MICROJET RECORDER (100mm) DATA SHEET PHC This recorder can record a maximum of 6 channels of DC voltage, mA, thermocouples and resistance bulbs. The adoption of an ink jet system makes it possible to record measured data in analog trace mode or to print in digital mode at a high speed. This 100mm-wide recorder performs recording clearly in 6 different colors. FEATURES 1. Compact size Compact and lightweight design, 199mm in depth and about 2.1kg in mass{weight}. 2.High quality recording • Ink jet system is used for recording and printing meas-ured data in 6 different colors at a high speed. Operating noise is minimized. • Six continuous traces without pen offset are possible with this compact size of recorder; a unique recording system is used for the first time in the industry. • Scale of each channel is printed on chart paper, eliminating the need for scales. 3.Easy setting of input signals DC voltage input (5mV span, 50V max.), 12 kinds of thermocouples (Type B, R, S, K, E, J, T, N, W, L, U, PN) and resistance bulbs (Pt100) can be set for each channel. 4.Digital printing Beside the analog recording of measured data, digital printing is also available (periodic printing, list printing, alarm printing, daily report printing, message printing). • Periodic data printing: Channel No., date, time, unit, chart speed, measured value • List printing: Date, time, unit, recording range, scaling value, alarm set value, chart speed, Tag No. • Alarm printing: Channel No., alarm type, on/off time, output relay No. • Daily report and totalized data printing: Printing of maximum, minimum, average and total of data measured during maximum 24 hours • Message printing : 10 messages, 16-character userentered messages 5.Interactive key operation Fluorescent indicator is used to clearly indicate alphanumeric characters and symbols. Input mode, recording range, alarm value, chart speed, etc., can be set according to the comments indicated by the display and operating keys. No bothersome operation is required. 6.Easy handling • A cartridge type recording device is used for easy replacement. • Allow to draw out the chart paper while recording. • Shortage of ink is detected in early stages and an alarm is given to the operator. • The end of chart paper is detected and indicated on the front panel display. • Shortage of ink and the end of chart paper alarm output is possible. 7. Full variety of functions • Alarm relay output/external control (record start/stop, chart speed change, data printing, message printing). This unit can easily be connected to the recorder by user (option). • Chart paper illumination lamp (option): The result of printing can be checked even in lower light. • Burnout function is provided as a standard. • Various recording: Enlarged/reduced recording, autorange recording, zone recording. • Calculation: Square root extraction, subtraction, engineering unit conversion, logarithm. • Language: Selectable 3 languages in display and printing. • Passcode security is configurable. • Transmission function: RS-485 (option) • The message print and alarm print function are operational, even when the recording mode is off. • All parameters of recording format, daily report, totalize, message and periodic data printing can be printed out. EDSX10-61j Date Nov. 24, 2011 PHC Input system 3 or 6 continuous recording 6 intermittent recording Thermocouple input...B, R, S, K, E, J, T, N, W, L, U, PN Resistance bulb input...Pt100Ω DC voltage input...50mV, 500mV, 5V, 50V range DC current input...4 to 20mA DC, 10 to 50mA DC (Shunt resistor (option) need to be connected to the terminal) Max. input voltage: • Thermocouple, resistance bulb and DC voltage (50mV, 500mV range) ...±10VDC or less • DC voltage input (5V, 50V range) ...