FUJI EDSX10-61J

MICROJET RECORDER (100mm)
DATA SHEET
PHC
This recorder can record a maximum of 6 channels of
DC voltage, mA, thermocouples and resistance bulbs.
The adoption of an ink jet system makes it possible to
record measured data in analog trace mode or to print in
digital mode at a high speed. This 100mm-wide recorder
performs recording clearly in 6 different colors.
FEATURES
1. Compact size
Compact and lightweight design, 199mm in depth and
about 2.1kg in mass{weight}.
2.High quality recording
• Ink jet system is used for recording and printing
meas-ured data in 6 different colors at a high speed.
Operating noise is minimized.
• Six continuous traces without pen offset are possible
with this compact size of recorder; a unique recording system is used for the first time in the industry.
• Scale of each channel is printed on chart paper, eliminating the need for scales.
3.Easy setting of input signals
DC voltage input (5mV span, 50V max.), 12 kinds of
thermocouples (Type B, R, S, K, E, J, T, N, W, L, U, PN)
and resistance bulbs (Pt100) can be set for each channel.
4.Digital printing
Beside the analog recording of measured data, digital
printing is also available (periodic printing, list printing,
alarm printing, daily report printing, message printing).
• Periodic data printing: Channel No., date, time, unit,
chart speed, measured value
• List printing: Date, time, unit, recording range, scaling value, alarm set value, chart speed, Tag No.
• Alarm printing: Channel No., alarm type, on/off time,
output relay No.
• Daily report and totalized data printing: Printing of
maximum, minimum, average and total of data
measured during maximum 24 hours
• Message printing : 10 messages, 16-character userentered messages
5.Interactive key operation
Fluorescent indicator is used to clearly indicate alphanumeric characters and symbols.
Input mode, recording range, alarm value, chart speed,
etc., can be set according to the comments indicated
by the display and operating keys. No bothersome operation is required.
6.Easy handling
• A cartridge type recording device is used for easy
replacement.
• Allow to draw out the chart paper while recording.
• Shortage of ink is detected in early stages and an
alarm is given to the operator.
• The end of chart paper is detected and indicated on
the front panel display.
• Shortage of ink and the end of chart paper alarm output is possible.
7. Full variety of functions
• Alarm relay output/external control (record start/stop,
chart speed change, data printing, message printing).
This unit can easily be connected to the recorder by
user (option).
• Chart paper illumination lamp (option): The result of
printing can be checked even in lower light.
• Burnout function is provided as a standard.
• Various recording: Enlarged/reduced recording, autorange recording, zone recording.
• Calculation: Square root extraction, subtraction, engineering unit conversion, logarithm.
• Language: Selectable 3 languages in display and
printing.
• Passcode security is configurable.
• Transmission function: RS-485 (option)
• The message print and alarm print function are operational, even when the recording mode is off.
• All parameters of recording format, daily report,
totalize, message and periodic data printing can be
printed out.
EDSX10-61j
Date
Nov. 24, 2011
PHC
Input system
3 or 6 continuous recording
6 intermittent recording
Thermocouple input...B, R, S, K, E, J, T,
N, W, L, U, PN
Resistance bulb input...Pt100Ω
DC voltage input...50mV, 500mV, 5V,
50V range
DC current input...4 to 20mA DC, 10 to
50mA DC
(Shunt resistor (option) need to be
connected to the terminal)
Max. input voltage:
• Thermocouple, resistance bulb and
DC voltage (50mV, 500mV range)
...±10VDC or less
• DC voltage input (5V, 50V range)
...±100VDC or less
Input signal setting and change:
Setting and change of input signal between thermocouple, resistance bulb
and DC voltage (50mV, 500mV, 5V, 50V
range) is possible for each channel by
the setting pin in the instrument.
Setting of recording range:
Setting is possible within the reference
range by using the keyboard.
Burnout function:When thermocouple or resistance bulb
input is disconnected, the recording is
deflected to 100%.
Reference range:
Input points:
Input signal:
Kind
Thermocouple
Resistance
bulb
DC voltage
Note:N
W
L
U
PN
Pt100
Z fold 15.08m
Continuous recording type
5 to 400 mm/h, continuous recording
401 to 1500 mm/h, intermittent recording
Intermittent recording type
5 to 1500 mm/h
Each can be set in 1 mm/h steps.
