Teccor® brand Thyristors Thyristor and Rectifier Testing Using Curve Tracers Introduction Limitations, Accuracy, and Correlation One of the most useful and versatile instruments for testing semiconductor devices is the curve tracer (CT). Tektronix is the best known manufacturer of curve tracers and produces four basic models: 575, 576, 577 and 370. These instruments are specially adapted CRT display screens with associated electronics such as power supplies, amplifiers, and variable input and output functions that allow the user to display the operating characteristics of a device in an easy-to-read, standard graph form. Operation of Tektronix CTs is simple and straightforward and easily taught to non-technical personnel. Although widely used by semiconductor manufacturers for design and analytical work, the device consumer will find many uses for the curve tracer, such as incoming quality control, failure analysis, and supplier comparison. Curve tracers may be easily adapted for go-no go production testing. Tektronix also supplies optional accessories for specific applications along with other useful hardware. Although the curve tracer is a highly versatile device, it is not capable of every test that one may wish to perform on TFNJDPOEVDUPSEFWJDFTTVDIBTEWEUTFDPOEBSZSFWFSTF breakdown, switching speeds, and others. Also, tests at WFSZIJHIDVSSFOUTBOEPSWPMUBHFTBSFEJGmDVMUUPDPOEVDU accurately and without damaging the devices. A special high-current test fixture available from Tektronix can extend operation to 200 A pulsed peak. Kelvin contacts available on the 576 and 577 eliminate inaccuracy in voltage measured at high current (VTM) by sensing voltage drop due to contact resistance and subtracting from the reading. Tektronix Equipment Although Tektronix no longer produces curve tracer model 575, many of the units are still operating in the field, and it is still an extremely useful instrument. The 576, 577 and 370 are current curve tracer models and are more streamlined in their appearance and operation. The 577 is a less elaborate version of the 576, yet retains all necessary test functions. The following basic functions are common to all curve tracers: t 1PXFSTVQQMZsupplies positive DC voltage, negative DC voltage, or AC voltage to bias the device. Available power is varied by limiting resistors. t 4UFQHFOFSBUPS supplies current or voltage in precise steps to control the electrode of the device. The number, polarity, and frequency of steps are selectable. t )PSJ[POUBMBNQMJmFS displays power supply voltage as applied to the device. Scale calibration is selectable. t 7FSUJDBMBNQMJmFS displays current drawn from the supply by the device. Scale calibration is selectable. Curve tracer controls for beam position, calibration, pulse operation, and other functions vary from model to model. The basic theory of operation is that for each curve one terminal is driven with a constant voltage or current and the other one is swept with a half sinewave of voltage. The driving voltage is stepped through several values, and a different trace is drawn on each sweep to generate a family of curves. Accuracy of the unit is within the published manufacturer’s specification. Allow the curve tracer to warm up and stabilize before testing begins. Always expand the horizontal or vertical scale as far as possible to increase the resolution. Be judicious in recording data from the screen, as the trace line width and scale resolution factor somewhat limit the accuracy of what may be read. Regular calibration checks of the instrument are recommended. Some users keep a selection of calibrated devices on hand to verify instrument operation when in doubt. Re-calibration or adjustment should be performed only by qualified personnel. Often discrepancies exist between measurements taken on different types of instrument. In particular, most semiconductor manufacturers use high-speed, computerized test equipment to test devices. They test using very short pulses. If a borderline unit is then measured on a curve tracer, it may appear to be out of specification. The most common culprit here is heat. When a semiconductor device increases in temperature due to current flow, certain characteristics may change, notably gate characteristics on SCRs, gain on transistors, leakage, and so on. It is very difficult to operate the curve tracer in such a way as to eliminate the heating effect. Pulsed or single-trace operation helps reduce this problem, but care should be taken in comparing curve tracer measurements to computer tests. Other factors such as stray capacitances, impedance matching, noise, and device oscillation also may create differences. Safety (Cautions and Warnings) Adhere rigidly to Tektronix safety rules supplied with each curve tracer. No attempt should be made to defeat any of the safety interlocks on the device as the curve tracer can produce a lethal shock. Also, older 575 models do not have the safety interlocks as do the new models. Take care never to touch any device or open the terminal while energized. WARNING: Devices on the curve tracer may be easily damaged from electrical overstress. ©2008 Littelfuse, Inc. Specifications are subject to change without notice. Please refer to http://www.littelfuse.com for current information. Thyristor and Rectifier Testing Using Curve Tracers AN1006 AN1006 Teccor® brand Thyristors AN1006 Follow these rules to avoid destroying devices: t 'BNJMJBSJ[FZPVSTFMGXJUIUIFFYQFDUFENBYJNVNMJNJUT of the device. t -JNJUUIFDVSSFOUXJUIUIFWBSJBCMFSFTJTUPSUPUIF minimum necessary to conduct the test. t *ODSFBTFQPXFSTMPXMZUPUIFTQFDJmFEMJNJU t 8BUDIGPSEFWJDFiSVOBXBZwEVFUPIFBUJOH t "QQMZBOEJODSFBTFHBUFPSCBTFESJWFTMPXMZBOEJO small steps. t $POEVDUUFTUTJOUIFNJOJNVNUJNFSFRVJSFE General Test Procedures Read all manuals before operating a curve tracer. Perform the following manufacturer’s equipment check: 1. Turn on and warm up curve tracer, but turn off, or down, all power supplies. Model 576 Curve Tracer Procedures The following test procedures are written for use with the model 576 curve tracer. (Figure AN1006.1) See “Model 370 Curve Tracer Procedure Notes” on page AN1006-16 and “Model 577 Curve Tracer Procedure Notes” on page AN1006-18 for setting adjustments required when using model 370 and 577 curve tracers. The standard 575 model lacks AC mode, voltage greater than 200 V, pulse operations, DC mode, and step offset controls. The 575 MOD122C does allow voltage up to 400 V, including 1500 V in an AC mode. Remember that at the time of design, the 575 was built to test only transistors and diodes. Some ingenuity, experience, and external hardware may be required to test other types of devices. For further information or assistance in device testing on Tektronix curve tracers, contact the Littelfuse Applications Engineering group. 2. Correctly identify terminals of the device to be tested. Refer to the manufacturer’s guide if necessary. 3. Insert the device into the test fixture, matching the device and test terminals. 3FNPWFIBOETGSPNUIFEFWJDFBOEPSDMPTFJOUFSMPDL cover. "QQMZSFRVJSFECJBTBOEPSESJWF 6. Record results as required. 