NI PXI-4132 Specifications Precision Source-Measure Unit This document provides the specifications for the NI PXI-4132 precision source-measure unit (SMU). Specifications are subject to change without notice. For the most recent NI PXI-4132 specifications, visit ni.com/manuals. National Instruments defines the capabilities and performance of its Test & Measurement instruments as Specifications, Typical Specifications, and Characteristic or Supplemental Specifications. Data provided in this document are Specifications unless otherwise noted. Specifications characterize the warranted performance of the instrument within the recommended calibration interval and under the stated operating conditions. Typical Specifications are specifications met by the majority of the instrument within the recommended calibration interval and under the stated operating conditions. The performance of the instrument is not warranted. Characteristic or Supplemental Specifications describe basic functions and attributes of the instrument established by design or during development and not evaluated during Verification or Adjustment. They provide information that is relevant for the adequate use of the instrument that is not included in the previous definitions. Unless otherwise noted, specifications are valid under the following conditions: • Ambient temperature 23 °C ± 5 °C • After 30 minute warm-up time • niDCPower Auto Zero property/attribute set to On To access the NI PXI-4132 documentation, navigate to Start»Programs»National Instruments»NI-DCPower»Documentation. Device Capabilities Table 1 and Figure 1 illustrate the voltage and the current source and sink ranges of the NI PXI-4132. I+ 100mA Table 1. NI PXI-4132 Current Source and Sink Ranges DC Voltage Ranges (CAT I) ±10 V ±100 V DC Current Source and Sink Ranges 10 μA 100 μA 1 mA 10 mA 100 mA Caution Shock hazards exist when voltage levels are greater than 30 VRMS, 42.4 V peak, or 60 VDC. Use extreme caution when a shock hazard is present. Always ensure the test system is de-energized before connecting or disconnecting the backshell assembly or cables from the NI PXI-4132. Quadrant II Sink Quadrant I Source 50mA 20mA V– –100V 20V 40V –20mA –40V –20V –50mA Quadrant III Source V+ 100V Quadrant IV Sink –100mA I– Figure 1. NI PXI-4132 Quadrant Diagram, characteristic SMU Specifications1 Voltage Programming Accuracy/Resolution Accuracy ± (% of output + offset) Range 1 Year 23 °C ± 5 °C Resolution Peak to Peak Noise, typical (0.1 Hz to 10 Hz) ±10 V 50 μV 0.025% + 3.0 mV 70 μV ±100 V 500 μV 0.025% + 10 mV 300 μV Current Programming Accuracy/Resolution2 Accuracy ± (% of output + offset) Range 1 Year 23 °C ± 5 °C Resolution Peak to Peak Noise, typical (0.1 Hz to 10 Hz) 10 μA 500 pA 0.034% + 2.0 nA 90 pA 100 μA 5 nA 0.034% + 20 nA 900 pA 1 mA 50 nA 0.034% + 0.2 μA 9 nA 10 mA 500 nA 0.034% + 2.0 μA 90 nA 100 mA 5 μA 0.034% + 20 μA 900 nA Voltage Measurement