MSP-300 series

MSP-300
300W Single Output Medical Type
series
■ Features :
‧Universal AC input / Full range
‧Built-in active PFC function, PF>0.95
‧High efficiency up to 89%
‧Withstand 300VAC surge input for 5 seconds
‧Protections: Short circuit / Overload / Over voltage / Over temperature
‧Built-in constant current limiting circuit
‧1U low profile 41mm
‧Medical safety approved (MOOP level)
‧Built-in cooling fan ON-OFF control
‧Built-in DC OK signal
‧Built-in remote ON-OFF control
‧Standby [email protected]
‧Built-in remote sense function
‧No load power consumption<0.5W (Note.6)
‧5 years warranty
SPECIFICATION
MODEL
OUTPUT
MSP-300-3.3
MSP-300-5
MSP-300-7.5
MSP-300-12
MSP-300-15
MSP-300-24
MSP-300-36
MSP-300-48
DC VOLTAGE
3.3V
5V
7.5V
12V
15V
24V
36V
48V
RATED CURRENT
60A
60A
40A
27A
22A
14A
9A
7A
CURRENT RANGE
0 ~ 60A
0 ~ 60A
0 ~ 40A
0 ~ 27A
0 ~ 22A
0 ~ 14A
0 ~ 9A
0 ~ 7A
RATED POWER
198W
300W
300W
324W
330W
336W
324W
336W
RIPPLE & NOISE (max.) Note.2 80mVp-p
90mVp-p
100mVp-p
120mVp-p
150mVp-p
150mVp-p
250mVp-p
250mVp-p
VOLTAGE ADJ. RANGE
4.3 ~ 5.8V
±2.0%
6.8 ~ 9V
±2.0%
10.2 ~ 13.8V
±1.0%
13.5 ~ 18V
±1.0%
21.6 ~ 28.8V
±1.0%
28.8 ~ 39.6V
±1.0%
40.8 ~ 55.2V
±1.0%
2.8 ~ 3.8V
VOLTAGE TOLERANCE Note.3 ±2.5%
LINE REGULATION
±0.5%
±0.5%
±0.5%
±0.3%
±0.3%
±0.2%
±0.2%
±0.2%
LOAD REGULATION
±1.0%
±1.0%
±1.0%
±0.5%
±0.5%
±0.5%
±0.5%
±0.5%
SETUP, RISE TIME
1000ms, 50ms/230VAC
HOLD UP TIME (Typ.)
16ms/230VAC
88%
87%
88%
89%
16ms/115VAC at full load
VOLTAGE RANGE
Note.5 85 ~ 264VAC
FREQUENCY RANGE
47 ~ 63Hz
INPUT
PF>0.95/230VAC
EFFICIENCY (Typ.)
80%
AC CURRENT (Typ.)
INRUSH CURRENT (Typ.)
4.5A/115VAC
35A/115VAC
OVERLOAD
PROTECTION
FUNCTION
120 ~ 370VDC
POWER FACTOR (Typ.)
LEAKAGE CURRENT
OVER VOLTAGE
2500ms, 50ms/115VAC at full load
PF>0.99/115VAC at full load
82%
86%
88%
2.25A/230VAC
70A/230VAC
Earth leakage current < 450μA/264VAC , Touch leakage current < 100μA/264VAC
105 ~ 135% rated output power
Protection type : Constant current limiting, recovers automatically after fault condition is removed
6 ~ 7V
9.4 ~ 10.9V
14.4 ~ 16.8V 18.8 ~ 21.8V 30 ~ 34.8V
41.4 ~ 48.6V
3.96 ~ 4.62V
57.6 ~ 67.2V
Protection type : Shut down o/p voltage, re-power on to recover
OVER TEMPERATURE
5V STANDBY
Shut down o/p voltage, recovers automatically after temperature goes down
5VSB : [email protected] ; tolerance±5%, ripple : 50mVp-p(max.)
DC OK SIGNAL
PSU turns on : 3.3 ~ 5.6V ; PSU turns off : 0 ~ 1V
REMOTE CONTROL
RC+ / RC-: 4 ~ 10V or open = power on ; 0 ~ 0.8V or short = power off
FAN CONTROL (Typ.)
Load 35±15% or RTH2≧50℃ Fan on
WORKING TEMP.
-40 ~ +70 ℃ (Refer to "Derating Curve")
20 ~ 90% RH non-condensing
WORKING HUMIDITY
ENVIRONMENT STORAGE TEMP., HUMIDITY -40 ~ +85 ℃ , 10 ~ 95% RH
TEMP. COEFFICIENT
± 0.03%/ ℃ (0 ~ 50 ℃ )
SAFETY &
EMC
(Note 4)
OTHERS
VIBRATION
10 ~ 500Hz, 5G 10min./1cycle, 60min. each along X, Y, Z axes
SAFETY STANDARDS
ISOLATION LEVEL
ANSI/AAMI ES60601-1, IEC60601-1 approved
Primary-Secondary: 2×MOOP, Primary-Earth: 1×MOOP
WITHSTAND VOLTAGE
I/P-O/P:4KVAC
ISOLATION RESISTANCE
EMC EMISSION
I/P-O/P, I/P-FG, O/P-FG:100M Ohms / 500VDC / 25 ℃ / 70% RH
Compliance to EN55011 (CISPR11) Class B, EN61000-3-2,-3
EMC IMMUNITY
Compliance to EN61000-4-2,3,4,5,6,8,11, EN60601-1-2
MTBF
176Khrs min.
