Medical Safety Agency Information

Condor Application Note 5/00
SAFETY AGENCY INFORMATION - MEDICAL
AN-114
SAFETY AGENCY INFORMATION
Medical Equipment, Including Dental
Equipment
Standard
IEC/EN/CSA 60601-1
UL 60601-1
Earth Leakage
Electric Strength
Creepage Distance
N.C. = 500 µA (6) Input to Ground = 1500 V ac (1) Input to Ground = 4 mm (2)
Input to Output = 4000 V ac (1,3) Input to Output = 8 mm (2,3)
N.C. = 300 µA (6)
IEC ELECTROMAGNETIC COMPATIBILITY (EMC)
DISTURBANCE
Fundamental Definitions and Terms
Description of Environment
Compatibility Levels for Systems
Radiated and Conducted Phenomena
Compatibility Levels in Industrial Plants
Limits for Harmonic Current
Limitation Voltage Fluctuation/Flicker
High Current Voltage Fluctuation/Flicker
General Specification - Overview
Electrostatic Discharge Test
Radiated RFI Immunity
Electrical Fast Transients/Burst
Mains Surges
Conducted RFI
Harmonics and Inter-Harmonics
Mains Frequency Magnetic Field
Pulsed Magnetic Field
Damped Oscillatory Magnetic Field
Supply Voltage Dips and Interruptions
Oscillatory Waves Immunity
GENERAL
IEC1000-1-1
IEC1000-2-1
IEC1000-2-2
IEC1000-2-3
IEC1000-2-4
IEC1000-3-2
IEC1000-3-3
IEC1000-3-5
IEC1000-4-1
IEC1000-4-2
IEC1000-4-3
IEC1000-4-4
IEC1000-4-5
IEC1000-4-6
IEC1000-4-7
IEC1000-4-8
IEC1000-4-9
IEC1000-4-10
IEC1000-4-11
IEC1000-4-12
Condor D.C. Power Supplies, Inc., 2311 Statham Parkway, Oxnard, CA 93033
800-235-5929 • Tel: 805-486-4565 • Fax: 805-487-8911• www.condorpower.com
Condor Application Note 5/00
Information Technology Equipment, Including Electrical Business Equipment
Standard
Touch Current
Electric Strength
Creepage Distance
IEC/EN/UL/CSA Handheld=0.75mA Input to Ground=1500Vac(4) Input to Ground=2.5mm(5)
Movable = 3.5 mA Input to Output = 3000 V ac (3,4) Input to Output = 6.4 mm (3,5)
60950-1
Stationary = 3.5 mA
NOTES
1. For measured voltages up to 250 V rms or dc.
2. For measured voltages up to 250 V rms or 300 V dc.
3. For SELV outputs.
4. For measured voltages up to 354 Vpk or dc.
5. For measured voltages up to 300 V rms or dc and based on Pollution Degree 2, Material Group III.
6. For single fault condition, maximum = 1 mA.
HIPOT TESTING
All Condor power supplies receive a hipot test to the following limits:
Test Location
Input to Ground
Input to Output
Output to Ground
Test Voltage
1800 Vac or
1800 Vac or
500
Vdc
2500
2500
Vdc
Vdc
Note: Hipot tests can be a destructive test; especially when very high voltages are used. A
catastrophic failure of the unit may result if a test voltage greater than 1800 Vac is applied
between input and output circuits. The components providing the isolation from input to
output cannot be tested while installed in the power supply without risk of damage to the
unit. All components providing input to output isolation are tested using a minimum of
5000 V ac.