Accuracy/Resolution Accuracy ± (% of reading + offset) Range 1 Year 23 °C ± 5 °C Resolution ±10 V 10 μV 0.02% + 2.0 mV ±100 V 100 μV 0.02% + 5.0 mV Current Measurement Accuracy/Resolution Accuracy ± (% of reading + offset) Range 1 2 1 Year 23 °C ± 5 °C Resolution 10 μA 10 pA 0.028% + 1.0 nA 100 μA 100 pA 0.028% + 10 nA 1 mA 1 nA 0.028% + 0.1 μA 10 mA 10 nA 0.028% + 1.0 μA 100 mA 100 nA 0.020% + 10 μA Temperature coefficient (Tempco) is 15% of accuracy specification per °C. Minimum programmable current limit/level is 2% of range. NI PXI-4132 Specifications 2 ni.com Additional Specifications Figures 2 through 4 illustrate the step response of the NI PXI-4132 for different loads. Settling Time, typical...................... <300 μs Settled to 0.1% of final value (1 V step at 50% load of current range) 1.2 1 Amplitude (Volts) Transient Response, typical ............ Recovers to <0.1% of voltage range within 100 μs after a change in load current from 10% to 90% of current range Wideband Source Noise, typical.............................................. 8 mVp-p into resistive load <1 mV RMS (20 Hz to 20 MHz bandwidth) 0.8 0.6 0.4 0.2 <50 pF 3.3 μF 10 μF 0 –0.2 0 Remote Sense.................................. Add 0.5% of HI lead drop to voltage accuracy specification Maximum Lead Drop ................. Up to 1 V drop per lead 0.5 1.0 1.5 2.0 Time (milliseconds) Figure 2. 100 mA Range Step Response, typical Load Regulation Voltage ....................................... 0.5 mV per mA of output load using Local Sense Current........................................ 0.01% of range per volt of output change 1.2 Amplitude (Volts) 1 Guard Offset Voltage, typical ......... <4 mV (Current ≤ 10 mA) Isolation Voltage (continuous), characteristic Channel-to-earth Ground............ 150 VDC, CAT I, verified by dielectric withstand test, 5 s 0.8 0.6 0.4 0.2 <50 pF 330 nF 1 μF 0 –0.2 0 0.1 0.2 0.3 0.4 0.5 Time (milliseconds) Caution Do not connect to MAINs. Do not connect to signals or use for measurements within CAT II, III, or IV. Figure 3. 10 mA Range Step Response, typical 1.2 Amplitude (Volts) 1 0.8 0.6 0.4 10 μA 0.2 100 μA 1 mA <50 pF <50 pF <50 pF 330 pF 3.3 nF 33 nF 10 nF 100 nF 1 nF 0 –0.2 0 0.1 0.2 0.3 0.4 0.5 Time (milliseconds) Figure 4. 1 mA, 100 μA, and 10 μA Range Step Response, typical © National Instruments Corporation 3 NI PXI-4132 Specifications RMS (ppm of range) 600 Voltage Current 10 60 1 6 0.1 0.01 0.6 0.1 1 Peak-to-Peak (ppm of range) Output Triggers (Events) 100 Types .......................................... Source Complete, Sequence Iteration Complete, Sequence Engine Done, Measure Complete Destinations ................................ PXI Trigger lines 0–73 Polarity................................... Configurable Pulse Width............................ Configurable between 150 ns and 1.6 μs 10 Figure 6 illustrates the programming flow in NI-DCPower using Sequence source mode with automatic measurements. For more information about programming the NI PXI-4132, refer to the NI DC Power Supplies and SMUs Help. NPLC Figure 5. Measurement Noise vs. Aperture Time, typical Supplemental Specifications Sequence Loop Count Measurement Speed1 Num Steps Table 2. Maximum Operation Rates Per Second for 60 Hz (50 Hz) ADC Aperture Time Measure to Host 3490 (2900) 1900 (1700) 1/8 PLC 470 (390) 425 (360) 1 PLC 59.9 (49.9) 59.0 (49.3) “Step” Source Trigger Start Sequence Trigger Advance Trigger Source-measure to Host2 1/64 PLC “Sequence” Source Source Measure Delay Source Complete Event Measure Complete Event Figure 6. NI-DCPower Programming Flow Maximum Source Update Rate2 ......4,200 updates/s Additional Information Trigger In to Source Delay ..............500 ns Recommended Calibration Interval ............................................ One year Triggers Physical Characteristics Input Triggers Dimensions............................................ 3U, one-slot, PXI/cPCI module; 2.0 cm × 13.0 cm × 21.6 cm (0.8 in. × 5.1 in. × 8.5 in.) Types...........................................Start, Source, Sequence Advance, Measure Sources........................................PXI Trigger lines 0–73 Polarity ...................................Configurable Weight............................................. 295 g (10.4 oz) Minimum Pulse Width ...........100 ns Destinations4 ...............................PXI Trigger lines 0–73 Front panel connectors.................... COMBICON, 5.08 mm (8 position) Polarity ...................................Active high (unconfigurable) Note Front panel connectors can accept wire gauges from 12 AWG to 28 AWG. Pulse Width ............................150 ns 1 2 3 4 Sequence Sequence Iteration Engine Complete Done Event Event Does not include load dependent settling time; niDCPower Auto Zero property/attribute set to Off. Source-measure to Host and Maximum Source Update Rate are performed with the source delay set to 200 μs. This is sufficient for the output to settle within 1% of the requested level with a simple resistive load. As you adjust the source delay for your application’s requirements, maximum rates vary. Pulse widths and logic levels compliant with PXI specification. Input triggers can be reexported. NI PXI-4132 Specifications 4 ni.com Power Requirements PXI power requirement ...................10 W at 5 V, 1 W at 3.3 V, 2 W at 12 V Environment NI PXI-4132 Precision SMU A Maximum altitude ...........................2,000 m (at 25 °C ambient temperature) B OUTPUT STATUS REMOTE SENSE Pollution Degree..............................2 100 VDC MAX 100 mA MAX Indoor use only. C Operating Environment LO Ambient temperature range .............0 °C to 55 °C (Tested in accordance with IEC 60068-2-1 and IEC 60068-2-2.) SENSE LO D E GUARD F HI G H Relative humidity range ..................10% to 70%, noncondensing; derate 1.3% per °C above 40 °C (Tested in accordance with IEC 60068-2-56.) GUARD I J SENSE HI OUTPUT CAT I Storage Environment Ambient temperature range .............