RPT-60 series 60W Triple Output Medical Type MODEL:RPT-60C OUTPUT FUNCTION TEST NO 1 2 3 4 5 TEST ITEM SPECICATION TEST CONDITION RIPPLE & NOISE V1: 80 mVp-p (Max ) I/P: 230VAC RESULT V1: 15 mVp-p (Max ) VERDICT V2: 100 mVp-p (Max ) 17 mVp-p (Max ) mVp-p (Max ) O/P:FULL LOAD Ta:25℃ V2: V3: 150 V3: 31 mVp-p (Max ) OUTPUT VOLTAGE V1: 3 %~ -2 % (Max) I/P: 115 VAC / 264 VAC V1: 0.25 %~ -0.25 % TOLERANCE V2: 6 %~ -6 % (Max) V2: 2.9 %~ -2.9 % V3: 8 %~ -8 % (Max) O/P:FULL/ MIN Ta:25℃ V3: 4.6 %~ -4.6 % V1: 0.5 %~ -0.5 % (Max) I/P: 115 VAC ~ 264 VAC V1: 0.12 %~ -0.12 % V2: 2 %~ -2 % (Max) V2: 0.04 %~ -0.04 % V3: 2 %~ -2 % (Max) O/P:FULL LOAD Ta:25℃ V3: 0.4 %~ -0.4 % V1: 1.5 %~ -1.5 % (Max) I/P: 230 VAC V1: 0.24 %~ -0.24 % V2: 3 %~ -3 % (Max) V2: 0.3 %~ -0.3 % V3: 4 %~ -4 % (Max) O/P:FULL ~MIN LOAD Ta:25℃ V3: 2 %~ -2 % V1: 1.5 %~ -1.5 % (Max) I/P: 230 VAC V1: 0 %~ 0 % V2: 3 %~ -3 % (Max) O/P: Testing O/P 60%LOAD V2: 1.4 %~ -1.4 % V3: 2.5 %~ -2.5 % LINE REGULATION LOAD REGULATION CROSS REGULATION V3: 4 %~ -4 % (Max) LOAD Other O/P 40%LOAD P P P P P Change Ta:25℃ 6 SET UP TIME 230VAC: 200 ms (Max) I/P: 230 VAC 230VAC/ 165 ms 115 VAC: 200 ms (Max) I/P: 115 VAC 115VAC/ 158 ms O/P:FULL LOAD Ta:25℃ 7 RISE TIME 230VAC: 15 ms (Max) I/P: 230 VAC 230VAC/ 7 ms 115VAC: 15 ms (Max) I/P: 115 VAC 115VAC/ 7 ms O/P:FULL LOAD Ta:25℃ 8 HOLD UP TIME 230VAC: 70 ms (TYP) I/P: 230 VAC 230VAC/ 79 ms 115VAC: 14 ms (TYP) I/P: 115 VAC 115VAC/ 15.2 ms <5 % O/P:FULL LOAD Ta:25℃ 9 OVER/UNDERSHOOT TEST I/P: 230 VAC < +5% TEST: O/P:FULL LOAD Ta:25℃ 10 DYNAMIC LOAD V1: 1000 mVp-p I/P: 230 VAC P P P P 118 mVp-p O/P:FULL /Min LOAD P 90%DUTY/1KHZ Ta:25℃ Model: RPT-60C Test Report 1 / 5 RPT-60 series 60W Triple Output Medical Type INPUT FUNCTION TEST NO 1 TEST ITEM SPECICATION TEST CONDITION INPUT VOLTAGE RANGE 90VAC~264 VAC I/P:TESTING RESULT VERDICT 71V~264V O/P:FULL LOAD Ta:25℃ I/P: TEST: OK LOW-LINE-3V= 87V P HIGH-LINE+15%=300 V O/P:FULL/MIN LOAD ON: 30 Sec . OFF: 30 Sec 10MIN ( AC POWER ON/OFF NO DAMAGE ) 2 3 INPUT FREQUENCY RANGE 47HZ ~63 HZ I/P: 90VAC ~ 264 VAC NO DAMAGE OSC O/P:FULL~MIN LOAD Ta:25℃ 79% (TYP) EFFICIENCY TEST: OK P I/P: 230 VAC 79.1 % O/P:FULL LOAD Ta:25℃ 4 INPUT CURRENT P 230V/ 0.7 A (TYP) I/P: 230 VAC I= 0.62 A/ 230 VAC 115V/ 1.1 A (TYP) I/P: 115 VAC I= 0.97 A/ 115 VAC P O/P:FULL LOAD Ta:25℃ 5 INRUSH CURRENT 230V/ 60 A (TYP) I/P: 230 VAC I= 49 A/ 230 VAC 115V/ 30 A(TYP) I/P: 115 VAC I= 25 A/ 115 VAC COLD START 6 LEAKAGE CURRENT < < 150 100 P O/P:FULL LOAD Ta:25℃ uA/ 264 VAC I/P: 264 VAC L-FG: 117 uA for N-FG: 117 uA leakage current O/P:Min LOAD Ta:25℃ uA/264 VAC I/P: 264 VAC L-V+ : 79 uA for O/P:Min LOAD