7. Disconnect all power to the device before removing. Thyristor and Rectifier Testing Using Curve Tracers ©2008 Littelfuse, Inc. Specifications are subject to change without notice. Please refer to http://www.littelfuse.com for current information. Teccor® brand Thyristors TYPE 576 TEKTRONIX, INC. CURVE TRACER AN1006 AN1006 VERTICAL PORTLAND, ORE, U.S.A. PER V E R T DIV PER DISPLAY OFFSET H O R I Z DIV CRT PER S T E P ()k DIV 9m PER DIV HORIZONTAL HORIZONTAL VOLTAGE CONTROL Note: All Voltage Settings Will Be Referenced to "Collector" COLLECTOR SUPPLY STEP GENERATOR VARIABLE COLLECTOR SUPPLY VOLTAGE RANGE AMPLITUDE MAX PEAK POWER (POWER DISSIPATION) STEP/OFFSET AMPLITUDE (AMPS/VOLTS) OFFSET STEP/OFFSET POLARITY STEP FAMILY RATE TERMINAL JACKS C C B B E E TERMINAL SELECTOR MT2/ANODE VARIABLE COLLECTOR SUPPLY VOLTAGE GATE/TRIGGER LEFT-RIGHT SELECTOR FOR TERMINAL JACKS MT1/CATHODE KELVIN TERMINALS USED WHEN MEASURING VTM OR VFM Figure AN1006.1 Tektronix Model 576 Curve Tracer ©2008 Littelfuse, Inc. Specifications are subject to change without notice. Please refer to http://www.littelfuse.com for current information. Thyristor and Rectifier Testing Using Curve Tracers Teccor® brand Thyristors AN1006 Power Rectifiers The rectifier is a unidirectional device which conducts when forward voltage (above 0.7 V) is applied. To connect the rectifier: Kelvin fixture is not used, an error in measurement of VFM will result due to voltage drop in fixture. If a Kelvin fixture is not available, Figure AN1006.3 shows necessary information to wire a test fixture with Kelvin connections. t %VFUPUIFDVSSFOUMJNJUBUJPOTPGTUBOEBSEDVSWFUSBDFS model 576, VFM cannot be tested at rated current without a Tektronix model 176 high-current module. The procedure below is done at IT(RMS) = 10 A (20 APK). This test parameter allows the use of a standard curve tracer and still provides an estimate of whether VFM is within specification. 1. Connect Anode to Collector Terminal (C). 2. Connect Cathode to Emitter Terminal (E). To begin testing, perform the following procedures. Procedure 1:VRRM and IRM To measure the VRRM and IRM parameter: SOCKET 1. Set Variable Collector Supply Voltage Range to 1500 V.(2000 V on 370) 2. Set )PSJ[POUBM knob to sufficient scale to allow viewing of trace at the required voltage level (100 V/DIV for 400 V and 600 V devices and 50 V/DIV for 200 V devices). 3. Set Mode to Leakage. 4. Set Vertical knob to 100 μA/DIV. (Due to leakage setting, the CRT readout will be 100 nA per division.) SOCKET PINS One set of pins wired to Collector (C), Base (B), and Emitter (E) Terminals 5. Set Terminal Selector to Emitter Grounded-Open Base. Socket used must have two sets of pins 6. Set Polarity to (-). 7. Set Power Dissipation to 2.2 W. (2 W on 370) The pins which correspond to the anode and cathode of the device are wired to the terminals marked CSENSE (MT2/Anode) and E SENSE (MT1/Cathode). The gate does not require a Kelvin connection. 8. Set Left-Right Terminal Jack Selector to correspond with location of test fixture. 9. Increase Variable Collector Supply Voltage to the rated VRRM of the device and observe the dot on the CRT. Read across horizontally from the dot to the vertical current scale. This measured value is the leakage current. (Figure AN1006.2) PER V E R T IRM 100 nA DIV H O R I Z DIV PER S T E P ()k DIV 9m PER DIV Figure AN1006.2 IRM = 340 nA at VRRM = 600 V Procedure 2:VFM Before testing, note the following: t ",FMWJOUFTUmYUVSFJTSFRVJSFEGPSUIJTUFTU*GB Thyristor and Rectifier Testing Using Curve Tracers Instructions for Wiring Kelvin Socket To measure the VFM parameter: 1. Set Variable Collector Supply Voltage Range to 15 Max Peak Volts. (16 V on 370) 2. Set )PSJ[POUBM knob to 0.5 V/DIV. 3. Set Mode to Norm. PER VRRM Figure AN1006.3 100 V 4. Set Vertical knob to 2 A/DIV. 5. Set Power Dissipation to 220 W (100 W on 577). 6. Set Polarity to (+). 7. Set Left-Right Terminal Jack Selector to correspond with location of test fixture. 8. Increase Variable Collector Supply Voltage until current reaches 20 A. WARNING: Limit test time to 15 seconds maximum. To measure VFM, follow along horizontal scale to the point where the trace crosses the 20 A axis. The distance from the left-hand side of scale to the crossing point is the VFM value. (Figure AN1006.4) Note: Model 370 current is limited to 10 A. ©2008 Littelfuse, Inc. Specifications are subject to change without notice. Please refer to http://www.littelfuse.com for current information. Teccor® brand Thyristors Note: The CRT screen readout should show 1% of the maximum leakage current if the vertical scale is divided by 1,000 when leakage current mode is used. PER V E R T VFM 2 A Procedure 2:VDRM / IDRM DIV PER H O R I Z 500 mV DIV IT PER S T E P ()k DIV 9m PER DIV Figure AN1006.4 VFM = 1 V at IPK = 20 A To measure the VDRM and IDRM parameter: 1. Set Left-Right Terminal Jack Selector to correspond with location of test fixture. 2. Set Variable Collector Supply Voltage to the rated VDRM of the device and observe the dot on CRT. Read across horizontally from the dot to the vertical current scale. This measured value is the leakage current. (Figure AN1006.5) WARNING: Do NOT exceed VDRM/VRRM rating of SCRs, Triacs, or Quadracs. These devices can be damaged. SCRs PER SCRs are half-wave unidirectional rectifiers turned on when current is supplied to the gate terminal. If the current supplied to the gate is to be in the range of 12 μA and 500 μA, then a sensitive SCR is required; if the gate current is between 1 mA and 50 mA, then a non-sensitive SCR is required. V E R T PER H O R I Z VDRM PER S T E P IDRM 1. Connect Anode to Collector Terminal (C). 2. Connect Cathode to Emitter Terminal (E). Note: When sensitive SCRs are being tested, a 1 kΩ resistor must be connected between the gate and the cathode, except when testing IGT. Figure AN1006.5 To begin testing, perform the following procedures. Procedure 3: VRRM / IRRM Procedure 1:VDRM / VRRM / IDRM / IRRM To measure the VRRM and IRRM parameter: To measure the VDRM7RRM*DRM, and IRRM parameter: 1. Set Polarity to (-). 2. Set )PSJ[POUBM knob to sufficient scale to allow viewing of trace at the required voltage level. (The 100 V/DIV scale should be used for testing devices having a VDRM value of 600 V or greater; the 50 V/DIV scale for testing parts rated from 300 V to 500 V, and so on.) 100 V DIV To connect the rectifier: 1. Set Variable Collector Supply Voltage Range to appropriate Max Peak Volts for device under test. (Value selected should be equal to or greater than the device’s VDRM rating.) 100 nA DIV ()k DIV 9m PER DIV IDRM = 350 nA at VDRM = 600 V 2. Repeat Steps 1 and 2 (VDRM, IDRM) except substitute VRRM value for VDRM. (Figure AN1006.6) PER V E R T IRRM 100 nA DIV PER VRRM H O R I Z 100 V DIV 3. Set Mode to Leakage. PER S T E P 4. Set Polarity to (+). 5. Set Power Dissipation to 0.5 W.(0.4 W on 370) ()k DIV 9m PER DIV 6. Set Terminal Selector to Emitter Grounded-Open Base. 7. Set Vertical knob to approximately ten times the maximum leakage current (IDRM, IRRM) specified for the device. (For sensitive SCRs, set to 50 μA.) ©2008 Littelfuse, Inc. Specifications are subject to change without notice. Please refer to http://www.littelfuse.com for current information. Figure AN1006.6 IRRM = 340 nA at VRRM = 600 V Thyristor and Rectifier Testing Using Curve Tracers AN1006 AN1006 Teccor® brand Thyristors AN1006 Procedure 4: VTM To measure the VTM parameter: 1. Set Terminal Selector to Step Generator-Emitter Grounded. 