INTERSIL AN1690.0

Application Note 1690
Electronics Meets the Challenges of Patient Monitors
The medical industry is keeping pace with other major
markets in terms of integration. Single units such as blood
pressure cuffs, ECGs and oxygen sensors have successfully
entered consumer markets, but there is increasing interest in
patient monitors, which encompass all of those monitoring
abilities and more.
piezo-resistive pressure transducer. For simplicity, several of
these biometric modules may utilize common digital, power
and IO subsystems.
Blood Pressure
In blood pressure biometric modules, the most critical function
is the pressure sensor circuit. Here precision amplifiers are
used to detect very small signals from the transducer and
amplify them to a level suitable for ADC processing. This is
typically followed by an active filter to limited unwanted noise
at higher frequencies. Amplifiers with low noise, low drift and
high gain are necessary to minimize measurement errors and
ensure accurate readings. See Figure 2 for a system block
diagram.
The medical market is generally grouped into three major
subcategories: home, clinical and imaging. Home medical is
predominately lower cost, portable equipment that has lower
performance requirements. Clinical and hospital grade
equipment is generally higher performance and therefore
more expensive. Imaging systems are generally very large
mainframe bases systems, with the exception of ultrasound,
that is rapidly expanding into portable and cart-based systems.
Fundamentally, most of these systems are analog sensor
measurement systems, but applied to biometric functions,
such as blood pressure, body temperature or heart rate. These
biometric sensors are designed to measure physical events,
such as temperature, pressure, light and flow. After
measurement, the system converts them into a corresponding
voltage or current. The signal is then conditioned and digitized
for processing and analysis. On the control side, the signal is
converted back to a current or voltage and applied to an
actuator to control such things as air flow, oxygen or
temperature.
Most commonly used piezo-resistive silicon pressure sensor in
medical applications is the Wheatstone bridge. The pressure
sensing element combines resistors and an etched diaphragm
structure to provide an electrical signal that changes with
pressure. As the diaphragm moves under pressure, stress is
concentrated in specific areas of the silicon element. The
result is a small voltage that changes proportional to the
pressure applied to the diaphragm. This bridge signal is then
amplified using precision op amps prior to ADC conversion.
Key questions to ask when recommending an amplifier arewhat is the required accuracy and what are the required
voltages? Hospital grade equipment has different
requirements than portable home based systems while
pressure sensors have varying sensitivities and voltage
requirements. The amplifier will generally be selected to
match the requirement of the sensor. The ISL28127 and
ISL28217 are excellent op amps for ±5V pressure sensor amp
gain frontends due to their low noise and low DC offset and
drift.
A typical high end patient monitor system has 5 basic
subsystems: ECG, pulse oximetry, blood pressure, body
temperature and respiration. Typically the most critical
components in each system are the sensor circuits.
Each module uses a different sensor and signal conditioning
circuit. For example, the ECG uses electrodes to measure the
electric pulse from the heart. The pulse oximetry (SpO2) uses a
light-emitting diode and light sensor to measure oxygen
content. Blood pressure is typically measured using a
B IO M E T R IC
A N A L O G /D IG IT A L M O D U L E S
S Y S T E M I/F
& CONTROL
E C G /E K G
IN A M P
SPO2
P R E C IS IO N
OP AMP
A U D IO A M P
IN T E R F A C E
PROCESSOR
BLO O D
PRESSURE
U S B /K B D
LEVEL XLATO R
MEMORY
VO LTAG E
REF
A /D
U S E R IN T E R F A C E
BODY
TEMPERATURE
DCP
R E S P IR A T O R Y
S W IT C H /
MUX
TO UCH
SCREEN
CONTROL
A M B IE N T L IG H T
SENSOR
D IS P L A Y I/F
V ID E O D R IV E R
POW ER
CORE &
I/O P O W E R
BATTERY
MANAGEMENT
SYSTEM
POW ER
FUEL
GAUGE
A C /D C
BATTERY
CHARGER
LCD
C O M M U N IC A T IO N
O R TELEM ETRY
OR
W IR E L E S S
ETHERNET
FIGURE 1. BLOCK DIAGRAM OF A PATIENT MONITOR
October 28, 2011
AN1690.0
1
CAUTION: These devices are sensitive to electrostatic discharge; follow proper IC Handling Procedures.
Copyright Intersil Americas Inc. 2011. All Rights Reserved.
1-888-INTERSIL or 1-888-468-3774 | Intersil (and design) is a trademark owned by Intersil Corporation or one of its subsidiaries.
All other trademarks mentioned are the property of their respective owners.
