ZBXYAF - First Sensor

ZBXYAF
Oxygen sensor control circuit board
FEATURES
· Control circuit board for all XYA series
oxygen sensors
· Measuring ranges 0...25 % or 0...100 %
oxygen or custom adjustable range
· Can be calibrated in normal air or any
known O2 concentration
· Automatic or manual calibration
· Linear 0...10 V, 4...20 mA and RS232
output of measured oxygen content
· High accuracy
SPECIFICATIONS
Maximum ratings
Supply voltage
Current consumption
Load limitation (4...20 mA)
Temperature limits
Storage
Operating
E / 11666 / B
24 VDC ±10 %
600 mA @ 24 VDC
100...600 Ω
-10...70 °C
-10...70 °C
1/5
www.first-sensor.com
www.sensortechnics.com
ZBXYAF
Oxygen sensor control circuit board
PERFORMANCE CHARACTERISTICS (VS = 24 V)
Characteristics
Min.
Measuring ranges (oxygen partial pressure)5
Accuracy
Repeatability
Output resolution
0...10 V
4...20 mA
RS232
Reaction time (adaptive output filtering in normal air)
Output inactive start up delay (heater warm up)
Initial warm up time (till stable output)
Typ.
1
5
Max.
Unit
1000
1
0.5
0.01
0.01
0.01
mbar
%FS
1
60
10
V
mA
%
s
min
OUTLINE DRAWING & ELECTRICAL CONNECTION
67
60
1 2 3 4 5
Calibration span pot
Auto cal.
Manual cal.
100 % O2
25 % O2
Sensor connections
67 60
Pin
Connection
1
Sensor heater GND (yellow)4
2
Sensor heater + (yellow)4
3
Sensor sense (blue)
Configuration jumper
links
4
Sensor common (black)
5
Sensor pump (red)
Sensor LED
6
24 VDC ±10 %
7
GND1
Calibration switch
8
4...20 mA output
9
0...10 VDC output
Power LED
10 Calibrate
Power and output
connections
11 Cycle2
12 RS232 Tx3
13 RS232 Rx3
6 7 8 9 10 11 12 13
M4 mounting holes
Heater
adjust
pot
V
Monitoring
device GND
100...
600
External calibration
switch (if required)
Supply
GND
Notes:
1. Due to high current flow in the supply GND, when monitoring the 0...10 V output it is recommended that a separate GND wire
for the measurement system is taken from pin 7. This removes errors due to voltage drops in the power supply connections.
2. A digital 3.3 VDC logic output cycles at the same frequency as the electrochemical pumping action of the oxygen sensing cell
during normal operation, thus providing a real time sensor health check. If the output ceases to cycle the sensor has entered a
start-up or error state. This provides fault proof operation.
3. When connecting the interface board via the RS232 connection ensure Tx goes to Rx of the PC and Rx goes to Tx of the PC.
Baud rate should be set to 9600.
4. Heater coil has no polarity.
5. Prolonged operation below 1 mbar oxygen partial pressure can damage the sensing element.
E / 11666 / B
2/5
www.first-sensor.com
www.sensortechnics.com
ZBXYAF
Oxygen sensor control circuit board
OPERATING AND CALIBRATION NOTES
Heater adjustment:
Check the sensor heater settings: Using a volt meter measure the voltage across pin 1 and pin 2. The heater
voltage should be 4.35 ±0.1 VDC measured as close to the sensor as possible.The heater voltage is set by
adjusting the HEATER ADJUST pot (see OUTLINE DRAWING on page 2).
Automatic calibration:
1. Ensure the ZBXYAF is configured for automatic calibration. See CONFIGURATION.
2. Place the sensor probe in the calibration gas, typically normal air. If automatic calibration should take place
with a gas of a different known O2 concentration see RS232 USER CONFIGURABLE OPTIONS on page 5.
3. Allow the output to stabilise for at least 5 min. (10 min. if powering from cold).
4. Apply GND to the CALIBRATE input (PIN 10) or press the on-board calibration switch for a minimum 12 s.
During this 12 s the CYCLE output (PIN 11) and the green LED will go high/on, blink rapidly, go high/on, go
low/off then return to cycling normally to indicate normal operation has resumed. At this point remove GND
from PIN 10 or release the calibration switch.
5. The output will now track to the correct value for the calibration gas depending on the measurement range
the ZBXYAF is configured for.