2. Set Polarity to (+). PER V E R T 2 A DIV PER VTM 3. Set Step/Offset Amplitude to twice the maximum IGT rating of the device (to ensure the device turns on). For sensitive SCRs, set to 2 mA. H O R I Z 500 mV DIV PER S T E P IPK 4. Set Max Peak Volts to 15 V. (16 V on 370) ()k DIV 9m PER DIV 5. Set Offset by depressing 0 (zero). 6. Set Rate by depressing Norm. 100 mA 20 7. Set Step Family by depressing Rep (repetitive). 8. Set Mode to DC. Figure AN1006.7 9. Set )PSJ[POUBM knob to 0.5 V/DIV. Procedure 5: IH 10. Set Power Dissipation to 220 W (100 W on 577). To measure the IH parameter: 11. Set Number of Steps to 1. (Set steps to 0 (zero) on 370.) 12. Set Vertical knob to a sufficient setting to allow the viewing of 2 times the IT(RMS) rating of the device (IT(peak)) on CRT. Before continuing with testing, note the following: (1) Due to the excessive amount of power that can be generated in this test, only parts with an IT(RMS) rating of 6 A or less should be tested on standard curve tracer. If testing devices above 6 A, a Tektronix model 176 high-current module is required. (2) A Kelvin test fixture is required for this test. If a Kelvin fixture is not used, an error in measurement of VTM will result due to voltage drop in the fixture. If a Kelvin fixture is not available, Figure AN1006.3 shows necessary information to wire a test fixture with Kelvin connectors. 13. Set Left-Right Terminal Jack Selector to correspond with the location of the test fixture. 14. Increase Variable Collector Supply Voltage until current reaches rated IT(peak), which is twice the IT(RMS) rating of theSCR under test. Note: Model 370 current is limited to 10 A. WARNING: Limit test time to 15 seconds maximum after the Variable Collector Supply has been set to IT(peak), After the Variable Collector Supply Voltage has been set to IT(peak), the test time can automatically be shortened by changing Step Family from repetitive to single by depressing the Single button. To measure VTM, follow along horizontal scale to the point where the trace crosses the IT(peak) value. The distance from the left-hand side of scale to the intersection point is the VTM value. (Figure AN1006.7) Thyristor and Rectifier Testing Using Curve Tracers VTM = 1.15 V at IT(peak) = 12 A 1. Set Polarity to (+). 2. Set Power Dissipation to 2.2 W. (2 W on 370) 3. Set Max Peak Volts to 75 V. (80 V on 370) 4. Set Mode to DC. 5. Set )PSJ[POUBM knob to Step Generator. 6. Set Vertical knob to approximately 10 percent of the maximum IH specified. Note: Due to large variation of holding current values, the scale may have to be adjusted to observe holding current. 7. Set Number of Steps to 1. 8. Set Offset by depressing 0(zero). (Press Aid and Oppose at the same time on 370.) 9. Set Step/Offset Amplitude to twice the maximum IGT of the device. 10. Set Terminal Selector to Step Generator-Emitter Grounded. 11. Set Step Family by depressing Single. 12. Set Left-Right Terminal Jack Selector to correspond with location of test fixture. 13. Increase Variable Collector Supply Voltage to maximum position (100). 14. Set Step Family by depressing Single. (This could possibly cause the dot on CRT to disappear, depending on the vertical scale selected.) 15. Change Terminal Selector from Step GeneratorEmitter Grounded to Open Base-Emitter Grounded. 16. Decrease Variable Collector Supply Voltage to the point where the line on the CRT changes to a dot. The position of the beginning point of the line, just before the line becomes a dot, represents the holding current value. (Figure AN1006.8) ©2008 Littelfuse, Inc. Specifications are subject to change without notice. Please refer to http://www.littelfuse.com for current information. Teccor® brand Thyristors PER V E R T PER DIV DIV PER PER H O R I Z H O R I Z DIV DIV PER PER S T E P IH Figure AN1006.8 V E R T 500 A 10 A S T E P IGT ()k DIV 9m PER DIV IH = 1.2 mA 50 mA ()k DIV 9m PER DIV Figure AN1006.9 5 K IGT = 25 μA Procedure 6: IGT and VGT Procedure 8: VGT To measure the IGT and VGT parameter: To measure the VGT parameter: 1. Set Polarity to (+). 1. Set Offset Multiplier to 0 (zero). (Press Aid and Oppose at the same time on 370.) 2. Set Number of Steps to 1. 3. Set Offset by depressing Aid. 2. Set Step Offset Amplitude to 20% rated VGT. 4. Set Offset Multiplier to 0 (zero). (Press Aid and Oppose at the same time on 370.) 3. Set Left-Right Terminal Jack Selector to correspond with location of test fixture. 5. Set Terminal Selector to Step Generator-Emitter Grounded. 4. Gradually increase Offset Multiplier until device reaches the conduction point. (Figure AN1006.10) Measure VGT by following horizontal axis to the point where the vertical line crosses axis. This measured value is VGT. (On 370, VGT will be numerically displayed on screen, under offset value.) 6. Set Mode to Norm. 7. Set Max Peak Volts to 15 V. (16 V on 370) 8. Set Power Dissipation to 2.2 W. (2 W on 370) For sensitive SCRs, set at 0.5 W. (0.4 W on 370) 9. Set )PSJ[POUBM knob to 2 V/DIV. PER 10. Set Vertical knob to 50 mA/DIV. V E R T 11. Increase Variable Collector Supply Voltage until voltage reaches 12 V on CRT. 50 mA DIV PER H O R I Z 12. After 12 V setting is completed, change )PSJ[POUBM knob to Step Generator. DIV PER Procedure 7: IGT VGT To measure the IGT parameter: 1. Set Step/Offset Amplitude to 20% of maximum rated IGT. Note: RGK should be removed when testing IGT 2. Set Left-Right Terminal Jack Selector to correspond with location of the test fixture. Figure AN1006.10 S T E P 200 mV ()k DIV 9m PER DIV 250m VGT = 580 mV 3. Gradually increase Offset Multiplier until device reaches the conduction point. (Figure AN1006.9) Measure IGT by following horizontal axis to the point where the vertical line crosses axis. This measured value is IGT. (On 370, IGT will be numerically displayed on screen under offset value.) ©2008 Littelfuse, Inc. Specifications are subject to change without notice. Please refer to http://www.littelfuse.com for current information. Thyristor and Rectifier Testing Using Curve Tracers AN1006 AN1006 Teccor® brand Thyristors AN1006 Triacs current. (Figure AN1006.11) Triacs are full-wave bidirectional AC switches turned on when current is supplied to the gate terminal of the device. If gate control in all four quadrants is required, then a sensitive gate Triac is needed, whereas a standard Triac can be used if gate control is only required in Quadrants I through III. PER V E R T PER H O R I Z 100 V DIV To connect the Triac: 1. Connect the Gate to the Base Terminal (B). 50 nA DIV PER S T E P VDRM IDRM 2. Connect MT1 to the Emitter Terminal (E). 3. Connect MT2 to the Collector Terminal (C). ()k DIV 9m PER DIV To begin testing, perform the following procedures. Figure AN1006.11 Procedure 1: (+)VDRM, (+)IDRM, (-)VDRM, (-)IDRM Note: The (+) and (-) symbols are used to designate the polarity MT2 with reference to MT1. To measure the (+)VDRM, (+)IDRM, (-)VDRM, and (-)IDRM parameter: 1. Set Variable Collector Supply Voltage Range to appropriate Max Peak Volts for device under test. (Value selected should be equal to the device’s VDRM rating.) WARNING: DO NOT exceed VDRM/VRRM rating of SCRs, Triacs, or Quadracs. These devices can be damaged. 