MKW10M SERIES - MINMAX Technology

®
MKW10M SERIES
DC/DC CONVERTER 10W, Reinforced Insulation, Medical Safety
FEATURES
►4200VAC reinforced Insulation
►Insulation rated for 300VAC Working Voltage
►Medical Safety to UL/CSA/EN/IEC 60601-1 3rd Edition
►2 MOOP rated
►Wide 2:1 Input Voltage Range
►Fully regulated Output Voltage
►Low Leakage Current
►Operating Temp. Range –40°C to +75 °C
►Input Filter meets EN 55022,class A and FCC, level A
►Overload Protection
►2”x 1” Plastic Package
►3 Years Product Warranty
PRODUCT OVERVIEW
The MINMAX MKW10M series is a new range of high performance DC/DC converter modules with a reinforced insulation system .The I/O- isolation voltage
is specified for 4200VACrms.The product comes in a compact 2”x1” industry standard package. All 15 models features wide 2:1 input voltage range and
fully regulated output voltage.The MKW10M DC/DC converters offer an economical solution for demanding applications in industrial and medical
instrumentation requesting a certified supplementary or reinforced insulation system to comply with industrial or latest medical safety standards.
Model Selection Guide
Model
Number
MKW10-12S05M
MKW10-12S051M
MKW10-12S12M
MKW10-12D12M
MKW10-12D15M
MKW10-24S05M
MKW10-24S051M
MKW10-24S12M
MKW10-24D12M
MKW10-24D15M
MKW10-48S05M
MKW10-48S051M
MKW10-48S12M
MKW10-48D12M
MKW10-48D15M
Input
Voltage
(Range)
Output
Voltage
VDC
VDC
5
5.1
12
±12
±15
5
5.1
12
±12
±15
5
5.1
12
±12
±15
12
(9 ~ 18)
24
(18 ~ 36)
48
(36 ~ 75)
Output Current
Input Current
Max.
@Max. Load
@No Load
Reflected
Ripple
Current
mA
1600
1600
835
±417
±333
2000
2000
835
±417
±333
2000
2000
835
±417
±333
mA(typ.)
907
907
1044
1042
1028
559
559
516
516
508
280
280
258
258
254
mA (typ.)
mA(typ.)
Max. capacitive
Load
μ F
@Max. Load
1000
30
100
470
220#
1000
20
50
470
220#
1000
10
25
Efficiency
(typ.)
470
220#
%
74
75
80
80
81
75
76
81
81
82
75
76
81
81
82
# For each output
Input Specifications
Parameter
Model
12V Input Models
24V Input Models
48V Input Models
12V Input Models
24V Input Models
48V Input Models
12V Input Models
24V Input Models
48V Input Models
Input Surge Voltage (1 sec. max.)
Start-Up Threshold Voltage
Under Voltage Shutdown
Short Circuit Input Power
Internal Power Dissipation
Conducted EMI
All Models
E-mail:[email protected]
2014/09/23 REV:3
Tel:886-6-2923150
Page 1 of 4
Min.
Typ.
Max.
Unit
-0.7
25
---0.7
50
---0.7
100
--7
8
9
13
15
18
VDC
30
33
36
----8.5
----16
----34
----3000
mW
----4000
mW
Compliance to EN 55022,class A and FCC part 15,class A
®
MKW10M SERIES
DC/DC CONVERTER 10W, Reinforced Insulation, Medical Safety
Output Specifications
Parameter
Output Voltage Setting Accuracy
Output Voltage Balance
Line Regulation
Load Regulation
Ripple & Noise
Min.Load
Over Load Protection
Transient Recovery Time
Transient Response Deviation
Temperature Coefficient
Short Circuit Protection
Conditions
Min.
Typ.
Max.
----±1.0
Dual Output, Balanced Loads
--±0.5
±2.0
Vin=Min. to Max.
--±0.3
±0.5
Io=15% to 100%
--±0.5
±1.0
Io=5% to 100%
--±0.6
±1.2
5V & 5.1V Output Models
----100
0-20 MHz Bandwidth
Other Output Models
----150
No minimum Load Requirement
120
150
----300
600
25% Load Step Change
--±3
±5
--±0.02
±0.05
Continuous
Unit
%Vnom.
%
%
%
%
mV P-P
mV P-P
%
μ sec
%
%/℃
Isolation, Safety Standards
Parameter
I/O Isolation Voltage (reinforced)
Leakage Current
Conditions
Min.
Typ.
Max.
Unit
60 Seconds
4200
---
---
VACrms
240VAC, 60Hz
---
---
10
μ A
I/O Isolation Resistance
500 VDC
10
---
---
GΩ
I/O Isolation Capacitance
100KHz, 1V
80
pF
Safety Standards
Approvals
--60
cUL/UL60950-1, CSA C22.2 No. 60950-1-03
UL60601-1,CSA C22.2 No.601-1,
IEC/EN 60950-1, IEC/EN 60601-1 3rd Edition, 2 MOOP
IEC60950-1 CB report, cUL/UL 60950-1 certificate
UL60601-1 UL certificate
General Specifications
Parameter
Conditions
Min.
Typ.
Max.
Unit
MIL-HDBK-217F@25℃, Ground Benign
120
1,000,000
150
---
180
---
KHz
Hours
Conditions
Min.
Max.
Unit
Natural Convection
-40
+75
℃
Case Temperature
---
+95
℃
Storage Temperature Range
-50
+125
℃
Humidity (non condensing)
---
95
% rel. H
---
4000
m
260
℃
Switching Frequency
MTBF(calculated)
Environmental Specifications
Parameter
Operating Ambient Temperature Range
(See Power Derating Curve)
Altitude
Cooling
Free-Air convection
Lead Temperature
(1.5mm from case for 10Sec.)
