ML200MHI Datasheet - Micropower Direct

ML200MHI Series
Compact SMT, 2W
Medical Approved
DC/DC Converters
Electrical Specifications
Specifications typical @ +25°C, nominal input voltage & rated output current, unless otherwise noted. Specifications subject to change without notice.
Key Features:
• 2W Output Power
• 4 kVAC rms Isolation
• Reinforced Insulation
• Low Leakage Current
• Compact SMT Case
• Single & Dual Outputs
• 2.0 MH MTBF
• Industry Standard Pin-Out
• EN 60950 Approved
• EN 60601 Approved
Input
Parameter
Conditions
Input Voltage Range
Input Filter
Reverse Polarity Input Current
Leakage Current
Min.
Typ.
5 VDC Input
4.5
5.0
12 VDC Input 10.8
12.0
24 VDC Input 21.6
24.0
Internal Capacitor
Max.
5.5
13.2
26.4
Units
0.3
2.0
A
µA
Max.
±4.0
±1.0
±1.5
Units
%
%
%
150
200
15
±0.02
mV P - P
mV P - P
mV rms
%/°C
Typ.
Max.
15
80
20
100
Units
VAC rms
VPk
VAC
G⍀
pF
kHz
Typ.
Max.
+80
+90
+125
Units
°C
°C
°C
95
%
VDC
Output
Parameter
Output Voltage Accuracy
Output Voltage Balance
Line Regulation
Load Regulation, See Note 1
Ripple & Noise (20 MHz), See Note 2
Ripple & Noise (20 MHz)
Ripple & Noise (20 MHz)
Temperature Coefficient
Output Short Circuit
Conditions
Min.
Typ.
±2.0
Dual Outputs, Balanced Loads
±0.1
Vin = Min to Max
±1.2
See Model Selection Guide
100
Over Line, Load & Temp.
±0.01
Momentary (0.5 Sec.)
General
Parameter
Isolation Voltage, Rated
Isolation Test Voltage
Reinforced Insulation Working Voltage
Isolation Resistance
Isolation Capacitance
Switching Frequency
Conditions
60 Seconds
Flash Tested For 1 Sec
500 VDC
100 kHz, 1V
Min.
4,000
6,000
300
10
50
Environmental
Parameter
Operating Temperature Range
Operating Temperature Range
Storage Temperature Range
Cooling
Humidity
Physical
Case Size
Case Material
Weight
MicroPower Direct
292 Page Street
Suite D
Stoughton, MA 02072
USA
Conditions
Ambient
Case
Min.
-25
-50
Free Air Convection
RH, Non-condensing
See Mechanical Drawing (Page 2)
Non-Conductive Black Plastic (UL94-V0)
0.14 Oz (3.75g)
Reliability Specifications
Parameter
MTBF
Safety Approvals
Moisture Sensitivity Level
Conditions
Min.
Typ.
Max.
MIL HDBK 217F, 25°C, Gnd Benign
2.0
UL 60601, UL 60950, EN 60601, EN 60950
IPC/JEDEC J-STD-020C Level 2
Units
MHours
Absolute Maximum Ratings
T: (781) 344-8226
F: (781) 344-8481
E: [email protected]
W: www.micropowerdirect.com
Parameter
Conditions
Input Voltage Surge (1 Sec)
Peak Reflow Temperature
Lead Temperature
Internal Power Dissipation
5 VDC Input
12 VDC Input
24 VDC Input
See Note 3
1.5 mm From Case For 10 Sec
All Models
Min.
-0.7
-0.7
-0.7
Typ.
Max.
