Eclipse© Aurora

®
AURORA
TM
Guided Wave Radar Level Transmitter
and Magnetic Level Indicator
DESCRIPTION
FULL REDUNDANCY
Aurora combines the operation of a conventional float operated Magnetic level indicator with the leading edge technology of Guided Wave Radar. The result is a true level measurement redundancy in a single 3" or 4" chamber design.
The Eclipse Guided Wave Radar is a 2-wire loop powered
24 V DC liquid level transmitter utilizing Time Domain
Reflectometry technology (TDR) to perform level measurement independent from media characteristics and process
conditions. The AuroraTM is a completely self-contained unit
for side mounting to a tank or vessel with threaded or
flanged pipe connections.
Flag-type
indication
F E AT U R E S
Shuttle-type
indication
• Complete redundant system whereby the measuring
results of the Eclipse can be continuously checked
against the level indication of the Magnetic Level
Indicator.
• Pro-active maintenance can be planned ahead of time
based upon the comparison of the measuring results of
the two systems.
• No calibration required on either measuring system.
• 2-wire, intrinsically safe loop powered level transmitter.
• HART®, AMS®, Foundation Fieldbus and PACTware communication protocol.
• Up to 5,7 m (224") measuring range.
• Up to 103 bar (1500 psi) – optional up to 310 bar (4500 psi).
• Up to +400 °C (+750 °F) process temperature - non condensing applications.
• Up to 155 bar @ +345 °C (2250 psi @ +650 °F) for saturated steam applications.
• Suited for SIL 1 and SIL 2 loops (full FMEDA report available for Eclipse transmitter) – optional SIL 2/3.
FE T Y I
NT
SA
EGRITY
A G E N C Y A P P R O VA L S
Approvals
ATEX II 3 G EEx nA II T6, non sparking
ATEX II 1 G EEx ia IIC T4, intrinsically safe ➀
ATEX II 1/2 G D EEx d[ia] IIC T6, explosion proof
Stoomwezen Secondary level safety device for steamdrums
TÜV
WHG § 19, overfill prevention
AIB
VLAREM II – 5.17.7
FM/CSA➁
Non Incendive / Intrinsically safe / Explosion proof
LRS
Lloyds Register of Shipping (marine applications)
Russian Authorisation Standards
GOST/
GOSGORTECHNADZOR➁
V EL
LE
Agency
ATEX
A P P L I C AT I O N S
MEDIA: Clean liquids; hydrocarbons to water-based media
(dielectric 1.4-100).
INTERFACE: Consult factory.
VESSELS: Most process or storage vessels up to rated
probe temperature and pressure.
CONDITIONS: All level measurement and control applications
including process conditions exhibiting visible vapors, foam,
surface agitation, bubbling or boiling, high fill/empty rates, low
level and varying dielectric media.
➀ Fisco ATEX, intrinsically safe for units with Fieldbus Foundation
➁ Consult factory for proper partnumbers
Worldwide level and flow solutions
P R I N C I P L E O F O P E R AT I O N
The coaxial GWR probe is separated from the float by
means of a baffle plate. The baffle plate assures the free
travel of the float by rising and falling level. The propagation
and reflection of high frequency pulses to determine liquid
level happens within the coaxial probe and is not impeded
by the magnetic field of the magnets mounted inside the
float of the indicator. These float magnets couple with the
individual flags as level moves up and down within the pipe
column.
Coaxial
GWR probe
non-magnetic
chamber wall
Liquid
level
Indicating
flags
Float
magnet
Float
Baffle
plate
GWR
probe
Float
Top View
ELECTRICAL WIRING
Power supply
GP / intrinsically safe / Explosion proof: min 11 V DC
0%
100 %
Standard shielded twisted cable (recommended but not
needed when wired as per NAMUR NE 21 for field strengths
up to 10 V/m).
Galvanic Barrier
max 28,6 V DC @ 94 mA
for intrinsically safe units
max 17,5 V DC @ 380 mA for Foundation Fieldbus units
(not needed for GP Dust Ex and Explosion proof models)
Hart® key
Ex
2
Non Ex
OPTIONS
Hot
Magnetic level indicators can be
ordered with several different tracing
options to heat the external chamber.