Recording cycle: Intermitter recording...30 sec/all points
Continuous recording...Depends on
chart speed.
<Calculation formula>
400
Recording
=
cycle [sec] Chart speed [mm/h]
(not faster than 2 seconds.)
Measuring cycle: Up to 3 inputs...160ms
6 inputs...320ms
Service life of ink:(Depends on operating conditions)
About 6 months for 6 points of linear
recording at 20 mm/h of chart speed
Chart handling: Tear off without disturbance of recording.
Chart length:
Chart speeds:
SPECIFICATIONS
Reference range
400 to 1760°C
0 to 1760°C
0 to 1760°C
-200 to 1370°C
-200 to
800°C
-200 to 1100°C
-200 to
400°C
0 to 1300°C
0 to 1760°C
-200 to
900°C
-200 to
400°C
0 to 1300°C
B
R
S
K
E
J
T
N
W
L
U
PN
Pt100
-200
to
600°C
-50
-500
-5
-50
to
to
to
to
+50mV
+500mV
+5V
+50V
Reference
752 to
32 to
32 to
-328 to
-328 to
-328 to
-328 to
32 to
32 to
-328 to
-328 to
32 to
-328
to
range
3200°F
3200°F
3200°F
2498°F
1472°F
2012°F
752°F
2372°F
3200°F
1652°F
752°F
2372°F
1112°F
Scaling is possible
within the range of
-32767 to +32767
(decimal point may be
put as necessary)
: NICROSIL-NISIL (IEC584)
: + side 5% Re, -side 26% Re .W (Hoskins Mfg. Co.,
U.S.A.)
: +side Fe, -side Cu.Ni alloy (DIN43710)
: +side Cu, -side Cu.Ni alloy (DIN43710)
: Platinel
: DIN IEC751
Recording system
Writing system: Ink jet system, 6 colors
Chart width:
100mm
Recording color: No. 1 channel (orange), No. 2 channel
(green), No. 3 channel (purple), No. 4
channel (red), No. 5 channel (black),
No. 6 channel (blue)
Recording color can be assigned for
each channel.
2
Indicating system
Indication:
Fluorescent indication (blue-green),
20 characters x 2 lines
Characters indicated:
5 x 7 dots, 4.16mm high, 2.25mm wide
Contents of indication:
(1)Measured value:
Temperature...1 digit below decimal point;
Voltage...6 digits (including sign
and decimal point)
Measured value of No.1 channel to
No.6 can be indicated simultaneously.
(2)Channel No.: 1 digit (1 to 6)
(3)Engineering unit: Max. 7 digits (°C,
°F, %, Pa, bar, ppm, m3/h, etc.)
(4)Tag No....8 characters
(5)Time: Year, month, day, hour, minute
(6)Status indication: Record ON, chart
end, battery alarm, alarm, ink shortage alarm, burnout, carriage failure
Configuration:
These can be set according to the
comments indicated by operating keys
as follows,
Passcode
Main chart speed
Sub chart speed
Alarm setting
Record mode (trend/logging)
Recording range
Input signal
List print request
Tag No.
Daily report setting
Totalize function
Communication parameter
Date and time setting
Ink monitor clear
Illumination on/off
Message definition
Measured value shift
Printing system
Writing system: Ink jet system, 6 colors
Periodic data printing:
Measured value, unit, date, time, time
line, chart speed, channel No.
List printing:
(1)Measured value list (date, time,
channel No., measured value, unit)
(2)Parameter list (date, time, channel
No., recording range, scaling, unit,
alarm set value, chart speed, Tag No.)
(3)Test pattern (all characters and
color patterns)
Message printing:10 m e s s a g e s , 16 - ch a r a c t e r u s e rentered messages
Alarm printing: Channel No., alarm type ( H, L, RH,
RL), output relay No., on/off time
Burnout printing: Burnout channel No. and time
Other:
Ink short age message, automatic
range selection mark, recording start
mark, chart speed change mark
Note:Printing is not available for more than 401 mm/h (continuous
recording), or more than 51 mm/h (intermittent recording).
