WALL PSRL0402DMX-0512

PSRL0402DM Series
400 Watt with Airflow
Dual Output Medical Approved
AC/DC Medical Power Supply with Active PFC
FEATURES
● 2 Year Warranty
● Both Medical & ITE Safety Approvals
● Power Factor Corrected to EN61000-3-2 Class D
● 3.3/ 5/ 12/ 24V Dual Output Optional Combinations
● U-Chassis Type Drive Max 250W Under Air Convection
● U-Chassis & Enclosed with Built-in Fan Mechanical Options
● 1U Height Size and High Power Density: 6.25 Watts/cu Inches
SPECIFICATIONS: PSRL0402DM Series
E Type (Enclosed with built-in Fan): 9(L) x 5(W) x 1.6(H) inches.
U Type (U-Chassis Type): 8(L) x 5(W) x 1.6(H) inches.
o
All specifications are based on 25 C, Nominal Input Voltage, and Maximum Output Current unless otherwise noted.
We reserve the right to change specifications based on technological advances.
INPUT SPECIFICATIONS
Input Voltage
90 – 264VAC Full Range
Input Frequency
47 to 63Hz
Input Current
6.35A at 90VAC full load
Inrush Current
35A max @ 230VAC with full load and cold start.
Leakage Current
Medical degree 300µA; ITE degree 1.5mA.
Remote ON/OFF
Designated as RSW on the CN3, requires a low signal to inhibit output.
OUTPUT SPECIFICATIONS
Output Voltage
See Table
Output Power Range
400 Watts max with airflow. (See Note 4)
Output Adjustability
Output user adjustable ±5% minimum.
Total Regulation
±5%
Output Current
See Table
Ripple & Noise (peak to peak)
±1%
Transient Response
Returns to within 1% in less than 2.5ms for a 50% load change and the peak transient does not excess 5%.
Hold-Up Time
20ms min. at 80% of full load.
Overshoot
Turn-on/off not exceed 5% over nominal voltage.
Turn On Delay
1 second maximum at 120VAC.
Remote Sense
Designated as RS+ and RS- on the CN3.
PROTECTION
Over Voltage Protection
Latching down will occur when output voltage exceeds 130% and recycle AC input to reset.
Short Circuit Protection
Trip without damage and auto-recovery.
Over-Temperature Protection
Protected in the event of excessive operating ambient 85°C and automatic recovery.
Over-Power Protection
C.C. mode 110-140% and auto-recovery.
Input Voltage Protection
Power shut down under 80 ±5VAC, and recovered over 86VAC.
Input Circuit Protection (Primary)
Two T8A/250V fuses inserted.
GENERAL SPECIFICATIONS
Switching Frequency
30KHz fixed frequency
Efficiency
75% minimum (measuring at 230Vand full load).
1500 VAC input line to chassis (2mA DC cut off current); 4000VAC between primary and secondary windings.
Withstand Voltage
Primary to core 1500VAC. All for 3 seconds.
Burn In
45 ±5°C for one hour @ 230VAC with full load.
PFC
Active power factor correction meets EN61000-3-2 class D.
Designated as PG on the CN3 will go high 100-500ms after regulation and goes low 1ms before loss of
Power Good
regulation.
Power Supply On
Green LED designated as LED 1 on the PCB.
Grounding Test
Apply 40A from ground pin to the earthed connection point. Max allowable resistance is 0.1 ohm.
Bi-color green LED in front panel (E Type only). Any protection occurred or RSW applied low signal will emit
LED Display
orange.
Wall Industries, Inc. 5 Watson Brook Road Exeter, NH 03833
603-778-2300 www.wallindustries.com Fax 603-778-9797
PSRL0402DM Series
400 Watt with Airflow
Dual Output Medical Approved
AC/DC Medical Power Supply with Active PFC
SPECIFICATIONS (CONTINUED)
ENVIRONMENTAL SPECIFICATIONS
Operating Temperature
Storage Temperature
Operating Humidity
Storage Humidity
Vibration
Cooling
Fan Drive
MTBF
PHYSICAL SPECIFICATIONS
Weight
Dimensions
Warranty
SAFETY
Emissions
Safety Regulations
0°C to +70°C ambient, de-rating at 2.5% per degree from 50°C to 70°C.
-20°C to +85°C
5% to 90% RH, non-condensing
5% to 95% RH, non-condensing
Frequency 5 ~ 50Hz, acceleration ±7.35 m/(s x s) on X, Y, and Z axis.
U Type (U-Chassis): 400W max with 23CFM airflow or 250W max under convection cooling.
E Type (Enclosed with side built-in fan): 400W
12VDC/400mA is available to drive an external fan.
100,000 hours (according to MIL-HBK-217F) at 30°C.
