermocouple Temperature (Ambient +608C, VS = –5 V, +15 V)
Thermocouple
Temperature
8C
Type J
Voltage
mV
AD596
Output
mV
Type K
Voltage
mV
Type J
Voltage
mV
AD596
Output
mV
Type K
Voltage
mV
AD597
Output
mV
–200
–180
–160
–140
–120
–7.890
–7.402
–6.821
–6.159
–5.426
–1370
–1282
–1177
–1058
–925
–5.891
–5.550
–5.141
–4.669
–4.138
–1446
–1362
–1262
–1146
–1016
500
520
540
560
580
27.388
28.511
29.642
30.782
31.933
5000
5203
5407
5613
5821
20.640
21.493
22.346
23.198
24.050
5066
5276
5485
5694
5903
–100
–80
–60
–40
–20
–4.632
–3.785
–2.892
–1.960
–.995
–782
–629
–468
–299
–125
–3.553
–2.920
–2.243
–1.527
–.777
–872
–717
–551
–375
–191
600
620
640
660
680
33.096
34.273
35.464
36.671
37.893
6031
6243
6458
6676
6897
24.902
25.751
26.599
27.445
28.288
6112
6321
6529
6737
6944
–10
0
10
20
25
–.501
0
.507
1.019
1.277
–36
54
146
238
285
–.392
0
.397
.798
1.000
–96
0
97
196
245
700
720
740
750
760
39.130
40.382
41.647
42.283
–
7120
7346
7575
7689
–
29.128
29.965
30.799
31.214
31.629
7150
7355
7560
7662
7764
30
40
50
60
80
1.536
2.058
2.585
3.115
4.186
332
426
521
617
810
1.203
1.611
2.022
2.436
3.266
295
395
496
598
802
780
800
820
840
860
–
–
–
–
–
–
–
–
–
–
32.455
33.277
34.095
34.909
35.718
7966
8168
8369
8569
8767
100
120
140
160
180
5.268
6.359
7.457
8.560
9.667
1006
1203
1401
1600
1800
4.095
4.919
5.733
6.539
7.338
1005
1207
1407
1605
1801
880
900
920
940
960
–
–
–
–
–
–
–
–
–
–
36.524
37.325
38.122
38.915
39.703
8965
9162
9357
9552
9745
200
220
240
260
280
10.777
11.887
12.998
14.108
15.217
2000
2201
2401
2602
2802
8.137
8.938
9.745
10.560
11.381
1997
2194
2392
2592
2794
980
1000
1020
1040
1060
–
–
–
–
–
–
–
–
–
–
40.488
41.269
42.045
42.817
43.585
9938
10130
10320
10510
10698
300
320
340
360
380
16.325
17.432
18.537
19.640
20.743
3002
3202
3402
3601
3800
12.207
13.039
13.874
14.712
15.552
2996
3201
3406
3611
3817
1080
1100
1120
1140
1160
–
–
–
–
–
–
–
–
–
–
44.439
45.108
45.863
46.612
47.356
10908
11072
11258
11441
11624
400
420
440
460
480
21.846
22.949
24.054
25.161
26.272
3999
4198
4398
4598
4798
16.395
17.241
18.088
18.938
19.788
4024
4232
4440
4649
4857
1180
1200
1220
1240
1250
–
–
–
–
–
–
–
–
–
–
48.095
48.828
49.555
50.276
50.633
11805
11985
12164
12341
12428
REV. B
AD597
Output
mV
Thermocouple
Temperature
8C
–3–
AD596/AD597
Excluding calibration errors, the above transfer function is accurate to within 1°C from +80°C to +550°C for the AD596 and
–20°C to +350°C for the AD597. The different temperature
ranges are due to the differences in J and K type thermocouple
curves.