–40 °C to 70 °C (Tested in accordance with IEC 60068-2-1 and IEC 60068-2-2.) K Relative humidity range ..................5% to 95%, noncondensing (Tested in accordance with IEC 60068-2-56.) Shock and Vibration A B C D E F Status Indicator Sense Status Indicator Output Low Terminal Sense Low Terminal Guard Terminal Output High Terminal G H I J K Operational shock............................30 g peak, half-sine, 11 ms pulse (Tested in accordance with IEC 60068-2-27. Test profile developed in accordance with MIL-PRF-28800F.) Guard Terminal Guard Terminal Guard Terminal Sense High Terminal Chassis Ground Random vibration Figure 7. NI PXI-4132 Front Panel Operating ....................................5 Hz to 500 Hz, 0.3 grms Nonoperating ..............................5 Hz to 500 Hz, 2.4 grms (Tested in accordance with IEC 60068-2-64. Nonoperating test profile exceeds the requirements of MIL-PRF-28800F, Class 3.) © National Instruments Corporation 5 NI PXI-4132 Specifications Accessories Compliance and Certifications Visit ni.com for more information about the following accessory. Safety This product meets the requirements of the following standards of safety for electrical equipment for measurement, control, and laboratory use: Caution For safety, always operate the NI PXI-4132 with suitably rated cables and the backshell kit provided in the shipping kit. Operating the device without the provided backshell may expose users to high voltage. • IEC 61010-1, EN 61010-1 • UL 61010-1, CSA 61010-1 Note For UL and other safety certifications, refer to the product label or the Online Product Certification section. Caution The protection provided by this equipment may be impaired if it is used in a manner not described in the product documentation. 100V CAT I Electromagnetic Compatibility This product meets the requirements of the following EMC standards for electrical equipment for measurement, control, and laboratory use: Figure 8. Assembled High Voltage Backshell for the NI PXI-4132 Table 3. NI Accessory for the NI PXI-4132 Accessory Additional Connector and Backshell Kit for the NI PXI-4132 Part Number (P/N) 781175-01 COMBICON, 5.08 mm (8 position) Description Mating connector for I/O NI PXI-4132 Specifications Phoenix Contact • EN 55011 (CISPR 11): Group 1, Class A emissions • AS/NZS CISPR 11: Group 1, Class A emissions • FCC 47 CFR Part 15B: Class A emissions • ICES-001: Class A emissions Note For EMC compliance, operate this device with shielded cables and the ferrite included in the NI PXI-4132 shipping kit. CE Compliance This product meets the essential requirements of applicable European Directives as follows: Table 4. Third-Party Accessory for the NI PXI-4132 Accessory EN 61326 (IEC 61326): Class A emissions; Basic immunity Note For the standards applied to assess the EMC of this