GSM40B series

40W AC-DC High Reliability Medical Adaptor
(
GSM40B
series
for 5~9V)
Features
Universal AC input / Full range
2 pole AC inlet IEC320-C8
Medical safety approved (2 x MOPP between primary to secondary)
Suitable for BF application with appropriate system consideration
Low leakage current <50uA
No load power consumption<0.1W
Energy efficiency level VI(Except 5~9V for Level V)
Comply with EISA 2007/DoE,NRCan, AU/NZ MEPS,
Applications
Mobile clinical workstation
Oral irrigator
Portable hemodialysis machine
Breath Machine
Medical computer monitor
EU ErP and meet CoC Version 5
Built-in active PFC function
High efficiency up to 91%
Fanless design with -30~+70℃ working temperature
Class Ⅱ power (without earth pin)
Protections: Short circuit / Overload / Over voltage / Over temperature
Fully enclosed plastic case
LED indicator for power on
100% full load burn-in test
Optional lock type DC plug
3 years warranty
Description
GSM40B is a highly reliable, 40W desktop style single-output green medical adaptor series. This product
is equipped with a 2-pin (no FG) standard IEC320-C8 power plug, adopting the input range from 80VAC to
264VAC. The entire series supplies different output voltages between 5VDC and 48VDC that can satisfy
the demands for various kinds of medical electrical devices. The circuitry design meets the international
medical standards (2*MOPP), having an ultra low leakage current (<50 uA), fitting the medical devices in
direct electrical contact with the patients.
With the efficiency up to 91% and the extremely low no-load power consumption below 0.1W, GSM40B is
compliant with USA EISA 2007/DoE, Canada NRCan, Australia and New Zealand MEPS, EU ErP,and meet
Code of Conduct (CoC) Version 5. The supreme feature allows the adaptor to save the energy when it is either
under the operating mode or the standby mode. The entire series utilizes the 94V-0 flame retardant plastic
case, providing the double insulation that effectively prevents electrical shock. GSM40B is approved with
the international medical safety certificates.
Model Encoding
GSM 40 B 05 - P1J
DC plug type
{
P1J: Plug for standard model,2.1ψ x5.5ψ x11mm,c+,tuning fork type
Optional plug type available by customer requested
Output voltage
IEC320-C8 AC inlet
Output wattage
Series name
File Name:GSM40B-SPEC 2015-06-17
GSM40B
40W AC-DC High Reliability Medical Adaptor
series
SPECIFICATION
ORDER NO.
GSM40B05-P1J GSM40B07-P1J GSM40B09-P1J GSM40B12-P1J GSM40B15-P1J GSM40B18-P1J GSM40B24-P1J GSM40B48-P1J
SAFETY MODEL NO.
DC VOLTAGE
OUTPUT
GSM40B05
Note.2 5V
GSM40B12
GSM40B15
GSM40B18
GSM40B24
GSM40B48
9V
12V
15V
18V
24V
48V
5A
5.34A
4.45A
3.34A
2.67A
2.22A
1.67A
0.84A
CURRENT RANGE
0 ~ 5A
0 ~ 5.34A
0 ~ 4.45A
0 ~ 3.34A
0 ~ 2.67A
0 ~ 2.22A
0 ~ 1.67A
0 ~ 0.84A
RATED POWER (max.)
25W
40W
40W
40W
40W
40W
40W
40W
RIPPLE & NOISE (max.) Note.3 100mVp-p
100mVp-p
100mVp-p
100mVp-p
100mVp-p
150mVp-p
180mVp-p
240mVp-p
VOLTAGE TOLERANCE Note.4 ±5.0%
LINE REGULATION
Note.5 ±1.0%
±5.0%
±1.0%
±5.0%
±5.0%
±1.0%
±5.0%
±3.0%
±1.0%
±3.0%
±3.0%
±1.0%
±3.0%
±3.0%
±1.0%
±3.0%
±2.5%
±1.0%
±2.5%
±2.5%
±1.0%
±2.5%
89%
90%
91%
25.2 ~ 32.4V
50.4 ~ 64.8V
±5.0%
SETUP, RISE TIME
Note.6 1000ms, 30ms / 230VAC
HOLD UP TIME (Typ.)
VOLTAGE RANGE
50ms / 230VAC
15ms / 115VAC at full load
Note.7 80 ~ 264VAC 113 ~ 370VDC
FREQUENCY RANGE
81%
AC CURRENT (Typ.)
