® MIHW2000 SERIES DC/DC CONVERTER 3W, Reinforced Insulation, Medical Safety FEATURES ►Reinforced Insulation rated for 300VAC Working Voltage ►I/O-isolation Voltage 4000VACrms ►Industrial & Medical Safety Approval ►Ultra-wide Input Voltage Ranges: 9-40VDC, 18-80VDC and 36-160VDC ►Fully regulated Output Voltage ►Low Leakage Current ►Operating Temp. Range –40°C to +85 °C ►Input Filter meets EN 55022, class A ►Short Circuit Protection ►3 Years Product Warranty PRODUCT OVERVIEW The MINMAX MIHW2000 series is a range of high performance DC/DC converter modules with a reinforced insulation system .The I/O- isolation voltage is specified for 4000VACrms.The product comes in a small DIP-24 package. All models features an ultra-wide input voltage range including 36-160VDC for railroad applications. Full SMD design with exclusive use of ceramic capacitors guarantees a high reliability with calculated MTBF of >1 Mio. hours. These high isolation DC/DC converters are the perfect solution solution for many demanding applications in industrial and railroad systems, in medical instrumentation, everywhere where a certified supplementary or reinforced insulation system is required to comply with specific industrial or medical safety standards. Model Selection Guide Model Input Output Number Voltage Voltage (Range) VDC MIHW2022 MIHW2023 MIHW2026 MIHW2027 MIHW2032 MIHW2033 MIHW2036 MIHW2037 MIHW2042 MIHW2043 MIHW2046 MIHW2047 24 (9 ~ 40) 48 (18 ~ 80) 110 (36 ~ 160) VDC 5 12 ±12 ±15 5 12 ±12 ±15 5 12 ±12 ±15 Output Current Input Current Reflected Max. capacitive Ripple Load Max. Min. @Max. Load @No Load Current mA 600 250 ±125 ±100 600 250 ±125 ±100 600 250 ±125 ±100 mA 90 37.5 ±18.8 ±15 90 37.5 ±18.8 ±15 90 37.5 ±18.8 ±15 mA(typ.) 160 151 151 151 80 75 75 75 35 33 33 33 mA(typ.) mA (typ.) 20 15 10 8 5 3 Efficiency (typ.) @Max. Load uF 1000 470 220# 220# 1000 470 220# 220# 1000 470 220# 220# % 78 83 83 83 78 83 83 83 78 83 83 83 # For each output Input Specifications Parameter Model 24V Input Models 48V Input Models 110V Input Models 24V Input Models 48V Input Models 110V Input Models 24V Input Models 48V Input Models 110V Input Models Input Surge Voltage (1 sec. max.) Start-Up Voltage Under Voltage Shutdown Reverse Polarity Input Current Short Circuit Input Power Internal Power Dissipation Conducted EMI E-mail:[email protected] 2011/09/22 REV:4 All Models Tel:886-6-2923150 Page 1 of 4 Min. Typ. Max. Unit -0.7 --50 -0.7 --100 -0.7 --180 8 8.5 9 VDC 13 15 17 26 30 34 ----8.5 ----16 ----32 ----0.3 A ----2000 mW ----2500 mW Compliance to EN 55022,class A and FCC part 15,class A ® MIHW2000 SERIES DC/DC CONVERTER 3W, Reinforced Insulation, Medical Safety Output Specifications Parameter Conditions Min. Typ. Max. --- ±0.5 ±1.0 % Dual Output, Balanced Loads --- ±0.5 ±2.0 % Line Regulation Vin=Min. to Max. --- ±0.3 ±0.5 % Load Regulation Io=25% to 100% --- ±0.5 ±1.0 % 5V Output Models --- 75 100 mV P-P Other Output Models --- 100 150 mV P-P Over Line, Load & Temp. --- --- 180 mV P-P --- --- 15 mV rms --- 150 500 uS --- ±3 ±6 % --- ±0.02 ±0.05 %/℃ 120 150 --- % Output Voltage Accuracy Output Voltage Balance Ripple & Noise (20MHz) Ripple & Noise (20MHz) Ripple & Noise (20MHz) Transient Recovery Time 25% Load Step Change Transient Response Deviation Temperature Coefficient Over Load Protection Foldback Short Circuit Protection Unit Continuous Isolation, Safety Standards Parameter I/O Isolation Voltage (rated) Conditions Min. Typ. Max. Unit 60 Seconds 4000 --- --- VACrms Flash tested for 1 Second 6000 --- --- VPK 240VAC, 60Hz --- --- 2 uA I/O Isolation Resistance 500 VDC 10 --- --- GΩ I/O Isolation Capacitance 100KHz, 1V 13 pF I/O Isolation Test Voltage Leakage Current Safety Standards Approvals --7 cUL/UL60950-1, CSA C22.2 No. 60950-1-03 UL60601-1,CSA C22.2 No.601-1 IEC/EN 60950-1, IEC/EN 60601-1 IEC60950-1 CB report, cUL/UL 60950-1 certificate UL60601-1 UL certificate General Specifications Parameter Conditions Min. Typ. Max. Unit --- 150 --- KHz 1,000,000 --- --- Hours Switching Frequency MTBF(calculated) MIL-HDBK-217F@25℃, Ground Benign Input Fuse 24V Input Models 48V Input Models 110V Input Models 1000mA Slow-Blow Type 500mA Slow-Blow Type 300mA Slow-Blow Type Environmental Specifications Parameter Operating Temperature Range (with Derating) Conditions Min. Max. Unit Ambient -40 +85 ℃ ℃ Case Temperature --- +95 Storage Temperature Range -50 +125 ℃ Humidity (non condensing) --- 95 % rel. H 260 ℃ Cooling Free-Air convection Lead Temperature (1.5mm from case for 10Sec.) --- E-mail:[email protected] 2011/09/22 REV:4 Tel:886-6-2923150 Page 2 of 4 ® MIHW2000 SERIES DC/DC CONVERTER 3W, Reinforced Insulation, Medical Safety Power Derating Curve 100 Natural convection 80 100LFM 200LFM 60 400LFM 40 20 0 ~ -40 0 20 40 60 80 100 110 Ambient Temperature ] Notes 1 Specifications typical at Ta=+25℃, resistive load, nominal input voltage and rated output current unless otherwise noted. 