±100VDC or less Input signal setting and change: Setting and change of input signal between thermocouple, resistance bulb and DC voltage (50mV, 500mV, 5V, 50V range) is possible for each channel by the setting pin in the instrument. Setting of recording range: Setting is possible within the reference range by using the keyboard. Burnout function:When thermocouple or resistance bulb input is disconnected, the recording is deflected to 100%. Reference range: Input points: Input signal: Kind Thermocouple Resistance bulb DC voltage Note:N W L U PN Pt100 Z fold 15.08m Continuous recording type 5 to 400 mm/h, continuous recording 401 to 1500 mm/h, intermittent recording Intermittent recording type 5 to 1500 mm/h Each can be set in 1 mm/h steps. Recording cycle: Intermitter recording...30 sec/all points Continuous recording...Depends on chart speed. <Calculation formula> 400 Recording = cycle [sec] Chart speed [mm/h] (not faster than 2 seconds.) Measuring cycle: Up to 3 inputs...160ms 6 inputs...320ms Service life of ink:(Depends on operating conditions) About 6 months for 6 points of linear recording at 20 mm/h of chart speed Chart handling: Tear off without disturbance of recording. Chart length: Chart speeds: SPECIFICATIONS Reference range 400 to 1760°C 0 to 1760°C 0 to 1760°C -200 to 1370°C -200 to 800°C -200 to 1100°C -200 to 400°C 0 to 1300°C 0 to 1760°C -200 to 900°C -200 to 400°C 0 to 1300°C B R S K E J T N W L U PN Pt100 -200 to 600°C -50 -500 -5 -50 to to to to +50mV +500mV +5V +50V Reference 752 to 32 to 32 to -328 to -328 to -328 to -328 to 32 to 32 to -328 to -328 to 32 to -328 to range 3200°F 3200°F 3200°F 2498°F 1472°F 2012°F 752°F 2372°F 3200°F 1652°F 752°F 2372°F 1112°F Scaling is possible within the range of -32767 to +32767 (decimal point may be put as necessary) : NICROSIL-NISIL (IEC584) : + side 5% Re, -side 26% Re .W (Hoskins Mfg. Co., U.S.A.) : +side Fe, -side Cu.Ni alloy (DIN43710) : +side Cu, -side Cu.Ni alloy (DIN43710) : Platinel : DIN IEC751 Recording system Writing system: Ink jet system, 6 colors Chart width: 100mm Recording color: No. 1 channel (orange), No. 2 channel (green), No. 3 channel (purple), No. 4 channel (red), No. 5 channel (black), No. 6 channel (blue) Recording color can be assigned for each channel. 2 Indicating system Indication: Fluorescent indication (blue-green), 20 characters x 2 lines Characters indicated: 5 x 7 dots, 4.16mm high, 2.25mm wide Contents of indication: (1)Measured value: Temperature...1 digit below decimal point; Voltage...6 digits (including sign and decimal point) Measured value of No.1 channel to No.6 can be indicated simultaneously. (2)Channel No.: 1 digit (1 to 6) (3)Engineering unit: Max. 7 digits (°C, °F, %, Pa, bar, ppm, m3/h, etc.) (4)Tag No....8 characters (5)Time: Year, month, day, hour, minute (6)Status indication: Record ON, chart end, battery alarm, alarm, ink shortage alarm, burnout, carriage failure Configuration: These can be set according to the comments indicated by operating keys as follows, Passcode Main chart speed Sub chart speed Alarm setting Record mode (trend/logging) Recording range Input signal List print request Tag No. Daily report setting Totalize function Communication parameter Date and time setting Ink monitor clear Illumination on/off Message definition Measured value shift Printing system Writing system: Ink jet system, 6 colors Periodic data printing: Measured value, unit, date, time, time line, chart speed, channel No. List printing: (1)Measured value list (date, time, channel No., measured value, unit) (2)Parameter list (date, time, channel No., recording range, scaling, unit, alarm set value, chart speed, Tag No.) (3)Test pattern (all characters and color patterns) Message printing:10 m e s s a g e s , 16 - ch a r a c t e r u s e rentered messages Alarm printing: Channel No., alarm type ( H, L, RH, RL), output relay No., on/off time Burnout printing: Burnout channel No. and time Other: Ink short age message, automatic range selection mark, recording start mark, chart speed change mark