RPS-60 series

RPS-60
60W Single Output Medical Type
series
■ Features :
4"×2" miniature size
Universal AC input / Full range
Low leakage current <100 μA
Protections: Short circuit / Overload / Over voltage
Cooling by free air convection
Medical safety approved (2 x MOPP between primary to secondary)
UL60950-1/IEC60950-1/EN60950-1 ITE safety approved
No load power consumption<0.75W
Fixed switch frequency at 100KHz
Suitable for BF application with appropriate system consideration
3 years warranty
..
SPECIFICATION
MODEL
RPS-60-3.3
RPS-60-5
RPS-60-12
RPS-60-15
RPS-60-24
RPS-60-48
DC VOLTAGE
3.3V
5V
12V
15V
24V
48V
RATED CURRENT
10A
10A
5A
4A
2.5A
1.25A
CURRENT RANGE
0 ~ 11A
0 ~ 11A
0 ~ 5.5A
0 ~ 4.4A
0 ~ 2.75A
0 ~ 1.375A
33W
50W
60W
60W
60W
60W
55W
66W
66W
66W
66W
RIPPLE & NOISE (max.) Note.2 80mVp-p
80mVp-p
120mVp-p
150mVp-p
240mVp-p
300mVp-p
VOLTAGE ADJ. RANGE
3.1 ~ 3.6V
VOLTAGE TOLERANCE Note.3 ±2.0%
4.75 ~ 5.5V
±2.0%
11.4 ~ 13.2V
±2.0%
13.5 ~ 16.5V
±2.0%
22.8 ~ 27.6V
±1.0%
45.6 ~ 52.8V
±1.0%
LINE REGULATION
±0.5%
±0.5%
±0.5%
±0.5%
±0.5%
±0.5%
LOAD REGULATION
±1.0%
±1.0%
±1.0%
±1.0%
±1.0%
±1.0%
SETUP, RISE TIME
500ms, 30ms/230VAC
HOLD UP TIME (Typ.)
50ms/230VAC
VOLTAGE RANGE
90 ~ 264VAC
FREQUENCY RANGE
47 ~ 63Hz
EFFICIENCY (Typ.)
74%
84%
85%
86%
AC CURRENT (Typ.)
INRUSH CURRENT (Typ.)
1.8A/115VAC
28.4 ~ 32.4V
55.2 ~ 64.8V
RATED POWER
PEAK LOAD(10sec.)
OUTPUT
INPUT
LEAKAGE CURRENT
Note.4 36.3W
500ms, 30ms/115VAC at full load
12ms/115VAC at full load
127 ~ 370VDC
79%
83%
1 A/230VAC
COLD START 60A/230VAC
30A/115VAC
Note.8 Earth leakage current < 150 μA/264VAC , Touch current < 100μ A/264VAC
OVER LOAD
PROTECTION
OVER VOLTAGE
WORKING TEMP.
115 ~ 150% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
3.8 ~ 5V
5.75 ~ 6.75V
13.8 ~ 16.2V
17.25 ~ 20.25V
Protection type : Shut down o/p voltage, re-power on to recover
-20 ~ +70℃ (Refer to "Derating Curve")
WORKING HUMIDITY
ENVIRONMENT STORAGE TEMP., HUMIDITY
TEMP. COEFFICIENT
VIBRATION
20 ~ 90% RH non-condensing
SAFETY STANDARDS
SAFETY & ISOLATION LEVEL
WITHSTAND VOLTAGE
EMC
UL60950-1, TUV EN60950-1, ANSI/AAMI ES60601-1, TUV EN60601-1, IEC60601-1 approved
(Note 5)
OTHERS
ISOLATION RESISTANCE
EMC EMISSION
10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
Primary-Secondary: 2xMOPP, Primary-Earth:1xMOPP, Secondary-Earth:1xMOPP
I/P-O/P:4KVAC I/P-FG:2KVAC O/P-FG:1.5KVAC
I/P-O/P, I/P-FG, O/P-FG:100M Ohms / 500VDC/ 25℃/ 70% RH
Compliance to EN55011 (CISPR11), EN55022 (CISPR22) Class B, EN61000-3-2,-3
EMC IMMUNITY
Compliance to EN61000-4-2,3,4,5,6,8,11, EN55024, EN60601-1-2, EN61204-3, medical level, criteria A
MTBF
353.6K hrs min. MIL-HDBK-217F (25℃)
101.6*50.8*29mm (L*W*H)
0.15Kg; 96pcs/15.4Kg/0.89CUFT
DIMENSION
PACKING
NOTE
-40 ~ +85℃, 10 ~ 95% RH
±0.03%/℃ (0 ~ 45℃)
1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25℃ of ambient temperature.
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uf & 47uf parallel capacitor.
3. Tolerance : includes set up tolerance, line regulation and load regulation.
4. 33% Duty cycle maximum within every 30 seconds. Average output power should not exceed the rated power.
5. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets
EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”
(as available on http://www.meanwell.com)
6. Length of set up time is measured at cold first start. Turning ON/OFF the power supply may lead to increase of the set up time.
7. Heat Sink HS1,HS2 can not be shorted.
8. Touch current was measured from primary input to DC output.
File Name:RPS-60-SPEC 2015-07-17
RPS-60
60W Single Output Medical Type
Mechanical Specification
series
Unit:mm
101.6
3.175
95.25
50.8
44.45
FS1
3.175
HS2
AC FUSE
T2.5/250V
3
2
1
CN1
SVR1
1
2
3
4
FS2
AC FUSE
T2.5/250V
CN2
HS1
M1
M2
3 max.
29
4-ψ3.3
DC Output Connector (CN2) : JST B4P-VH or equivalent
AC Input Connector (CN1) : JST B3P-VH or equivalent
Pin No.
1
2
3
Assignment
AC/N
No Pin
AC/L
Mating Housing
Pin No.
1,2
3,4
Terminal
JST VHR
or equivalent
JST SVH-21T-P1.1
or equivalent
Mating Housing
Terminal
JST VHR
or equivalent
JST SVH-21T-P1.1
or equivalent
: Grounding Required
1.HS1,HS2 cannot be shorted.
2.M1 is safety ground. For better EMC performance,
Please secure an electrical connection between
M1,M2 and chassis grounding.
Block Diagram
RECTIFIERS
&
FILTER
EMI
FILTER
I/P
Assignment
+V
-V
fosc : 100KHz
RECTIFIERS
&
FILTER
POWER
SWITCHING
+V
-V
DETECTION
CIRCUIT
FG
O.L.P.
CONTROL
O.V.P.
Output Derating
Output Derating VS Input Voltage
100
100
90
80
80
70
LOAD (%)
LOAD (%)
60
40
20
-20
0
10
20
30
40
50
AMBIENT TEMPERATURE (℃)
60 65 70 (HORIZONTAL)
60
50
40
90
95
100
115
120
140
160
180
200
220
240
264
INPUT VOLTAGE (VAC) 60Hz
File Name:RPS-60-SPEC 2015-07-17