Tracing systems require that a special
blanket (custom designed to meet the
customers specifications) cover the
entire chamber and tracing equipment.
Cold
To facilitate operation where the
product is kept cold via chillers,
refrigerants and condensers, a frost
extension option and low temperature insulation are offered. The frost
extension is constructed of durable
acrylic plastic and suitable for contact with media as cold as -200 °C
(-320 °F) such as liquid nitrogen.
Consult factory for ordering information.
S E L E C T I O N D ATA
A complete measuring system consists of:
1. Order code for the Aurora - Magnetic Level Indicator
2. Order code for the Aurora - Guided Wave Radar Transmitter
3. Specify separately
- operating and min S.G. of the media
- max operating pressure and process temperature.
3
1. Order code for the Aurora - Magnetic Level Indicator
BASIC MODEL NUMBER
Min. S.G.
Code
B
B
B
B
B
B
B
B
B
B
G
G
G
G
G
G
G
G
G
G
A
B
C
D
E
F
G
H
J
K
40 °C
(100 °F)
27,6 (400)
41,3 (600)
55,2 (800)
41,3 (600)
51,7 (750)
75,8 (1100)
75,8 (1100)
62,0 (900)
103 (1500)
103 (1500)
0,75
0,65
0,50
0,76
0,76
0,65
0,50
0,75
0,65
0,50
Pressure rating in bar (psi)
200 °C
315 °C
(400 °F)
(600 °F)
26,6 (386)
23,4 (340)
24,3 (352)
18,1 (262)
32,3 (469)
24,1 (349)
39,9 (579)
35,2 (510)
49,9 (723)
43,9 (637)
44,5 (645)
33,1 (480)
44,5 (645)
33,1 (480)
59,9 (868)
52,8 (765)
60,7 (880)
45,2 (655)
60,7 (880)
45,2 (655)
400 °C
(750 °F)
22,2 (322)
4,1 (60)
5,5 (80)
33,3 (483)
41,6 (603)
7,6 (110)
7,6 (110)
49,9 (724)
10,3 (150)
10,3 (150)
float
mat.
cage
size
316 SST
Titanium
Titanium
316 SST
316 SST(*)
Titanium
Titanium
316 SST(*)
Titanium(*)
Titanium(*)
3"
3"
4"
3"
3"
3"
4"
4"
4"
4"
(*) pressurised float
MATERIALS OF CONSTRUCTION
Code with
Code with
scale in %
scale in cm
of span
B
C
E
F
H
J
L
M
Code
flags only
A
D
G
K
Flanges
Cage
Indication rail
Carbon steel
316/316L SST (1.4401/1.4404)
Carbon steel
316/316L SST (1.4401/1.4404)
316/316L SST
(1.4401/1.4404)
Aluminium
316 SST (1.4401)
CAGE AND FLANGE RATING
A
B
C
D
E
F
150 lbs
300 lbs
600 lbs
900 lbs
1500 lbs
2500 lbs (max 345 bar (5000 psi))
PROCESS CONNECTION – SIZE
2
1"
3
1 1/2"
4
2"
1
2
3
4
5
6
7
PN
PN
PN
PN
PN
PN
PN
B
C
D
16
25/40
63
100
160
250
320
EN 1092-1 Type B1
EN 1092-1 Type B1
EN 1092-1 Type B2
EN 1092-1 Type B2
DIN 2638 Form E
DIN 2628 Form E
DIN 2629 Form E
DN 25
DN 40
DN 50
DIN sizes only in combination with flanged process conn.