DIMENSION
199*105*41mm (L*W*H)
0.95Kg;15pcs/15.3Kg/0.69CUFT
PACKING
NOTE
I/P-FG:2KVAC
O/P-FG:0.5KVAC
MIL-HDBK-217F (25 ℃ )
1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25℃ of ambient temperature.
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uf & 47uf parallel capacitor.
3. Tolerance : includes set up tolerance, line regulation and load regulation.
4. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets
EMC directives. For guidance on how to perform these EMC tests, please refer to EMI testing of component power supplies.
(as available on http://www.meanwell.com)
5. Derating may be needed under low input voltages. Please check the derating curve for more details.
6. No load power consumption<0.5W when RC- & RC+ (CN100 pin4,6) 0 ~ 8V or short.
File Name:MSP-300-SPEC 2014-12-12
MSP-300
300W Single Output Medical Type
Mechanical Specification
Case No.980A
series
Unit:mm
4-M4 L=6
12.25
168.5
199
12.25
168.5
120
42.5
1
Terminal Pin No. Assignment
Pin No.
1
2
3
Assignment
AC/L
AC/N
FG
3
4
FAN
5
7
LED
CN100
SVR
Mating Housing
Terminal
HRS DF11-8DS
or equivalent
HRS DF11-**SC
or equivalent
167.5
12.25
167.5
12.25
17max.
4-M4 L=6
fosc : 70KHz
Block Diagram
ACTIVE
INRUSH
CURRENT
LIMITING
EMI
FILTER
I/P
Air flow
direction
6
Connector Pin No. Assignment (CN100) :
HRS DF11-8DP-2DS or equivalent
Pin No. Assignment
1
AUX
2
AUXG
3
DC-OK
4
RC5
GND
6
RC+
7
+S
8
-S
4-M4 L=5
2
Pin No.
Assignment
4,5
DC OUTPUT -V
6,7
DC OUTPUT +V
AUX
POWER
RECTIFIERS
&
FILTER
AUX
AUXG
+S
RECTIFIERS
&
PFC
POWER
SWITCHING
RECTIFIERS
&
FILTER
+V
-V
O.T.P.
O.L.P.
-S
DC-OK
DETECTION
CIRCUIT
FG
PFC
CONTROL
O.V.P.
PWM
CONTROL
RC+
REMOTE
CONTROL
Derating Curve
RC-
Output Derating VS Input Voltage
100
100
90
80
80
70
60
50
60
40
50
20
-40
40
0
10
20
30
40
50
AMBIENT TEMPERATURE (℃ )
60
70
(HORIZONTAL)
85
100
125
135
155
264
INPUT VOLTAGE (V) 60Hz
File Name:MSP-300-SPEC 2014-12-12
MSP-300
300W Single Output Medical Type
series
Function Description of CN100
Pin No. Function Description
Auxiliary voltage output, 4.75~5.25V, reference to pin 2(AUXG). The maximum load current is 0.3A. This output has the built-in oring
1
AUX diodes and is not controlled by the "remote ON/OFF control".
Auxiliary voltage output ground. The signal return is isolated from the output terminals (+V & -V).
DC-OK signal is a TTL level signal, referenced to pin5(DC-OK GND). High when PSU turns on.
2
3
4
5
6
AUXG
DC-OK
RCGND
RC+
7
+S
Turns the output on and off by electrical or dry contact between pin 4 (RC-), Short: Power OFF, Open: Power ON.
Positive sensing. The +S signal should be connected to the positive terminal of the load. The +S and -S leads should be twisted in pair to
minimize noise pick-up effect. The maximum line drop compensation is 0.5V.
8
-S
Negative sensing. The -S signal should be connected to the negative terminal of the load. The -S and +S leads should be twisted in pair to
minimize noise pick-up effect. The maximum line drop compensation is 0.5V.
Remote control ground.
This pin connects to the negative terminal(-V). Return for DC-OK signal output.
Function Manual
1.Remote Sense
The remote sensing compensates voltage drop on the
load wiring up to 0.5V.
CN100
1
2
3
4
5
1
7 CN100
6
AUX DC-OK GND
2 AUXG RC-
RC+
+S
7
-S
8
1 7
2 8
-S
(Pin8)
LOAD
-
+
Sense lines should
be tw isted in pairs
+S
(Pin7)
Fig 1.1
2.DC-OK Signal
DC-OK signal is a TTL level signal. High when PSU turns on.
Between DC-OK(pin6) and GND(pin4)
Output Status
3.3 ~ 5.6V
ON
0 ~ 1V
OFF
CN100
1
2
3
4
5
6
1
7 CN100
AUX DC-OK GND
2 AUXG RC-
RC+
+S
7
-S
8
1 7
2 8
Fig 2.1
File Name:MSP-300-SPEC 2014-12-12
MSP-300
300W Single Output Medical Type
series
3.Remote Control
The PSU can be turned ON/OFF by using the "Remote
ON/OFF" function
Between RC+(pin3) and RC-(pin5)
Output Status
SW ON (Short)
OFF
SW OFF (Open)
ON
CN100
1
2
3
4
5
6
7 CN100
1
AUX DC-OK GND
2 AUXG RC-
RC+
+S
7
-S
8
1 7
2 8
SW
Fig 3.1
File Name:MSP-300-SPEC 2014-12-12