product, refer to the Online Product Certification section. Caution You must install mating connectors according to local safety codes and standards and according to the specifications provided by the manufacturer. You are responsible for verifying the safety compliance of third-party connectors and their usage according to the relevant standard(s), including UL and CSA in North America and IEC and VDE in Europe. Manufacturer • P/N • 2006/95/EC; Low-Voltage Directive (safety) • 2004/108/EC; Electromagnetic Compatibility Directive (EMC) 1825378 6 ni.com Online Product Certification Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for this product, visit ni.com/certification, search by model number or product line, and click the appropriate link in the Certification column. Environmental Management National Instruments is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial not only to the environment but also to NI customers. For additional environmental information, refer to the NI and the Environment Web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document. Waste Electrical and Electronic Equipment (WEEE) EU Customers At the end of the product life cycle, all products must be sent to a WEEE recycling center. For more information about WEEE recycling centers, National Instruments WEEE initiatives, and compliance with WEEE Directive 2002/96/EC on Waste Electrical and Electronic Equipment, visit ni.com/environment/weee. ⬉ᄤֵᙃѻક∵ᶧࠊㅵ⧚ࡲ⊩ ˄Ё RoHS˅ Ёᅶ᠋ National Instruments ヺড়Ё⬉ᄤֵ ᙃѻકЁ䰤ࠊՓ⫼ᶤѯ᳝ᆇ⠽䋼ᣛҸ (RoHS)DŽ݇Ѣ National Instruments Ё RoHS ড়㾘ᗻֵᙃˈ 䇋ⱏᔩ ni.com/environment/rohs_chinaDŽ (For information about China RoHS compliance, go to ni.com/environment/rohs_china.) © National Instruments Corporation 7 NI PXI-4132 Specifications National Instruments, NI, ni.com, and LabVIEW are trademarks of National Instruments Corporation. Refer to the Terms of Use section on ni.com/legal for more information about National Instruments trademarks. Other product and company names mentioned herein are trademarks or trade names of their respective companies. For patents covering National Instruments products/technology, refer to the appropriate location: Help»Patents in your software, the patents.txt file on your media, or the National Instruments Patent Notice at ni.com/patents. © 2009 National Instruments Corporation. All rights reserved. 375128B Nov09 NI PXI-4132 仕様 高精度ソースメジャーユニット このドキュメントには、NI PXI-4132 高精度ソースメジャーユニット(SMU)の仕様が記載されています。仕様は 事前の通知なしに変更されることがあります。最新の NI PXI-4132 の仕様については、ni.com/manuals を参照し てください。 ナショナルインスツルメンツでは、テスト・計測用機器の機能および性能を「仕様」、「標準仕様」 、および「特性 または補足仕様」として定義しています。このドキュメントに記載されたデータは注釈がない限り「仕様」です。 「仕様」は推奨キャリブレーション間隔内において、記載された動作条件下で保証される計測器の性能を示しま す。 「標準仕様」は推奨キャリブレーション間隔内において、記載された動作条件下で大多数の計測器が満たす仕様を 示します。計測器の性能は保証されません。 「特性または補足仕様」は、設計または開発中に特定された計測器の基本的機能および属性を示し、検証または調 整中に評価されたものではありません。これには、前述の定義に含まれていない、計測器の標準的な使用に関す る情報が記載されています。 特に注釈がない限り、これらの仕様は以下の条件に対して有効です。 • • • 周囲温度 : 23 ℃ ± 5 ℃ ウォームアップ時間 : 30 分 niDCPower オートゼロプロパティ / 属性を ON に設定 NI PXI-4132 のドキュメントにアクセスするには、スタート→すべてのプログラム→ National Instruments → NI-DCPower →ドキュメントを選択します。 デバイス機能 表 1 および図 1 は、NI PXI-4132 の電圧および電流ソー スとシンクレンジを示します。 I+ 100mA 表 1 NI PXI-4132 の電流ソースおよびシンクレンジ DC 電圧レンジ(CAT I) ±10 V ±100 V DC 電流ソース / ╙ੑ⽎㒢 䉲䊮䉪 シンクレンジ 10 μA 100 μA 1 mA 10 mA 100 mA 注意 電圧レベルが 30 VRMS、42.4 V peak、 または 60 VDC を超える場合、感電する危険 性があります。感電の危険性がある場合は特に 注意してください。NI PXI-4132 をバックシェ ルアセンブリまたはケーブルに接続 / 接続解 除する前に、必ずテストシステムの電源が落 ちていることを確認してください。 50mA ╙৻⽎㒢 䉸䊷䉴 20mA V– –100V V+ 20V 40V –20mA –40V –20V –50mA ╙ਃ⽎㒢 䉸䊷䉴 100V ╙྾⽎㒢 䉲䊮䉪 –100mA I– 図 1 NI PXI-4132 の象限図、特性 SMU 仕様1 電圧プログラミング確度 / 分解能 確度 ±(出力値の % + オフセット) レンジ 1年 23 ℃ ± 5 ℃ 分解能 ピーク間ノイズ、標準 (0.1 Hz ~ 10 Hz) ±10 V 50 μV 0.025% + 3.0 mV 70 μV ±100 V 500 μV 0.025% + 10 mV 300 μV 電流プログラミング確度 / 分解能2 確度 ±(出力値の % + オフセット) レンジ 1年 23 ℃ ± 5 ℃ 分解能 ピーク間ノイズ、標準 (0.1 Hz ~ 10 Hz) 10 μA 500 pA 0.034% + 2.0 nA 90 pA 100 μA 5 nA 0.034% + 20 nA 900 pA 1 mA 50 nA 0.034% + 0.2 μA 9 nA 10 mA 500 nA 0.034% + 2.0 μA 90 nA 100 mA 5 μA 0.034% + 20 μA 900 nA 電圧測定確度 / 分解能 確度 ±(読み取り値の % + オフセット) レンジ 1年 23 ℃ ± 5 ℃ 分解能 ±10 V 10 μV 0.02% + 2.0 mV ±100 V 100 μV 0.02% + 5.0 mV 電流測定確度 / 分解能 確度 ±(読み取り値の % + オフセット) レンジ 1年 23 ℃ ± 5 ℃ 分解能 10 μA 10 pA 0.028% + 1.0 nA 100 μA 100 pA 0.028% + 10 nA 1 mA 1 nA 0.028% + 0.1 μA 10 mA 10 nA 0.028% + 1.0 μA 100 mA 100 nA 0.020% + 10 μA 1 温度係数(Tempco)は、1 ℃あたりの確度仕様の 15%。 2 プログラム可能な最小電流制限 / レベルはレンジの 2%。 NI PXI-4132 仕様 2 ni.com/jp 図 2 ~ 4 は、異なる負荷に対する NI PXI-4132 のス テップ応答を示します。 追加仕様 整定時間、標準 ................................<300 μs 。最終値の 0.1% に整定(電流レ ンジの 50% 負荷で 1 V ステップ) 1.2 1 過渡応答、標準 ................................ 電流レンジで 10% ~ 90% の負荷電流変化 後、100 μs 内で電圧レ ンジの <0.1% に回復 ᝄ䋨V䋩 0.8 広帯域ソースノイズ、標準...........8 mVp-p (負荷抵抗) <1 mV RMS(20 Hz ~ 20 MHz 帯域幅) 0.6 0.4 0.2 <50 pF 3.3 µF 10 µF 0 リモートセンス ................................ 電圧確度仕様に 0.5% の HI リード降下を追 加 –0.2 0 0.5 1.0 1.5 2.0 ᤨ㑆䋨䊚䊥⑽䋩 最大リード降下 ...........................1 リードにつき最大 1 V の降下 図 2 100 mA レンジステップ応答、標準 負荷変動 1.2 電圧 .................................................0.5 mV(ローカルセン ス使用時の出力負荷の 1 mA あたり) 1 0.8 ᝄ䋨V䋩 電流 ................................................. レンジの 0.01% (出力変更の 1 ボルト あたり) ガードオフセット電圧、標準 .....<4 mV (電流 ≤ 10 mA) 0.6 0.4 0.2 <50 pF 330 nF 1 µF 0 絶縁電圧(連続)、特性 –0.2 チャンネル / アース間 ..............150 VDC、CAT I、 5 秒間の耐電圧試験で 確認済み 0 0.1 0.2 0.3 0.4 0.5 ᤨ㑆䋨䊚䊥⑽䋩 図 3 10 mA レンジステップ応答、標準 注意 電源に接続しないでください。CAT II、 III、または IV で、信号を接続したり測定用に 使用したりしないでください。 1.2 1 ᝄ䋨V䋩 0.8 0.6 0.4 10 µA 0.2 100 µA 1 mA <50 pF <50 pF <50 pF 330 pF 3.3 nF 33 nF 10 nF 100 nF 1 nF 0 –0.2 0 0.1 0.2 0.3 0.4 0.5 ᤨ㑆䋨䊚䊥⑽䋩 図 4 1 mA、100 μA、および 10 μA レンジステップ 応答、標準 © National Instruments Corporation 3 NI PXI-4132 仕様 トリガ 入力トリガ 600 㔚 㔚ᵹ 10 60 1 6 0.1 0.01 タイプ ............................................ 開始、ソース、シーケ ンスアドバンス、測定 入力元 ............................................ PXI トリガライン 0 ~ 73 䊏䊷䉪㑆 䊧䊮䉳䈱ppm RMS 䊧䊮䉳䈱ppm 100 極性 ........................................... 