Topology
16 × 1 multiplexer
Contents
About These Specifications...................................................................................................... 1
Input Characteristics................................................................................................................. 2
RF Performance Characteristics................................................................................................3
Dynamic Characteristics........................................................................................................... 4
Physical Characteristics............................................................................................................ 5
Environment..............................................................................................................................5
Shock and Vibration..................................................................................................................5
Diagrams................................................................................................................................... 5
Accessories................................................................................................................................7
Compliance and Certifications..................................................................................................8
Safety................................................................................................................................ 8
Electromagnetic Compatibility......................................................................................... 8
CE Compliance ................................................................................................................ 9
Online Product Certification............................................................................................. 9
Environmental Management............................................................................................. 9
About These Specifications
Specifications characterize the warranted performance of the instrument under the stated
operating conditions.
Typical Specifications are specifications met by the majority of the instrument under the stated
operating conditions and are tested at 23 °C ambient temperature. Typical specifications are
not warranted.
All voltages are specified in DC, ACpk, or a combination unless otherwise specified.
Caution The protection provided by the NI 2748 can be impaired if it is used in a
manner not described in this document.
Input Characteristics
All input characteristics are DC, ACrms, or a combination unless otherwise specified.
Maximum
............................................................................
switching voltage
30 V
............................................................................
Maximum
switching current
0.5 A
(per channel)
............................................................................
Maximum
carry current
0.5 A
(per channel)
Maximum RF power
............................................................................
10 W
Caution The switching power is limited by the maximum switching current and
the maximum voltage. Channel-to-common switching power must not exceed 10 W.
Note National Instruments recommends against switching active RF signals. As a
relay actuates, the channel is momentarily unterminated. Some RF sources can be
damaged by reflections if their outputs are not properly terminated. Refer to your RF
source documentation for more information.
Note Switching active RF signals below 0 dBm may degrade signal integrity and
decrease relay life. For more information about switching active RF signals, visit
ni.com/info and enter Info Code RFSwitching.
Typical DC path resistance
....................................................................
Initial
<1 Ω
End-of-life
....................................................................
≥2 Ω
Path resistance is a combination of relay contact resistance and trace resistance. Contact
resistance typically remains low for the life of a relay. At the end of relay life, the contact
resistance rises rapidly above 1.0 Ω.
2
|
ni.com |
NI PXIe-2748 Specifications
RF Performance Characteristics
Characteristic
............................................................................
impedance (Z0)
50 Ω nominal
Insertion loss
....................................................................
≤1 GHz
1.5 dB
≤3 GHz
....................................................................
3 dB
Voltage standing wave ratio (VSWR)
≤1
....................................................................
GHz
1.4
≤3 GHz
....................................................................
1.6
Typical isolation
≤1 GHz
....................................................................
>60 dB
≤3 GHz
....................................................................
>40 dB
Typical channel-to-channel skew
............................................................................
<160 ps
Refer to the following figures for typical insertion loss, typical VSWR, and typical isolation,
respectively.
Figure 1. Typical Insertion Loss
Frequency (GHz)
0
0.5
1
1.5
2
2.5
3
0.00E+00
Typical Insertion Loss
–5.00E–01
–1.00E+00
–1.50E+00
–2.00E+00
–2.50E+00
–3.00E+00
NI PXIe-2748 Specifications
|
© National Instruments
|
3
Figure 2. Typical VSWR
1.50E+00
1.45E+00
1.40E+00
Typical VSWR
1.35E+00
1.30E+00
1.25E+00
1.20E+00
1.15E+00
1.10E+00
1.05E+00
1.00E+00
0
0.5
1
1.5
2
2.5
3
Frequency (GHz)
Figure 3. Typical Isolation
Frequency (GHz)
0
1
2
3
4
0
–10
–20
Typical Isolation
–30
–40
–50
–60
–70
–80
–90
–100
Dynamic Characteristics
............................................................................
Maximum relay operate time
10 ms
Note Certain applications may require additional time for proper settling. Refer to
the NI Switches Help at ni.com/manuals for more information about including
additional settling time.
............................................................................
Expected mechanical relay life
1 × 107 cycles
4
|
ni.com |
NI PXIe-2748 Specifications
Physical Characteristics
Relay
............................................................................
type
Electromechanical, latching
I/O connectors
............................................................................
17 MCX jacks, gold plated
PXI
............................................................................
Express power requirement
3.96 W at 3.3 V, 0.2 W at 12 V
Dimensions (L × W × H)
............................................................................
3U, one slot, PXI/cPCI module
21.6 cm × 2.0 cm × 13.0 cm
(8.5 in. × 0.8 in. × 5.1 in.)
Weight
............................................................................
330 g (11.30 oz)
Environment
............................................................................
Operating temperature
0 °C to 55 °C
Storage temperature
............................................................................
-40 °C to 70 °C
Relative humidity
............................................................................
5% to 85%, noncondensing
Pollution Degree
............................................................................
2
Maximum altitude
............................................................................
2,000 m
Indoor use only.
Shock and Vibration
............................................................................
Operational Shock
30 g peak, half-sine, 11 ms pulse (Tested in
accordance with IEC 60068-2-27. Test profile
developed in accordance with
MIL-PRF-28800F.)
Random Vibration
Operating
....................................................................
5 Hz to 500 Hz, 0.3 grms
Nonoperating
....................................................................
5 Hz to 500 Hz, 2.4 grms (Tested in accordance
with IEC 60068-2-64. Nonoperating test
profile exceeds the requirements of
MIL-PRF-28800F, Class 3.)
