thermatel® tg1/tg2

THERMATEL®
TG1/TG2
Thermal
dispersion switch
DESCRIPTION
Thermatel® TG1/TG2 switches consist of electronics in a
DIN rail housing and a remote sensor with aluminium or
stainless steel sensor housing (max 500 m (1640 ft) away
from electronics).
TG1/TG2 switches can easily be adjusted to detect flow
(gases and liquids), level or liquid-liquid interface. Both units
are 2-wire 24 V DC powered and intrinsically safe approved.
The TG1 offers standard LED flow indication, the TG2 offers
LED flow indication per NAMUR NE 44.
For FLOW/LEVEL/INTERFACE applications
F E AT U R E S
• Easy field calibration
•
•
•
•
•
•
•
•
•
– pre-calibration from factory at
request.
Variable flow or Flow / No flow detection of gases and
liquids.
Excellent low flow sensitivity.
Continuous diagnostics detect sensor fault.
Continuous monitoring of flow rate versus setpoint via
LED.
mA output provides repeatable indication of flow rate
and fault detection.
Optional retractable fitting for dismantling under process
conditions.
Process conditions up to +450 °C (+850 °F) and 414 bar
(6000 psi).
Unique spherical tip design option ideal for liquids or
high viscosity applications.
Suited for SIL1 and SIL2 loops (full FMEDA report available).
Spherical type sensor
DIN rail
FE T Y I
NT
SA
Twin tip type sensor
EGRITY
V EL
LE
MEDIA: all types of gases and liquids.
VESSELS: pipe sizes down to 1/4". Max sensor length up to
3,3 m. Can be installed at any angle vertically/horizontally.
CONDITIONS: Can be used on conductive and non conductive media, very light density to heavy viscous media (up
to 10.000 cP). Can be set to ignore foam, aeration, turbulence, and cavitation.
A P P L I C AT I O N S
A G E N C Y A P P R O VA L S
Agency
Approval
ATEX
II 1 G EEx ia IIB T5
Russian Authorisation Standards➀
Other approvals are available, consult factory for more details
➀ Consult factory for proper model numbers and classifications.
Worldwide level and flow solutions
TECHNOLOGY
The unit utilises Magnetrol’s proven thermal dispersion technology. The sensor consists of two RTD (Resistance Temperature
Detector) elements. One is the reference and the second is heated to a temperature above the process temperature. The electronics detect the temperature difference between the two elements. The temperature difference is greatest in air, then decreases when cooling occurs due to a change in media. An increase in the flow rate further decreases the temperature difference.
The set point is adjusted for the switch to alarm at the desired temperature difference. Once the set point is reached, the relay
will change state.
Level
Flow
No Flow/Low Flow
In the absence of flow/low flow,
the self-heated sensor creates a
temperature differential between
the two sensors.
Flow
As media flows increases across
the sensing assembly, heat is
dissipated and temperature differential decreases .
A P P L I C AT I O N S
FLOW
Thermatel® TG1/TG2 switches may be installed in a variety
of flow applications as shown in the illustration below.
Flow/No Flow can be detected in an input line to a primary
tank, or in an output line. They may be installed for overflow
detection in a pipe connected to an overflow tank or
installed in a drain line for Wet/Dry indication. In addition,
due to the capability to detect liquids or gases, the
Thermatel® flow switch may be installed in a gas vent to
detect off-gas from the primary tank.
Flow / No flow
- Liquid or Gas flow detection
- Maintain a minimum flow rate
- Pump protection
- Cooling air/water
- Lubrication systems
- Chemical feed pumps
- Detect presence of flow
- Relief valves
- Flare lines
- Water for injection (WFI)
- Filtration systems
- Separation systems
- CIP systems
- Air, CO2, N2 flow
2
Switch at set flow rate
Additive flow
Leak detection
LEVEL
Thermatel® TG1/TG2 switches can be installed in a variety
of level applications as shown in the illustration below. High
or low level alarm applications can be installed in either vertical or horizontal mountings. Input line
Output line
Overflow
tank
High Level
As media contacts the
sensing assembly, heat
is absorbed by the fluid,
decreasing the temperature differential.
Off gas
Gas vent
Overflow
Low Level
In the absence of media, the
self-heated sensor tip creates a
temperature differential
between the two sensors.
