Configuration ......................................... 4x1 multiplexer
RF Performance Characteristics
Characteristic impedance (Z0)................ 50 Ω nominal
Insertion loss
≤2.5 GHz......................................... <0.6 dB
≤4 GHz............................................ <0.9 dB
VSWR
≤2.5 GHz......................................... <1.3
≤4 GHz............................................ <1.5
Channel-to-channel isolation
≤2.5 GHz......................................... >60 dB
≤4 GHz............................................ >55 dB
Input Characteristics
All input characteristics are DC, ACrms, or a combination unless otherwise
specified.
Maximum switching voltage.................. 30 V
(channel-to-channel and channel-to-ground)
Maximum switching current .................. 0.33 A
National Instruments™, NI™, and ni.com™ are trademarks of National Instruments Corporation. Product and company
names mentioned herein are trademarks or trade names of their respective companies. For patents covering
National Instruments products, refer to the appropriate location: Help»Patents in your software, the patents.txt file
on your CD, or ni.com/patents.
ni.com
© 2003 National Instruments Corp. All rights reserved.
April 2003
323536A-01
Maximum carry current ..........................0.33 A
Maximum switching power ...................10 W
National Instruments recommends against switching active RF signals. As a relay
actuates, the channel is momentarily unterminated. Some RF sources can be damaged
by reflections if their outputs are not properly terminated. Consult your RF source
documentation for more information.
Note
Maximum RF carry power .....................10 W
DC path resistance
Initial................................................<0.2 Ω
End of life ........................................>1 Ω
Path resistance is a combination of relay contact resistance and trace
resistance. Contact resistance typically remains low for the life of a relay.
At the end of relay life, the contact resistance rises rapidly above 1.0 Ω.
Dynamic Characteristics
Relay operate time (at 20 °C) .................15 ms
Release time (at 20 °C) ...........................15 ms
Expected relay life
Mechanical ......................................5,000,000 cycles
Electrical..........................................100,000 cycles
(maximum load)
Trigger Characteristics
Input trigger
Sources ............................................PXI trigger lines 0–7 and STAR
Minimum pulse width......................70 ns
Output trigger
Destinations .....................................PXI trigger lines 0–7
Pulse width ......................................1 µs
Specifications for the NI PXI-2591
2
ni.com
Physical Characteristics
Relay type .............................................. Electromechanical, non-latching
I/O connectors ........................................ 5 SMA jacks
Contact material ..................................... Gold
Dimensions (W × H × D) ....................... 3 cm × 10 cm × 16 cm
(0.8 in. × 3.9 in. × 6.3 in.)
Weight .................................................... 225 g
(8 oz)
Environment
Operating temperature............................ 0 °C to 50 °C
Storage temperature ............................... –20 °C to 70 °C
Relative humidity ................................... 5% to 85% noncondensing
Pollution Degree .................................... 2
CH0
CH1
COM
CH2
CH3
Figure 1. NI PXI-2591 Power-On State
Compliance and Certifications
Safety
This product is designed to meet the requirements of the following
standards of safety for electrical equipment for measurement, control
and laboratory use:
Note
•
IEC 61010-1, EN 61010-1
•
UL 3111-1, UL 61010B-1
•
CAN/CSA C22.2 No. 1010.1
For UL and other safety certifications, refer to the product label or to ni.com.
© National Instruments Corporation
3
Specifications for the NI PXI-2591
Electromagnetic Compatibility
Emissions................................................EN 55011 Class A at 10 m
FCC Part 15A above 1 GHz
Immunity ................................................EN 61326:1997 + A2:2001,
Table 1
EMC/EMI ...............................................CE, C-Tick and FCC Part 15
(Class A) Compliant
Note
For EMC compliance, you must operate this device with shielded cabling.
CE Compliance
This product meets the essential requirements of applicable European
Directives, as amended for CE Marking, as follows:
Low-Voltage Directive (safety)..............73/23/EEC
Electromagnetic Compatibility
Directive (EMC) .....................................89/336/EEC
Refer to the Declaration of Conformity (DoC) for this product for any additional
regulatory compliance information. To obtain the DoC for this product, click Declarations
of Conformity Information at ni.com/hardref.nsf/.
Note
*323536A-01*
323536A-01
Apr03