RPD-75 75W Dual Output Medical Type series Features : Universal AC input/Full range Low leakage current<200uA Protections: Short circuit / Overload / Over voltage Free air convection for rated power and 23.5CFM forced air convection for peak load UL60601-1/IEC60601-1/EN60601-1 medical safety approved Fixed switching frequency at 65KHz 3 years warranty EN60601-1 IEC60601-1 SPECIFICATION OUTPUT RPD-75B RPD-75A MODEL OUTPUT NUMBER CH1 CH2 CH1 CH2 DC VOLTAGE 5V 12V 5V 24V RATED CURRENT 7A 3A 5A 2A CURRENT RANGE 1 ~ 9.5A 0.3 ~ 4A 1 ~ 6.8A 0.2 ~ 2.7A RATED POWER 71W PEAK LOAD (23.5CFM) 95.5W 73W 98.8W 120mVp-p RIPPLE & NOISE (max.) Note.2 80mVp-p VOLTAGE ADJ. RANGE CH1: 4.75 ~ 5.5V VOLTAGE TOLERANCE Note.3 2.0% INPUT 0.5% 1.0% LOAD REGULATION 1.5% 3.0% 1.5% 3.0% SETUP, RISE TIME 500ms, 30ms/230VAC HOLD UP TIME (Typ.) 80ms/230VAC VOLTAGE RANGE 90 ~ 264VAC FREQUENCY RANGE 47 ~ 63Hz 77% EFFICIENCY (Typ.) 20ms/115VAC at full load 127 ~ 370VDC 79% 1.5A/115VAC 1A/230VAC COLD START 25A/115VAC 50A/230VAC Earth leakage current <200uA / 264VAC, Patient leakage current <100uA/264VAC 140 ~ 180% rated output power Protection type : Hiccup mode, recovers automatically after fault condition is removed OVER VOLTAGE CH1: 5.75 ~ 6.75V Protection type : Shut down o/p voltage, re-power on to recover WORKING TEMP. -20 ~ +70 TEMP. COEFFICIENT VIBRATION SAFETY STANDARDS WITHSTAND VOLTAGE (Refer to output load derating curve) 20 ~ 90% RH non-condensing -40 ~ +85 , 10 ~ 95% RH 0.03%/ (0 ~ 45 ) 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes UL60601-1, TUV EN60601-1, IEC60601-1 approved I/P-O/P:4KVAC I/P-FG:1.5KVAC O/P-FG:1.5KVAC I/P-O/P, I/P-FG, O/P-FG:100M Ohms / 500VDC / 25 / 70% RH EMI CONDUCTION & RADIATION Compliance to EN55011 (CISPR11), EN55022 (CISPR22) Class B Compliance to EN61000-3-2,-3 HARMONIC CURRENT EMS IMMUNITY Compliance to EN61000-4-2,3,4,5,6,8,11; ENV50204, EN55024, EN60601-1-2, EN61000-6-2, EN61204-3, heavy industry level, EN61204-3 medical level, criteria A ISOLATION RESISTANCE MTBF NOTE 500ms, 30ms/115VAC at full load OVERLOAD WORKING HUMIDITY ENVIRONMENT STORAGE TEMP., HUMIDITY OTHERS 6.0% 1.0% PROTECTION (Note 4) 6.0% 200mVp-p 0.5% LEAKAGE CURRENT EMC CH1: 4.75 ~ 5.5V 2.0% LINE REGULATION AC CURRENT (Typ.) INRUSH CURRENT (Typ.) SAFETY & 80mVp-p DIMENSION 569.9K hrs min. MIL-HDBK-217F (25 127*76.2*31mm (L*W*H) PACKING 0.25Kg; 63pcs/16.7Kg/1.35CUFT ) 1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature. 2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uf & 47uf parallel capacitor. 3. Tolerance : includes set up tolerance, line regulation and load regulation. 4. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets EMC directives. For guidance on how to perform these EMC tests, please refer to EMI testing of component power supplies. (as available on http://www.meanwell.com) 5. Length of set up time is measured at cold first start. Turning ON/OFF the power supply may lead to increase of the set up time. 6. Heat Sink HS1,HS2,HS3 can not be shorted. File Name:RPD-75-SPEC 2011-03-31 RPD-75 75W Dual Output Medical Type Mechanical Specification series Unit:mm FAN 5.7 15cm 23.5CFM min. HS3 CN2 3 2 1 64.75 76.2 1 2 3 4 5 6 7 8 HS1 CN1 HS2 AC Input Connector (CN1) : JST B3P-VH or equivalent Pin No. 1 2 3 AC FUSE T2.5/250V Assignment AC/N No Pin AC/L Mating Housing Terminal JST VHR or equivalent JST SVH-21T-P1.1 or equivalent FS2 FS1 4115.8 5.6 DC Output Connector (CN2) : JST B8P-VH or equivalent 3.8 Pin No. 1,2 3,4,5 6,7 8 127 Assignment V1 COM V2 NC Mating Housing Terminal JST VHR or equivalent JST SVH-21T-P1.1 or equivalent 3max. 31 HS1,HS2,HS3 can not be shorted Block Diagram fosc : 65KHz RECTIFIERS & FILTER EMI FILTER I/P POWER SWITCHING RECTIFIERS & FILTER +V2 RECTIFIERS & FILTER +V1 COM DETECTION CIRCUIT FG O.L.P. CONTROL O.V.P. Derating Curve Output Derating VS Input Voltage Ta=25 100 133 23.5CFM Forced air Cooling 100 90 80 Convection Cooling 70 LOAD (%) LOAD (%) 80 50 20 -20 0 10 20 30 40 50 AMBIENT TEMPERATURE ( ) 60 70 (HORIZONTAL) 60 50 40 90 95 100 115 120 140 160 180 200 220 240 264 INPUT VOLTAGE (VAC) 60Hz File Name:RPD-75-SPEC 2011-03-31