Low Flow
Body
Interface
High level
alarm
Low level alarm
output line
Overflow
tank
Pump protection
- High level
- Low level
- Interface between different media
- Oil/water
- Liquid/foam
- Suitable for any liquid level detection including:
- High viscosity
- High solids content
- Aeration
- Foam
- Insensitive to dielectric, specific gravity, viscosity
SENSOR DESIGNS
Thermatel offers two sensor tip designs: the twin tip and the
unique spherical tip. Both designs have similar operating
ranges. Both detect flow or level at approximately the same
rate. However, the spherical tip responds faster to a loss of
flow or a dry condition.
Twin tip
Spherical tip
SPHERICAL TIP
The sensing elements are bonded directly to the wall of the tip, providing protection of the sensors.
The spherical tip is recommended for all types of applications: general purpose use, liquid flow applications, high viscosity and
applications where buildup can occur. The spherical tip is suitable for process pressures up to 41,4 bar (600 psi) and can handle process temperatures up to +200 °C (+400 °F).
TWIN TIP
The sensing elements are mounted at the end of each tip.
The twin tip is preferred for air flow applications and is available in corrosion resistant materials including Hastelloy C and
Monel. The twin tip is suitable for process pressures up to 207 bar (3000 psi) and can handle process temperatures up to
+200 °C (+400 °F).
HIGH TEMPERATURE/HIGH PRESSURE (TMH)
This twin tip sensor is suitable for process pressures up to 414 bar (6000 psi) and can handle process temperatures up to
+450 °C (+850 °F).
MINI SENSOR (TMM)
This twin tip sensor is suitable for installing in smaller pipe sizes. Available with 1/2", 3/4" and 1" NPT connections. The twin tip
design provides minimal blockage of the pipe.
LOW FLOW BODY (TML)
This version is used for even lower flow rates with 1/4" and 1/2" connections.
Typical air flow
Typical water flow
18.00
15.00
16.00
14.00
13.00
12.00
10.00
mA
mA
14.00
8.00
6.00
12.00
11.00
10.00
4.00
0
15
Twin Tip
30
45
Spherical Tip
60 m/s
Hastelloy
9.00
8.00
0,0
0,3
0,6
Spherical Tip
0,9
1,2
Twin Tip
1,5 m/s
Hastelloy
Model
Sensor design
Recommended for
TMB
Spherical tip
Same as TMA but can be used with integral electronics up to +200 °C (+400 °F)
TMA
TMC
TMD
TMH
TMM
TML
Spherical tip
Twin tip
Twin tip
Twin tip
Mini twin tip
Low flow body
Best sensitivity for liquid flows / suitable for gas flow – resists heavy coating
Best sensitivity for air/gas flows – resists light coating
Same as TMC but can be used with integral electronics up to +200 °C (+400 °F)
High temperature and/or high pressure conditions – resists light coating
Installation in small pipe sizes – resists light coating
Detection/control of extreme low flows – resists light coating
3
DIMENSIONS IN mm (inches)
DIN rail
55 (2.16)
75 (2.95)
DELAY
POWER
ALARM
ERROR
LOW
FAIL-SAFE
35 (1.38)
Sensor housing
HIGH
®
1
(4 10
.33
)
Rotation clearance
76 (3.00)
max. 500 m (1640 ft)
cable is ordered separately
70
(2.75)
Single cable entry
Separate or seal
welded flange
4
ø 118 (4.63)
Overhead clearance 70 (2.75)
114
(4.49)
Threaded:
46 (1.81)
Welded flange: 60 (2.36)
Heat extension: 223 (8.78)
EXPEDITE SHIP PLAN (ESP)
Several models are available for quick shipment, within max. 4 weeks after factory receipt of purchase order, through the
Expedite Ship Plan (ESP).
Models covered by ESP service are conveniently colour coded in the selection data charts.
To take advantage of ESP, simply match the colour coded model number codes (standard dimensions apply).
ESP service may not apply to orders of ten units or more. Contact your local representative for lead times on larger volume
orders, as well as other products and options.
