NI PXIe-2738 Specifications

DEVICE SPECIFICATIONS
NI PXIe-2738
256-Crosspoint Relay Matrix
This document lists specifications for the NI PXIe-2738 (NI 2738) matrix relay card. All
specifications are subject to change without notice. Visit ni.com/manuals for the most current
specifications.
Topology
............................................................................2-wire 8 × 32 matrix
Contents
About These Specifications...................................................................................................... 1
Input Characteristics................................................................................................................. 2
RF Performance Characteristics................................................................................................3
Dynamic Characteristics........................................................................................................... 4
Physical Characteristics............................................................................................................ 4
Environment..............................................................................................................................5
Operating Environment.....................................................................................................5
Storage Environment.........................................................................................................5
Shock and Vibration..................................................................................................................5
Diagrams................................................................................................................................... 6
Accessories................................................................................................................................7
Derating NI 2738 Load at >50 °C.............................................................................................8
Module Load Derating at >50 °C......................................................................................8
Compliance and Certifications..................................................................................................9
Safety................................................................................................................................ 9
Electromagnetic Compatibility......................................................................................... 9
CE Compliance .............................................................................................................. 10
Online Product Certification........................................................................................... 10
Environmental Management........................................................................................... 10
About These Specifications
Specifications characterize the warranted performance of the instrument under the stated
operating conditions.
Typical Specifications are specifications met by the majority of the instrument under the stated
operating conditions and are tested at 23 °C ambient temperature. Typical specifications are
not warranted.
All voltages are specified in DC, ACpk, or a combination unless otherwise specified.
Caution The protection provided by the NI 2738 can be impaired if it is used in a
manner not described in this document.
Input Characteristics
Maximum switching voltage
....................................................................
Channel-to-channel
100 V
Channel-to-ground
....................................................................
100 V, CAT O
Note This module is rated for Measurement Category I. It is intended to carry
signal voltages no greater than 70 Vrms, 100 Vpk, or 100 VDC. This module can
withstand up to 500 V impulse voltage. Do not use this module for connection to
signals or for measurements within Categories II, III, or IV. Do not connect to
MAINS supply circuits (for example, wall outlets) of 115 VAC or 230 VAC.
Measurement Categories CAT I and CAT O (Other) are equivalent and are for measurements
performed on circuits not directly connected to the electrical distribution system referred to as
MAINS voltage. This category is for measurement of voltages from specially protected
secondary circuits. Such voltage measurements include signal levels, special hardware,
limited-energy parts of hardware, circuits powered by regulated low-voltage sources, and
electronics.
Caution When hazardous voltages (>42.4 Vpk/60 VDC) are present on any
channel, safety low-voltage (≤42.4 Vpk/60 VDC) cannot be connected to any other
channel.
............................................................................
Maximum switching power
60 W, 62.5 VA
Caution The switching power is limited by the maximum switching current, the
maximum voltage, and must not exceed 60 W, 62.5 VA.
Maximum current
Switching
....................................................................2 A (per channel)
Carry
....................................................................
2 A (per channel)
Minimum switching load
............................................................................
20 mV/1 mA
Note Switching inductive loads (for example, motors and solenoids) can produce
high voltage transients in excess of the module’s rated voltage. Without additional
protection, these transients can interfere with module operation and impact relay life.
2
|
ni.com
|
NI PXIe-2738 Specifications
For more information about transient suppression, visit ni.com/info and enter the
Info Code relayflyback.
DC
............................................................................
isolation resistance
1 GΩ (typical at 25º C)
(channel-to-COM)
DC path resistance
Initial
....................................................................
<1.3 Ω
End-of-life
....................................................................
≥2.3 Ω
Path resistance is a combination of relay contact resistance and trace resistance. Measure path
resistance by combining resistance of the high and low signal paths from one row to one
column. Contact resistance typically remains low for the life of a relay. At the end of relay life,
the contact resistance rises rapidly above ≥2.3 Ω.
Thermal
............................................................................
EMF
<10 µV
RF Performance Characteristics
Bandwidth (-3 dB, typical at 23º C)
....................................................................
1-wire 50 Ω termination
≤ 10 MHz
2-wire 100 Ω termination
....................................................................
