DSOX4PWR Power Measurement Application User`s Guide

DSOX4PWR Power
Measurement
Application
User's Guide
s1
Notices
© Agilent Technologies, Inc. 2007-2009,
2011-2012
No part of this manual may be reproduced in
any form or by any means (including electronic storage and retrieval or translation
into a foreign language) without prior agreement and written consent from Agilent
Technologies, Inc. as governed by United
States and international copyright laws.
Manual Part Number
Version 03.00.0000
Edition
October 13, 2012
Available in electronic format only
Agilent Technologies, Inc.
1900 Garden of the Gods Road
Colorado Springs, CO 80907 USA
Trademarks
Microsoft®, MS-DOS®, Windows®, Windows 2000®, and Windows XP® are U.S.
registered trademarks of Microsoft Corporation.
Adobe®, Acrobat®, and the Acrobat
Logo® are trademarks of Adobe Systems
Incorporated.
Warranty
The material contained in this document is provided “as is,” and is subject to being changed, without notice,
in future editions. Further, to the maximum extent permitted by applicable
law, Agilent disclaims all warranties,
either express or implied, with regard
to this manual and any information
contained herein, including but not
limited to the implied warranties of
merchantability and fitness for a particular purpose. Agilent shall not be
liable for errors or for incidental or
consequential damages in connection with the furnishing, use, or performance of this document or of any
information contained herein. Should
Agilent and the user have a separate
written agreement with warranty
terms covering the material in this
document that conflict with these
terms, the warranty terms in the separate agreement shall control.
Safety Notices
CAUTION
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like
that, if not correctly performed or
adhered to, could result in damage
to the product or loss of important
data. Do not proceed beyond a
CAUTION notice until the indicated
conditions are fully understood and
met.
Technology Licenses
The hardware and/or software described in
this document are furnished under a license
and may be used or copied only in accordance with the terms of such license.
Restricted Rights Legend
If software is for use in the performance of a
U.S. Government prime contract or subcontract, Software is delivered and licensed as
“Commercial computer software” as
defined in DFAR 252.227-7014 (June 1995),
or as a “commercial item” as defined in FAR
2.101(a) or as “Restricted computer software” as defined in FAR 52.227-19 (June
1987) or any equivalent agency regulation or
contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies’ standard commercial license
terms, and non-DOD Departments and
Agencies of the U.S. Government will
receive no greater than Restricted Rights as
2
defined in FAR 52.227-19(c)(1-2) (June
1987). U.S. Government users will receive
no greater than Limited Rights as defined in
FAR 52.227-14 (June 1987) or DFAR
252.227-7015 (b)(2) (November 1995), as
applicable in any technical data.
WA R N I N G
A WARNING notice denotes a
hazard. It calls attention to an
operating procedure, practice, or
the like that, if not correctly performed or adhered to, could result
in personal injury or death. Do not
proceed beyond a WARNING
notice until the indicated conditions are fully understood and
met.
DSOX4PWR Power Measurement Application User's Guide
Power Measurement Application—At a Glance
The DSOX4PWR Power Measurement and Analysis for the InfiniiVision
4000 X- Series oscilloscopes lets you quickly and easily analyze switching
power supply efficiency and reliability.
Input Line Analysis
AC
Input
Rectifier
Block
Output Analysis
Transient Response Analysis
Turn On Analysis
Power Device Analysis
Power Transfer
and Isolation
Output
DC
Output
PWM
Controller
Modulation Analysis
Regulation
and Control
Figure 1
Switch-Mode Power Supply (SMPS) Block Diagram and Types of
Measurements
With the Power Measurement Application, you can:
• Measure switching loss and conduction loss at the switching device (to
help improve efficiency).
• Analyze dI/dt and dV/dt slew rate (for reliable operation).
• Automate oscilloscope setup for ripple measurements (to eliminate
tedious manual oscilloscope set up).
• Perform pre- compliance testing to IEC 61000- 3- 2 standards (to reduce
compliance testing time).
• Analyze line power with total harmonic distortion, true power,
apparent power, power factor, and crest factor tests (to quickly provide
power quality information).
• Measure output noise (ripple).
• Analyze modulation using the on- time and off- time information of a
Pulse Width Modulation (PWM) signal (to help characterize the active
power factor).
DSOX4PWR Power Measurement Application User's Guide
3
• Measure how well a circuit rejects ripple coming from the input power
supply at various frequencies with the Power Supply Rejection Ratio
(PSRR) measurement.
The power measurement and analysis license, along with the oscilloscope,
high- voltage differential probe, current probe, probe deskew fixture, and
passive probe, form a complete power measurement system for power
supply design and testing.
Included with DSOX4PWR, at no additional charge, is a license for the
U1881A PC- based power analysis software package which provides
additional offline power measurements and report generation.
This guide describes:
• Chapter 1, “Prerequisites,” starting on page 9
• Chapter 2, “Getting Started,” starting on page 15
• Chapter 3, “Performing Power Analysis,” starting on page 25
• Chapter 4, “Automatic Power Measurements,” starting on page 63
4
DSOX4PWR Power Measurement Application User's Guide
Contents
Power Measurement Application—At a Glance
1
Prerequisites
Safety
9
Oscilloscope Requirements 9
Bandwidth Requirements 10
Memory Requirements 10
Software Version Requirements
11
Probe Requirements 11
Voltage Probe 11
Current Probe 12
De-Skewing the Voltage and Current Probes
2
3
12
Getting Started
Step 1: Access the Power Measurement Application
Step 2: Perform channel deskew
16
Step 3: Select the type of power analysis
20
Step 4: Make DUT connections and set up signals
20
Step 5: Change the analysis settings (if available)
22
Step 6: Apply the analysis
22
Step 7: View the analysis results
3
15
23
Performing Power Analysis
Input Measurements 25
Power Quality 25
DSOX4PWR Power Measurement Application User's Guide
5
Current Harmonics 29
Inrush Current 34
Switching Measurements
Switching Loss 36
Slew Rate 41
Modulation 44
36
Output Measurements 47
Output Ripple 47
Turn On/Turn Off 49
Transient Response 52
Power Supply Rejection Ratio (PSRR)
Efficiency 58
4
Automatic Power Measurements
Power Factor
Real Power
63
64
Apparent Power
64
Reactive Power
64
Crest Factor
65
Phase Angle
65
Output Ripple
Input Power
Efficiency
66
66
Peak Current
Transient
65
65
Output Power
6
55
67
67
Turn On Time
67
Turn Off Time
67
DSOX4PWR Power Measurement Application User's Guide
Power Loss
68
Power Loss/Cyc
Energy Loss
68
68
Index
DSOX4PWR Power Measurement Application User's Guide
7
8
DSOX4PWR Power Measurement Application User's Guide
DSOX4PWR Power Measurement Application
User's Guide
1
Prerequisites
Safety 9
Oscilloscope Requirements 9
Probe Requirements 11
This chapter describes safety considerations and the necessary
requirements for using the Power Measurement Application.
Safety
WA R N I N G
When connecting to a circuit with hazardous voltages, make sure the probes and
other components are used within their ratings. Refer to the probes' and other
components' documentation.
Oscilloscope Requirements
The DSOX4PWR Power Measurement Application works with the
4000 X- Series digital storage oscilloscopes (DSO).
• The InfiniiVision 4000 X- Series oscilloscopes come in 200 MHz,
350 MHz, 500 MHz, 1 GHz, and 1.5 GHz bandwidth models, with 2 Mpts
or 4 Mpts of memory.
Characteristics of the power supply under test determine the oscilloscope
bandwidth and memory required.
• “Bandwidth Requirements" on page 10
• “Memory Requirements" on page 10
s1
9
1
Prerequisites
• “Software Version Requirements" on page 11
Bandwidth Requirements
The bandwidth requirements of the oscilloscope and probe are driven by
the slew rate (rise/fall times) of the switching device.
For oscilloscopes with Gaussian response (typical for 1 GHz and lower
bandwidth oscilloscopes), the oscilloscope's rise time is commonly related
to the oscilloscope's bandwidth using the formula:
rise time = 0.35/bandwidth
To measure an input signal's rise time with ±5% error, the oscillosope's
rise time must be 1/3 of the input signal's rise time. Therefore, the
oscilloscope bandwidth required is:
BW = [ 0.35 / (input signal rise time / 3) ]
For example, a switching device whose rise time is 10 ns requires
oscilloscope (and probe) bandwidth of 105 MHz.
