TOTAL-POWER MPS-30-15

TOTAL POWER INT'L
30W Single Output for Medical Type
MPS-30
series
Features :
Universal AC input / Full range
Low leakage current
0.3mA
Protections:Short circuit/Over load/Over voltage/Over temperature
Fixed switching frequency at 100KHz
Low cost
High reliability
3 years warranty
SPECIFICATION
MODEL
OUTPUT
MPS-30-5
MPS-30-12
MPS-30-15
MPS-30-24
MPS-30-27
DC VOLTAGE
5V
12V
15V
24V
27V
48V
RATED CURRENT
5A
2.5A
2A
1.2A
1.1A
0.6A
CURRENT RANGE
0 ~ 5A
0 ~ 2.5A
0 ~ 2A
0 ~ 1.2A
0 ~ 1.1A
0 ~ 0.6A
RATED POWER
25W
30W
30W
28.8W
29.7W
28.8W
120mVp-p
150mVp-p
240mVp-p
240mVp-p
240mVp-p
RIPPLE & NOISE (max.) Note.2 80mVp-p
VOLTAGE TOLERANCE Note.3 3.0%
LINE REGULATION
LOAD REGULATION
INPUT
3.0%
3.0%
3.0%
3.0%
2.0%
1.0%
1.0%
0.5%
0.5%
0.5%
0.5%
2.0%
1.0%
1.0%
1.0%
1.0%
1.0%
SETUP, RISE, HOLD TIME
500ms, 30ms, 50ms/230VAC
VOLTAGE RANGE
88 ~ 264VAC auto switch
FREQUENCY RANGE
47 ~ 63Hz
EFFICIENCY(Typ.)
72%
AC CURRENT
INRUSH CURRENT(max.)
0.8A/115VAC
LEAKAGE CURRENT
OVER LOAD
PROTECTION OVER VOLTAGE
OVER TEMPERATURE
500ms, 30ms, 10ms/115VAC at full load
120 ~ 370VDC
75%
76%
77%
78%
78%
29.7 ~ 36.45V
52.8 ~ 64.8V
0.5A/230VAC
COLD START 35A/230VAC
<0.3mA / 264VAC
Above 105% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
5.5 ~ 6.75V
13.2 ~ 16.2V
16.5 ~ 20.25V
Protection type : Shut down o/p voltage, re-power on to recover
Tj 135 typically (U1) Detect on main control IC
26.4 ~ 32.4V
WORKING TEMP.
Protection type : Shut down o/p voltage, recovers automatically after temperature goes down
-10 ~ +50 (Refer to output load derating curve)
WORKING HUMIDITY
20 ~ 90% RH non-condensing
ENVIRONMENT STORAGE TEMP., HUMIDITY
TEMP. COEFFICIENT
VIBRATION
SAFETY STANDARDS
WITHSTAND VOLTAGE
MPS-30-48
-20 ~ +85 , 10 ~ 95% RH
0.03%/ (0 ~ 50
10 ~ 500Hz, 2G 10min./1cycle, Period for 60min.each along X, Y, Z axes
UL2601-1, TUV EN60601-1, IEC601-1 Approved
I/P-O/P:4KVAC I/P-FG:1.5KVAC O/P-FG:SHORT
I/P-O/P, I/P-FG:100M Ohms/500VDC
SAFETY & ISOLATION RESISTANCE
EMI CONDUCTION & RADIATION Compliance to EN55011 (CISPR11) Class B
EMC
(Note 4)
OTHERS
HARMONIC CURRENT
Compliance to EN61000-3-2,-3
EMS IMMUNITY
MTBF
Compliance to EN61000-4-2,3,4,5,6,8,11; ENV50204, EN55024, EN60601-1-2 Medical level, criteria A
DIMENSION
PACKING
NOTE
547Khrs min. MIL-HDBK-217F (25
101.6*65.8*23.5mm (L*W*H)
0.16Kg; 90pcs/15.8Kg/1.02CUFT
)
1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature.
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uf & 47uf parallel capacitor.
3. Tolerance : includes set up tolerance, line regulation and load regulation.
4. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets
EMC directives.
MPS-30
30W Single Output For Medical Type
Mechanical Specification
series
Unit:mm
101.6
66.1
95.2
18
3.2
3.2
51.4
6 2 .6
6 5 .8
CN2
CN1
4
3
2
1
M3
L=
5
4
3
2
1
6m
m
FUSE*2
T2.5A 250V
3 .2
23.5 max.
4-
DC Output Connector (CN2) : Molex 41791-4 or equivalent
AC Input Connector (CN1) : Molex 41791-5 or equivalent
Pin No.
1
2,4
3
5
Assignment
FG
No Pin
AC/N
AC/L
Mating Housing
Terminal
Molex 2139
or equivalent
Molex 2478
or equivalent
Pin No.
1,2
3,4
Assignment
-V
+V
Mating Housing
Molex 2139
or equivalent
Terminal
Molex 2478
or equivalent
Block Diagram
fosc : 100KHz
RECTIFIERS
&
FILTER
EMI
FILTER
I/P
POWER
SWITCHING
RECTIFIERS
&
FILTER
PWM
CONTROL
+V
-V
DETECTION
CIRCUIT
FG
CROWBAR
OVP
Output Derating
Static Characteristics
OUTPUT VOLTAGE (V)
100
80
LOAD (%)
60
40
20
350
28
300
24
250
20
200
16
150
12
100
50
8
-10
0
10
20
30
40
AMBIENT TEMPERATURE (
50
)
60
(HORIZONTAL)
60
80
100
120
140
160
180
200
INPUT VOLTAGE (V) 60Hz
220
240
260
OUTPUT RIPPLE (mVp-p)
Ta=25
32