2. Set )PSJ[POUBM knob to sufficient scale to allow viewing of trace at the required voltage level. (The 100 V/DIV scale should be used for testing devices having a VDRM rating of 600 V or greater; the 50 V/DIV scale for testing parts rated from 30 V to 500 V, and so on.) 3. Set Mode to Leakage. 4. Set Polarity to (+). 5. Set Power Dissipation to 0.5 W. (0.4 W on 370) 6. Set Terminal Selector to Emitter Grounded-Open Base. 7. Set Vertical knob to ten times the maximum leakage current (IDRM) specified for the device. Note: The CRT screen readout should show 1% of the maximum leakage current. The vertical scale is divided by 1,000 when leakage mode is used. (+)IDRM = 205 nA at (+)VDRM = 600 V Procedure 3: (-)VDRM, (-)IDRM To measure the (-)VDRM and (-)IDRM parameter: 1. Set Polarity to (-). 2. Repeat Procedures 1 and 2. (Read measurements from upper right corner of the screen.) Procedure 4: VTM (Forward and Reverse) To measure the VTM (Forward and Reverse) parameter: 1. Set Terminal Selector to Step Generator-Emitter Grounded. 2. Set Step/Offset Amplitude to twice the maximum IGT rating of the device (to insure the device turns on). 3. Set Variable Collector Supply Voltage Range to 15 V Max Peak volts. (16 V on 370) 4. Set Offset by depressing 0 (zero). 5. Set Rate by depressing Norm. 6. Set Step Family by depressing Rep (Repetitive). 7. Set Mode to Norm. 8. Set )PSJ[POUBM knob to 0.5 V/DIV. 9. Set Power Dissipation to 220 W (100 W on 577). 10. Set Number of Steps to 1. 11. Set Step/Offset Polarity to non-inverted (button extended; on 577 button depressed). Procedure 2: (+)VDRM, (+)IDRM 12. Set Vertical knob to a sufficient setting to allow the viewing of 1.4 times the IT(RMS) rating of the device [IT(peak) on CRT]. To measure the (+)VDRM and (+)IDRM parameter: Note the following: 1. Set Left-Right Terminal Jack Selector to correspond with location of the test fixture. t %VFUPUIFFYDFTTJWFBNPVOUPGQPXFSUIBUDBOCF generated in this test, only parts with an IT(RMS) rating of 8 A or less should be tested on standard curve tracer. If testing devices above 8 A, a Tektronix model 176 highcurrent module is required. 2. Increase Variable Collector Supply Voltage to the rated VDRM of the device and observe the dot on the CRT. Read across horizontally from the dot to the vertical current scale. This measured value is the leakage Thyristor and Rectifier Testing Using Curve Tracers t ",FMWJOUFTUmYUVSFJTSFRVJSFEGPSUIJTUFTU*GB ©2008 Littelfuse, Inc. Specifications are subject to change without notice. Please refer to http://www.littelfuse.com for current information. Teccor® brand Thyristors Kelvin fixture is not used, an error in measurement of VTM will result due to voltage drop in fixture. If a Kelvin fixture is not available, Figure AN1006.3 shows necessary information to wire a test fixture with Kelvin connections. Procedure 7: IH(Forward and Reverse) To measure the IH (Forward and Reverse) parameter: 1. Set Step/Offset Amplitude to twice the IGT rating of the device. 2. Set Power Dissipation to 10 W. Procedure 5: VTM (Forward) 3. Set Max Peak Volts to 75 V. (80 V on 370) To measure the VTM (Forward) parameter: 4. Set Mode to DC. 1. Set Polarity to (+). 5. Set )PSJ[POUBM knob to Step Generator. 2. Set Left-Right Terminal Jack Selector to correspond with location of test fixture. 6. Set Vertical knob to approximately 10% of the maximum IH specified. 3. Increase Variable Collector Supply Voltage until current reaches rated IT(peak), which is 1.4 times IT(RMS) rating of the Triac under test. Note: Model 370 current is limited to 10 A. Note: Due to large variation of holding current values, the scale may have to be adjusted to observe holding current. 7. Set Number of Steps to 1. WARNING: Limit test time to 15 seconds maximum. After the Variable Collector Supply Voltage has been set to IT(peak), the test time can automatically be set to a short test time by changing Step Family from repetitive to single by depressing the Single button. To measure VTM, follow along horizontal scale to the point where the trace crosses the IT(peak) value. The distance from the left-hand side of scale to the crossing point is the VTM value. (Figure AN1006.12) 8. Set Step/Offset Polarity to non-inverted (button extended, on 577 button depressed). 9. Set Offset by depressing 0 (zero). (Press Aid and Oppose at same time on 370.) 10. Set Terminal Selector to Step Generator-Emitter Grounded. Procedure 8: IH(Forward) To measure the IH (Forward) parameter: 1. Set Polarity to (+). 2. Set Left-Right Terminal Jack Selector to correspond with location of test fixture. PER V E R T 2 A DIV PER H O R I Z VTM 500 mV DIV PER IPK S T E P ()k DIV 9m PER DIV Figure AN1006.12 100 mA 20 3. Increase Variable Collector Supply Voltage to maximum position (100). 4. Set Step Family by depressing Single. This could possibly cause the dot on the CRT to disappear, depending on the vertical scale selected). 5. Decrease Variable Collector Supply Voltage to the point where the line on the CRT changes to a dot. The position of the beginning point of the line, just before the line becomes a dot, represents the holding current value. (Figure AN1006.13) VTM (forward) = 1.1 V at IPK = 11.3 A (8 A rms) PER V E R T Procedure 6: VTM (Reverse) DIV To measure the VTM (Reverse) parameter: PER H O R I Z 1. Set Polarity to (-). DIV 2. Set Left-Right Terminal Jack Selector to correspond with the location of the test fixture. 3. Increase Variable Collector Supply Voltage until current reaches rated IT(peak). 4. Measure VTM(Reverse) similar to Figure AN1006.12, except from upper right hand corner of screen. PER S T E P ©2008 Littelfuse, Inc. ()k DIV 9m PER DIV IH Figure AN1006.13 Specifications are subject to change without notice. Please refer to http://www.littelfuse.com for current information. 5 mA 50 mA 100m IH (Forward) = 8.2 mA Thyristor and Rectifier Testing Using Curve Tracers AN1006 AN1006 Teccor® brand Thyristors AN1006 Procedure 9: IH(Reverse) To measure the IH (Reverse) parameter: PER V E R T 1. Set Polarity to (-). 50 mA DIV 2. Repeat Procedure 7 measuring IH(Reverse). (Read measurements from upper right corner of the screen.) PER H O R I Z DIV Procedure 10: IGT PER S T E P To measure the IGT parameter: 1. Set Polarity to (+). IGT ()k DIV 9m PER DIV 2. Set Number of Stepsto 1. (Set number of steps to 0 (zero) on 370.) 3. Set Offset by depressing Aid. (On 577, also set Zero button to Offset. Button is extended.) 4. Set Offset Multiplier to 0 (zero). (Press Aid and Oppose at same time on 370.) 5. Set Terminal Selector to Step Generator-Emitter Grounded. Figure AN1006.14 5 mA 10 IGT in Quadrant I = 18.8 mA Procedure 12: IGT - Quadrant II [MT2 (+) Gate (-)] To measure the IGT - Quadrant II parameter: 6. Set Mode to Norm. 1. Set Step/Offside Polarity by depressing Invert (release button on 577). 7. Set Max Peak Volts to 15 V. (16 V on 370) 2. Set Polarity to (+). 