Memory Requirements
The memory requirements of the oscilloscope depend on the time range
and the types of signals to capture:
memory depth = time range * sample rate
• For switching device signals: If you need to capture the switching
signals for the duration of half the mains cycle (60 Hz), with a slew
rate of 50 ns (using a sampling rate that is four times the required
bandwidth), memory depth = 8.333 ms * 21 MHz * 4 = 699972 points.
With InfiniiVision 3000 X- Series oscilloscopes, the sampling rate is
determined by time range setting. In the above case, the sample rate in
High Resolution mode for the time range of 8.333 ms is 100 MSa/s;
therefore, the memory depth needed is 833300 points.
• For input AC line signals: You need to capture a few cycles in order to
plot the FFT. Resolution of the FFT plot = sampling rate / data size.
The expected harmonics are in multiples of 50/60 Hz.
10
DSOX4PWR Power Measurement Application User's Guide
1
Prerequisites
Because the input signals have low frequency components, a high
sampling rate is unnecessary. For example, the RTCA- DO- 160E
specification states that a sampling rate of 100 kSa/s and higher should
be sufficient. For a 60 Hz signal, to capture 10 cycles you need to
capture a duration of 83.33 ms.
The InfiniiVision 3000 X- Series oscilloscopes set the sampling rate to be
10 MSa/s for the above time range. The memory depth required is
83330 points with an FFT resolution of 4.77 Hz.
Software Version Requirements
Table 1
Oscilloscope Software Version Required
Oscilloscope Family
Software Version Required
InfiniiVision 4000 X-Series
3.00 or later
Probe Requirements
• “Voltage Probe" on page 11
• “Current Probe" on page 12
• “De- Skewing the Voltage and Current Probes" on page 12
Voltage Probe
You can use the following voltage probes:
• Agilent N2791A differential probe, 25 MHz, 700 V dynamic range.
• Agilent N2790A differential probe with AutoProbe interface, 100 MHz,
1.4 kV dynamic range.
• Agilent N2792A differential probe, 200 MHz bandwidth, 20 V dynamic
range.
• Agilent N2793A differential probe, 800 MHz bandwidth, 15 V dynamic
range.
• Agilent N2891A high- voltage differential probe, 70 MHz bandwidth, 7 kV
dynamic range.
DSOX4PWR Power Measurement Application User's Guide
11
1
Prerequisites
• Agilent 1141A differential probe, 200 MHz bandwidth, 400 V dynamic
range.
• Agilent 10070D passive probe 1:1, 20 MHz bandwidth, 400 V max. input
(for power supply noise measurement and for Power Supply Rejection
Ratio measurement).
• Agilent N2870A passive probe, 1:1, 35 MHz bandwidth, 55 V max input
(for power supply noise measurement and for Power Supply Rejection
Ratio measurement).
For voltage probe bandwidth requirements, see “Bandwidth
Requirements" on page 10.
The probe's voltage range required depends on the input signals to
measure. An AC- DC switch mode power supply needs a high voltage range
probe because the switching signals and input line signals can go up to
700 Vpp. For a DC- DC switch mode power supply, a smaller probe voltage
range is sufficient because the signal amplitudes are much smaller.
A passive probe is typically used to measure DC output and transient
response.
Current Probe
You can use the following Agilent AC/DC current probes:
• 1147B 50 MHz bandwidth, 15A peak.
• N2893A 100 MHz bandwidth, 30A peak.
• N2780A 2 MHz bandwidth, 500A peak.
• N2781A 10 MHz bandwidth, 150A peak.
• N2782A 50 MHz bandwidth, 30A peak.
• N2783A 100 MHz bandwidth, 30A peak.
For current probe bandwidth requirements, see “Bandwidth
Requirements" on page 10.
De-Skewing the Voltage and Current Probes
To ensure accurate power loss measurements, use the U1880A deskew
fixture to adjust the skew for any time delay differences between the
current probe and voltage probe signal paths.
12
DSOX4PWR Power Measurement Application User's Guide
Prerequisites
1
The procedure on de- skewing probes is described in Chapter 2, “Getting
Started,” starting on page 15.
Table 2
U1880A Deskew Fixture Environmental Characteristics
Temperature
Operating: –10 °C to +55 °C
Non-operating: –20 °C to +60 °C
Humidity
Operating: 95% RH at 40 °C for 24 hr
Non-operating: 90% RH at 65 °C for 24 hr
Altitude
Operating: to 4,570 m (15,000 ft)
Non-operating: to 15,244 m (50,000 ft)
Indoor use
Rated for indoor use only
DSOX4PWR Power Measurement Application User's Guide
13
1
14
Prerequisites
DSOX4PWR Power Measurement Application User's Guide
DSOX4PWR Power Measurement Application
User's Guide
2
Getting Started
Step 1: Access the Power Measurement Application 15
Step 2: Perform channel deskew 16
Step 3: Select the type of power analysis 20
Step 4: Make DUT connections and set up signals 20
Step 5: Change the analysis settings (if available) 22
Step 6: Apply the analysis 22
Step 7: View the analysis results 23
This chapter gives an overview of the steps you must take when first
performing power measurements.
Step 1: Access the Power Measurement Application
To access the power measurements application on the oscilloscope:
1 Press the [Analyze] key.
2 Press Features; then, select Power Application.
3 Press Features again to enable power measurements.
s1
15
2
Getting Started
Next
• "Step 2: Perform channel deskew" on page 16
Step 2: Perform channel deskew
To make accurate power loss measurements, you must perform current
and voltage channel deskew using the U1880A deskew fixture. The channel
deskew procedure calibrates the time delay between current and voltage
probes.
You need to perform the deskew procedure once initially, and you should
re- run the procedure when any part of the hardware setup changes (for
example, a different probe, different oscilloscope channel, etc.) or when
the ambient temperature changes.
To perform the channel deskew:
1 First, demagnetize and zero- adjust the current probe. Refer to the
current probe's documentation for instructions on how to do this.
16
DSOX4PWR Power Measurement Application User's Guide
Getting Started
2
2 Make connections to the U1880A deskew fixture:
Small Loop
Large Loop
For current probes:
•
•
•
•
1147B (50 MHz, 15A)
N2893A (10 MHz, 15A)
N2782A (50 MHz, 30A)
N2783A (100 MHz, 30A)
• N2780A (2 MHz, 500A)
• N2781A (10 MHz, 150A)
Connect high-voltage
differential probe to either:
• J5 (2.54 mm connector)
• J6 and J7 (alligator type)
• J2 (2.54 mm connector)
• J3 and J4 (alligator type)
a Connect D+ and D- of the high- voltage differential probe to the
deskew fixture.
b Connect the current probe to the current loop with the direction of
the arrow pointing towards the current flow.
c Make sure the switch on the deskew fixture is set to the appropriate
side of the fixture (either "small loop" or "large loop").
d Connect the deskew fixture to a USB port on your oscilloscope or a
PC using a USB cable. The USB port supplies power to the deskew
fixture.
3 In the Power Application main menu, press Signals.
DSOX4PWR Power Measurement Application User's Guide
17
2
Getting Started
4 In the Power Application Signals Menu, press the Voltage softkey and
use the Entry knob to select the analog channel probing the voltage
signal.
5 Press the Current softkey and use the Entry knob to select the analog
channel probing the current signal.
6 Press the Deskew softkey.
7 In the Power Application Deskew Menu, press Auto Deskew.
NOTE
Use the lowest attenuation setting on the high voltage differential probes whenever
possible because the voltage levels on the deskew fixture are very small. Using a higher
attenuation setting may yield inaccurate skew values (and affect the measurements made)
because the noise level is magnified as well.
When the deskew process completes, you see a message indicating
whether the deskew was successful, and if so, the settings being used.
18
DSOX4PWR Power Measurement Application User's Guide
Getting Started
2
8 Press the Back Back/Up key twice to return to the Power Application
main menu.