6. Calibration is complete and all stored memory values are retained on power loss.
If calibration is required with a different gas of known O2 concentration and access to the RS232 menus with a
PC is not achievable to change the automatic calibration percentage, a manual calibration may be performed.
Manual calibration:
1. Ensure the ZBXYAF is configured for manual calibration. See CONFIGURATION.
2. Place the sensor probe in the calibration gas, typically normal air.
3. Allow the output to stabilise for at least 5 min. (10 min. if powering from cold).
4. Apply GND to the CALIBRATE input (PIN 10) or press the on-board calibration switch for a minimum 5 s or
until the CYCLE output and green LED blink at a steady 1 Hz. Remove GND from PIN 10 or release the
calibration switch. Calibration is now initialised.
5. Adjust the calibrate span pot until the output equals the correct value of the calibration gas concentration
(see table below for examples of calibration values).
6. Re-apply GND to PIN 10 or press the calibration switch for a minimum 5 s. During the 5 s the CYCLE
output/LED will blink rapidly, go high/on, go low/off then return to cycling normally to indicate normal
operation has resumed. At this point remove GND from PIN 10 or release the calibration switch.
7. The output will now track to the correct value for the calibration gas depending on the measurement range
the ZBXYAF is configured for.
8. Calibration is complete and all stored memory values are retained on power loss.
Calibration values
O2 concentration in
normal atmosphere
(1013 mbar pressure and
typical humidity levels)
0.1... 25 % Vol. O2
0.1... 100 % Vol. O2
0.1... 25 % Vol. O2
0.1... 100 % Vol. O2
20.7 %
8.28 VDC
2.07 VDC
17.25 mA
7.34 mA
90 %
-
9.0 V
-
18.4 mA
25 %
10.0 VDC
-
20 mA
-
5%
2.0 VDC
0.5 VDC
7.2 mA
4.8 mA
E / 11666 / B
0...10 V output
4...20 mA output
3/5
www.first-sensor.com
www.sensortechnics.com
ZBXYAF
Oxygen sensor control circuit board
CONFIGURATION
The ZBXYAF boards can be user configured to measuring ranges of 0.1... 25 % Vol. O2 and 0.1...100 % Vol. O2.
The entire measurement range is linear in both cases. Factory default is 0.1...25 % Vol. O2. The ZBXYAF may
be reconfigured at any time by adjusting the position of the header pin jumper links on the interface PCB (see
OUTLINE DRAWING on page 2):
· Power down the ZBXYAF board
· Adjust the position of the jumper links to the desired configuration. Each jumper link must be placed in one
of the two positions, either manual (MAN) or automatic (AUTO) calibration and 100 % or 25 % measuring
range. Thin nosed pliers should be used to remove and replace the jumper links. Ensure the jumper links
are correctly seated before reapplying the power.
WARNING:
Prior to re-configuration the ZBXYAF must be powered down. The jumper links must also be
repositioned correctly and in the correct orientation. Failure to adhere to the above could result in product
damage. Products damaged due to incorrect configuration will not be covered under warranty.
RS232 OPERATION
With the RS232 outputs connected to a PC or any other RS232 compatible device the user has the ability to
access two modes of operation, continuous data streaming and the menu screens. Recommended programs
for communicating via PC serial RS232 are Hyperterminal (Windows default), Teraterminal and PuTTY. A
freeware PuTTY program can be downloaded from http://the.earth.li/~sgtatham/putty/latest/x86/putty.exe
Continuous data streaming
On power up, after the initial 60 s heater delay, the ZBXYAF will automatically begin outputting the measured
O2 concentration and sensor Td as both an averaged and raw value. The averaged values give a stable output
with the amount of averaging user variable whilst the raw unaveraged values allow the user to detect sudden
oxygen changes. The averaged value is the measurement output on both the 4...20 mA and 0...10 VDC outputs.
The sensor Td value is the measure of the partial pressure of oxygen in the measurement gas. The O2
concentration (%) is the Td value scaled by the stored calibration value. To stop or restart the data streaming ‘s’
(lower or upper case) should be sent to the unit. Data streaming automatically ceases during calibration.
Menu screens
If the ZBXYAF receives an enter character from the connected PC or device it automatically enters the menu
password screen and stops outputting O2 % and Td values. After the correct password is entered followed by
the enter character, the menu screens are accessed. The menu screens are primarily for diagnostics and
information although there are user configurable options that may be changed. These are the automatic O2
calibration %, the amount of output filtering (averaging) and the analog output ranges. All three processes are
described below. The menu access password may also be changed by the user.