Note:Printing is not available for more than 401 mm/h (continuous recording), or more than 51 mm/h (intermittent recording). Performance and characteristics Accuracy and resolution: Performance under reference condit i o n (23± 2° C, 65± 10%RH, power voltage and frequency variation ±1%, warm-up time 30 minutes or more, vertical mounting, free from the effect of external noise) Indication (digital) Resolution 0.1°C 0.1°C 0.1°C ±(0.15% 0.1°C +1 digit) 0.1°C (without reference 0.1°C junction 0.1°C compen0.1°C sation 0.1°C error) 0.1°C 0.1°C 0.1°C Input Th e r- B moR couple S K E J T N W L U PN Resistance Pt100 bulb DC -50 voltage -500 -5 -50 Accuracy ±(0.15% +1 digit) to to to to +50mV ±(0.15% +500mV +1 digit) +5V +50V Withstand voltage: Input terminal - input terminal...500V AC, 1 minute Power supply terminal - ground... 2000V AC, 1 minute Input terminal - ground...500V AC, 1 minute Power terminal - input terminal...500V AC, 1 minute Between alarm terminals...750V AC, 1 minute (L eakage current 5mA or less, however, if the power supply is 24V DC, Leakage current of “Power supply terminal - ground” is 10mA or less) Reference junction compensation accuracy: K, E, J, T, N, L, U, PN ............. ±0.5°C (In case of minus input measurement: ±1.2°C) R, S, B, W . ............................ ±1°C (In case of minus input measurement: ±2.4°C) Common mode noise rejection: 120dB at 50, 60Hz ±0.1Hz Series mode noise rejection: 30dB at 50, 60Hz ±0.1Hz Physical data Mounting: Panel (may be inclined up to 30 ° backwards from the vertical.) Recording Accuracy Resolution α=90~60° Indication accuracy, 0.1mm ±0.25% miniof record- mum ing span 0.1°C 10µV 100µV 1mV 10mV Note: Indication accuracy is in % of reference range. Indication accuracy of B type TC is ±(0.36% + 1digit) between 400°C to 600°C. Indication accuracy of all type TC is ±(0.36% + 1digit) between -200°C to -100°C. Input resistance: Thermocouple:>10MΩ 50mV range: >10MΩ 500mV range: >100kΩ 5V and 50V range: >1MΩ Chart speed accuracy: ±0.1% (expansion and contraction of paper is not included) Clock accuracy: ±50ppm or less (monthly error; about 2 minutes) Imsulation resistance: 100MΩ or more (between each terminal and earth, at 500V DC) Two more recorders can be mounted side by side. Material: Case...Steel plate Front door frame...Polycarbonate with glass wool Mass{weight}: Approx. 2.1 kg (without option) Approx. 2.2 kg (with option) Case size: Bezel 144x144mm Depth 199mm Cutout 137x137mm Finish color: Case...Black, Front door frame...Black External terminals:Screw terminal (M4 screw) Power requirement 1)S upply voltage 100 to 240V AC products (9digit of code symbol = “D”) • Power supply voltage: 100V (-15%) to 240V (+10%) AC (Free power supply) • Supply frequency: 50/60Hz both employable • Power consumption: 100V AC with all options approximately 26VA 2)S upply voltage 24V DC products (9digit of code symbol = “L”) • Power supply voltage: 24V (±10%) DC • Power consumption: 26.4V DC with all options 26VA or less 3 PHC Operating environment (for devices to operate continuously) Usage environment:Indoor Temperature limits: 0 to 50°C Humidity limits: 20 to 80%RH, non condensing is required (temperature x humidity<3200) Vibration: 10 to 60Hz, 0.2m/s2{0.02G} or less Mounting position: Front inclination 0°, rear inclination 30°, left/right inclination 0° Signal source resistance: Thermocouple input...1kΩ or less Voltage input...Less than 0.1% of input resistance Resistance bulb input...Less than 10 Ω per line (Resistance of each wire of 3-wire system should be balanced with others.) Warm-up time: 30 min or more Shock: No external shock Environmental protection: IEC IP50(Front) / 20(Terminal) [Not covered by UL approval] Installation category: Ⅱ Pollution degree: 2 Operating altitude: 2000m max. Operating environment influence Effect of power supply fluctuation : (1)Supply voltage 100 to 240V AC products (9th digit of code symbol = “D”) Voltage variation: 85 to 264V AC (frequeucy: 50/60Hz) 100VAC base Change in indication… ±(0.1%+1 digit) max. Change in recording… ±0.2% of recording span, max. Frequency variation: 47 to 63Hz (power voltage: 100V AC) 50Hz base Change in indication… ±(0.1%+1 digit) max. Change in recording… ±0.2% of recording span, max. (2)Supply voltage 24V DC products (9th digit of code symbol = “L”) With 21.6 to 26.4V DC fluctuation, 24V DC base Change in indication....