Application Note 1690
System Management
V MON
Gain
Precision
amp
LCD
Controller
+ Vs
Precision
amp
-Vs
ADC
DAC
Class D
Amp
Audio Amp
VOLTA GE
REFERENC E
PRE CISI ON
AMPS
uC, uP, or
FPGA
-Vs
Sensor Compensation
VOLTA GE
REFERENCE
Interface
R S-232
Low-Pass Filter
+ Vs
Pressure
Sensor
S equencers
RTC
5V/12V/24V
BATTE RY
CHARGERS/
SAFETY
DCP
LDO
AC/DC
Module
DC-DC
FIGURE 2. BLOCK DIAGRAM OF A BLOOD PRESSURE SYSTEM
DIAGNOSTIC ECG MULTI-LEAD (e.g. 12-lead)
[1 of n* channels shown below]
Gain / Hi-Pass Filter
Lead 1
+5V
+5V
+5V
Low Pass Filter
+5V
1 of n ADC
+5V
M
U
X
IN-AMP
-5V
+5V
Lead 1
-5V
Lead 2
Lead 3
Core &
I/O Power
PACE
ARTIFACT
DETECTION
ADC
ISOLATION
VOLTAGE
REFERENCE
SIGNAL
PROCESSOR
FPGA
+5V
ISOLATION
LEAD
DETECTION
-5V
VMON
AC/ DC
Battery
Managment
System
Power
FIGURE 3. BLOCK DIAGRAM OF AN ECG SYSTEM
Electrocardiogram (ECG)
There are several precision amplifier and instrumentation
amplifier opportunities in ECG applications. Shown in Figure 3 is
a diagnostic (or clinical) ECG with up to 12-leads. Key blocks for
lead devices are the electrode gain amplifier, high pass filter
(usually 0.5Hz), low pass filter (around 150Hz) and right leg drive
circuit. For ECG, each electrode requires a precision
instrumentation amp to extract a very small signal that rides on a
300-700mV common mode voltage. Typically, this amplifier will
use a higher supply voltage to enable a high gain without railing
the amplifier in the presence of the high common mode voltage
from the body. This can be a discrete instrumentation amp or an
integrated instrumentation amplifier. Second and third stage
active filter amplifiers are needed to set a very specific band
(0.5Hz - 150Hz) to capture the ECG QRS wave signal. Typically,
these will be low noise 5V amplifiers with good bandwidth. In
2
addition, low noise, low power amplifiers are needed for the right
leg drive feedback function.
In multi-channel systems, such as a 12-lead ECG monitor, it is
common to multiplex signals into a common ADC. The key
requirements for the multiplexer (mux) are typically low
on-resistance and low charge injection. Generally a specific mux
is selected to match the voltage requirements of the filter
amplifiers and the ADC. It is also common for multichannel ECGs
to have automated lead detection to enable multi-configuration
operations. Generally, a low on-resistance switch is used in this
circuit as well.
Multiplexers like the ISL43681 and ISL43640 are excellent
choices for medical devices as they can operate from 2V to 12V.
In addition, they have low on-resistance of 39Ω to 60Ω, which
lowers distortion and reduces kick-back voltage from the ADC.
Their low charge injection of 0.3pC to 2pC reduces error
AN1690.0
October 28, 2011
Application Note 1690
contribution on charge redistribution ADCs. They are also low
power and available in the very small QFN package.
For 3-lead portable ECG applications, Iow power instrumentation
amps from a CMOS process may be a better selection. Since
CMOS inputs naturally provide a high impedance input, the need
for external buffers is eliminated, saving cost. A good choice for
the input amplifier can deliver active feedback with a very
precise base line compensation voltage, low 1/F noise,
extremely low offset and low drift versus temperature.
Pulse Oximetry (SpO2)
Oxygen is carried in the blood in hemoglobin and is one of the
key vital signs needing detection. Pulse oximetry takes
advantage of the fact the blood absorbs certain wavelengths of
light differently when it is oxygenated than when it is oxygen
deprived.
The wavelength that marks the difference in absorption to
identify oxygen concentration is approximately 805nm for adult
hemoglobin. Therefore, we will use two other wavelengths-one
above and one below to calculate the percentage of oxygenated
red blood cells. Usually 660nm and 940nm are used. A high
impedance, low bias current op amp is needed to process the
photodiodes that receive the signal at these wavelengths. The
ADC also needs to have the fast throughput of a 16-bit device.
The DC and background noise is subtracted out, and the pulsed
signals are then scaled. Extensive over-sampling, filtering and
signal processing are needed to eliminate noise, such as
movement artifacts from the small signals, as well as allowing
for the pulse rate to be measured.
Conclusion
The medical market has developed enough to support the needs
of a wide range of amplifiers, multiplexers, converters,
references, interface products and power supply products. The
portable instruments need lowest power and meticulous battery
management above and beyond the stringent noise and gain
requirements. The non-portable instruments need reliable
isolation and patient protection in addition to their precise
measurement capability.
It is difficult, if not impossible, to offer the best solution for all
situations. Every trade-off of power consumption, available
supplies, required resolution, portability, noise and more will play
into the selection of the key components. Understanding the
implications of device choice will ensure that your system safely
delivers quality for the lifetime of the system.
FIGURE 4. BLOCK DIAGRAM FOR PULSE OXIMETRY SYSTEM
Intersil Corporation reserves the right to make changes in circuit design, software and/or specifications at any time without notice. Accordingly, the reader is
cautioned to verify that the Application Note or Technical Brief is current before proceeding.
For information regarding Intersil Corporation and its products, see www.intersil.com
3
AN1690.0
October 28, 2011