Ta:25℃ L-V-: 79 uA N-V+: 79 uA N-V-: 79 uA earth touch leakage current P P PROTECTION FUNCTION TEST NO 1 TEST ITEM SPECICATION TEST CONDITION OVER LOAD PROTECTION 115%~ 150% I/P: 230 VAC I/P: 115 VAC O/P:TESTING Ta:25℃ 2 OVER VOLTAGE PROTECTION CH1: 5.75V~ 6.75V I/P: 230 VAC I/P: 115 VAC 3 SHORT PROTECTION RESULT VERDICT 131 %/ 230 VAC 125 %/ 115 VAC P Hiccup Mode 6.4 V/ 230 VAAC 6.6 V/ 115 VAC O/P:MIN LOAD Ta:25℃ Shunt down Re- power ON SHORT EVERY OUTPUT I/P: 264 VAC NO DAMAGE 1 HOUR NO DAMAGE O/P:FULL LOAD Ta:25℃ Hiccup Mode Model: RPT-60C P P Test Report 2 / 5 RPT-60 series 60W Triple Output Medical Type ENVIRONMENT TEST NO 1 TEST ITEM SPECICATION TEMPERATURE RISE TEST MODEL:RPT-60A TEST CONDITION RESULT VERDICT 1. ROOM AMBIENT BURN-IN:1HRS I/P: 230VAC O/P: FULL LOAD Ta= 29.5 ℃ 2. HIGH AMBIENT BURN-IN:2 HRS I/P: 230VAC O/P: FULL LOAD Ta= 39.6℃ 2 3 OVER LOAD BURN-IN TEST LOW TEMPERATURE ROOM AMBIENT HIGH AMBIENT Ta= 29.5 ℃ Ta= 39.6 ℃ 56.5℃ 66.6℃ NO Position P/N 1 2 LF1 BD1 LF158 US4K 4A/800V SHI 59.3℃ 68.6℃ 3 C5 100U/400V NCC 105℃ 57.3℃ 67.8℃ 4 T1 COIL TF-1419 LS 76.4℃ 87.3℃ 5 C36 47U/50V RUB 105℃ YXF 68.4℃ 79.2℃ 6 ZD1 P6KE100A PAN 73.7℃ 84.4℃ 7 Q1 2SK2545 6A/600V TOS 95.4℃ 106.2℃ 8 D1 EGP20J 2A/600V ZOW 86.4℃ 97.6℃ 9 U1 1230D 63.0℃ 76.2℃ 10 D100 MBR20H100CT 20A/100V 102.6℃ 112.6℃ 11 D200 YG805C10R 20A/100V FUJI 93.8℃ 103.0℃ 12 D300 31DQ10 3A/100V 93.4℃ 101.6℃ 13 C105 1800U/10V RUB 105℃ ZLH 81.1℃ 91.6℃ 14 C205 390U/35V RUB 105℃ ZLH 83.9℃ 94.4℃ 15 C305 470U/16V NCC 105℃ KY 72.6℃ 84.4℃ TEST:OK NO DAMAGE I/P: 230 VAC 1 HOUR ( MIN ) O/P: V1=5.7A V2=2.5A V3=0.55A Ta:25℃ TURN ON AFTER 2 HOUR TURN ON TEST P I/P: 230 VAC O/P: 100﹪LOAD P TEST:OK P Ta= -20℃ 4 AFTER 12 HOURS I/P: 272 VAC HIGH TEMPERATURE IN CHAMBER ON CONTROL 40℃ O/P:FULL LOAD Ta= 40℃ NO DAMAGE + 0.03 %(0~50℃) HUMIDITY= HIGH VOLTAGE TURN ON TEST 5 TEMPERATURE VIBRATION TEST P 95 %R.H I/P: 230 VAC + 0.01%(0~50℃) P O/P:FULL LOAD COEFFICIENT 6 TEST:OK HIGH HUMIDITY 1 Carton & 1 Set TEST:OK (1) Waveform: Sine Wave (2) Frequency:10~500Hz (3) Sweep Time:10min/sweep cycle P (4) Acceleration:2G (5) Test Time:1 hour in each axis (X.Y.Z) (6) Ta:25℃ Model: RPT-60C Test Report 3 / 5 RPT-60 series 60W Triple Output Medical Type SAFETY TEST NO 1 2 3 4 TEST ITEM SPECICATION TEST CONDITION RESULT WITHSTAND VOLTAGE I/P-O/P: 4 KVAC/min I/P-O/P: 4.2 KVAC/min I/P-O/P: 2.11 mA I/P-FG: 2 KVAC/min I/P-FG: 2.4 KVAC/min I/P-FG: 1.46 mA O/P-FG: 1.5 KVAC/min O/P-FG: 1.8 KVAC/min Ta:25℃ O/P-FG: 1.46 mA I/P-O/P:500VDC>100MΩ I/P-O/P: 500 VDC I/P-O/P: 30 GΩ I/P-FG: 500VDC>100MΩ I/P-FG: 500 VDC O/P-FG:500VDC>100MΩ O/P-FG: 500 Ta:25℃ VDC I/P-FG: 30 GΩ O/P-FG: 30 