U Type (U-Chassis): 1.3Kg
E Type (Enclosed with side built-in fan): 1.6Kg
U Type (U-Chassis): 8(L) x 5(W) x 1.6(H) inches.
E Type (Enclosed with side built-in fan): 9(L) x 5(W) x 1.6(H) inches.
2 years
FCC part15, CISPR 22 Class B, Conducted.
Approved to UL60950-1/ 60601-1, CSA C22.2 No. 60950-1-03/ 601.1-M90, TUV EN60950-1/ 60601-1,
CE Mark (LVD) EN61204-3/ 60601-1-2/ 61000-3-2,3, and IEC61000-4 Series Regulations and CB.
OUTPUT VOLTAGE / CURRENT RATING CHART
Model
PSRL0402DMx-0312
PSRL0402DMx-0324
PSRL0402DMx-0512
PSRL0402DMx-0524
PSRL0402DMx-1224
Output Voltage
Range
V1: +3.3 VDC
V2: +12 VDC
V1: +3.3 VDC
V2: +24 VDC
V1: +5 VDC
V2: +12 VDC
V1: +5 VDC
V2: +24 VDC
V1: +12 VDC
V2: +24 VDC
Output Current
Convection
22.95CFM
30A
40A
16.7A
25A
30A
40A
8.34A
12.5A
30A
40A
16.7A
25A
30A
40A
8.34A
12.5A
16.7A
25A
8.3A
12.5A
Regulation
Ripple & Noise
± 5%
± 5%
± 5%
± 5%
± 5%
± 5%
± 5%
± 5%
± 5%
± 5%
± 1%
± 1%
± 1%
± 1%
± 1%
± 1%
± 1%
± 1%
± 1%
± 1%
NOTES
1. PSRL0402DM Series is designated as PSRL0402DMx-y where x can be U-chassis (Type U), U-chassis with top cover
(Type C), or Enclosed with side built-in Fan (Type E), y can be 0312, 0324, 0512, 0524 and 1224 for output voltage.
2. 10% minimum load is required for all outputs to maintain the ripple and regulation.
3. Output is fully isolated.
4. Total combined power of V1 and V2: 400W max. with 22.95 CFM forced air and 250W convection cooling for
PSRL0402DMU-1224; 400W for PSRL0402DME-1224 by self-cooling. 300W max. with 22.95 CFM forced air and 200W
convection cooling for other PSRL0402DMU-y models; 300W for other PSRL0402DME-y models by self-cooling.
Wall Industries, Inc. 5 Watson Brook Road Exeter, NH 03833
603-778-2300 www.wallindustries.com Fax 603-778-9797
PSRL0402DM Series
400 Watt with Airflow
Dual Output Medical Approved
AC/DC Medical Power Supply with Active PFC
MECHANICAL DRAWINGS
PSRL0402DMU Series (U-Chassis Type): 8(L) x 5(W) x 1.6(H) inches; Weight: 1.3kg; Option: Top Cover.
PSRL0402DME Series (Enclosed with built-in Fan Type): 9(L) x 5(W) x 1.6(H) inches; Weight: 1.6kg.
Output Pin Assignment
V1
V2
RTN
Howder
Pins 1 – 2
Pin 6
Pins 3 – 5
Molex
Pins 1 – 6
Pins 14 – 16
Pins 7 – 13
Wall Industries, Inc. 5 Watson Brook Road Exeter, NH 03833
603-778-2300 www.wallindustries.com Fax 603-778-9797
PSRL0402DM Series
400 Watt with Airflow
Dual Output Medical Approved
AC/DC Medical Power Supply with Active PFC
I/O CONNECTOR PIN ASSIGNMENT
Input Connector (CN1):
PSRL0402DMU Series: mating Molex Part No. 09-91-0700 equivalent (7 pin, 5 used), or Howder Terminal block (HD-121-3P).
PSRL0402DME Series: IEC320 or equivalent Snap-in mounting type or DINKLE Terminal block (DT-35-A02W-03).
Output Connector (CN2):
Mating Molex 16 pins (09-91-1600), or Howder (HD-121-6P) M3.5, 8 pins terminal block, 9.5MM Center.
Output Pin Assignment:
(See table)
Logic signal connectors (CN3):
Mating JST XHP-9 or equivalent (CHYAO SHIUNN JS-2001-05) Mating Pins: JST SXH-002T-P0.6 for AWG 30 to 26.
Mounting Inserts:
6-32, M4 4 Places individually with maximum penetration 0.15 inches on bottom side and 0.25 inch on both sides
Wall Industries, Inc. 5 Watson Brook Road Exeter, NH 03833
603-778-2300 www.wallindustries.com Fax 603-778-9797