TEMPERATURE PROPORTIONAL OUTPUT MODE
The AD596/AD597 can be used to generate a temperature
proportional output of 10 mV/°C when operated with J and K
type thermocouples as shown in Figure 1. Thermocouples produce low level output voltages which are a function of both the
temperature being measured and the reference or cold junction
temperature. The AD596/AD597 compensates for the cold
junction temperature and amplifies the thermocouple signal to
produce a high level 10 mV/°C voltage output which is a function only of the temperature being measured. The temperature
stability of the part indicates the sensitivity of the output voltage
to changes in ambient or device temperatures. This is typically
0.02°C/°C over the +25°C to +100°C recommended ambient
temperature range. The parts will operate over the extended
ambient temperature ranges from –55°C to +125°C, but thermocouple nonlinearity at the reference junction will degrade the
temperature stability over this extended range. Table I is a list of
ideal AD596/AD597 output voltages as a function of Celsius
temperature for type J and K ANSI standard thermocouples
with package and reference junction at 60°C. As is normally the
case, these outputs are subject to calibration and temperature
sensitivity errors. These tables are derived using the ideal transfer functions:
European DIN FE-CuNi thermocouple vary slightly from ANSI
type J thermocouples. Table I does not apply when these types
of thermocouples are used. The transfer functions given previously and a thermocouple table should be used instead.
Figure 1 also shows an optional trimming network which can be
used to change the device’s offset voltage. Injecting or sinking
200 nA from Pin 3 will offset the output approximately 10 mV
(1°C).
The AD596/AD597 can operate from a single supply from 5 V
to 36 V or from split supplies totalling 36 V or less as shown.
Since the output can only swing to within 2 V of the positive
supply, the usable measurement temperature range will be restricted when positive supplies less than 15 V for the AD597
and 10 V for the AD596 are used. If the AD596/AD597 is to be
used to indicate negative Celsius temperatures, then a negative
supply is required.
Common-mode voltages on the thermocouple inputs must
remain within the common-mode voltage range of the AD596/
AD597, with a return path provided for the bias currents. If the
thermocouple is not remotely grounded, then the dotted line
connection shown in Figure 1 must be made to one of the thermocouple inputs. If there is no return path for the bias currents,
the input stage will saturate, causing erroneous output voltages.
AD596 output = (Type J voltage + 301.5 µV) × 180.57
AD597 output = (Type K voltage) × 245.46
CONSTANTAN
(ALUMEL)
IRON
(CHROMEL)
+5V TO +30V
AD596/
AD597*
+15V
100kV
OPTIONAL
OFFSET 10kV
ADJUST
100kV
VOUT
SPAN OF
+5V TO +30V
1MV
–15V
In this configuration, the AD596/AD597 H package option has
circuitry which detects the presence of an open thermocouple. If
the thermocouple loop becomes open, one or both of the inputs
to the device will be deprived of bias current causing the output
to saturate. It is this saturation which is detected internally and
used to activate the alarm circuitry. The output of this feature
has a flexible format which can be used to source or sink up to
20 mA of current. The collector (+ALM) should not be allowed
to become more positive than (–VS + 36 V), however, it may be
permitted to be more positive than +VS. The emitter voltage
(–ALM) should be constrained such that it does not become
more positive than 4 V below +VS. If the alarm feature is not
used, this pin should be connected to Pins 4 or 5 as shown in
Figure 1. The alarm function is unavailable on the AR package
option.
0.01mF
0.01mF
0 TO –25V
*H PACKAGE PINOUT SHOWN
Figure 1. Temperature Proportional Output Connection
The offsets and gains of these devices have been laser trimmed
to closely approximate thermocouple characteristics over measurement temperature ranges centered around 175°C with the
AD596/AD597 at an ambient temperature between 25°C and
100°C. This eliminates the need for additional gain or offset
adjustments to make the output voltage read:
VOUT = 10 mV/°C × (thermocouple temperature in °C) (within
specified tolerances).
