NI PXIe-1491 Specifications

DEVICE SPECIFICATIONS
NI PXIe-1491
This document lists specifications for the NI PXIe-1491 high-definition multimedia interface
(HDMI) analyzer.
Specifications are warranted under the following conditions: you have installed software
support for the NI PXIe-1491. Install one of the following software applications to
automatically install software support: NI Picture Quality Analysis (PQA) 1.5 or later,
NI Video Measurement Suite (VMS) 3.2 or later, or NI AudioMASTER 6.3 or later.
Note The features and functions of the NI PXIe-1491 depend on the software
support you install.
Specifications describe the warranted, traceable product performance over ambient
temperature ranges of 0 ºC to 55 ºC, unless otherwise noted.
Typical values describe useful product performance beyond specifications that are not covered
by warranty and do not include guardbands for measurement uncertainty or drift. Typical
values may not be verified on all units shipped from the factory. Unless otherwise noted,
typical values cover the expected performance of units over ambient temperature ranges of
23 °C ± 5 °C with a 90% confidence level, based on measurements taken during development
or production.
Nominal values (or supplemental information) describe additional information about the
product that may be useful, including expected performance that is not covered under
Specifications or Typical values. Nominal values are not covered by warranty.
Specifications are subject to change without notice. For the most recent NI PXIe-1491
specifications, visit ni.com/manuals.
Caution Refer to the Read Me First: Safety and Electromagnetic Compatibility
document for important safety and electromagnetic compatibility information. To
obtain a copy of this document online, visit ni.com/manuals and search for the
document title.
To access documentation for the software applications that support the NI PXIe-1491, navigate
to Start»All Programs»National Instruments»<Application>»Documentation, where
<Application> stands for NI PQA 1.5 or later, NI VMS 3.2 or later, or NI AudioMASTER 6.3
or later, depending on which applications you have installed.
Caution To ensure the specified EMC performance, operate this product only with
HDMI-compliant cables and accessories.
Contents
Hardware Block Diagram......................................................................................................... 2
HDMI Input (HDMI IN connector).......................................................................................... 3
Monitor Output (MONITOR OUT connector)......................................................................... 4
Video Triggers...........................................................................................................................4
Audio Triggers.......................................................................................................................... 4
Software.................................................................................................................................... 5
Power........................................................................................................................................ 5
Dimensions and Weight............................................................................................................ 5
Front Panel Connectors and Indicators..................................................................................... 6
Environment..............................................................................................................................8
Operating Environment.....................................................................................................8
Storage Environment.........................................................................................................8
Shock and Vibration..................................................................................................................9
Compliance and Certifications..................................................................................................9
Safety................................................................................................................................ 9
Electromagnetic Compatibility......................................................................................... 9
CE Compliance .............................................................................................................. 10
Online Product Certification........................................................................................... 10
Environmental Management........................................................................................... 10
Hardware Block Diagram
The following figure shows the hardware block diagram of the NI PXIe-1491.
Figure 1. NI PXIe-1491 Hardware Block Diagram
S/PDIF for ARC
HDMI IN
HDMI Type A
TMDS + HDCP
DDC/CEC/HEAC
MONITOR OUT
HDMI Type A
DDC/CEC
|
ni.com
100BASE-T Tx/Rx
Receiver
Block
Video Clock + Data
Audio Clock + Data
Multimedia
Processor
Clocking
Sources
TMDS + HDCP Re-encryption
LAN LINK
HEC LINK
RJ-45
2
Signal
Conditioning
Ethernet
Switch
| NI PXIe-1491 Specifications
PXI Express Bus
and
Signal Interface
HDMI Input (HDMI IN connector)
HDMI
............................................................................
functions1
Video, audio, CEC, ARC, HEC, HPD
............................................................................
TMDS
clock rate2
25 MHz to 225 MHz
Video
............................................................................
clock rate
25 MHz to 165 MHz
HDMI version
............................................................................
1.4b
Maximum resolution3
60
....................................................................
Hz (HDTV)
1080p
30 Hz (3D HDTV)
....................................................................