Performance and characteristics
Accuracy and resolution:
Performance under reference condit i o n (23± 2° C, 65± 10%RH, power
voltage and frequency variation ±1%,
warm-up time 30 minutes or more,
vertical mounting, free from the effect
of external noise)
Indication (digital)
Resolution
0.1°C
0.1°C
0.1°C
±(0.15%
0.1°C
+1 digit)
0.1°C
(without
reference 0.1°C
junction
0.1°C
compen0.1°C
sation
0.1°C
error)
0.1°C
0.1°C
0.1°C
Input
Th e r- B
moR
couple
S
K
E
J
T
N
W
L
U
PN
Resistance
Pt100
bulb
DC
-50
voltage -500
-5
-50
Accuracy
±(0.15%
+1 digit)
to
to
to
to
+50mV ±(0.15%
+500mV +1 digit)
+5V
+50V
Withstand voltage:
Input terminal - input terminal...500V
AC, 1 minute
Power supply terminal - ground...
2000V AC, 1 minute
Input terminal - ground...500V AC, 1
minute
Power terminal - input terminal...500V
AC, 1 minute
Between alarm terminals...750V AC, 1
minute
(L eakage current 5mA or less, however, if the power supply is 24V DC,
Leakage current of “Power supply
terminal - ground” is 10mA or less)
Reference junction compensation accuracy:
K, E, J, T, N, L, U, PN ............. ±0.5°C
(In case of minus input measurement:
±1.2°C)
R, S, B, W . ............................ ±1°C
(In case of minus input measurement:
±2.4°C)
Common mode noise rejection:
120dB at 50, 60Hz ±0.1Hz
Series mode noise rejection:
30dB at 50, 60Hz ±0.1Hz
Physical data
Mounting:
Panel (may be inclined up to 30 ° backwards from the vertical.)
Recording
Accuracy
Resolution
α=90~60°
Indication
accuracy, 0.1mm
±0.25%
miniof record- mum
ing span
0.1°C
10µV
100µV
1mV
10mV
Note: Indication accuracy is in % of reference range.
Indication accuracy of B type TC is ±(0.36% + 1digit) between
400°C to 600°C.
Indication accuracy of all type TC is ±(0.36% + 1digit) between
-200°C to -100°C.
Input resistance: Thermocouple:>10MΩ
50mV range: >10MΩ
500mV range: >100kΩ
5V and 50V range: >1MΩ
Chart speed accuracy:
±0.1% (expansion and contraction of
paper is not included)
Clock accuracy: ±50ppm or less (monthly error; about
2 minutes)
Imsulation resistance:
100MΩ or more (between each terminal and earth, at 500V DC)
Two more recorders can be mounted
side by side.
Material:
Case...Steel plate
Front door frame...Polycarbonate with
glass wool
Mass{weight}:
Approx. 2.1 kg (without option)
Approx. 2.2 kg (with option)
Case size:
Bezel 144x144mm
Depth 199mm
Cutout 137x137mm
Finish color:
Case...Black, Front door frame...Black
External terminals:Screw terminal (M4 screw)
Power requirement
1)S upply voltage 100 to 240V AC products (9digit of
code symbol = “D”)
• Power supply voltage:
100V (-15%) to 240V (+10%) AC (Free
power supply)
• Supply frequency:
50/60Hz both employable
• Power consumption:
100V AC with all options approximately 26VA
2)S upply voltage 24V DC products (9digit of code
symbol = “L”)
• Power supply voltage:
24V (±10%) DC
• Power consumption:
26.4V DC with all options 26VA or less
3
PHC
Operating environment
(for devices to operate continuously)
Usage environment:Indoor
Temperature limits:
0 to 50°C
Humidity limits: 20 to 80%RH, non condensing is required
(temperature x humidity<3200)
Vibration:
10 to 60Hz, 0.2m/s2{0.02G} or less
Mounting position:
Front inclination 0°, rear inclination
30°, left/right inclination 0°
Signal source resistance:
Thermocouple input...1kΩ or less
Voltage input...Less than 0.1% of input resistance
Resistance bulb input...Less than 10 Ω
per line (Resistance of each wire of
3-wire system should be balanced
with others.)
