GSM06E series

6W AC-DC High Reliability Medical Adaptor
GSM06E
series
EN 60601-1
IEC 60601-1
8
■ Features
Universal AC input / Full range
2 pole EURO plug
Medical safety approved (2 × MOPP between primary to secondary)
Suitable for BF application with appropriate system consideration
Low leakage current <50μA
No load power consumption<0.3W
Energy efficiency Level Ⅴ
Comply with EU ErP
Class Ⅱ power (without earth pin)
Protections: Short circuit / Overload / Over voltage
Fully enclosed plastic case
Optional lock type DC plug
100% full load burn-in test
3 years warranty
■ Applications
Blood glucose meter
Blood pressure meter
Nebulizer
Inhaler
Portable medical device
■ Description
GSM06E is a highly reliable, 6W wall-mounted style single-output green medical adaptor series. This
product is equipped with a 2-pin (no FG) standard European power plug, adopting the input range from
85VAC to 264VAC. The entire series supplies different output voltages between 5VDC and 24VDC that
can satisfy the demands for various kinds of miniature medical devices. The circuitry design meets the
international medical standards (2 x MOPP), having an ultra low leakage current (<50μA), fitting the medical
devices in direct electrical contact with the patients.
With the efficiency up to 82% and the extreme low no-load power consumption below 0.3W. GSM06E is
compliant with EU ErP. The supreme feature allows the adaptor to save the energy when it is either under
the operating mode or the standby mode. The entire series utilizes the 94V-0 flame retardant plastic
case, providing the double insulation that effectively prevents electrical shock. GSM06E is approved
with the international medical safety certificates.
■ Model Encoding
GSM 06 E 05 - P1J
DC plug type
{
P1J : Plug for standard model, 2.1ψ × 5.5ψ ×11mm, c+, tuning fork type
Optional plug type available by customer requested
Output voltage
European 2 pin AC plug
Output wattage
Series name
File Name:GSM06E-SPEC 2015-05-12
GSM06E
6W AC-DC High Reliability Medical Adaptor
series
SPECIFICATION
GSM06E05-P1J GSM06E06-P1J GSM06E07-P1J GSM06E09-P1J GSM06E12-P1J GSM06E15-P1J GSM06E18-P1J GSM06E24-P1J
ORDER NO.
SAFETY MODEL NO.
DC VOLTAGE
OUTPUT
GSM06E05
Note.2 5V
GSM06E06
GSM06E07
GSM06E09
GSM06E12
GSM06E15
GSM06E18
GSM06E24
6V
7.5V
9V
12V
15V
18V
24V
0.25A
RATED CURRENT
1.2A
1A
0.8A
0.66A
0.5A
0.4A
0.33A
CURRENT RANGE
0 ~ 1.2A
0 ~ 1A
0 ~ 0.8A
0 ~ 0.66A
0 ~ 0.5A
0 ~ 0.4A
0 ~ 0.33A
0 ~ 0.25A
RATED POWER
6W
6W
6W
6W
6W
6W
6W
6W
RIPPLE & NOISE (max.) Note.3 50mVp-p
50mVp-p
80mVp-p
80mVp-p
100mVp-p
120mVp-p
150mVp-p
180mVp-p
VOLTAGE TOLERANCE Note.4 ±5.0%
±5.0%
±5.0%
±5.0%
±5.0%
±5.0%
±5.0%
±4.0%
LINE REGULATION
Note.5 ±0.5%
±0.5%
±0.5%
±0.5%
±0.5%
±0.5%
±0.5%
±0.5%
LOAD REGULATION
Note.6 ±5.0%
±5.0%
±5.0%
±5.0%
±3.0%
±3.0%
±3.0%
±2.0%
74%
76%
77%
79%
80%
82%
SETUP, RISE, HOLD UP TIME 1000ms, 50ms, 12ms at full load
85 ~ 264VAC 120 ~ 370VDC
VOLTAGE RANGE
INPUT
FREQUENCY RANGE
47 ~ 63Hz
EFFICIENCY (Typ.)
68%
AC CURRENT
INRUSH CURRENT (max.)
0.18A / 100VAC
LEAKAGE CURRENT(max.)
OVERLOAD
PROTECTION
OVER VOLTAGE
WORKING TEMP.
WORKING HUMIDITY
ENVIRONMENT STORAGE TEMP., HUMIDITY
TEMP. COEFFICIENT
VIBRATION
SAFETY STANDARDS
SAFETY & WITHSTAND VOLTAGE
ISOLATION RESISTANCE
EMC
(Note. 7)
EMC EMISSION
EMC IMMUNITY
MTBF
OTHERS
CONNECTOR
NOTE
74%
30A / 230VAC
Touch current < 50μA/264VAC
>105% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
110 ~ 140% rated output power
Protection type : Clamp by zener diode
0 ~ +50℃(Refer to "Derating Curve")
20% ~ 90% RH non-condensing
-20 ~ +85℃, 10 ~ 95% RH
±0.04% / ℃ (0 ~ 40℃)
10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
IEC60950-1, TUV EN60601-1, EN60601-1-11 approved
I/P-O/P:5656VDC
I/P-O/P:100M Ohms / 500VDC / 25℃/ 70% RH
Compliance to EN55011, EN61000-3-2,3
Compliance to EN61000-4-2,3,4,5,6,8,11, EN60601-1-2 medical level, criteria A
DIMENSION
PACKING
PLUG
500Khrs min. MIL-HDBK-217F(25℃)
32*66*42.5mm (L*W*H)
100g ; 90pcs / 10Kg / CARTON
See page 3 ; Other type available by customer requested
CABLE
See page 3 ; Other type available by customer requested
1.All parameters are specified at 230VAC input, rated load, 25℃ 70% RH ambient.
2.DC voltage: The output voltage set at point measure by plug terminal & 50% load.
3.Ripple & noise are measured at 20MHz by using a 12" twisted pair terminated with a 0.1uf & 47uf capacitor.
4.Tolerance: includes set up tolerance, line regulation, load regulation.
5.Line regulation is measured from low line to high line at rated load.
6.Load regulation is measured from 0% to 100% rated load.
7.The power supply is considered as an independent unit, but the final equipment still need to re-confirm that the whole system complies with the
EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”
(as available on http://www.meanwell.com)
File Name:GSM06E-SPEC 2015-05-12
GSM06E
6W AC-DC High Reliability Medical Adaptor
Derating Curve
series
Static Characteristics
100
100
LOAD (%)
LOAD (%)
90
70
80
70
60
50
40
0
10
20
30
40
50
85
AMBIENT TEMPERATURE (℃)
100
110
135
155
230 264
INPUT VOLTAGE (V) 60Hz
Mechanical Specification
Plug Assignment
Unit:mm
Standard plug: P1J
34.5±1
P1J
NAME PLATE
4ψ
13.6
P/N
OUTPUT
CENTER
+
18.3
19.8
18
Optional lock type plug: P2S
SWITCHCRAFT S761K plug equivalent
UL2468 22AWG 1220 +150/-0mm for 5~9V
UL2468 24AWG 1830 +150/-0mm for 12~24V
19±0.5
62.5±1
42.5±1
66±1
11
C"+"
32±1
ID 2.1 x OD 5.5
Installation Manual
Please refer to : http://www.meanwell.com/webnet/search/InstallationSearch.html
File Name:GSM06E-SPEC 2015-05-12