–4–
REV. B
AD596/AD597
SETPOINT CONTROL MODE
The AD596/AD597 can be connected as a setpoint controller as
shown in Figure 2. The thermocouple voltage is cold junction
compensated, amplified, and compared to an external setpoint
voltage. The relationship between setpoint voltage and temperature is given in Table I. If the temperature to be controlled is
within the operating range (–55°C to +125°C) of the device, it
can monitor its own temperature by shorting the inputs to
ground. The setpoint voltage with the thermocouple inputs
grounded is given by the expressions:
+VS
–
G
+
0.01mF
ICE
POINT
COMP
–
G
+
+
+A
AD596/
AD597*
AD596 Setpoint Voltage = °C × 9.6 mV/°C + 42 mV
AD597 Setpoint Voltage = °C × 10.1 mV/°C – 9.1 mV
*H PACKAGE PINOUT SHOWN
The input impedance of the setpoint pin of the AD596/AD597
is approximately 50 kΩ. The temperature coefficient of this
resistance is ± 15 ppm/°C. Therefore, the 100 ppm/°C 5 kΩ pot
shown in Figure 2 will only introduce an additional ± 1°C degradation of temperature stability over the +25°C to +100°C ambient temperature range.
TEMPERATURE
CONTROLLED CONSTANTAN
(ALUMEL)
REGION
R
VREF
SETPOINT
VOLTAGE
SETPOINT
VOLTAGE
5kV
100ppm/8C
HYSTERESIS
(OPTIONAL)
HEATER
DRIVER
*H PACKAGE PINOUT SHOWN
Figure 2. Setpoint Control Mode
Switching hysteresis is often used in setpoint systems of this type
to provide noise immunity and increase system reliability. By
reducing the frequency of on-off cycling, mechanical component
wear is reduced leading to enhanced system reliability. This can
easily be implemented with a single external resistor between
Pins 7 and 3 of the AD596/AD597. Each 200 nA of current
injected into Pin 3 when the output switches will cause about
1°C of hysteresis; that is:
RHYST (Ω) =
0.01mF
–VS
Figure 3. Stand-Alone Temperature Transducer
Temperature Proportional Output Connection
TEMPERATURE
COMPARATOR
0.01mF OUTPUT
+V
AD596/
AD597*
IRON
(CHROMEL)
Simply omit the thermocouple and connect the inputs (Pins 1
and 2) to common. The output will now reflect the compensation voltage and hence will indicate the AD596/AD597 temperature. In this three terminal, voltage output, temperature
sensing mode, the AD596/AD597 will operate over the full
extended –55°C to +125°C temperature range. The output
scaling will be 9.6 mV per °C with the AD596 and 10.1 mV per
°C with the AD597. Additionally there will be a 42 mV offset
with the AD596 causing it to read slightly high when used in
this mode.
THERMOCOUPLE CONNECTIONS
The connection of the thermocouple wire and the normal wire
or printed circuit board traces going to the AD596/AD597
forms an effective reference junction as shown in Figure 4. This
junction must be kept at the same temperature as the AD596/
AD597 for the internal cold junction compensation to work
properly. Unless the AD596/AD597 is in a thermally stable
enclosure, the thermocouple leads should be brought in directly
to Pins 1 and 2.
REFERENCE JUNCTION
LIMITING RESISTOR
TO
LED
0.01mF
CONSTANTAN
(ALUMEL)
IRON
+VS
(CHROMEL)
AD596/
AD597*
NOTE:
A BIAS RETURN PATH
FROM PINS 1 AND 2
OF LESS THAN 1kV
IMPEDANCE MUST BE
PROVIDED.
V OUT
1
×
200 nA °CHYST
VOUT
In the setpoint configuration, the AD596/AD597 output is
saturated at all times, so the alarm transistor will be ON regardless of whether there is an open circuit or not. However, –ALM
must be tied to a voltage below (+VS – 4 V) for proper operation
of the rest of the circuit.