Warm-up time:
30 min or more
Shock:
No external shock
Environmental protection:
IEC IP50(Front) / 20(Terminal)
[Not covered by UL approval]
Installation category: Ⅱ
Pollution degree: 2
Operating altitude:
2000m max.
Operating environment influence
Effect of power supply fluctuation :
(1)Supply voltage 100 to 240V AC products (9th digit
of code symbol = “D”)
Voltage variation: 85 to 264V AC
(frequeucy: 50/60Hz) 100VAC base
Change in indication… ±(0.1%+1
digit) max.
Change in recording… ±0.2% of
recording span, max.
Frequency variation: 47 to 63Hz
(power voltage: 100V AC) 50Hz base
Change in indication… ±(0.1%+1
digit) max.
Change in recording… ±0.2% of
recording span, max.
(2)Supply voltage 24V DC products (9th digit of code
symbol = “L”)
With 21.6 to 26.4V DC fluctuation,
24V DC base
Change in indication....±(0.1% + 1
digit) max.
Change in recording....±0.2% of recording span, max.
Input signal source resistance or wiring resistance influence:
Thermocouple... 10µV per 100Ω
Vo l t a g e i n p u t . . . Va r i a t i o n o f 0 . 1 %
change of resistance
Change in indication... ±(0.1%+1
digit) max.
Change in recording.... ±0.2% of recording span, max.
Reistance bulb...Variation of resistance with changes in 10Ω per wire
Change in indication... ±(0.1%+1
digit) max.
4
Change in recording... ±0.2% of recording span, max.
(3 wires should be balanced.)
Temperature influence:
Change in indication... ±(0.3%+1digit)
/10°C, max.
Change in recording... ±0.5%/10°C,
max.
Mounting position influence:
Inclination within 30°
Change in indication... ±(0.1%+1
digit) max.
Change in recording... ±0.2% of recording span, max.
Vibration influence:
Linear vibration with 10 to 60Hz of
frequency and 0.2m/s 2{0.02G} of acceleration is applied to each of 3 directions for 2 hours.
Change in indication...±(0.1%+1
digit) max.
Change in recording... ±0.2% of recording span, max.
Effect of external noise:
Normal mode noise (50,60Hz±0.1Hz)
... 30dB or more
Common mode noise (50,60Hz±0.1Hz)
… 120dB or more
Chart paper influence:
Standard temperature/humidity: 20°C,
65%RH
Expansion at 85%RH... 0.4% max.
Contraction at 35%RH... 0.5% max.
Alarm
Setting from keyboard
levels:
Max. 4 levels for each channel
High(H), Low(L), High-rate of
change(RH), Low-rate of change(RL)
Alarm action indication:
Kind of alarm and output relay No. are
indicated for each channel at occurrence of alarm.
Printing:
Channel No., kind of alarm, output relay No. and on/off time are printed on
chart paper.
Output:
See optional specifications.
Hysteresis:
Approx. 0.5% of recording span
Alarm timing:
Recognition; 1 second (worst case)
Action; additional 1 second (worst case)
Alarm latch:
Hold the alarm display and alarm output.
Others:
Shortage of ink and the end of chart
paper alarm output is possible.
Setting method:
Number of alarm
Alarm types:
Transportation/storage condition
(Detach a recording head from the main unit before transportation)
Temperature limits:
-10 to +60°C
Humidity limits: 5 to 90%RH, non condensing is required
Vibration:
10 to 60Hz, 2.45m/s2{0.25G}
Shock:
294m/s2{30G} or less
FUNCTIONS
1. Chart illumination:
LED
2.Alarm output/3-point external control:
This unit can be mounted from the
rear side of the recorder.
(1) Alarm output (DO):
6 points of relay contact N.O. (1a) or
N.C. (1b) output for individual channel
operation or common operation
Contact capacity:
N.O. contact 240V AC/3A
30V DC/3A
(resistive load)
N.C. contact 125V AC/0.4A
30V DC/2A
(resistive load)
Only following specifications obtain
UL approval.
1 a c o n t a c t 3 0 V AC
/30V DC 3A
(resistive load)
(2) External control (DI):
The following control is possible with
external contact signal.
• Recording start/stop;
Recording start/stop is effected by
contact signal. Recording is started
when contact is closed and stopped
when contact is open.