PROCESS CONNECTION - TYPE
A
B
D
F
1
Threaded NPT-F
Socket weld
ANSI RF Slip on flanges up to 600 lbs rating
ANSI RJ Weld Neck flanges for 600 lbs up to 2500 lbs rating
EN/DIN Weld Neck flanges
MEASURING RANGE (center-to-center)
English ranges (dimensions as per specified inch dimension)
0
0
0
0
0
0
0
0
0
0
A
B
C
D
E
14"
32"
48"
60"
72"
/
/
/
/
/
356
813
1219
1524
1829
mm
mm
mm
mm
mm
0
0
0
0
0
0
0
0
F
G
H
I
84"
96"
108"
120"
/
/
/
/
2134
2438
2743
3048
mm
mm
mm
mm
Metric ranges (specify per cm increments)
0 3 0
4 1 0
5 7 0
B G
4
minimum 30 cm (11.81")
maximum 410 cm (161") - for 7MS
maximum 570 cm (224") - for 7MD/7MR
complete order code for the Aurora - Magnetic Level Indicator
Dimensions in mm (inches)
1/2"
NPT plugged vent
1/2"
NPT plugged vent
E
E
Measuring
range
Measuring
range
254 (10)➀
1/2"
NPT plugged drain
1/2"
B
Flanged
➀
254 (10)➀
NPT plugged drain
B
Threaded/Welded
For S.G. < 0.8 and/or flange rating > 1500 lbs / PN 250, dimension will increase
Flanged 3" cages
Size
Connection
B
E
1"
NPT/SW - 150/300 lbs
69 (2.72)
178 (7.00)
1" 1/2"
NPT/SW - 150/300 lbs
81 (3.19)
178 (7.00)
2"
NPT/SW - 150/300 lbs
84 (3.31)
178 (7.00)
1"
RF Slip on flange - 150/300 lbs
170 (6.69)
178 (7.00)
1" 1/2"
RF Slip on flange - 150/300 lbs
180 (7.09)
178 (7.00)
2"
RF Slip on flange - 150/300 lbs
185 (7.28)
178 (7.00)
DN 25
PN 16/25/40 flange EN 1092-1 Type B1
170 (6.69)
178 (7.00)
DN 40
PN 16/25/40 flange EN 1092-1 Type B1
180 (7.09)
178 (7.00)
DN 50
PN 16 flange EN 1092-1 Type B1
185 (7.28)
178 (7.00)
DN 50
PN 25/40 flange EN 1092-1 Type B1
188 (7.40)
178 (7.00)
5
Flanged 4" cages
Size
Connection
B
E
1"
NPT/SW - 150/300/600 lbs
82 (3.23)
178 (7.00)
1" 1/2"
NPT/SW - 150/300/600 lbs
94 (3.70)
178 (7.00)
2"
NPT/SW - 150/300/600 lbs
97 (3.82)
178 (7.00)
1"
NPT - 900/1500 lbs
92 (3.62)
205 (8.07)
1" 1/2"
NPT - 900/1500 lbs
94 (3.70)
215 (8.46)
2"
NPT - 900/1500 lbs
97 (3.82)
225 (8.86)
1"
NPT - 2500 lbs
99 (3.90)
275 (10.83)
1" 1/2"
NPT - 2500 lbs
102 (4.02)
285 (11.22)
2"
NPT - 2500 lbs
112 (4.41)
295 (11.61)
1"
SW - 900/1500 lbs
99 (3.90)
205 (8.07)
1" 1/2"
SW - 900/1500 lbs
102 (4.02)
215 (8.46)
2"
SW - 900/1500 lbs
112 (4.41)
225 (8.86)
1"
SW - 2500 lbs
99 (3.90)
275 (10.83)
1" 1/2"
SW - 2500 lbs
102 (4.02)
285 (11.22)
2"
SW - 2500 lbs
112 (4.41)
295 (11.61)
1"
RF Slip on flange - 150/300/600 lbs
185 (7.28)
178 (7.00)
1" 1/2"
RF Slip on flange - 150/300/600 lbs
200 (7.87)
178 (7.00)
2"
RF Slip on flange - 150/300/600 lbs
200 (7.87)
178 (7.00)
1"
RJ Weld Neck flange - 600 lbs
185 (7.28)
178 (7.00)
1" 1/2"
RJ Weld Neck flange - 600 lbs
200 (7.87)
178 (7.00)
2"
RJ Weld Neck flange - 600 lbs
200 (7.87)
178 (7.00)
1"
RJ Weld Neck flange - 900 lbs
195 (7.68)
190 (7.48)
1" 1/2"
RJ Weld Neck flange - 900 lbs
205 (8.07)
190 (7.48)
2"
RJ Weld Neck flange - 900 lbs
225 (8.86)
190 (7.48)
1"
RJ Weld Neck flange - 1500 lbs
195 (7.68)
205 (8.07)
1" 1/2"
RJ Weld Neck flange - 1500 lbs
205 (8.07)
205 (8.07)
2"
RJ Weld Neck flange - 1500 lbs
225 (8.86)
205 (8.07)
1"
RJ Weld Neck flange - 2500 lbs
211 (8.31)
275 (10.83)
1" 1/2"
RJ Weld Neck flange - 2500 lbs
235 (9.25)
275 (10.83)
2"
RJ Weld Neck flange - 2500 lbs
250 (9.84)
275 (10.83)
DN 25
PN 16/25/40 flange EN 1092-1 Type B1
185 (7.28)
178 (7.00)
DN 40
PN 16/25/40 flange EN 1092-1 Type B1