0.01mF
*H PACKAGE PINOUT SHOWN
STAND-ALONE TEMPERATURE TRANSDUCER
The AD596/AD597 may be configured as a stand-alone Celsius
thermometer as shown in Figure 3.
REV. B
VOUT
9.6mV/8C
GND
–VS
Figure 4. PCB Connections
To ensure secure bonding, the thermocouple wire should be
cleaned to remove oxidization prior to soldering. Noncorrosive
resin flux is effective with iron, constantan, chromel, and
alumel, and the following solders: 95% tin–5% silver, or 90%
tin–10% lead.
–5–
AD596/AD597
temperature differences will result in a direct error at the output. In the temperature proportional mode, the alarm feature
will only activate in the event of an open thermocouple or system transient which causes the device output to saturate.
Self-Heating errors will not effect the operation of the alarm but
two cases do need to be considered. First, after a fault is corrected and the alarm is reset, the AD596/AD597 must be allowed to cool before readings can again be accurate. This can
take 5 minutes or more depending upon the thermal environment seen by the device. Second, the junction temperature of
the part should not be allowed to exceed 150°C. If the alarm
circuit of the AD596/AD597 is made to source or sink 20 mA
with 30 V across it, the junction temperature will be 90°C above
ambient causing the die temperature to exceed 150°C when
ambient is above 60°C. In this case, either the load must be
reduced, or a heat sink used to lower the thermal resistance.
SINGLE AND DUAL SUPPLY CONNECTIONS
In the single supply configuration as used in the setpoint controller of Figure 2, any convenient voltage from +5 V to +36 V
may be used, with self-heating errors being minimized at lower
supply levels. In this configuration, the –VS connection at Pin 5
is tied to ground. Temperatures below zero can be accommodated in the single supply setpoint mode, but not in the single
supply temperature measuring mode (Figure 1 reconnected for
single supply). Temperatures below zero can only be indicated
by a negative output voltage, which is impossible in the single
supply mode.
Common-mode voltages on the thermocouple inputs must
remain below the positive supply, and not more than 0.15 V
more negative than the minus supply. In addition, a return path
for the input bias currents must be provided. If the thermocouple is not remotely grounded, then the dotted line connections in Figures 1 and 2 are mandatory.
TEMPERATURE READOUT AND CONTROL
Figure 6 shows a complete temperature indication and control
system based on the AD596/AD597. Here the AD596/AD597 is
being used as a closed-loop thermocouple signal conditioner
and an external op amp is used to implement setpoint. This has
two important advantages. It provides a high level (10 mV/°C)
output for the A/D panel meter and also preserves the alarm
function for open thermocouples.
STABILITY OVER TEMPERATURE
The AD596/AD597 is specified for a maximum error of ± 4°C at
an ambient temperature of 60°C and a measuring junction
temperature at 175°C. The ambient temperature stability is
specified to be a maximum of 0.05°C/°C. In other words, for
every degree change in the ambient temperature, the output will
change no more than 0.05 degrees. So, at 25°C the maximum
deviation from the temperature-voltage characteristic of Table I
is ± 5.75°C, and at 100°C it is ± 6°C maximum (see Figure 5). If
the offset error of ± 4°C is removed with a single offset adjustment, these errors will be reduced to ± 1.75°C and ± 2°C max.
The optional trim circuit shown in Figure 1 demonstrates how
the ambient offset error can be adjusted to zero.
The A/D panel meter can easily be offset and scaled as shown to
read directly in degrees Fahrenheit. If a two temperature calibration scheme is used, the dominant residual errors will arise
from two sources; the ambient temperature rejection (typically
± 2°C over a 25°C to 100°C range) and thermocouple nonlinearity typical +1°C from 80°C to 550°C for type J and +1°C
from –20°C to 350°C for type K.
An external voltage reference is used both to increase the stability of the A/D converter and supply a stable reference for the
setpoint voltage.