INRUSH CURRENT (Typ.)
1A / 115VAC
OVERLOAD
OVER VOLTAGE
1500ms, 30ms / 115VAC at full load
47 ~ 63Hz
EFFICIENCY (Typ.)
LEAKAGE CURRENT(max.)
PROTECTION
GSM40B09
7.5V
RATED CURRENT
LOAD REGULATION
INPUT
GSM40B07
85.5%
86%
105 ~ 160% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
5.25 ~ 6.75V
7.88 ~ 10.13V 9.45 ~ 12.15V 12.6 ~ 16.2V 15.75 ~ 20.25V 18.9 ~ 24.3V
Protection type : Shut down o/p voltage, re-power on to recover
WORKING TEMP.
-30 ~ +60℃ (Refer to "Derating Curve")
WORKING HUMIDITY
20% ~ 90% RH non-condensing
-40 ~ +85℃, 10 ~ 95% RH
±0.03% / ℃ (0 ~ 50℃)
SAFETY STANDARDS
ISOLATION
LEVEL
SAFETY &
WITHSTAND VOLTAGE
EMC
(Note. 8)
ISOLATION RESISTANCE
Primary-Secondary: 2xMOPP
I/P-O/P:4KVAC O/P-FG:SHORT
CONNECTOR
NOTE
88.5%
30A / 115VAC
65A / 230VAC
Touch current < 50 A/264VAC
ENVIRONMENT STORAGE TEMP., HUMIDITY
TEMP. COEFFICIENT
VIBRATION
OTHERS
88%
0.5A / 230VAC
10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
ANSI/AAMI ES60601-1 / ES60601-1-11, TUV EN60601-1 / 60601-1-11 approved
EMC EMISSION
I/P-O/P:100M Ohms / 500VDC / 25℃/ 70% RH
Compliance to EN55011(CISPR11) class B, EN61000-3-2,3, FCC PART 15 class B
EMC IMMUNITY
Compliance to EN61000-4-2,3,4,5,6,8,11, EN55024, EN60601-1-2, EN61204-3 medical level, criteria A
MTBF
DIMENSION
740K hrs min. MIL-HDBK-217F(25℃)
125*50*31.5mm (L*W*H)
0.29Kg; 40pcs/12.6Kg/1.05CUFT
PACKING
PLUG
See page 3 ; Other type available by customer requested
CABLE
See page 3 ; Other type available by customer requested
1. All parameters are specified at 230VAC input, rated load, 25℃ 70% RH ambient.
2. DC voltage: The output voltage set at point measure by plug terminal & 50% load.
3. Ripple & noise are measured at 20MHz by using a 12" twisted pair terminated with a 0.1uf & 47uf capacitor.
4. Tolerance: includes set up tolerance, line regulation, load regulation.
5. Line regulation is measured from low line to high line at rated load.
6. Length of set up time is measured at first cold start. Turning ON/OFF the power supply may lead to increase of the set up time.
7. Derating may be needed under low input voltages. Pleas check the derating curve for more details.
8. The power supply is considered as an independent unit, but the final equipment still need to re-confirm that the whole system complies with the
EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”
(as available on http://www.meanwell.com)
File Name:GSM40B-SPEC 2015-06-17
GSM40B
40W AC-DC High Reliability Medical Adaptor
Derating Curve
series
Static Characteristics
100
100
90
80
LOAD (%)
LOAD (%)
80
60
40
70
60
50
20
40
-30
0
10
20
30
40
50
60
70
(HORIZONTAL)
AMBIENT TEMPERATURE (℃)
80
95
100
110
120
140
160
180
200
220
240
264
INPUT VOLTAGE (VAC) 60Hz
Mechanical Specification
Case No. GSM60B
Unit:mm
Ul2464 16AWG 1000±50mm for 5 ~ 7.5V
UL1185 16AWG 1000±50mm for 9 ~ 12V
UL1185 16AWG 1500±50mm for 15 ~ 48V
31.5
50
POWER LED
70±10mm
11±0.5mm
ID 2.1 x OD 5.5
Outside
Inside
Plug Assignment
Standard plug: P1J
P1J
P/N
OUTPUT
CENTER
+
Optional lock type plug: P2S
SWITCHCRAFT S761K plug equivalent
Installation Manual
Please refer to : http://www.meanwell.com/webnet/search/InstallationSearch.html
File Name:GSM40B-SPEC 2015-06-17