200 (7.87)
178 (7.00)
DN 50
PN 16 flange EN 1092-1 Type B1
200 (7.87)
178 (7.00)
DN 50
PN 25/40 flange EN 1092-1 Type B1
203 (7.99)
178 (7.00)
DN 25
PN 63 flange EN 1092-1 Type B2
203 (7.99)
178 (7.00)
DN 40
PN 63 flange EN 1092-1 Type B2
217 (8.54)
178 (7.00)
DN 50
PN 63 flange EN 1092-1 Type B2
217 (8.54)
178 (7.00)
DN 25
PN 100 flange EN 1092-1 Type B2
203 (7.99)
178 (7.00)
DN 40
PN 100 flange EN 1092-1 Type B2
217 (8.54)
178 (7.00)
DN 50
PN 100 flange EN 1092-1 Type B2
223 (8.78)
178 (7.00)
DN 25
PN 160 DIN 2638 Form E
205 (8.07)
205 (8.07)
DN 40
PN 160 DIN 2638 Form E
219 (8.62)
205 (8.07)
DN 50
PN 160 DIN 2638 Form E
230 (9.06)
205 (8.07)
DN 25
PN 250 DIN 2628 Form E
212 (8.35)
205 (8.07)
DN 40
PN 250 DIN 2628 Form E
235 (9.25)
205 (8.07)
DN 50
PN 250 DIN 2628 Form E
240 (9.45)
205 (8.07)
DN 25
PN 320 DIN 2629 Form E
225 (8.86)
275 (10.83)
DN 40
PN 320 DIN 2629 Form E
243 (9.57)
275 (10.83)
DN 50
PN 320 DIN 2629 Form E
255 (10.04)
275 (10.83)
6
2. Order code for the Aurora - Eclipse 705 Guided Wave Radar transmitter
TRANSMITTER
7
7
7
7
Type ➀
Eclipse - blind transmitter
Eclipse - transmitter with digital display and keypad
Eclipse - blind transmitter
Eclipse - transmitter with digital display and keypad
3
4
5
6
➀
Signal output
4-20 mA with HART® communication
4-20 mA with HART® communication
Foundation Fieldbus® communication
Foundation Fieldbus® communication
Power
24 V DC
2-wire
loop powered
Standard electronics: SFF > 85 %. Consult factory for SIL enhanced electronics: SFF > 91 %
MOUNTING/CLASSIFICATION (Consult factory for FM/CSA approvals)
1
Integral, General purpose (&IS: FM/CSA)
2
Remote, General purpose (&IS: FM/CSA)
A
Integral, ATEX II 1 G EEx ia II C T4 - FISCO ATEX, intrinsically safe for units with Fieldbus Foundation
B
Integral, ATEX II 1 G EEx ia II C T4 - FISCO ATEX, intrinsically safe for units with Fieldbus Foundation
C
Integral, ATEX II 1/2 G D EEx d[ia] II C T6
D
Remote, ATEX II 1/2 G D EEx d[ia] II C T6
Integral, ATEX II 3 G EEx nA II T6
E
F
Remote, ATEX II 3 G EEx nA II T6
HOUSING
1
Cast aluminium dual compartiment, 3/4" NPT cable entry (2 entries – one plugged)
2
Cast aluminium dual compartiment, M20 x 1,5 cable entry (2 entries – one plugged)
3
Stainless steel dual compartiment, 3/4" NPT cable entry (2 entries – one plugged)
4
Stainless steel dual compartiment, M20 x 1,5 cable entry (2 entries – one plugged)
PROBE TYPE - all coaxial type, overfill safe GWR probes
R
7MR - Overfill safe GWR probe
D
7MD - High Temp / High Pressure (HTHP) GWR probe
S
7MS - Saturated steam GWR probe
(dielectric range ≥ 1,4) - WHG approved
(dielectric range ≥ 2,0) - WHG approved
(dielectric range ≥ 10,0) - Stoomwezen approved
PROBE MATERIAL
A
316 / 316 L (1.4401/1.4404) stainless steel
B
Hastelloy C (2.4819)
not for “S” probe (7MS)
C
Monel (2.4360)
not for “S” probe (7MS)
PROCESS SEAL-MATERIAL ➀
For the 7MR GWR probe ➁
0
1
2
Viton® GFLT - for universal use / steam applications
min -40°C (-40 °F) / max +200 °C (+400 °F)
EPDM (Ethylene Propylene) - for e.g. caustic soda applications min -50°C (-60 °F) / max +125 °C (+250 °F)
min -40°C (-40 °F) / max +200 °C (+400 °F)
Kalrez 4079 - for aggressive media
➀
Consult factory for alternative seal materials
➁
For ammonia/chlorine applications use the 7MD GWR
probe.