±(0.1% + 1 digit) max. Change in recording....±0.2% of recording span, max. Input signal source resistance or wiring resistance influence: Thermocouple... 10µV per 100Ω Vo l t a g e i n p u t . . . Va r i a t i o n o f 0 . 1 % change of resistance Change in indication... ±(0.1%+1 digit) max. Change in recording.... ±0.2% of recording span, max. Reistance bulb...Variation of resistance with changes in 10Ω per wire Change in indication... ±(0.1%+1 digit) max. 4 Change in recording... ±0.2% of recording span, max. (3 wires should be balanced.) Temperature influence: Change in indication... ±(0.3%+1digit) /10°C, max. Change in recording... ±0.5%/10°C, max. Mounting position influence: Inclination within 30° Change in indication... ±(0.1%+1 digit) max. Change in recording... ±0.2% of recording span, max. Vibration influence: Linear vibration with 10 to 60Hz of frequency and 0.2m/s 2{0.02G} of acceleration is applied to each of 3 directions for 2 hours. Change in indication...±(0.1%+1 digit) max. Change in recording... ±0.2% of recording span, max. Effect of external noise: Normal mode noise (50,60Hz±0.1Hz) ... 30dB or more Common mode noise (50,60Hz±0.1Hz) … 120dB or more Chart paper influence: Standard temperature/humidity: 20°C, 65%RH Expansion at 85%RH... 0.4% max. Contraction at 35%RH... 0.5% max. Alarm Setting from keyboard levels: Max. 4 levels for each channel High(H), Low(L), High-rate of change(RH), Low-rate of change(RL) Alarm action indication: Kind of alarm and output relay No. are indicated for each channel at occurrence of alarm. Printing: Channel No., kind of alarm, output relay No. and on/off time are printed on chart paper. Output: See optional specifications. Hysteresis: Approx. 0.5% of recording span Alarm timing: Recognition; 1 second (worst case) Action; additional 1 second (worst case) Alarm latch: Hold the alarm display and alarm output. Others: Shortage of ink and the end of chart paper alarm output is possible. Setting method: Number of alarm Alarm types: Transportation/storage condition (Detach a recording head from the main unit before transportation) Temperature limits: -10 to +60°C Humidity limits: 5 to 90%RH, non condensing is required Vibration: 10 to 60Hz, 2.45m/s2{0.25G} Shock: 294m/s2{30G} or less FUNCTIONS 1. Chart illumination: LED 2.Alarm output/3-point external control: This unit can be mounted from the rear side of the recorder. (1) Alarm output (DO): 6 points of relay contact N.O. (1a) or N.C. (1b) output for individual channel operation or common operation Contact capacity: N.O. contact 240V AC/3A 30V DC/3A (resistive load) N.C. contact 125V AC/0.4A 30V DC/2A (resistive load) Only following specifications obtain UL approval. 1 a c o n t a c t 3 0 V AC /30V DC 3A (resistive load) (2) External control (DI): The following control is possible with external contact signal. • Recording start/stop; Recording start/stop is effected by contact signal. Recording is started when contact is closed and stopped when contact is open. • Chart speed change; Selection between normal and remote chart speeds is effected by contact signal. Remote chart speed is selected when contact is closed and normal when contact is open. • Measured value printing; Measured value list printing (date, time, channel No., measured value, unit) is effected by contact signal. Printing is started when contact is closed. • Message printing Note: For external control, use a dry