+2.08C
+1.758C
MAXIMUM
A traditional requirement for the design of setpoint control
thermocouple systems has been to configure the system such
that the appropriate action is taken in the event of an open
thermocouple. The open thermocouple alarm pin with its flexible current-limited output format supports this function when
the part operates in the temperature proportional mode. In
addition, if the thermocouple is not remotely grounded, it is
possible to program the device for either a positive or negative
full-scale output in the event of an open thermocouple. This is
done by connecting the bias return resistor directly to Pin 1 if a
high output voltage is desired to indicate a fault condition. Alternately, if the bias return is provided on the thermocouple lead
connected to Pin 2, an open circuit will result in an output low
reading. Figure 6 shows the ground return connected to Pin 1
so that if the thermocouple fails, the heater will remain off. At
the same time, the alarm circuit lights the LED signalling the
need to service the thermocouple. Grounding Pin 2 would lead
to low output voltage saturation, and in this circuit would result
in a potentially dangerous thermal runaway under fault conditions.
+0.88C
0
TYPICAL
–0.88C
MAXIMUM
–1.758C
–2.08C
258C
608C
1008C
Figure 5. Drift Error vs. Temperature
THERMAL ENVIRONMENTAL EFFECTS
The inherent low power dissipation of the AD596/AD597 keeps
self-heating errors to a minimum. However, device output is
capable of delivering ± 5 mA to an external load and the alarm
circuitry can supply up to 20 mA. Since the typical junction to
ambient thermal resistance in free air is 150°C/W, significant
temperature difference between the package pins (where the
reference junction is located) and the chip (where the cold junction temperature is measured and then compensated) can exist
when the device is operated in a high dissipation mode. These
–6–
REV. B
AD596/AD597
+V
TEMPERATURE
CONSTANTAN
(ALUMEL)
IRON
(CHROMEL)
READOUT 8F
470V
–
LCD DISPLAY
+
+V
AD596/
AD597*
HEATER
ICL7136
45.2kV
IN HI
10kV
IN LO
1.27MV
+V
*H PACKAGE PINOUT SHOWN
5V
40.2kV
REF HI
AD584
10kV
REF LO
SET-POINT
ADJUST
5kV
–
10kV
1kV
OP07
+
10MV
Figure 6. Temperature Measurement and Control
REV. B
–7–
120V AC
AD596/AD597
OUTLINE DIMENSIONS
Dimensions shown in inches and (mm).
10-Pin Metal Can
(TO-100)
0.185 (4.70)
0.165 (4.19)
0.160 (4.06)
0.110 (2.79)
0.250 (6.35) MIN
0.050 (1.27) MAX
6
7
0.370 (9.40)
0.335 (8.51)
0.335 (8.51)
0.305 (7.75)
5
0.115
(2.92)
BSC
C831b–5–2/98
REFERENCE PLANE
0.750 (19.05)
0.500 (12.70)
8
4
0.045 (1.14)
0.027 (0.69)
9
3
2
0.019 (0.48)
0.016 (0.41)
10
1
0.230 (5.84)
BSC
0.021 (0.53)
0.016 (0.41)
0.040 (1.02) MAX
0.045 (1.14)
0.010 (0.25)
0.034 (0.86)
0.027 (0.69)
36° BSC
BASE & SEATING PLANE
8-Lead Small Outline (SOIC)
(SO-8)
0.1968 (5.00)
0.1890 (4.80)
0.1574 (4.00)
0.1497 (3.80)
PIN 1
0.0098 (0.25)
0.0040 (0.10)
8
5
1
4
0.2440 (6.20)
0.2284 (5.80)
0.0688 (1.75)
0.0532 (1.35)
8°
0° 0.0500 (1.27)
0.0160 (0.41)
PRINTED IN U.S.A.
0.0500 0.0192 (0.49)
SEATING (1.27)
0.0098 (0.25)
PLANE BSC 0.0138 (0.35) 0.0075 (0.19)
0.0196 (0.50)
x 45°
0.0099 (0.25)
–8–
REV. B