For the 7MD GWR probe
N
Borosilicate - for non condensing applications
min -195 °C (-320 °F) / max +400 °C (+750 °F)
For the 7MS GWR probe
8
PEEK - for saturated steam applications
min -40 °C (-40 °F) / max +345 °C (+650 °F)
MEASURING RANGE (Select the same measuring range as per page 4)
English ranges (dimensions as per specified inch dimension)
84" / 2134
0 0 F
0 0 A
14" / 356 mm
G
96"
/ 2438
0
0
0 0 B
32" / 813 mm
0
0
H
108"
/ 2743
0 0 C
48" / 1219 mm
0
0
I
120"
/ 3048
0 0 D
60" / 1524 mm
0 0 E
mm
mm
mm
mm
72" / 1829 mm
Metric ranges (specify per cm increments)
0 3 0
4 1 0
5 7 0
7
minimum 30 cm (11.81")
maximum 410 cm (161") - for 7MS
maximum 570 cm (224") - for 7MD/7MR
complete order code for the Aurora - Eclipse 705 Guided Wave Radar transmitter
Viton® is a registered trademark of Dupont Performance Elastomers
7
ECLIPSE 705 TRANSMITTER
Dimensions in mm (inches)
83
(3.28)
105
(4.12)
83
(3.28)
105
(4.12)
102
(4.00)
102
(4.00)
256
(10.08)
256
(10.08)
45°
45°
2 cable
entries
2 cable
entries
168 (6.61)
Mounting
Flange
7MD: 277 (10.90)
7MS: 240 ( 9.45)
Mounting
Flange
Probe
Insertion
Length
Probe
Insertion
Length
7MD/7MS
with flanged connection
7MR
with flanged connection
83
(3.28)
105
(4.12)
45° View
102
(4.00)
70
(2.75)
2 cable
entries
76
(3.00)
51
(2.00)
89
(3.50)
95
(3.75)
45°
214
(8.43)
840
(33.00)
2 holes
126
(4.94)
108
(4.25)
Remote Eclipse housing
111
(4.38)
Eclipse Housing,
(45° View)
8
T R A N S M I T T E R S P E C I F I C AT I O N S
FUNCTIONAL/PHYSICAL
Description
Power (at terminals)
Specification
General Purpose / ATEX Intrinsically Safe: 11 to 28,6 V DC
ATEX Explosion Proof (with Intrinsically Safe probe) 11 to 36 V DC
Foundation Fieldbus (FISCO ATEX Exi): 9 to 17,5 V DC
Foundation Fieldbus (General purpose & Exd): 9 to 32 V DC
4-20 mA with HART®, 3,8 mA to 20,5 mA useable (meets NAMUR NE 43) or
Foundation Fieldbus H1 (ITK Ver. 4)
Signal Output
Span
300 to 5700 mm (12 to 224") (7MS: max 4100 mm (161"))
Resolution
Analog: 0,01 mA
Display: 0,1 cm (inch)
630 Ω @ 20,5 mA - 24 V DC
Loop Resistance
Damping
Diagnostic Alarm
User Interface
Display
Menu Language
Housing Material
Approvals
SIL (Safety Integrity
Level)
Adjustable 0-10 s
Adjustable 3,6 mA, 22 mA, HOLD
3-button keypad and/or HART® communicator, Foundation Fieldbus, AMS® or PACTware®
2-line x 8-character LCD
English/Spanish/French/German
IP 66/Aluminium A356T6 (< 0.20 % copper) or stainless steel
ATEX II 1 G EEx ia II C T4, intrinsically safe – for non Foundation Fieldbus units
FISCO ATEX, intrinsically safe - for Foundation Fieldbus units
ATEX II 1/2 G D EEx d[ia] II C T6 - T85 °C, explosion proof for all units➀
ATEX II 3 G EEx nA II T6, non sparking – for non Foundation Fieldbus units