A complete measuring system consists of:
S E L E C T I O N D ATA
1. THERMATEL® DIN RAIL electronics and sensor housing
2. Connecting cable
3. THERMATEL® sensor
4. Optional: Order code for thread-on mounting flanges
5. Optional: Retractable probe assembly, consult factory for details
6. Optional: Factory calibration, consult factory
1. Order code for Thermatel® DIN RAIL ELECTRONICS
T G 1
T G 2
Electronics with standard LED flow indication
Electronics with LED flow indication per NAMUR NE 44
1
OUTPUT
2 Amp SPDT alarm relay with mA output signal (non linear / non scaleable)
POWER SUPPLY
2
24 V DC
MOUNTING
0 D
Remote DIN rail mounted electronics
T
2
6
SENSOR HOUSING / CABLE ENTRY
IP 65, Cast aluminium, M20 x 1,5 cable entry
IP 65, Cast aluminium, 3/4" NPT cable entry
IP 65, Cast stainless steel, 3/4" NPT cable entry
APPROVAL
A 0
T G
1 2 0 D
A 0
ATEX II 1 G EEx ia IIB T5, intrinsically safe
complete order code for Thermatel® TG1/TG2 electronics
X = product with a specific customer requirement
2. Order code for connecting cable (standard 2-wire shielded instrument cable – 0,50 mm2)
1 3 7
3 2 2 2
0 0 1 - 5 0 0 From 1 m (3.28 ft) min. to 500 m (1640 ft) max.
Specify in increments of 1 m (3.28 ft)
complete order code for connecting cable
5
CONNECTIONS
Insertion
length
NPT connection
Insertion
length
BSP connection
Insertion
length
Welded flange ANSI - EN (DIN)
Threaded
DIMENSIONS IN mm (inches) – TMA/TMB/TMC/TMD
Standard: 46 (1.81)
With heat extension: 223 (8.78)
Standard: 60 (2.36)
With heat extension: 223 (8.78)
Insertion length
NPT connection
Insertion length
BSP connection
Insertion
length
ø 22,9 (0.90)
ø 22,9 (0.90)
Threaded Sensor
Flanged Sensor
P R E S S U R E / T E M P E R AT U R E R AT I N G – T M A / T M B / T M C / T M D
Sensor
Material
code
TMC, TMD
A, D
TMC, TMD
B
TMC, TMD
C
TMA, TMB
A
Insertion length
All
@ +40 °C (+100 °F)
41,4 bar (600 psi)
Maximum process pressure
@ +120 °C (+250 °F)
@ +200 °C (+400 °F)
33,8 bar (490 psi)
= minimum length
207 bar (3000 psi)
170 bar (2460 psi)
= minimum length
207 bar (3000 psi)
181 bar (2627 psi)
> minimum length
> minimum length
= minimum length
> minimum length
128 bar (1850 psi)
28,6 bar (415 psi)
148 bar (2140 psi)
105 bar (1517 psi)
91,0 bar (1320 psi)
103 bar (1500 psi)
90,6 bar (1313 psi)
80,7 bar (1170 psi)
82,8 bar (1200 psi)
70,3 bar (1020 psi)
65,5 bar (950 psi)
172 bar (2500 psi)
TMA/TMB sensors
147 bar (2125 psi)
161 bar (2340 psi)
137 bar (1980 psi)
TMC/TMD sensors with material code A or D
Insertion length = minimum length
6
Process pressure (bar)
Process pressure (bar)
Insertion length > minimum length
Process temperature (°C)
Process temperature (°C)
S E L E C T I O N D A T A ( C O N T. )
3. Order code for Thermatel® TG1/TG2 – STANDARD SENSOR
T
T
T
T
M
M
M
M
A
B
C
D
Spherical tip
Spherical tip
Twin tip
Twin tip
-
standard
with heat extension
