How to configure trigger output signals

Agilent B2961A/B2962A 6.5 Digit Low Noise Power Source
How to configure trigger output signals
This material shows how to configure the instrument to output the trigger signals at the specified timing using Agilent
B2961A/B2962A Power Source, through an example to source the trigger output signal before the transient device action.
Figure 1 shows the advanced trigger operation flowchart of the B2961A/B2962A Power Source. During the trigger
operation, the B2961A/B2962A Power Source can output trigger signals at the following six timings to synchronize the
other channels or instruments. Please note that the B2961A/B2962A Power Source has the independent trigger system
for both Transient (Source) and Acquire (Monitor) actions respectively and the capability to control both actions
individually.
*RST
ABORt
INITiate etc.
A
INITiated?
Idle
ARM
Layer
ARM
Events
TRIGger
Layer
No
Yes
ARM
Count=0
Event
Detector
Trigger
Output
Before
in ARM
Trigger
Output
After
in ARM
TRIGger
Count=0
B
Trigger
Output
Before
in TRIGger
Trigger
Output
After
in TRIGger
Yes
TRIGger
Events
Event
Detector
No
Last
Count?
Yes
No
Last
Count?
TRIGger
Delay
TRIGger
Count+1
ARM Delay
ARM
Count+1
Trigger
Output
Before
Device Action
A
B
Device
Action
Trigger
Output
After
Device Action
Figure 1. Advanced trigger operation flowchart of the B2961/B2962A Power Source
In order to output the trigger signals, it is required to specify the timing to output the trigger signals, the digital I/O pin
number and the digital I/O properties such as the polarity, the trigger type, etc. The level of the signal is fixed to 5 V. The
digital I/O properties define the actual waveform of the trigger output signals. Figure 2 shows the examples of the trigger
signal to be outputted by specifying the digital I/O properties.
Width
Example 1
Example
Type
Polarity
Position
1
Edge
Positive
Both
2
Edge
Positive
Before
Example 3
3
Level
Positive
N/A
Example 4
4
Edge
Negative
Both
5
Edge
Negative
After
Example 5
6
Level
Negative
N/A
5V
Example 2
Example 6
Action
(Arm, Trigger, Device)
Figure 2. Examples of the trigger signal to be outputted
Time
Agilent B2961A/B2962A 6.5 Digit Low Noise Power Source
In order to know how to set up the trigger output signals with front panel operation, the procedure is shown to configure
the instrument to output the trigger signal from the digital I/O pin 1 at the beginning of the transient device action.
(Please see Figure 3)
TRIGger Count=0
TRIGger
Events
No
Event
Detector
Width (200 us)
Yes
Last
Count?
EXT1
5V
Trigger Output
on Transient
TRIGger Delay
TRIGger Count+1
Device Action
TRIGger Layer
Transient
Device Action
Time
Figure 3. Example to output the trigger signal
Setting on the trigger output at the beginning of the transient device action
Step 1.
Step 2.
If you aren’t on the top of the Function menu, press
repeatedly to return to the top level.
On the top level of the Function menu
Any tree structure can’t be seen
In the middle level of the Function menu
Some tree structure can be seen
Press
and then press
to open Trigger Configuration dialogue.
(1) Press Trigger
(2) Press Config
Step 3.
Press
and select
(1) Press
to specify the device action, the trigger output of which is configured.
(2) Press ACTION
Agilent B2961A/B2962A 6.5 Digit Low Noise Power Source
Step 4.
Rotate
to select the action and press
to edit it. Then press
to specify the
transient action, the trigger output of which is configured.
(1) Rotate and press
Action
Step 5.
Rotate
to edit
to select the channel and press
(2) Press TRANS.
to edit it. Then press
to select the
channel 1 as the channel, the trigger output of which is configured.
(1) Rotate and press
Channel
Step 6.
Rotate
to edit
(2) Press Ch 1
to select the trigger output status and press
on.
(1) Rotate and press
Trigger Output State
to edit
(2) Press ON
to edit it. Then press
to set it
Agilent B2961A/B2962A 6.5 Digit Low Noise Power Source
Step 7.
Rotate
to select the trigger output pin and press
to edit it. Then press
the digital I/O pin 1 as the trigger output pin.
(1) Rotate and press
Trigger Output Pin
Step 8.
Press
to edit
(2) Press EXT1
to make the modification effective.
(1) Press OK to make the modification effective
to select
Agilent B2961A/B2962A 6.5 Digit Low Noise Power Source
Setting the digital I/O pin properties for trigger output
Step 9.
If you aren’t on the top of the Function menu, press
repeatedly to return to the top level.
On the top level of the Function menu
Any tree structure can’t be seen
In the middle level of the Function menu
Some tree structure can be seen
Step 10. Press
,
,
, and then press
to open DIO Configuration
dialogue.
(1) Press More…
(2) Press I/O
(3) Press DIO
(4) Press Config
Step 11. Press
and select
to specify the digital I/O pin number, the properties of which is
configured.
(1) Press
Step 12. Rotate
(2) Press Pin 1
to select the input/output function and press
to edit it. Then press
it to the trigger output.
(1) Rotate and press
to edit
the input/output function
(2) Press TRIGGER OUT
to set
Agilent B2961A/B2962A 6.5 Digit Low Noise Power Source
Step 13. Rotate
to select the polarity and press
to edit it. Then press
to set it to positive
polarity.
(1) Rotate and press
the polarity
Step 14. Rotate
to edit
(2) Press POS.
to select the output trigger type and press
to edit it. Then press
to set it
to the edge trigger.
(1) Rotate and press
the output trigger type
Step 15. Rotate
to edit
(2) Press EDGE
to select the output trigger timing and press
to edit it. Then press
enable the trigger output at the beginning of the specified action.
(1) Rotate and press
the output trigger timing
to edit
(2) Press BEFORE
to
Agilent B2961A/B2962A 6.5 Digit Low Noise Power Source
Step 16. Rotate
to select the pulse width of the output trigger and press
Then enter 200 us to set the pulse width of the output trigger to 200 us.
(1) Rotate and press
to edit
the pulse width of the output trigger
Step 17. Press
(2) Enter the pulse width
to make the modification effective.
(1) Press OK to make the modification effective
to edit it.