FM and CSA, Non incendive, intrinsically safe (FISCO) and explosion proof
STOOMWEZEN – Secondary level safety device for steamdrums
TÜV – WHG § 19, VLAREM II 5.17-7
LRS – Lloyds Register of Shipping (marine applications)
GOST-K/GGTN-K – ROSTECH/FSTS – Russian Authorisation Standards
Standard
electronics
Enhanced
electronics
Functional safety to SIL 1 / SIL 2 in accordance to 61508 – SFF > 85 %
– full FMEDA reports and declaration sheets available at request
Functional safety to SIL 2 / SIL 3 in accordance to 61508 – SFF > 91 %
– full FMEDA reports and declaration sheets available at request
Ui = 28,4 V, li = 94 mA, Pi = 1 W
Ui = 17,5 V, li = 380 mA, Pi = 5,32 W (Foundation Fieldbus)
Ci = 2,2 nF, Li = 3 µH
Ci = 0,24 nF, Li = 3 µH (Foundation Fieldbus)
ANSI/ISA-571.03 SA1 (Shock), ANSI/ISA-571.03 VC2 (Vibration)
2,70 kg net; 3,20 kg gross – amplifier only
5,70 kg net; 6,20 kg gross – amplifier only
H 214 mm (8.43") x W 111 mm (4.38") x D 188 m
Electrical Data
Equivalent Data
Shock/Vibration Class
Net and Gross Weight
Cast aluminium
Stainless steel
Overall Dimensions
➀ ATEX, explosion proof units use EEx d bushing material STYCAST 2057 FR
PERFORMANCE
Description
Specification
Reference Conditions with a 1,8 m (72")
coaxial type GWR probe
Coaxial/twin lead probes
Linearity➁
Single lead probes
➁
Coaxial/twin lead probes
Accuracy
Single lead probes
7MT interface
Resolution
Repeatability
Hysteresis
Response Time
Warm-up Time
Ambient Temp.
Reflection from liquid, with dielectric in center of selected range, at +20 °C (70 °F) with
CFD threshold➀
Process Dielectric Effect
Operating Temp. Effect
Humidity
Electromagnetic Compatibility
< 0,1 % of probe length or 2,5 mm (0.1"), whichever is greater
< 0,3 % of probe length or 8 mm (0.3"), whichever is greater
< 0,1 % of probe length or 2,5 mm (0.1"), whichever is greater
± 0,5 % of probe length or 13 mm (0.5"), whichever is greater
± 25 mm (1")
± 2,5 mm (0.1")
< 2,5 mm (0.1")
< 2,5 mm (0.1")
< 1 second
< 5 seconds
-40 °C to +80 °C (-40 °F to +175 °F)
– blind transmitter
-20 °C to +70 °C (-5 °F to +160 °F)
– with digital display
-40 °C to +70 °C (-40 °F to +160 °F)
– for EEx ia and EEx d[ia] with blind transmitter
-20 °C to +70 °C (-5 °F to +160 °F)
– for EEx ia and EEx d[ia] with digital display
< 7,5 mm (0.3") within selected range
Approx. +0,02 % of probe length/°C for probes ≥ 2,5 m (8')➂
0-99 %, non-condensing
Meets CE requirements (EN-61000-6-4, EN 61000-6-2) and NAMUR NE 21
(Single and Twin-Rod probe must be used in metallic vessel or stillwell)
➀ May degrade for 7MD probe or with fixed threshold.
➁ Top 600 mm (24") of twin rod probe: 30 mm (1.18").
