Datasheet

100 kV
X-RAY MICROFOCUS X-RAY SOURCE
SOURCE L10321
TLSOF0185
FEATURES
APPLICATIONS
●Wide X-ray beam angle: 118 °
●Short FOD (Focus to Object Distance): 7.3 mm
●Non-destructive inspection
●X-ray CT
●Focal spot size: 5 µm (at 4 W)
[Applicable objects]
X-ray image with higher magnification ratio
The focal spot of 5 µm of the sealed type X-ray tube offers sharp
and clear X-ray images even at a high magnification.
●No high voltage cable connection required
High voltage power supply is integrated.
•Electronic component
•Printed circuit board
•Plastic component
•Metal component
SPECIFICATIONS
GENERAL
Parameter
X-ray tube voltage setting range
X-ray tube current setting range
X-ray tube voltage operational range 1
X-ray tube current operational range 1
Maximum output
X-ray focal spot size (Nominal value) 2
X-ray output window material / Thickness
X-ray beam angle 3
Focus to object distance (FOD)
Target material
Weight 4
Communication method
Description / Value
0 to 100
0 to 200
40 to 100
10 to 200
20
5 to 30
Beryllium/0.5
Approx. 118 / 105
Approx. 7.3
Tungsten
Approx. 9.5
Interface: RS-232C (9-pin D-sub connector)
Unit
kV
µA
kV
µA
W
µm
mm
degree
mm
—
kg
—
Description / Value
+24 (+2.4, -0)
Less than 120
Continuous rating
+10 to +40
0 to +50
20 to 85 (No condensation)
Unit
V
W
—
°C
°C
%
Description
EN 50581 Category 9
Emission limits: CISPR 11 Group 1 Class A
IEC/EN 61326-1
Immunity requirements: Table 2
Unit
—
Description
PC / AT compatible
Windows® XP, 7
RS-232C interface
Unit
—
—
—
RATINGS
Parameter
Input voltage (DC)
Power consumption
Rated output
Operating ambient temperature
Storage ambient temperature
Operating and storage humidity
REGULATION AND STANDARD
Parameter
RoHS directive
EMC
—
CONTROL SOFTWARE 5
Parameter
Applicable PC
Applicable OS
Interface
NOTE: 1See the graph of the X-ray tube Voltage / Current operation range.
2This focal spot size changes depending on the output.
3Reference value: With 50 % of maximum X-ray emission.
4This weight includes the accessories of approx. 0.25 kg.
5The control software is provided as a sample software for the purpose of MFX operation.
The performance of the software is not guaranteed.
!
PRECAUTIONS TO USE
· This microfocus X-ray source generates X-rays harmful to the human body. Use sufficient caution when handling the equipment to avoid
direct or inadvertent exposure to X-rays.
Install the X-ray source or the X-ray tube unit in an X-ray shielded cabinet or room equipped with safety interlock functions to prevent
accidental exposure to X-rays.
OPERATIONAL CAUTION
· This microfocus X-ray source generates X-rays and must therefore be used only under the supervision of qualified personnel.
· This microfocus X-ray source shall be used in compliance with health and safety regulations enforced in order to prevent health hazards
problems due to ionizing radiation.
X-RAY TUBE VOLTAGE / CURRENT OPERATION RANGE
250
TXPRB0027EB
TUBE CURRENT (µA)
200
150
100
50
10
40
* The X-ray tube voltage guaranteed range is 40 kV to 100 kV.
* Operation is not guaranteed when the tube current is below 10 µA.
50
60
70
80
90
100
TUBE VOLTAGE (kV)
DIMENSIONAL OUTLINE (Unit: mm)
4 × M6 DEPTH 8
(MOUNTING
SCREW HOLES)
147.0 ± 0.5
86 ± 1
X-RAY FOCAL
SPOT
78 ± 1
FOD
7.3
147.0 ± 0.5
50
° 55
100
4.5
33
X-RAY BEAM ANGLE
118°
°
INTERLOCK-2
289 ± 1
147.0 ± 0.5
INTERLOCK-1
195 ± 1
COOLING FAN
23
X-RAY FOCAL
SPOT
4 × M2 DEPTH 3
POWER DC 24 V
RS-232C
12
PROTECTIVE EARTH TERMINAL
85 MIN.
(REFERENCE)*
167 ± 1
4 × M6 DEPTH 8
(MOUNTING
SCREW HOLES)
172 ± 1
4 × M6 DEPTH 8
(MOUNTING
SCREW HOLES)
9.5
147.0 ± 0.5
147.0 ± 0.5
147.0 ± 0.5
,
* Connection cables aren t included.