GΩ ISOLATION RESISTANCE GROUNDING CONTINUITY FG(PE) TO CHASSIS APPROVAL OR TRACE < 100 mΩ TUV: Certificate NO :R 50112371 VERDICT P NO DAMAGE P NO DAMAGE 40 A / 2min Ta:25℃ 3 mΩ P P UL: File NO :E227340 E.M.C TEST NO 1 2 3 4 TEST ITEM SPECICATION TEST CONDITION RESULT HARMONIC EN61000-3-2 I/P: 230 VAC/50HZ PASS CLASS A O/P:FULL LOAD Ta:25℃ EN55022 EN55011 I/P: 230 VAC (50HZ) PASS CLASS B O/P:FULL/50% LOAD Ta:25℃ Test by certified Lab CONDUCTION RADIATION E.S.D 5 E.F.T EN55022 EN55011 I/P: 230 VAC (50HZ) PASS O/P:FULL LOAD Ta:25℃ Test by certified Lab EN61000-4-2 MEDICAL I/P: 230 VAC/50HZ CRITERIA A EN61000-4-4 MEDICAL SURGE P P O/P:FULL LOAD Ta:25℃ I/P: 230 VAC/50HZ P CRITERIA A O/P:FULL LOAD Ta:25℃ INPUT: 2KV 6 P CLASS B AIR:8KV / Contact:6KV VERDICT IEC61000-4-5 MEDICAL I/P: 230 VAC/50HZ P CRITERIA A O/P:FULL LOAD Ta:25℃ L-N :1KV P L,N-PE:2KV 7 Test by certified Lab & Test Report Prepare M.T.B.F & LIFE CYCLE CALCULATION NO 1 TEST ITEM SPECICATION CAPACITOR SUPPOSE C105 IS THE MOST CRITICAL COMPONENT I/P: 230VAC O/P:FULL LOAD Ta= 25 ℃ LIFE TIME= 77256 HRS LIFE CYCLE I/P: 230VAC 2 MTBF TEST CONDITION O/P:FULL LOAD Ta= 50 ℃ RESULT VERDICT P LIFE TIME= 26543 HRS MIL-HDBK-217F NOTICES2 PARTS COUNT P TOTAL FAILURE RATE: 677.8K HRS Model: RPT-60C Test Report 4 / 5 RPT-60 series 60W Triple Output Medical Type COMPONENT STRESS TEST NO 1 TEST ITEM SPECICATION TEST CONDITION RESULT Power Transistor Q1 Rated I/P:High-Line +3V = 267 V (1) 488 V ( D to S) or (C to E) Peak Voltage K2545 : 600V O/P: (1)Full Load Turn on (2) 518 V 6A VERDICT P (2) Output Short Ta:25℃ 2 Diode Peak Voltage D100 Rated I/P:High-Line +3V = 267 V (1) 54 V MBR20H100 : 100V 20A O/P: (1)Full Load Turn on (2) 51 V D200 Rated (1) 107 V BYQ28X-200 : 200V 10 A (2) 124 V D300 Rated (1) 226 V HER305 : 400V 3A (2) 221 V (1) 468 V (2)Output Short Ta:25℃ 3 Clamp Diode Peak Voltage D1 Rated I/P:High-Line +3V = 267 V EGP20J : 600V 2A O/P: (1) Dynamic Load P 90%Duty/1KHz P Ta:25℃ 4 Input Capacitor Voltage C5 Rated : 100u / 400V/ 105℃ I/P:High-Line +3V = 267 V (1) 370 V O/P: (1)Full Load Turn on /Off (2) 370 V (2) Min load Turn on /Off (3) 370 V P (3)Full Load /Min load Change Ta:25℃ 5 Control IC Voltage Test U1 Rated I/P:High-Line +3V = 267 V (1) 12.65 V 1230D : 18V O/P: (1)Full Load Turn on /Off (2) 11.86 V (2) Min load Turn on /Off (3) 12.65 V P (3)Full Load /Min load Change Ta:25℃ DATE 2006/4/3 2006/6/2 2006/8/16 SAMPLE TEST RESULT TESTER APPROVAL RD SAMPLE PASS VINCENT TSENG MAX LIN PASS VINCENT TSENG MAX LIN PASS VINCENT TSENG MAX LIN PRODUCT SAMPLE W0605A33 PRODUCT SAMPLE W0607C20 2003/12/12 Model: RPT-60C A50-F023 Test Report 5 / 5