8. Set Power Dissipation to 10 W. 9. Set Step Family by depressing Single. 3. Set observed dot to bottom right corner of CRT grid by turning the horizontal position knob. When Quadrant II testing is complete, return dot to original position. 10. Set )PSJ[POUBM knob to 2 V/DIV. 4. Repeat Procedure 11. 11. Set Vertical knob to 50 mA/DIV. 12. Set Step/Offset Polarity to non-inverted position (button extended, on 577 button depressed). Procedure 13: IGT - Quadrant III [MT2 (-) Gate (-)] To measure the IGT - Quadrant III parameter: 13. Set Variable Collector Supply Voltage until voltage reaches 12 V on CRT. 1. Set Polarity to (-). 14. After 12 V setting is completed, change )PSJ[POUBM knob to Step Generator. 2. Set Step/Offset Polarity to non-inverted position (button extended, on 577 button depressed). 3. Repeat Procedure 11. (Figure AN1006.15) Procedure 11: IGT - Quadrant I [MT2 (+) Gate (+)] To measure the IGT - Quadrant I parameter: PER 1. Set Step/Offset Amplitude to approximately 10% of rated IGT. IGT 2. Set Left-Right Terminal Jack Selector to correspond with location of test fixture. 50 mA DIV PER H O R I Z 3. Gradually increase Offset Multiplier until device reaches conduction point. (Figure AN1006.14) Measure IGT by following horizontal axis to the point where the vertical line passes through the axis. This measured value is IGT. (On 370, IGT is numerically displayed on screen under offset value.) DIV PER S T E P ()k DIV 9m PER DIV Figure AN1006.15 Thyristor and Rectifier Testing Using Curve Tracers V E R T 5 mA 10 IGT in Quadrant III = 27 mA ©2008 Littelfuse, Inc. Specifications are subject to change without notice. Please refer to http://www.littelfuse.com for current information. Teccor® brand Thyristors Procedure 14: IGT - Quadrant IV [MT2 (-) Gate (+)] PER V E R T To measure the IGT - Quadrant IV parameter: 50 mA DIV 1. Set Polarity to (-). PER 2. Set Step/Offset Polarity by depressing Invert (release button on 577). H O R I Z VGT DIV 3. Set observed dot to top left corner of CRT grid by turning the )PSJ[POUBM position knob. When Quadrant IV testing is complete, return dot to original position. PER 4. Repeat Procedure 11. ()k DIV 9m PER DIV Procedure 15: VGT To measure the VGT parameter: 1. Set Polarity to (+). 2. Set Number of Steps to 1. (Set steps to 0 (zero) on 370.) 500 mV S T E P Figure AN1006.16 100m VGT in Quadrant I = 780 mV Procedure 17: VGT - Quadrant II [MT2 (+) Gate (-)] To measure the VGT - Quadrant II parameter: 3. Set Offset by depressing Aid. (On 577, also set 0 (zero) button to Offset. Button is extended.) 1. Set Step/Offset Polarity by depressing Invert (release button on 577). 4. Set Offset Multiplier to 0 (zero). (Press Aid and Oppose at same time on 370.) 2. Set Polarity to (+). 5. Set Terminal Selector to Step Generator-Emitter Grounded. 3. Set observed dot to bottom right corner of CRT grid by turning the )PSJ[POUBM position knob. When Quadrant II testing is complete, return dot to original position. 6. Set Mode to Norm. 4. Repeat Procedure 16. 7. Set Max Peak Volts to 15 V. (16 V on 370) 8. Set Power Dissipation to 10 W. Procedure 18: VGT - Quadrant III [MT2 (-) Gate (-)] 9. Set Step Family by depressing Single. To measure the VGT - Quadrant III parameter: 10. Set )PSJ[POUBM knob to 2 V/DIV. 1. Set Polarity to (-). 11. Set Step/Offset Polarity to non-inverted position (button extended, on 577 button depressed). 2. Set Step/Offset Polarity to non-inverted position (button extended, on 577 button depressed). 12. Set Current Limit to 500 mA (not available on 577). 3. Repeat Procedure 16. (Figure AN1006.17) 13. Increase Variable Collector Supply Voltage until voltage reaches 12 V on CRT. PER V E R T 14. After 12 V setting is complete, change )PSJ[POUBM knob to Step Generator. VGT Procedure 16: VGT - Quadrant I [MT2 (+) Gate (+)] PER H O R I Z To measure the VGT - Quadrant I parameter: DIV PER 1. Set Step/Offset Amplitude to 20% of rated VGT. S T E P 2. Set Left-Right Terminal Jack Selector to correspond with location of test fixture. 3. Gradually increase Offset Multiplier until device reaches conduction point. (Figure AN1006.16) Measure VGT by following horizontal axis to the point where the vertical line passes through the axis. This measured value will be VGT. (On 370, VGT will be numerically displayed on screen under offset value.) 50 mA DIV ()k DIV 9m PER DIV Figure AN1006.17 500 mV 100m VGT in Quadrant III = 820 mV Procedure 19: VGT - Quadrant IV [MT2 (-) Gate (+)] To measure the VGT - Quadrant IV parameter: 1. Set Polarity to (-). 2. Set Step/Offset Polarity by depressing Invert (release button on 577). ©2008 Littelfuse, Inc. Specifications are subject to change without notice. Please refer to http://www.littelfuse.com for current information. Thyristor and Rectifier Testing Using Curve Tracers AN1006 AN1006 Teccor® brand Thyristors AN1006 3. Set observed dot to top left corner of CRT grid by turning the )PSJ[POUBM position knob. When testing is complete in Quadrant IV, return dot to original position. 4. Repeat Procedure 16. vertical current scale.) This measured value is the leakage current. (Figure AN1006.18) WARNING: Do NOT exceed VDRM/VRRM rating of SCRs, Triacs, or Quadracs. These devices can be damaged. Quadracs PER V E R T Quadracs are simply Triacs with an internally-mounted DIAC. As with Triacs, Quadracs are bidirectional AC switches which are gate controlled for either polarity of main terminal voltage. 50 nA DIV PER H O R I Z 50 V DIV To connect the Quadrac: 1. Connect Trigger to Base Terminal (B). 2. Connect MT1 to Emitter Terminal (E). PER S T E P VDRM IDRM 3. Connect MT2 to Collector Terminal (C). ()k DIV 9m PER DIV To begin testing, perform the following procedures. Figure AN1006.18 Procedure 1: (+)VDRM, (+)IDRM, (-)VDRM, (-)IDRM Note: The (+) and (-) symbols are used to designate the polarity of MT2 with reference to MT1. To measure the (+)VDRM, (+)IDRM, (-)VDRM, and (-)IDRM parameter: 1. Set Variable Collector Supply Voltage Range to appropriate Max Peak Volts for device under test. (Value selected should be equal to or greater than the device’s VDRM rating). (+)IDRM = 51 nA at (+)VDRM = 400 V Procedure 3: (-)VDRM and (-)IDRM To measure the (-)VDRM and (-)IDRM parameter: 1. Set Polarity to (-). 2. Repeat Procedures 1 and 2. (Read measurements from upper right corner of screen). Procedure 4: VBO, IBO,VBO (Quadrac Trigger DIAC or Discrete DIAC) 2. Set )PSJ[POUBM knob to sufficient scale to allow viewing of trace at the required voltage level. (The7%*7 scale should be used for testing devices having a VDRM rating of 600 V or greater; the 50 V/DIV scale for testing parts rated from 300 V to 500 V, and so on). 1. Connect MT1 to Emitter Terminal (E). 3. Set Mode to Leakage. 3. Connect Trigger Terminal to MT2 Terminal through a 10 Ω resistor. 4. Set Polarity to (+). 5. Set Power Dissipation to 0.5 W. (0.4 W on 370) To connect the Quadrac: 2. Connect MT2 to Collector Terminal (C). To measure the VBO, IBO, and ΔVBO parameter: 6. Set Terminal Selector to Emitter Grounded-Open Base. 1. Set Variable Collector Supply Voltage Range to 75 Max Peak Volts.(80 V on 370) 7. Set Vertical knob to ten times the maximum leakage current (IDRM) specified for the device. 