The deskew values are saved in the oscilloscope until a factory default or
secure erase is performed. The next time you run the Power Application,
you can use the saved deskew values or perform the deskew again.
Generally, you need to perform the deskew again when part of the test
setup changes (for example, a different probe, different oscilloscope
channel, etc.) or when the ambient temperature has changed.
See Also
Next
•
"U1880A Deskew Fixture User's Guide".
• "Step 3: Select the type of power analysis" on page 20
DSOX4PWR Power Measurement Application User's Guide
19
2
Getting Started
Step 3: Select the type of power analysis
1 In the Power Application main menu, press the Analysis softkey; then,
turn the Entry knob to select the type of power analysis.
The following types of power analysis are available:
• Current Harmonics.
• Efficiency.
• Inrush Current.
• Modulation.
• Power Quality.
• Switching Loss.
• Transient Response.
• Turn On/Turn Off.
• Output Ripple.
• Power Supply Rejection Ratio (PSRR).
• Slew Rate
Next
• "Step 4: Make DUT connections and set up signals" on page 20
Step 4: Make DUT connections and set up signals
For each type of power analysis, there is a Signals softkey and menu for
specifying the oscilloscope channels being used and setting other related
options.
1 In the Power Application main menu, press the Signals softkey.
2 In the Power Application Signals Menu, connect your probes to the
device under test and to the oscilloscope as shown in the connection
diagram.
20
DSOX4PWR Power Measurement Application User's Guide
2
Getting Started
3 In the above example, you would press the Voltage and Current softkeys
and make sure the proper analog channel is selected.
NOTE
Be sure to select the proper attenuation factor used for the voltage probe.
The attenuation factor multiplied by the probe's maximum output voltage gives the
maximum input signal. For example, the N2791A probe's maximum output voltage is ±7V,
so a 100:1 attenuation ratio gives a maximum input signal of ±700V.
NOTE
Also, be sure to select the proper attenuation factor used for the current probe.
4 If other softkeys are present for setting related options, like the Cycles
softkey in the above example, use them to specify the appropriate
settings.
DSOX4PWR Power Measurement Application User's Guide
21
2
Getting Started
5 If it is present, press the Auto Setup softkey to automatically scale and
position the voltage and current channels and perhaps set the time/div.
6 Press the
menu.
Next
Back
Back/Up key to return to the Power Application main
• "Step 5: Change the analysis settings (if available)" on page 22
Step 5: Change the analysis settings (if available)
If there are settings available for the type of power analysis chosen, there
will be a Settings softkey or other softkeys in the Power Application main
menu. To specify the power analysis settings:
1 In the Power Application main menu, press the Settings softkey or other
softkeys to make the appropriate settings for the type of analysis being
performed.
For example, the Current Harmonics settings menu looks like:
For descriptions of the settings available for each type of power
analysis, see Chapter 3, “Performing Power Analysis,” starting on page
25.
2 When you have fininshed changing the settings, return to the Power
Application main menu (perhaps by pressing the
necessary).
Next
Back
Back/Up key if
• "Step 6: Apply the analysis" on page 22
Step 6: Apply the analysis
Each type of power analysis provides an Apply softkey for starting the
analysis.
1 In the Power Application main menu, press Apply.
22
DSOX4PWR Power Measurement Application User's Guide
Getting Started
Next
2
• "Step 7: View the analysis results" on page 23
Step 7: View the analysis results
Once a power analysis has completed, you can view the results in the
following ways:
• By viewing the power analysis results on screen.
• By adding automatic power measurements.
Viewing Power
Analysis Results
on Screen
Adding Automatic
Power
Measurements
Power analysis results are displayed on the oscilloscope screen.
For example, here is a Current Harmonics analysis result:
Just like adding automatic measurements of voltage (peak- to- peak, max,
min, etc.) and time (frequency, period, rise time, fall time, etc.), you can
also add automatic power measurements. See Chapter 4, “Automatic Power
Measurements,” starting on page 63.
DSOX4PWR Power Measurement Application User's Guide
23
2
Getting Started
See Also
To learn more about the individual types of power analysis, their input
signals, their settings, and their results, see:
• Chapter 3, “Performing Power Analysis,” starting on page 25
24
DSOX4PWR Power Measurement Application User's Guide
DSOX4PWR Power Measurement Application
User's Guide
3
Performing Power Analysis
Input Measurements 25
Switching Measurements 36
Output Measurements 47
This chapter describes the types of power analysis you can perform with
the Power Measurement Application, the proper probing connections to
the device under test, signal setup, settings, and results.
Input Measurements
• "Power Quality" on page 25
• "Current Harmonics" on page 29
• "Inrush Current" on page 34
Power Quality
The Power Quality analysis shows the quality of the AC input line.
Some AC current may flow back into and back out of the load without
delivering energy. This current, called reactive or harmonic current, gives
rise to an "apparent" power which is larger than the actual power
consumed. Power quality is gauged by these measurements: power factor,
apparent power, true power, reactive power, crest factor, and phase angle
of the current and voltage of the AC line.
Signals Setup
1 With the Power Quality analysis selected in the Power Application main
menu, press the Signals softkey.
s1
25
3
Performing Power Analysis
2 Connect your probes to the device under test and to the oscilloscope as
shown in the connection diagram.
Figure 2
Typical Configuration for Input Line Analysis Tests
a Connect D+ of the voltage probe to the live wire of the AC input.
b Connect D- of the voltage probe to the neutral wire of the AC input.
c On the voltage probe, select the appropriate attenuation ratio.
d Connect the current probe to the live wire of the AC input with the
direction of the arrow pointing towards the current flow.
e Connect the voltage and current probes to the oscilloscope input
channels.
3 Press the Voltage and Current softkeys and make sure the proper analog
channel is selected.
4 Make sure the proper probe attenuation factors are set in the
oscilloscope for the voltage and current probes.
5 Press the Cycles softkey; then, turn the Entry knob to select the desired
number of cycles to capture in one acquisition.
6 Press the Auto Setup softkey to automatically set the vertical scale and
position of the voltage and current channels.
Also displayed it the power waveform which is the math multiply
operator of the voltage and current waveforms.
26
DSOX4PWR Power Measurement Application User's Guide
3
Performing Power Analysis
7 Press the
menu.
Settings
Back
Back/Up key to return to the Power Application main
1 In the Power Application main menu, press the Type: softkey; then, turn
the Entry knob to select the type of measurement to make in the power
quality analysis:
• Power Factor — Ratio of the actual power to the apparent power.
• Real (Actual) Power — The portion of power flow that, averaged over
a complete cycle of the AC waveform, results in net transfer of
energy in one direction.
• Apparent Power — The portion of power flow due to stored energy,
which returns to the source in each cycle.
• Reactive Power — The difference between apparent power and real
power due to reactance.
DSOX4PWR Power Measurement Application User's Guide
27
3
Performing Power Analysis
• Crest Factor — Crest factor is the ratio between the instantaneous
peak current/voltage required by the load and the RMS
current/voltage (RMS stands for Root Mean Square, which is a type
of average).
• Phase Angle — In the power triangle (the right triangle where
apparent_power2 = real_power2 + reactive_power2), phase angle is the
angle between the apparent power and the real power, indicating the
amount of reactive power.
Analysis Results
To perform the analysis, press Apply in the Power Application main menu.
When the analysis has completed, results are displayed.
The voltage and current waveforms are displayed as well as the input
power waveform (waveform math multiply of the voltage and current).
Also displayed are the power quality measurements you have selected and
applied:
• "Power Factor" on page 63
28
DSOX4PWR Power Measurement Application User's Guide
Performing Power Analysis
3
• "Real Power" on page 64
• "Apparent Power" on page 64
• "Reactive Power" on page 64
• "Crest Factor" on page 65
• "Phase Angle" on page 65
Power quality measurements are calculated using the captured voltage and
current waveforms over the number of cycles specified.
Automatic
Measurements
You can add these relevant automatic measurements using the [Meas] key
and menu.
Automatic Voltage measurements (refer to oscilloscope User's Guide for
more information):
• AC - RMS
Current Harmonics
Switching power supplies draw a range of harmonics from the AC mains.