standard
with heat extension
+120
+200
+120
+200
°C
°C
°C
°C
(+250
(+400
(+250
(+400
°F)
°F)
°F)
°F)
MATERIAL OF CONSTRUCTION FOR SENSOR AND PROCESS CONNECTION
A
B
C
316/316L (1.4401/1.4404) stainless steel
Hastelloy® C (2.4819)
Monel® (2.4360)
– TMC/TMD only
– TMC/TMD only
PROCESS CONNECTION – SIZE/TYPE
Threaded
1
2
2
1 0
1 0
2 0
3/4" NPT
1" NPT
1" BSP (G 1")
2
2
2
3
3
3
4
4
4
3
4
5
3
4
5
3
4
5
0
0
0
0
0
0
0
0
0
1"
1"
1"
1 1/2"
1 1/2"
1 1/2"
2"
2"
2"
150
300
600
150
300
600
150
300
600
lbs
lbs
lbs
lbs
lbs
lbs
lbs
lbs
lbs
B
B
C
C
D
D
D
D
B
C
B
C
A
B
D
E
0
0
0
0
0
0
0
0
DN
DN
DN
DN
DN
DN
DN
DN
PN
PN
PN
PN
PN
PN
PN
PN
16/25/40
63/100
16/25/40
63/100
16
25/40
63
100
ANSI flanges
EN (DIN) flanges
0
0
0
0
0
3
0
25
25
40
40
50
50
50
50
ANSI
ANSI
ANSI
ANSI
ANSI
ANSI
ANSI
ANSI
ANSI
RF
RF
RF
RF
RF
RF
RF
RF
RF
EN
EN
EN
EN
EN
EN
EN
EN
1092-1
1092-1
1092-1
1092-1
1092-1
1092-1
1092-1
1092-1
INSERTION LENGTH – MINIMUM
0
0
0
0
0
T M
max
max
max
max
0
0
0
0
0
5
6
7
6
8
5 cm (2")
5,5 cm (2.17")
7 cm (2.76")
5,5 cm (2.17")
7,5 cm (3")
0
0
0
0
0
3
6
7
8
7
9
0
Minimum 6 cm (2.36")
Minimum 7 cm (2.76")
Minimum 8 cm (3.15")
Minimum 7 cm (2.76")
Minimum 9 cm (3.54")
Maximum 330 cm (130")
Type
Type
Type
Type
Type
Type
Type
Type
Sensor
TMA, TMB
TMC, TMD
A
B2
A
B2
A
A
B2
B2
Process connection
NPT
flanged
BSP
NPT, flanged
BSP
INSERTION LENGTH – SELECTABLE – Specify per cm (0.39") increment
Sensor
TMA, TMB
TMC, TMD
all
Process connection
NPT
flanged
BSP
NPT, flanged
BSP
all
complete order code for Thermatel® TG1/TG2 STANDARD SENSOR
X = product with a specific customer requirement
7
DIMENSIONS in mm (inches) – HYGIENIC TMA/TMB
TMA: 29 (1.14)
TMB: 192 (7.56)
Insertion Length
Ø 22,9 (0.90)
TMA: 35 (1.38)
Insertion Length
Ø 22,9 (0.90)
TMA: 29 (1.14)
Varivent®
DIN 11851
Tri-Clamp®
Insertion Length
Ø 22,9 (0.90)
P R E S S U R E / T E M P E R AT U R E R AT I N G – H Y G I E N I C T M A / T M B
Note: max. process pressure is downrated to the design pressure of the selected process connection.
@ +40 °C (+100 °F)
@ +120 °C (+250 °F)
33,8 bar (490 psi)
Process pressure (bar)
41,4 bar (600 psi)
Maximum process pressure
Process temperature (°C)
8
@ +200 °C (+400 °F)
28,6 bar (415 psi)
S E L E C T I O N D A T A ( C O N T. )
3. Order code for Thermatel® TG1/TG2 – HYGIENIC SENSOR
T M A
T M B
Spherical tip - standard
Spherical tip - with heat extension
max +120 °C (+250 °F)
max +200 °C (+400 °F)
SENSOR MATERIAL (0,82 Μm Ra (32 Ra) surface finish)
A
316/316L (1.4401/1.4404) stainless steel
PROCESS CONNECTION – SIZE/TYPE ➀
3
4
B
C
D
V
T
T
S
S
S
V
0
0
0
0
0
0
1 1/2" Tri-Clamp®
2" Tri-Clamp®
DN 25 DIN 11851 - only available with TMA
DN 40 DIN 11851 - only available with TMA
DN 50 DIN 11851 - only available with TMA
DN 65 Varivent® type N - only available with TMA
➀ Consult factory for other process connections (NEUMO BioControl®, G1A, ...)
