MSP-300 300W Single Output Medical Type series ■ Features : ‧Universal AC input / Full range ‧Built-in active PFC function, PF>0.95 ‧High efficiency up to 89% ‧Withstand 300VAC surge input for 5 seconds ‧Protections: Short circuit / Overload / Over voltage / Over temperature ‧Built-in constant current limiting circuit ‧1U low profile 41mm ‧Medical safety approved (MOOP level) ‧Built-in cooling fan ON-OFF control ‧Built-in DC OK signal ‧Built-in remote ON-OFF control ‧Standby [email protected] ‧Built-in remote sense function ‧No load power consumption<0.5W (Note.6) ‧5 years warranty SPECIFICATION MODEL OUTPUT MSP-300-3.3 MSP-300-5 MSP-300-7.5 MSP-300-12 MSP-300-15 MSP-300-24 MSP-300-36 MSP-300-48 DC VOLTAGE 3.3V 5V 7.5V 12V 15V 24V 36V 48V RATED CURRENT 60A 60A 40A 27A 22A 14A 9A 7A CURRENT RANGE 0 ~ 60A 0 ~ 60A 0 ~ 40A 0 ~ 27A 0 ~ 22A 0 ~ 14A 0 ~ 9A 0 ~ 7A RATED POWER 198W 300W 300W 324W 330W 336W 324W 336W RIPPLE & NOISE (max.) Note.2 80mVp-p 90mVp-p 100mVp-p 120mVp-p 150mVp-p 150mVp-p 250mVp-p 250mVp-p VOLTAGE ADJ. RANGE 4.3 ~ 5.8V ±2.0% 6.8 ~ 9V ±2.0% 10.2 ~ 13.8V ±1.0% 13.5 ~ 18V ±1.0% 21.6 ~ 28.8V ±1.0% 28.8 ~ 39.6V ±1.0% 40.8 ~ 55.2V ±1.0% 2.8 ~ 3.8V VOLTAGE TOLERANCE Note.3 ±2.5% LINE REGULATION ±0.5% ±0.5% ±0.5% ±0.3% ±0.3% ±0.2% ±0.2% ±0.2% LOAD REGULATION ±1.0% ±1.0% ±1.0% ±0.5% ±0.5% ±0.5% ±0.5% ±0.5% SETUP, RISE TIME 1000ms, 50ms/230VAC HOLD UP TIME (Typ.) 16ms/230VAC 88% 87% 88% 89% 16ms/115VAC at full load VOLTAGE RANGE Note.5 85 ~ 264VAC FREQUENCY RANGE 47 ~ 63Hz INPUT PF>0.95/230VAC EFFICIENCY (Typ.) 80% AC CURRENT (Typ.) INRUSH CURRENT (Typ.) 4.5A/115VAC 35A/115VAC OVERLOAD PROTECTION FUNCTION 120 ~ 370VDC POWER FACTOR (Typ.) LEAKAGE CURRENT OVER VOLTAGE 2500ms, 50ms/115VAC at full load PF>0.99/115VAC at full load 82% 86% 88% 2.25A/230VAC 70A/230VAC Earth leakage current < 450μA/264VAC , Touch leakage current < 100μA/264VAC 105 ~ 135% rated output power Protection type : Constant current limiting, recovers automatically after fault condition is removed 6 ~ 7V 9.4 ~ 10.9V 14.4 ~ 16.8V 18.8 ~ 21.8V 30 ~ 34.8V 41.4 ~ 48.6V 3.96 ~ 4.62V 57.6 ~ 67.2V Protection type : Shut down o/p voltage, re-power on to recover OVER TEMPERATURE 5V STANDBY Shut down o/p voltage, recovers automatically after temperature goes down 5VSB : [email protected] ; tolerance±5%, ripple : 50mVp-p(max.) DC OK SIGNAL PSU turns on : 3.3 ~ 5.6V ; PSU turns off : 0 ~ 1V REMOTE CONTROL RC+ / RC-: 4 ~ 10V or open = power on ; 0 ~ 0.8V or short = power off FAN CONTROL (Typ.) Load 35±15% or RTH2≧50℃ Fan on WORKING TEMP. -40 ~ +70 ℃ (Refer to "Derating Curve") 20 ~ 90% RH non-condensing WORKING HUMIDITY ENVIRONMENT STORAGE TEMP., HUMIDITY -40 ~ +85 ℃ , 10 ~ 95% RH TEMP. COEFFICIENT ± 0.03%/ ℃ (0 ~ 50 ℃ ) SAFETY & EMC (Note 4) OTHERS VIBRATION 10 ~ 500Hz, 5G 10min./1cycle, 60min. each along X, Y, Z axes SAFETY STANDARDS ISOLATION LEVEL ANSI/AAMI ES60601-1, IEC60601-1 approved Primary-Secondary: 2×MOOP, Primary-Earth: 1×MOOP WITHSTAND VOLTAGE I/P-O/P:4KVAC ISOLATION RESISTANCE EMC EMISSION I/P-O/P, I/P-FG, O/P-FG:100M Ohms / 500VDC / 25 ℃ / 70% RH Compliance to EN55011 (CISPR11) Class B, EN61000-3-2,-3 EMC IMMUNITY Compliance to EN61000-4-2,3,4,5,6,8,11, EN60601-1-2 MTBF 176Khrs min. DIMENSION 199*105*41mm (L*W*H) 0.95Kg;15pcs/15.3Kg/0.69CUFT PACKING NOTE I/P-FG:2KVAC O/P-FG:0.5KVAC MIL-HDBK-217F (25 ℃ ) 1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25℃ of ambient temperature. 2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uf & 47uf parallel capacitor. 3. Tolerance : includes set up tolerance, line regulation and load regulation. 4. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets EMC directives. For guidance on how to perform these EMC tests, please refer to EMI testing of component power supplies. (as available on http://www.meanwell.com) 5. Derating may be needed under low input voltages. Please check the derating curve for more details. 6. No load power consumption<0.5W when RC- & RC+ (CN100 pin4,6) 0 ~ 8V or short. File Name:MSP-300-SPEC 2014-12-12 MSP-300 300W Single Output Medical Type Mechanical Specification Case No.980A series Unit:mm 4-M4 L=6 12.25 168.5 199 12.25 168.5 120 42.5 1 Terminal Pin No. Assignment Pin No. 1 2 3 Assignment AC/L AC/N FG 3 4 FAN 5 7 LED CN100 SVR Mating Housing Terminal HRS DF11-8DS or equivalent HRS DF11-**SC or equivalent 167.5 12.25 167.5 12.25 17max. 4-M4 L=6 fosc : 70KHz Block Diagram ACTIVE INRUSH CURRENT LIMITING EMI FILTER I/P Air flow direction 6 Connector Pin No. Assignment (CN100) : HRS DF11-8DP-2DS or equivalent Pin No. Assignment 1 AUX 2 AUXG 3 DC-OK 4 RC5 GND 6 RC+ 7 +S 8 -S 4-M4 L=5 2 Pin No. Assignment 4,5 DC OUTPUT -V 6,7 DC OUTPUT +V AUX POWER RECTIFIERS & FILTER AUX AUXG +S RECTIFIERS & PFC POWER SWITCHING RECTIFIERS & FILTER +V -V O.T.P. O.L.P. -S DC-OK DETECTION CIRCUIT FG PFC CONTROL O.V.P. PWM CONTROL RC+ REMOTE CONTROL Derating Curve RC- Output Derating VS Input Voltage 100 100 90 80 80 70 60 50 60 40 50 20 -40 40 0 10 20 30 40 50 AMBIENT TEMPERATURE (℃ ) 60 70 (HORIZONTAL) 85 100 125 135 155 264 INPUT VOLTAGE (V) 60Hz File Name:MSP-300-SPEC 2014-12-12 MSP-300 300W Single Output Medical Type series Function Description of CN100 Pin No. Function Description Auxiliary voltage output, 4.75~5.25V, reference to pin 2(AUXG). The maximum load current is 0.3A. This output has the built-in oring 1 AUX diodes and is not controlled by the "remote ON/OFF control". Auxiliary voltage output ground. The signal return is isolated from the output terminals (+V & -V). DC-OK signal is a TTL level signal, referenced to pin5(DC-OK GND). High when PSU turns on. 2 3 4 5 6 AUXG DC-OK RCGND RC+ 7 +S Turns the output on and off by electrical or dry contact between pin 4 (RC-), Short: Power OFF, Open: Power ON. Positive sensing. The +S signal should be connected to the positive terminal of the load. The +S and -S leads should be twisted in pair to minimize noise pick-up effect. The maximum line drop compensation is 0.5V. 8 -S Negative sensing. The -S signal should be connected to the negative terminal of the load. The -S and +S leads should be twisted in pair to minimize noise pick-up effect. The maximum line drop compensation is 0.5V. Remote control ground. This pin connects to the negative terminal(-V). Return for DC-OK signal output. Function Manual 1.Remote Sense The remote sensing compensates voltage drop on the load wiring up to 0.5V. CN100 1 2 3 4 5 1 7 CN100 6 AUX DC-OK GND 2 AUXG RC- RC+ +S 7 -S 8 1 7 2 8 -S (Pin8) LOAD - + Sense lines should be tw isted in pairs +S (Pin7) Fig 1.1 2.DC-OK Signal DC-OK signal is a TTL level signal. High when PSU turns on. Between DC-OK(pin6) and GND(pin4) Output Status 3.3 ~ 5.6V ON 0 ~ 1V OFF CN100 1 2 3 4 5 6 1 7 CN100 AUX DC-OK GND 2 AUXG RC- RC+ +S 7 -S 8 1 7 2 8 Fig 2.1 File Name:MSP-300-SPEC 2014-12-12 MSP-300 300W Single Output Medical Type series 3.Remote Control The PSU can be turned ON/OFF by using the "Remote ON/OFF" function Between RC+(pin3) and RC-(pin5) Output Status SW ON (Short) OFF SW OFF (Open) ON CN100 1 2 3 4 5 6 7 CN100 1 AUX DC-OK GND 2 AUXG RC- RC+ +S 7 -S 8 1 7 2 8 SW Fig 3.1 File Name:MSP-300-SPEC 2014-12-12