Standard limits are set for these harmonics because these harmonics can
travel back to the supply grid and cause problems with other devices on
the grid.
Use the Current Harmonics analysis to test a switching power supply's
current harmonics to pre- compliance standard of IEC61000- 3- 2 (Class A,
B, C, or D). The analysis presents up to 40 harmonics.
Signals Setup
1 With the Current Harmonics analysis selected in the Power Application
main menu, press the Signals softkey.
2 Connect your probes to the device under test and to the oscilloscope as
shown in the connection diagram.
DSOX4PWR Power Measurement Application User's Guide
29
3
Performing Power Analysis
Figure 3
Typical Configuration for Input Line Analysis Tests
a Connect D+ of the voltage probe to the live wire of the AC input.
b Connect D- of the voltage probe to the neutral wire of the AC input.
c On the voltage probe, select the appropriate attenuation ratio.
d Connect the current probe to the live wire of the AC input with the
direction of the arrow pointing towards the current flow.
e Connect the voltage and current probes to the desired oscilloscope
channels.
3 Press the Voltage and Current softkeys and make sure the proper analog
channels are selected.
4 Make sure the proper probe attenuation factors are set in the
oscilloscope for the voltage and current probes.
5 Press the Cycles softkey; then, turn the Entry knob to select the desired
number of cycles to capture in one acquisition.
6 Press the Auto Setup softkey to automatically scale and position the
voltage and current channels and set the appropriate time/div.
Also set automatically is the Hanning FFT window (for better frequency
resolution and low spectral leakage). If you choose to set up signals
manually, you can select other FFT windows for analysis, such as the
30
DSOX4PWR Power Measurement Application User's Guide
Performing Power Analysis
3
Blackman- Harris window (for minimal spectral leakage) or the
Hamming window (for better frequency resolution and moderate
spectral leakage).
7 Press the
menu.
Settings
Back
Back/Up key to return to the Power Application main
1 In the Power Application main menu, press the Settings softkey.
2 In the Power Application Current Harmonics Menu, make the
appropriate settings.
Table 3
Current Harmonics Analysis Settings
Setting
Description
Line Frequency
Enter the line frequency.
Current Harmonics
Standard
Select the standard to perform compliance testing on the current
harmonics.
• IEC 61000-3-2 Class A — for balanced three-phase equipment,
household appliances (except equipment identified as Class D), tools
excluding portable tools, dimmers for incandescent lamps, and audio
equipment.
• IEC 61000-3-2 Class B — for portable tools.
• IEC 61000-3-2 Class C — for lighting equipment. Class C requires a
power factor calculation that happens when the Apply softkey (in the
Power Application main menu) is pressed. For this reason, you are
only allowed to select Class C when the Power Application is disabled
— it forces you to press Apply (again) to perform the analysis.
• IEC 61000-3-2 Class D — for equipment having a specified power
according less than or equal to 600 W, of the following types: personal
computers and personal computer monitors, television receivers.
Display
Choose how to display harmonics:
• Table.
• Bar Chart.
• Off — Harmonics measurement results are not displayed.
DSOX4PWR Power Measurement Application User's Guide
31
3
Performing Power Analysis
After the analysis has been performed, you can return to the settings
menu to change the display type and, if results are displayed in Bar
Chart or Table form, press the Scroll Harmonics softkey and use the
Entry knob to scroll current harmonics analysis results.
3 When you have fininshed changing the settings, press the
key to return to the Power Application main menu.
Analysis Results
Back
Back/Up
To perform the analysis, press Apply in the Power Application main menu.
When the analysis has completed, results are displayed.
Table 4
Current Harmonics Test Results
FFT waveform
32
Shows the frequency components in the input current. The FFT
is calculated using the Hanning window.
DSOX4PWR Power Measurement Application User's Guide
Performing Power Analysis
Table 4
3
Current Harmonics Test Results (continued)
Harmonic, Actual Value
(RMS), Limit (RMS), Margin,
Pass/Fail Status
For the first 40 harmonics, these values are displayed:
• Actual Value (RMS) — the measured value in the units
specified by the Harmonics Unit parameter.
• Limit (RMS) — the limit specified by the selected Current
Harmonics Standard parameter.
• Margin — the margin specified by the selected Current
Harmonics Standard parameter.
• Pass/Fail Status — whether the value passes or fails
according to the selected Current Harmonics Standard.
Rows in the table or bars in the chart are color-coded
according to pass/fail values.
Marginal results are greater than 85% of the limit but less
than 100% of the limit.
THD (Total Harmonic
Distortion)
Where:
• Xn = voltage or current of each harmonic
• X1 = fundamental voltage or current value
Saving
Harmonics Test
Results
To save current harmonics test results to a USB storage device
1 Press the [Save/Recall] key.
2 In the Save/Recall Menu, press the Save softkey.
3 In the Save Menu, press the Format softkey and turn the Entry knob to
select Power Harmonics data (*.csv) .
4 Press the softkey in the second position and use the Entry knob to
navigate to the save location. See the oscilloscope User's Guide for
information on navigating storage locations.
5 Finally, press the Press to Save softkey.
A message indicating whether the save was successful is displayed.
Automatic
Measurements
You can add these relevant automatic measurements using the [Meas] key
and menu.
Automatic Power App measurements:
• "Apparent Power" on page 64
DSOX4PWR Power Measurement Application User's Guide
33
3
Performing Power Analysis
• "Crest Factor" on page 65
Automatic Voltage measurements (refer to oscilloscope User's Guide for
more information):
• AC - RMS
Inrush Current
The Inrush current analysis measures the peak inrush current of the
power supply when the power supply is first turned on.
Signals Setup
1 With the Inrush analysis selected in the Power Application main menu,
press the Signals softkey.
2 Connect your probes to the device under test and to the oscilloscope as
shown in the connection diagram.
Figure 4
Typical Configuration for Inrush Current Analysis Tests
a Connect D+ of the voltage probe to the live wire of the AC input.
b Connect D- of the voltage probe to the neutral wire of the AC input.
c On the voltage probe, select the appropriate attenuation ratio.
d Connect the current probe to the live wire of the AC input with the
direction of the arrow pointing towards the current flow.
e Connect the voltage and current probes to the oscilloscope input
channels.
3 Press the Voltage and Current softkeys and make sure the proper analog
channels are selected.
34
DSOX4PWR Power Measurement Application User's Guide
Performing Power Analysis
3
4 Make sure the proper probe attenuation factors are set in the
oscilloscope for the voltage and current probes.
5 Press the Expected softkey; then, turn the Entry knob to specify the
expected inrush current amplitude. This sets the vertical scale of the
channel probing current.
6 Press the Max Vin softkey; then, turn the Entry knob to specify the
maximum input voltage. This sets the vertical scale of the channel
probing voltage.
7 Press the
menu.
Analysis Results
Back
Back/Up key to return to the Power Application main
To perform the analysis, press Apply in the Power Application main menu.
Follow the onscreen instructions. When the analysis has completed, results
are displayed.
DSOX4PWR Power Measurement Application User's Guide
35
3
Performing Power Analysis
The voltage and current waveforms are displayed. Also displayed is this
automatic power measurements:
• "Peak Current" on page 67
Switching Measurements
• "Switching Loss" on page 36
• "Slew Rate" on page 41
• "Modulation" on page 44
Switching Loss
The Switching Loss analysis calculates the power dissipated in the
switching cycles across the switching device. Typical power losses include:
• Switching losses that occur during switching of Vds and Id.
36
DSOX4PWR Power Measurement Application User's Guide
3
Performing Power Analysis
• Conduction losses that occur when the switching device (MOSFET) is
ON.
turn-OFF
switching
period
ON state
OFF state
turn-ON
switching
period
ON state
Id(t)
ON state
forward
voltage drop
current probe
null offset
switching
level
t
Vds(t)
0
P(t) = Vds(t) * Id(t)
t
0
conduction loss
Figure 5
switching loss
false losses due to
probe null offset
switching loss
conduction loss
Loss Occurrence in the Power Device
Design engineers use this information to improve the power conversion
efficiency of the power supply.
Switching loss is also used to quantify the power loss that is transferred
to the heat sink of the power device.