contact. Contact capacity: 12V DC, 0.05A, N.O. (1a) contact 3.Transmission function: RS - 4 8 5 i n t e r fa c e fo r t r a n s m i tt i n g meas-ured value and receiving the condition of setting. Transmission method Half-duplex bit serial Synchronizing method Start-stop synchronizing Code Binary Data length, 8 bits Parity: odd/even/none Stop bit: 1 or 2 Transmission speed 2400, 4800, 9600, 19200 bps Number of units connected Max. 31 units Transmission distance Max. 1 km Remarks: When connecting through RS-232C, be sure to use a 232 to 485 converter. The following shows a recommended converter. Maker: System Sakom Co., Ltd., Japan Tel: +81-3-3797-0211 Type: KS-485 Function Range setting Description Recording range can be set for each channel. Input setting Any input can be set for each channel. Skip function Used to skip recording, indication and alarm at any measuring point. List printing function Optional specifications Measured value list Date, time, and measured value unit can be printed. Parameter list Date, time, recording range, scaling, unit, kind of input, alarm set value, chart speed, and Tag No. can be printed. Test pattern All characters and color patterns can be printed. Periodic data printing function Time, date, chart speed, measured value and unit can be printed at fixed intervals. Printing can be enabled/disabled from keyboard. Message printing Maximum 10 messages, 16-character userentered messages can be printed. Alarm printing function Time, channel No., kind of alarm, and output relay No. can be printed when alarm is on or off. Unit indication Engineering units such as °C, °F, %, mV, mA, Pa, l, etc., are indicated (setting from key board). Scaling function Scaling with DC voltage input is possible. (Setting of decimal point is also possible within the range of -32767 to +32767). Subtract function Difference between any channels is recorded (channel is set from keyboard). Logarithm Measurd value can be displyed and printed by 10 n power. Auto-range recording Recording range is automatically changed for recording in the event of overrange or underrange (setting with keyboard). This function is not available for combination of zone recording and enlarged/reduced recording. Zone recording Recording area is divided into a maximum of 3 zones for recording. This function is not available for combination of automatic range selection and enlarged/reduced recording. Enlarged/reduced recording A part of recording area of each channel is expanded or contracted for recording. This function is not available for combination of automatic range selection and zone recording. Square-root extraction function Square-root extraction of DC voltage input is possible. Daily report function Measured value of every hour for maximum one day (24 data) in each channel is stored for printing. Maximum, minimum and average values are also printed at the same time. ONOFF operation, ON-OFF of each channel and operation start time/stop time can be set from keyboard. Totalize function Integrated value of every hour for maximum one day (24 data) in each channel is stored for printing (integration in 1 sec steps). Possible to print total value only. Total value is also printed at the same time. ON-OFF operation, ON-OFF of each channel and operation start time/stop time can be set from keyboard. Measured value shift Shift the zero point and inclination of the measured value so that the measured value can be adjusted according to other instruments. Memory backup Set data and clock function are protected by built-in lithium battery (expected battery life, about 10 years under normal temperature). Input filter Response is delayed according to sudden changes in input of each channel (1st order lag filter). Time constant setting range: 0 to 900 sec (setting from keyboard). Burnout function When thermocouple or resistance bulb input is disconnected, it is deflected 100%. Also, it is indicated and printed at the same time. Passcode 4 digits passcode security is available. Language English, German, or French is selectable for display and printing. 