➂ Accuracy may degrade slightly < 2,5 m (8')
Top 1220 mm (48") of single rod: application dependant.
9
P R O B E S P E C I F I C AT I O N S
Description
Materials
Probe
Process seal
Probe diameter
Mounting
Process Connection
Probe length
Transition Zone➀
Top
Bottom
Max. Process Temp.
Max. Process Pressure
Dielectric Range – Max. Viscosity
7MR: overfill safe GWR probe
316/316L (1.4401/1.4404)
Hastelloy C® (2.4819) or Monel® (2.4360)
Viton® GFLT, EPDM or Kalrez 4079 (Consult factory for alternatives)
Inside rod: 8 mm (0.3125") – Outer tube: 22 mm (0.875")
In-tank mounting / external cage mounting (WHG approved)
Threaded: 3/4" NPT or 1" BSP (G1)
Flanged: Various ANSI, EN/DIN or torque tube mating flanges
From 600 mm to 6100 mm (24 to 240"), selectable per 10 mm
0 mm (0")
εr: 2,0 = 150 mm (6")/εr: 80 = 25 mm (1")
+200 °C @ 18 bar (+400 °F @ 270 psi)
70 bar @ +20 °C (1000 psi @ +70 °F) – see table at page 11
1,4 to 100 – 200 cP
Viton® is a registered trademark of Dupont Performance Elastomers
Description
Materials
Probe
Process seal
Spacers
Probe diameter
Mounting
Process Connection
Probe length (selectable per 10 mm)
Top
Transition Zone➀
Bottom
Process Temp.
Max. Process Pressure
Max. Viscosity
Dielectric Range
Vacuum service
10
Max
Min
7MD: high pressure/high temperature
7MS: saturated steam GWR probe
GWR probe (overfill safe)
(overfill safe)
316/316L (1.4401/1.4404)
Borosilicate/inconel X750
PEEK with Aegis PF 128
Ceramic
Inside rod: 8 mm (0.125") – Outer tube: 22 mm (0.875")
In-tank mounting / external cage mounting (7MD – WHG / 7MS – Stoomwezen approved)
Threaded: 3/4" NPT or 1" BSP (G1)
Flanged: Various ANSI, EN/DIN or “proprietary” mating flanges
From 600 mm to 6100 mm (24 to 240")
From 1100 mm to 4500 mm (43 to 177")
25 mm (1")
εr: 2,0 = 150 mm (6") / εr: 80 = 25 mm (1")
εr: 10 = 150 mm (6") / εr: 80 = 25 mm (1")
+400 °C @ 135 bar (+750 °F @ 2000 psi)
-196 °C @ 135 bar (-320 °F @ 2000 psi)
345 bar @ +20 °C (5000 psig @ +70 °F)
200 cP
2 to 100
Full vacuum
(Helium leak < 10-8 cc/s @ 1 atmosphere vacuum)
+345 °C @ 155 bar (+650 °F @ 2250 psi)
-15 °C @ 205 bar (0 °F @ 3000 psi)
155 bar @ +345 °C (2250 psi @ +650 °F)
10 to 100
Negative pressure but not up to full vacuum
M A G N E T I C L E V E L I N D I C AT O R – S P E C I F I C AT I O N S
Measured value
Measuring range
Liquid level (consult factory for liquid-liquid interface)
From 300 mm (11.81") up to 5700 mm (224")
Indicators
Metal flag (red/white) – all indicators are hermetically sealed and assured by the “Insta-Seal” technology
Scale
Specific gravity
Visual indication
Type
Materials
Materials
Size
Rating
Configuration
Float
Cage
Insulation material
Process connections
Design
Constructions
In cm or % of span
As low as 0.5 kg/dm3
Visible from a distance up to 30 m (100 feet)
With magnetic flux ring – assembly – see page 2
316 SST (1.4401), Titanium, others at request
316/316L SST (1.4401/1.4404), others at request
3" or 4" depending configuration
Up to 2500 lbs / PN 320 class ratings
Side/side connection with 1/2" NPT plugged vent and drain
Weather resistant silicone cloth (high temperature application)
Polyurethane + aluminium jacket with polymeric frost extension for flag-rail (cryogenic application)
Threaded, socket welded or flanged
All cages are designed to meet the European PED (Pressure Equipment Directives)
97/23 EC guidelines
Standard commercial design
NACE construction
Others at request: eg. IBR, ASME - ANSI B31.3
T E M P E R AT U R E - P R E S S U R E R AT I N G F O R E C L I P S E P R O B E S E A L S
350
60
Process Pressure (bar)
Process Pressure (bar)
70
50
40
30
20
10
300
250
200
150
100
50
0
0
-40
-20
0
20
40
60
80
100
120
140
160
180
Process Temperature (°C)
7MR: overfill prevention GWR probe
200
-150
-100
-50
0
50
100
150
200
250
300
350
400
Process Temperature (°C)
7MD: high temperature / high pressure GWR probe
7MS: saturated steam GWR probe
11
QUALITY ASSURANCE - ISO 9001:2000
THE QUALITY ASSURANCE SYSTEM IN PLACE AT MAGNETROL GUARANTEES THE HIGHEST LEVEL OF QUALITY DURING THE DESIGN,
THE CONSTRUCTION AND THE SERVICE OF CONTROLS.