Signals Setup
1 With the Switching Loss analysis selected in the Power Application main
menu, press the Signals softkey.
2 Connect your probes to the device under test and to the oscilloscope as
shown in the connection diagram.
DSOX4PWR Power Measurement Application User's Guide
37
3
Performing Power Analysis
Figure 6
Typical Configuration for Power Device Analysis Tests
a Connect D+ of the voltage probe to the source of the MOSFET.
b Connect D- of the voltage probe to the drain of the MOSFET.
c On the voltage probe, select the appropriate attenuation ratio.
d Connect the current probe to the drain of the MOSFET with the
direction of the arrow pointing towards the current flow.
e Connect the voltage and current probes to the oscilloscope input
channels.
3 Press the Voltage and Current softkeys and make sure the proper analog
channel is selected.
4 Make sure the proper probe attenuation factors are set in the
oscilloscope for the voltage and current probes.
5 Press the Auto Setup softkey to automatically set the vertical scale and
position of the voltage and current channels.
6 Press the
menu.
Settings
Back
Back/Up key to return to the Power Application main
1 In the Power Application main menu, press the Settings softkey.
2 In the Power Application Switching Loss Menu, make the appropriate
settings.
38
DSOX4PWR Power Measurement Application User's Guide
Performing Power Analysis
Table 5
3
Switching Loss Analysis Settings
Setting
Description
V Ref
Enter the switching level for the switching edges. The value is in
percentage of the maximum switch voltage.
You can adjust this value to ignore noise floors.
This value specifies the threshold that is used to determine the
switching edges.
I Ref
Enter the switching level for the start of switching edges. The value is in
percentage of the maximum switch current.
You can adjust this value to ignore noise floors or null offset that is
difficult to eliminate in current probes.
This value specifies the threshold that is used to determine the
switching edges.
DSOX4PWR Power Measurement Application User's Guide
39
3
Performing Power Analysis
Table 5
Switching Loss Analysis Settings (continued)
Setting
Description
Conduction
Choose how to calculate conduction:
• Voltage waveform — The Power waveform uses the original data, and
the calculation is: P = V x I
• Rds(on) — The Power waveform includes error correction:
• In the On Zone (where the voltage level is below V Ref) – the
Power calculation is: P = Id2 x Rds(on)
Specify Rds(on) using the additional softkey.
• In the Off Zone (where the current level is below I Ref) – the Power
calculation is: P = 0 Watt.
• Vce(sat) — The Power waveform includes error correction:
• In the On Zone (where the voltage level is below V Ref) – the
Power calculation is: P = Vce(sat) x Ic
Specify Vce(sat) using the additional softkey.
• In the Off Zone (where the current level is below I Ref) – the Power
calculation is: P = 0 Watt.
3 When you have fininshed changing the settings, press the
key to return to the Power Application main menu.
Analysis Results
Back
Back/Up
To perform the analysis, press Apply in the Power Application main menu.
When the analysis has completed, results are displayed.
40
DSOX4PWR Power Measurement Application User's Guide
Performing Power Analysis
3
The voltage and current waveforms are displayed, as well as the power
waveform (waveform math multiply of the voltage and current). Also
displayed are these automatic power measurements and statistics:
• "Power Loss" on page 68
• "Power Loss/Cyc" on page 68
• "Energy Loss" on page 68
Automatic
Measurements
You can add these relevant automatic measurements using the [Meas] key
and menu.
Automatic Time measurements (refer to oscilloscope User's Guide for more
information):
• Frequency
Slew Rate
The Slew Rate analysis measures the rate of voltage or current change
during switching.
DSOX4PWR Power Measurement Application User's Guide
41
3
Performing Power Analysis
Signals Setup
1 With the Slew Rate analysis selected in the Power Application main
menu, press the Signals softkey.
2 Connect your probes to the device under test and to the oscilloscope as
shown in the connection diagram.
Figure 7
Typical Configuration for Power Device Analysis Tests
a Connect D+ of the voltage probe to the source of the MOSFET.
b Connect D- of the voltage probe to the drain of the MOSFET.
c On the voltage probe, select the appropriate attenuation ratio.
d Connect the current probe to the drain of the MOSFET with the
direction of the arrow pointing towards the current flow.
e Connect the voltage and current probes to the oscilloscope input
channels.
3 Press the Voltage and Current softkeys and make sure the proper analog
channel is selected.
4 Make sure the proper probe attenuation factors are set in the
oscilloscope for the voltage and current probes.
5 Press the Auto Setup softkey to automatically set the vertical scale and
position of the voltage and current channels.
6 Press the
menu.
42
Back
Back/Up key to return to the Power Application main
DSOX4PWR Power Measurement Application User's Guide
3
Performing Power Analysis
Settings
Analysis Results
1 In the Power Application main menu, press the Source softkey; then,
turn the Entry knob to select either dV/dt or dI/dt as the source for the
slew rate analysis.
To perform the analysis, press Apply in the Power Application main menu.
When the analysis has completed, results are displayed.
The voltage and current waveforms are displayed along with a
differentiate math function waveform that shows the slew rate.
Max and Min measurements on the differentiate math function waveform
are added and displayed.
DSOX4PWR Power Measurement Application User's Guide
43
3
Performing Power Analysis
Table 6
Slew Rate Test Results
dV/dt
[ y(n) - y(n-1) ] / [ x(n) - x(n-1) ], measures the slew rate of Vds of
the power device (MOSFET).
dI/dt
[ y(n) - y(n-1) ] / [ x(n) - x(n-1) ], measures the slew rate of Id of the
power device (MOSFET).
Modulation
The Modulation analysis measures the control pulse signal to a switching
device (MOSFET) and observes the trending of the pulse width, duty cycle,
period, frequency, etc. of the control pulse signal in response to different
events.
Signals Setup
1 With the Modulation analysis selected in the Power Application main
menu, press the Signals softkey.
2 Connect your probes to the device under test and to the oscilloscope as
shown in the connection diagram.
Figure 8
Continuous Mode Connection for Modulation Analysis Tests
a Connect D+ of the voltage probe to the gate of the MOSFET.
b Connect D- of the voltage probe to the source of the MOSFET.
c On the voltage probe, select the appropriate attenuation ratio.
d Connect the current probe to the drain of the MOSFET.
e Connect the voltage and current probes to oscilloscope input
channels.
44
DSOX4PWR Power Measurement Application User's Guide
3
Performing Power Analysis
3 Press the Voltage and Current softkeys and make sure the proper analog
channels are selected.
4 Make sure the proper probe attenuation factors are set in the
oscilloscope for the voltage and current probes.
5 Press the Duration softkey; then, turn the Entry knob to specify the time
to capture signals. This sets the time scale of the oscilloscope.
6 Press the Auto Setup softkey to automatically set the vertical scale and
position of the voltage and current channels.
7 Adjust the trigger level to capture waveforms at the same place in each
cycle (in other words, stabilize the waveform display).
8 Press the
menu.
Settings
Back
Back/Up key to return to the Power Application main
1 In the Power Application main menu, press the Source softkey; then,
turn the Entry knob to select either Voltage or Current as the source
for the modulation analysis.
2 Press the Type: softkey; then, turn the Entry knob to select the type of
measurement to make in the modulation analysis:
• Average
• RMS - AC
• Ratio
• Period
• Frequency
• +Width
• - Width
• Duty Cycle
• Rise Time
DSOX4PWR Power Measurement Application User's Guide
45
3
Performing Power Analysis
• Fall Time
Analysis Results
To perform the analysis, press Apply in the Power Application main menu.
When the analysis has completed, results are displayed.
The Measurement Trend math waveform vizualization is used to plot the
measurement variation for each cycle of the modulation waveforms.
Automatic
Measurements
You can add these relevant automatic measurements using the [Meas] key
and menu.
Automatic Voltage measurements (refer to oscilloscope User's Guide for
more information):
• Average
• AC - RMS
• Ratio
Automatic Time measurements (refer to oscilloscope User's Guide for more
information):
46
DSOX4PWR Power Measurement Application User's Guide
Performing Power Analysis
3
• Period
• Frequency
• +Width
• - Width
• Duty Cycle
• Rise Time
• Fall Time
Output Measurements
• "Output Ripple" on page 47
• "Turn On/Turn Off" on page 49
• "Transient Response" on page 52
• "Power Supply Rejection Ratio (PSRR)" on page 55
• "Efficiency" on page 58
Output Ripple
The Output Ripple analysis measures the ripple noise of the power supply
output.