≤ 10 MHz
Open channel isolation (typical at 23º C, 50 Ω
termination )
10 kHz
....................................................................
≥ 84 dB
100 kHz
....................................................................
≥ 71 dB
1 MHz
....................................................................
≥ 50 dB
Channel-to-channel crosstalk (typical at 23º C,
50 Ω termination )
10 kHz
....................................................................
≤ -69 dB
100 kHz
....................................................................
≤ -52 dB
1 MHz
....................................................................
≤ -33 dB
NI PXIe-2738 Specifications
|
© National Instruments
|
3
Dynamic Characteristics
Relay Operate Time
....................................................................
Typical
1 ms
Maximum
....................................................................
3.3 ms
Simultaneous drive limit
............................................................................
90 relays
Expected mechanical relay life
............................................................................
1 × 108 cycles
Expected electrical relay life
≤30 mV, ≤10 mA resistive
....................................................................
2.5 × 106 cycles
30 V, 1 A
....................................................................
5 × 105 cycles
30 V, 2 A
....................................................................
1 × 105 cycles
60 VDC, 1 ADC resistive
....................................................................
1 × 105 cycles
Note Relays are field replaceable. Refer to the NI Switches Help at ni.com/
manuals for more information about replacing a failed relay.
Note Certain applications may require additional time for proper settling. Refer to
the NI Switches Help at ni.com/manuals for more information about including
additional settling time.
Physical Characteristics
............................................................................
Relay type
Electromechanical, non-latching
Relay contact material
............................................................................
Palladium-ruthenium, gold covered
Front panel connector
............................................................................
160 DIN 41612, 160 positions, male
Power requirement
PXI Express
12 V
............................................................
24 W
3.3 V
............................................................
3.6 W
Dimensions (L × W × H)
............................................................................
3U, one slot, PXI/cPCI module, PXI Express
compatible 21.6 cm × 2.0 cm × 13.0 cm
(8.5 in. × 0.8 in. × 5.1 in.)
Weight
............................................................................423 g (14.9 oz)
4
|
ni.com |
NI PXIe-2738 Specifications
Environment
Maximum
............................................................................
altitude
2,000 m (800 mbar) (at 25 °C ambient
temperature)
Pollution Degree
............................................................................
2
Indoor use only.
Operating Environment
............................................................................
Ambient temperature range
0 °C to 55 °C (Tested in accordance with
IEC 60068-2-1 and IEC 60068-2-2. Meets
MIL-PRF-28800F Class 3 low temperature
limit and MIL-PRF-28800F Class 2 high
temperature limit.)
Relative
............................................................................
humidity range
10% to 90%, noncondensing (Tested in
accordance with IEC 60068-2-56.)
Related Information
Derating NI 2738 Load at >50 °C on page 8
Storage Environment
Ambient
............................................................................
temperature range
-40 °C to 71 °C (Tested in accordance
with IEC 60068-2-1 and IEC 60068-2-2. Meets
MIL-PRF-28800F Class 3 limits.)
Relative
............................................................................
humidity range
5% to 95%, noncondensing (Tested in
accordance with IEC 60068-2-56.)
Shock and Vibration
............................................................................
Operating shock
30 g peak, half-sine, 11 ms pulse (Tested in
accordance with IEC 60068-2-27. Meets
MIL-PRF-28800F Class 2 limits.)
Random vibration
Operating
....................................................................
5 Hz to 500 Hz, 0.3 grms
Nonoperating
....................................................................
5 Hz to 500 Hz, 2.4 grms (Tested in accordance
with IEC 60068-2-64. Nonoperating test
profile exceeds the requirements of
MIL-PRF-28800F, Class 3.)
NI PXIe-2738 Specifications
|
© National Instruments
|
5
Diagrams
The following figure shows the NI 2738 power-on state.
Figure 1. NI 2738 Power-On State
(C0+, C0–)
(C1+, C1–)
(C31+, C31–)
(R0+, R0–)
(R1+, R1–)
(R6+, R6–)
(C1–)
(C1+)
(R7+, R7–)
(R6+)
(R6–)
The following figure shows the NI 2738 connector pinout.