INSERTION LENGTH – Specify per cm (0.39") increment
0 0 7
3 3 0
T M
0
Minimum 7 cm (2.76")
Maximum 330 cm (130")
complete order code for Thermatel® TG1/TG2 HYGIENIC SENSOR
X = product with a specific customer requirement
9
DIMENSIONS IN mm (inches) – TMH
251
(9.88)
Insertion
length
BSP
connection
254
(10.00)
273
(10.75)
Insertion
length
NPT
connection
Insertion length
Flanged
connection
Ø 21,9
(0.86)
Ø 21,9
(0.86)
P R E S S U R E / T E M P E R AT U R E R AT I N G – T M H
@ +40 °C (+100 °F)
@ +120 °C (+250 °F)
339 bar (4920 psi)
@ +200 °C (+400 °F)
295 bar (4280 psi)
Process pressure (bar)
414 bar (6000 psi)
Maximum process pressure
Process temperature (°C)
10
@ +450 °C (+850 °F)
233 bar (3380 psi)
S E L E C T I O N D A T A ( C O N T. )
3. Order code for Thermatel® TG1/TG2 – HIGH TEMPERATURE / HIGH PRESSURE SENSOR
T M H
High temperature / high pressure twin tip – max +450 °C (+850 °F) / max 414 bar (6000 psi)➀
➀ Not available with retractable probe assembly.
MATERIAL OF CONSTRUCTION FOR SENSOR AND PROCESS CONNECTION
A
B
316/316L (1.4401/1.4404) stainless steel
Hastelloy® C (2.4819)
PROCESS CONNECTION – SIZE/TYPE
Threaded
1
2
2
1 0
1 0
2 0
3/4" NPT
1" NPT
1" BSP (G 1")
ANSI flanges
2 3 0
1"
2 4 0
1"
2 5 0
1"
2 7 0
1"
3 3 0
1 1/2"
3 4 0
1 1/2"
3 5 0
1 1/2"
3 7 0
1 1/2"
3 8 0
1 1/2"
4 3 0
2"
4 4 0
2"
4 5 0
2"
4 7 0
2"
4 8 0
2"
EN (DIN) flanges
B B 0
DN 25
B C 0
DN 25
B G 0
DN 25
C B 0
DN 40
C C 0
DN 40
C G 0
DN 40
C J 0
DN 40
D A 0
DN 50
D B 0
DN 50
D D 0
DN 50
D E 0
DN 50
D G 0
DN 50
D J 0
DN 50
150 lbs
300 lbs
600 lbs
900/1500 lbs
150 lbs
300 lbs
600 lbs
900/1500 lbs
2500 lbs
150 lbs
300 lbs
600 lbs
900/1500 lbs
2500 lbs
PN
PN
PN
PN
PN
PN
PN
PN
PN
PN
PN
PN
PN
16/25/40
63/100
250
16/25/40
63/100
250
400
16
25/40
63
100
250
400
INSERTION LENGTH – MINIMUM
0 0 6 5,5 cm (2.17")
0 0 7 7 cm (2.76")
0 0 8 7,5 cm (3")
0
0
0
0
T M H
0
ANSI
ANSI
ANSI
ANSI
ANSI
ANSI
ANSI
ANSI
ANSI
ANSI
ANSI
ANSI
ANSI
ANSI
EN
EN
EN
EN
EN
EN
EN
EN
EN
EN
EN
EN
EN
RF
RF
RF
RF
RF
RF
RF
RF
RF
RF
RF
RF
RF
RF
1092-1
1092-1
1092-1
1092-1
1092-1
1092-1
1092-1
1092-1
1092-1
1092-1
1092-1
1092-1
1092-1
Type
Type
Type
Type
Type
Type
Type
Type
Type
Type
Type
Type
Type
A
B2
B2
A
B2
B2
B2
A
A
B2
B2
B2
B2
Process connection
NPT
flanged
BSP
INSERTION LENGTH – SELECTABLE – Specify per cm (0.39") increment
0
0
0
9
7
8
9
1
Minimum 7 cm (2.76")
Minimum 8 cm (3.15")
Minimum 9 cm (3.54")
Maximum 91 cm (36")
Process connection
NPT
flanged
BSP
all
complete order code for Thermatel® TG1/TG2
HIGH TEMPERATURE /HIGH PRESSURE SENSOR
X = product with a specific customer requirement
11
S E L E C T I O N D A T A ( C O N T. )
3. Order code for Thermatel® TG1/TG2 – MINI SENSOR
T M M
Mini twin tip – max +120 °C (+250 °F)
MATERIAL OF CONSTRUCTION FOR SENSOR AND PROCESS CONNECTION
A
316/316L (1.4401/1.4404) stainless steel
PROCESS CONNECTION – SIZE/TYPE