• Chart speed change;
Selection between normal and remote chart speeds is effected by
contact signal. Remote chart speed
is selected when contact is closed
and normal when contact is open.
• Measured value printing;
Measured value list printing (date,
time, channel No., measured value,
unit) is effected by contact signal. Printing is started when contact is closed.
• Message printing
Note: For external control, use a dry contact.
Contact capacity: 12V DC, 0.05A,
N.O. (1a) contact
3.Transmission function:
RS - 4 8 5 i n t e r fa c e fo r t r a n s m i tt i n g
meas-ured value and receiving the
condition of setting.
Transmission method
Half-duplex bit serial
Synchronizing method
Start-stop synchronizing
Code
Binary Data length, 8 bits
Parity: odd/even/none
Stop bit: 1 or 2
Transmission speed
2400, 4800, 9600, 19200 bps
Number of units connected
Max. 31 units
Transmission distance
Max. 1 km
Remarks: When connecting through RS-232C, be sure to use a
232 to 485 converter.
The following shows a recommended converter.
Maker: System Sakom Co., Ltd., Japan
Tel:
+81-3-3797-0211
Type: KS-485
Function
Range setting
Description
Recording range can be set for each channel.
Input setting
Any input can be set for each channel.
Skip function
Used to skip recording, indication and alarm at
any measuring point.
List printing function
Optional specifications
Measured
value list
Date, time, and measured value unit can be
printed.
Parameter list
Date, time, recording range, scaling, unit, kind
of input, alarm set value, chart speed, and Tag
No. can be printed.
Test pattern
All characters and color patterns can be printed.
Periodic data
printing function
Time, date, chart speed, measured value and
unit can be printed at fixed intervals. Printing
can be enabled/disabled from keyboard.
Message
printing
Maximum 10 messages, 16-character userentered messages can be printed.
Alarm printing
function
Time, channel No., kind of alarm, and output
relay No. can be printed when alarm is on or
off.
Unit indication
Engineering units such as °C, °F, %, mV, mA,
Pa, l, etc., are indicated (setting from key
board).
Scaling function
Scaling with DC voltage input is possible. (Setting of decimal point is also possible within the
range of -32767 to +32767).
Subtract function
Difference between any channels is recorded
(channel is set from keyboard).
Logarithm
Measurd value can be displyed and printed by
10 n power.
Auto-range
recording
Recording range is automatically changed for
recording in the event of overrange or underrange (setting with keyboard). This function is
not available for combination of zone recording
and enlarged/reduced recording.
Zone recording
Recording area is divided into a maximum of 3
zones for recording.
This function is not available for combination
of automatic range selection and enlarged/reduced recording.
Enlarged/reduced
recording
A part of recording area of each channel is expanded or contracted for recording.
This function is not available for combination of
automatic range selection and zone recording.
Square-root extraction function
Square-root extraction of DC voltage input is
possible.
Daily report
function
Measured value of every hour for maximum
one day (24 data) in each channel is stored
for printing. Maximum, minimum and average
values are also printed at the same time. ONOFF operation, ON-OFF of each channel and
operation start time/stop time can be set from
keyboard.
Totalize function
Integrated value of every hour for maximum
one day (24 data) in each channel is stored for
printing (integration in 1 sec steps). Possible
to print total value only. Total value is also
printed at the same time. ON-OFF operation,
ON-OFF of each channel and operation start
time/stop time can be set from keyboard.
Measured value
shift
Shift the zero point and inclination of the measured value so that the measured value can be
adjusted according to other instruments.
Memory backup
Set data and clock function are protected by
built-in lithium battery (expected battery life,
about 10 years under normal temperature).
Input filter
Response is delayed according to sudden
changes in input of each channel (1st order lag
filter). Time constant setting range: 0 to 900
sec (setting from keyboard).
Burnout function
When thermocouple or resistance bulb input is
disconnected, it is deflected 100%. Also, it is
indicated and printed at the same time.
Passcode
4 digits passcode security is available.
Language
English, German, or French is selectable for
display and printing.
5
PHC
Alarm latch
function
The alarm display and alarm output are held
even after the cause of alarming was gone.
ON-OFF operation can be set from keyboard. Cancellation of the held alarm can be made
from external control (DI).