1080p
60 Hz UXGA graphics
....................................................................
1600 × 1200
Color depth
............................................................................
8 bits to 12 bits per color channel4
Chroma formats
....................................................................
RGB
4:4:4
YCbCr
....................................................................
4:4:4 or 4:2:2
............................................................................
Audio
channels5
8-channel LPCM or encoded audio
Audio
............................................................................
sample rate
32 kHz to 192 kHz
ARC channels
............................................................................
2
ARC sample rates
............................................................................
32, 44.1, or 48 kHz
ARC sample rate accuracy, typical
............................................................................
35 ppm
HDCP version
............................................................................
1.4
HEC data rate, nominal
............................................................................
100 Mbps
Note To modify the capabilities of the NI PXIe-1491 as presented to the HDMI
source, use the NI EDID Update Utility. For example, you might want to do so in
order to match the HDMI capabilities of the NI PXIe-1491 to those of a device
under test.
1
2
3
4
5
This feature depends on the software support you install for the NI PXIe-1491.
The TMDS clock rate is higher than the video clock rate for deep color video formats.
The actual resolution in 3D depends on the 3D mode. Higher frame rates are possible for lower
spatial resolutions.
Deep color video is 10 bits or 12 bits per channel.
Depending on software support, you can program this feature.
NI PXIe-1491 Specifications
|
© National Instruments
| 3
Monitor Output (MONITOR OUT connector)
............................................................................
Function6
Repeater
............................................................................HDMI IN connector
Source
............................................................................
HDMI functions
Video, audio
Video Triggers
............................................................................
Trigger type
Start
............................................................................
Sources
Immediate, PXI_TRIG<0..7>
............................................................................
Detection
Immediate, PXI_TRIG<0..7>
............................................................................
Detection start
Edge (rising or falling)
............................................................................
Minimum detectable pulse width
64 ns
............................................................................
Destinations7
PXI_TRIG<0...7>
Audio Triggers
............................................................................
Trigger type
Start
............................................................................
Sources
Immediate, PXI_TRIG<0..7>, or Audio
channel <0..7>
Detection8
Audio
....................................................................
channel
Level crossing
....................................................................
PXI_TRIG<0..7>
Edge (rising or falling)
............................................................................
Minimum detectable pulse width
64 ns
............................................................................
Destinations9
PXI_TRIG<0...7>
6
7
8
9
4
|
The HDMI monitor output provides the content from the HDMI IN connector. The
video format and frame rate for the monitor output are the same as for the input.
200 ns pulse width, typical.
This feature depends on the software support you install for the NI PXIe-1491. Depending on
support, you can program the level and hysteresis.
200 ns pulse width, typical.
ni.com
|
NI PXIe-1491 Specifications
Software
Driver
............................................................................
software
The NI PXIe-1491 is supported through
application software. When you install the
software support, you are prompted to install
any drivers required by the software.
............................................................................
Application software
Install one of the following software
applications to automatically install software
support for the NI PXIe-1491: NI Picture
Quality Analysis (PQA) 1.5 or later, NI Video
Measurement Suite (VMS) 3.2 or later, or
NI AudioMASTER 6.3 or later.
............................................................................
Application development
Controlling the NI PXIe-1491 is supported in
environments
LabVIEW via NI PQA or NI VMS. NI CEC,
NI EDID, and NI HPD provide additional
functionality.
............................................................................
Testing sequence environment
NI AudioMASTER, NI PQA, and NI VMS
support controlling the NI PXIe-1491 through
NI TestStand.
Note Refer to the NI PXIe-1491 Getting Started Guide for more information about
software support for the environments in which you can program the NI PXIe-1491.
Power
+3.3
............................................................................
VDC
2.3 A
+12 VDC
............................................................................
2.4 A
+5 Vaux
............................................................................
None
Total
............................................................................
power
30 W
Dimensions and Weight
Dimensions
............................................................................3U, One Slot, PXI Express Module
21.6 × 2.0 × 13.0 cm (8.5 × 0.8 × 5.1 in.)
Weight
............................................................................