Threaded
0
1
2
1 0
1 0
1 0
1/2" NPT
3/4" NPT
1" NPT
INSERTION LENGTH – MINIMUM
0 0 3
2,5 cm (1")
0 0 5
3 3 0
Minimum 5 cm (2")
Maximum 330 cm (130")
INSERTION LENGTH – SELECTABLE – Specify per cm (0.39") increment
T M M
A
1 0
complete order code for Thermatel® TG1/TG2 MINI SENSOR
X = product with a specific customer requirement
D I M E N S I O N S I N m m ( i n c h e s ) & P R E S S U R E / T E M P E R AT U R E R AT I N G – T M M
46
(1.81)
Insertion length
= 2,5 cm (1")
> 2,5 cm (1")
Insertion
length
Maximum process pressure
@ +40 °C (+100 °F)
@ +120 °C (+250 °F)
128 bar (1850 psi)
105 bar (1517 psi)
207 bar (3000 psi)
170 bar (2460 psi)
Insertion length = minimum length
Insertion length > minimum length
Process pressure (bar)
ø 16
(0.63)
Process temperature (°C)
RECOMMENDED FLOW RANGES – TMM
Pipe size
Water
Air
1"
3,8 to 1600 l/h (1 to 420 GPH)
5 to 290 Nm3/h (3 to 170 SCFM)
1/2"
3/4"
12
0,75 to 680 l/h (0.2 to 180 GPH)
2 to 900 l/h (0.5 to 240 GPH)
0,85 to 120 Nm3/h (0.5 to 70 SCFM)
2,5 to 170 Nm3/h (1.5 to 100 SCFM)
S E L E C T I O N D A T A ( C O N T. )
3. Order code for Thermatel® TG1/TG2 – LOW FLOW BODY SENSOR
T M L
Low flow body – max +120 °C (+250 °F) / max 400 bar (5800 psi)
MATERIAL OF CONSTRUCTION FOR SENSOR AND PROCESS CONNECTION
A
316/316L (1.4401/1.4404) stainless steel
PROCESS CONNECTION – SIZE/TYPE
Threaded
T
V
T
V
1
1
0
0
0
0
0
0
1/4"
1/2"
1/4"
1/2"
NPT-F
NPT-F
BSP (G 1/4")
BSP (G 1/2")
MOUNTING BRACKET
0 0 0
1 0 0
T M L
A
0 0
0
None
With carbon steel mounting bracket
complete order code for Thermatel® TG1/TG2 LOW FLOW BODY SENSOR
X = product with a specific customer requirement
D I M E N S I O N S I N m m ( i n c h e s ) & P R E S S U R E / T E M P E R AT U R E R AT I N G – T M L
optional mounting
bracket
2 holes
ø 9,5 (0.37)
51
(2.00)
70
(2.75)
Maximum process pressure
@ +40 °C (+100 °F)
400 bar (5800 psi)
76
(3.00)
89
(3.50)
328 bar (4760 psi)
32
(1.25)
95
(3.75)
Process pressure (bar)
1/4" or 1/2"
NPT or BSP
@ +120 °C (+250 °F)
51 (2.0)
Process temperature (°C)
RECOMMENDED FLOW RANGES – TML
Size
1/4"
1/2"
flow body
flow body
Water
0,02 to 5,7 l/h (0.0055 to 1.5 GPH)
0,04 to 11,5 l/h (0.01 to 3 GPH)
Air
0,006 to 5,75 Nm3/h (0.21 to 200 SCFH)
0,015 to 11,5 Nm3/h (0.53 to 400 SCFH)
13
S E L E C T I O N D A T A ( C O N T. )
4. Optional sensor mounting flanges
Thread-on mounting flanges can only be used in combination with 3/4" NPT process connection sensor.
Consult factory for other sizes or materials.
Thread-on flanges for use with 3/4" NPT-M connections
ANSI B16.5 flanges
1"
1"
150 lbs RF
150 lbs RF
300 lbs RF
1 1/2"
300 lbs RF
3"
300 lbs RF
2"
4"
6"
1"
1
1/2"
2"
004-5867-001
004-5867-022
150 lbs RF
6"
004-5867-021
150 lbs RF
3"
4"
316/316L SST
004-5867-041
150 lbs RF
2"
Carbon steel
150 lbs RF
1 1/2"
300 lbs RF
300 lbs RF
300 lbs RF
600 lbs RF
600 lbs RF
600 lbs RF
Part No.
