AGILENT N2780A

N2780A Series AC/DC Current Probes
A wide selection of current probes to meet
your application's needs
Data Sheet
Hybrid technology for AC and DC measurements
Using hybrid technology that includes a Hall-effect sensor and an AC current
transformer, the probes provide accurate measurement of DC or AC currents up to
500 Arms (for model N2780A) or DC–100 MHz (for model N2783A), without breaking into the circuit. Using split core construction, the probe easily clips on and off
of a conductor.
Wide range of applications
Figure 1 N2780A Series current probes
with N2779A power supply
• Various bandwidths: DC to 2 MHz,
10 MHz, 50 MHz, 100 MHz
• DC and AC measurements
• Superior 1% accuracy and high
signal-to-noise ratio
• Overload-protect function prevents
probe damage from excessive input
• Direct connection to high-impedance
1 MΩ BNC input of oscilloscope
• “Demagnetize” button to remove any
residual magnetism that builds up in
the magnetic core
• External power supply (N2779A) lets
you connect up to three N278xA
current probes to a single power
supply
Compatible with any oscilloscope
with a high-impedance BNC input, the
new N2780A Series current probes
offer accurate and reliable solution for
measuring DC and AC currents.
The current probes feature broad measurement ranges (up to 500 A), flat frequency response, low noise and low insertion loss that make the probes ideal for current measurements in areas such as measuring steady state or transient current
of motor drives, switching power supplies, inverters, controllers, sensors, disk
drives, LCD displays, electronic ballasts and amplifiers. The high signal-to-noise
ratio of the N2782A and N2783A makes them ideal for making low-level current
measurements in milliampere ranges.
Accurate current measurement
A built-in DEMAG (demagnetize) function allows the removal of any residual
magnetism that has built up in the magnetic core due to power on/off switching
or excessive input current. In addition, voltage offset or temperature drift on the
probe can be easily corrected by using the zero adjustment control.
Figure 2 N2783A, N2780A, N2781A and N2782A current probe (from left to right)
-10
Maximum input current (Arms)
35
Amplitude (dB)
-20
-30
-40
-50
-60
30
25
20
15
10
5
1
10k
100
1M
10
100M
100
1k
10k
Figure 3 Frequency response of N2783A
1M
10M
12
9
Deviation (%)
1
100m
10m
6
3
0
-3
-6
0.001
1m
0.01
0.1
1
10
Input Current (A)
100
1k
10k
100k
1M
10M
100M
AC 50 Hz
Frequency (Hz)
Figure 5 Insertion impedance of N2783A
DC
Figure 6. Amplitude accuracy characteristics of N2783A
Note: For the characteristic plots of other current probe models, refer to the N2780A Series current probe user’s manual.
N2779A 3-channel power supply specifications
Applicable current probes:
Number of power supply connectors:
Connector type:
Output voltage:
Maximum rated power:
Input power requirement:
2
100M
Figure 4 Continuous maximum input rating of N2783A
10
Input Impedance (ohm)
100k
Frequency (Hz)
Frequency (Hz)
N2780A, N2781A, N2782A, N2783A and N2774A
3
LEMO inc./FFA..0S.304.CNAC42Z
DC (12 V, 2.5 A)
170 VA
AC 100-240 VAC, 50/60 Hz, 125-170 VAC
100
Model number
N2780A
N2781A
N2782A
N2783A
Bandwidth (-3dB)
Risetime
Maximum current (continuous) RMS
Maximum peak current (non-continuous)
Lowest measurable current
(at ±3% accuracy at DC, scope
set to 1 mV/div and high-resolution
mode on)
Output voltage rate
Amplitude accuracy*
(DC and 45 to 66 Hz, rated current)
Noise (measured with 20 MHz
bandwidth limit filter on the scope)
Temperature coefficient for
sensitivity (within a range of 0°C
to 40 °C or 32 °F to 104 °F)
Effect of external magnetic fields
(in a DC to 60 Hz, 400 A/m
magnetic field)
Maximum rated power
DC to 2 MHz
175 ns or less
500 A
700 A peak
20 mA
DC to 10 MHz
35 ns or less
150 A
300 A peak;
20 mA
DC to 50 MHz
7 ns or less
30 A
50 A peak
5 mA
DC to 100 MHz
3.5 ns or less
30 A
50 A peak
5 mA
0.01 V/A (100:1)
±1.0% rdg.
± 500 mA
Equivalent to
25 mArms or less
±2% or less
0.01 V/A (100:1)
±1.0% rdg.
± 100 mA
Equivalent to
25 mArms or less
±2% or less
0.1 V/A (10:1)
±1.0% rdg.