208 g (7.3 oz)
NI PXIe-1491 Specifications |
© National Instruments
| 5
Front Panel Connectors and Indicators
This section describes the connectors and LED indicators on the NI PXIe-1491.
Figure 2. NI PXIe-1491 Front Panel
HDMI Analyzer
ACCESS
ACTIVE
SYNC
HDMI
IN
MONITOR
OUT
LAN LINK/ACT
HEC LINK/ACT
6
|
ni.com
| NI PXIe-1491 Specifications
Table 1. NI PXIe-1491 Front Panel Connectors
Connector
HDMI IN
Function(s)
Connector Type
HDMI-compliant input. Depending on software
support, supports audio return channel (ARC) and
Ethernet over HDMI (HEC).
HDMI type A
MONITOR OUT HDMI-compliant repeater output from the HDMI IN
connector.
HEC/LAN
HDMI type A
Provides 100BASE-T access to HEC from HDMI IN
RJ-45
connector path. Does not serve as an end-point adapter.
Table 2. NI PXIe-1491 Front Panel LEDs
LED
Indications
ACCESS
Indicates the following basic hardware statuses of the module:
•
OFF—The module is not yet initialized or detected a problem with a
power rail.
•
AMBER—The module is being accessed. Accessed means that the
device setup registers are being written to in order to control the device.
•
GREEN—The device is ready to be programmed.
•
RED—The device exceeded approved operating temperature and
thermal shutdown occurred.
ACTIVE
Indicates the following active states:
•
OFF—The device is not armed, triggered, or acquiring a waveform.
•
AMBER—The device is armed and waiting for a trigger.
•
GREEN—The device received a Reference trigger. A green LED also
indicates that the device is acquiring a waveform.
•
RED—The device detected an error. You must use the application
software to access the device to determine the cause of the error. The
LED remains red until the error condition is removed.
SYNC
Indicates the following states for the selected input:
•
OFF—No valid input signal from the source is detected, or the device is
not yet in an operational state.
•
GREEN—The device detected a valid input signal from the source.
NI PXIe-1491 Specifications |
© National Instruments
| 7
Table 2. NI PXIe-1491 Front Panel LEDs (Continued)
LED
Indications
LAN
LINK/ACT
Indicates the following states for the selected input:
•
GREEN—There is a link over the Ethernet cable, but there is no active
communication.
•
BLINKING GREEN—There is active communication over the Ethernet
cable.
HEC
LINK/ACT
Indicates the following states for the selected input:
•
YELLOW—The device is not yet functional, or there is a link over the
HDMI cable, but there is no active communication.
•
BLINKING YELLOW—There is active communication over the HDMI
cable.
Environment
Maximum
............................................................................
altitude
2,000 m (800 mbar) (at 25 °C ambient
temperature)
............................................................................
Pollution Degree
2
Indoor use only.
Operating Environment
Ambient temperature range
0 °C to 55 °C (Tested in accordance with
............................................................................
IEC-60068-2-1 and IEC-60068-2-2.)
............................................................................
Relative humidity range
10% to 90%, noncondensing (Tested in
accordance with IEC-60068-2-56.)
Storage Environment
Ambient
temperature range
-40 °C to 71 °C (Tested in accordance
............................................................................
with IEC-60068-2-1 and IEC-60068-2-2.
Meets MIL-PRF-28800F Class 3 limits.)
Relative humidity range
5% to 95%, noncondensing (Tested in
............................................................................
accordance with IEC-60068-2-56.)
8
|
ni.com
| NI PXIe-1491 Specifications
Shock and Vibration
Operating shock
............................................................................
30 g peak, half-sine, 11 ms pulse (Tested in
accordance with IEC-60068-2-27. Meets
MIL-PRF-28800F Class 2 limits.)
Random vibration
....................................................................
Operating
5 Hz to 500 Hz, 0.3 grms
Nonoperating
....................................................................
5 Hz to 500 Hz, 2.4 grms (Tested in accordance
with IEC-60068-2-64. Nonoperating test
profile exceeds the requirements of
MIL-PRF-28800F, Class 3.)