2. Set )PSJ[POUBM knob to 10 V/DIV. Note: The CRT readout should show 1% of the maximum leakage current. The vertical scale is divided by 1,000 when the leakage mode is used. Procedure 2: (+)VDRM and (+)IDRM To measure the (+)VDRM and (+)IDRM parameter: 3. Set Vertical knob to 50 μA/DIV. 4. Set Polarity to AC. 5. Set Mode to Norm. 6. Set Power Dissipation to 0.5 W. (0.4 W on 370) 7. Set Terminal Selector to Emitter Grounded-Open Base. 1. Set Left-Right Terminal Jack Selector to correspond with the location of the test fixture. 2. Increase Variable Collector Supply Voltage to the rated VDRM of the device and observe the dot on the CRT. (Read across horizontally from the dot to the Thyristor and Rectifier Testing Using Curve Tracers ©2008 Littelfuse, Inc. Specifications are subject to change without notice. Please refer to http://www.littelfuse.com for current information. Teccor® brand Thyristors Procedure 5: VBO (Positive and Negative) Note the following: To measure the VBO (Positive and Negative) parameter: 1. Set Left-Right Terminal Jack Selector to correspond with the location of the test fixture. 2. Set Variable Collector Supply Voltage to 55 V (65 V on 370) and apply voltage to the device under test (D.U.T.) using the Left Hand Selector Switch. The peak voltage at which current begins to flow is the VBO value. (Figure AN1006.19) Due to the excessive amount of power that can be generated in this test, only parts with an IT(RMS) rating of 8 A or less should be tested on standard curve tracer. If testing devices above 8 A, a Tektronix model 176 highcurrent module is required. A Kelvin test fixture is required for this test. If a Kelvin fixture is not used, an error in measurement of VTM will result due to voltage drop in fixture. If a Kelvin fixture is not available, Figure AN1006.3 shows necessary information to wire a test fixture with Kelvin connections. PER V E R T 50 A DIV VBO +IBO PER H O R I Z 10 V DIV PER IBO +VBO S T E P ()k DIV 9m PER DIV Figure AN1006.19 (+)VBO = 35 V; (-)VBO = 36 V; (±)IBO" Procedure 6: IBO (Positive and Negative) To measure the IBO (Positive and Negative) parameter, at the VBO point, measure the amount of device current just before the device reaches the breakover point. The measured current at this point is the IBO value. Note: If IBO is less than 10 μA, the current cannot readily be seen on curve tracer. To measure the VTM (Forward and Reverse) parameter: 1. Set Terminal Selector to Emitter Grounded-Open Base. 2. Set Max Peak Volts to 75 V. (80 V on 370) 3. Set Mode to Norm. 4. Set )PSJ[POUBM knob to 0.5 V/DIV. 5. Set Power Dissipation to 220 watts (100 watts on a 577). 6. Set Vertical knob to a sufficient setting to allow the viewing of 1.4 times the IT(RMS) rating of the device IT(peak) on the CRT. Procedure 9: VTM(Forward) To measure the VTM (Forward) parameter: 1. Set Polarity to (+). 2. Set Left-Right Terminal Jack Selector to correspond with the location of the test fixture. 3. Increase Variable Collector Supply Voltage until current reaches rated IT(peak), which is 1.4 times the IT(RMS) rating of the Triac under test. Procedure 7: ΔVBO (Voltage Breakover Symmetry) Note: Model 370 current is limited to 10 A. To measure the ΔVBO (Voltage Breakover Symmetry) parameter: WARNING: Limit test time to 15 seconds maximum. 1. Measure positive and negative VBO values per Procedure 5. 2. Subtract the absolute value of VBO (-) from VBO (+). 4. To measure VTM, follow along horizontal scale to the point where the trace crosses the IT(peak) value. This horizontal distance is the VTM value. (Figure AN1006.20) The absolute value of the result is: PER V E R T ΔVBO = [ I+VBO I - I -VBO I ] 1 A DIV Procedure 8: VTM (Forward and Reverse) To test VTM, the Quadrac must be connected the same as when testing VBO, IBO, and ΔVBO. PER VTM H O R I Z 500 mV DIV PER To connect the Quadrac: S T E P IPK 1. Connect MT1 to Emitter Terminal (E). ()k DIV 9m PER DIV 2. Connect MT2 to Collector Terminal (C). 3. Connect Trigger Terminal to MT2 Terminal through a 10 Ω resistor. Figure AN1006.20 ©2008 Littelfuse, Inc. Specifications are subject to change without notice. Please refer to http://www.littelfuse.com for current information. VTM (Forward) = 1.1 V at IPK = 5.6 A Thyristor and Rectifier Testing Using Curve Tracers AN1006 AN1006 Teccor® brand Thyristors AN1006 Procedure 10: VTM(Reverse) PER V E R T To measure the VTM (Reverse) parameter: DIV 1. Set Polarity to (-). PER 2. Set Left-Right Terminal Jack Selector to correspond with the location of the test fixture. H O R I Z S T E P IH ()k DIV 9m PER DIV Procedure 11: IH(Forward and Reverse) 1. Set Power Dissipation to 50 W. Figure AN1006.21 IH (Forward) = 18 mA Procedure 13: IH(Reverse) To measure the IH (Reverse) parameter: 2. Set Max Peak Volts to 75 V. (80 V on 370) 1. Set Polarity to (-). 3. Set Mode to DC. 2. Continue testing per Procedure 12 for measuring IH (Reverse). 4. Set )PSJ[POUBM knob to 5 V/DIV. 5. Set Vertical knob to approximately 10% of the maximum IH specified. Note: Due to large variations of holding current values, the scale may have to be adjusted to observe holding current. 6. Set Terminal Selector to Emitter Grounded-Open Base. SIDACs The SIDAC is a bidirectional voltage-triggered switch. Upon application of a voltage exceeding the SIDAC breakover voltage point, the SIDAC switches on through a negative resistance region (similar to a DIAC) to a low on-state voltage. Conduction continues until current is interrupted or drops below minimum required holding current. Procedure 12: IH(Forward) To connect the SIDAC: To measure the IH (Forward) parameter: 1. Connect MT1 to the Emitter Terminal (E). 1. Set Polarity to (+). 2. Connect MT2 to the Collector Terminal (C). 2. Set Left-Right Terminal Jack Selector to correspond with the location of the test fixture. To begin testing, perform the following procedures. 3. Increase Variable Collector Supply Voltage to maximum position (100). Note: Depending on the vertical scale being used, the dot may disappear completely from the screen. 4. Decrease Variable Collector Supply Voltage to the point where the line on the CRT changes to a dot. The position of the beginning point of the line, just before the line changes to a dot, represents the IH value. (Figure AN1006.21) V PER 4. Measure VTM(Reverse) the same as in Procedure 8. (Read measurements from upper right corner of screen). To measure the IH (Forward and Reverse) parameter: 5 DIV 3. Increase Variable Collector Supply Voltage until current reaches rated IT(peak). For these steps, it is again necessary to connect the Trigger to MT2 through a 10 Ω resistor. The other connections remain the same. 5 mA Procedure 1: (+) VDRM, (+)IDRM, (-)VDRM, (-)IDRM Note: The (+) and (-) symbols are used to designate the polarity of MT2 with reference to MT1. To measure the (+)VDRM, (+)IDRM, (-)VDRM, and (-)IDRM parameter: 1. Set Variable Collector Supply Voltage Range to 1500 Max Peak Volts. 2. Set )PSJ[POUBM knob to 50 V/DIV. 3. Set Mode to Leakage. 4. Set Polarity to (+). 