Parameter
copy
Set parameters on any channel can be copied
to any other channels.
SCOPE OF DELIVERY
Recorder, panel mounting bracket, accessories (ink cartridge (1), chart paper (1), ink absorption cloth (1)).
Instruction manual (1).
Note:Ink cartridge is not mounted on the recorder at the time of delivery.
Spare parts
Item
CODE SYMBOLS
Ink cartridge
0 0 3 -
V
Power Supply
100 to 240V AC 50/60Hz Note 1)
24V DC
D
L
Chart paper illumination
Without
With
A
B
Alarm output/external control
Without
6-point alarm output (N.O.)/
3-point external control Note 2)
6-point alarm output (N.C.)/
3-point external control Note 1)
0
1
2
Y
R
Transmission function
Without
With RS-485 Note 1)
Note 1) This is not conformed to UL61010-1:2001approval.
Note 2) T
his is conformed toUL approval with conditions attached.
Please refer to "Optional specifications" on page 5.
Remarks:Input signal
Setting prior to delivery is as follows.
• Thermocouple K: 0 to 1200°C
Note: Contact Fuji Electric for additional features not listed such as
Flow integration record and Calculation of input signals.
6
1 pc
1 box (6 charts)
Description
Recording points
3 continuous recording
6 continuous recording
6 intermittent recording
3 3
6 6
7 6
Unit of quantity for sale
Chart paper (0 to 50,
PEX00DL1-5000B
50 uniform division)
1 2 3 4 5 6 7 8 - 9 10 11 12 13
PHC
Part No.
PHZH1002
Other (optional items)
Item
Type
Specification
Shunt resistor
PHZT1101
For 10Ω ± 0.1%
PHZK1601
With 6-point alarm output(N.O.)/
3-point external control
PHZK1611
With 6-point alarm output(N.C.)/
3-point external control
Alarm output/
external control unit
The product conforms to the requirements of the
Electromagnetic compatibility Directive 89/336/EEC as
detailed within the technical construction file number
TN510406. The applicables standards used to demonstrate
compliance are:EN 50081-1 :-1992
Conducted and Radiated emissions
EN 50082-1 :-1992
Radiated immunity, ESD and FBT
PHC
OUTLINE DIAGRAMS (Unit:mm)
2
Mounting bracket
174.4
Alarm/external
control unit
(option)
136.4
Panel
144
Input termianl
30
136.4
Power terminal
t
t
Transmission
termianl
(option)
41
144
158
199
26
Panel coutout
When mounting multiple unit
When mounting one unit
137+1.5
0
n : Number of units
to be mounted
137+1.5
0
137+1.5
0
[144 x n–6]+20
Note:
When panel mounting, two bracket are necessary
either on the top and bottom of the recorder.
CONNECTION DIAGRAMS
Alarm/external
control unit
(option)
Input terminal
Resistance
bulb
–
–
+ Thermocouple
Voltage
11
21
12
11
¨ Input 1
23
22
21
¨ Input 2
33
32
31
¨ Input 3
43
42
41
¨ Input 4
53
52
51
¨ Input 5
63
62
61
¨ Input 6
SHLD TRX2 TRX1 ¨ Transmission
–
+
(option)
DI 2
12
22
13
23
14
24 Alarm 1
15
25 Alarm 2
16
26 Alarm 3
17
27 Alarm 4
18
28 Alarm 5
19
29 Alarm 6
+
13
DI 1
DI 1
DI 2
DI 3
Note:
Power terminal
AC
AC
G
Power source
100 to 240V AC
50/60Hz
DI 3
+
G
Power source
24V DC
Record start or message print 1
Chart speed change or message point 2
Measured value printing or alarm latch cancel or message print 3
Alarm relay contact is selectable N.O. or N.C. by Code Symbols.
7
PHC
Caution on Safety
*Before using this product, be sure to read its instruction manual in advance.
International Sales Div
Sales Group
Gate City Ohsaki, East Tower, 11-2, Osaki 1-chome,
Shinagawa-ku, Tokyo 141-0032, Japan
http://www.fujielectric.com
Phone: 81-3-5435-7280, 7281 Fax: 81-3-5435-7425
http://www.fujielectric.com/products/instruments/
Information in this catalog is subject to change without notice.