6
|
ni.com
|
NI PXIe-2738 Specifications
Figure 2. NI 2738 Connector Pinout
CO–
CO+
C2–
C2+
C4–
C4+
C6–
C6+
C8–
C8+
C10–
C10+
C12–
C12+
C14–
C14+
C16–
C16+
C18–
C18+
C20–
C20+
C22–
C22+
C24–
C24+
C26–
C26+
C28–
C28+
C30–
C30+
B32
A32
B31
A31
B30
A30
B29
A29
B28
A28
B27
A27
B26
A26
B25
A25
B24
A24
B23
A23
B22
A22
B21
A21
B20
A20
B19
A19
B18
A18
B17
A17
C0–
C32
D32
E32
C31
D31
E31
C30
D30
E30
C29
D29
E29
C28
D28
E28
C27
D27
E27
C26
D26
E26
C25
D25
E25
C24
D24
E24
C23
D23
E23
C22
D22
E22
C21
D21
E21
C20
D20
E20
C19
D19
E19
C18
D18
E18
C17
D17
E17
C16
C1+
C1–
R0+
C3+
C3–
R0–
C5+
C5–
R1+
C7+
C7–
R1–
C9+
C9–
R2+
C11+
C11–
R2–
C13+
C13–
R3+
C15+
C15–
R3–
C17+
C17–
R0+
C19+
C19–
R0–
C21+
C21–
R1+
C23+
C23–
R1–
C25+
C25–
R2+
C27+
C27–
R2–
C29+
C29–
R3+
C31+
C31–
R3–
C1+
C1–
C30+
C0–
C0+
C2–
C2+
C4–
C4+
C6–
C6+
C8–
C8+
C10–
C10+
C12–
C12+
C14–
C14+
C16–
C16+
C18–
C18+
C20–
C20+
C22–
C22+
C24–
C24+
C26–
C26+
C28–
C28+
C30–
C30+
A17
C16
D16
E16
C15
D15
E15
C14
D14
E14
B16
A16
B15
A15
B14
A14
C13
D13
E13
C12
D12
E12
B13
A13
B12
A12
C11
D11
E11
B11
A11
C10
D10
E10
B10
A10
C9
D9
E9
C8
D8
E8
B9
A9
B8
A8
C7
D7
E7
C6
D6
E6
C5
D5
E5
C4
D4
E4
B7
A7
B6
A6
B5
A5
B4
A4
C3
D3
E3
B3
A3
C2
D2
E2
C1
D1
E1
B2
A2
B1
A1
C1+
C1–
R4+
C3+
C3–
R4–
C5+
C5–
R5+
C7+
C7–
R5–
C9+
C9–
R6+
C11+
C11–
R6–
C13+
C13–
R7+
C15+
C15–
R7–
C17+
C17–
R4+
C19+
C19–
R4–
C21+
C21–
R5+
C23+
C23–
R5–
C25+
C25–
R6+
C27+
C27–
R6–
C29+
C29–
R7+
C31+
C31–
R7–
Accessories
Table 1. NI Accessories for the NI 2738
Accessory
Part Number
DIN160 to 50-pin D-SUB switch cable, 1 m
782417-03
DIN160 to DIN160 switch cable, 1 m
782417-02
NI PXIe-2738 Specifications
|
© National Instruments
|
7
Table 1. NI Accessories for the NI 2738 (Continued)
Accessory
Part Number
DIN160 to bare wire switch cable, 1 m
782417-01
Relay replacement kit for IM02PNS Relays
781089-10
Derating NI 2738 Load at >50 °C
To verify you are operating the NI 2738 within supported 2-wire derating conditions, complete
the following steps:
1. Use the following equation to calculate the load:
Load = Channels1 × Current1 2 + Channels2 × Current2 2 + ... + Channels�
× Current� 2
where Channels is the number of channels that simultaneously carry a signal, Current, for
1, ... , n.
2.
Verify that the load at your ambient operating temperature falls within the shaded region
of the following figure.
Figure 3. NI 2738 Load Derating, 2-Wire
35
Power Load (Amps 2 x Channels)
30
25
20
15
10
5
0
15
20
25
35
40
45
50
Ambient Temperature (°C)
Module Load Derating at >50 °C
The following examples calculate supported derating conditions for the NI 2738.
8
|
ni.com |
NI PXIe-2738 Specifications
55
Example 1
where
(3 × 22) + (4 × 1.72) = 23.6�2 × channels
3 channels carry 2 A
4 channels carry 1.7 A
You can use this module at ambient temperatures between 0 °C and
55 °C.