5. Set Power Dissipation to 2.2 W. (2 W on 370) 6. Set Terminal Selector to Emitter Grounded-Open Base. 7. Set Vertical knob to 50 μA/DIV. (Due to leakage mode, the CRT readout will show 50 nA.) Thyristor and Rectifier Testing Using Curve Tracers ©2008 Littelfuse, Inc. Specifications are subject to change without notice. Please refer to http://www.littelfuse.com for current information. Teccor® brand Thyristors Procedure 2: (+)VDRM and (+)IDRM Procedure 5: VBO To measure the (+)VDRM and (+)IDRM parameter: To measure the VBO parameter, increase Variable Collector Supply Voltage until breakover occurs. (Figure AN1006.23) The voltage at which current begins to flow and voltage on CRT does not increase is the VBO value. 1. Set Left-Right Terminal Jack Selector to correspond with the location of the test fixture. 2. Increase Variable Collector Supply Voltage to the rated VDRM of the device and observe the dot on the CRT. Read across horizontally from the dot to the vertical current scale. This measured value is the leakage current. (Figure AN1006.22) PER V E R T +IBO VBO PER V E R T DIV H O R I Z 50 V +VBO IBO DIV S T E P IDRM Figure AN1006.22 ()k DIV 9m PER DIV IDRM = 50 nA at VDRM = 90 V Procedure 3: (-) VDRMand (-) IDRM To measure the (-)VDRM and (-)IDRM parameter: 1. Set Polarity to (-). 2. Repeat Procedures 1 and 2. (Read measurements from upper right corner of the screen). Procedure 4: VBO and IBO To measure the VBO and IBO parameter: 1. Set Variable Collector Supply Voltage Range to 1500 Max Peak Volts. (2000 V on 370) 2. Set )PSJ[POUBM knob to a sufficient scale to allow viewing of trace at the required voltage level (50 V/DIV for 95 V to 215 V VBO range devices and 100 V/DIVfor devices having VBO ≥ 15 V). H O R I Z 50 V PER S T E P ()k DIV 9m PER DIV PER VDRM PER DIV 50 nA PER 50 A DIV Figure AN1006.23 (+)VBO = 100 V; (-)VBO = 100 V; (±)IBO" Procedure 6: IBO To measure the IBO parameter, at the VBO point, measure the amount of device current just before the device reaches the breakover mode. The measured current at this point is the IBO value. Note: If IBO is less than 10 μA, the current cannot readily be seen on the curve tracer. Procedure 7: IH(Forward and Reverse) To measure the IH (Forward and Reverse) parameter: 1. Set Variable Collector Supply Voltage Range to 1500 Max Peak Volts (400 V on 577; 2000 V on 370). 2. Set )PSJ[POUBM knob to a sufficient scale to allow WJFXJOHPGUSBDFBUUIFSFRVJSFEWPMUBHFMFWFM7%*7 for devices with VBO range from 95 V to 215 V and 100 7%*7GPSEFWJDFTIBWJOH7BO ≥ 215 V). 3. Set Vertical knob to 20% of maximum holding current specified. 3. Set Vertical knob to 50 A/DIV. 4. Set Polarity to AC. 4. Set Polarity to AC. 5. Set Mode to Norm. 5. Set Mode to Norm. 6. Set Power Dissipation to 220 W (100 W on 577). 6. Set Power Dissipation to 10 W. 7. Set Terminal Selector to Emitter Grounded-Open Base. 7. Set Terminal Selector to Emitter Grounded-Open Base. 8. Set Left-Right Terminal Jack Selector to correspond with location of test fixture. 8. Set Left-Right Terminal Jack Selector to correspond with the location of the test fixture. WARNING: Limit test time to 15 seconds maximum. 9. Increase Variable Collector Supply Voltage until device breaks over and turns on. (Figure AN1006.24) ©2008 Littelfuse, Inc. Specifications are subject to change without notice. Please refer to http://www.littelfuse.com for current information. Thyristor and Rectifier Testing Using Curve Tracers AN1006 AN1006 Teccor® brand Thyristors AN1006 PER V E R T 20 mA 3. To measure VTM, follow along horizontal scale to the point where the trace crosses the IT(peak) value. This horizontal distance is the VTM value. (Figure AN1006.25) DIV PER H O R I Z IH 50 V PER V E R T DIV 500 mA DIV IH PER S T E P PER H O R I Z ()k DIV 9m PER DIV Figure AN1006.24 DIV VTM IH = 48 mA in both forward and reverse directions IH is the vertical distance between the center horizontal axis and the beginning of the line located on center vertical axis. 500 mV PER S T E P IPK Figure AN1006.25 ()k DIV 9m PER DIV VTM (Forward) = 950 mV at IPK = 1.4 A Procedure10: VTM(Reverse) Procedure 8: VTM(Forward and Reverse) To measure the VTM (Forward and Reverse) parameter: To measure the VTM (Reverse) parameter: Set Polarity to (-). 1. Set Variable Collector Supply Voltage Range to 350 Max Peak Volts. (400 V on 370) Repeat Procedure 8 to measure VTM(Reverse). 2. Set )PSJ[POUBM knob to 0.5 V/DIV. DIACs 3. Set Vertical knob to 0.5 A/DIV. 4. Set Polarity to (+). 5. Set Mode to Norm. DIACs are voltage breakdown switches used to trigger-on Triacs and non-sensitive SCRs in phase control circuits. 6. Set Power Dissipation to 220 W (100 W on 577). Note: DIACs are bi-directional devices and can be connected in either direction. 7. Set Terminal Selector to Emitter Grounded-Open Base. To connect the DIAC: Before continuing with testing, note the following: Connect one side of the DIAC to the Collector Terminal (C). t ",FMWJOUFTUmYUVSFJTSFRVJSFEGPSUIJTUFTU*GB Kelvin fixture is not used, an error in measurement of VTM will result due to voltage drop in fixture. If a Kelvin fixture is not available, Figure AN1006.3 shows necessary information to wire a test fixture with Kelvin Connections. Connect other side of the DIAC to the Emitter Terminal (E). To continue testing, perform the following procedures. Procedure 9:VTM(Forward) To measure the VTM (Forward) parameter: 1. Set Left-Right Terminal Jack Selector to correspond with the location of the test fixture. 2. Increase Variable Collector Supply Voltage until current reaches rated IT(peak), which is 1.4 times the IT(RMS) rating of the SIDAC. Note: Model 370 current is limited. Set to 400 mA. Check for 1.1 V MAX. WARNING: Limit test time to 15 seconds. Thyristor and Rectifier Testing Using Curve Tracers To begin testing, perform the following procedures. Procedure 1: Curve Tracer Setup To set the curve tracer and begin testing: 1 Set Variable Collector Supply Voltage Range to 75 Max Peak Volts. (80 V on 370) 2. Set )PSJ[POUBM knob to sufficient scale to allow viewing of trace at the required voltage level (10 V to 20 V/DIV depending on device being tested). 3. Set Vertical knob to 50 μA/DIV. 4. Set Polarity to AC. 5. Set Mode to Norm. 6. Set Power Dissipation to 0.5 W. (0.4 W on 370) 7. Set Terminal Selector to Emitter Grounded-Open Base. ©2008 Littelfuse, Inc. Specifications are subject to change without notice. Please refer to http://www.littelfuse.com for current information. Teccor® brand Thyristors Procedure 2: VBO Model 370 Curve Tracer Procedure Notes To measure the VBO parameter: 1. Set Left-Right Terminal Jack Selector to correspond with the location of the test fixture. 