2 Transient recovery time is measured to within 1% error band for a step change in output load of 75% to 100%. 3 Ripple & Noise measurement bandwidth is 0-20 MHz. 4 These power converters require a minimum output loading to maintain specified regulation, operation under no-load conditions will not damage these 5 All DC/DC converters should be externally fused at the front end for protection. 6 Other input and output voltage may be available, please contact factory. 7 Specifications subject to change without notice. modules; however, they may not meet all specifications listed. Package Specifications 3.9 10.5 [0.41] [0.15] Pin Connections 0.2 [0.008] Mechanical Dimensions 0.60 [0.024] 20.32 [0.80] 2.54 [0.10] 31.8 [1.25] +Vin 11 No Pin Common 12 -Vout No Pin 13 +Vout -Vout 15 No Pin +Vout 23 -Vin -Vin 24 -Vin -Vin 5.08 [0.20] 2.0 [0.08] ►All dimensions in mm (inches) ►Tolerance: X.X±0.25 Physical Characteristics Case Size : Case Material : 31.8x20.3x10.5mm (1.25x0.8x0.41 Inches) Non-Conductive Black Plastic (flammability to UL 94V-0 rated) Weight : 16.2g 2011/09/22 REV:4 Tel:886-6-2923150 Page 3 of 4 (X.XX±0.01) X.XX±0.13 ( X.XXX±0.005) ►Pin E-mail:[email protected] Dual Output 13 2.54 [0.10] 15 23 +Vin 12 Bottom View 24 Single Output 1 20.3 [0.80] 11 15.22 [0.60] 1 Pin diameter 0.6 ±0.05 (0.024±0.002) ® MIHW2000 SERIES DC/DC CONVERTER 3W, Reinforced Insulation, Medical Safety Test Configurations Input Reflected-Ripple Current Test Setup Input reflected-ripple current is measured with a inductor Lin (4.7uH) and Cin (220uF, ESR < 1.0Ω at 100 KHz) to simulate source impedance. Capacitor Cin, offsets possible battery impedance. Current ripple is measured at the input terminals of the module, measurement bandwidth is 0-500 KHz. To Oscilloscope + + Battery +Vin Lin DC / DC Converter Current Probe Cin +Out -Vin Load -Out Peak-to-Peak Output Noise Measurement Test Use a Cout 0.47uF ceramic capacitor. Scope measurement should be made by using a BNC socket, measurement bandwidth is 0-20 MHz. Position the load between 50 mm and 75 mm from the DC/DC Converter. +Vin +Out Copper Strip Single Output DC / DC Converter -Vin Cout -Out Scope Resistive Load +Out Dual Output DC / DC Converter Com. -Vin Copper Strip Copper Strip +Vin Scope Cout Cout -Out Resistive Load Copper Strip Scope Copper Strip Design & Feature Considerations Overcurrent Protection To provide protection in a fault (output overload) condition, the unit is equipped with internal current limiting circuitry and can endure current limiting for an unlimited duration. At the point of current-limit inception, the unit shifts from voltage control to current control. The unit operates normally once the output current is brought back into its specified range. Input Source Impedance The power module should be connected to a low ac-impedance input source. Highly inductive source impedances can affect the stability of the power module. In applications where power is supplied over long lines and output loading is high, it may be necessary to use a capacitor on the input to insure startup. By using a good quality low Equivalent Series Resistance (ESR < 1.0Ω at 100 kHz) capacitor of a 4.7uF for the 24V input devices, a 2.2uF for the 48V devices and a 1uF for the 110V devices, capacitor mounted close to the power module helps ensure stability of the unit. + DC Power Source +Vin + +Out DC / DC Converter Load Cin - -Vin -Out Output Ripple Reduction A good quality low ESR capacitor placed as close as practicable across the load will give the best ripple and noise performance. To reduce output ripple, it is recommended to use 3.3uF capacitors at the output. + +Vin - +Out Single Output DC / DC Converter DC Power Source -Vin + Cout +Vin DC Power Source Load -Out - +Out Dual Output DC / DC Com. Converter -Vin -Out Cout Load Load Cout Maximum Capacitive Load The MIHW2000 series has limitation of maximum connected capacitance on the output. The power module may operate in current limiting mode during start-up, affecting the ramp-up and the startup time. Connect capacitors at the point of load for best performance. The maximum capacitance can be found in the data sheet. Thermal Considerations Many conditions affect the thermal performance of the power module, such as orientation, airflow over the module and board spacing. To avoid exceeding the maximum temperature rating of the components inside the power module, the case temperature must be kept below 95℃. The derating curves are determined from measurements obtained in a test setup. Position of air velocity probe and thermocouple 15mm / 0.6in 50mm / 2in Air Flow DUT 18, Sin Sin Road, An-Ping Industrial District, Tainan 702, Taiwan Tel: 886-6-2923150 Fax: 886-6-2923149 E-mail: [email protected] Minmax Technology Co., Ltd. 2011/09/22 REV:4 Page 4 of 4