5 PHC Alarm latch function The alarm display and alarm output are held even after the cause of alarming was gone. ON-OFF operation can be set from keyboard. Cancellation of the held alarm can be made from external control (DI). Parameter copy Set parameters on any channel can be copied to any other channels. SCOPE OF DELIVERY Recorder, panel mounting bracket, accessories (ink cartridge (1), chart paper (1), ink absorption cloth (1)). Instruction manual (1). Note:Ink cartridge is not mounted on the recorder at the time of delivery. Spare parts Item CODE SYMBOLS Ink cartridge 0 0 3 - V Power Supply 100 to 240V AC 50/60Hz Note 1) 24V DC D L Chart paper illumination Without With A B Alarm output/external control Without 6-point alarm output (N.O.)/ 3-point external control Note 2) 6-point alarm output (N.C.)/ 3-point external control Note 1) 0 1 2 Y R Transmission function Without With RS-485 Note 1) Note 1) This is not conformed to UL61010-1:2001approval. Note 2) T his is conformed toUL approval with conditions attached. Please refer to "Optional specifications" on page 5. Remarks:Input signal Setting prior to delivery is as follows. • Thermocouple K: 0 to 1200°C Note: Contact Fuji Electric for additional features not listed such as Flow integration record and Calculation of input signals. 6 1 pc 1 box (6 charts) Description Recording points 3 continuous recording 6 continuous recording 6 intermittent recording 3 3 6 6 7 6 Unit of quantity for sale Chart paper (0 to 50, PEX00DL1-5000B 50 uniform division) 1 2 3 4 5 6 7 8 - 9 10 11 12 13 PHC Part No. PHZH1002 Other (optional items) Item Type Specification Shunt resistor PHZT1101 For 10Ω ± 0.1% PHZK1601 With 6-point alarm output(N.O.)/ 3-point external control PHZK1611 With 6-point alarm output(N.C.)/ 3-point external control Alarm output/ external control unit The product conforms to the requirements of the Electromagnetic compatibility Directive 89/336/EEC as detailed within the technical construction file number TN510406. The applicables standards used to demonstrate compliance are:EN 50081-1 :-1992 Conducted and Radiated emissions EN 50082-1 :-1992 Radiated immunity, ESD and FBT PHC OUTLINE DIAGRAMS (Unit:mm) 2 Mounting bracket 174.4 Alarm/external control unit (option) 136.4 Panel 144 Input termianl 30 136.4 Power terminal t t Transmission termianl (option) 41 144 158 199 26 Panel coutout When mounting multiple unit When mounting one unit 137+1.5 0 n : Number of units to be mounted 137+1.5 0 137+1.5 0 [144 x n–6]+20 Note: When panel mounting, two bracket are necessary either on the top and bottom of the recorder. CONNECTION DIAGRAMS Alarm/external control unit (option) Input terminal Resistance bulb – – + Thermocouple Voltage 11 21 12 11 ¨ Input 1 23 22 21 ¨ Input 2 33 32 31 ¨ Input 3 43 42 41 ¨ Input 4 53 52 51 ¨ Input 5 63 62 61 ¨ Input 6 SHLD TRX2 TRX1 ¨ Transmission – + (option) DI 2 12 22 13 23 14 24 Alarm 1 15 25 Alarm 2 16 26 Alarm 3 17 27 Alarm 4 18 28 Alarm 5 19 29 Alarm 6 + 13 DI 1 DI 1 DI 2 DI 3 Note: Power terminal AC AC G Power source 100 to 240V AC 50/60Hz DI 3 + G Power source 24V DC Record start or message print 1 Chart speed change or message point 2 Measured value printing or alarm latch cancel or message print 3 Alarm relay contact is selectable N.O. or N.C. by Code Symbols. 7 PHC Caution on Safety *Before using this product, be sure to read its instruction manual in advance. International Sales Div Sales Group Gate City Ohsaki, East Tower, 11-2, Osaki 1-chome, Shinagawa-ku, Tokyo 141-0032, Japan http://www.fujielectric.com Phone: 81-3-5435-7280, 7281 Fax: 81-3-5435-7425 http://www.fujielectric.com/products/instruments/ Information in this catalog is subject to change without notice. Printed in Japan