Signals Setup
1 With the Output Ripple analysis selected in the Power Application main
menu, press the Signals softkey.
2 Connect your probes to the device under test and to the oscilloscope as
shown in the connection diagram.
DSOX4PWR Power Measurement Application User's Guide
47
3
Performing Power Analysis
Figure 9
Typical Configuration for Output Voltage Ripple Test
a Connect the voltage probe (passive or differential) to the DC output
of the power supply.
b Connect the voltage probe to an oscilloscope input channel.
3 Press the Voltage softkey and make sure the proper analog channel is
selected.
4 Make sure the proper probe attenuation factor is set in the oscilloscope
for the voltage probe.
5 Press the Duration softkey; then, turn the Entry knob to select the time
scale of the measurement.
6 Press the Auto Setup softkey to automatically set the vertical scale and
position of the voltage channel as well as the time scale.
7 Press the
menu.
Analysis Results
Back
Back/Up key to return to the Power Application main
To perform the analysis, press Apply in the Power Application main menu.
When the analysis has completed, results are displayed.
48
DSOX4PWR Power Measurement Application User's Guide
Performing Power Analysis
3
The output voltage waveforms is displayed along with this automatic
power measurement:
• "Output Ripple" on page 65
Turn On/Turn Off
The Turn On analysis determines how fast a turned on power supply takes
to reach 90% of its steady state output.
The Turn Off analysis determines how fast a turned off power supply
takes to reduce its output voltage to 10% of maximum.
Signals Setup
1 With the Turn On/Turn Off analysis selected in the Power Application main
menu, press the Signals softkey.
2 Connect your probes to the device under test and to the oscilloscope as
shown in the connection diagram.
DSOX4PWR Power Measurement Application User's Guide
49
3
Performing Power Analysis
Figure 10
Typical Configuration for Turn On/Off Analysis Tests
a Connect D+ of the input voltage probe to the live wire of the AC
input.
b Connect D- of the input voltage probe to the neutral wire of the AC
input.
c On the input voltage probe, select the appropriate attenuation ratio.
d Connect the output voltage probe (passive or differential) to the DC
output of the power supply.
e Connect the voltage probes to the oscilloscope input channels.
3 Press the Input V and Output V softkeys and make sure the proper analog
channels are selected.
4 Make sure the proper probe attenuation factors are set in the
oscilloscope for the voltage probes.
5 Press the Duration softkey; then, turn the Entry knob to select the time
scale of the measurement.
6 Press the Max Vin softkey; then, turn the Entry knob to specify the
maximum input voltage.
Enter the maximum (peak- to- peak) source voltage amplitude. The
source voltage will be used to trigger the oscilloscope in "Turn On
Time" test.
50
DSOX4PWR Power Measurement Application User's Guide
3
Performing Power Analysis
This value is used to adjust the vertical scale of the channel probing
the oscilloscope input voltage.
7 Press the Steady Vout softkey; then, turn the Entry knob to specify the
expected steady state output DC voltage of the power supply.
This value is used to adjust the vertical scale of the channel probing
the oscilloscope output voltage.
8 Press the
menu.
Settings
Back
Back/Up key to return to the Power Application main
1 In the Power Application main menu, press the Test softkey; then, turn
the Entry knob to select whether turn on or turn off analysis is
performed:
• Turn On — measures the time taken to get the output voltage of the
power supply after the input voltage is applied.
• Turn Off — measures the time taken for the output voltage of the
power supply to turn off after the input voltage is removed.
Analysis Results
To perform the analysis, press Apply in the Power Application main menu.
Follow the instructions on screen. When the analysis has completed,
results are displayed.
DSOX4PWR Power Measurement Application User's Guide
51
3
Performing Power Analysis
The input and output voltage waveforms are displayed. Also displayed is
this automatic power measurements:
• "Turn On Time" on page 67
• "Turn Off Time" on page 67
Transient Response
The Transient Response analysis determines how fast a power supply's
output voltage responds to change at the output load. This time is from
when the output voltage first exits the settling band to when it last enters
the settling band.
Signals Setup
1 With the Transient Response analysis selected in the Power Application
main menu, press the Signals softkey.
2 Connect your probes to the device under test and to the oscilloscope as
shown in the connection diagram.
52
DSOX4PWR Power Measurement Application User's Guide
3
Performing Power Analysis
Figure 11
Typical Configuration for Power Output Transient Response
a Connect the voltage probe (passive or differential) to the DC output
of the power supply.
b Connect the voltage probe to an oscilloscope input channel.
c Connect the current probe to the output load of the power supply.
The change in the load current will be used to trigger the
oscilloscope to capture the transients.
d Connect the current probe to an oscilloscope input channel.
3 Press the Voltage and Current softkeys and make sure the proper analog
channel is selected.
4 Make sure the proper probe attenuation factors are set in the
oscilloscope for the voltage and current probes.
5 Press the Duration softkey; then, turn the Entry knob to select the time
scale of the measurement.
6 Press the Overshoot softkey; then, turn the Entry knob to specify the %
of overshoot of the output voltage.
This value will be used to determine the settling band value for the
transient response and to adjust the vertical scale of the oscilloscope.
7 Press the Steady Vout softkey; then, turn the Entry knob to specify the
expected steady state output DC voltage of the power supply.
DSOX4PWR Power Measurement Application User's Guide
53
3
Performing Power Analysis
This value is used along with the overshoot percentage to specify the
settling band for the transient response and to adjust the vertical scale
of the oscilloscope.
8 Press the
menu.
Settings
Back
Back/Up key to return to the Power Application main
1 In the Power Application main menu, press the Settings softkey.
2 In the Power Application Transient Menu, make the appropriate
settings.
Table 7
Transient Response Analysis Settings
Setting
Description
Initial I
Enter the initial load current value. The initial load current will be used as
a reference and to trigger the oscilloscope.
New I
Enter the new load current value. The new load current will be used as a
reference and to trigger the oscilloscope.
3 When you have fininshed changing the settings, press the
key to return to the Power Application main menu.
Analysis Results
Back
Back/Up
To perform the analysis, press Apply in the Power Application main menu.
Follow the instructions on screen. When the analysis has completed,
results are displayed.
54
DSOX4PWR Power Measurement Application User's Guide
Performing Power Analysis
3
The voltage and current waveforms are displayed. There are start and end
time stamps that mark the measured area. Also displayed is this automatic
power measurements:
• "Transient" on page 67
Power Supply Rejection Ratio (PSRR)
The Power Supply Rejection Ratio (PSRR) test is used to determine how
well a voltage regulator rejects ripple noise over different frequency range.
This analysis provides a signal from the oscilloscope's waveform generator
that sweeps its frequency. This signal is used to inject ripple to the DC
voltage that feeds the voltage regulator.
The AC RMS ratio of the input over the output is measured and is plotted
over the range of frequencies.
DSOX4PWR Power Measurement Application User's Guide
55
3
Performing Power Analysis
There are many different ways to measure PSRR. Because the oscilloscope
has a higher noise floor and lower sensitivity than a network analyzer, it
is difficult to measure PSRR any better than - 60 dB. The PSRR test using
the oscilloscope is usually acceptable for spot- checking overall PSRR
behavior of a power supply under test.
Signals Setup
1 With the Power Supply Rejection Ratio (PSRR) analysis selected in the Power
Application main menu, press the Signals softkey.
2 Connect your probes to the device under test and to the oscilloscope as
shown in the connection diagram.
Figure 12
Typical Configuration for Power Supply Rejection Ratio Analysis
a Connect one voltage probe (passive or differential) to the input of the
low- dropout (LDO) regulator (and ground).
b Connect a second voltage probe (passive or differential) to the output
of the low- dropout (LDO) regulator (and ground).
c Connect the waveform generator output to the modulated power
supply.