2. Set Variable Collector Supply Voltage to 55 V (65 V for 370) and apply voltage to device under test (D.U.T.), using Left-Right-Selector Switch. The peak voltage at which current begins to flow is the VBO value. (Figure AN1006.26) PER V E R T 50 A Because the curve tracer procedures in this application note are written for the Tektronix model 576 curve tracer, certain settings must be adjusted when using model 370. Variable Collector Supply Voltage Range and Power Dissipation controls have different scales than model 576. The following table shows the guidelines for setting Power Dissipation when using model 370. (Figure AN1006.27) Model 576 Model 370 If power dissipation is 0.1 W, set at 0.08 W. If power dissipation is 0.5 W, set at 0.4 W. If power dissipation is 2.2 W, set at 2 W. If power dissipation is 10 W, set at 10 W. If power dissipation is 50 W, set at 50 W. If power dissipation is 220 W, set at 220 W. DIV +IBO PER H O R I Z 10 V DIV PER IBO VBO +VBO S T E P ()k DIV 9m PER DIV Although the maximum power setting on the model 370 curve tracer is 200 W, the maximum collector voltage available is only 400 V at 220 W. The following table shows the guidelines for adapting Collector Supply Voltage Range settings for model 370 curve tracer procedures: Model 576 Figure AN1006.26 Procedure 3: (+)VBO = 35 V; (-)VBO = 36 V; (±)IBO" IBO"BOE$BOOPU#F3FBE&BTJMZ IBO To measure the IBO parameter, at the VBO point, measure the amount of device current just before the device reaches the breakover mode. The measured current at this point is the IBO value. Note: If IBO is less than 10 μA, the current cannot readily be seen on the curve tracer. Procedure 4: ΔVBO(Voltage Breakover Symmetry) To measure the ΔVBO (Voltage Breakover Symmetry) parameter: 1. Measure positive and negative values of VBO as shown in Figure AN1006.26. Model 370 If voltage range is 15 V set at 16 V. If voltage range is 75 V set at 80 V. If voltage range is 350 V set at 400 V. If voltage range is 1500 V set at 2000 V The following table shows the guidelines for adapting terminal selector knob settings for model 370 curve tracer procedures: Model 576 Model 370 If Step generator (base) is emitter grounded then Base Step generator is emitter common. If Emitter grounded is open base then Base open is emitter common. 2. Subtract the absolute value of VBO(-) from VBO(+). The absolute value of the result is: ΔVBO = [ I +VBO I - I -VBO I ] ©2008 Littelfuse, Inc. Specifications are subject to change without notice. Please refer to http://www.littelfuse.com for current information. Thyristor and Rectifier Testing Using Curve Tracers AN1006 AN1006 Teccor® brand Thyristors AN1006 HORIZONTAL VOLTAGE CONTROL PROGRAMMABLE CURVE TRACER DISPLAY INTENSITY SETUP MEMORY Note: All Voltage Settings Will Be Referenced to "Collector" STEP GENERATOR HORIZONTAL VOLTS/DIV VERTICAL CURRENT/DIV VERT/DIV CURSOR POLARITY STEP/OFFSET AMPLITUDE COLLECTOR HORZ/DIV CURSOR CRT STEP/OFFSET POLARITY STEP/OFFSET AMPLITUDE (AMPS/VOLTS) PER STEP OFFSET OFFSET OFFSET OR gm/DIV CURSOR POSITION AUX SUPPLY GPIB PLOTTER MEASUREMENT STEP FAMILY AUX SIPPLY COLLECTOR SUPPLY VARIABLE COLLECTOR SUPPLY VOLTAGE RANGE TERMINAL JACKS CONFIGURATION COLLECTOR SUPPLY MAX PEAK VOLTS C C POLARITY VARIABLE GATE/TRIGGER B LEFT B SENSE B E E BOTH B SENSE VARIABLE COLLECTOR SUPPLY VOLTAGE E SENSE E SENSE RIGHT MAX PEAK POWER (POWER DISSIPATION) C SENSE C SENSE MT2/ANODE MAX PEAK POWER WATTS POWER LEFT-RIGHT SELECTOR FOR TERMINAL JACKS Figure AN1006.27 MT1/CATHODE KELVIN TERMINALS USED WHEN MEASURING V TM OR V FM TERMINAL SELECTOR Tektronix Model 370 Curve Tracer Model 577 Curve Tracer Procedure Notes Because the curve tracer procedures in this application note are written for the Tektronix model 576 curve tracer, certain settings must be adjusted when using model 577. Model 576 curve tracer has separate controls for polarity (AC,+,-) and mode (Norm, DC, Leakage), whereas Model 577 has only a polarity control. The following table shows the guidelines for setting Collector Supply Polarity when using model 577. (Figure AN1006.28) Model 576 Model 577 If using Leakage mode along with polarity setting of +(NPN) and -(PNP),[vertical scale divided by 1,000], set Collector Supply Polarity to either +DC or -DC, depending on polarity setting specified in the procedure. The vertical scale is read directly from the scale on the control knob. If using DC mode along with either +(NPN) or -(PNP) polarity, set Collector Supply Polarity to either +DC or -DC depending on polarity specified. If using Norm mode along with either +(NPN) or -(PNP) polarity, set Collector Supply Polarity to either +(NPN) or -(PNP) per specified procedure. If using Norm mode with AC polarity, 0OFEJGGFSFODFCFUXFFONPEFMTBOEJTUIF4UFQ Offset Polarity setting. The polarity is inverted when the button is depressed on the Model 576 curve tracer. 5IF.PEFMJTPQQPTJUFUIF4UFQ0GGTFU1PMBSJUZJT “inverted” when the button is extended and “Normal” Thyristor and Rectifier Testing Using Curve Tracers set Collector Supply Polarity to AC. XIFOUIFCVUUPOJTEFQSFTTFE5IF4UFQ0GGTFU1PMBSJUZ is used only when measuring IGT and VGT of Triacs and Quadracs in Quadrants l through lV. ©2008 Littelfuse, Inc. Specifications are subject to change without notice. Please refer to http://www.littelfuse.com for current information. Teccor® brand Thyristors Also, the Variable Collector Supply Voltage Range and Power Dissipation controls have different scales than model 576. The following table shows the guidelines for setting Power Dissipation when using model 577. Model 576 Although the maximum power setting on model 576 curve tracer is 220 W (compared to 100 W for model 577), the maximum collector current available is approximately the same. This is due to the minimum voltage range on model 577 curve tracer being 6.5 V compared to 15 V for model 576. The following table shows the guidelines for adapting Collector Voltage Supply Range settings for model 577 curve tracer procedures: Model 577 If power dissipation is 0.1 W, set at 0.15 W. If power dissipation is 0.5 W, set at 0.6 W. If power dissipation is 2.2 W, set at 2.3 W. If power dissipation is 10 W, set at 9 W. If power dissipation is 50 W, set at 30 W. If power dissipation is 220 W, set at 100 W. Model 576 Model 577 If voltage range is 15 V set at either 6.5 V or 25 V, depending on parameter being tested. Set at 6.5 V when measuring VTM (to allow maximum collector current) and set at 25 V when measuring IGT and VGT. If voltage range is 75 V set at 100 V. If voltage range is 1500 V, set at 1600 V. BRIGHTNESS STORE INTENSITY CRT Avoid extremely bright display FOCUS Adjust for best focus BEAM FINDER VARIABLE COLLECTOR SUPPLY VOLTAGE RANGE VARIABLE COLLECTOR SUPPLY VOLTAGE POWER STEP FAMILY VARIABLE COLLECTOR% STEP GENERATOR SECTION STEP/OFFSET AMPLIFIER MAX PEAK VOLTS NUMBER OF STEPS MAX PEAK POWER (POWER DISSIPATION) OFFSET MULTI Watch high power settings. Can damage device under test STEP/OFFSET POLARITY POLARITY COLLECTOR SUPPLY POLARITY POSITION DISPLAY Indicates Collector Supply Disabled HORIZONTAL VOLTAGE CONTROL Note: All Voltage Settings Will Be Referenced to "Collector" STEP RATE POSITION COLLECTOR SUPPLY Terminal Selector TERMINAL JACKS MT2/ANODE GATE/TRIGGER MT1/CATHODE C C SENSE C B E C SENSE B E SENSE E Indicates Dangerous Voltages on Test jacks E SENSE VERTICAL (off) LEFT-RIGHT SELECTOR FOR TERMINAL JACKS LEFT RIGHT KELVIN TERMINALS USED WHEN MEASURING VTM OR VFM VARIABLE VOLTAGE LOOPING COMPENSATION Figure AN1006.28 STEP GEN OUTPUT VARIABLE OUTPUT VERTICAL CURRENT SUPPLY GROUND EXT BASE OR EMIT INPUT Tektronix Model 577 Curve Tracer ©2008 Littelfuse, Inc. Specifications are subject to change without notice. Please refer to http://www.littelfuse.com for current information. Thyristor and Rectifier Testing Using Curve Tracers AN1006 AN1006