SENSORTECHNICS ZBXYA025M0

ZBXYA
Oxygen sensor interface boards
FEATURES
· Control circuit boards for all
XYA series oxygen sensors
· Oxygen partial pressure range
1...250 mbar or 1...1000 mbar
(0...25 % or 0...100% oxygen)
· Functional testing and calibration in
ambient air
· Linear 0...10 V or 4...20 mA output of
measured oxygen content
· High accuracy
· Low power consumption
SPECIFICATIONS
Maximum ratings
Supply voltage
Heater supply voltage
Stand by
Power consumption (interface)
ZBXYA...M...
ZBXYA...D...
Temperature limits
Storage
Operating
May 2008 / 638
24 VDC ±20 %
4.35 ±0.1 VDC (1.85 A)
2.0 VDC
250 mW
500 mW
-10...50 °C
-10...60 °C
1/5
www.sensortechnics.com
ZBXYA
Oxygen sensor interface boards
PERFORMANCE CHARACTERISTICS
Characteristics
Min.
Typ.
Max.
Measuring ranges
ZBXYA025...
1
250
(oxygen partial pressure)
ZBXYA100...
1
1000
Unit
mbar
Accuracy
2
%FS
Resolution
0.04
V
OUTLINE DRAWING & ELECTRICAL CONNECTION
ZBXYA...D...
Dimensions:
80 x 100 mm,
Ø 6.5 mm mounting holes
May 2008 / 638
Connector:
Screw connector block
2/5
www.sensortechnics.com
ZBXYA
Oxygen sensor interface boards
OUTLINE DRAWING & ELECTRICAL CONNECTION (cont.)
Operating and Calibration Notes for ZBXYA...D
Heater adjustment:
1. Check the sensor heater settings: Using a volt meter measure the voltage across H+ and H-. The heater voltage should be
4.35 ±0.1 VDC. If using long cables (>1m) measure this voltage as close to the sensor as possible.
2. If the heater voltage needs to be altered then follow the next steps.
3. With the power off, remove the daughter board, and expose the heater voltage setting pot.
4. Apply power to the board.
5. Using a volt meter measure the heater voltage across H+ and H- and adjust the pot until the voltage is 4.35 ±0.1 VDC.
6. Switch the power off and re-assemble the daughter board.
Auto-calibration of the sensor and board (only available for ZBXYA025D...):
Note: Auto-calibration assumes that the sensor is in a normal atmosphere (1013 mbar pressure, 20.95 % O2 and typical
humidity levels giving an O2 concentration of 20.7 %. The board adjusts its output to 20.7 %/25 % of full-scale.
1. Ensure the sensor is in ambient air (typically 20.7 % O2). Remove or purge application environment.
2. Switch the power on and leave for 10 Minutes and allow to stabilise.
3. The green LED should be blinking, this indicates normal operation.
4. Voltage output devices: Measure the voltage at the K1 output and the 0 V (measuring circuit) terminal.
The reading should be approx. 8.28 V.
Current output devices: Measure the current at K1 output terminal through an external load resistor (typ. 200 Ω).
The reading should be approx. 17.25 mA.
5. If the reading is not correct then press and hold the calibration button for a minimum of 10 s until the reading is correct.
This may also be achieved using an external calibration switch (Normally Open) placed across GND (0 V Supply)
and K2 (see outline drawing).
6. Once released the green LED should start blinking again to indicate normal operation.
7. The system is now calibrated. (Please note that system calibration is lost on power loss and steps 1-8 must be repeated).
8. Note that the auto-calibration will only work if the output is within 20 % of the correct target value. If this is not the case,
the PCB will not attempt to calibrate and will indicate this by setting the green LED constantly on (until removal of power). If
this occurs it may be necessary to manually calibrate the system as described below.
Manual calibration of the sensor and board:
Manual calibration may be performed at any known O2 content including normal air.
1. Ensure the sensor is in an atmosphere with known O2 content (see table for calibration values).
2. Switch the power on and leave for 10 Minutes and allow to stabilise.
3. The green LED should be blinking, this indicates normal operation.
4. Voltage output devices: Measure the voltage at the K1 output and the 0 V (measuring circuit) terminal.
Current output devices: Measure the current at K1 output terminal through an external load resistor (typ. 200 Ω).
5. If the reading is not correct then adjust the calibration setting pot until the desired output is produced. Turn the pot clockwise to decrease the output value and counter-clock-wise to increase it. Because it takes some time for the output signal
to stabilise following an adjustment, make small calibration changes and allow to settle for 30 seconds each time. See
table for calibration values.
6. If the board is calibrated manually it will retain its calibration after the power is switched off.
Calibration values
O2 concentration in normal
atmosphere (1013 mbar pressure)
ZBXYA025...0
ZBXYA100...0
ZBXYA025...4
ZBXYA100...4
20.7 %
8.28 VDC
2.07 VDC
17.25 mA
7.34 mA
90 %
-
9.0 V
-
18.4 mA
25 %
10.0 VDC
-
20 mA
-
5%
2.0 VDC
0.5 VDC
7.2 mA
4.8 mA
Board Test:
1. The green LED should be blinking to indicate normal operation.
2. If the Test button is pressed a 20 % drop in the reading indicates normal operation.
May 2008 / 638
3/5
www.sensortechnics.com
ZBXYA
Oxygen sensor interface boards
OUTLINE DRAWING & ELECTRICAL CONNECTION (cont.)
ZBXYA...M...
Dimensions:
63 x 108 mm,
Ø 4 mm mounting holes
Connector:
15 pin male D connector
Operating and Calibration Notes for ZBXYA...M
1. Reset of the ZBXYA...M interface board is possible by putting the power off/on.
2. A proper operation of a connected sensor and electronic circuit can be tested with the test switch or with an external
signal. After switching the test function the output signal should indicate a 20% lower value.
3. LED indication: proper sensor operation is indicated by blinking of LED D3.
4. Calibration: sensor and circuit can be calibrated in normal air. After a loss of power the ZBXYA...M circuit stores the
calibration value.
May 2008 / 638
4/5
www.sensortechnics.com
ZBXYA
Oxygen sensor interface boards
ORDERING INFORMATION
Series
Options
Example:
ZBXYA
ZBXYA
Measuring range
Type
Type
025:
1...250 mbar
(oxygen partial pressure)
M: requires external power supply
for sensor heating element
0:
100:
1...1000 mbar
(oxygen partial pressure)
D:
4: 4...20 mA
025
includes power supply
for sensor heating element
D
0...10 V
0
Note: Custom specific options are widely available!
Please contact your nearest Sensortechnics sales office for further information.
Sensortechnics reserves the right to make changes to any products herein. Sensortechnics does not assume any liability arising out of the
application or use of any product or circuit described herein, neither does it convey any license under its patent rights nor the rights of others.
May 2008 / 638
5/5
www.sensortechnics.com