Example 2
where
(8 × 22) = 32�2 × channels
8 channels carry 2 A
You can use this module at ambient temperatures between 0 °C and
50 °C.
Compliance and Certifications
Safety
This product is designed to meet the requirements of the following electrical equipment safety
standards for measurement, control, and laboratory use:
•
IEC 61010-1, EN 61010-1
•
UL 61010-1, CSA 61010-1
Note For UL and other safety certifications, refer to the product label or the Online
Product Certification section.
Electromagnetic Compatibility
This product meets the requirements of the following EMC standards for electrical equipment
for measurement, control, and laboratory use:
•
EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
•
EN 55011 (CISPR 11): Group 1, Class A emissions
•
EN 55022 (CISPR 22): Class A emissions
•
EN 55024 (CISPR 24): Immunity
NI PXIe-2738 Specifications
|
© National Instruments
|
9
•
•
•
•
AS/NZS CISPR 11: Group 1, Class A emissions
AS/NZS CISPR 22: Class A emissions
FCC 47 CFR Part 15B: Class A emissions
ICES-001: Class A emissions
Note In the United States (per FCC 47 CFR), Class A equipment is intended for
use in commercial, light-industrial, and heavy-industrial locations. In Europe,
Canada, Australia, and New Zealand (per CISPR 11), Class A equipment is intended
for use only in heavy-industrial locations.
Note Group 1 equipment (per CISPR 11) is any industrial, scientific, or medical
equipment that does not intentionally generate radio frequency energy for the
treatment of material or inspection/analysis purposes.
Note For EMC declarations, certifications, and additional information, refer to the
Online Product Certification section.
CE Compliance
This product meets the essential requirements of applicable European Directives, as follows:
•
2006/95/EC; Low-Voltage Directive (safety)
•
2014/30/EU; Electromagnetic Compatibility Directive (EMC)
Online Product Certification
Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance
information. To obtain product certifications and the DoC for this product, visit ni.com/
certification, search by model number or product line, and click the appropriate link in the
Certification column.
Environmental Management
NI is committed to designing and manufacturing products in an environmentally responsible
manner. NI recognizes that eliminating certain hazardous substances from our products is
beneficial to the environment and to NI customers.
For additional environmental information, refer to the Minimize Our Environmental Impact
web page at ni.com/environment. This page contains the environmental regulations and
directives with which NI complies, as well as other environmental information not included in
this document.
Waste Electrical and Electronic Equipment (WEEE)
EU Customers At the end of the product life cycle, all NI products must be
disposed of according to local laws and regulations. For more information about
how to recycle NI products in your region, visit ni.com/environment/weee.
10
|
ni.com
|
NI PXIe-2738 Specifications
电子信息产品污染控制管理办法(中国 RoHS)
中国客户 National Instruments 符合中国电子信息产品中限制使用某些有害物
质指令(RoHS)。关于 National Instruments 中国 RoHS 合规性信息,请登录
ni.com/environment/rohs_china。(For information about China RoHS
compliance, go to ni.com/environment/rohs_china.)
NI PXIe-2738 Specifications |
© National Instruments
|
11
Refer to the NI Trademarks and Logo Guidelines at ni.com/trademarks for information on National Instruments trademarks.
Other product and company names mentioned herein are trademarks or trade names of their respective companies. For patents
covering National Instruments products/technology, refer to the appropriate location: Help»Patents in your software, the
patents.txt file on your media, or the National Instruments Patent Notice at ni.com/patents. You can find information about
end-user license agreements (EULAs) and third-party legal notices in the readme file for your NI product. Refer to the Export
Compliance Information at ni.com/legal/export-compliance for the National Instruments global trade compliance policy and
how to obtain relevant HTS codes, ECCNs, and other import/export data. NI MAKES NO EXPRESS OR IMPLIED WARRANTIES
AS TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY ERRORS.
U.S. Government Customers: The data contained in this manual was developed at private expense and is subject to the
applicable limited rights and restricted data rights as set forth in FAR 52.227-14, DFAR 252.227-7014, and DFAR 252.227-7015.
© 2014 National Instruments. All rights reserved.
376202A-01