TXPRA0027EA
RELATED PRODUCTS
X-RAY IMAGE INTENSIFIER DIGITAL CAMERA UNIT C7336-05/-52
The C7336 series consist of a high resolution, high contrast 4-inch X-ray image intensifier (X-ray I.I.) and a 2.8 megapixel CMOS image sensor.
The X-ray I.I. used has a fixed field-of-view of 100 mm diameter or a 4 inches/2 inches adjustable field-of-view and an input window made of thin aluminum which is
excellent in X-ray transmission and causes less scattering of X-rays. These features
allow real-time detection at X-ray energy levels from about 20 keV.
The captured images can be transferred to PC directly by interface of IEEE1394b.
X-CUBE™ (COMPACT X-RAY CCD CAMERA) H8480, H8481, H8953
TXPRF0020
X-CUBEs are compact X-ray CCD camera designed for non-destructive inspection,
which make X-ray imaging as easy as an ordinary CCD camera in handling. The
H8480 and H8953 use a 2/3 type CCD coupled to large-diameter tapered FOPs
which are coated with CsI. The H8481 uses a straight type FOP instead of the large
FOP, achieving a high resolution of 20 Lp/mm.
Left: H8480 Center: H8953 Right: H8481
FOS (Fiber optic plate coated with X-ray scintillator)
The FOS is an optical device for X-ray imaging, fabricated by coating an X-ray scintillator material over a fiber optic plate consisting
of more than tens of million glass fibers each a few micrometers in diameter. The FOS provides higher sensitivity and resolution
than currently used sensitized paper films and also allows real-time digital radiography when directly coupled to a commercially
available CCD. The fiber optic plate used in the FOS has excellent X-ray absorption characteristics, so that X-rays penetrating the
X-ray scintillator and directly entering the CCD are minimized to less than 1 %. This protects the CCD from the deterioration and increased noise caused by X-ray irradiation, assuring a long service life and maintaining high image quality.
Various sizes and shapes of FOS are available to meet your particular needs, including tapered FOP types.
■Structure
TMCPF0072
Protection Film
Scintillator : CsI(Tl)
FOP
(Fiber optic plate)
2D Image-Device
Windows ® is a registered trademark of Microsoft Corporation in the United States and/or other countries.
Other product and software names mentioned herein may be either registered trademarks or trademarks of their respective owners.
Subject to local technical requirements and regulations, availability of products included in this promotional material may vary. Please consult with our sales office.
Information furnished by HAMAMATSU is believed to be reliable. However, no responsibility is assumed for possible inaccuracies or omissions. Specifications are
subject to change without notice. No patent rights are granted to any of the circuits described herein. ©2015 Hamamatsu Photonics K.K.
HAMAMATSU PHOTONICS K.K.
www.hamamatsu.com
HAMAMATSU PHOTONICS K.K., Electron Tube Division
314-5, Shimokanzo, Iwata City, Shizuoka Pref., 438-0193, Japan, Telephone: (81)539/62-5248, Fax: (81)539/62-2205
U.S.A.: Hamamatsu Corporation: 360 Foothill Road, Bridgewater. N.J. 08807-0910, U.S.A., Telephone: (1)908-231-0960, Fax: (1)908-231-1218 E-mail: [email protected]
Germany: Hamamatsu Photonics Deutschland GmbH: Arzbergerstr. 10, D-82211 Herrsching am Ammersee, Germany, Telephone: (49)8152-375-0, Fax: (49)8152-2658 E-mail: [email protected]
France: Hamamatsu Photonics France S.A.R.L.: 19, Rue du Saule Trapu, Parc du Moulin de Massy, 91882 Massy Cedex, France, Telephone: (33)1 69 53 71 00, Fax: (33)1 69 53 71 10 E-mail: [email protected]
United Kingdom: Hamamatsu Photonics UK Limited: 2 Howard Court, 10 Tewin Road, Welwyn Garden City, Hertfordshire AL7 1BW, United Kingdom, Telephone: (44)1707-294888, Fax: (44)1707-325777 E-mail: [email protected]
North Europe: Hamamatsu Photonics Norden AB: Torshamnsgatan 35 SE-164 40 Kista, Sweden, Telephone: (46)8-509-031-00, Fax: (46)8-509-031-01 E-mail: [email protected]
TXPR1013E02
Italy: Hamamatsu Photonics Italia S.r.l.: Strada della Moia, 1 int. 6, 20020 Arese (Milano), Italy, Telephone: (39)02-93581733, Fax: (39)02-93581741 E-mail: [email protected]
NOV. 2015 IP
China: Hamamatsu Photonics (China) Co., Ltd.: B1201 Jiaming Center, No.27 Dongsanhuan Beilu, Chaoyang District, Beijing 100020, China, Telephone: (86)10-6586-6006, Fax: (86)10-6586-2866 E-mail: [email protected]