One example of a modulated power supply is the "TS200" Option 1A
from Accel Instruments. You can also use an injection transformer to
inject the oscilloscope's WaveGen signal to a power supply output that
is connected to the low- dropout (LDO) regulator. In this case, the
injection transformer and power supply combination replaces a
modulated power supply (like TS200).
3 Press the Input V and Output V softkeys and make sure the proper analog
channels are selected.
56
DSOX4PWR Power Measurement Application User's Guide
Performing Power Analysis
3
4 Make sure the proper probe attenuation factors are set in the
oscilloscope for the voltage probes.
5 Press the
menu.
Settings
Back
Back/Up key to return to the Power Application main
1 In the Power Application main menu, press the Settings softkey.
2 In the Power Application PSRR Menu, make the appropriate settings.
Table 8
Switching Loss Analysis Settings
Setting
Description
Min Freq
Sets the start sweep frequency value. The measurement is displayed on
a log scale, so you can select from decade values.
Max Freq
Sets the end sweep frequency value. The measurement is displayed on a
log scale, so you can select from decade values in addition to the
maximum frequency of 20 MHz.
Max Ratio
Specifies the vertical scale of the PSRR math waveform.
Amplitude
Sets the amplitude value for the waveform generator.
Output Load
Sets the waveform generator expected output load impedance.
3 When you have fininshed changing the settings, press the
key to return to the Power Application main menu.
Analysis Results
Back
Back/Up
To perform the analysis, press Apply in the Power Application main menu.
When the analysis has completed, results are displayed.
DSOX4PWR Power Measurement Application User's Guide
57
3
Performing Power Analysis
The input and output voltage waveforms and the PSRR math waveform are
displayed. Tracking X and Y cursors are also displyed to show the ratio
dB values at various frequencies.
Efficiency
Efficiency analysis tests the overall efficiency of the power supply by
measuring the output power over the input power. This analysis requires a
4- channel oscilloscope because input voltage, input current, output voltage,
and output current are measured.
Signals Setup
1 With the Efficiency analysis selected in the Power Application main
menu, press the Signals softkey.
2 Connect your probes to the device under test and to the oscilloscope as
shown in the connection diagram.
58
DSOX4PWR Power Measurement Application User's Guide
Performing Power Analysis
Figure 13
3
Typical Configuration for Efficiency Analysis Tests
a Connect D+ of the input voltage probe to the live wire of the AC
input.
b Connect D- of the input voltage probe to the neutral of the AC input.
c On the input voltage probe, select the appropriate attenuation ratio.
d Connect the input current probe to the live wire of the AC input
with the direction of the arrow pointing towards the current flow.
e Connect D+ of the output voltage probe to the input path of the load.
f Connect D- of the output voltage probe to the return path of the
load.
g On the output voltage probe, select the appropriate attenuation ratio.
h Connect the output current probe to the input path of the load with
the direction of the arrow pointing towards the current flow.
i Connect the voltage and current probes to the oscilloscope input
channels.
3 Press the Input V, Input I, Output V, and Output I softkeys and make sure the
proper analog channels are selected.
4 Make sure the proper probe attenuation factors are set in the
oscilloscope for the voltage and current probes.
5 Press the Duration softkey; then, turn the Entry knob to specify the time
to capture signals. This sets the time scale of the oscilloscope.
DSOX4PWR Power Measurement Application User's Guide
59
3
Performing Power Analysis
6 Press the Auto Setup softkey to automatically set the vertical scale and
position of the voltage and current channels.
7 Press the
menu.
Analysis Results
Back
Back/Up key to return to the Power Application main
To perform the analysis, press Apply in the Power Application main menu.
When the analysis has completed, results are displayed.
The input voltage, input current, output voltage, and output current
waveforms are displayed, as well as the input power waveform (waveform
math multiply of the input voltage and current). Also displayed are these
automatic power measurements and statistics:
• "Input Power" on page 65
• "Output Power" on page 66
• "Efficiency" on page 66
60
DSOX4PWR Power Measurement Application User's Guide
Performing Power Analysis
Automatic
Measurements
3
You can add these relevant automatic measurements using the [Meas] key
and menu.
Automatic Power App measurements:
• "Real Power" on page 64
• "Apparent Power" on page 64
• "Reactive Power" on page 64
• "Power Factor" on page 63
• "Phase Angle" on page 65
Automatic Voltage measurements (refer to oscilloscope User's Guide for
more information):
• AC - RMS
• DC - RMS
• Maximum
• Minimum
• Peak- Peak
Automatic Time measurements (refer to oscilloscope User's Guide for more
information):
• Frequency
• Phase
DSOX4PWR Power Measurement Application User's Guide
61
3
62
Performing Power Analysis
DSOX4PWR Power Measurement Application User's Guide
DSOX4PWR Power Measurement Application
User's Guide
4
Automatic Power Measurements
Power Factor 63
Real Power 64
Apparent Power 64
Reactive Power 64
Crest Factor 65
Phase Angle 65
Output Ripple 65
Input Power 65
Output Power 66
Efficiency 66
Peak Current 67
Transient 67
Turn On Time 67
Turn Off Time 67
Power Loss 68
Power Loss/Cyc 68
Energy Loss 68
Power Factor
Ratio of the actual AC line power to the apparent power.
Real Power / Apparent Power
The power factor measurement is made using two source inputs, the
voltage waveform and the current waveform, and it also requires a math
multiply waveform of the voltage and current waveforms.
s1
63
4
Automatic Power Measurements
Real Power
The portion of power flow that, averaged over a complete cycle of the AC
waveform, results in net transfer of energy in one direction.
The real power measurement is made on one source input that represents
power. This is typically a math multiply waveform of the voltage and
current waveforms.
Apparent Power
The portion of AC line power flow due to stored energy, which returns to
the source in each cycle.
IRMS * VRMS
The apparent power measurement is made using two source inputs, the
voltage waveform and the current waveform.
Reactive Power
The difference between apparent power and real power due to reactance.
Using the power triangle (the right triangle where apparent_power2 =
real_power2 + reactive_power2):
Measured in VAR (Volts- Amps- Reactive)
64
DSOX4PWR Power Measurement Application User's Guide
4
Automatic Power Measurements
The reactive power measurement is made using two source inputs, the
voltage waveform and the current waveform, and it also requires a math
multiply waveform of the voltage and current waveforms.
Crest Factor
Crest factor is the ratio between the instantaneous peak AC line
current/voltage required by the load and the RMS current/voltage.
Select the voltage source for V crest factor: Vpeak / VRMS
Select the current source for I crest factor: Ipeak / IRMS
Phase Angle
In the power triangle (the right triangle where apparent_power2 =
real_power2 + reactive_power2), phase angle is the angle between the
apparent power and the real power, indicating the amount of reactive
power. Small phase angles equate to less reactive power.
The phase angle measurement is made using two source inputs, the
voltage waveform and the current waveform, and it also requires a math
multiply waveform of the voltage and current waveforms.
Output Ripple
VMax - VMin
The output ripple measurement is made on one source input that is the
output voltage waveform.
Input Power
Input V * Input I
DSOX4PWR Power Measurement Application User's Guide
65
4
Automatic Power Measurements
The input power measurement is made using two source inputs, the input
voltage waveform and the input current waveform, and it also requires a
math multiply waveform of the voltage and current waveforms.
The input power measurement requires that you specify the channels
probing the input voltage, input current, output voltage, and output
current in the Power Application Signals Menu and that you perform the
automated signals setup by pressing the Auto Setup softkey in the same
menu.
Output Power
Output V * Output I
The output power measurement is made using two source inputs, the
output voltage waveform and the output current waveform, and it also
requires a math multiply waveform of the voltage and current waveforms.
The output power measurement requires that you specify the channels
probing the input voltage, input current, output voltage, and output
current in the Power Application Signals Menu and that you perform the
automated signals setup by pressing the Auto Setup softkey in the same
menu.
Efficiency
Output power / input power.
The efficiency measurement is made on one source input that represents
input power. This is typically a math multiply waveform of the input
voltage and input current waveforms. This measurement also requires the
output voltage waveform and the output current waveform specified in the
signals setup for the Efficiency power analysis.
The efficiency measurement requires that you specify the channels probing
the input voltage, input current, output voltage, and output current in the
Power Application Signals Menu and that you perform the automated
signals setup by pressing the Auto Setup softkey in the same menu.