---
E-mail:[email protected]
2014/09/23 REV:3
Tel:886-6-2923150
Page 2 of 4
®
MKW10M SERIES
DC/DC CONVERTER 10W, Reinforced Insulation, Medical Safety
Power Derating Curve
100
Natural
Convection
20LFM
Output Power (%)
80
100LFM
60
200LFM
400LFM
40
20
0
~
-40
0
20
40
60
80
100 110
Ambient Temperature C
Notes
1
2
3
4
5
6
Specifications typical at Ta=+25℃, resistive load, nominal input voltage and rated output current unless otherwise noted.
Transient recovery time is measured to within 1% error band for a step change in output load of 75% to 100%.
We recommend to protect the converter by a slow blow fuse in the input supply line.
Other input and output voltage may be available, please contact factory.
That “natural convection” is about 20LFM but is not equal to still air (0 LFM).
Specifications are subject to change without notice.
Package Specifications
Mechanical Dimensions
Pin Connections
5.08
[0.20]
5.08
[0.20]
Bottom View
5
4
Pin
Single Output
Dual Output
1
+Vin
+Vin
2
-Vin
-Vin
3
+Vout
+Vout
4
No Pin
Common
5
-Vout
-Vout
50.8 [2.00]
1.00 [ 0.04]
1
45.72 [1.80]
2
12.0
[0.47]
►
All dimensions in mm (inches)
3
►
Tolerance: X.X±0.25 (X.XX±0.01)
10.16
[0.40]
10.16
[0.40]
2.54
[0.10]
25.4 [1.00]
Physical Characteristics
Case Size
:
50.8x25.4x12.0mm (2.0x1.0x0.47 inches)
Case Material
:
Non-Conductive Black Plastic (flammability to UL 94V-0 rated)
Pin Material
:
Copper Alloy with Gold Plate Over Nickel Subplate
Weight
:
24.5g
E-mail:[email protected]
2014/09/23 REV:3
Tel:886-6-2923150
Page 3 of 4
X.XX±0.13 ( X.XXX±0.005)
►
Pin diameter  1.0 ±0.05 (0.04±0.002)
®
MKW10M SERIES
DC/DC CONVERTER 10W, Reinforced Insulation, Medical Safety
Test Setup
Input Reflected-Ripple Current Test Setup
Input reflected-ripple current is measured with a inductor Lin (4.7μ H) and Cin (220μ F, ESR < 1.0Ω at 100 KHz) to simulate source impedance. Capacitor Cin,
offsets possible battery impedance. Current ripple is measured at the input terminals of the module, measurement bandwidth is 0-500 KHz.
To Oscilloscope
+
+
Battery
+Vin
Lin
DC / DC
Converter
Current
Probe
Cin
+Out
-Vin
Load
-Out
Peak-to-Peak Output Noise Measurement Test
Use a Cout 0.47μ F ceramic capacitor. Scope measurement should be made by using a BNC socket, measurement bandwidth is 0-20 MHz. Position the load
between 50 mm and 75 mm from the DC/DC Converter.
+Vin
+Out
Copper Strip
Single Output
DC / DC
Converter
-Vin
Cout
-Out
Scope
+Vin
+Out
Copper Strip
Dual Output
DC / DC
Converter
Com.
Copper Strip
-Vin
-Out
Resistive
Load
Copper Strip
Scope
Cout
Cout
Resistive
Load
Scope
Copper Strip
Technical Notes
Input Source Impedance
The power module should be connected to a low ac-impedance input source. Highly inductive source impedances can affect the stability of the power module. In
applications where power is supplied over long lines and output loading is high, it may be necessary to use a capacitor on the input to insure startup. By using a
good quality low Equivalent Series Resistance (ESR < 1.0Ω at 100 kHz) capacitor of a 10μ F for the 12V input devices and a 4.7μ F for the 24V input devices and a
2.2μ F for the 48V devices, capacitor mounted close to the power module helps ensure stability of the unit.
+
DC Power
Source
+Vin
+
+Out
DC / DC
Converter
Load
Cin
-
-Vin
-Out
Output Ripple Reduction
A good quality low ESR capacitor placed as close as practicable across the load will give the best ripple and noise performance. To reduce output ripple, it is
recommended to use 3.3μ F capacitors at the output.
+
+Vin
DC Power
Source
-
+Out
Single Output
DC / DC
Converter
-Vin
+
Cout
+Vin
DC Power
Source
Load
-Out
-
+Out
Dual Output
DC / DC Com.
Converter
-Vin
-Out
Cout
Load
Load
Cout
Maximum Capacitive Load
The MKW10M series has limitation of maximum connected capacitance on the output. The power module may operate in current limiting mode during start-up,
affecting the ramp-up and the startup time. Connect capacitors at the point of load for best performance. The maximum capacitance can be found in the data
sheet.
Thermal Considerations
Many conditions affect the thermal performance of the power module, such as orientation, airflow over the module and board spacing. To avoid exceeding the
maximum temperature rating of the components inside the power module, the case temperature must be kept below 95℃. The derating curves are determined
from measurements obtained in a test setup.
Position of air velocity
probe and thermocouple
15mm / 0.6in
50mm / 2in
Air Flow
DUT
18, Sin Sin Road, An-Ping Industrial District, Tainan 702, Taiwan
Tel: 886-6-2923150 Fax: 886-6-2923149 E-mail: [email protected]
Minmax Technology Co., Ltd.
2014/09/23 REV:3
Page 4 of 4