構成可能 最小パルス幅 .......................... 100 ns 出力先4 .......................................... PXI トリガライン 0 ~ 73 0.6 0.1 1 極性 ........................................... アクティブ HIGH (構成不可) 10 NPLC パルス幅 ................................... 150 ns 図 5 測定ノイズ対アパーチャ遅延、標準 出力トリガ(イベント) タイプ ............................................ ソース完了、 シーケンス反復完了、 シーケンスエンジン完 了、測定完了 補足仕様 測定速度1 表 2 60 Hz における 1 秒あたりの最大動作レート (括弧内は 50 Hz でのレート) ADC アパー チャ遅延 測定から ホストへ 出力先 ............................................ PXI トリガライン 0 ~ 73 極性 ........................................... 構成可能 ソース測定から ホスト 2 へ 1/64 PLC 3490 (2900) 1900 (1700) 1/8 PLC 470 (390) 425 (360) 1 PLC 59.9 (49.9) 59.0 (49.3) パルス幅 ................................... 150 ns ~ 1.6 μs に構成可能 最大ソースアップデート レート2................................................ 4,200 アップデート /s トリガ入力から ソースまでの遅延 ........................... 500 ns 1 負荷に依存する整定時間は含まれません。niDCPower オートゼロプロパティ / 属性は OFF に設定されていま す。 2 ソース測定からホスト、および最大ソースアップデートレートは、ソース遅延が 200 μs に設定された状態で実 行します。簡単な負荷抵抗で、要求されたレベルの 1% 以内に出力を整定するにはこれで十分です。アプリ ケーションの要求に合わせてソース遅延を調整する場合、最大レートは一定ではありません。 3 パルス幅および論理レベルは PXI 仕様に準拠。 4 入力トリガは再エクスポート可能。 NI PXI-4132 仕様 4 ni.com/jp 図 6 は、シーケンスソースモードで自動測定を使用し た場合の NI-DCPower のプログラミングフローを示し ます。NI PXI-4132 でのプログラミングについては、 『NI DC 電源および SMU ヘルプ』を参照してください。 䉲䊷䉬䊮䉴 䉲䊷䉬䊮䉴 䊦䊷䊒 䉦䉡䊮䊃 㐿ᆎ 䊃䊥䉧 䉲䊷䉬䊮䉴 䉝䊄䊋䊮䉴 䊃䊥䉧 NI PXI-4132 Precision SMU 䉴䊁䉾䊒㩷 䉴䊁䉾䊒 䉴䊁䉾䊒 ᢙ A B OUTPUT STATUS 䉸䊷䉴 䊃䊥䉧 REMOTE SENSE 100 VDC MAX 100 mA MAX 䉸䊷䉴 䉸䊷䉴 ㆃᑧ ᷹ቯ 䉸䊷䉴 ᷹ቯቢੌ ቢੌ 䉟䊔䊮䊃 䉟䊔䊮䊃 C 䉲䊷䉬䊮䉴 䉲䊷䉬䊮䉴 ᓳቢੌ 䉣䊮䉳䊮 䉟䊔䊮䊃 ቢੌ 䉟䊔䊮䊃 LO SENSE LO D E GUARD F 図 6 NI-DCPower プログラミングフロー HI G H 追加情報 GUARD I 推奨キャリブレーション 間隔 ......................................................1 年 J SENSE HI 物理特性 OUTPUT CAT I 外形寸法 .............................................3U、1 スロット、 PXI/cPCI モジュール、 2.0 cm × 13.0 cm × K 21.6 cm (0.8 in. × 5.1 in. × 8.5 in.) 重量 ......................................................295 g (10.4 oz) フロントパネルコネクタ ..............COMBICON、 5.08 mm(8 ピン) メモ フロントパネルコネクタは、12 AWG から 28 AWG までのワイヤゲージに対応しま す。 A ステータスインジケータ F 出力 HIGH 端子 B SENSE ステータスインジ G ガード端子 ケータ H ガード端子 C 出力 LOW 端子 I ガード端子 D SENSE LOW 端子 J SENSE HIGH 端子 E ガード端子 K シャーシグランド 図 7 NI PXI-4132 フロントパネル © National Instruments Corporation 5 NI PXI-4132 仕様 所要電力 アクセサリ PXI 所要電力 ..................................... 10 W(5 V 時)、 1 W(3.3 V 時)、 2 W(12 V 時) 以下のアクセサリの詳細については、ni.com/jp を参 照してください。 注意 安全対策として、NI PXI-4132 は出荷 キットに含まれている適切な定格ケーブルお よびバックシェルキットを使用して操作して ください。付属のバックシェルを取り付けず にデバイスを操作すると、ユーザが高電圧に さらされる恐れがあります。 環境 最大使用高度 .................................... 2,000 m (周囲温度 25 ℃時) 汚染度 ................................................. 2 室内使用のみ。 動作環境 周囲温度範囲 .................................... 0 ~ 55 ℃ (IEC 60068-2-1/ IEC 60068-2-2 に準拠 して試験済み) 100V CAT I 相対湿度範囲 .................................... 10 ~ 70%、結露なき こと。40 ℃以上では 1 ℃あたり 1.3% の ディレーティング (IEC 60068-2-56 に準 拠して試験済み) 図 8 組み立てられた NI PXI-4132 の高電圧バックシェル 保管環境 表 3 NI PXI-4132 対応の NI アクセサリ 周囲温度範囲 ................................ –40 ℃~ 70 ℃ (IEC 60068-2-1/ IEC 60068-2-2 に準拠 して試験済み) 相対湿度範囲 .................................... 5 ~ 95%、 結露なきこと (IEC 60068-2-56 に準 拠して試験済み) 耐衝撃 / 振動 動作時衝撃 .............................................. 最大 30 g(半正弦波) 、 11 ms パルス (IEC 60068-2-27 に準拠 して試験済み。 MIL-PRF-28800F に準拠 してテストプロファイ ルを確立。 ) ランダム振動 動作時 ............................................ 5 ~ 500 Hz、0.3 grms 非動作時 ........................................5 ~ 500 Hz、2.4 grms (IEC 60068-2-64 に準 拠して試験済み。非動 作時テストプロファイ ルは MIL-PRF-28800F、 Class 3 の要件を上回 る。) NI PXI-4132 仕様 アクセサリ 製品番号(P/N) NI PXI-4132 用追加コネ クタおよびバックシェ ルキット 781175-01 注意 地域の安全コードと基準、および製造 元によって提供された規格に従ってメイトコ ネクタを取り付ける必要があります。他社製 コネクタの安全適合指令、また該当する基準 (北米の UL および CSA、ヨーロッパの IEC お よび VDE を含む)に従った使用方法を確認し てください。 表 4 NI PXI-4132 対応の他社製アクセサリ アクセサリ 説明 製造元 P/N COMBICON、 I/O 用メイ Phoenix Contact 1825378 5.08 mm (8 ピン) 6 トコネクタ ni.com/jp 認可および準拠 環境管理 安全性 ナショナルインスツルメンツは、環境に優しい製品の 設計および製造に努めています。NI は、製品から特定 の有害物質を除外することが、環境のみならず NI のお 客様にとって有益であると考えています。 この製品は、計測、制御、実験に使用される電気装置 に関する以下の規格および安全性の必要条件を満たし ます。 • • IEC 61010-1、EN 61010-1 UL 61010-1、CSA 61010-1 環境の詳細な情報については、ni.com/environment (英語)の NI and the Environment を参照してくださ い。このページには、ナショナルインスツルメンツが準 拠する環境規制および指令、およびこのドキュメント メモ UL およびその他の安全保証について に含まれていないその他の環境に関する情報が記載さ は、製品ラベルまたは「オンライン製品認証」 れています。 セクションを参照してください。 廃電気電子機器(WEEE) 注意 製品ドキュメントに記載されている手 順以外の方法で使用した場合、この装置に装 備されている保護機能が正常に動作しない場 合があります。 欧州のお客様へ 製品寿命を過ぎたすべての 製品は、必ず WEEE リサイクルセンターへ送 付してください。WEEE リサイクルセンター、 ナショナルインスツルメンツの WEEE への取 り組み、および廃電気電子機器に関する WEEE 指令 2002/96/EC との準拠について は、ni.com/environment/weee(英語)を 参照してください。 電磁両立性 この製品は、計測、制御、実験に使用される電気装置 に関する以下の EMC 規格の必要条件を満たします。 • EN 61326 (IEC 61326): Class A エミッション、 ⬉ᄤֵᙃѻક∵ᶧࠊㅵ⧚ࡲ⊩ ˄Ё RoHS˅ 基本イミュニティ • EN 55011 (CISPR 11): Group 1、Class A エミッ National Instruments ヺড়Ё⬉ᄤֵ ᙃѻકЁ䰤ࠊՓ⫼ᶤѯ᳝ᆇ⠽䋼ᣛҸ (RoHS)DŽ݇Ѣ National Instruments Ё RoHS ড়㾘ᗻֵᙃˈ 䇋ⱏᔩ ni.com/environment/rohs_chinaDŽ Ёᅶ᠋ ション • AS/NZS CISPR 11: Group 1、Class A エミッショ • • FCC 47 CFR Part 15B: Class A エミッション ICES-001: Class A エミッション ン (For information about China RoHS compliance, go to ni.com/environment/rohs_china.) メモ 製品の EMC 決定に適用する基準に関 しては、「オンライン製品認証」セクションを 参照してください。 メモ EMC に適合させるには、このデバイ スを NI PXI-4132 キットに同梱されるシールド ケーブルおよびフェライトと併用してくださ い。 CE マーク準拠 この製品は、該当する EC 理事会指令による基本的要 件に適合しています。 • • 2006/95/EC、低電圧指令(安全性) 2004/108/EC、電磁両立性指令(EMC) オンライン製品認証 その他の適合規格については、適合宣言(DoC)を参 照してください。この製品の製品認証および適合宣言を 入手するには、ni.com/certification(英語)にア クセスして型番または製品ラインで検索し、保証の欄 の該当するリンクをクリックしてください。 © National Instruments Corporation 7 NI PXI-4132 仕様 National Instruments、NI、ni.com、および LabVIEW は National Instruments Corporation (米国ナショナルインスツルメンツ社)の商標です。National Instruments の商標の詳細については、 ni.com/legal の「Terms of Use」セクションを参照してください。本文書中に記載されたその他の 製品名および企業名は、それぞれの企業の商標または商号です。National Instruments の製品 / 技術を保 護する特許については、ソフトウェアで参照できる特許情報 ( ヘルプ→特許情報 )、メディアに含まれてい る patents.txt ファイル、または「National Instruments Patent Notice」(ni.com/patents)のうち、 該当するリソースから参照してください。 © 2009 National Instruments Corporation. All rights reserved. 375128B-01 2009 年 11 月