004-5867-023
004-5867-043
004-5867-052
004-5867-002
004-5867-032
004-5867-003
004-5867-024
004-5867-004
004-5867-025
004-5867-005
004-5867-042
004-5867-044
004-5867-026
004-5867-006
004-5867-027
004-5867-007
004-5867-028
004-5867-008
004-5867-029
004-5867-009
004-5867-030
004-5867-010
004-5867-051
004-5867-046
004-5867-031
004-5867-033
004-5867-034
004-5867-035
004-5867-053
004-5867-036
004-5867-037
004-5867-038
004-5867-039
004-5867-040
004-5867-050
consult factory
004-5867-048
consult factory
004-5867-045
004-5867-049
Hastelloy C
consult factory
E L E C T R O N I C S S P E C I F I C AT I O N S
Description
Specifications
Power consumption
Flow range
5 W max.
TMA-A, TMB-A, TMC-A, TMD-A, TMM: 0,003 to 1,5 m/s (0.01 to 5.0 FPS) – water
0,03 to 150 m/s (0.1 to 500 FPS) – air
Power supply
19,2 to 28,8 V DC
TMM:
see table on page 12
TML:
see table on page 13
TMC-B, TMC-C, TMD-B, TMD-C,TMH: 0,003 to 0,3 m/s (0.01 to 1.0 FPS) – water
0,03 to 150 m/s (0.1 to 500 FPS) – air
Output
Alarm
Continuous
User interface
Set point
LED indication
Approvals
Error
Range selection
Power
Error
Alarm
SIL (Safety Integrity Level)
Housing material
Net weight
14
2 A SPDT relay
mA output (non linear, non scaleable)
3,6 mA (Low Level Fail-Safe) – 22 mA (High Level Fail-safe)
Adjustable via potentiometer located on DIN Rail housing
Selectable in probe electronics
LED’s for Power/Alarm status
Red LED blinks in case of error
4 x green LED's
1 x yellow LED
1 x red LED
all LED's OFF
–
–
–
–
for safe/ (normal) condition
indicates when flow or level is approaching the alarm set point
indicates an alarm condition (TG1)
indicates an alarm condition (TG2)
ATEX II 1 G EEx ia IIB T5
Other approvals are available, consult factory for more details
Functional safety to SIL1 as 1oo1 / SIL2 as 1oo2 in accordance to IEC 61508 –
SFF of 79,4 % – full FMEDA reports and declaration sheets available
DIN Rail: IP 20, polycarbonate / Sensor housing: IP 65, Aluminium or Stainless Steel
Aluminium: 1,6 kg (3.5 lbs) – electronics only
Stainless steel: 4,0 kg (8.8 lbs) – electronics only
PERFORMANCE
Description
Specification
Ambient temperature
-40 °C to +70 °C (-40 °F to +160 °F)
Storage: -50 °C to +75 °C (-58 °F to +170 °F)
0-99 %, non-condensing
Meets CE requirements (EN 61326: 1997 + A1 + A2)
Response time
Repeatability
Humidity
Electromagnetic compatibility
1-10 s typical (dependent on sensor type, application and set point)
< 1 % @ constant temperature
S E N S O R S P E C I F I C AT I O N S
Description
Materials
Sensor diameter
Process connection
Sensor length
Process temperature
Max process pressure
Description
Spherical tip - Twin tip sensors
INDUSTRIAL TMA/TMB - TMC/TMD
316/316L (1.4401/1.4404)
Hastelloy® C (2.4819) – TMC/TMD only
Monel® (2.4360) – TMC/TMD only
22,9 mm (0.90")
Threaded: NPT or BSP
Flanged: various ANSI or EN (DIN) flanges
5 - 330 cm (2" - 130")
TMA/TMC: -70 °C to +120 °C (-100 °F to +250 °F)
TMB/TMD: -70 °C to +200 °C (-100 °F to +400 °F)
See info on page 6
Mini twin tip sensor
TMM
Materials
316/316L (1.4401/1.4404)
Process temperature
-70 °C to +120 °C (-100 °F to +250 °F)
Sensor diameter
Process connection
Sensor length
Max process pressure
Description
Materials
Surface finish
Sensor diameter
Process connection
Sensor length
Process temperature
Max process pressure
16 mm (0.63")
21,9 mm (0.86")
5,5 - 91 cm (2.17" - 36")
-70 °C to +450 °C (-100 °F to +850 °F)
See info on page 10
Low flow body
TML
or 1/2" pipe size
Threaded: 1/4" or 1/2" NPT-F or BSP
Not applicable
See info on page 12
0,82 µm Ra (32 Ra)
316/316L (1.4401/1.4404)
Hastelloy® C (2.4819)
1/4"
Threaded: 1/2", 3/4" or 1" NPT
2,5 - 330 cm (1" - 130")
316/316L (1.4401/1.4404)
HTHP sensor
TMH
See info on page 13
Spherical tip sensors
HYGIENIC TMA/TMB
22,9 mm (0.90")
Tri-Clamp®, DIN 11851, Varivent®
7 - 330 cm (2.76" - 130")
TMA: -70 °C to +120 °C (-100 °F to +250 °F)
TMB: -70 °C to +200 °C (-100 °F to +400 °F)
See info on page 8
15
QUALITY ASSURANCE - ISO 9001:2008
THE QUALITY ASSURANCE SYSTEM IN PLACE AT MAGNETROL GUARANTEES THE HIGHEST LEVEL OF QUALITY DURING THE DESIGN,
THE CONSTRUCTION AND THE SERVICE OF CONTROLS.