9.0
18.0
30.0
245
260
650
Caution: Exceeding Absolute Maximum Ratings may damage the module. These are not continuous operating ratings.
www.micropowerdirect.com
Units
VDC
°C
°C
mW
www.micropowerdirect.com
Model Selection Guide
Model
Number
ML205MS-05HI
ML205MS-12HI
ML205MS-15HI
ML205MD-12HI
ML205MD-15HI
ML212MS-05HI
ML212MS-12HI
ML212MS-15HI
ML212MD-12HI
ML212MD-15HI
ML224MS-05HI
ML224MS-12HI
ML224MS-15HI
ML224MD-12HI
ML224MD-15HI
Input
Output
Current
Voltage (VDC)
Current (mA)
Voltage
Current
Nominal
Range
Full-Load No-Load (VDC) (mA, Max) (mA, Min)
5
4.50 - 5.50
606
90
5.0
400.0
8.0
5
4.50 - 5.50
600
90
12.0
165.0
3.0
5
4.50 - 5.50
605
90
15.0
133.0
2.5
5
4.50 - 5.50
553
90
±12.0
±83.0
±1.5
5
4.50 - 5.50
542
90
±15.0
±66.0
±1.0
12
10.8 - 13.2
253
40
5.0
400.0
8.0
12
10.8 - 13.2
250
40
12.0
165.0
3.0
12
10.8 - 13.2
252
40
15.0
133.0
2.5
12
10.8 - 13.2
224
40
±12.0
±83.0
±1.5
12
10.8 - 13.2
220
40
±15.0
±66.0
±1.0
24
21.6 - 26.4
126
30
5.0
400.0
8.0
24
21.6 - 26.4
125
30
12.0
165.0
3.0
24
21.6 - 26.4
126
30
15.0
133.0
2.5
24
21.6 - 26.4
112
30
±12.0
±83.0
±1.5
24
21.6 - 26.4
110
30
±15.0
±66.0
±1.0
Notes:
1. Load regulation is measured for a load change of 20% to 100%.
2. When measuring output ripple, it is recommended that an external
0.47 µF ceramic capacitor be placed from the +Vout pin to the
-Vout pin for single output units and from each output to common
for dual output units. For noise sensitive applications, the use of
3.3 µF capacitors will reduce the output ripple.
3. The recommended time above liquidous (TL) is <60 seconds at 217 °C.
For more information, please contact the factory.
4. Operation at no-load will not damage these units. However, they
may not meet all specifications.
5. For dual output units, the maximum capacitive load is given for
each output.
6. Dual output units may be connected to provide a 24 VDC or 30 VDC
output. To do this, connect the load across the positive (+Vout) and
negative (-Vout) outputs and float the output common.
7. The converter should be connected to a low ac-impedance source.
An input source with a highly inductive impedance may affect the
stability of the converter. In applications where the converter output
loading is high and input power is supplied over long lines, it may
be necessary to use a capacitor on the input to insure start-up.
In this case, it is recommended that a low ESR (ESR <1.0 at
100 kHz) capacitor be mounted close to the converter. For 5V input
units a 2.2 µF is recommended, for 12V input units, a 1.0 µF; and
for 24V units a 0.47 µF.
8. It is recommended that a fuse be used on the input of a power
supply for protection. See the table above for the correct rating.
Capacitive
Load
Efficiency
Load
Regulation
(%, Typ)
(µF, Max)
(%, Max)
330
330
330
±100
±100
330
330
330
±100
±100
330
330
330
±100
±100
12
10
10
10
10
12
10
10
10
10
12
10
10
10
10
66
66
66
72
73
66
66
66
74
75
66
66
66
74
75
Fuse Rating
Slow-Blow
(mA)
1,000
1,000
1,000
1,000
1,000
500
500
500
500
500
200
200
200
200
200
Derating Curve
Mechanical Dimensions
Pin Connections
Pin
Single
Dual
1
-Vin
-Vin
7
NC
NC
8
NC
Common
9
+Vout
+Vout
10
-Vout
-Vout
16
+Vin
+Vin
NC = No Connection
Mechanical Notes:
• All dimensions are typical in inches (mm)
• Tolerance x.xx = ±0.01 (±0.25)
MicroPower Direct
We Power Your Success - For Less!
292 Page Street Ste D Stoughton, MA 02072
•
TEL: (781) 344-8226
•
FAX: (781) 344-8481
•
E-Mail: [email protected]