Agilent B2902/12A Precision Source/Measure Unit How to make a sweep measurement with 2 channel sources This material shows how to perform a sweep measurement with 2 channel sources through an example of IdVg characterization of FET. Figure 1 illustrates the connection and condition supposed in the example of measuring FET using a member of the B2902A or B2912A. Ch1 A High Force Gate - V Ch2 - Drain V - Source A - Low Force DUT : FET Ch1 Source Mode Start Value Stop Value Points Current Limit : Gate : Voltage :0V :2V : 101 : 1 uA Ch2 Source Mode Start Value Stop Value Points Current Limit : Drain : Voltage :2V :2V : 101 : 100 mA Figure 1. Connection and condition supposed in the example Figure 2 shows the timing chart for the sweep measurement using 2 channel sources with the front panel operation. The specified source values are sourced immediately after turning on each . Then, when you press , the instrument will make a sweep measurement. If it is necessary, you can specify any measurement trigger delay time which is the wait time after sourcing each source value and before making a measurement. The measurement time consists of Measurement Speed and some overhead time. Measurement Speed is the parameter specified by the user. Overhead time includes the time to change the measurement range, etc. Ch 2 V or I Output ON :Measurement Source Value Start/Stop Value Output Off State TMD TM TMD TM TMD TM Time TMD Ch 1 V or I Stop Value Source Value TMD TMD TM TM TM Start Value Output Off State Output ON Press Trigger TMD : Measurement Trigger Delay Time Internal Trigger Time TM : Measurement Time Figure 2. Timing chart for the sweep measurement using 2 channel sources Agilent B2902/12A Precision Source/Measure Unit Performing a sweep measurement with 2 channel sources 1. Setting up Channel 1 Step 1. Press Step 2. Press repeatedly until Single View for Channel 1 is shown in the display. to edit the channel 1 source function, and then select (1) Press Mode Step 3. Press Press (2) Enter Source Value to edit the channel 1 limit value, and then enter 1 uA to set it to 1 uA. (1) Press Limit Step 5. Press current. (2) Press VOLTS (V) to edit the channel 1 source value, and then enter 0 V to set it to 0 V. (1) Press Source Step 4. to set it to the voltage source. (2) Enter Limit Value to configure the channel 1 measurement parameter, and then select (1) Press Measure (2) Press AMPS (I) to set it to the Agilent B2902/12A Precision Source/Measure Unit Step 6. Press Sub-Panel. to change the keys shown in Assist keys, and then press (1) Press Show Sweep Step 7. Press , then press (2) Sweep Sub-Panel appears to turn on Single Linear Sweep Mode. After turning on Single Linear Sweep Mode, you can see Source Shape which shows the single linear sweep mode. (1) Press LINEAR SINGLE Step 8. Rotate to show Sweep Source Shape to select Channel 1 Sweep Parameters and set them up as below. (Start: 0 V, Stop: 2 V, Points: 101, Step: 20 mV) Agilent B2902/12A Precision Source/Measure Unit 2. Setting up Channel 2 Step 1. Press Step 2. Press to show Single View for Channel 2. to edit the channel 2 source function, and then select (1) Press Mode Step 3. Press Press (2) Enter Source Value to edit the channel 2 limit value, and then enter 100 mA to set it to 100 mA. (1) Press Limit Step 5. Press current. (2) Press VOLTS (V) to edit the channel 2 source value, and then enter 0 V to set it to 0 V. (1) Press Source Step 4. to set it to the voltage source. (2) Enter Limit Value to configure the channel 2 measurement parameter, and then select (1) Press Measure (2) Press AMPS (I) to set it to the Agilent B2902/12A Precision Source/Measure Unit Step 6. Rotate and press to edit the channel 2 sweep parameters. Then press Linear Sweep Mode. (1) Rotate and press to edit Channel 2 Sweep Parameters Step 7. Rotate (2) Press LINEAR SINGLE to select Channel 2 Sweep Parameters and set them up as below. (Start: 2 V, Stop: 2 V, Points: 101, Step: 0 mV) 3. Setting up Graph View Step 1. Press to show Graph View Step 2. Rotate and press (1) Rotate Type field to edit the X-axis data type to select X-axis Data (2) Press Type to edit X-axis Data to turn on Single Agilent B2902/12A Precision Source/Measure Unit Step 3. Press , and then select to set the X-axis data type to the channel 1 voltage. (1) Press More … 4. (2) Press Ch 1 V (V1) Performing measurement Step 1. Press measurement. for both channels to source the voltage, and then press (The status information will show to perform a during the measurement.) Step 2. Press to adjust the scale of the graph after finishing the measurement. Now you can see the measurement result on the GUI of the B2902/12A as bellow. Step 3. If you’d like to change the graph scaling of Y-axis from LINEAR to LOG, rotate and press Y-axis scaling. (1) Rotate and press Y-axis scaling to select (2) Press LOG to select Agilent B2902/12A Precision Source/Measure Unit Step 4. If you’d like to change the graph scale value of Y-axis, rotate and press (1) Rotate and press Y-axis scale value to