66
DSOX4PWR Power Measurement Application User's Guide
4
Automatic Power Measurements
Peak Current
The Peak Current can be a positive or negative value, so the result is the
larger of the measured maximum or minimum.
The peak current measurement is made on one source input that is the
current waveform.
Transient
Transient response time = t2 – t1, where:
• t1 = The first time a voltage waveform exits the settling band.
• t2 = The last time it enters into the settling band.
• Settling band = +/- overshoot % of the steady state output voltage.
The transient measurement is made using X cursors on the output voltage
signal.
Turn On Time
Turn On time = t2 - t1, where:
• t1 = AC input voltage rises to 10% of its maximum amplitude (Start
Time).
• t2 = DC output voltage rises to 90% of its maximum amplitude (End
Time).
The turn on time measurement is made using X cursors on two source
inputs, the input voltage waveform and the output voltage waveform.
Turn Off Time
Turn Off time = t2 - t1, where:
• t1 = AC input voltage goes below 10% of its positive peak (or negative
peak which ever occurs first) (Start Time).
DSOX4PWR Power Measurement Application User's Guide
67
4
Automatic Power Measurements
• t2 = DC output voltage drops to 10% of its steady state value (End
Time).
The turn on time measurement is made using X cursors on two source
inputs, the input voltage waveform and the output voltage waveform.
Power Loss
Pn = Vdsn * Idn, where n is each sample.
The power loss measurement is made on one source input that represents
power. This is typically a math multiply waveform of the voltage and
current waveforms.
Power Loss/Cyc
Pn = (Vdsn * Idn) * (Time range of zoom window) * (Counter measurement
of the voltage of the switching signal), where n is each sample.
The power loss per cycle measurement is made on one source input that
represents power. This is typically a math multiply waveform of the
voltage and current waveforms.
This measurement operates when in zoom mode and the counter
measurement is installed on the voltage of the switching signal.
Energy Loss
=  (Vdsn * Idn) * sample size, where n is each sample.
The energy loss measurement is made on one source input that represents
power. This is typically a math multiply waveform of the voltage and
current waveforms.
68
DSOX4PWR Power Measurement Application User's Guide
Index
Symbols
% overshoot, 53
current probe attenuation factor
parameter, 21
current switching level, 39
Hanning window, 30
harmonic, 33
high-voltage differential probe, 11
Numerics
D
I
10070D passive probe, 12
1141A differential probe, 12
1147B current probe, 12
4000 X-Series oscilloscopes, 11
deskew (channel), performing, 16
deskew fixture (U1880A), 12, 17
device under test, connection to, 20
dI/dt, 3, 44
dI/dt test results, 44
differential probe, high-voltage, 11
duration for turn on/off analysis, 50
duration of Efficiency analysis, 59
duration of load change, 53
duration of Modulation analysis, 45
duration of output ripple analysis, 48
Duty Cycle, power modulation analysis, 45
dV/dt, 3, 44
dV/dt test results, 44
I crest factor, 65
IEC 61000-3-2 standard, 3, 31
initial load current, 54
input power, 65
inrush current, 34
inrush current analysis, 34
inrush current analysis signals, 34
inrush current test results, 35
A
accessing the Power Measurement
Application, 15
apparent power, 3, 27, 64
at a glance, 3
Average, power modulation analysis, 45
B
bandwidth (oscilloscope) requirements, 10
Blackman-Harris window, 31
block diagram, switch-mode power
supply, 3
C
channel deskew, performing, 16
conduction, 40
conduction loss, 3, 37
configuration (test) parameters, 22
connection to device under test, 20
crest factor, 3, 28, 65
current harmonics, 29
current harmonics analysis settings, 31
current harmonics analysis signals, 29
current harmonics plot parameter, 31
current harmonics standard parameter, 31
current harmonics test results, 32
current probe, 12
E
L
line frequency parameter, 31
load transient response test results, 54
M
efficiency analysis signals, 58
efficiency, power app automatic
measurement, 66
energy loss, 68
expected inrush current, 35
maximum source voltage, 50
memory (oscilloscope) requirements, 10
modulation, 44
modulation analysis, 3
modulation analysis settings, 45
modulation analysis signals, 44
F
N
Fall Time, power modulation analysis, 46
FFT waveform (plot), 32
Frequency, power modulation analysis, 45
N2780A current probe, 12
N2781A current probe, 12
N2782A current probe, 12
N2783A current probe, 12
N2790A high-voltage differential
probe, 11
N2791A high-voltage differential
probe, 11
N2792A differential probe, 11
N2793A differential probe, 11
N2870A passive probe, 12
G
getting started, 15
H
Hamming window, 31
DSOX4PWR Power Measurement Application User's Guide
69
Index
N2891A high-voltage differential
probe, 11
N2893A current probe, 12
negative pulse width, power modulation
analysis, 45
new load current parameter, 54
noise floor, 39
notices, 2
null offset, 39
number of cycles, 26, 30
prerequisites, 9
probe (oscilloscope) requirements, 11
probe (oscilloscope), current, 12
probe (oscilloscope), high-voltage
differential, 11
probe (oscilloscope), passive, 12
PSRR (Power Supply Rejection Ratio), 4
PWM (pulse width modulation), 3
O
Ratio, power modulation analysis, 45
Rds (dynamic ON resistance)
waveform, 40
reactive power, 27, 64
real (actual) power, 27, 64
requirements, current probe, 12
requirements, deskew fixture, 12
requirements, high-voltage differential
probe, 11
requirements, oscilloscope, 9
requirements, oscilloscope bandwidth, 10
requirements, oscilloscope memory, 10
requirements, oscilloscope probe, 11
requirements, oscilloscope software
version, 11
requirements, passive probe, 12
results (test), viewing, 23
ripple measurements, 3
Rise Time, power modulation analysis, 45
RMS - AC, power modulation analysis, 45
running tests, 22
oscilloscope bandwidth requirements, 10
oscilloscope memory requirements, 10
oscilloscope probe requirements, 11
oscilloscope requirements, 9
oscilloscope software version
requirements, 11
output noise, 3
output power, 66
output ripple, 65
output ripple analysis signals, 47
output voltage ripple, 47
output voltage ripple test results, 48
overall efficiency, 58
overall efficiency test results, 60
overall system efficiency, 66
overview, 3
P
parameters, test configuration, 22
passive probe, 12
peak current, 67
percent overshoot, 53
Period, power modulation analysis, 45
phase angle, 28, 65
positive pulse width, power modulation
analysis, 45
power factor, 3, 27, 63
power loss, 68
power loss per cycle, 68
power measurements, more about, 25
power quality, 3, 25
power quality analysis settings, 27
power quality analysis signals, 25
power quality test results, 28
pre-compliance testing, 3
70
R
switching loss analysis signals, 37
switching loss test results, 40
switch-mode power supply block
diagram, 3
T
test configuration, 22
test results, viewing, 23
test selection, 20
tests, running, 22
THD (total harmonic distortion), 33
total harmonic distortion, 3
trademarks, 2
transient response, 52
transient response analysis, 52
transient response analysis settings, 54
transient response analysis signals, 52
transient response time, 67
true power, 3
turn off time, 51, 67
turn off time test results, 51
turn on time, 51, 67
turn on time test results, 51
turn on/off analysis, 49
turn on/off analysis settings, 51
turn on/off analysis signals, 49
U
U1880A deskew fixture, 12, 17
V
S
safety, 9
selecting tests, 20
slew rate analysis settings, 43
slew rate analysis signals, 42
slew rate of Id in power device, 44
slew rate of Vds in power device, 44
slew rate power analysis, 41
software (oscilloscope) version
requirements, 11
software version (oscilloscope)
requirements, 11
steady state DC output voltage, 51, 53
switching loss, 3, 36
switching loss analysis settings, 38
V crest factor, 65
viewing test results, 23
viewing waveforms, 23
voltage (high) differential probe, 11
voltage probe attenuation factor
parameter, 21
voltage switching level, 39
W
waveforms, viewing, 23
window parameter (FFT analysis), 30
DSOX4PWR Power Measurement Application User's Guide