35203A - Sensors

35203A
Platinum Series
User Configurable ±1g to ±5g
Triaxial - Digital Accelerometer
0.494
[12,55]
0.167 0.354
[4,25] [9,00]
0.354
[9,00]
0.354
[9,00]
0.354
[9,00]
0.354
[9,00]
0.354
[9,00]
3 mm x 0,5 mm Thread
0.138 [3,50] Deep
16 Places on 2 Views
0.354
[9,00]
0.575
[14,60]
X
1
1.437
3 [36,50]
2
0.689
[17,50]
2.049
[52,05]
Rear View
Bottom View
Spring and cable
added with
Option Tnnn
Accelerometer Connections
Dimensions in inches [mm]
Mounting Instructions
Use two 3,0 x 0,5 mm thread machine screws to attach to the
bottom or rear 35203A mount faces. Be careful not to bottom
screws in tapped holes as this can strip threads.
➤ A light film of oil or silicone grease can be used to enhance
contact between mounting surfaces.
➤ A locking compound should be used on all threads to prevent
screws loosening due to vibrations.
➤
INSTRUMENTS
Summit Instruments, Inc.
2 2 3 6 N . C l eve l a n d - M a s s i l l o n R d .
A k ro n , O h i o 4 4 3 3 3 - 1 2 5 5 U S A
( 3 3 0 ) 6 5 9 - 3 3 1 2 FA X : ( 3 3 0 ) 6 5 9 - 3 2 8 6
How to Use the 35203A
If a Summit Instruments 35250A PC Interface Adapter is
available, connect its sensor connector to the 35203A, apply
power to the 35250A, and connect a computer to the 35250A
RS232 connector. A positive acceleration (labeled on the
accelerometer cover) causes a positive change on the
corresponding output.
➤ If a 35250A is not available, connect a +8 to 30V DC power
source to +VS Connect power source ground to the Gnd pin.
Connect a computer RS485 interface to the two RS485 pins on
the 35203A.
➤ Optionally connect a voltmeter or other instrument to each
35203A Analog Output and Signal-. For best results use a
converter with an input impedance greater than 1MΩ.
➤ Keep cable lengths between an analog output and an instrument
as short as practical. Use shielded twisted pair wiring, especially
near electric motors or other noise sources.
➤ Download Instrument Configuration Utility (ICU) software from
www.summitinstruments.com and use it to configure the 35203A.
When transmitting High Speed Data to a PC, sample frequency
should be at least 10 times the filter bandwidth when linear
interpolation is used for signal reconstruction. Built-in low-pass
filtering can be used to lower the noise floor and eliminate
spurious peak measurements from short-duration shock
impulses with insignificant energy content.
➤
Specifications*
Accelerometer(s)
Maximum Full Scale Range
Sens. Drift 25°C to TMIN or TMAX
Zero G Drift 25°C to TMIN or TMAX
Alignment
Transverse Sensitivity
Nonlinearity
Upper Cutoff Frequency
Noise Density
Temperature Sensor
Range
Resolution
Accuracy
Digital Signal Processor
Internal Word Size
Sensor Scan Rate
Analog Outputs
Voltage Swing
Impedance to GND
Nonlinearity
Digital Output Word Size
IRIG-106 PCM Output Rate
Power Supply (VS)
Input voltage limits
Input voltage - operating
Input current
Rejection Ratio
Temperature Range (TA)
Shock survival
Powered
Unpowered
Mass
* Data subject to change without notice.
Min
Typical
±5
±7.5
±0.5
±0.2
±2.0
0.25
0.2
‡
1000
0.225
1
‡
-55
0.25
100
-20
+8
80
-40
-500
-1000
6 7 8 9
Terminal Block
2
0.187
[4,76]
SUMMIT
1 2 3 4 5
0.240
[6,10]
3
1
1 2 3 4 5 6 7 8 9
0.25
±2
130
50
120
85
†Scale linearly with range
Pin
1
2
3
4
5
6
7
8
9
Description
Cable Wire
Analog Output 1+
Analog Output 2+
Analog Output 3+
SignalRS485RS485+
Aux
+VS (Power)
Gnd
Brown
Red
Orange
Yellow
Green
Blue
Violet
Grey
White*
* connected to braided shield at D-Shell Pin
Three - Year Limited Warranty
Max
Units
10,000
0.325
g
%
g
degrees
%
% FSR
Hz
mg/√Hz
‡
±3
°C
°C
°C
32
42,500
bits
Hz
4.75
220
0.15
16
3
V
Ω
% FSR
bits
Mbit/sec
+60
+30
V
V
mA
dB
°C
+125
85
+500
+1000
g
g
g
‡ Low-pass 3 dB cutoff
Summit Instruments, Inc. (SI) warrants this product against
defects in materials and workmanship for a period of three years
from date of shipment. During the warranty period, SI will, at its
option, either repair or replace products which prove to be
defective.
Exclusions
This warranty does not apply to defects or damage resulting from:
• Improper installation or maintenance.
• Customer supplied software or interfacing.
• Unauthorized modification or misuse.
• Operation outside of the product specifications.
Limitations
The warranty set forth above is exclusive and no other warranty,
whether written or oral, is expressed or implied. Summit
Instruments specifically disclaims the implied warranties of
merchantability and fitness for a particular purpose.
Some jurisdictions do not allow limitations on how long an
implied warranty lasts, so the above limitation or exclusion may
not apply to you. However, any implied warranty of
merchantability or fitness is limited to the one-year duration of
this written warranty.
Warranty Service
To obtain service during the warranty period products must be
returned, transportation prepaid, to SI.
25203A-D001 Rev. 06/05