select to modify Y-axis scale value. (2) Enter the scale such as 10 pA Configuring the measurement speed In the default setting, the instrument selects the appropriate measurement speed and range automatically to get the fine accuracy. However, you can also specify these parameters on the GUI of the B2902/12A to meet a variety of the requirement to the measurement conditions. For example, let’s try to change the measurement speed to NORMAL to make a measurement more carefully. If you select NORMAL, the aperture time is set to 1 PLC. Here, PLC stands for power line cycle and the specified number of power line cycles is used per a measurement. Step 1. Press Step 2. Press repeatedly until Single View for Channel 2 is shown in the display. to edit the channel 2 measurement speed, and then select (If you can’t see in Assist keys, press (1) Press Speed to set it to NORMAL. to change the keys shown in Assist keys.) (2) Press NORMAL Agilent B2902/12A Precision Source/Measure Unit Configuring the measurement range operation The parameters which configure the measurement range operation can be displayed in Range Sub-panel in Single View, although Sweep Sub-Panel is shown at this moment. In the default setting, the B2902/12A performs the current measurement using 1 uA current minimum measurement range with AUTO range operation. With AUTO range operation, the B2902/12A selects the proper range for the measurement with specified minimum measurement range so that you don’t need to take care about it. To know how to change the measurement range setting, try to configure to use 10 nA current minimum measurement range with AUTO range operation. Step 1. Press repeatedly until Single View for Channel 2 is shown in the display. Step 2. Press to show Range Sub-Panel. (If you can’t see change the keys shown in Assist keys.) (1) Press More... (if necessary) Step 3. Rotate and press in Assist keys, press to (2) Press Hide Sweep to edit the channel 2 current minimum measurement range, and then select to set it to 10 nA. (1) Rotate and press to edit Current Minimum Measurement Range (2) Press 10nA Agilent B2902/12A Precision Source/Measure Unit If you’d like to fix the measurement range, you can select FIXED range operation as below. Step 4. Rotate and press to edit the channel 2 current measurement range operation. Then Select to set it to FIXED. (1) Rotate and press to edit Current Measurement Range Operation (2) Press FIXED Configuring the measurement trigger delay time The trigger parameters including the measurement trigger delay time and the trigger period can be displayed in Trigger Sub-panel in Single View, although Range Sub-Panel is shown at this moment. In the default setting, the trigger type is set to the automatic trigger type (AUTO) so that you don’t need to specify these trigger parameters. If you’d like to specify the measurement trigger delay time, take the following steps. Step 1. Press repeatedly until Single View for Channel 1 is shown in the display. Step 2. Press to show Trigger Sub-Panel. (If you can’t see change the keys shown in Assist keys.) (1) Press More... (if necessary) Step 3. Press in Assist keys, press (2) Press Show Trigger to edit the channel 1 trigger type, and then select (1) Press (2) Press SYNC to set it to SYNC. to Agilent B2902/12A Precision Source/Measure Unit Step 4. Rotate to select Channel 1 Trigger Parameters and set them up as below. (Source Trigger Count: 101, Measurement Trigger Count: 101, Measurement Trigger Delay Time: 100ms) Note) Source and Measurement Trigger Count should be the same number as Sweep Points. Step 5. Press to show Single View for Channel 2 is shown in the display. Step 6. Rotate and press to edit the channel 2 trigger type, and then select (1) Rotate and press Trigger Type Step 7. Rotate to edit to set it to SYNC. (2) Press SYNC to select Channel 2 Trigger Parameters and set them up as below. (Source Trigger Count: 101, Measurement Trigger Count: 101, Measurement Trigger Delay Time: 100ms) Note) Source and Measurement Trigger Count should be the same number as Sweep Points. Now you’ve configured 100 ms measurement trigger delay time. Agilent B2902/12A Precision Source/Measure Unit Viewing the list of measurement results The measurement results including the measurement time stamp can be referred by the following steps. Step 1. Step 2. If you aren’t on the top of the Function menu, press repeatedly to return to the top level. On the top level of the Function menu Any tree structure can’t be seen In the middle level of the Function menu Some tree structure can be seen If you’d like to see the list of measurement results, press , then press Measure Result dialogue. (1) Press Result (2) Press Measure Step 3. Rotate and press (1) Rotate field Step 4. to select Channel field. Then press to select Channel Rotate and press (1) Rotate and press Data field (2) Press Ch 2 to select Data field. Then rotate to select to select Channel 2. to scroll the data list. (2) Rotate or press Assist key to scroll the list to open