Changing the menu access password
The password is factory set to ‘default’. This however may be changed to a user specific password.
· Connect the ZBXYAF via the RS232 interface to the PC.
· Press Enter then enter your current security password. Press Enter to access the menu screen.
· In the Configuration menu (menu 2) enter ‘3’ to access the password menu screen.
· Enter the new password then press Enter to save.
· The new password is now stored in memory and is retained on power loss.
Pressing ESC returns the screen to the previous menu.
E / 11666 / B
4/5
www.first-sensor.com
www.sensortechnics.com
ZBXYAF
Oxygen sensor control circuit board
RS232 USER CONFIGURABLE OPTIONS
Note:
When navigating the menu screens pressing ESC returns to the previous screen.
Pressing ESC on the front menu screen returns the output to Continuous Data Streaming.
Changing the automatic calibration value
The system is factory set to automatically calibrate to 20.7 % O2 to allow simple calibration in normal air. This
value takes into account average humidity in the atmosphere. If calibration with a gas of a different known O2
concentration is required the factory set value may be changed via the RS232 interface.
· Connect the ZBXYAF via the RS232 interface to the PC.
· Press Enter then enter your security password. Press Enter to access the menu screen.
· In the configuration menu (menu 2) enter the auto calibration value screen (option 1 - Enter Auto Calib).
· The number entered should be the oxygen concentration (%) of the calibration gas to 2 decimal places.
Press Enter to save.
· The new automatic calibration value is now stored in memory. This value is retained on power loss.
Variable output filtering (Td averaging)
The ZBXYAF is factory default to use adaptive output filtering to give an optimum balance between output
stability and response to oxygen changes. However this balance may be altered by the customer to suit the
needs of the application.
· Connect the ZBXYAF via the RS232 interface to the PC.
· Press Enter then enter your security password. Press Enter to access the menu screen.
· In the Configuration menu (menu 2) enter the Td average screen (option 2 - Enter Td Averaging).
· The number entered should be between 0 and 200. 0 for adaptive filtering (recommended), 1 for a very fast
and dynamic output response but relatively unstable to 200 for an extremely stable output but slow
response to oxygen changes. Press Enter to save.
· The new averaging value is now stored in memory. This value is retained on power loss.
Adjusting a custom measurement range for analog outputs (4...20mA and 0...10VDC)
The ZBXYAF is factory default to output a range of 0.1... 25 % Vol. O2 via its two analogue outputs. This range
can be expanded to 0.1... 100 % Vol. O2 (see CONFIGURATION above). When the unit is reconfigured to output
0.1... 100 % Vol. O2 the user also has the option to fully customise the output via RS232. This is extremely
useful in applications where the O2 variation is within a narrow band as it allows the analog outputs to be
tailored to this limited range.
· Ensure the ZBXYAF is configured for 0.1... 100 % Vol. O2 operation (see CONFIGURATION above).
· Connect the ZBXYAF via the RS232 interface to the PC.
· Press Enter then enter your security password. Press Enter to access the menu screen.
· In the configuration menu (menu 2) enter the maximum range screen (option 3 - Enter O2 Max Range).
· The number entered should be between 1.00 and 100.00 to represent the maximum output range.
· Press Enter to save, then ESC to return to the configuration menu.
· Enter the minimum range screen (option 4 - Enter O2 Min Range).
· The number entered should be between 0.00 and 99.00 to represent the minimum output range.
· Press enter to save. The new ranges are now stored in memory and are retained on power loss.
The custom range adjustment does not apply to the RS232 output and is overruled if the unit is reconfigured for
0.1... 25 % Vol. O2 operation.
ERROR CONDITIONS
If the oxygen sensor is not connected up correctly or is damaged the ZBXYAF will highlight this by blinking the
CYCLE output (Pin 11) and green LED in a 3 short blinks 1 long blink pattern. An error code is also displayed
on the RS232 output. If an error condition occurs the unit should be powered down and all wiring checked
before reapplying the power.
First Sensor reserves the right to make changes to any products herein. First Sensor does not assume any liability arising out of the application
or use of any product or circuit described herein, neither does it convey any license under its patent rights nor the rights of others.
E / 11666 / B