Diagrams
The following figure shows the NI PXIe-2748 power-on state.
NI PXIe-2748 Specifications
|
© National Instruments
|
5
Figure 4. NI PXIe-2748 Power-On State
CH0
CH1
CH2
CH3
CH4
CH5
CH6
CH7
COM
CH8
CH9
CH10
CH11
CH12
CH13
CH14
CH15
The following figure shows the NI PXIe-2748 front panel connector.
6
|
ni.com
|
NI PXIe-2748 Specifications
Figure 5. NI PXIe-2748 Front Panel
3 GHz 16 x 1
50 Ω Multiplexer
CH
0
CH
8
CH
1
CH
9
CH
2
CH
10
CH
3
CH
11
COM
CH
4
CH
12
CH
5
CH
13
CH
6
CH
14
CH
7
CH
15
MAX
10W
Accessories
Visit ni.com for more information about the following accessories.
NI PXIe-2748 Specifications
|
© National Instruments
|
7
Table 1. NI Accessories for the NI PXIe-2748
Accessory
Length
Part Number
MCX-MCX cable
0.3 m
188374-0R3
MCX-MCX cable
1.0 m
188374-01
MCX-BNC cable
0.3 m
188375-0R3
MCX-BNC cable
1.0 m
188375-01
MCX-SMB cable
0.3 m
188376-0R3
MCX-SMB cable
1.0 m
188376-01
MCX-SMA cable
0.3 m
188377-0R3
MCX-SMA cable
1.0 m
188377-01
Compliance and Certifications
Safety
This product is designed to meet the requirements of the following electrical equipment safety
standards for measurement, control, and laboratory use:
•
IEC 61010-1, EN 61010-1
•
UL 61010-1, CSA 61010-1
Note For UL and other safety certifications, refer to the product label or the Online
Product Certification section.
Electromagnetic Compatibility
This product meets the requirements of the following EMC standards for electrical equipment
for measurement, control, and laboratory use:
•
EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
•
EN 55011 (CISPR 11): Group 1, Class A emissions
•
EN 55022 (CISPR 22): Class A emissions
•
EN 55024 (CISPR 24): Immunity
•
AS/NZS CISPR 11: Group 1, Class A emissions
•
AS/NZS CISPR 22: Class A emissions
•
FCC 47 CFR Part 15B: Class A emissions
•
ICES-001: Class A emissions
Note In the United States (per FCC 47 CFR), Class A equipment is intended for
use in commercial, light-industrial, and heavy-industrial locations. In Europe,
8
|
ni.com
|
NI PXIe-2748 Specifications
Canada, Australia, and New Zealand (per CISPR 11), Class A equipment is intended
for use only in heavy-industrial locations.
Note Group 1 equipment (per CISPR 11) is any industrial, scientific, or medical
equipment that does not intentionally generate radio frequency energy for the
treatment of material or inspection/analysis purposes.
Note For EMC declarations, certifications, and additional information, refer to the
Online Product Certification section.
CE Compliance
This product meets the essential requirements of applicable European Directives, as follows:
•
2014/35/EU; Low-Voltage Directive (safety)
•
2014/30/EU; Electromagnetic Compatibility Directive (EMC)
Online Product Certification
Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance
information. To obtain product certifications and the DoC for this product, visit ni.com/
certification, search by model number or product line, and click the appropriate link in the
Certification column.
Environmental Management
NI is committed to designing and manufacturing products in an environmentally responsible
manner. NI recognizes that eliminating certain hazardous substances from our products is
beneficial to the environment and to NI customers.
For additional environmental information, refer to the Minimize Our Environmental Impact
web page at ni.com/environment. This page contains the environmental regulations and
directives with which NI complies, as well as other environmental information not included in
this document.
Waste Electrical and Electronic Equipment (WEEE)
EU Customers At the end of the product life cycle, all NI products must be
disposed of according to local laws and regulations. For more information about
how to recycle NI products in your region, visit ni.com/environment/weee.
电子信息产品污染控制管理办法(中国 RoHS)
中国客户 National Instruments 符合中国电子信息产品中限制使用某些有害物
质指令(RoHS)。关于 National Instruments 中国 RoHS 合规性信息,请登录
ni.com/environment/rohs_china。(For information about China RoHS
compliance, go to ni.com/environment/rohs_china.)
NI PXIe-2748 Specifications
|
© National Instruments
|
9
Refer to the NI Trademarks and Logo Guidelines at ni.com/trademarks for information on National Instruments trademarks.
Other product and company names mentioned herein are trademarks or trade names of their respective companies. For patents
covering National Instruments products/technology, refer to the appropriate location: Help»Patents in your software, the
patents.txt file on your media, or the National Instruments Patent Notice at ni.com/patents. You can find information about
end-user license agreements (EULAs) and third-party legal notices in the readme file for your NI product. Refer to the Export
Compliance Information at ni.com/legal/export-compliance for the National Instruments global trade compliance policy and
how to obtain relevant HTS codes, ECCNs, and other import/export data. NI MAKES NO EXPRESS OR IMPLIED WARRANTIES
AS TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY ERRORS.
U.S. Government Customers: The data contained in this manual was developed at private expense and is subject to the
applicable limited rights and restricted data rights as set forth in FAR 52.227-14, DFAR 252.227-7014, and DFAR 252.227-7015.
© 2015 National Instruments. All rights reserved.
376209A-01
Jun15