Compliance and Certifications
Safety
This product is designed to meet the requirements of the following electrical equipment safety
standards for measurement, control, and laboratory use:
•
IEC 61010-1, EN 61010-1
•
UL 61010-1, CSA 61010-1
Note For UL and other safety certifications, refer to the product label or the Online
Product Certification section.
Electromagnetic Compatibility
This product meets the requirements of the following EMC standards for electrical equipment
for measurement, control, and laboratory use:
EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
•
•
EN 55011 (CISPR 11): Group 1, Class A emissions
•
EN 55022 (CISPR 22): Class A emissions
•
EN 55024 (CISPR 24): Immunity
•
AS/NZS CISPR 11: Group 1, Class A emissions
•
AS/NZS CISPR 22: Class A emissions
•
FCC 47 CFR Part 15B: Class A emissions
•
ICES-001: Class A emissions
Note In the United States (per FCC 47 CFR), Class A equipment is intended for
use in commercial, light-industrial, and heavy-industrial locations. In Europe,
Canada, Australia, and New Zealand (per CISPR 11), Class A equipment is intended
for use only in heavy-industrial locations.
NI PXIe-1491 Specifications |
© National Instruments
| 9
Note Group 1 equipment (per CISPR 11) is any industrial, scientific, or medical
equipment that does not intentionally generate radio frequency energy for the
treatment of material or inspection/analysis purposes.
Note For EMC declarations, certifications, and additional information, refer to the
Online Product Certification section.
CE Compliance
This product meets the essential requirements of applicable European Directives, as follows:
•
2006/95/EC; Low-Voltage Directive (safety)
•
2004/108/EC; Electromagnetic Compatibility Directive (EMC)
Online Product Certification
To obtain product certifications and the DoC for this product, visit ni.com/certification, search
by model number or product line, and click the appropriate link in the Certification column.
Environmental Management
NI is committed to designing and manufacturing products in an environmentally responsible
manner. NI recognizes that eliminating certain hazardous substances from our products is
beneficial to the environment and to NI customers.
For additional environmental information, refer to the Minimize Our Environmental Impact
web page at ni.com/environment. This page contains the environmental regulations and
directives with which NI complies, as well as other environmental information not included in
this document.
Waste Electrical and Electronic Equipment (WEEE)
EU Customers At the end of the product life cycle, all products must be sent to a
WEEE recycling center. For more information about WEEE recycling centers,
National Instruments WEEE initiatives, and compliance with
WEEE Directive 2002/96/EC on Waste Electrical and Electronic Equipment, visit
ni.com/environment/weee.
电子信息产品污染控制管理办法(中国 RoHS)
中国客户 National Instruments 符合中国电子信息产品中限制使用某些有害物
质指令(RoHS)。关于 National Instruments 中国 RoHS 合规性信息,请登录
ni.com/environment/rohs_china。(For information about China RoHS
compliance, go to ni.com/environment/rohs_china.)
10
| ni.com
|
NI PXIe-1491 Specifications
Refer to the NI Trademarks and Logo Guidelines at ni.com/trademarks for information on National Instruments trademarks.
Other product and company names mentioned herein are trademarks or trade names of their respective companies. For patents
covering National Instruments products/technology, refer to the appropriate location: Help»Patents in your software, the
patents.txt file on your media, or the National Instruments Patent Notice at ni.com/patents. You can find information about
end-user license agreements (EULAs) and third-party legal notices in the readme file for your NI product. Refer to the Export
Compliance Information at ni.com/legal/export-compliance for the National Instruments global trade compliance policy and
how to obtain relevant HTS codes, ECCNs, and other import/export data. NI MAKES NO EXPRESS OR IMPLIED WARRANTIES
AS TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY ERRORS.
U.S. Government Customers: The data contained in this manual was developed at private expense and is subject to the
applicable limited rights and restricted data rights as set forth in FAR 52.227-14, DFAR 252.227-7014, and DFAR 252.227-7015.
© 2012—2014 National Instruments. All rights reserved.
375837B-01
Jun14