OUR QUALITY ASSURANCE SYSTEM IS APPROVED AND CERTIFIED TO ISO 9001:2000 AND OUR TOTAL COMPANY IS COMMITTED TO
PROVIDING FULL CUSTOMER SATISFACTION BOTH IN QUALITY PRODUCTS AND QUALITY SERVICE.
PRODUCT WARRANTY
:2000
ALL MAGNETROL MECHANICAL LEVEL CONTROLS ARE WARRANTED FREE OF DEFECTS IN MATERIALS AND WORKMANSHIP FOR FIVE
FULL YEARS (ELECTRONICS ONE FULL YEAR) FROM THE DATE OF ORIGINAL FACTORY SHIPMENT.
IF RETURNED WITHIN THE WARRANTY PERIOD; AND, UPON FACTORY INSPECTION OF THE CONTROL, THE CAUSE OF THE CLAIM IS DETERMINED TO BE COVERED
UNDER THE WARRANTY; THEN, MAGNETROL INTERNATIONAL WILL REPAIR OR REPLACE THE CONTROL AT NO COST TO THE PURCHASER (OR OWNER) OTHER THAN
TRANSPORTATION.
MAGNETROL SHALL NOT BE LIABLE FOR MISAPPLICATION, LABOR CLAIMS, DIRECT OR CONSEQUENTIAL DAMAGE OR EXPENSE ARISING FROM THE INSTALLATION
OR USE OF THE EQUIPMENT. THERE ARE NO OTHER WARRANTIES EXPRESSED OR IMPLIED, EXCEPT, SPECIAL WRITTEN WARRANTIES COVERING SOME MAGNETROL
PRODUCTS.
BULLETIN N°:
EFFECTIVE:
SUPERSEDES:
UNDER RESERVE OF MODIFICATIONS
Heikensstraat 6, 9240 Zele, België
Tel. +32 (0)52.45.11.11 • Fax. +32 (0)52.45.09.93 • E-Mail: [email protected]
DEUTSCHLAND
Alte Ziegelei 2-4, D-51491 Overath
Tel. 02204 / 9536-0 • Fax. 02204 / 9536-53 • E-Mail: [email protected]
FRANCE
40 - 42, rue Gabriel Péri, 95130 Le Plessis Bouchard
Tél. 01.34.44.26.10 • Fax. 01.34.44.26.06 • E-Mail: [email protected]
ITALIA
Via Arese 12, I-20159 Milano
Tel. (02) 607.22.98 (R.A.) • Fax. (02) 668.66.52 • E-Mail: [email protected]
UNITED
KINGDOM
Unit 1 Regent Business Centre, Jubilee Road Burgess Hill West Sussex RH 15 9TL
Tel. (01444) 871313 • Fax (01444) 871317 • E-Mail: [email protected]
INDIA
E-22, Anand Niketan, New Delhi - 110 021
Tel. 91 (11) 41661840 • Fax 91 (11) 41661843 • E-Mail: [email protected]
www.magnetrol.com
BENELUX
OUR NEAREST REPRESENTATIVE
BE 57-138.3
NOVEMBER 2005
December 2002