OUR QUALITY ASSURANCE SYSTEM IS APPROVED AND CERTIFIED TO ISO 9001:2008 AND OUR TOTAL COMPANY IS COMMITTED TO
PROVIDING FULL CUSTOMER SATISFACTION BOTH IN QUALITY PRODUCTS AND QUALITY SERVICE.
PRODUCT WARRANTY
:2008
ALL MAGNETROL ELECTRONIC AND ULTRASONIC LEVEL CONTROLS ARE WARRANTED FREE OF DEFECTS IN MATERIALS AND WORKMANSHIP FOR ONE FULL YEAR FROM THE DATE OF ORIGINAL FACTORY SHIPMENT. IF RETURNED WITHIN THE WARRANTY PERIOD; AND, UPON FACTORY INSPECTION OF THE CONTROL, THE CAUSE OF THE CLAIM IS DETERMINED TO BE COVERED UNDER THE WARRANTY; THEN, MAGNETROL INTERNATIONAL WILL REPAIR OR
REPLACE THE CONTROL AT NO COST TO THE PURCHASER (OR OWNER) OTHER THAN TRANSPORTATION.
MAGNETROL SHALL NOT BE LIABLE FOR MISAPPLICATION, LABOR CLAIMS, DIRECT OR CONSEQUENTIAL DAMAGE OR EXPENSE ARISING FROM THE INSTALLATION
OR USE OF THE EQUIPMENT. THERE ARE NO OTHER WARRANTIES EXPRESSED OR IMPLIED, EXCEPT, SPECIAL WRITTEN WARRANTIES COVERING SOME MAGNETROL
PRODUCTS.
®
UNDER RESERVE OF MODIFICATIONS
Heikensstraat 6, 9240 Zele, België -Belgique
Tel. +32 (0)52.45.11.11 • Fax. +32 (0)52.45.09.93 • E-Mail: [email protected]
DEUTSCHLAND
Alte Ziegelei 2-4, D-51491 Overath
Tel. +49 (0)2204 / 9536-0 • Fax. +49 (0)2204 / 9536-53 • E-Mail: [email protected]
INDIA
B-506, Sagar Tech Plaza, Saki Naka Junction, Andheri (E), Mumbai - 400072
Tel. +91 22 2850 7903 • Fax. +91 22 2850 7904 • E-Mail: [email protected]
ITALIA
Via Arese 12, I-20159 Milano
Tel. +39 02 607.22.98 • Fax. +39 02 668.66.52 • E-Mail: [email protected]
RUSSIA
198095 Saint-Petersburg, Marshala Govorova street, house 35A, office 427
Tel. +7-812.702.70.87 • E-Mail: [email protected]
U.A.E.
UNITED
KINGDOM
DAFZA Office 5EA 722 • PO Box 293671 • Dubai
Tel. +971-4-6091735 • Fax +971-4-6091736 • E-Mail: [email protected]
Unit 1 Regent Business Centre, Jubilee Road Burgess Hill West Sussex RH 15 9TL
Tel. +44 (0)1444 871313 • Fax +44 (0)1444 871317 • E-Mail: [email protected]
www.magnetrol.com
BENELUX
FRANCE
BULLETIN N°:
EFFECTIVE:
SUPERSEDES:
OUR NEAREST REPRESENTATIVE
BE 54-105.7
DECEMBER 2014
November 2012