± 10 mA
Equivalent to
2.5 mArms or less
±2% or less
0.1 V/A (10:1)
±1.0% rdg.
± 10 mA
Equivalent to
2.5 mArms or less
±2% or less
Equivalent to a
maximum of 800 mA
Equivalent to a
maximum of 150 mA
Equivalent to a
maximum of 20 mA
Equivalent to a
maximum of 5 mA
7.2 VA (with rated
current)
DC ±12 V ±0.5 V
20 mm dia. (0.79” dia.)
Sensor cable:
Approx. 2 m (78.7”)
Power supply cable:
Approx. 1 m (39.4”)
5.5 VA (with rated
current)
DC ±12 V ±1 V
20 mm dia. (0.79” dia.)
Sensor cable:
Approx. 2 m (78.7”)
Power supply cable:
Approx. 1 m (39.4”)
5.6 VA (with rated
current)
DC ±12 V ±0.5 V
5 mm dia. (0.2” dia.)
Sensor cable:
Approx. 1.5 m (78.7”)
Power supply cable:
Approx. 1 m (39.4”)
5.3 VA (with rated
current)
DC ±12 V ±0.5 V
5 mm dia. (0.2” dia.)
Sensor cable:
Approx. 1.5 m (78.7”)
Power supply cable:
Approx. 1 m (39.4”)
Rated supply voltage
Diameter of measurable conductors
Cable lengths
Note*: The amplitude accuracy specifications are guaranteed at 23°C ± 3°C (or 73°F ± 5°F)
Compatible Oscilloscopes
Ordering Information
Any oscilloscope offering 1 MΩ BNC
input including Agilent 1000 Series,
InfiniiVision 6000 and 7000 Series, and
Infiniium 8000 and 9000 Series. You
must select the input impedance of
the oscilloscope to be 1 MΩ in order to
make accurate measurements. If the
oscilloscope you are using does not
have a 1 MΩ input impedance setting,
you can purchase the Agilent E2697A
50Ω to 1 MΩ adapter.
N2780A
2 MHz/500A AC/DC current probe
N2781A
10 MHz/150A AC/DC current probe
N2782A
50 MHz/30A AC/DC current probe
N2783A
100 MHz/30A AC/DC current probe
N2779A
3-channel power supply for
N2780A Series current probes
3
Remove all doubt
Agilent Email Updates
www.agilent.com/find/emailupdates
Get the latest information on the
products and applications you select.
Agilent Direct
www.agilent.com/find/agilentdirect
Quickly choose and use your test
equipment solutions with confidence.
www.lxistandard.org
LXI is the LAN-based successor to
GPIB, providing faster, more efficient
connectivity. Agilent is a founding
member of the LXI consortium.
Our repair and calibration services
will get your equipment back to you,
performing like new, when promised. You will get full value out of
your Agilent equipment throughout its lifetime. Your equipment
will be serviced by Agilent-trained
technicians using the latest factory
calibration procedures, automated
repair diagnostics and genuine parts.
You will always have the utmost
confidence in your measurements.
For information regarding self maintenance of this product, please
contact your Agilent office.
Agilent offers a wide range of additional expert test and measurement services for your equipment,
including initial start-up assistance,
onsite education and training, as
well as design, system integration,
and project management.
For more information on repair and
calibration services, go to:
www.agilent.com/find/removealldoubt
www.agilent.com
www.agilent.com/find/probes
For more information on Agilent Technologies’ products, applications or services,
please contact your local Agilent office.
The complete list is available at:
www.agilent.com/find/contactus
Americas
Canada
Latin America
United States
(877) 894-4414
305 269 7500
(800) 829-4444
Asia Pacific
Australia
China
Hong Kong
India
Japan
Korea
Malaysia
Singapore
Taiwan
Thailand
1 800 629 485
800 810 0189
800 938 693
1 800 112 929
0120 (421) 345
080 769 0800
1 800 888 848
1 800 375 8100
0800 047 866
1 800 226 008
Europe & Middle East
Austria
01 36027 71571
Belgium
32 (0) 2 404 93 40
Denmark
45 70 13 15 15
Finland
358 (0) 10 855 2100
France
0825 010 700
Germany
07031 464 6333
Ireland
1890 924 204
Israel
972-3-9288-504/544
Italy
39 02 92 60 8484
Netherlands
31 (0) 20 547 2111
Spain
34 (91) 631 3300
Sweden
0200-88 22 55
Switzerland
0800 80 53 53
United Kingdom 44 (0) 118 9276201
Other European Countries:
www.agilent.com/find/contactus
Revised: October 1, 2008
Product specifications and descriptions
in this document subject to change
without notice.
© Agilent Technologies, Inc. 2009
Printed in USA, June 26, 2009
5989-6432EN